Abstract

Objective

The purpose of this study was to critically appraise and summarize the evidence for reliability of muscle strength and muscle power assessment in patients with neuromuscular diseases (NMDs) using isokinetic dynamometry.

Methods

PubMed, CINAHL, and Embase electronic databases were searched from inception to March 8, 2022. Studies designed to evaluate reliability of muscle strength and power measurements using isokinetic dynamometry were included in this review. First, the methodological quality of the studies was assessed according to the Consensus-Based Standards for the Selection of Health Measurement Instruments guidelines. Next, the quality of measurement properties was determined. Finally, the methodological quality and quality of measurement properties of the studies were combined to obtain a best-evidence synthesis.

Results

A best-evidence synthesis of reliability was performed in 11 studies including postpoliomyelitis syndrome (n = 5), hereditary motor and sensory neuropathy (n = 2), motor neuron diseases (n = 1), myotonic dystrophy (n = 1), and groups of pooled NMDs (n = 2). A best-evidence synthesis on measurement error could not be performed. Quality of evidence on reliability ranged from high in postpoliomyelitis syndrome to very low in hereditary motor and sensory neuropathy, motor neuron diseases, and groups of pooled NMDs. The most frequently used outcome measure was peak torque, which was reliable in all populations (intraclass correlation coefficient >0.7).

Conclusion

The quality of evidence for reliability of isokinetic dynamometry was found to vary substantially among different NMDs. High quality of evidence has been obtained only in patients with postpoliomyelitis syndrome. Further research is needed in the majority of known NMDs to determine reliability and validity of isokinetic dynamometry.

Impact

The ability of isokinetic dynamometers to capture clinically relevant changes in muscle strength and muscle power in NMDs remains unclear. Isokinetic dynamometry results in NMDs should be interpreted with caution.

Introduction

Neuromuscular disorders (NMDs) are a heterogeneous group of acquired and genetic diseases that may encompass all levels of dysfunction of the peripheral nervous system—that is, the anterior horn cells, the peripheral nerve, and the neuromuscular junction—as well as muscle itself.1 The large majority of NMDs are characterized by muscle weakness.1–3 Muscle function assessments are generally used to determine disease severity, gain insight into disease trajectory, and assess therapeutic efficacy.4,5 To meet these objectives, reliable and sensitive outcome measures for muscle strength and muscle power are required.6

Various methods are available to assess muscle function. Manual muscle testing and handheld dynamometry (HHD) are most commonly used for quantitative strength measurements due to their low cost and ease of use.7 Both measures provide insight into the maximum strength in a fixed position (ie, isometric strength).8 Manual muscle testing uses a 6-point manual scale (the Medical Research Council scale) to evaluate muscle strength ranging from grade 0 (no contraction noticeable) to grade 5 (complete range of motion against gravity and hold the test position against maximum resistance).9 HHD registers the force one can produce against a portable handheld dynamometer in a fixed position and allows quantification in newtons.10 Despite the broad and longstanding application of manual muscle testing and HHD in clinical practice, both techniques have limitations; manual muscle testing lacks sensitivity to differentiate between moderate weakness to normal strength,6,11,12 whereas HHD reliability is limited in stronger muscle groups.7 In studies in which ceiling effects may come into play, alternative techniques such as isokinetic dynamometry might represent better options.

Isokinetic dynamometry is considered the “gold standard” of muscle strength and muscle power measurements in orthopedic and neurological patients.13–16 Isokinetic dynamometers (IDs) have superior sensitivity and have higher reliability compared with HHD testing in adults who are healthy.13 IDs are designed to quantify both isometric and isokinetic strength.17 Isometric strength measurements are performed with the limb in a fixed position measuring the muscle’s ability to develop static force, measured as torque in newton-meters. Isokinetic strength is defined as the ability of a muscle to develop dynamic force. Power (watts) refers to the ability to produce force over a specified period of time and range.8 Isokinetic muscle strength and muscle power are more reflective of daily life activities than isometric measurements.18,19

To properly assess the added value of isokinetic dynamometry in the evaluation of muscle function in patients with NMDs, the reliability of the method needs to be assessed.20 According to the Consensus-Based Standards for the Selection of Health Measurement Instruments (COSMIN) guidelines, reliability refers to the degree to which measurements are free from error. The domain of reliability is subdivided into the measurement properties of reliability and measurement error.21 The proportion of the total variance of the measurement due to true differences is defined as reliability, expressed in Intraclass Correlation Coefficients (ICCs) or Pearson r.21,22 In addition to the reliability of results, it is important to know which part of the difference between repeated measurements is attributed to systematic and random error, the measurement error.21,23 The variability between repeated measurements can be expressed in different standards, including the standard error of measurement (SEM), limits of agreement, and the smallest detectable change.22,23

Several studies have explored the reliability of strength and power measurements using IDs in patients with NMD, but no systematic review to our knowledge has been performed on the methodological quality of measurement properties and applicability of isokinetic dynamometry. Here, we conducted a literature review to determine the current level of evidence for reliability of isometric and isokinetic muscle strength and muscle power assessment in patients with NMDs using isokinetic dynamometers.

Methods

Data Sources and Searches

We searched relevant articles in the electronic databases of PubMed, CINAHL, and Embase up to March 8, 2022. Reference lists of included articles were additionally searched for relevant studies. The search was not constrained by either timespan or language. Depending on the search engine MESH terms, Emtree words and keywords were implemented in the search string. We used the search filter as described by Terwee et al24 to optimize the yield of relevant articles. The full search strategy is listed in Supplementary Appendix 1.

Study selection.
Figure

Study selection.

Study Selection

One author (D.R.W.) performed the database search. Subsequently, 2 reviewers (D.R.W. and T.R.) independently selected articles that met the inclusion criteria. Articles were eligible for inclusion when the following 4 criteria were met: the aim of the article was to describe measurement properties of strength and/or power measurements using an isokinetic dynamometer; the articles described reliability and/or measurement error; the population tested included individuals with neuromuscular disease(s); and articles were written in either the English or Dutch language. The 3 exclusion criteria were as follows: no full-text article was available, the particular dynamometer used was incompletely specified, and isokinetic dynamometric measurements were used to validate other tests or evaluate interventions.

We screened all articles on title and abstract. Articles potentially eligible for the study were attained in full text and further examined for inclusion. Articles selected by both reviewers were included. Articles selected by only 1 reviewer were discussed until consensus was reached. If needed, a third reviewer (B.B.) was consulted.

Data Extraction and Quality Assessment

Data Extraction

Descriptive data and data on measurement properties were extracted from each study by 1 author (D.R.W.). Three steps were taken after data extraction, based on the COSMIN guidelines.21 First, we assessed the methodological quality of the studies. Second, we determined the quality of measurement properties. Finally, we combined both methodological quality and quality measurement properties of the studies in a best-evidence synthesis.

