-
PDF
- Split View
-
Views
-
Cite
Cite
Kyu Kim, Dae-Young Kim, Jiwon Seo, Iksung Cho, Chi Young Shim, Geu-Ru Hong, Jong-Won Ha, The ratio of measured and reference effective orifice areas for discriminating prosthetic aortic valve obstruction, European Heart Journal - Cardiovascular Imaging, Volume 24, Issue 2, February 2023, Pages 232–240, https://doi.org/10.1093/ehjci/jeac206
- Share Icon Share
Abstract
We aimed to evaluate the efficacy of the measured effective orifice area (EOA)/reference EOA ratio in discriminating mechanical prosthetic aortic valve (PAV) obstruction.
This is a retrospective study of 193 mechanical PAV patients with an elevated mean transprosthetic pressure gradient (PG) over 20 mmHg or peak velocity over 3 m/s. Of those, 143 patients were objectively proven PAV obstruction with cardiac computed tomography or surgical inspection. The EOA was measured using the continuity equation, and the reference EOA values were obtained from previous guidelines. The measured/reference EOA ratio was significantly lower in the obstruction group (0.63 ± 0.18 vs. 0.86 ± 0.17; P < 0.001). The EOA ratio added incremental value for discriminating obstruction from the conventional parameters recommended in the guidelines. Receiver operating characteristic curve analysis revealed that the measured/reference EOA ratio discriminated PAV obstruction from those without obstruction [area under the curve (AUC), 0.840; 95% confidence interval, 0.783–0.898; P < 0.001]. A cutoff of 0.71 had 73.4% sensitivity and 82.0% specificity. The novel diagnostic algorithm adding the EOA ratio had similar accuracy to previous guideline algorithms, including reference EOA, and conventional Doppler parameters (AUC, 0.763 vs. 0.731; P = 0.309). In patients with a large PAV (≥23 mm), the novel algorithm had higher accuracy than the previous algorithm (AUC, 0.788 vs. 0.642; P = 0.019).
The ratio of measured/reference EOA adds incremental value over conventional Doppler parameters and might be helpful for distinguishing PAV obstruction.

Prosthetic aortic valves, without or with obstruction (A), novel algorithm for diagnosing pannus (B), and receiver operating characteristic curves of novel and previous guideline algorithm for detecting pannus (C). *Reference effective orifice area, †measured effective orifice area, ‡pannus formation.
Introduction
Valvular heart disease affects more than 100 million people worldwide with an increasing burden.1 Despite improved treatment strategies, valve replacement does not provide a definitive cure and prosthetic valve diseases persist.2 Among these, prosthetic aortic valve (PAV) obstruction derived from a pannus or thrombus is a life-threatening complication (Graphical abstract A).3 Doppler transthoracic echocardiography (TTE) is the standard method for assessing prosthetic valve function because it is non-invasive, radiation-free, and cost-effective.4 Several TTE parameters such as Doppler velocity index (DVI), acceleration time (AT), ejection time (ET) ratio, transprosthetic jet contour, and effective orifice area (EOA) are used in PAV assessment.5–8 However, the presence of PAV induces acoustic shadowing and reverberation artefacts, leading to suboptimal examination for PAV function. Because mechanical valves have more severe artefacts, evaluating PAV obstruction is challenging.7 In addition, a high transprosthetic pressure gradient (PG) or velocity alone is not proof of PAV obstruction. This phenomenon may be secondary to prosthesis–patient mismatch (PPM), a high-flow condition, PAV regurgitation, or localized high jet velocity.4 Previous guidelines recommend the additional use of computed tomography (CT), transesophageal echocardiography (TEE), or fluoroscopy.9 However, cardiac CT is associated with iodine contrast and radiation exposure, and TEE is semi-invasive. In particular, when prosthetic valve dysfunction is caused by pannus, the diagnosis is challenging using echocardiography.10 Recent recommendations from the European Association of Cardiovascular Imaging have described the use of reference EOA values, but clinical field validation is lacking.11 This study aimed to evaluate the efficacy of reference EOA values for discriminating PAV obstruction and construct the novel diagnostic algorithm using reference EOA values.
Methods
Study population
This study was a retrospective study from single tertiary hospital. A total of 996 patients with mechanical PAV and elevated transprosthetic velocity (>3.0 m/s) or mean systolic PG (≥20 mmHg) were identified. Patients with bioprosthetic AV were not included in this study. Among them, 223 underwent CT or cardiac surgery within 3 months. Patients with active endocarditis involving PAV, without reference EOA values or PAV information, with poor image quality for calculating Doppler parameters were excluded. After exclusion, 193 patients with PAV were analyzed; of them, 143 were diagnosed with PAV obstruction and 50 were diagnosed have intact PAV function, classified as no PAV obstruction group (Figure 1). Diagnosis for PAV obstruction was made by surgical inspection or cardiac CT findings. In case of discordance between surgical inspection and CT, we considered surgical inspection first. CT scan and referring patients for surgery were performed at discretion of the responsible physician. Patient data were obtained from medical records. The reference EOA values for each aortic valve (AV) prosthesis were obtained from previous guidelines and studies (see Supplementary data online, Table S1).2,9,11,12 Then, projected indexed EOA values were calculated as the reference EOA from the guidelines/body surface area (BSA), which was described in recent guideline from European Association of Cardiovascular Imaging.11 Patients with a projected indexed EOA ≤ 0.65 [≤0.55 in body mass index (BMI) ≥ 30 kg/m2] were defined as having severe PPM and 0.66–0.85 (0.56–0.70 in BMI ≥ 30 kg/m2) as moderate PPM. We performed sensitivity analysis to test if our findings were robust. First, we analyzed patients diagnosed with PAV obstruction based only on surgical findings. Second, we defined the PAV obstruction as in case of limited motion of the PAV leaflet by CT or surgical inspection. The Institutional Review Board of Severance Hospital approved this study, which was conducted in compliance with the Declaration of Helsinki. The requirement for informed consent was waived.

Study flow. MPG; mean pressure gradient, PAV; prosthetic aortic valve.
Echocardiography
All echocardiograms were analyzed in a blinded fashion by trained cardiologists. According to current guidelines,11,13 the stroke volume was calculated using the following formula: annulus diameter2 × 0.785 × velocity-time integral of LVOT (VTILVOT). The transprosthetic peak velocity, peak, and mean transprosthetic PG were measured using continuous-wave Doppler VTI. The EOA was measured as stroke volume/Velocity-time integral of aortic valve (VTIAV) by continuity equation. The ratio of measured/reference EOA was calculated using mean reference EOA value of each prosthetic AV from previous guidelines.9,11 The transprosthetic jet contour was qualitatively estimated and classified into 3 categories (triangular and early peaking, intermediate, rounded, and symmetric).9,11 The DVI was calculated as VTILVOT/VTIAV. AT was defined as the time interval from the beginning of systolic flow to the peak velocity. ET was measured as the time from onset to the end of systolic flow across the PAV.7 All measurements represent an average of three cardiac cycles for patients with sinus rhythm and an average of five cardiac cycles for patients with atrial fibrillation.11 Intraclass correlation coefficients (ICCs) were calculated to assess the inter- and intra-observer variability of the measured/reference EOA ratio and the difference between the measured LVOT diameter and reference annulus diameters of the PAV. A total of 20 samples (included 10 without PAV obstructions, and 10 PAV obstructions) were assessed by the same observer on different occasions in a random order and by another observer. We then constructed novel hierarchical diagnostic algorithm models for PAV obstruction by pannus and compared the novel algorithm with previously recommended algorithms (see Supplementary data online, Figure S1).9,11 Fluoroscopic data and EOA changes during follow-up, which are described in 2016 recommendation, were not included for the purpose of this study to evaluate incremental value of single echocardiographic parameter, measured/reference EOA ratio. We performed a subgroup analysis of patients expected to have difficulties in the accurate assessment of PAV function: large-size PAV (≥23 mm), atrial fibrillation, double valve replacement (DVR), and lower transprosthetic PG (<40 mmHg).
Cardiac CT
All CT scans were performed with a dual-source CT scanner (SOMATOM Definition Flash; Siemens Healthcare, Forchheim, Germany). Scans were performed in retrospective ECG-gated data acquisition mode using the triple-phase injection method (70 mL of iopamidol followed by 30 mL of 30% blended iopamidol with saline and 20 mL of saline at 5 mL/s). Images were generated using iterative reconstruction (sinogram-affirmed iterative reconstruction). Image reconstruction was performed with a medium kernel (I36f), and the reconstruction slice thickness was 0.75 mm with 0.5-mm increments. For all patients, the datasets of transverse plane image were reconstructed every 10% of the cardiac cycle. Image analysis was performed using 3D software (Aquarius iNtuition, Ver. 4.4.11, TeraRecon, San Mateo, CA, USA). The presence of subprosthetic soft tissue mass with low-attenuation (pannus) was assessed.14 When CT attenuation of a subvalvular mass was lower than that of the interventricular septum, thrombus was favoured over pannus. Limited motion of a mechanical PAV was considered present when motion of a leaflet or leaflets was restricted, with an opening angle which was reduced more than 4° compared to the manufacturer’s value.15 All CT analyses were independently performed by radiologists blinded to clinical information and echocardiographic results.
