Abstract

Background 

The SelectSecure lumenless 3830 pacing lead is often considered to be the pacing lead of choice for transvenous pacing in children because of its small diameter, lead strength, and reliable long-term sensing and pacing characteristics. One of the potential long-term pitfalls of a sturdy pacing lead is relative retraction with growth in children resulting in late lead dislodgement.

Case summary 

We report two cases of late SelectSecure 3830 lead dislodgement at 11.8 years (Case 1) and 8.8 years (Case 2), respectively, post the initial implantation. Case 1 was diagnosed with congenital complete heart block (CHB) at 9 months old when he presented with unconfirmed diphtheria infection. Case 2 was diagnosed with CHB at 14 weeks of age with positive maternal anti-Ro antibodies. Both patients underwent implantation of a transvenous permanent pacemaker implantation with Medtronic SelectSecure 3830 lead due to symptomatic bradycardia. Apart from a pulse generator change at 8.5 years (Case 1) and 7 years (Case 2), respectively, post-implant due to normal battery depletion, both patients are well in the interim.

Discussion 

As part of the pacemaker follow-up for rapidly growing children, we recommend more frequent surveillance of lead ‘tautness’ by chest radiography especially in children with CHB with no underlying heart rhythm.

Learning points

  • The 4.1 Fr SelectSecure 3830 pacing lead is a sturdy lead that has reliable long-term sensing and pacing characteristics.

  • One of the potential pitfalls of a sturdy pacing lead is relative retraction with linear growth in children leading to late lead dislodgement.

  • The combination of the creation of atrial loop and the use of slow absorbing ligatures to anchor the lead has been reported to improve the longevity of the pacing leads.

  • Regular chest radiograph and electronic performance check are essential to ensure optimum performance.

Introduction

With the advances in pacemaker generator and lead designs over the past three decades, there has been a gradual shift towards placement of transvenous pacing systems in young children with encouraging short and medium to long-term outcomes.1,2 However, the life-long pacing requirement in linearly growing paediatric patients presents ongoing unique challenges to lead selection and placement.

The SelectSecure 3830 lead (Medtronic Inc., Minneapolis, MN, USA) is often the pacing lead of choice for transvenous pacing in children because of its small diameter and reliable long-term sensing and pacing abilities, therefore, avoiding frequent lead revisions.1,3–5

This 4.1 French (Fr) bipolar lead is the thinnest lead available since 2000. It is lumenless with a cable cathode conductor and an anode conductor that is externally wound around in helical fashion. The conductors are separated by a layered insulation of ethylene–tetrafluoro–ethylene. The use of composite insultation materials allows this thin lead to have high crush resistance. The lead has active fixation with a steroid eluting non-retractable screw-in tip. The lead is easily delivered using the 8.4 Fr deflectable delivery catheter (SelectSite C304, Medtronic Inc., Minneapolis, MN, USA) or can be delivered through pre-shaped guiding catheters or even peel-apart introducer sheaths in small children and infants.6

One of the potential long-term pitfalls of this very sturdy pacing lead is retraction and lead dislodgement with linear growth in children. Several studies have reported complications related to other brands of leads dislodging in paediatric patients during early3–5 and medium-term1 follow-up. We report two cases of late lead dislodgement of 4.1 Fr SelectSecure 3830 in two patients with congenital complete heart block (CHB).

Timeline

Case 1

Case 2

Case presentation

Case 1

A 15-year-old boy was diagnosed with congenital CHB at 9 months of age when he was admitted with a suspected but unconfirmed diphtheria throat infection. During that admission, he was noted to be persistently bradycardic ranging from 60 to 80 b.p.m. with evidence of CHB on his electrocardiogram (ECG). He was otherwise healthy with no comorbidities. He underwent an implantation of transvenous VVIR pacemaker in 2004 when he was aged 3 years and 8 months as he became progressively bradycardic with resting ventricular rate below 45 b.p.m. He had a mild degree of lethargy but was otherwise well and his physical examination was normal with no evidence of heart failure.