Quality Assessment

Two reviewers (D.R.W. and T.R.) independently assessed the methodological quality of the included studies following the COSMIN risk of bias checklist.25 This is a standardized tool to assess the methodological quality of studies on measurement properties. The risk of bias checklist contains 10 measurement properties that can be scored on a 4-point scale.25

We rated the methodological quality of the relevant measurement properties (ie, reliability and measurement error) as very good, adequate, doubtful, inadequate, or not applicable. The lowest rating of any standard in the box was then used according to the “worst score counts” principle.21 In case of disagreement regarding the methodological quality, the 2 principal reviewers discussed the paper in detail until consensus was reached. If not, a third reviewer (B.B.) was consulted.

The quality of the measurement properties was rated sufficient, indeterminate, or insufficient based on criteria provided by the COSMIN guidelines (Tab. 1).

Table 1

Quality Criteria for Good Measurement Propertiesa

Measurement PropertyRatingCriteria
ReliabilitySufficientICC or weighted κ ≥ 0.70
IndeterminateICC or weighted κ not reported
InsufficientICC or weighted κ < 0.70
Measurement errorSufficientSDC or LoA < minimal important change
IndeterminateMinimal important change not defined
InsufficientSDC or LoA > minimal important change
Measurement PropertyRatingCriteria
ReliabilitySufficientICC or weighted κ ≥ 0.70
IndeterminateICC or weighted κ not reported
InsufficientICC or weighted κ < 0.70
Measurement errorSufficientSDC or LoA < minimal important change
IndeterminateMinimal important change not defined
InsufficientSDC or LoA > minimal important change
a

LoA = limits of agreement; SDC = smallest detectable change.

Table 1

Quality Criteria for Good Measurement Propertiesa

Measurement PropertyRatingCriteria
ReliabilitySufficientICC or weighted κ ≥ 0.70
IndeterminateICC or weighted κ not reported
InsufficientICC or weighted κ < 0.70
Measurement errorSufficientSDC or LoA < minimal important change
IndeterminateMinimal important change not defined
InsufficientSDC or LoA > minimal important change
Measurement PropertyRatingCriteria
ReliabilitySufficientICC or weighted κ ≥ 0.70
IndeterminateICC or weighted κ not reported
InsufficientICC or weighted κ < 0.70
Measurement errorSufficientSDC or LoA < minimal important change
IndeterminateMinimal important change not defined
InsufficientSDC or LoA > minimal important change
a

LoA = limits of agreement; SDC = smallest detectable change.

Data Synthesis and Analysis

Quality of Evidence

To summarize the quality of evidence, we performed a best-evidence synthesis based on methodological quality and quality of measurement properties. The level of evidence was described per diagnosis, type of measurement (ie, isometric or isokinetic), and outcome measure (ie, peak torque, power, angle at peak, maximum isometric torque). We graded the quality of evidence as high, moderate, low, or very low based on the Grading of Recommendations Assessment, Development and Evaluation approach.21 Starting with the assumption of high-quality evidence, scores were downgraded for the factors risk of bias, inconsistency, imprecision, and indirectness.

Results

Included Studies

We identified a total of 11 articles for this study (the flowchart is shown in the Figure). The kappa agreement between the reviewers in selecting articles after application of the inclusion and exclusion criteria was 0.92.

Study Characteristics

The topics of the included 11 articles were postpoliomyelitis syndrome (5 articles),26–30 hereditary motor and sensory neuropathy (HMSN) (2 articles),31,32 myotonic dystrophy type 1 (1 article),33 and motor neuron disease (1 article)34; 2 studies contained information on a group with more than 1 NMD35,36 (Tab. 2). All studies enrolled adult patients ranging in age from 24 to 77 years. Types of dynamometers used were Biodex (multi-joint) System 3 (Pro),26–28,33,35 Biodex System 4 Pro,33 Cybex 2,29,32 Cybex 6000,36 Kin-Com,30 Kin-Com 2,34 and the Lido Active Multi-Joint 2.31 All studies reported details of testing position and fixation used. Three studies reported the number of patients unable to perform all measurement procedures because of muscle weakness.26,29,31

Table 2

Study Characteristics and Test Proceduresa

StudyDiagnosisNo. of Participants (Men/Women)Age, yDynamometerTime Between Test and RetestMeasurement
IsometricIsokinetic
Mean (SD)RangeMovementOutcomeSpeedMovementOutcome
Andersen31 (1996)HMSN7 (3/4)4624–70Lido Active Multi-Joint 2At least 24 h30, 60, 180°/sAnkle dorsal flexion-plantar flexionPeak torque
Brogårdh et al26 (2015)Postpoliomyelitis syndrome28 (16/12)6829–81Biodex System 3 Pro14 dShoulder abduction, elbow flexionHighest MVC60°/sElbow flexion-extensionPeak torque
Fillyaw et al32 (1989)HMSN15 (10/5)48.3 (9.6)Cybex 2 isokinetic dynamometer1–2 moShoulder flexion- extension, elbow flexion- extension, knee flexion- extension, ankle dorsal flexion-plantarflexionMaximum isometric torque
Flansbjer and Lexell28 (2010)Postpoliomyelitis syndrome30 (19/11)6351–77Biodex Multi-Joint System 3 Pro7 dKnee extensionHighest MVC60°/sKnee flexion-extensionPeak torque
Flansbjer et al27 (2011)Postpoliomyelitis syndrome31b (18/13)6350–77Biodex Multi-Joint System 3 Pro7 dAnkle dorsal flexionHighest MVC30°/sAnkle dorsal flexionPeak torque
Griffin et al34 (1994)Motor neuron diseases11c (NA)NAKin-Com 21 wk30°,120°/sIsokinetic knee flexion-extensionAverage torque
Horemans et al30 (2004)Postpoliomyelitis syndrome65 (23/42)5236–68Kin-Com3 wkKnee extensionHighest MVC
Kilfoil and St Pierre29 (1993)Postpoliomyelitis syndrome8d (4/4)Men: 47.5 (10); women: 56.8 (16)Cybex 21–3 wk1.05,2.09, 3.14,4.18 rads∙s−1Knee flexion-extensionPeak torque,
angle at peak torque,
maximum power, average power
Knak et al33 (2020)Noncongenital myotonic dystrophy78 (39/39)40 (10)Biodex System 3 Pro and System 4 Pro1–2 wkKnee flexion- extension, hip flexion- extension, ankle plantar flexion, and dorsal flexionMVC
Tiffreau et al36 (2003)Multiple NMDs pooled14 (7/7)Men: 35.2 (8.4); women: 38.7 (11.2)Cybex 60007 d10°/sKnee flexion-extension (continuous passive movement mode)Mean work, maximum work
Tiffreau et al35 (2007)Multiple NMDs pooled15 (3/12)40.416–67Biodex 3 isokinetic dynamometerAt least
2 h
10°,30°/sKnee flexion-extension (continuous passive movement mode)Peak torque,
angle at peak torque, work, power
StudyDiagnosisNo. of Participants (Men/Women)Age, yDynamometerTime Between Test and RetestMeasurement
IsometricIsokinetic
Mean (SD)RangeMovementOutcomeSpeedMovementOutcome
Andersen31 (1996)HMSN7 (3/4)4624–70Lido Active Multi-Joint 2At least 24 h30, 60, 180°/sAnkle dorsal flexion-plantar flexionPeak torque
Brogårdh et al26 (2015)Postpoliomyelitis syndrome28 (16/12)6829–81Biodex System 3 Pro14 dShoulder abduction, elbow flexionHighest MVC60°/sElbow flexion-extensionPeak torque
Fillyaw et al32 (1989)HMSN15 (10/5)48.3 (9.6)Cybex 2 isokinetic dynamometer1–2 moShoulder flexion- extension, elbow flexion- extension, knee flexion- extension, ankle dorsal flexion-plantarflexionMaximum isometric torque
Flansbjer and Lexell28 (2010)Postpoliomyelitis syndrome30 (19/11)6351–77Biodex Multi-Joint System 3 Pro7 dKnee extensionHighest MVC60°/sKnee flexion-extensionPeak torque
Flansbjer et al27 (2011)Postpoliomyelitis syndrome31b (18/13)6350–77Biodex Multi-Joint System 3 Pro7 dAnkle dorsal flexionHighest MVC30°/sAnkle dorsal flexionPeak torque
Griffin et al34 (1994)Motor neuron diseases11c (NA)NAKin-Com 21 wk30°,120°/sIsokinetic knee flexion-extensionAverage torque
Horemans et al30 (2004)Postpoliomyelitis syndrome65 (23/42)5236–68Kin-Com3 wkKnee extensionHighest MVC
Kilfoil and St Pierre29 (1993)Postpoliomyelitis syndrome8d (4/4)Men: 47.5 (10); women: 56.8 (16)Cybex 21–3 wk1.05,2.09, 3.14,4.18 rads∙s−1Knee flexion-extensionPeak torque,
angle at peak torque,
maximum power, average power
Knak et al33 (2020)Noncongenital myotonic dystrophy78 (39/39)40 (10)Biodex System 3 Pro and System 4 Pro1–2 wkKnee flexion- extension, hip flexion- extension, ankle plantar flexion, and dorsal flexionMVC
Tiffreau et al36 (2003)Multiple NMDs pooled14 (7/7)Men: 35.2 (8.4); women: 38.7 (11.2)Cybex 60007 d10°/sKnee flexion-extension (continuous passive movement mode)Mean work, maximum work
Tiffreau et al35 (2007)Multiple NMDs pooled15 (3/12)40.416–67Biodex 3 isokinetic dynamometerAt least
2 h
10°,30°/sKnee flexion-extension (continuous passive movement mode)Peak torque,
angle at peak torque, work, power
a

