Abstract

BACKGROUND

Low concentrations of cardiac troponin (cTn) have been recommended for rapid rule-out of acute myocardial infarction (AMI). We examined the Beckman Coulter Access high-sensitivity cardiac troponin I (hs-cTnI) assay to identify a single test threshold that can safely rule out AMI.

METHODS

This analysis used stored samples collected in 2 prospective observational studies. In all, 1871 patients presenting to a tertiary emergency department with symptoms of acute coronary syndrome had blood taken for measurement of cTnI on presentation. The endpoint was type 1 myocardial infarction (T1MI). Sensitivity and negative predictive value (NPV) were calculated for hs-cTnI values below the 99th percentile.

RESULTS

Ninety-eight patients had T1MI (5.2%), and 638 (34.1%) patients had an hs-cTnI <2 ng/L (limit of detection), with sensitivity of 99.0% (95% CI, 94.4%–100%) and NPV of 99.8% (95% CI, 99.1%–100%). No hs-cTnI value above a concentration of 2 ng/L achieved sensitivity of 99%. However, an NPV of 99.5% was achieved at values <6 ng/L. A cutoff <6 ng/L enabled 1475 (78.8%) patients to be ruled out on presentation with sensitivity of 93.9% (95% CI, 87.1%–97.7%).

CONCLUSIONS

A single baseline cTn <2 ng/L measured with the Access hs-cTnI assay performed well for rule-out of AMI. This cutoff concentration identified 99% of patients with AMI and could reduce the number of patients requiring lengthy assessment. A cutoff of <6 ng/L yielded a high NPV but missed more cases of AMI than would be acceptable to clinicians.

This article is published and distributed under the terms of the Oxford University Press, Standard Journals Publication Model (https://dbpia.nl.go.kr/journals/pages/open_access/funder_policies/chorus/standard_publication_model)
You do not currently have access to this article.