The implantation procedure was performed under general anaesthesia. A sub-pectoral pocket was created. The left subclavian vein was punctured and a 59 cm SelectSecure 3830 lead was delivered to his right ventricular outflow tract (RVOT) septum using an 8.4 Fr deflectable sheath. The lead was tested for adequate sensing and pacing threshold. Of note, no atrial loop was created this time in order to potentially avoid loop migration (Figure 1A). The implanted lead was anchored in the sub-pectoral pocket with a non-absorbable suture and connected to the generator. The wound was closed with a subcuticular suture in a standard manner. The patient was reviewed in pacemaker clinic at 6 weeks post-initial implantation, then at 6 months interval within the 1st year of implantation and annually beyond 1st year.

Serial chest X-rays/fluoroscopy images in Case 1. (A) At initial implantation of pacemaker system with no right atrial loop created with the redundant lead. (B) At interval pulse generator change, the pacing lead was noted to be within the right ventricle but away from the right ventricular outflow tract septum. (C) At lead extraction and pacemaker system upgrade, pacing lead was noted to have withdrawn to right atrium.
Figure 1

Serial chest X-rays/fluoroscopy images in Case 1. (A) At initial implantation of pacemaker system with no right atrial loop created with the redundant lead. (B) At interval pulse generator change, the pacing lead was noted to be within the right ventricle but away from the right ventricular outflow tract septum. (C) At lead extraction and pacemaker system upgrade, pacing lead was noted to have withdrawn to right atrium.

He required a pulse generator change at 12 years of age due to normal battery depletion. During the box-change admission, the pacing lead was noted to be within the right ventricle but away from the RVOT septum on chest radiography (CXR) (Figure 1B). His ECG showed good ventricular pacing with appropriate ventricular capture (Figure 2A). No intervention was made to the lead as sensing and pacing thresholds were completely satisfactory at that time.

Serial electrocardiogram in Case 1. (A) Baseline electrocardiogram showing ventricular pacing spikes with appropriate ventricular capture. (B) Electrocardiogram at the presentation with pacemaker output failure showing pacing spikes without ventricular capture.
Figure 2

Serial electrocardiogram in Case 1. (A) Baseline electrocardiogram showing ventricular pacing spikes with appropriate ventricular capture. (B) Electrocardiogram at the presentation with pacemaker output failure showing pacing spikes without ventricular capture.

During a routine electrical performance check at 15 years and 5 months of age his pacemaker was noted to have loss of capture. The patient was asymptomatic with ventricular rate of 57 b.p.m., with normal blood pressure and no evidence of heart failure. The case was discussed at our Joint Cardiac Conference with agreement to proceed with lead extraction and pacemaker system upgrade. On admission, it was noted the pacing lead was withdrawn to right atrium (RA) on an echocardiogram and as shown on fluoroscopy (Figure 1C), and his ECG showed pacing spikes without ventricular capture (Figure 2B).

Lead extraction procedure

After unsuccessful simple traction approach, we used a Bulldog Lead Extender (Cook Medical) and One-Tie Compression Coil (Cook Medical) for lead control. An 11 Fr, and 13 Fr Evolution Mechanical Dilator Sheath (Cook Medical), was advanced over the lead to dissect fibrous and periosteal adhesions up to the innominate vein and superior vena cava junction while applying constant traction on the lead. We had to resort to the femoral approach using a needle-eye snare to extract a calcified Select Secure lead completely. Haemostasis was achieved with a purse-string suture applied to the entry site. During the same procedure, he had an uneventful upgrade to a transvenous dual-chamber pacemaker system via the left subclavian vein.

At his latest follow-up, 3 years after the implantation of the dual-chamber pacemaker system, the patient is clinically well with normal cardiovascular exam. His pacing check was satisfactory with atrial threshold of 0.75V @ 0.4 ms, right ventricular threshold of 1.0V @ 0.4 ms with 99% ventricular pacing and 26% atrial pacing.

