Aims

The miniaturized leadless pacemaker has recently emerged as a bradycardia therapy in humans, and many patients may need at least two pacemakers in their lifetime. Thus, the present study assessed the effect of two leadless pacemakers in the right ventricle (RV) on cardiac function in a swine model.

Methods and results

Fourteen mini pigs were chronically studied. Each animal received two sequential leadless pacemakers in the RV with 1 month between two separate implantations. All animals were then followed for 6 months. Cardiac function was assessed with echocardiography at baseline, the second implantation, and the end of the 6-month follow-up. A necropsy at the end of the study was performed to measure the length of the fibrous tissue covering the devices and assess tricuspid valve integrity. Four animals died during the surgical procedure, and one animal died of infection during the follow-up period. In the remaining nine animals, there were no significant changes in left ventricular ejection fraction, aortic time integral, cardiac output, and left ventricular size from baseline to the end of the study. The mean length of fibrous tissue covering Micra devices was 14.3 ± 7.8 mm. No tricuspid valve injury was observed.

Conclusion

Multiple leadless pacemakers implanted in the RV do not appear to impact cardiac function. Therefore, it appears safe to implant multiple leadless pacemakers in the same heart.

What's new?

  • Transcatheter leadless pacemaker.

  • Majority of pacemaker patients need at least two devices in their life time. The present study demonstrated the feasibility of implanting two leadless pacemakers in the right ventricle in an animal model.

  • The study raised several safety issues that should be considered during clinical practice.

Introduction

Conventional pacemakers with transvenous leads in the right ventricle (RV) have been used clinically for more than 50 years. While pacemakers have demonstrated effectiveness in treating bradycardia, complications associated with pacemaker implantation can cause patient injury and increased healthcare costs. The most common complications after pacemaker implantation include pacing lead dislodgement and fracture, pocket haematoma and infection, and tricuspid valve impairment.111 The investigation by Kirkfeldt et al.10 found that 9.5% of 5492 patients who received a pacemaker experienced at least one complication.

A transcatheter pacemaker with a significantly smaller size compared with conventional pacemaker has been developed for bradycardia therapy because of the advancement of technology in the areas of miniaturization, batteries, electronic packaging, biocompatibility, ultra-low power circuitry, and communication systems. Due to its small size, the transcatheter pacemaker has been designed to be implanted inside the RV. Leadless pacemakers such as Nanostim™ and Micra™ have recently been investigated clinically in patients with bradycardia worldwide.1216

It has been estimated that the average patient age at implantation is 75 years, and the average pacemaker now lasts ∼11 years.17 Therefore, many pacemaker patients will need at least two devices in their life time. Concerns have been raised that multiple devices in the heart may compromise cardiac function. Thus, in the present study, it was investigated whether two transcatheter pacemakers (Micra) sequentially implanted in the RV impacted cardiac function in a swine model.

Methods

The study protocol was reviewed and approved by the Ethical Committee of Animal Experimental Center in Fuwai hospital and complied with the institutional and national guidelines for the care and use of laboratory animals.

General procedures

A total of 14 mini pigs (45–55 kg, either sex) were used. Animals were sedated with mixture of ketamine (35 mg/kg, Jiangsu Hengrui Medicine Co. Ltd, Jiangsu, China) and diazepam (1.5 mg/kg, Sanjing Pharmaceutical Co. Ltd, Heilongjiang, China) and then intubated for artificial respiration with a respirator (Siemens Servo 900C; Siemens Medical Solutions, Erlangen, Germany) and under continuous cardiac-respiratory monitoring. Anaesthesia was maintained by intravenous infusion of 3.5% pentobarbital (Sigma-Aldrich Co. LLC., MO, USA). Ringer's solution was supplied whenever necessary depending on blood gas–electrolyte analysis. Cardiac functional assessment was conducted with transthoracic echocardiography before implantation. After implanting the first Micra device, animals were allowed to recover for 1 month. Each pig was then implanted with a second device, and during the procedure, cardiac functional assessment with echocardiography was repeated. Following the second implantation, animals were followed with daily monitoring of temperature, activity, and dietary intake for additional 6 months. At the end of the study, animals were anaesthetized, the locations of the Micra pacemakers were documented using fluoroscopy, and cardiac function assessment was repeated. Then, animals were sacrificed with 3 M KCl (Zizhu Pharmaceutical Co. Ltd, Beijing, China), the chest was opened, the epicardial surface was examined, and the RV was carefully opened to examine the devices in the RV. Tissue near the electrode of each device was fixed, sliced, and stained for histological analysis.