HMSN = hereditary motor and sensory neuropathy; MVC = maximum voluntary contraction (ie, peak torque); NA = not available; NMDs = neuromuscular diseases.

b

Seventeen individuals were unable to perform the measurement with the more affected limb because of muscle weakness.

c

Five patients with motor neuron diseases were included in the reliability assessment.

d

Three patients were not eligible for bilateral evaluation.

Table 2

Study Characteristics and Test Proceduresa

StudyDiagnosisNo. of Participants (Men/Women)Age, yDynamometerTime Between Test and RetestMeasurement
IsometricIsokinetic
Mean (SD)RangeMovementOutcomeSpeedMovementOutcome
Andersen31 (1996)HMSN7 (3/4)4624–70Lido Active Multi-Joint 2At least 24 h30, 60, 180°/sAnkle dorsal flexion-plantar flexionPeak torque
Brogårdh et al26 (2015)Postpoliomyelitis syndrome28 (16/12)6829–81Biodex System 3 Pro14 dShoulder abduction, elbow flexionHighest MVC60°/sElbow flexion-extensionPeak torque
Fillyaw et al32 (1989)HMSN15 (10/5)48.3 (9.6)Cybex 2 isokinetic dynamometer1–2 moShoulder flexion- extension, elbow flexion- extension, knee flexion- extension, ankle dorsal flexion-plantarflexionMaximum isometric torque
Flansbjer and Lexell28 (2010)Postpoliomyelitis syndrome30 (19/11)6351–77Biodex Multi-Joint System 3 Pro7 dKnee extensionHighest MVC60°/sKnee flexion-extensionPeak torque
Flansbjer et al27 (2011)Postpoliomyelitis syndrome31b (18/13)6350–77Biodex Multi-Joint System 3 Pro7 dAnkle dorsal flexionHighest MVC30°/sAnkle dorsal flexionPeak torque
Griffin et al34 (1994)Motor neuron diseases11c (NA)NAKin-Com 21 wk30°,120°/sIsokinetic knee flexion-extensionAverage torque
Horemans et al30 (2004)Postpoliomyelitis syndrome65 (23/42)5236–68Kin-Com3 wkKnee extensionHighest MVC
Kilfoil and St Pierre29 (1993)Postpoliomyelitis syndrome8d (4/4)Men: 47.5 (10); women: 56.8 (16)Cybex 21–3 wk1.05,2.09, 3.14,4.18 rads∙s−1Knee flexion-extensionPeak torque,
angle at peak torque,
maximum power, average power
Knak et al33 (2020)Noncongenital myotonic dystrophy78 (39/39)40 (10)Biodex System 3 Pro and System 4 Pro1–2 wkKnee flexion- extension, hip flexion- extension, ankle plantar flexion, and dorsal flexionMVC
Tiffreau et al36 (2003)Multiple NMDs pooled14 (7/7)Men: 35.2 (8.4); women: 38.7 (11.2)Cybex 60007 d10°/sKnee flexion-extension (continuous passive movement mode)Mean work, maximum work
Tiffreau et al35 (2007)Multiple NMDs pooled15 (3/12)40.416–67Biodex 3 isokinetic dynamometerAt least
2 h
10°,30°/sKnee flexion-extension (continuous passive movement mode)Peak torque,
angle at peak torque, work, power
StudyDiagnosisNo. of Participants (Men/Women)Age, yDynamometerTime Between Test and RetestMeasurement
IsometricIsokinetic
Mean (SD)RangeMovementOutcomeSpeedMovementOutcome
Andersen31 (1996)HMSN7 (3/4)4624–70Lido Active Multi-Joint 2At least 24 h30, 60, 180°/sAnkle dorsal flexion-plantar flexionPeak torque
Brogårdh et al26 (2015)Postpoliomyelitis syndrome28 (16/12)6829–81Biodex System 3 Pro14 dShoulder abduction, elbow flexionHighest MVC60°/sElbow flexion-extensionPeak torque
Fillyaw et al32 (1989)HMSN15 (10/5)48.3 (9.6)Cybex 2 isokinetic dynamometer1–2 moShoulder flexion- extension, elbow flexion- extension, knee flexion- extension, ankle dorsal flexion-plantarflexionMaximum isometric torque
Flansbjer and Lexell28 (2010)Postpoliomyelitis syndrome30 (19/11)6351–77Biodex Multi-Joint System 3 Pro7 dKnee extensionHighest MVC60°/sKnee flexion-extensionPeak torque
Flansbjer et al27 (2011)Postpoliomyelitis syndrome31b (18/13)6350–77Biodex Multi-Joint System 3 Pro7 dAnkle dorsal flexionHighest MVC30°/sAnkle dorsal flexionPeak torque
Griffin et al34 (1994)Motor neuron diseases11c (NA)NAKin-Com 21 wk30°,120°/sIsokinetic knee flexion-extensionAverage torque
Horemans et al30 (2004)Postpoliomyelitis syndrome65 (23/42)5236–68Kin-Com3 wkKnee extensionHighest MVC
Kilfoil and St Pierre29 (1993)Postpoliomyelitis syndrome8d (4/4)Men: 47.5 (10); women: 56.8 (16)Cybex 21–3 wk1.05,2.09, 3.14,4.18 rads∙s−1Knee flexion-extensionPeak torque,
angle at peak torque,
maximum power, average power
Knak et al33 (2020)Noncongenital myotonic dystrophy78 (39/39)40 (10)Biodex System 3 Pro and System 4 Pro1–2 wkKnee flexion- extension, hip flexion- extension, ankle plantar flexion, and dorsal flexionMVC
Tiffreau et al36 (2003)Multiple NMDs pooled14 (7/7)Men: 35.2 (8.4); women: 38.7 (11.2)Cybex 60007 d10°/sKnee flexion-extension (continuous passive movement mode)Mean work, maximum work
Tiffreau et al35 (2007)Multiple NMDs pooled15 (3/12)40.416–67Biodex 3 isokinetic dynamometerAt least
2 h
10°,30°/sKnee flexion-extension (continuous passive movement mode)Peak torque,
angle at peak torque, work, power
a