Case 2

A 15-year-old girl who was born prematurely at 30 weeks of gestation and was noted postnatally to have 2:1 atrioventricular block. She was subsequently diagnosed with CHB at 14 weeks of age with positive maternal anti-Ro antibodies. She was well and thriving with no sign of heart failure at that stage with a baseline ventricular rate of 80 b.p.m. However, she became progressively more bradycardic with resting ventricular rate between 45 and 50 b.p.m. along with near syncopal episodes and subjectively reduced energy levels. Her cardiovascular examination was normal with no evidence of heart failure. She underwent a transvenous VVIR pacemaker insertion with a 59 cm Select Secure 3830 pacing lead implanted on the RVOT septum in 2007 at the age of 6 years and 9 months. The redundant lead was pushed into create a loop in the RA (Figure 3A) and the implanted lead anchored in the sub-pectoral pocket with a non-absorbable suture. The patient was reviewed in pacemaker clinic at 6 weeks post-initial implantation, then at 6 months of interval within the 1st year of implantation and annually beyond 1st year.

Serial chest X-rays/fluoroscopy images in Case 2. (A) At initial implantation of pacemaker system with a right atrial loop creation using the redundant lead. (B) At interval pulse generator change, note the tip of the lead at right ventricular outflow tract septum. (C) At pacemaker change, note the pacing lead has withdrawn to right atrium.
Figure 3

Serial chest X-rays/fluoroscopy images in Case 2. (A) At initial implantation of pacemaker system with a right atrial loop creation using the redundant lead. (B) At interval pulse generator change, note the tip of the lead at right ventricular outflow tract septum. (C) At pacemaker change, note the pacing lead has withdrawn to right atrium.

She underwent a pulse generator change at 14 years of age due to expected battery depletion. Pacing check showed stable low lead capture threshold. Fluoroscopy image showed the position of the tip of the lead at RVOT septum; however, the extent of the RA loop was not imaged at this time (Figure 3B).

During a routine pacemaker check at 15 years and 6 months of age her pacemaker was noted to have lost ventricular capture. Her ventricular rate was 49 b.p.m. and the patient was asymptomatic. Fluoroscopy showed that the lead had dislodged and retracted to the RA (Figure 3C). The ECG showed that the retracted ventricular lead appears to capture the atrium (Figure 4). A month later, she opted to have implantation of a Medtronic Micra™ leadless pacemaker so that she could engage in physical contact sports once the old box was explanted. The SelectSecure 3830 lead was left in situ after careful consideration of the risks vs. benefits of removing a near decade-old lead. After 2 years of follow-up, she remains well with a ventricular threshold of 0.63V @ 0.4 ms with 23% ventricular pacing and she is regularly involved in contact sports.

Electrocardiogram for Case 2 at the time of lead dislodgement showing the pacing spike appears to capture the atrium.
Figure 4

Electrocardiogram for Case 2 at the time of lead dislodgement showing the pacing spike appears to capture the atrium.

Discussion

About 1% of all pacemakers are implanted into children.7,8 The epicardial pacemaker implantation is indicated in the presence of complex congenital heart defects with intracardiac shunts or absence of appropriate cardiac cavity, in small babies where the veins are too small to accommodate the passing of a pacing lead or in children with venous occlusion. Epicardial leads are traditionally associated with higher lead failure, shorter battery longevity, and rare but serious cardiac strangulation and coronary compression.8–12 Traditional transvenous leads, on the other hand, carry a significant risk of venous thrombosis.13–15

The transvenous approach to pacemaker implantation in children and adolescents is an appealing alternative to the epicardial approach and has generally favourable outcomes.

Newer refined transvenous leads such as the bipolar, screw-in Medtronic SelectSecure 3830 with a smaller diameter may result in fewer long-term venous complications.1–4,16 However, children are more prone to mechanical complications because of their active lifestyle, higher frequency of traumatic contacts, and somatic growth as shown in our cases.

Lead redundancy or loops were created where possible to allow for patient linear growth at the discretion of the operator. No atrial loop was created in Case 1 at the initial implantation as the loop migrated to the RVOT and required retraction. There was an atrial loop created in Case 2; however, a fluoroscopy image showing the tip of the lead in desired position did not capture the atrial loop at subsequent pulse generator change.

The combination of the creation of an atrial loop to the ventricular lead and the use of slow absorbing ligatures to anchor the lead has been reported to improve the longevity of the pacing leads.2,12,17,18

However, the length of redundant lead introduced by creating a loop may still not be sufficient for linear growth in some patients as our cases show.19 Furthermore, large atrial loops in addition to migrating into the right ventricle can also become adherent to the atrial walls. There have also been mixed results with late lead advancement to recreate atrial loop.1,20 This may well be because of the lead being bound down by fibrous tissues intravascularly after many years.