Statistical analysis
Continuous variables are presented as mean ± standard deviation (SD) and compared using Student’s t-test. If the variables were not normally distributed, they were presented as median (interquartile range) and compared using Wilcoxon rank-sum test. Categorical variables are presented as number (percentage and compared using the χ2 test. The incremental value of the measured/reference EOA ratio over conventional Doppler parameters for discriminating PAV obstruction was assessed using logistic regression models. The R-squared and global χ2 scores for each model were calculated. Prognostic values were compared using the likelihood-ratio test. A receiver operating characteristic (ROC) curve was plotted to determine the accuracy of each Doppler parameter and the best cutoff values for identifying PAV obstruction. We also developed a hierarchical algorithm model using the reference EOA value and compared it with previous diagnostic algorithm models.9,11 The area under the ROC curve (AUC) was calculated for each algorithm. The accuracy of the diagnostic algorithm was compared using DeLong’s test.16 All statistical analyses were performed with R version 4.1.0 (The R Foundation for Statistical Computing; www.R-project.org).
Results
Baseline characteristics and echocardiographic parameters
The patients’ baseline characteristics are presented in Table 1. The details of the PAVs are presented in Supplementary data online, Table S2. The patients with PAV obstruction were older than those of no PAV obstruction group [median age: 65.0 (54.0–69.0) vs. 58.0 (52.0–67.0) years; P = 0.036], and female was more common (76.2 vs. 40.0%; P < 0.001). Blood pressure, heart rate, BMI, heart rhythm, and the proportion of projected PPM did not differ significantly between the groups. The method of diagnosis for PAV obstruction was mainly based on surgical inspection in PAV obstruction group (65.7 vs. 26.0%), but CT in no PAV obstruction group (34.3 vs. 74.0%, P < 0.001). Both CT and surgical inspection were performed in 91 (47.2%) of 213 patients. Three of 91 patients showed discordant findings; CT was negative, but pannus was detected in surgical inspection. There was one case of PAV thrombi and it was combined with pannus. In no PAV obstruction group, 13 of 50 (26.0%) patients underwent surgery. Indications of surgery in patients without PAV obstruction were paravalvular regurgitation in 2, and severe dysfunction of mitral or tricuspid valves in 11. In PAV obstruction group, 94 of 143 (65.7%) patients underwent cardiac surgery whereas 49 (34.3%) patients did not.
. | No PAV obstruction (n = 50) . | PAV obstruction (n = 143) . | P value . |
---|---|---|---|
Age, years | 58.0 (52.0–67.0) | 65.0 (54.0–69.0) | 0.036 |
Sex | <0.001 | ||
ȃMale | 30 (60.0%) | 34 (23.8%) | |
ȃFemale | 20 (40.0%) | 109 (76.2%) | |
Height, cm | 165.1 ± 8.8 | 158.5 ± 8.2 | <0.001 |
Weight, kg | 59.5 (54.0–72.0) | 55.0 (50.0–62.5) | 0.002 |
BSA, m2 | 1.67 (1.54–1.84) | 1.55 (1.46–1.68) | <0.001 |
BMI, kg/m2 | 22.5 (20.4–25.7) | 22.3 (20.3–24.4) | 0.503 |
ȃ ≥ 30 | 2 (4.0%) | 4 (2.8%) | >0.999 |
ȃ < 30 | 48 (96.0%) | 139 (97.2%) | >0.999 |
Systolic blood pressure, mmHg | 121.0 ± 17.8 | 121.3 ± 16.1 | 0.926 |
Diastolic blood pressure, mmHg | 70.9 ± 9.2 | 68.7 ± 10.5 | 0.193 |
Heart rate, beats per minute | 68.0 (60.0–80.0) | 67.0 (60.0–76.0) | 0.326 |
Rhythm | 0.178 | ||
ȃSinus rhythm | 20 (40.0%) | 42 (29.4%) | |
ȃAtrial fibrillation | 27 (54.0%) | 97 (67.8%) | |
ȃPaced | 3 (6.0%) | 4 (2.8%) | |
Prosthetic valve size | 21.0 (19.0–23.0) | 21.0 (19.0–21.0) | 0.032 |
ȃRange | 19–25 | 19–25 | |
ȃLarge size (≥23 mm) | 18 (36.0%) | 16 (11.2%) | <0.001 |
Prosthetic valve type | |||
ȃBileaflet | 49 (98.0%) | 137 (95.8) | 0.783 |
ȃMonoleaflet | 1 (2.0%) | 6 (4.2%) | 0.783 |
ȃSupra-annular type | 6 (12.0%) | 12 (8.4%) | 0.636 |
Time interval from previous AVR, years | 11.9 ± 9.2 | 19.3 ± 7.8 | <0.001 |
Reference EOA | 1.50 (1.30–1.70) | 1.40 (1.20–1.50) | 0.005 |
Projected indexed EOA, cm2/m2 | 0.90 (0.75–0.98) | 0.91 (0.76–0.99) | 0.991 |
ȃModerate patient–prosthesis mismatch | 12 (24.0%) | 36 (25.2%) | >0.999 |
ȃSevere patient–prosthesis mismatch | 4 (8.0%) | 14 (9.8%) | 0.927 |
DVR status | 33 (66.0%) | 119 (83.2%) | 0.018 |
Limited motion of PAV leaflet | 0 (0%) | 84 (58.7%) | <0.001 |
Concomitant aortic valve regurgitation | 0.091 | ||
ȃNo, mild | 45 (90.0%) | 139 (97.2%) | |
ȃModerate or severe | 5 (10.0%) | 4 (2.8%) | |
Method of diagnosis | <0.001 | ||
ȃSurgical inspection | 13 (26.0%) | 94 (65.7%) | |
ȃCardiac CT | 37 (74.0%) | 49 (34.3%) | |
Indication for reoperation | <0.001 | ||
ȃPAV obstruction | 0 (0%) | 94 (65.7%) | |
ȃPAV paravalvular regurgitation | 2 (4.0%) | 0 (0%) | |
ȃMitral or tricuspid valvular dysfunction | 11 (22.0%) | 29 (20.3%) | |
Redo aortic valve replacement | 2 (4.0%) | 94 (65.7%) | <0.001 |
Echocardiography | |||
ȃEOA, cm2 | 1.26 (1.00–1.50) | 0.80 (0.68–0.99) | <0.001 |
ȃMeasured/reference EOA ratio | 0.86 ± 0.17 | 0.63 ± 0.18 | <0.001 |
ȃEOA/BSA, cm2/m2 | 0.76 (0.64–0.90) | 0.51 (0.43–0.64) | <0.001 |
ȃLVEF, % | 62.0 (57.0–68.0) | 66.0 (61.0–71.0) | 0.009 |
ȃȃ < 45% | 5 (10.0%) | 9 (6.3%) | 0.580 |
ȃLVOT diameter, mm | 21.7 (19.9–23.3) | 20.5 (19.0–21.2) | <0.001 |