Implantation of Micra device

In each animal, the skin of the right-side neck was prepared by tincture of iodine solution, and then the right jugular vein was exposed. An incision was made on the exposed right jugular vein to allow the vein access, and a modified Micra introducer was then tracked to right atrium over a guide wire under fluoroscopy. The modification was to replace the 56 cm sheath with a 15 cm solid plastic 24 Fr sheath while keeping the same valve. After withdrawing the guide wire and dilator, the Micra delivery system was inserted into the right atrium through the introducer. The delivery system contained the Micra device (Medtronic plc., MN, USA) that was 25.9 mm long and 6.7 mm in diameter with four tines for fixation (Figure 1A).
Micra device and fluoroscopic imaging with Micra devices: (A) an example of the Micra device and (B) fluoroscopic imaging of two Micra devices implanted in the RV.
Figure 1

Micra device and fluoroscopic imaging with Micra devices: (A) an example of the Micra device and (B) fluoroscopic imaging of two Micra devices implanted in the RV.

Then, the delivery system was articulated to access the RV through the tricuspid valve. Once the system reached the RV apex, the Micra device was deployed from the delivery system allowing the fixation tines to embed into the cardiac tissue. A fixation test was performed under fluoroscopy to ensure that at least two of the four tines were embedded in the myocardium. Once the device was adequately fixed in the RV, the tether was gently pulled out of the delivery system to free the device. Then, the delivery system was removed from the introducer, and the introducer was withdrawn from the jugular vein. The jugular vein was repaired, and the subcutaneous tissue and skin were sutured. One month later, the second Micra device was implanted in an identical manner ∼1 cm from the previously implanted Micra as shown in Figure 1B.

Data collection and analyses

Procedure time and fluoroscopy time

The implantation procedure times were recorded from skin cut to closure. The procedure fluoroscopy times in minutes were also recorded.

Echocardiography

In all sedated animals, transthoracic echocardiography was conducted with a 3.5-MHz phased-array transducer of the echo machine (GE VividE9 Dimension, GE Health, Milwaukee, WI). Two-dimensional and Doppler examinations were performed to make the measurements of cardiac four-chamber left ventricular end-diastolic diameter (LVEDD), left ventricular end-diastolic volume (LVEDV), left ventricular end-systolic volume (LVESV), left ventricular ejection fraction (LVEF, 2-D Simpson's method), aortic velocity time integral (VTI), cardiac output (CO, = stroke volume × heart rate), the aortic pre-ejection interval (APEI), and pulmonic pre-ejection interval (PPEI). Aortic pre-ejection interval is the time from Q-wave onset on the electrocardiogram (ECG) to aortic valve opening as measured by echocardiography. Pulmonic pre-ejection interval is the time from Q-wave on the ECG to the pulmonic valve opening. All echocardiographic values were obtained from the average of measurements from six continuous normal cardiac beats.

Complications

All complications including implantation failure, device dislodgement, implantation-related infection, and death were documented.

Post-mortem

A necropsy was performed immediately after the animal was sacrificed. After the epicardium was examined, the RV was opened along the line from the tricuspid valve (between the anterior and the posterior leaflets) to the RV apex. The locations of Micra devices were photographed with a digital camera, and the length of fibrotic tissue covering each implanted device was measured. The length of fibrous tissue covering the Micra was calculated by measuring the distance from the device proximal end to the edge of fibrous adhesion (Figure 2), and subtracting this value from the device total length (25.9 mm).
Fibrous tissue attaching around Micra devices observed at necropsy. Arrow indicates the fibrous tissue attachment.
Figure 2

Fibrous tissue attaching around Micra devices observed at necropsy. Arrow indicates the fibrous tissue attachment.

Histology examinations of cardiac tissue

The Micra devices were removed from the heart, and then the implantation sites in the RV were excised transmurally. Samples of the myocardium in the right and left ventricles that were far away from the implantation sites were also collected as control in each heart. These samples were fixed in 10% neutral buffered formalin. Fixed tissue samples were sectioned transmurally (endocardium to epicardium) with steps of 200 µm intervals, and slides were stained with haematoxylin and eosin (HE) and Masson's trichrome stain for microscopic examination.