Gasparini et al21 suggested to leave redundant lead loop within the inferior vena cava (IVC) to allow for further growth by shortening excess loop. However, the IVC loop has been associated with IVC thrombosis and obstruction,22 which may lead to Budd–Chiari syndrome hence we avoid that approach at our centre.

Conclusion

Our two case studies emphasize that over the longer term the Medtronic SelectSecure 3830 lead is subject to the usual problems related to transvenous leads. Long-term follow-up for patients with this lead should include pacemaker checks at regular intervals to ensure optimum pacemaker and pacing lead performance, and CXR every 2 years to pre-empt the problem of lead retraction and sudden loss of pacing. This matters most especially in children with CHB with no underlying heart rhythm. Further studies on the long-term outcome of SelectSecure 3830 leads in growing children are essential.

Lead author biography

Li Yen Ng graduated with an honoured medical degree from the Royal College of Surgeon in Ireland. I am currently a 3rd year specialist registrar training in the Paediatric Cardiology with the Royal College of Physician in Ireland. My area of interests in Paediatric Cardiology includes advanced cardiac imaging, cardio-genetics, and electrophysiology.

Supplementary material

Supplementary material is available at European Heart Journal - Case Reports online.

Slide sets: A fully edited slide set detailing this case and suitable for local presentation is available online as Supplementary data.

Consent: The authors confirm that written consent for submission and publication of this case report, including images and associated text, has been obtained from the patients in line with COPE guidance.

Conflict of interest: none declared.

Funding: none declared.

References

1

De Filippo
P
,
Giofre
F
,
Leidi
C
,
Senni
M
,
Ferrari
P.
Transvenous pacing in pediatric patients with bipolar lumenless lead: ten-year clinical experience
.
Int J Cardiol
2018
;
255
:
45
49
.

2

Konta
L
,
Chubb
MH
,
Bostock
J
,
Rogers
J
,
Rosenthal
E.
Twenty-seven years experience with transvenous pacemaker implantation in children weighing <10 kg
.
Circ Arrhythm Electrophysiol
2016
;
9
:
e003422
.

3

Chakrabarti
S
,
Morgan
GJ
,
Kenny
D
,
Walsh
KP
,
Oslizlok
P
,
Martin
RP
et al.
Initial experience of pacing with a lumenless lead system in patients with congenital heart disease
.
Pacing Clin Electrophysiol
2009
;
32
:
1428
1433
.

4

Khan
A
,
Zelin
K
,
Karpawich
PP.
Performance of the lumenless 4.1-Fr diameter pacing lead implanted at alternative pacing sites in congenital heart: a chronic 5-year comparison
.
Pacing Clin Electrophysiol
2010
;
33
:
1467
1474
.

5

Gammage
MD
,
Lieberman
RA
,
Yee
R
,
Manolis
AS
,
Compton
SJ
,
Khazen
C
,
for the Worldwide SelectSecure Clinical Investigators
et al.
Multi-center clinical experience with a lumenless, catheter-delivered, bipolar, permanent pacemaker lead: implant safety and electrical performance
.
Pacing Clin Electro
2006
;
29
:
858
865
.

6

Kenny
D
,
Walsh
KP.
Noncatheter-based delivery of a single-chamber lumenless pacing lead in small children
.
Pacing Clin Electro
2007
;
30
:
834
838
.

7

Bink-Boelkens MTHE
.
Cardiac pacing in infants and children
.
Neth J Cardiol
1992
;
5
:
199
202
.

8

Kerstjens-Frederikse
MW
,
Bink-Boelkens
MT
,
de Jongste
MJ
,
Homan van der Heide
JN.
Permanent cardiac pacing in children: morbidity and efficacy of follow-up
.
Int J Cardiol
1991
;
33
:
207
214
.

9

Eyskens
B
,
Mertens
L
,
Moerman
P
,
Ector
H
,
Daenen
W
,
Gewilling
M.
Cardiac strangulation, a rare complication of epicardial pacemaker leads during growth
.
Heart
1997
;
77
:
288
289
.