ȃLVOT diameter difference from valve size | 0.45 (−0.30–1.00) | −0.10 (−1.25–0.20) | <0.001 |
ȃLVOT VTI, cm | 22.7 ± 4.8 | 22.7 ± 4.6 | 0.994 |
ȃPeak LVOT velocity, m/s | 1.09 (0.97–1.29) | 1.08 (0.93–1.21) | 0.296 |
ȃStroke volume, mL | 83.2 (72.0–103.1) | 69.2 (60.7–83.4) | <0.001 |
ȃStroke volume index, mL/m2 | 49.4 (41.4–60.1) | 44.7 (38.6–51.7) | 0.019 |
ȃDynamic LVOT obstruction | 0 (0%) | 8 (5.6%) | 0.195 |
ȃAV VTI, cm | 66.1 (57.6–75.3) | 88.0 (72.1–101.3) | <0.001 |
ȃPeak transprosthetic velocity, m/s | 3.33 (3.09–3.67) | 4.10 (3.63–4.71) | <0.001 |
ȃPeak transprosthetic pressure, mmHg | 44.5 (38.0–54.0) | 67.0 (52.0–88.2) | <0.001 |
ȃMean transprosthetic pressure, mmHg | 25.5 (21.0–30.0) | 37.4 (29.3–50.5) | <0.001 |
ȃDoppler velocity index | 0.34 (0.29–0.41) | 0.26 (0.23–0.30) | <0.001 |
ȃAcceleration time of aortic velocity, msec | 83.4 ± 22.5 | 98.9 ± 23.0 | <0.001 |
ȃEjection time, msec | 282.8 ± 44.2 | 293.5 ± 35.1 | 0.128 |
ȃAcceleration time/ejection time ratio | 0.30 ± 0.07 | 0.34 ± 0.06 | <0.001 |
ȃCW Doppler jet contour | <0.001 | ||
ȃȃTriangle, early peaking | 11 (22.0%) | 17 (11.9%) | |
ȃȃIntermediate | 38 (76.0%) | 78 (54.5%) | |
ȃȃRound, symmetrical | 1 (2.0%) | 48 (33.6%) |
. | No PAV obstruction (n = 50) . | PAV obstruction (n = 143) . | P value . |
---|---|---|---|
Age, years | 58.0 (52.0–67.0) | 65.0 (54.0–69.0) | 0.036 |
Sex | <0.001 | ||
ȃMale | 30 (60.0%) | 34 (23.8%) | |
ȃFemale | 20 (40.0%) | 109 (76.2%) | |
Height, cm | 165.1 ± 8.8 | 158.5 ± 8.2 | <0.001 |
Weight, kg | 59.5 (54.0–72.0) | 55.0 (50.0–62.5) | 0.002 |
BSA, m2 | 1.67 (1.54–1.84) | 1.55 (1.46–1.68) | <0.001 |
BMI, kg/m2 | 22.5 (20.4–25.7) | 22.3 (20.3–24.4) | 0.503 |
ȃ ≥ 30 | 2 (4.0%) | 4 (2.8%) | >0.999 |
ȃ < 30 | 48 (96.0%) | 139 (97.2%) | >0.999 |
Systolic blood pressure, mmHg | 121.0 ± 17.8 | 121.3 ± 16.1 | 0.926 |
Diastolic blood pressure, mmHg | 70.9 ± 9.2 | 68.7 ± 10.5 | 0.193 |
Heart rate, beats per minute | 68.0 (60.0–80.0) | 67.0 (60.0–76.0) | 0.326 |
Rhythm | 0.178 | ||
ȃSinus rhythm | 20 (40.0%) | 42 (29.4%) | |
ȃAtrial fibrillation | 27 (54.0%) | 97 (67.8%) | |
ȃPaced | 3 (6.0%) | 4 (2.8%) | |
Prosthetic valve size | 21.0 (19.0–23.0) | 21.0 (19.0–21.0) | 0.032 |
ȃRange | 19–25 | 19–25 | |
ȃLarge size (≥23 mm) | 18 (36.0%) | 16 (11.2%) | <0.001 |
Prosthetic valve type | |||
ȃBileaflet | 49 (98.0%) | 137 (95.8) | 0.783 |
ȃMonoleaflet | 1 (2.0%) | 6 (4.2%) | 0.783 |
ȃSupra-annular type | 6 (12.0%) | 12 (8.4%) | 0.636 |
Time interval from previous AVR, years | 11.9 ± 9.2 | 19.3 ± 7.8 | <0.001 |
Reference EOA | 1.50 (1.30–1.70) | 1.40 (1.20–1.50) | 0.005 |
Projected indexed EOA, cm2/m2 | 0.90 (0.75–0.98) | 0.91 (0.76–0.99) | 0.991 |
ȃModerate patient–prosthesis mismatch | 12 (24.0%) | 36 (25.2%) | >0.999 |
ȃSevere patient–prosthesis mismatch | 4 (8.0%) | 14 (9.8%) | 0.927 |
DVR status | 33 (66.0%) | 119 (83.2%) | 0.018 |
Limited motion of PAV leaflet | 0 (0%) | 84 (58.7%) | <0.001 |
Concomitant aortic valve regurgitation | 0.091 | ||
ȃNo, mild | 45 (90.0%) | 139 (97.2%) | |
ȃModerate or severe | 5 (10.0%) | 4 (2.8%) | |
Method of diagnosis | <0.001 | ||
ȃSurgical inspection | 13 (26.0%) | 94 (65.7%) | |
ȃCardiac CT | 37 (74.0%) | 49 (34.3%) | |
Indication for reoperation | <0.001 | ||
ȃPAV obstruction | 0 (0%) | 94 (65.7%) | |
ȃPAV paravalvular regurgitation | 2 (4.0%) | 0 (0%) | |
ȃMitral or tricuspid valvular dysfunction | 11 (22.0%) | 29 (20.3%) | |
Redo aortic valve replacement | 2 (4.0%) | 94 (65.7%) | <0.001 |
Echocardiography | |||
ȃEOA, cm2 | 1.26 (1.00–1.50) | 0.80 (0.68–0.99) | <0.001 |
ȃMeasured/reference EOA ratio | 0.86 ± 0.17 | 0.63 ± 0.18 | <0.001 |
ȃEOA/BSA, cm2/m2 | 0.76 (0.64–0.90) | 0.51 (0.43–0.64) | <0.001 |
ȃLVEF, % | 62.0 (57.0–68.0) | 66.0 (61.0–71.0) | 0.009 |
ȃȃ < 45% | 5 (10.0%) | 9 (6.3%) | 0.580 |
ȃLVOT diameter, mm | 21.7 (19.9–23.3) | 20.5 (19.0–21.2) | <0.001 |
ȃLVOT diameter difference from valve size | 0.45 (−0.30–1.00) | −0.10 (−1.25–0.20) | <0.001 |
ȃLVOT VTI, cm | 22.7 ± 4.8 | 22.7 ± 4.6 | 0.994 |
ȃPeak LVOT velocity, m/s | 1.09 (0.97–1.29) | 1.08 (0.93–1.21) | 0.296 |
ȃStroke volume, mL | 83.2 (72.0–103.1) | 69.2 (60.7–83.4) | <0.001 |
ȃStroke volume index, mL/m2 | 49.4 (41.4–60.1) | 44.7 (38.6–51.7) | 0.019 |
ȃDynamic LVOT obstruction | 0 (0%) | 8 (5.6%) | 0.195 |
ȃAV VTI, cm | 66.1 (57.6–75.3) | 88.0 (72.1–101.3) | <0.001 |
ȃPeak transprosthetic velocity, m/s | 3.33 (3.09–3.67) | 4.10 (3.63–4.71) | <0.001 |
ȃPeak transprosthetic pressure, mmHg | 44.5 (38.0–54.0) | 67.0 (52.0–88.2) | <0.001 |
ȃMean transprosthetic pressure, mmHg | 25.5 (21.0–30.0) | 37.4 (29.3–50.5) | <0.001 |
ȃDoppler velocity index | 0.34 (0.29–0.41) | 0.26 (0.23–0.30) | <0.001 |
ȃAcceleration time of aortic velocity, msec | 83.4 ± 22.5 | 98.9 ± 23.0 | <0.001 |
ȃEjection time, msec | 282.8 ± 44.2 | 293.5 ± 35.1 | 0.128 |
ȃAcceleration time/ejection time ratio | 0.30 ± 0.07 | 0.34 ± 0.06 | <0.001 |
ȃCW Doppler jet contour | <0.001 | ||
ȃȃTriangle, early peaking | 11 (22.0%) | 17 (11.9%) | |
ȃȃIntermediate | 38 (76.0%) | 78 (54.5%) | |
ȃȃRound, symmetrical | 1 (2.0%) | 48 (33.6%) |
AVR, aortic valve replacement; CW, continuous wave; LVEF, left ventricular ejection fraction; LVOT, left ventricular outflow tract; VTI, velocity-time integral.