The tissue fibrosis surrounding the implantation sites were measured by a pathologist who was blinded to study design. Sections were analysed using a Q-Imaging Micro Publisher 5.0 RTV microscope camera (Q-Imaging, BC, Canada) on a microscope (Olympus BX51, Olympus Corporation, Japan) with a ×10 magnification objective. Tissue fibrosis was assessed with (i) the distance between the tip of the pacing electrode and the edge of tissue fibrosis, (ii) fibrosis area around the Micra device per slide view, and (iii) the fibrosis thickness around the tines of Micra devices.

Statistical analysis

Data collected in animals that survived the second device implantation were included in the statistical analysis. Data are presented as mean ± SD or percentage wherever applicable. A paired or unpaired t-test was used to compare numeral values. Longitudinal data were analysed with repeated measures ANOVA. A statistically significant difference required a probability of P < 0.05. All statistical analyses were conducted using statistical software package SPSS 19.0 (SPSS, Chicago, IL, USA).

Results

Implantation outcomes

Fourteen pigs underwent implantation procedures, of which 5 animals died prior to the end of the 6-month follow-up. The deaths of 4 animals were caused by procedure-related reasons, and 1 died due to chronic infection 139 days after the second Micra device implantation. The procedural-related complications included one acute device dislodgement, two atrial ruptures by the introducer, and one massive air embolism at the time of the second implantation.

Surgical procedure time and fluoroscopy time

In Figure 3 are shown the procedure times and fluoroscopy times for implantation. The total procedure time and fluoroscopy use were similar between the first and second implantations. The fluoroscopy time in the first 4 animals of 33.9 ± 11.4 min was significantly longer than the fluoroscopy time in the last 5 animals (8.9 ± 8.1 min, P < 0.001).
Procedure and fluoroscopy duration (minutes) of each animal. Four symbols and lines represent the time of procedure and fluoroscopy in the first and second implantations, respectively.
Figure 3

Procedure and fluoroscopy duration (minutes) of each animal. Four symbols and lines represent the time of procedure and fluoroscopy in the first and second implantations, respectively.

Assessment of cardiac function

Cardiac dimensions and functional measurements are summarized in Table 1. Comparing the 1- and 7-month measurements with baseline, there was no significant change in LVEF, VTI, and CO. Similarly, there was no significant change in left ventricular size in terms of LVESV, LVEDV, and LVEDD. The inter-ventricular mechanical delay was assessed by the difference between APEI and PPEI. The difference was 6.6 ± 3.7 ms at baseline, 7.3 ± 4.4 ms at the second implantation, and 9.6 ± 7.6 ms at the end of the study (P = 0.46 among three time points).

Table 1

Echocardiographic measurements

Implantation 1Implantation 2EndP-value
LVESV (mL)15.8 ± 8.313.4 ± 4.316.1 ± 5.00.14
LVEDV (mL)42.6 ± 10.243.8 ± 9.746.2 ± 7.80.73
LVEDD (mm)38.0 ± 2.938.0 ± 2.839.2 ± 4.90.70
LVEF (%)66.4 ± 5.169.7 ± 6.265.4 ± 7.40.39
VTI (cm)17.2 ± 3.216.9 ± 2.916.5 ± 2.80.89
CO (L)2.0 ± 0.42.1 ± 0.22.0 ± 0.20.91
Implantation 1Implantation 2EndP-value
LVESV (mL)15.8 ± 8.313.4 ± 4.316.1 ± 5.00.14
LVEDV (mL)42.6 ± 10.243.8 ± 9.746.2 ± 7.80.73
LVEDD (mm)38.0 ± 2.938.0 ± 2.839.2 ± 4.90.70
LVEF (%)66.4 ± 5.169.7 ± 6.265.4 ± 7.40.39
VTI (cm)17.2 ± 3.216.9 ± 2.916.5 ± 2.80.89
CO (L)2.0 ± 0.42.1 ± 0.22.0 ± 0.20.91

P-value: to test the time impact on the measurements at the first implant, the second implantation and the end of the study.

LVESV, left ventricular end-systolic volume; LVEDV, left ventricular end-diastolic volume; LVEDD, left ventricular end-diastolic diameter; LVEF, left ventricular ejection fraction; VTI, velocity time integral; CO, cardiac output.