HMSN = hereditary motor and sensory neuropathy; MVC = maximum voluntary contraction (ie, peak torque); NA = not available; NMDs = neuromuscular diseases.

b

Seventeen individuals were unable to perform the measurement with the more affected limb because of muscle weakness.

c

Five patients with motor neuron diseases were included in the reliability assessment.

d

Three patients were not eligible for bilateral evaluation.

Eight studies evaluated the reliability of isokinetic measurements; 5 of these studies tested knee flexion-extension,28,29,35,36 2 tested ankle plantar flexion and dorsal flexion,27,31 and 1 tested elbow flexion and extension.26 Six studies reported the reliability of isometric measurements: 4 for knee-flexion and/or extension,28,30,32,33 3 for ankle plantar flexion and dorsal flexion,27,32,33 2 for elbow flexion,26,32 and 2 for shoulder abduction.26,32

A total of 6 of 8 studies used peak torque in isokinetic measurements. The highest maximum voluntary contraction (MVC) or maximum isometric torque (ie, peak torque) was used in all 6 studies that contained isometric measurements.

Six studies evaluated measurement error; 4 of these studies reported on knee flexion and/or extension,28,30,32,33 3 reported on ankle dorsal flexion and plantar flexion,27,32,33 and 2 reported on elbow flexion, elbow extension, and shoulder abduction.26,29 A summary of all measurement properties is presented in Supplementary Appendix 2.

Methodological Quality

The methodological quality of the 11 selected articles is presented in Table 3. Eight studies demonstrated adequate quality,26–30,33,35,36 whereas 3 studies showed doubtful methodological quality on reliability.31,32,34 Five studies demonstrated an adequate methodological quality on measurement error,26–28,30,33 whereas 1 study was considered to have a doubtful methodological quality on measurement error.32 The main methodological flaws included the lack of information on patients being stable during the test–retest period, although it could be assumed in most studies. Three studies lacked sufficiently detailed description of the statistics used,29,31,32 which affected both reliability and measurement error scores. Supplementary Appendix 3. shows details of COSMIN scores on each criterion.

Table 3

Methodological Qualitya

DiagnosisStudyReliabilityMeasurement Error
Postpoliomyelitis syndromeBrogårdh et al26 (2015)AdequateAdequate
Flansbjer and Lexell28 (2010)AdequateAdequate
Flansbjer et al27 (2011)AdequateAdequate
Horemans et al30 (2004)AdequateAdequate
Kilfoil and St Pierre29 (1993)AdequateIndeterminate
HMSNAndersen31 (1996)DoubtfulIndeterminate
Fillyaw et al32 (1989)DoubtfulDoubtful
Motor neuron diseasesGriffin et al34 (1994)DoubtfulIndeterminate
Myotonic dystrophy type 1Knak et al33 (2020)AdequateAdequate
Pooled NMDsTiffreau et al36 (2003)AdequateIndeterminate
Tiffreau et al35 (2007)AdequateIndeterminate
DiagnosisStudyReliabilityMeasurement Error
Postpoliomyelitis syndromeBrogårdh et al26 (2015)AdequateAdequate
Flansbjer and Lexell28 (2010)AdequateAdequate
Flansbjer et al27 (2011)AdequateAdequate
Horemans et al30 (2004)AdequateAdequate
Kilfoil and St Pierre29 (1993)AdequateIndeterminate
HMSNAndersen31 (1996)DoubtfulIndeterminate
Fillyaw et al32 (1989)DoubtfulDoubtful
Motor neuron diseasesGriffin et al34 (1994)DoubtfulIndeterminate
Myotonic dystrophy type 1Knak et al33 (2020)AdequateAdequate
Pooled NMDsTiffreau et al36 (2003)AdequateIndeterminate
Tiffreau et al35 (2007)AdequateIndeterminate
a

HMSN = hereditary motor and sensory neuropathy; NMDs = neuromuscular diseases.

Table 3

Methodological Qualitya

DiagnosisStudyReliabilityMeasurement Error
Postpoliomyelitis syndromeBrogårdh et al26 (2015)AdequateAdequate
Flansbjer and Lexell28 (2010)AdequateAdequate
Flansbjer et al27 (2011)AdequateAdequate
Horemans et al30 (2004)AdequateAdequate
Kilfoil and St Pierre29 (1993)AdequateIndeterminate
HMSNAndersen31 (1996)DoubtfulIndeterminate
Fillyaw et al32 (1989)DoubtfulDoubtful
Motor neuron diseasesGriffin et al34 (1994)DoubtfulIndeterminate
Myotonic dystrophy type 1Knak et al33 (2020)AdequateAdequate
Pooled NMDsTiffreau et al36 (2003)AdequateIndeterminate
Tiffreau et al35 (2007)AdequateIndeterminate
DiagnosisStudyReliabilityMeasurement Error
Postpoliomyelitis syndromeBrogårdh et al26 (2015)AdequateAdequate
Flansbjer and Lexell28 (2010)AdequateAdequate
Flansbjer et al27 (2011)AdequateAdequate
Horemans et al30 (2004)AdequateAdequate
Kilfoil and St Pierre29 (1993)AdequateIndeterminate
HMSNAndersen31 (1996)DoubtfulIndeterminate
Fillyaw et al32 (1989)DoubtfulDoubtful
Motor neuron diseasesGriffin et al34 (1994)DoubtfulIndeterminate
Myotonic dystrophy type 1Knak et al33 (2020)AdequateAdequate
Pooled NMDsTiffreau et al36 (2003)AdequateIndeterminate
Tiffreau et al35 (2007)AdequateIndeterminate
a

HMSN = hereditary motor and sensory neuropathy; NMDs = neuromuscular diseases.