. | No PAV obstruction (n = 50) . | PAV obstruction (n = 143) . | P value . |
---|---|---|---|
Age, years | 58.0 (52.0–67.0) | 65.0 (54.0–69.0) | 0.036 |
Sex | <0.001 | ||
ȃMale | 30 (60.0%) | 34 (23.8%) | |
ȃFemale | 20 (40.0%) | 109 (76.2%) | |
Height, cm | 165.1 ± 8.8 | 158.5 ± 8.2 | <0.001 |
Weight, kg | 59.5 (54.0–72.0) | 55.0 (50.0–62.5) | 0.002 |
BSA, m2 | 1.67 (1.54–1.84) | 1.55 (1.46–1.68) | <0.001 |
BMI, kg/m2 | 22.5 (20.4–25.7) | 22.3 (20.3–24.4) | 0.503 |
ȃ ≥ 30 | 2 (4.0%) | 4 (2.8%) | >0.999 |
ȃ < 30 | 48 (96.0%) | 139 (97.2%) | >0.999 |
Systolic blood pressure, mmHg | 121.0 ± 17.8 | 121.3 ± 16.1 | 0.926 |
Diastolic blood pressure, mmHg | 70.9 ± 9.2 | 68.7 ± 10.5 | 0.193 |
Heart rate, beats per minute | 68.0 (60.0–80.0) | 67.0 (60.0–76.0) | 0.326 |
Rhythm | 0.178 | ||
ȃSinus rhythm | 20 (40.0%) | 42 (29.4%) | |
ȃAtrial fibrillation | 27 (54.0%) | 97 (67.8%) | |
ȃPaced | 3 (6.0%) | 4 (2.8%) | |
Prosthetic valve size | 21.0 (19.0–23.0) | 21.0 (19.0–21.0) | 0.032 |
ȃRange | 19–25 | 19–25 | |
ȃLarge size (≥23 mm) | 18 (36.0%) | 16 (11.2%) | <0.001 |
Prosthetic valve type | |||
ȃBileaflet | 49 (98.0%) | 137 (95.8) | 0.783 |
ȃMonoleaflet | 1 (2.0%) | 6 (4.2%) | 0.783 |
ȃSupra-annular type | 6 (12.0%) | 12 (8.4%) | 0.636 |
Time interval from previous AVR, years | 11.9 ± 9.2 | 19.3 ± 7.8 | <0.001 |
Reference EOA | 1.50 (1.30–1.70) | 1.40 (1.20–1.50) | 0.005 |
Projected indexed EOA, cm2/m2 | 0.90 (0.75–0.98) | 0.91 (0.76–0.99) | 0.991 |
ȃModerate patient–prosthesis mismatch | 12 (24.0%) | 36 (25.2%) | >0.999 |
ȃSevere patient–prosthesis mismatch | 4 (8.0%) | 14 (9.8%) | 0.927 |
DVR status | 33 (66.0%) | 119 (83.2%) | 0.018 |
Limited motion of PAV leaflet | 0 (0%) | 84 (58.7%) | <0.001 |
Concomitant aortic valve regurgitation | 0.091 | ||
ȃNo, mild | 45 (90.0%) | 139 (97.2%) | |
ȃModerate or severe | 5 (10.0%) | 4 (2.8%) | |
Method of diagnosis | <0.001 | ||
ȃSurgical inspection | 13 (26.0%) | 94 (65.7%) | |
ȃCardiac CT | 37 (74.0%) | 49 (34.3%) | |
Indication for reoperation | <0.001 | ||
ȃPAV obstruction | 0 (0%) | 94 (65.7%) | |
ȃPAV paravalvular regurgitation | 2 (4.0%) | 0 (0%) | |
ȃMitral or tricuspid valvular dysfunction | 11 (22.0%) | 29 (20.3%) | |
Redo aortic valve replacement | 2 (4.0%) | 94 (65.7%) | <0.001 |
Echocardiography | |||
ȃEOA, cm2 | 1.26 (1.00–1.50) | 0.80 (0.68–0.99) | <0.001 |
ȃMeasured/reference EOA ratio | 0.86 ± 0.17 | 0.63 ± 0.18 | <0.001 |
ȃEOA/BSA, cm2/m2 | 0.76 (0.64–0.90) | 0.51 (0.43–0.64) | <0.001 |
ȃLVEF, % | 62.0 (57.0–68.0) | 66.0 (61.0–71.0) | 0.009 |
ȃȃ < 45% | 5 (10.0%) | 9 (6.3%) | 0.580 |
ȃLVOT diameter, mm | 21.7 (19.9–23.3) | 20.5 (19.0–21.2) | <0.001 |
ȃLVOT diameter difference from valve size | 0.45 (−0.30–1.00) | −0.10 (−1.25–0.20) | <0.001 |
ȃLVOT VTI, cm | 22.7 ± 4.8 | 22.7 ± 4.6 | 0.994 |
ȃPeak LVOT velocity, m/s | 1.09 (0.97–1.29) | 1.08 (0.93–1.21) | 0.296 |
ȃStroke volume, mL | 83.2 (72.0–103.1) | 69.2 (60.7–83.4) | <0.001 |
ȃStroke volume index, mL/m2 | 49.4 (41.4–60.1) | 44.7 (38.6–51.7) | 0.019 |
ȃDynamic LVOT obstruction | 0 (0%) | 8 (5.6%) | 0.195 |
ȃAV VTI, cm | 66.1 (57.6–75.3) | 88.0 (72.1–101.3) | <0.001 |
ȃPeak transprosthetic velocity, m/s | 3.33 (3.09–3.67) | 4.10 (3.63–4.71) | <0.001 |
ȃPeak transprosthetic pressure, mmHg | 44.5 (38.0–54.0) | 67.0 (52.0–88.2) | <0.001 |
ȃMean transprosthetic pressure, mmHg | 25.5 (21.0–30.0) | 37.4 (29.3–50.5) | <0.001 |
ȃDoppler velocity index | 0.34 (0.29–0.41) | 0.26 (0.23–0.30) | <0.001 |
ȃAcceleration time of aortic velocity, msec | 83.4 ± 22.5 | 98.9 ± 23.0 | <0.001 |
ȃEjection time, msec | 282.8 ± 44.2 | 293.5 ± 35.1 | 0.128 |
ȃAcceleration time/ejection time ratio | 0.30 ± 0.07 | 0.34 ± 0.06 | <0.001 |
ȃCW Doppler jet contour | <0.001 | ||
ȃȃTriangle, early peaking | 11 (22.0%) | 17 (11.9%) | |
ȃȃIntermediate | 38 (76.0%) | 78 (54.5%) | |
ȃȃRound, symmetrical | 1 (2.0%) | 48 (33.6%) |
. | No PAV obstruction (n = 50) . | PAV obstruction (n = 143) . | P value . |
---|---|---|---|
Age, years | 58.0 (52.0–67.0) | 65.0 (54.0–69.0) | 0.036 |
Sex | <0.001 | ||
ȃMale | 30 (60.0%) | 34 (23.8%) | |
ȃFemale | 20 (40.0%) | 109 (76.2%) | |
Height, cm | 165.1 ± 8.8 | 158.5 ± 8.2 | <0.001 |
Weight, kg | 59.5 (54.0–72.0) | 55.0 (50.0–62.5) | 0.002 |
BSA, m2 | 1.67 (1.54–1.84) | 1.55 (1.46–1.68) | <0.001 |
BMI, kg/m2 | 22.5 (20.4–25.7) | 22.3 (20.3–24.4) | 0.503 |
ȃ ≥ 30 | 2 (4.0%) | 4 (2.8%) | >0.999 |
ȃ < 30 | 48 (96.0%) | 139 (97.2%) | >0.999 |
Systolic blood pressure, mmHg | 121.0 ± 17.8 | 121.3 ± 16.1 | 0.926 |
Diastolic blood pressure, mmHg | 70.9 ± 9.2 | 68.7 ± 10.5 | 0.193 |
Heart rate, beats per minute | 68.0 (60.0–80.0) | 67.0 (60.0–76.0) | 0.326 |
Rhythm | 0.178 | ||
ȃSinus rhythm | 20 (40.0%) | 42 (29.4%) | |
ȃAtrial fibrillation | 27 (54.0%) | 97 (67.8%) | |
ȃPaced | 3 (6.0%) | 4 (2.8%) | |
Prosthetic valve size | 21.0 (19.0–23.0) | 21.0 (19.0–21.0) | 0.032 |
ȃRange | 19–25 | 19–25 | |
ȃLarge size (≥23 mm) | 18 (36.0%) | 16 (11.2%) | <0.001 |
Prosthetic valve type | |||
ȃBileaflet | 49 (98.0%) | 137 (95.8) | 0.783 |
ȃMonoleaflet | 1 (2.0%) | 6 (4.2%) | 0.783 |
ȃSupra-annular type | 6 (12.0%) | 12 (8.4%) | 0.636 |
Time interval from previous AVR, years | 11.9 ± 9.2 | 19.3 ± 7.8 | <0.001 |
Reference EOA | 1.50 (1.30–1.70) | 1.40 (1.20–1.50) | 0.005 |
Projected indexed EOA, cm2/m2 | 0.90 (0.75–0.98) | 0.91 (0.76–0.99) | 0.991 |
ȃModerate patient–prosthesis mismatch | 12 (24.0%) | 36 (25.2%) | >0.999 |
ȃSevere patient–prosthesis mismatch | 4 (8.0%) | 14 (9.8%) | 0.927 |
DVR status | 33 (66.0%) | 119 (83.2%) | 0.018 |
Limited motion of PAV leaflet | 0 (0%) | 84 (58.7%) | <0.001 |
Concomitant aortic valve regurgitation | 0.091 | ||
ȃNo, mild | 45 (90.0%) | 139 (97.2%) | |
ȃModerate or severe | 5 (10.0%) | 4 (2.8%) | |
Method of diagnosis | <0.001 | ||
ȃSurgical inspection | 13 (26.0%) | 94 (65.7%) | |
ȃCardiac CT | 37 (74.0%) | 49 (34.3%) | |
Indication for reoperation | <0.001 | ||
ȃPAV obstruction | 0 (0%) | 94 (65.7%) | |
ȃPAV paravalvular regurgitation | 2 (4.0%) | 0 (0%) | |
ȃMitral or tricuspid valvular dysfunction | 11 (22.0%) | 29 (20.3%) | |
Redo aortic valve replacement | 2 (4.0%) | 94 (65.7%) | <0.001 |
Echocardiography | |||
ȃEOA, cm2 | 1.26 (1.00–1.50) | 0.80 (0.68–0.99) | <0.001 |
ȃMeasured/reference EOA ratio | 0.86 ± 0.17 | 0.63 ± 0.18 | <0.001 |
ȃEOA/BSA, cm2/m2 | 0.76 (0.64–0.90) | 0.51 (0.43–0.64) | <0.001 |
ȃLVEF, % | 62.0 (57.0–68.0) | 66.0 (61.0–71.0) | 0.009 |
ȃȃ < 45% | 5 (10.0%) | 9 (6.3%) | 0.580 |
ȃLVOT diameter, mm | 21.7 (19.9–23.3) | 20.5 (19.0–21.2) | <0.001 |
ȃLVOT diameter difference from valve size | 0.45 (−0.30–1.00) | −0.10 (−1.25–0.20) | <0.001 |
ȃLVOT VTI, cm | 22.7 ± 4.8 | 22.7 ± 4.6 | 0.994 |
ȃPeak LVOT velocity, m/s | 1.09 (0.97–1.29) | 1.08 (0.93–1.21) | 0.296 |
ȃStroke volume, mL | 83.2 (72.0–103.1) | 69.2 (60.7–83.4) | <0.001 |
ȃStroke volume index, mL/m2 | 49.4 (41.4–60.1) | 44.7 (38.6–51.7) | 0.019 |
ȃDynamic LVOT obstruction | 0 (0%) | 8 (5.6%) | 0.195 |
ȃAV VTI, cm | 66.1 (57.6–75.3) | 88.0 (72.1–101.3) | <0.001 |
ȃPeak transprosthetic velocity, m/s | 3.33 (3.09–3.67) | 4.10 (3.63–4.71) | <0.001 |
ȃPeak transprosthetic pressure, mmHg | 44.5 (38.0–54.0) | 67.0 (52.0–88.2) | <0.001 |
ȃMean transprosthetic pressure, mmHg | 25.5 (21.0–30.0) | 37.4 (29.3–50.5) | <0.001 |
ȃDoppler velocity index | 0.34 (0.29–0.41) | 0.26 (0.23–0.30) | <0.001 |
ȃAcceleration time of aortic velocity, msec | 83.4 ± 22.5 | 98.9 ± 23.0 | <0.001 |
ȃEjection time, msec | 282.8 ± 44.2 | 293.5 ± 35.1 | 0.128 |
ȃAcceleration time/ejection time ratio | 0.30 ± 0.07 | 0.34 ± 0.06 | <0.001 |
ȃCW Doppler jet contour | <0.001 | ||
ȃȃTriangle, early peaking | 11 (22.0%) | 17 (11.9%) | |
ȃȃIntermediate | 38 (76.0%) | 78 (54.5%) | |
ȃȃRound, symmetrical | 1 (2.0%) | 48 (33.6%) |
AVR, aortic valve replacement; CW, continuous wave; LVEF, left ventricular ejection fraction; LVOT, left ventricular outflow tract; VTI, velocity-time integral.