Table 1

Echocardiographic measurements

Implantation 1Implantation 2EndP-value
LVESV (mL)15.8 ± 8.313.4 ± 4.316.1 ± 5.00.14
LVEDV (mL)42.6 ± 10.243.8 ± 9.746.2 ± 7.80.73
LVEDD (mm)38.0 ± 2.938.0 ± 2.839.2 ± 4.90.70
LVEF (%)66.4 ± 5.169.7 ± 6.265.4 ± 7.40.39
VTI (cm)17.2 ± 3.216.9 ± 2.916.5 ± 2.80.89
CO (L)2.0 ± 0.42.1 ± 0.22.0 ± 0.20.91
Implantation 1Implantation 2EndP-value
LVESV (mL)15.8 ± 8.313.4 ± 4.316.1 ± 5.00.14
LVEDV (mL)42.6 ± 10.243.8 ± 9.746.2 ± 7.80.73
LVEDD (mm)38.0 ± 2.938.0 ± 2.839.2 ± 4.90.70
LVEF (%)66.4 ± 5.169.7 ± 6.265.4 ± 7.40.39
VTI (cm)17.2 ± 3.216.9 ± 2.916.5 ± 2.80.89
CO (L)2.0 ± 0.42.1 ± 0.22.0 ± 0.20.91

P-value: to test the time impact on the measurements at the first implant, the second implantation and the end of the study.

LVESV, left ventricular end-systolic volume; LVEDV, left ventricular end-diastolic volume; LVEDD, left ventricular end-diastolic diameter; LVEF, left ventricular ejection fraction; VTI, velocity time integral; CO, cardiac output.

Necropsy

Nine animals reached the study endpoint with a mean follow-up time of 215 ± 7 days. The mean length of fibrous tissue covering the Micra devices was 14.3 ± 7.8 mm. There were no observations of tricuspid valve injury in any of the 9 animals (0.0%). Furthermore, the fixation tines of the Micra device could be seen from the epicardium for 6 implanted devices in 5 animals (33.3% of all implanted Micra devices).

Based on the histological examination, the tissue fibrosis thickness measured from the edge of the tissue fibrosis to the endocardium where the Micra pacing electrode was placed against (as shown in Figure 4) was 595.0 ± 339.3 µm. The area (the percentage in each slide) of tissue fibrosis around Micra device was 23.8 ± 12.3%, and the average thickness around tines of the Micra devices was 362.1 ± 205.6 µm.
Tissue fibrosis beneath the Micra pacing electrode. Arrow indicates the thickness of the fibrosis measured from the endocardium where the device was placed.
Figure 4

Tissue fibrosis beneath the Micra pacing electrode. Arrow indicates the thickness of the fibrosis measured from the endocardium where the device was placed.

Discussion

The present study with animals chronically implanted with two Micra leadless pacemakers showed no significant changes in cardiac dimension and function in terms of echocardiographic parameters LVEDV, LVESV, LVEDD, LVEF, VTI, CO, and RV-LV synchrony. There was no observation of injury to the tricuspid valve. Additionally, the fluoroscopy time was significantly shortened as the implanters gained experience with the procedure. Histological examination showed tissue fibrosis surrounding the fixation tines of the Micra device. Overall, the present study demonstrates the feasibility of multiple Micra devices implanted in the RV.

Transcatheter leadless pacemakers like the Micra device have emerged as a result of technology advancement in the fields of electronic circuit, battery, power management, materials, miniature packaging, communication, and biocompatibility. There are several clinical advantages in implanting a small pacing device inside the ventricle for the treatment of bradycardia when compared with conventional pacemakers with transvenous leads. The transcatheter pacemaker without a transvenous lead and subcutaneous device pocket can avoid the complications such as lead dislodgement and fracture, and lead/device-related infection that have been frequently observed clinically.111

The longevity of conventional pacemakers in today's clinical practice varies from 8 to 10 years depending on different manufacturers and pattern of device service (programming). The average age of patients implanted with a pacemaker is 75 years and will survive on average 12 years.17,18 Therefore, many patients who receive their pacemaker will need a second pacemaker. Transcatheter pacemakers also have an estimated longevity of 10 years. Once the end of the device life approaches, there are two possibilities for the replacement of Micra device. The first is to implant another transcatheter pacemaker adjacent to the first one leaving the first one in place. To allow this approach, the Micra pacemaker has an off mode in which it cannot turn itself on under any circumstances. As shown in this study, placing two transcatheter pacemakers in the ventricle does not impact cardiac function. The second possibility is to extract the first transcatheter pacemaker and then implant a second one. It is expected that after 10 years, the device will be completely encapsulated making extraction more difficult. While it is possible to retrieve a transcatheter pacemaker before it is encapsulated,19 it is likely attempting to extract a completely encapsulated device would create additional risk to the patient. Because of this uncertainty around transcatheter pacemaker extraction, it seems prudent to leave the first transcatheter pacemaker and implant a second device next to it.