Quality of Measurement Properties

Eight studies reported on the reliability of isokinetic measurements26–29,31,34–36 (Tab. 4). All outcomes but angle at peak torque showed sufficient reliability (ICC = 0.7–0.99) on various speeds, joints and movements in postpoliomyelitis syndrome, HMSN, motor neuron diseases, and combinations of NMDs. The reliability of isometric measurements was determined in 6 studies.26–28,30,32,33 The highest MVC was sufficiently reliable in postpoliomyelitis syndrome (ICC = 0.92–0.98). In myotonic dystrophy type 1, torque had a sufficiently reliable outcome measure (ICC = 0.79–0.97). In HMSN, maximum isometric torque was sufficiently reliable in shoulder and elbow movements (ICC = 0.84–0.997).

Table 4

Synthesis of Results for Reliabilitya

DiagnosisType of MeasurementTotal No. of ParticipantsOutcome UsedICC RangeQuality of Measurement Properties: ReliabilityRisk of BiasInconsistencyImprecisionIndirectnessQuality of Evidence
Postpoliomyelitis syndromeIsometric154Highest MVC0.92–0.98Sufficient0000High
Isokinetic97Highest peak torque0.81–0.99Sufficient00−10Moderate
8Angle at peak torque0.20–0.91Insufficient−10−20Very low
8Average power0.88–0.98Sufficient−10−20Very low
8Maximum power0.84–0.98Sufficient−10−20Very low
HMSNIsometric30Maximum isometric torque0.84–0.96Sufficient−20−20Very low
Isokinetic4–7Highest peak torque0.975–0.997bSufficient−20−20Very low
Myotonic dystrophy type 1Isometric58–72Torque0.79–0.97Sufficient−10−10Low
Motor neuron diseasesIsokinetic5Average torque≥0.97Sufficient−20−20Very low
Multiple pooled NMDsIsokinetic15Highest peak torque0.92–0.99Sufficient−10−20Very low
15Angle at peak torque0.56–0.90Insufficient−10−20Very low
15Maximum power0.94–0.98Sufficient−10−20Very low
14Mean work0.78–0.84Sufficient−10−20Very low
14Maximum work0.7–0.83Sufficient−10−20Very low
DiagnosisType of MeasurementTotal No. of ParticipantsOutcome UsedICC RangeQuality of Measurement Properties: ReliabilityRisk of BiasInconsistencyImprecisionIndirectnessQuality of Evidence
Postpoliomyelitis syndromeIsometric154Highest MVC0.92–0.98Sufficient0000High
Isokinetic97Highest peak torque0.81–0.99Sufficient00−10Moderate
8Angle at peak torque0.20–0.91Insufficient−10−20Very low
8Average power0.88–0.98Sufficient−10−20Very low
8Maximum power0.84–0.98Sufficient−10−20Very low
HMSNIsometric30Maximum isometric torque0.84–0.96Sufficient−20−20Very low
Isokinetic4–7Highest peak torque0.975–0.997bSufficient−20−20Very low
Myotonic dystrophy type 1Isometric58–72Torque0.79–0.97Sufficient−10−10Low
Motor neuron diseasesIsokinetic5Average torque≥0.97Sufficient−20−20Very low
Multiple pooled NMDsIsokinetic15Highest peak torque0.92–0.99Sufficient−10−20Very low
15Angle at peak torque0.56–0.90Insufficient−10−20Very low
15Maximum power0.94–0.98Sufficient−10−20Very low
14Mean work0.78–0.84Sufficient−10−20Very low
14Maximum work0.7–0.83Sufficient−10−20Very low
a

HMSN = hereditary motor and sensory neuropathy; MVC = maximum voluntary contraction; NMDs = neuromuscular diseases.

b

r values.

Table 4

Synthesis of Results for Reliabilitya

DiagnosisType of MeasurementTotal No. of ParticipantsOutcome UsedICC RangeQuality of Measurement Properties: ReliabilityRisk of BiasInconsistencyImprecisionIndirectnessQuality of Evidence
Postpoliomyelitis syndromeIsometric154Highest MVC0.92–0.98Sufficient0000High
Isokinetic97Highest peak torque0.81–0.99Sufficient00−10Moderate
8Angle at peak torque0.20–0.91Insufficient−10−20Very low
8Average power0.88–0.98Sufficient−10−20Very low
8Maximum power0.84–0.98Sufficient−10−20Very low
HMSNIsometric30Maximum isometric torque0.84–0.96Sufficient−20−20Very low
Isokinetic4–7Highest peak torque0.975–0.997bSufficient−20−20Very low
Myotonic dystrophy type 1Isometric58–72Torque0.79–0.97Sufficient−10−10Low
Motor neuron diseasesIsokinetic5Average torque≥0.97Sufficient−20−20Very low
Multiple pooled NMDsIsokinetic15Highest peak torque0.92–0.99Sufficient−10−20Very low
15Angle at peak torque0.56–0.90Insufficient−10−20Very low
15Maximum power0.94–0.98Sufficient−10−20Very low
14Mean work0.78–0.84Sufficient−10−20Very low
14Maximum work0.7–0.83Sufficient−10−20Very low
DiagnosisType of MeasurementTotal No. of ParticipantsOutcome UsedICC RangeQuality of Measurement Properties: ReliabilityRisk of BiasInconsistencyImprecisionIndirectnessQuality of Evidence
Postpoliomyelitis syndromeIsometric154Highest MVC0.92–0.98Sufficient0000High
Isokinetic97Highest peak torque0.81–0.99Sufficient00−10Moderate
8Angle at peak torque0.20–0.91Insufficient−10−20Very low
8Average power0.88–0.98Sufficient−10−20Very low
8Maximum power0.84–0.98Sufficient−10−20Very low
HMSNIsometric30Maximum isometric torque0.84–0.96Sufficient−20−20Very low
Isokinetic4–7Highest peak torque0.975–0.997bSufficient−20−20Very low
Myotonic dystrophy type 1Isometric58–72Torque0.79–0.97Sufficient−10−10Low
Motor neuron diseasesIsokinetic5Average torque≥0.97Sufficient−20−20Very low
Multiple pooled NMDsIsokinetic15Highest peak torque0.92–0.99Sufficient−10−20Very low
15Angle at peak torque0.56–0.90Insufficient−10−20Very low
15Maximum power0.94–0.98Sufficient−10−20Very low
14Mean work0.78–0.84Sufficient−10−20Very low
14Maximum work0.7–0.83Sufficient−10−20Very low
a

HMSN = hereditary motor and sensory neuropathy; MVC = maximum voluntary contraction; NMDs = neuromuscular diseases.

b

r values.

Six studies reported on measurement error26–28,30,32,33; 3 of these studies reported the smallest detectable change. Of these, 2 studies investigated postpoliomyelitis and 1 study investigated myotonic dystrophy type 1.26,28,33 The quality of the measurement properties for measurement error were indeterminate due to the absence of reported anchor-based minimal important change.21 Additional data on measurement properties can be found in Supplementary Appendix 2.