Regarding echocardiographic parameters, patients in the PAV obstruction group had a lower measured EOA [0.80 (0.68–0.99) vs. 1.26 (1.00–1.50) cm2; P < 0.001] and measured/reference EOA ratio (0.63 ± 0.18 vs. 0.86 ± 0.17; P < 0.001). The PAV obstruction group had smaller PAV size, reference EOA value [1.40 (1.20–1.50) vs. 1.50 (1.30–1.70) cm2; P = 0.005], and lower stroke volume index [SVi, 44.7 (38.6–51.7) vs. 49.4 (41.4–60.1) mL/m2, P = 0.019]. Patients with DVR (with mitral prosthesis) showed lower SVi than isolated AVR [44.9 (38.2–51.8) vs. 50.1 (40.6–60.1) mL/m2, P = 0.018]. The patients with surgical inspection had significantly higher mean transprosthetic PGs [40.0 (29.1–53.0) vs. 29.0 (23.0–35.8) mmHg, P < 0.001] and lower measured/reference EOA ratio (0.65 ± 0.18 vs. 0.73 ± 0.22, P = 0.007). Conventional Doppler parameters, including DVI, AT, AT/ET ratio, and Doppler jet contour differed significantly between the two groups (Table 1). The ICC between the LVOT diameter and reference annulus diameter of the PAV was not reliable (ICC = 0.566).
Diagnostic accuracy of Doppler parameters
The Doppler parameters and ROC analysis are described in Table 2. All conventional Doppler parameters could discriminate between PAV obstruction patients and no PAV obstructions, except for an AT/ET ratio over 0.37. The largest area under the ROC curve was observed with the measured EOA [0.849; 95% confidence interval (CI), 0.788–0.910; P < 0.001], followed by the measured/reference EOA ratio (0.840; 95% CI, 0.783–0.898; P < 0.001). The best cutoff value was 1.07 for the measured EOA and 0.71 for the measured/reference EOA ratio. The best cutoff value of measured EOA was different according to PAV size [1.25 for large PAV (≥23 mm), 0.80 for small PAV]. However, those of measured/reference EOA ratio was consistent as 0.71. When the measured/reference EOA ratio cutoff was set to 0.70 for convenience, the sensitivity and specificity were 68.5 and 84.0%, respectively. The measured/reference EOA ratio added incremental value over conventional Doppler parameters, such as mean transprosthetic PG, DVI, and AT/ET ratio (Figure 2). The inter- (ICC = 0.981; P < 0.001) and intra-observer variabilities (ICC = 0.984, P < 0.001) for the measured/reference EOA ratio showed excellent agreement (see Supplementary data online, Figure S2). The mean difference between the observers was −0.014 ± 0.039. The mean difference for the same observer was −0.002 ± 0.038.

Incremental value of measured/reference effective orifice ration for discriminating prosthetic aortic valve obstruction.
Receiver operating characteristic curve analysis for PAV obstruction of Doppler parameters
. | AUC (95% CI) . | P Value . | Best cutoff value . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Measured EOA/reference EOA ratio | 0.840 (0.783–0.898) | <0.001 | 0.71 | 73.4 | 82.0 | 92.9 | 52.5 |
ȃMeasured EOA/reference EOA ratio <0.7 | 0.763 (0.699–0.827) | <0.001 | Positive | 68.5 | 84.0 | 92.5 | 48.3 |
Measured EOA vs. reference EOA | |||||||
ȃMeasured EOA < reference EOA—1SD | 0.733 (0.658–0.807) | <0.001 | Positive | 82.5 | 64.0 | 86.8 | 56.1 |
ȃMeasured EOA < reference EOA—2SD | 0.691 (0.635–0.747) | <0.001 | Positive | 46.2 | 92.0 | 94.3 | 37.4 |
Measured EOA, cm2 | 0.849 (0.788–0.910) | <0.001 | 1.07 | 83.2 | 70.0 | 88.8 | 59.3 |
ȃMeasured EOA <1.1 cm2 | 0.756 (0.684–0.828) | <0.001 | Positive | 83.2 | 68.0 | 88.1 | 58.6 |
ȃMeasured EOA <0.8 cm2 | 0.728 (0.679–0.778) | <0.001 | Positive | 49.7 | 96.0 | 97.3 | 40.0 |
Doppler velocity index | 0.766 (0.687–0.846) | <0.001 | 0.31 | 81 | 64.0 | 86.5 | 54.2 |
ȃDoppler velocity index <0.35 | 0.644 (0.570–0.717) | <0.001 | Positive | 88.7 | 40.0 | 80.8 | 55.6 |
ȃDoppler velocity index < 0.25 | 0.637 (0.576–0.698) | <0.001 | Positive | 39.4 | 88.0 | 90.3 | 33.8 |
Rounded, symmetrical CW jet contour | 0.658 (0.614–0.701) | <0.001 | Positive | 33.6 | 98.0 | 98.0 | 34.0 |
Acceleration time, msec | 0.691 (0.607–0.776) | <0.001 | 88 | 66.4 | 66.0 | 84.8 | 40.7 |
ȃAcceleration time ≥80 msec | 0.615 (0.537–0.693) | 0.001 | Positive | 76.9 | 46.0 | 80.3 | 41.1 |
ȃAcceleration time >100 msec | 0.637 (0.570–0.705) | <0.001 | Positive | 45.5 | 82.0 | 87.8 | 34.5 |
Acceleration time/LV ejection time ratio | 0.668 (0.575–0.762) | <0.001 | 0.33 | 55.9 | 76.0 | 87.0 | 37.6 |
ȃAcceleration time/LV ejection time ratio >0.32 | 0.637 (0.561–0.713) | <0.001 | Positive | 57.3 | 70.0 | 84.5 | 36.5 |
ȃAcceleration time/LV ejection time ratio >0.37 | 0.529 (0.465–0.593) | 0.200 | Positive | 23.8 | 82.0 | 79.1 | 27.3 |
. | AUC (95% CI) . | P Value . | Best cutoff value . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Measured EOA/reference EOA ratio | 0.840 (0.783–0.898) | <0.001 | 0.71 | 73.4 | 82.0 | 92.9 | 52.5 |
ȃMeasured EOA/reference EOA ratio <0.7 | 0.763 (0.699–0.827) | <0.001 | Positive | 68.5 | 84.0 | 92.5 | 48.3 |
Measured EOA vs. reference EOA | |||||||
ȃMeasured EOA < reference EOA—1SD | 0.733 (0.658–0.807) | <0.001 | Positive | 82.5 | 64.0 | 86.8 | 56.1 |
ȃMeasured EOA < reference EOA—2SD | 0.691 (0.635–0.747) | <0.001 | Positive | 46.2 | 92.0 | 94.3 | 37.4 |
Measured EOA, cm2 | 0.849 (0.788–0.910) | <0.001 | 1.07 | 83.2 | 70.0 | 88.8 | 59.3 |
ȃMeasured EOA <1.1 cm2 | 0.756 (0.684–0.828) | <0.001 | Positive | 83.2 | 68.0 | 88.1 | 58.6 |
ȃMeasured EOA <0.8 cm2 | 0.728 (0.679–0.778) | <0.001 | Positive | 49.7 | 96.0 | 97.3 | 40.0 |
Doppler velocity index | 0.766 (0.687–0.846) | <0.001 | 0.31 | 81 | 64.0 | 86.5 | 54.2 |
ȃDoppler velocity index <0.35 | 0.644 (0.570–0.717) | <0.001 | Positive | 88.7 | 40.0 | 80.8 | 55.6 |
ȃDoppler velocity index < 0.25 | 0.637 (0.576–0.698) | <0.001 | Positive | 39.4 | 88.0 | 90.3 | 33.8 |
Rounded, symmetrical CW jet contour | 0.658 (0.614–0.701) | <0.001 | Positive | 33.6 | 98.0 | 98.0 | 34.0 |
Acceleration time, msec | 0.691 (0.607–0.776) | <0.001 | 88 | 66.4 | 66.0 | 84.8 | 40.7 |
ȃAcceleration time ≥80 msec | 0.615 (0.537–0.693) | 0.001 | Positive | 76.9 | 46.0 | 80.3 | 41.1 |
ȃAcceleration time >100 msec | 0.637 (0.570–0.705) | <0.001 | Positive | 45.5 | 82.0 | 87.8 | 34.5 |
Acceleration time/LV ejection time ratio | 0.668 (0.575–0.762) | <0.001 | 0.33 | 55.9 | 76.0 | 87.0 | 37.6 |
ȃAcceleration time/LV ejection time ratio >0.32 | 0.637 (0.561–0.713) | <0.001 | Positive | 57.3 | 70.0 | 84.5 | 36.5 |
ȃAcceleration time/LV ejection time ratio >0.37 | 0.529 (0.465–0.593) | 0.200 | Positive | 23.8 | 82.0 | 79.1 | 27.3 |
AVR, aortic valve replacement; CW, continuous wave; LVEF, left ventricular ejection fraction; LVOT, left ventricular outflow tract; VTI, velocity-time integral.