The size of these pig hearts was significantly smaller than the typical human heart. The LVEDV in adult humans ranges from 58 to 120 mL20 compared with the LVEDV between 40 and 50 mL for the pigs in this study. Therefore, the impact of the devices in the hearts of these pigs is likely stronger than what would be seen in humans. The fact that no difference in cardiac function was found in these smaller animals indicates that there is even less chance of an impact of multiple devices in the larger human hearts.

There were a number of safety findings in this study. Firstly, there was no case of tricuspid valve injury observed in the present study. The tricuspid valve injury incidence has been reported for pacing and defibrillation leads, which may lead to the development of right ventricular heart failure clinically.21 Secondly, no large fibrous tissues were found around any of the Micra devices, which have been reported before around pacing/defibrillation leads. Therefore, the impact of fibrous attachment around the Micra device on cardiac contraction may be minimal. Additionally, the fluoroscope time in the present study significantly decreased below 10 min once implanters mastered the implantation techniques after few cases of implantation.

In the present study, Micra tines were visible from the epicardial surface in some of the animals. This was probably due in part to the fact that the pig hearts have no trabeculae, so the devices were implanted deep in the apex. Penetration into the pericardial space by helixes is a known phenomenon with helix electrodes, and rarely requires any further intervention clinically.22 Thus, it is expected that the tine penetrations will have similarly few consequences.

The present study demonstrates that (i) it is technically feasible to implant two sequential Micra devices in the RV, and (ii) there are no significant changes in cardiac function in terms of LVEF, VTI, CO, and RV-LV synchrony assessed by echocardiography. Thus, the results of the present study demonstrate that multiple transcatheter pacemakers can be implanted in the RV with no negative impact on cardiac function.

Limitations

While the present study was a chronic observation in cardiac function, the follow-up period of 6 months is limited when compared with the device longevity of 10 years at which the second implantation is required. Thus, the present study cannot provide the long-term information regarding cardiac function and adverse events over 6 months. Furthermore, the implantation procedure of Micra devices used in the present study is different from that used in humans, especially the route of vein access. The present study used the jugular venous access, while femoral venous access is used clinically. This delivery route created some procedure-related complications in the study, which will not likely be seen using the femoral route. Moreover, the main objective of cardiac functional assessment following multiple Micra device implantations in the present study would not be affected by the measurement of Micra device electrical parameters because no pacing therapy was provided in the present study.

Conclusions

The present study demonstrates the feasibility of implanting multiple Micra devices in the RV without a negative impact on cardiac function in a swine model. Considering the uncertainty of the tool for extraction of Micra devices in the future, it is possible to implant two or more than two Micra devices in the same ventricle once the end of life of the first implanted Micra device approaches in clinical practice.

Funding

The study was sponsored by a research grant from CRHF Division, Medtronic, plc.

Conflict of interest: X.Z. and M.B. are employees of Medtronic that makes pacemaker products.

Acknowledgements

The authors would like to thank Lepeng Zeng and Tiger Wang of Medtronic China (Shanghai) for their help in coordinating and monitoring the study.

References

1

Udo
EO
,
Zuithoff
NP
,
van Hemel
NM
,
de Cock
CC
,
Hendriks
T
,
Doevedans
PA
et al. .
Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study
.
Heart Rhythm
2012
;
9
:
728
35
.

2

Gul
EE
,
Kayrak
M
.
Common pacemaker problems: lead and pocket complications
. In:
Das
MK
(ed).
Modern Pacemakers-Present and Future
.
Rijeka, Croatia
:
Intech
,
2011
. pp.
299
318
.

3

Ellenbogen
KA
,
Hellkamp
AS
,
Wilkoff
BL
,
Camunas
JL
,
Love
JC
,
Hadjis
TA
et al. .
Complications arising after implantation of DDD pacemakers: the MOST experience
.
Am J Cardiol
2003
;
92
:
740
9
.

4

Bracke
FA
,
Meijer
A
,
van Gelder
LM
.
Lead extraction for capsule related infections: a single-centre experience
.
Europace
2004
;
6
:
243
7
.

5

Wiegand
UK
,
Bode
F
,
Bonnemeier
H
,
Eberhard
F
,
Schlei
M
,
Peters
W
.
Long-term complication rates in ventricular, single lead VDD, and dual chamber pacing
.
Pacing Clin Electrophysiol
2003
;
26
:
1961
9
.