Quality of Evidence

Evidence of high quality was found for isometric measurements of strength in postpoliomyelitis syndrome using the highest MVC as outcome measure. Isokinetic measurements in postpoliomyelitis syndrome showed only moderate quality of evidence when highest peak torque was used as the outcome measure. For other NMDs, quality of evidence was low to very low for both isometric and isokinetic measurements (Tab. 4) as a result of small sample sizes and doubtful methodological quality.

Reliability

Evidence of high quality was found for isometric measurements of strength in postpoliomyelitis syndrome using the highest MVC as outcome measure. Isokinetic measurements in postpoliomyelitis syndrome showed only moderate quality of evidence when highest peak torque was used as the outcome measure. For other NMDs, quality of evidence was low to very low for both isometric and isokinetic measurements (Tab. 4) as a result of small sample sizes and doubtful methodological quality.

Measurement Error

We could not perform a best-evidence synthesis of measurement error because the quality of the measurement property could not be determined.

Discussion

We found moderate to high evidence that measuring strength and power using an isokinetic dynamometer in postpoliomyelitis syndrome is reliable. For other NMDs, the quality of evidence on reliability was low to very low, mostly because of small sample sizes and doubtful methodological quality of the included studies. The quality of evidence for measurement error was indefinable for all NMDs because of the absence of reported anchor-based minimal clinically important differences.

In terms of reliability, the findings of this systematic review are similar to the results of reviews on adults who were healthy13 and for isokinetic knee strength measures in children.37 They are comparable with results of reliability studies in patients with fibromyalgia,38 heart failure,39 and chronic obstructive pulmonary disease,40 which all concluded that IDs are reliable tools for measuring strength and/or power. A previous study by Mhandi et al41 focusing on the use of IDs in NMDs summarized the reliability of isokinetic measurements in NMDs as high based solely on ICCs. The present study shows comparable results on ICCs and additionally provides the quality of evidence in different groups of NMDs.

Measurement errors of strength testing in NMDs are similar to studies of adults who were healthy for isokinetic measurements of knee (SEM = 5.4–17.3 N·m) and ankle (SEM = 0.9–5.7 N·m) and isometric knee strength (SEM = 6.2–8.1 N·m).13 Compared with adults who were healthy, patients with NMDs showed comparable to slightly higher measurement error, expressed as SEM%, which indicates the relative measurement error (SEM divided by the group mean × 100). SEM% is comparable in isometric knee measurements (6.4%–13.0% in NMDs vs 3.6%–15.9% in adults who were healthy) and slightly higher in isometric hip measurements (14.0%–15.0% in NMDs vs 6.5%–10.8% in adults who were healthy) and ankle measurements (12.4%–30.0% in NMDs vs 5.8%–13.8% in adults who were healthy).13 Higher SEM% compared with adults who were healthy may reflect the overall lower strength in patients with NMDs given that absolute SEMs are comparable.13

Multiple studies have showed the impact of force conductor placement, rest before and between measurements, and use of the handrail on strength and power measured. With respect to the former, Andersen reported that a 3-cm displacement of the force conductor impacted the outcome of plantar flexor strength measurement by as much as 23 N·m.31 Rest has been shown to impact strength measurement outcomes in both adults who were healthy and patients with NMDs.31,42 Nunes et al reported differences in strength outcome and reliability while using the handrails in isokinetic knee strength measurements.43 These findings demonstrate the importance of meticulous description of any protocol used to assess muscle function using isokinetic dynamometry.

Three studies included in our search outcome used the continuous passive mode (CPM) on 7 patients with HMSN31 and 2 groups with multiple NMDs and with strength ranging from Medical Research Council grade 2 to Medical Research Council grade 5.31,35,36 In isokinetic measurements using CPM, the dynamometer moves at a constant speed and supports the tested limb through a preset range of motion. CPM is described as a valid mode to measure torque, making it possible to measure strength in patients with Medical Research Council grades as low as 2 and 3.44,45 The included studies that used CPM showed similar reliability compared with the other studies, indicating IDs might not only be of use in stronger but also in weaker patients.

The described test procedures per study can be used to create reliable strength measurements, and the reported SEM allows to determine whether improvement and decrease in strength are true changes.

Clinicians can use the outcome of this study to set up a protocol for strength and power measurements in NMDs when using isokinetic dynamometers. The synthesis of results provided by the present investigation summarizes the current quality of evidence for the use of isokinetic dynamometers in measuring strength and/or power in NMDs.

Limitations

Limitations of the Systematic Review

This study used the COSMIN criteria to include studies. The COSMIN was developed to measure measurement properties of patient-reported outcomes. Several reviews show that the guideline can also be used to review studies on clinical outcome measures.46–48

The COSMIN guidelines only include studies with measurement properties as a specific research aim. Studies focusing on other specific aims such as validation of HHD that additionally reported on reliability were excluded.49 This review did not address feasibility. Although feasibility is not part of reliability, it affects the clinical applicability of isokinetic dynamometry. In this review, we compared outcomes obtained with different IDs, although a recent study did not show differences between the Biodex System 3 Pro and the Cybex Humac Norm 770. However, it is not enough evidence that there is no difference at all between IDs. Therefore, we cannot rule out the influence of differences between IDs.50 We therefore recommend the use of the same ID for specific patient groups.

Limitations of the Reviewed Studies

Multiple studies26–29,31 excluded patients who did not reach the preset velocity in isokinetic measurements. Thus, reliability and measurement error are determined for a subgroup of the population and do not represent the entire population. Only 2 studies included more than 50 patients, which may be considered an adequate number of patients to determine reliability estimates.51 With limited samples sizes and only a few studies at each NMD, only the highest MVC measured isometrically in post-poliomyelitis patients was not scored down for imprecision for having <100 participants.

One study used the Pearson r to measure reliability in isokinetic measurements in HMSN. Although the COSMIN guidelines score down on the methodological quality of the study, a correlation >0.7 is considered sufficient when scoring the measurement properties. Comparison with ICCs can be misleading even when ICCs and r point in the same direction of high correlation/reliability.

This systematic review provides an overview of strength and power testing in NMDs using isokinetic dynamometers. We found that the quality of evidence regarding the reliability of measuring strength and power with isokinetic dynamometers in neuromuscular diseases was variable between NMDs. Specifically, isometric measurements in postpoliomyelitis syndrome were found to have high reliability, whereas reliability was low to very low for isometric and isokinetic measurements in HMSN and other NMDs. This review has identified highest peak torque or highest MVC as the most commonly used and reliable outcome, whereas angle at peak torque should be considered an inappropriate outcome measure.

Recommendations

Studies with larger samples sizes are needed to create reliable protocols and useful reliability data for measuring patients with NMDs. An accurate description of the protocol used can improve reliability. To improve reliability, the authors recommend using the arms crossed on the chest for knee measurements and to standardize patient and force conductor placement. Highest peak torque/highest MVC show the highest quality of evidence and highest reliability and are therefore recommended as an outcome measure. For weaker patients, the use of lower velocities in isokinetic testing or measurements using CPM is recommended. In stronger patients, higher velocities are recommended because they have a higher correlation with walking. Finally, the authors recommend the use of standardized rest intervals between measurements to limit the effect of fatigue.