Receiver operating characteristic curve analysis for PAV obstruction of Doppler parameters
. | AUC (95% CI) . | P Value . | Best cutoff value . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Measured EOA/reference EOA ratio | 0.840 (0.783–0.898) | <0.001 | 0.71 | 73.4 | 82.0 | 92.9 | 52.5 |
ȃMeasured EOA/reference EOA ratio <0.7 | 0.763 (0.699–0.827) | <0.001 | Positive | 68.5 | 84.0 | 92.5 | 48.3 |
Measured EOA vs. reference EOA | |||||||
ȃMeasured EOA < reference EOA—1SD | 0.733 (0.658–0.807) | <0.001 | Positive | 82.5 | 64.0 | 86.8 | 56.1 |
ȃMeasured EOA < reference EOA—2SD | 0.691 (0.635–0.747) | <0.001 | Positive | 46.2 | 92.0 | 94.3 | 37.4 |
Measured EOA, cm2 | 0.849 (0.788–0.910) | <0.001 | 1.07 | 83.2 | 70.0 | 88.8 | 59.3 |
ȃMeasured EOA <1.1 cm2 | 0.756 (0.684–0.828) | <0.001 | Positive | 83.2 | 68.0 | 88.1 | 58.6 |
ȃMeasured EOA <0.8 cm2 | 0.728 (0.679–0.778) | <0.001 | Positive | 49.7 | 96.0 | 97.3 | 40.0 |
Doppler velocity index | 0.766 (0.687–0.846) | <0.001 | 0.31 | 81 | 64.0 | 86.5 | 54.2 |
ȃDoppler velocity index <0.35 | 0.644 (0.570–0.717) | <0.001 | Positive | 88.7 | 40.0 | 80.8 | 55.6 |
ȃDoppler velocity index < 0.25 | 0.637 (0.576–0.698) | <0.001 | Positive | 39.4 | 88.0 | 90.3 | 33.8 |
Rounded, symmetrical CW jet contour | 0.658 (0.614–0.701) | <0.001 | Positive | 33.6 | 98.0 | 98.0 | 34.0 |
Acceleration time, msec | 0.691 (0.607–0.776) | <0.001 | 88 | 66.4 | 66.0 | 84.8 | 40.7 |
ȃAcceleration time ≥80 msec | 0.615 (0.537–0.693) | 0.001 | Positive | 76.9 | 46.0 | 80.3 | 41.1 |
ȃAcceleration time >100 msec | 0.637 (0.570–0.705) | <0.001 | Positive | 45.5 | 82.0 | 87.8 | 34.5 |
Acceleration time/LV ejection time ratio | 0.668 (0.575–0.762) | <0.001 | 0.33 | 55.9 | 76.0 | 87.0 | 37.6 |
ȃAcceleration time/LV ejection time ratio >0.32 | 0.637 (0.561–0.713) | <0.001 | Positive | 57.3 | 70.0 | 84.5 | 36.5 |
ȃAcceleration time/LV ejection time ratio >0.37 | 0.529 (0.465–0.593) | 0.200 | Positive | 23.8 | 82.0 | 79.1 | 27.3 |
. | AUC (95% CI) . | P Value . | Best cutoff value . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Measured EOA/reference EOA ratio | 0.840 (0.783–0.898) | <0.001 | 0.71 | 73.4 | 82.0 | 92.9 | 52.5 |
ȃMeasured EOA/reference EOA ratio <0.7 | 0.763 (0.699–0.827) | <0.001 | Positive | 68.5 | 84.0 | 92.5 | 48.3 |
Measured EOA vs. reference EOA | |||||||
ȃMeasured EOA < reference EOA—1SD | 0.733 (0.658–0.807) | <0.001 | Positive | 82.5 | 64.0 | 86.8 | 56.1 |
ȃMeasured EOA < reference EOA—2SD | 0.691 (0.635–0.747) | <0.001 | Positive | 46.2 | 92.0 | 94.3 | 37.4 |
Measured EOA, cm2 | 0.849 (0.788–0.910) | <0.001 | 1.07 | 83.2 | 70.0 | 88.8 | 59.3 |
ȃMeasured EOA <1.1 cm2 | 0.756 (0.684–0.828) | <0.001 | Positive | 83.2 | 68.0 | 88.1 | 58.6 |
ȃMeasured EOA <0.8 cm2 | 0.728 (0.679–0.778) | <0.001 | Positive | 49.7 | 96.0 | 97.3 | 40.0 |
Doppler velocity index | 0.766 (0.687–0.846) | <0.001 | 0.31 | 81 | 64.0 | 86.5 | 54.2 |
ȃDoppler velocity index <0.35 | 0.644 (0.570–0.717) | <0.001 | Positive | 88.7 | 40.0 | 80.8 | 55.6 |
ȃDoppler velocity index < 0.25 | 0.637 (0.576–0.698) | <0.001 | Positive | 39.4 | 88.0 | 90.3 | 33.8 |
Rounded, symmetrical CW jet contour | 0.658 (0.614–0.701) | <0.001 | Positive | 33.6 | 98.0 | 98.0 | 34.0 |
Acceleration time, msec | 0.691 (0.607–0.776) | <0.001 | 88 | 66.4 | 66.0 | 84.8 | 40.7 |
ȃAcceleration time ≥80 msec | 0.615 (0.537–0.693) | 0.001 | Positive | 76.9 | 46.0 | 80.3 | 41.1 |
ȃAcceleration time >100 msec | 0.637 (0.570–0.705) | <0.001 | Positive | 45.5 | 82.0 | 87.8 | 34.5 |
Acceleration time/LV ejection time ratio | 0.668 (0.575–0.762) | <0.001 | 0.33 | 55.9 | 76.0 | 87.0 | 37.6 |
ȃAcceleration time/LV ejection time ratio >0.32 | 0.637 (0.561–0.713) | <0.001 | Positive | 57.3 | 70.0 | 84.5 | 36.5 |
ȃAcceleration time/LV ejection time ratio >0.37 | 0.529 (0.465–0.593) | 0.200 | Positive | 23.8 | 82.0 | 79.1 | 27.3 |
AVR, aortic valve replacement; CW, continuous wave; LVEF, left ventricular ejection fraction; LVOT, left ventricular outflow tract; VTI, velocity-time integral.
Accuracy of novel diagnostic algorithm using measured/reference EOA ratio
The diagnostic algorithm models and their ROC analyses are presented in Table 3 and Graphical abstract B and C. All diagnostic models showed significant accuracy for detecting PAV obstruction. The novel model had the highest AUC value (0.763, P < 0.001), 68.5% sensitivity, and 84.0% specificity. Compared with the 2009 algorithm, which did not include reference EOA values, the novel model had significantly higher AUC values (0.763 vs. 0.642; P = 0.003). The accuracies of the novel and 2016 algorithms were not significantly different (AUC, 0.763 vs. 0.731; P = 0.309). Despite its simplicity, the novel algorithm had a significantly higher AUC value than the 2016 recommendation (0.788 vs. 0.642; P = 0.019) in patients with a large PAV (≥23 mm) (Table 3). The sensitivity analyses with patients performed surgical inspection only and patients with limited leaflet motion showed similar results to those of the main findings (see Supplementary data online, Table S3).