6

Connolly
SJ
,
Kerr
CR
,
Gent
M
,
Roberts
RS
,
Yusuf
S
,
Gillis
AM
,
Canadian Trial of Physiologic Pacing Investigators
.
Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes
.
N Engl J Med
2000
;
342
:
1385
91
.

7

Smith
HJ
,
Fearnot
NE
,
Byrd
CL
,
Wilkoff
BL
,
Love
CJ
,
Sellers
TD
.
Five-years experience with intravascular lead extraction
.
Pacing Clin Electrophysiol
1994
;
17
(Pt. II)
:
2016
20
.

8

Ferguson
TB
,
Cl
K
,
Crimmins-Reda
P
.
The additional hospital costs generated in the management of complications of pacemaker and defibrillator implantations
.
J Thorac Cardiovasc Surg
1996
;
111
:
742
52
.

9

Poole
JE
,
Gleva
MJ
,
Mela
T
,
Chung
MK
,
Uslan
DZ
,
Borge
R
,
Replace Registry Investigators
.
Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures, results from the REPLACE registry
.
Circulation
2010
;
122
:
1553
61
.

10

Kirkfeldt
RE
,
Johansen
JB
,
Nohr
EA
,
Jorgensen
OD
,
Nielsen
JC
.
Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark
.
Eur Heart J
2014
;
35
:
1186
94
.

11

Klug
D
,
Balde
M
,
Pavin
D
,
Hidden-Lucet
F
,
Clementy
J
,
Sadoul
N
et al. .
Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study
.
Circulation
2007
;
116
:
1349
55
.

12

Reddy
VY
,
Knops
RE
,
Sperzel
J
,
Miller
MA
,
Petru
J
,
Simon
J
et al. .
Permanent leadless cardiac pacing: results of the LEADLESS trial. CirculationAHA.113.006987 Published online before print March 24
, 2014.

13

Knops
RE
,
Tjong
FVY
,
Neuzil
P
,
Sperzel
J
,
Miller
MA
,
Petru
J
et al. .
Chronic performance of leadless cardiac pacing: one year follow-up to the LEADLESS trial
.
J Am Coll Cardiol
2015
;
65
:
1497
504
.

14

Steinwender
C
,
Hönig
S
,
Saleh
K
,
Lambert
T
,
Kammler
J
,
Kypta
A
.
A first-in-man experience with a minimally invasive transcatheter pacemaker. Proceeding of CARDIOSTIM/EHRA EUROPACE2014, Abstract 83/1
.

15

Sperzel
J
,
Burri
H
,
Gras
D
,
Tjong
FV
,
Knops
RE
,
Hindricks
G
et al. .
State of the art of leadless pacing
.
Europace
2015
;
17
:
1508
13
.

16

Ritter
P
,
Duray
GZ
,
Zhang
S
,
Narasimhan
C
,
Soejima
K
,
Omar
R
et al. .
The rationale and design of the Micra Transcatheter Pacing Study: safety and efficacy of a novel miniaturized pacemaker
.
Europace
2015
;
17
:
807
13
.

17

Benkemoun
H
,
Sacrez
J
,
Lagrange
P
,
Amiel
A
,
Prakash
A
,
Himmrich
E
et al. .
Optimizing pacemaker longevity with pacing mode and settings programming: results from a pacemaker multicenter registry
.
Pacing Clinical Electrophysiology
2012
;
35
:
403
8
.

18

Arias
E
.
United States Life Tables, 2006. CDC National Vital Statistics Reports 2010 58 (Number 21)
.

19

Bonner
MD
.
Extraction of the Micra transcatheter pacemaker system. Heart Rhythm 2014 proceedings
. .

20

Clay
S
,
Alfakih
K
,
Radjenovic
A
,
Jones
T
,
Ridgway
JP
,
Sinvananthan
MU
.
Normal range of human left ventricular volumes and mass using steady state free precession MRI in the radial long axis orientation
.
MAGMA
2006
;
19
:
41
5
.

21

Loupy
A
,
Messika-Zeitoun
D
,
Cachier
A
,
Himbert
D
,
Brochet
E
,
Lung
B
et al. .
An unusual cause of pacemaker-induced severe tricuspid regurgitation
.
Eur J Echocardiography
2008
;
9
:
201
3
.

22

Hirschl
DA
,
Jain
VR
,
Spindola-Franco
H
,
Gross
JN
,
Haramati
LB
.
Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT
.
Pacing Clin Electrophysiol
2007
;
30
:
28
32
.

Author notes

The first two authors contributed equally to this study.