Further research on the use of CPM in isokinetic measurements is useful for the measurement of power in patients with lower strength who are unable to achieve the preset velocity.

Author Contributions

Concept/idea/research design: D.R. van der Woude, B. Bartels

Writing: D.R. van der Woude, B. Bartels

Data collection: D.R. van der Woude, T. Ruyten

Data analysis: D.R. van der Woude, T. Ruyten, B. Bartels

Project management: D.R. van der Woude, B. Bartels

Providing facilities/equipment: B. Bartels

Consultation (including review of manuscript before submitting): T. Ruyten, B. Bartels

Systematic Review Registration

This review was not registered.

Disclosures

The authors completed the ICMJE Form for Disclosure of Potential Conflicts of Interest and reported no conflicts of interest.

References

1.

Bhatt
JM
.
The epidemiology of neuromuscular diseases
.
Neurol Clin
.
2016
;
34
:
999
1021
.

2.

Johnson
NE
,
Quinn
C
,
Eastwood
E
,
Tawil
R
,
Heatwole
CR
.
Patient-identified disease burden in facioscapulohumeral muscular dystrophy
.
Muscle Nerve
.
2012
;
46
:
951
953
.

3.

Cornwall
KM
,
Butterfield
RJ
,
Hernandez
A
,
Heatwole
C
,
Johnson
NE
.
A qualitative approach to health related quality-of-life in congenital muscular dystrophy
.
J Neuromuscul Dis
.
2018
;
5
:
251
255
.

4.

Wadman
RI
,
Wijngaarde
CA
,
Stam
M
, et al.
Muscle strength and motor function throughout life in a cross-sectional cohort of 180 patients with spinal muscular atrophy types 1c–4
.
Eur J Neurol
.
2018
;
25
:
512
518
.

5.

Mcdonald
CM
,
Henricson
EK
,
Abresch
RT
, et al.
The 6-minute walk test and other endpoints in Duchenne muscular dystrophy: longitudinal natural history observations over 48 weeks from a multicenter study
.
Muscle Nerve
.
2013
;
48
:
343
356
.

6.

Aitkens
S
,
Lord
J
,
Bernauer
E
, et al.
Relationship of manual muscle testing to objective strength measurements
.
Muscle Nerve
.
1989
;
12
:
173
177
.

7.

Stark
T
,
Walker
B
,
Phillips
JK
,
Fejer
R
,
Beck
R
.
Hand-held dynamometry correlation with the gold standard isokinetic dynamometry: a systematic review
.
PM&R
.
2011
;
3
:
472
479
.

8.

Sapega
AA
,
Drillings
G
.
The definition and assessment of muscular power
.
J Orthop Sports Phys Ther
.
1983
;
5
:
7
9
.

9.

Hislop
HJ
.
Daniels & Worthinghams Muscle Testing: Techniques of Manual Examination
. 8th ed.
St. Louis, MO: Saunders Elsevier
;
2007
.

10.

Bohannon
RW
.
Hand-held dynamometry: adoption 1900–2005
.
Percept Mot Skills
.
2006
;
103
:
3
4
.

11.

Hébert
LJ
,
Remec
JF
,
Saulnier
J
,
Vial
C
,
Puymirat
J
.
The use of muscle strength assessed with handheld dynamometers as a non-invasive biological marker in myotonic dystrophy type 1 patients: a multicenter study
.
BMC Musculoskelet Disord
.
2010
;
11
:
72
. https://doi.org/10.1186/1471-2474-11-72.

12.

Andres
PL
,
Skerry
LM
,
Thornell
B
,
Portney
LG
,
Finison
LJ
,
Munsat
TL
.
A comparison of three measures of disease progression in ALS
.
J Neurol Sci
.
1996
;
139
:
64
70
.

13.

Chamorro
C
,
Armijo-Olivo
S
,
De La Fuente
C
,
Fuentes
J
,
Javier
CL
.
Absolute reliability and concurrent validity of hand held dynamometry and isokinetic dynamometry in the hip, knee and ankle joint: systematic review and meta-analysis
.
Open Med
.
2017
;
12
:
359
375
.

14.

Chamorro
C
,
Arancibia
M
,
Trigo
B
,
Arias-poblete
L
,
Jerez-Mayorga
D
.
Absolute reliability and concurrent validity of hand-held dynamometry in shoulder rotator strength assessment
.
Systematic review and meta-analysis
. Int J Environ Res Public Health.
2021
;
18
:
9293
. https://doi.org/10.3390/ijerph18179293.

15.

Fulcher
ML
,
Hanna
CM
,
Raina
EC
.
Reliability of handheld dynamometry in assessment of hip strength in adult male football players
.
J Sci Med Sport
.
2010
;
13
:
80
84
.

16.

Kristensen
OH
,
Stenager
E
,
Dalgas
U
.
Muscle strength and poststroke hemiplegia: a systematic review of muscle strength assessment and muscle strength impairment
.
Arch Phys Med Rehabil
.
2017
;
98
:
368
380
.

17.

Baltzopoulos
V
,
Brodie
DA
.
Isokinetic dynamometry
. Applications and limitations.
Sports Med
.
1989
;
8
:
101
116
.

18.

Lexell
J
,
Flansbjer
UB
,
Brogårdh
C
.
Isokinetic assessment of muscle function: our experience with patients afflicted with selected diseases of the nervous system
.
Isokinet Exerc Sci
.
2012
;
20
:
267
273
.

19.

Kamo
T
,
Asahi
R
,
Azami
M
, et al.
Rate of torque development and the risk of falls among community dwelling older adults in Japan
.
Gait Posture
.
2019
;
72
:
28
33
.

20.

de
Vet
HCW
,
Terwee
CB
,
Knol
DL
,
Bouter
LM
.
When to use agreement versus reliability measures
.
J Clin Epidemiol
.
2006
;
59
:
1033
1039
.

21.

Prinsen
CAC
,
Mokkink
LB
,
Bouter
LM
, et al.
COSMIN guideline for systematic reviews of patient-reported outcome measures
.
Qual Life Res
.
2018
;
27
:
1147
1157
.

22.

Hopkins
WG
.
Measures of reliability in sports medicine and science
.
Sport Med
.
2000
;
30
:
1
15
. https://doi.org/10.2165/00007256-200030010-00001.

23.

Bartlett
JW
,
Frost
C
.
Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables
.
Ultrasound Obstet Gynecol
.
2008
;
31
:
466
475
.

24.

Terwee
CB
,
Jansma
EP
,
Riphagen
II
,
De Vet
HCW
.
Development of a methodological PubMed search filter for finding studies on measurement properties of measurement instruments
.
Qual Life Res
.
2009
;
18
:
1115
1123
.

25.

Mokkink
LB
,
de
Vet
HCW
,
Prinsen
CAC
, et al.
COSMIN risk of bias checklist for systematic reviews of patient-reported outcome measures
.
Qual Life Res
.
2018
;
27
:
1171
1179
.

26.

Brogårdh
C
,
Flansbjer
UB
,
Carlsson
H
,
Lexell
J
.
Intra-rater reliability of arm and hand muscle strength measurements in persons with late effects of polio
.
PM R
.
2015
;
7
:
1035
1041
.