Receiver operating characteristic curve analysis of diagnostic algorithm for PAV obstruction in specified subgroups
. | AUC (95% CI) . | P value . | P value (vs. novel) . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Overall study population, n = 193 | |||||||
ȃNovel algorithm | 0.763 (0.699–0.827) | <0.001 | 68.5 | 84.0 | 92.5 | 48.3 | |
ȃ2016 recommendation | 0.731 (0.658–0.804) | <0.001 | 0.309 | 76.2 | 70.0 | 87.9 | 50.7 |
ȃ2009 recommendation | 0.642 (0.584–0.700) | <0.001 | 0.003 | 38.5 | 90.0 | 91.7 | 33.8 |
Large prosthetic AV (≥23 mm), n = 34 | |||||||
ȃNovel algorithm | 0.788 (0.649–0.927) | <0.001 | 68.8 | 88.9 | 84.6 | 76.2 | |
ȃ2016 recommendation | 0.642 (0.478–0.807) | 0.051 | 0.019 | 56.2 | 72.2 | 64.3 | 65.0 |
ȃ2009 recommendation | 0.510 (0.378–0.643) | 0.448 | 0.003 | 18.8 | 83.3 | 50.0 | 53.6 |
Low gradient (mean PG <40 mmHg), n = 129 | |||||||
ȃNovel algorithm | 0.668 (0.592–0.744) | <0.001 | 50.0 | 83.7 | 83.3 | 50.6 | |
ȃ2016 recommendation | 0.688 (0.606–0.771) | <0.001 | 0.591 | 66.2 | 71.4 | 79.1 | 56.5 |
ȃ2009 recommendation | 0.572 (0.512–0.632) | 0.018 | 0.033 | 22.5 | 91.8 | 81.8 | 42.1 |
DVR status, n = 152 | |||||||
ȃNovel diagnostic algorithm | 0.758 (0.679–0.836) | <0.001 | 69.7 | 81.8 | 93.3 | 42.9 | |
ȃ2016 recommendation | 0.724 (0.634–0.814) | <0.001 | 0.364 | 78.2 | 66.7 | 89.4 | 45.8 |
ȃ2009 recommendation | 0.649 (0.578–0.721) | <0.001 | 0.032 | 42.0 | 87.9 | 92.6 | 29.6 |
Aortic valve replacement only, n = 41 | |||||||
ȃNovel diagnostic algorithm | 0.754 (0.627–0.880) | <0.001 | 62.5 | 88.2 | 88.2 | 62.5 | |
ȃ2016 recommendation | 0.716 (0.574–0.857) | 0.004 | 0.582 | 66.7 | 76.5 | 80.0 | 61.9 |
ȃ2009 recommendation | 0.575 (0.474–0.676) | 0.097 | 0.014 | 20.8 | 94.1 | 83.3 | 45.7 |
Atrial fibrillation, n = 124 | |||||||
ȃNovel algorithm | 0.724 (0.631–0.817) | <0.001 | 67.0 | 77.8 | 91.5 | 39.6 | |
ȃ2016 recommendation | 0.662 (0.558–0.767) | <0.001 | 0.160 | 73.2 | 59.3 | 86.6 | 38.1 |
ȃ2009 recommendation | 0.620 (0.543–0.697) | 0.008 | 0.068 | 35.1 | 88.9 | 91.9 | 27.6 |
. | AUC (95% CI) . | P value . | P value (vs. novel) . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Overall study population, n = 193 | |||||||
ȃNovel algorithm | 0.763 (0.699–0.827) | <0.001 | 68.5 | 84.0 | 92.5 | 48.3 | |
ȃ2016 recommendation | 0.731 (0.658–0.804) | <0.001 | 0.309 | 76.2 | 70.0 | 87.9 | 50.7 |
ȃ2009 recommendation | 0.642 (0.584–0.700) | <0.001 | 0.003 | 38.5 | 90.0 | 91.7 | 33.8 |
Large prosthetic AV (≥23 mm), n = 34 | |||||||
ȃNovel algorithm | 0.788 (0.649–0.927) | <0.001 | 68.8 | 88.9 | 84.6 | 76.2 | |
ȃ2016 recommendation | 0.642 (0.478–0.807) | 0.051 | 0.019 | 56.2 | 72.2 | 64.3 | 65.0 |
ȃ2009 recommendation | 0.510 (0.378–0.643) | 0.448 | 0.003 | 18.8 | 83.3 | 50.0 | 53.6 |
Low gradient (mean PG <40 mmHg), n = 129 | |||||||
ȃNovel algorithm | 0.668 (0.592–0.744) | <0.001 | 50.0 | 83.7 | 83.3 | 50.6 | |
ȃ2016 recommendation | 0.688 (0.606–0.771) | <0.001 | 0.591 | 66.2 | 71.4 | 79.1 | 56.5 |
ȃ2009 recommendation | 0.572 (0.512–0.632) | 0.018 | 0.033 | 22.5 | 91.8 | 81.8 | 42.1 |
DVR status, n = 152 | |||||||
ȃNovel diagnostic algorithm | 0.758 (0.679–0.836) | <0.001 | 69.7 | 81.8 | 93.3 | 42.9 | |
ȃ2016 recommendation | 0.724 (0.634–0.814) | <0.001 | 0.364 | 78.2 | 66.7 | 89.4 | 45.8 |
ȃ2009 recommendation | 0.649 (0.578–0.721) | <0.001 | 0.032 | 42.0 | 87.9 | 92.6 | 29.6 |
Aortic valve replacement only, n = 41 | |||||||
ȃNovel diagnostic algorithm | 0.754 (0.627–0.880) | <0.001 | 62.5 | 88.2 | 88.2 | 62.5 | |
ȃ2016 recommendation | 0.716 (0.574–0.857) | 0.004 | 0.582 | 66.7 | 76.5 | 80.0 | 61.9 |
ȃ2009 recommendation | 0.575 (0.474–0.676) | 0.097 | 0.014 | 20.8 | 94.1 | 83.3 | 45.7 |
Atrial fibrillation, n = 124 | |||||||
ȃNovel algorithm | 0.724 (0.631–0.817) | <0.001 | 67.0 | 77.8 | 91.5 | 39.6 | |
ȃ2016 recommendation | 0.662 (0.558–0.767) | <0.001 | 0.160 | 73.2 | 59.3 | 86.6 | 38.1 |
ȃ2009 recommendation | 0.620 (0.543–0.697) | 0.008 | 0.068 | 35.1 | 88.9 | 91.9 | 27.6 |
AVR, aortic valve replacement; CW, continuous wave; LVEF, left ventricular ejection fraction; LVOT, left ventricular outflow tract; VTI, velocity-time integral.
Receiver operating characteristic curve analysis of diagnostic algorithm for PAV obstruction in specified subgroups
. | AUC (95% CI) . | P value . | P value (vs. novel) . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Overall study population, n = 193 | |||||||
ȃNovel algorithm | 0.763 (0.699–0.827) | <0.001 | 68.5 | 84.0 | 92.5 | 48.3 | |
ȃ2016 recommendation | 0.731 (0.658–0.804) | <0.001 | 0.309 | 76.2 | 70.0 | 87.9 | 50.7 |
ȃ2009 recommendation | 0.642 (0.584–0.700) | <0.001 | 0.003 | 38.5 | 90.0 | 91.7 | 33.8 |
Large prosthetic AV (≥23 mm), n = 34 | |||||||
ȃNovel algorithm | 0.788 (0.649–0.927) | <0.001 | 68.8 | 88.9 | 84.6 | 76.2 | |
ȃ2016 recommendation | 0.642 (0.478–0.807) | 0.051 | 0.019 | 56.2 | 72.2 | 64.3 | 65.0 |
ȃ2009 recommendation | 0.510 (0.378–0.643) | 0.448 | 0.003 | 18.8 | 83.3 | 50.0 | 53.6 |
Low gradient (mean PG <40 mmHg), n = 129 | |||||||
ȃNovel algorithm | 0.668 (0.592–0.744) | <0.001 | 50.0 | 83.7 | 83.3 | 50.6 | |
ȃ2016 recommendation | 0.688 (0.606–0.771) | <0.001 | 0.591 | 66.2 | 71.4 | 79.1 | 56.5 |
ȃ2009 recommendation | 0.572 (0.512–0.632) | 0.018 | 0.033 | 22.5 | 91.8 | 81.8 | 42.1 |
DVR status, n = 152 | |||||||
ȃNovel diagnostic algorithm | 0.758 (0.679–0.836) | <0.001 | 69.7 | 81.8 | 93.3 | 42.9 | |
ȃ2016 recommendation | 0.724 (0.634–0.814) | <0.001 | 0.364 | 78.2 | 66.7 | 89.4 | 45.8 |
ȃ2009 recommendation | 0.649 (0.578–0.721) | <0.001 | 0.032 | 42.0 | 87.9 | 92.6 | 29.6 |
Aortic valve replacement only, n = 41 | |||||||
ȃNovel diagnostic algorithm | 0.754 (0.627–0.880) | <0.001 | 62.5 | 88.2 | 88.2 | 62.5 | |
ȃ2016 recommendation | 0.716 (0.574–0.857) | 0.004 | 0.582 | 66.7 | 76.5 | 80.0 | 61.9 |
ȃ2009 recommendation | 0.575 (0.474–0.676) | 0.097 | 0.014 | 20.8 | 94.1 | 83.3 | 45.7 |
Atrial fibrillation, n = 124 | |||||||
ȃNovel algorithm | 0.724 (0.631–0.817) | <0.001 | 67.0 | 77.8 | 91.5 | 39.6 | |
ȃ2016 recommendation | 0.662 (0.558–0.767) | <0.001 | 0.160 | 73.2 | 59.3 | 86.6 | 38.1 |
ȃ2009 recommendation | 0.620 (0.543–0.697) | 0.008 | 0.068 | 35.1 | 88.9 | 91.9 | 27.6 |
. | AUC (95% CI) . | P value . | P value (vs. novel) . | Sensitivity, % . | Specificity, % . | Positive predictive value, % . | Negative predictive value, % . |
---|---|---|---|---|---|---|---|
Overall study population, n = 193 | |||||||
ȃNovel algorithm | 0.763 (0.699–0.827) | <0.001 | 68.5 | 84.0 | 92.5 | 48.3 | |
ȃ2016 recommendation | 0.731 (0.658–0.804) | <0.001 | 0.309 | 76.2 | 70.0 | 87.9 | 50.7 |
ȃ2009 recommendation | 0.642 (0.584–0.700) | <0.001 | 0.003 | 38.5 | 90.0 | 91.7 | 33.8 |
Large prosthetic AV (≥23 mm), n = 34 | |||||||
ȃNovel algorithm | 0.788 (0.649–0.927) | <0.001 | 68.8 | 88.9 | 84.6 | 76.2 | |
ȃ2016 recommendation | 0.642 (0.478–0.807) | 0.051 | 0.019 | 56.2 | 72.2 | 64.3 | 65.0 |
ȃ2009 recommendation | 0.510 (0.378–0.643) | 0.448 | 0.003 | 18.8 | 83.3 | 50.0 | 53.6 |
Low gradient (mean PG <40 mmHg), n = 129 | |||||||
ȃNovel algorithm | 0.668 (0.592–0.744) | <0.001 | 50.0 | 83.7 | 83.3 | 50.6 | |
ȃ2016 recommendation | 0.688 (0.606–0.771) | <0.001 | 0.591 | 66.2 | 71.4 | 79.1 | 56.5 |
ȃ2009 recommendation | 0.572 (0.512–0.632) | 0.018 | 0.033 | 22.5 | 91.8 | 81.8 | 42.1 |
DVR status, n = 152 | |||||||
ȃNovel diagnostic algorithm | 0.758 (0.679–0.836) | <0.001 | 69.7 | 81.8 | 93.3 | 42.9 | |
ȃ2016 recommendation | 0.724 (0.634–0.814) | <0.001 | 0.364 | 78.2 | 66.7 | 89.4 | 45.8 |
ȃ2009 recommendation | 0.649 (0.578–0.721) | <0.001 | 0.032 | 42.0 | 87.9 | 92.6 | 29.6 |