10

Mah
DY
,
Prakash
A
,
Porras
D
,
Fynn-Thompson
F
,
DeWitt
ES
,
Banka
P.
Coronary artery compression from epicardial leads: More common than we think
.
Heart Rhythm
2018
;
15
:
1439
1447
.

11

Kubus
P
,
Materna
O
,
Gebauer
RA
,
Matejka
T
,
Gebauer
R
,
Tlaskal
T
et al.
Permanent epicardial pacing in children: long-term results and factors modifying outcome
.
Europace
2012
;
14
:
509
514
.

12

Silvetti
MS
,
Drago
F
,
Grutter
G
,
De Santis
A
,
Di Ciommo
V
,
Ravà
L.
Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients
.
Europace
2006
;
8
:
530
536
.

13

Bar-Cohen
Y
,
Berul
CI
,
Alexander
ME
,
Fortescue
EB
,
Walsh
EP
,
Triedman
JK
et al.
Age, size, and lead factors alone do not predict venous obstruction in children and young adults with transvenous lead systems
.
J Cardiovasc Electrophysiol
2006
;
17
:
754
759
.

14

Sanjeev
S
,
Karpawich
PP.
Superior vena cava and innominate vein dimensions in growing children: an aid for interventional devices and transvenous leads
.
Pediatr Cardiol
2006
;
27
:
414
419
.

15

Figa
FH
,
McCrindle
BW
,
Bigras
JL
,
Hamilton
RM
,
Gow
RM.
Risk factors for venous obstruction in children with transvenous pacing leads
.
Pacing Clin Electro
1997
;
20
:
1902
1909
.

16

Campbell
RM
,
Raviele
AA
,
Hulse
EJ
,
Auld
DO
,
McRae
GJ
,
Tam
VK
et al.
Experience with a low profile bipolar, active fixation pacing lead in pediatric patients
.
Pacing Clin Electro
1999
;
22
:
1152
1157
.

17

Stojanov
P
,
Velimirovic
D
,
Hrnjak
V
,
Pavlovic
SU
,
Zivkovic
M
,
Djordjevic
Z.
Absorbable suture technique: solution to the growth problem in pediatric pacing with endocardial leads
.
Pacing Clin Electro
1998
;
21
:
65
68
.

18

Gabbarini
F
,
Golzio
PG
,
Agnoletti
G.
Selective site pacing in paediatric patients: how to recreate a new loop in the right atrium to follow body growth
.
Europace
2011
;
13
:
1055
1057
.

19

Gheissari
A
,
Hordof
AJ
,
Spotnitz
HM.
Transvenous pacemakers in children: relation of lead length to anticipated growth
.
Ann Thorac Surg
1991
;
52
:
118
121
.

20

Rudokas
MWP
,
Bowman
MD
,
Van Hare
TM
,
Silva
GF
,
Jennifer
NA.
Lead advancement in pediatric patients with select secure 3830 pacing leads
.
J Cardiol Curr Res
2014
;
1
:
1
.

21

Gasparini
M
,
Mantica
M
,
Galimberti
P
,
Coltorti
F
,
Ceriotti
C
,
Priori
SG.
Inferior vena cava loop of the implantable cardioverter defibrillator endocardial lead: a possible solution of the growth problem in pediatric implantation
.
Pacing Clin Electro
2000
;
23
:
2108
2112
.

22

Chandra
S
,
Ameta
D
,
Vijay
SK
,
Saran
RK.
Pacemaker lead induced inferior vena caval thrombosis leading to portal hypertension
.
Indian Pacing Electrophysiol J
2013
;
13
:
84
87
.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact [email protected]
Editor: Christoph Sinning,
Christoph Sinning
Editor
Search for other works by this author on:
Inga Voges,
Inga Voges
Editor
Search for other works by this author on:
Konstantinos Iliodromitis,
Konstantinos Iliodromitis
Editor
Search for other works by this author on:
Linh Ngo,
Linh Ngo
Editor
Search for other works by this author on:
Mariame Chakir
Mariame Chakir
Editor
Search for other works by this author on:

Supplementary data

Comments

0 Comments
Submit a comment
You have entered an invalid code
Thank you for submitting a comment on this article. Your comment will be reviewed and published at the journal's discretion. Please check for further notifications by email.