27.

Flansbjer
U
,
Drake
AM
,
Lexell
J
.
Reproducibility of ankle dorsiflexor muscle strength measurements in individuals with post-polio syndrome
.
Isokinet Exerc Sci
.
2011
;
19
:
55
61
.

28.

Flansbjer
UBB
,
Lexell
J
.
Reliability of knee extensor and flexor muscle strength measurements in persons with late effects of polio
.
J Rehabil Med
.
2010
;
42
:
588
592
.

29.

Kilfoil
MR
,
St Pierre
DMM
.
Reliability of Cybex II isokinetic evaluations of torque in post-poliomyelitis syndrome
.
Arch Phys Med Rehabil
.
1993
;
74
:
730
735
.

30.

Horemans
HL
,
Beelen
A
,
Nollet
F
, et al.
Reproducibility of maximal quadriceps strength and its relationship to maximal voluntary activation in postpoliomyelitis syndrome
.
Arch Phys Med Rehabil
.
2004
;
85
:
1273
1278
.

31.

Andersen
H
.
Reliability of isokinetic measurements of ankle dorsal and plantar flexors in normal subjects and in patients with peripheral neuropathy
.
Arch Phys Med Rehabil
.
1996
;
77
:
265
268
.

32.

Fillyaw
M
,
Badger
G
,
Bradley
W
.
Quantitative measures of neurological function in chronic neuromuscular diseases and ataxia
.
J Neurol Sci
.
1989
;
92
:
17
36
.

33.

Knak
KL
,
Sheikh
AM
,
Andersen
H
,
Witting
N
,
Vissing
J
.
Intrarater reliability and validity of outcome measures in myotonic dystrophy type 1
.
Neurology
.
2020
;
94
:
e2508
e2520
.

34.

Griffin
JW
,
Tooms
RE
,
Vander Zwaag
R
,
O’Toole
ML
,
Bertorini
TE
.
Eccentric and concentric muscle performance in patients with spastic paresis secondary to motor neuron disease. A preliminary report
.
Neuromuscul Disord
.
1994
;
4
:
131
138
.

35.

Tiffreau
V
,
Ledoux
I
,
Eymard
B
,
Thévenon
A
,
Hogrel
JY
.
Isokinetic muscle testing for weak patients suffering from neuromuscular disorders: a reliability study
.
Neuromuscul Disord
.
2007
;
17
:
524
531
.

36.

Tiffreau
V
,
Turlure
A
,
Viet
G
, et al.
Isokinetic dynamometry of weak muscles
.
Isokinet Exerc Sci
.
2003
;
11
:
13
20
.

37.

Muñoz-Bermejo
L
,
Pérez-Gómez
J
,
Manzano
F
,
Collado-Mateo
D
,
Villafaina
S
,
Adsuar
JC
.
Reliability of isokinetic knee strength measurements in children: a systematic review and meta-analysis
.
PLoS One
.
2019
;
14
:
1
15
. https://doi.org/10.1371/journal.pone.0226274.

38.

Adsuar
JC
,
Olivares
PR
,
del
Pozo-Cruz
B
,
Parraca
JA
,
Gusi
N
.
Test-retest reliability of isometric and isokinetic knee extension and flexion in patients with fibromyalgia: evaluation of the smallest real difference
.
Arch Phys Med Rehabil
.
2011
;
92
:
1646
1651
.

39.

Quittan
M
,
Wiesinger
GF
,
Crevenna
R
, et al.
Isokinetic strength testing in patients with chronic heart failure--a reliability study
.
Int J Sports Med
.
2001
;
22
:
40
44
.

40.

Frykholm
E
,
Géphine
S
,
Saey
D
, et al.
Inter-day test–retest reliability and feasibility of isokinetic, isometric, and isotonic measurements to assess quadriceps endurance in people with chronic obstructive pulmonary disease: a multicenter study
.
Chron Respir Dis
.
2018
;
16
:1479973118816497. https://doi.org/10.1177/1479973118816497.

41.

El Mhandi
L
,
Bethoux
F
.
Isokinetic testing in patients with neuromuscular diseases: a focused review
.
Am J Phys Med Rehabil
.
2013
;
92
:
163
178
.

42.

Stratford
PW
,
Bruulsema
A
,
Maxwell
B
,
Black
T
,
Harding
B
.
The effect of inter-trial rest interval on the assessment of isokinetic thigh muscle torque
.
J Orthop Sports Phys Ther
.
1990
;
11
:
362
366
.

43.

Nunes
JP
,
Cunha
PM
,
Mayhew
JL
, et al.
Influence of handgrip stabilization during isokinetic knee strength assessment in older women
.
Percept Mot Skills
.
2020
;
127
:
671
683
.

44.

Merlini
L
,
Dell’Accio
D
,
Holzl
A
,
Granata
C
.
Isokinetic muscle testing (IMT) in neuromuscular diseases. Preliminary report
.
Neuromuscul Disord
.
1992
;
2
:
201
207
.

45.

Nordez
A
,
Casari
P
,
Cornu
C
.
Accuracy of Biodex system 3 pro computerized dynamometer in passive mode
.
Med Eng Phys
.
2008
;
30
:
880
887
.

46.

Bartels
B
,
de
Groot
JF
,
Terwee
CB
.
The six-minute walk test in chronic pediatric conditions: a systematic review of measurement properties
.
Teach Stat
.
2013
;
16
:
60
61
.

47.

Drake
D
,
Kennedy
R
,
Wallace
E
.
The validity and responsiveness of isometric lower body multi-joint tests of muscular strength: a systematic review
.
Sport Med - Open
.
2017
;
3
:
23
. https://doi.org/10.1186/s40798-017-0091-2.

48.

Seager
A
,
French
H
,
Meldrum
D
.
Measurement properties of instruments for assessment of cervical spine function in infants with torticollis: a systematic review
.
Eur J Pediatr
.
2019
;
178
:
657
671
.

49.

Kilmer
DD
,
McCrory
MA
,
Wright
NC
,
Rosko
RA
,
Kim
HR
,
Aitkens
SG
.
Hand-held dynamometry reliability in persons with neuropathic weakness
.
Arch Phys Med Rehabil
.
1997
;
78
:
1364
1368
.

50.

De Araujo Ribeiro Alvares
JB
,
Rodrigues
R
,
De Azevedo Franke
R
, et al.
Inter-machine reliability of the Biodex and Cybex isokinetic dynamometers for knee flexor/extensor isometric, concentric and eccentric tests
.
Phys Ther Sport
.
2015
;
16
:
59
65
.

51.

de
Vet
HCW
,
Terwee
CB
,
Mokkink
LB
,
Knol
DL
.
Measurement in Medicine: A Practical Guide
. Cambridge, UK:
Cambridge University Press
;
2011
. https://doi.org/10.1017/CBO9780511996214.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.

Comments

0 Comments
Submit a comment
You have entered an invalid code
Thank you for submitting a comment on this article. Your comment will be reviewed and published at the journal's discretion. Please check for further notifications by email.