Aortic valve replacement only, n = 41 | |||||||
ȃNovel diagnostic algorithm | 0.754 (0.627–0.880) | <0.001 | 62.5 | 88.2 | 88.2 | 62.5 | |
ȃ2016 recommendation | 0.716 (0.574–0.857) | 0.004 | 0.582 | 66.7 | 76.5 | 80.0 | 61.9 |
ȃ2009 recommendation | 0.575 (0.474–0.676) | 0.097 | 0.014 | 20.8 | 94.1 | 83.3 | 45.7 |
Atrial fibrillation, n = 124 | |||||||
ȃNovel algorithm | 0.724 (0.631–0.817) | <0.001 | 67.0 | 77.8 | 91.5 | 39.6 | |
ȃ2016 recommendation | 0.662 (0.558–0.767) | <0.001 | 0.160 | 73.2 | 59.3 | 86.6 | 38.1 |
ȃ2009 recommendation | 0.620 (0.543–0.697) | 0.008 | 0.068 | 35.1 | 88.9 | 91.9 | 27.6 |
AVR, aortic valve replacement; CW, continuous wave; LVEF, left ventricular ejection fraction; LVOT, left ventricular outflow tract; VTI, velocity-time integral.
Discussion
The main study findings were as follows: (i) patients with PAV obstruction had smaller measured EOA values and smaller PAV sizes than without PAV obstruction; (ii) the measured/reference EOA ratio had an incremental value over the conventional Doppler parameters; (iii) diagnostic models using reference EOA values had higher accuracy than previous model based on the 2009 recommendations;9 and (iv) for large valves, the novel diagnostic algorithm model using the measured EOA/reference EOA ratio had a higher AUC than the previous algorithm based on the 2016 recommendation.11 Using reference EOA values, a more accurate identification of PAV dysfunction may be possible.
Demographic and echocardiographic characteristics of patients with PAV obstruction
In this study, more patients with PAV obstruction were female, had smaller PAV sizes, and more DVR. Most of PAV obstruction was derived from pannus. The detailed mechanism of pannus formation has not been fully demonstrated, but previous studies reported young age, small sized PAV, and DVR were the risk factors for pannus formation and described the inflammatory reaction and changes of subaortic hemodynamics as possible explanation.17,18
Similar to native valve stenosis, most conventional Doppler parameters are applicable to PAV, but some inherent caveats and pitfalls caused by the nature of prosthetic valves exist.19,20 Even a normally functioning PAV has some degree of blood flow obstruction, resulting in an elevated transprosthetic PG. Among all parameters, measured EOA and measured/reference EOA ratio had the highest AUC in this study. The ROC curve–derived optimal cutoff values for each Doppler parameter were similar to previous recommendations.9,11 The AUC of the AT/ET ratio was relatively low compared with that reported by Zekry et al.7 Our study included more patients with atrial fibrillation and a larger study population. Even multiple cardiac cycles were averaged; atrial fibrillation impacted ET and caused inaccuracies.
As the LVOT diameter is squared in the EOA measurement by the CE, small measurement errors are significant in the EOA calculation. In this study, there was no significant difference in the VTILVOT, but the LVOT diameter was significantly lower in the PAV obstruction group. There was significant difference in LVOT diameter between reference size and measurement. As previously reported, substitution of the LVOT diameter with the labelled prosthesis size was found to be inappropriate.21 Circumferential or focal pannus formation may lead to small LVOT size and lower EOA value, without leaflet motion limitations.
Incremental value of reference EOA
Two-dimensional TTE remains the method of choice for evaluating PAV function, but there are unmet needs to be overcome. Measured EOA is a single reliable parameter for detecting PAV obstruction, but it has different cutoff values according to size of PAV. On the other hand, measured/reference EOA ratio had consistent cutoff value, regardless of PAV size. Therefore measured/reference EOA ratio was considered to be more clinically useful. A recent recommendation described the efficacy of comparing the measured EOA to the reference EOA rather than using fixed cut-off values of measured EOA.2,4,12 Use of the reference EOA value is theoretically correct as every PAV has its own hemodynamic characteristics, but there are a few validation studies in real-world practice. As the recent expert consensus for optimal valve prosthesis selection insisted on presenting each prosthesis’s hemodynamic performance uniformly,22 the use of reference values for evaluating PAV performance is expected to increase. Our study showed incremental values of using reference EOA over conventional Doppler parameters. Furthermore, diagnostic algorithm models that included reference EOA was more accurate than the 2009 recommendation, which did not include the reference EOA. In addition, our algorithm showed a higher AUC value for large-sized PAV than 2016 recommendations. Larger-sized PAV had larger reference EOA values and higher standard deviations and might result in wider reference values and might have reduced the diagnostic accuracy in 2016 recommendations. These results probably related with the higher SD values in larger sized PAV.
Limitations
The present study has some limitations. First, it was a retrospective study from a single tertiary center, which inherently limits the generalizability of its results. However, this study had a larger population than previous studies and various subgroups, and the sensitivity analysis showed consistent results. Second, there may be errors in the EOA calculation using the continuity equation. Measuring the LVOT diameter may be challenging in patients with prosthetic valves. Since our study had a high proportion of double-valve replacements, these errors might be magnified. To overcome this limitation, we performed intra-and interobserver variability analyses and demonstrated consistency in the EOA values. However, inter- and intraobserver reproducibility was studied from repeated measurement of same images rather than repeated acquisition and measurement. Therefore, variability might be underestimated. Third, some subgroup analyses included a small number of subjects. However, as we had a larger population, even the small subgroup had similar numbers compared with previous studies. Fourth, although surgical confirmation of the presence of pannus would be the gold standard, not all patients were surgically confirmed in our study. In addition, patients with strong suspicion of PAV obstruction had undergone CT or surgery, which might lead to ascertainment bias. Although the accuracy of CT was validated in previous studies,11 there might be false positives or negatives. However, there were only small numbers of false positive or negative in CT, compared with surgical finding. We also performed a sensitivity analysis using only surgical data, which showed similar results to the main findings.
Conclusion
The ratio of measured/reference EOA has incremental value over conventional Doppler parameters. This parameter might be helpful for distinguishing PAV obstruction. The simple novel diagnostic algorithm model using the measured/reference EOA ratio had higher accuracy than the algorithm from previous recommendations, especially for large-sized PAV.
Supplementary data
Supplementary data are available at European Heart Journal - Cardiovascular Imaging online.
Acknowledgements
MID (Medical Illustration & Design), a part of the Medical Research Support Services of Yonsei University College of Medicine, for providing excellent support with medical illustration.
Funding
None declared.
Data availability
Data would be available upon reasonable request.
References
Author notes
Conflict of interest: None declared.