H E Sostman, K A Manley, A melting-point-of gallium apparatus for thermometer calibration., Clinical Chemistry, Volume 24, Issue 8, 1 August 1978, Pages 1331–1334, https://doi.org/10.1093/clinchem/24.8.1331
- Share Icon Share
Abstract
We have investigated the equilibrium melting point of gallium as a temperature fixed-point at which to calibrate small thermistor thermometers, such as those used to measure temperature in enzyme reaction analysis and other temperature-dependent biological assays. We have determined that the melting temperature of "6N" (99.999% pure) gallium is 29.770 +/- 0.002 degrees C, and that the constant-temperature plateau can be prolonged for several hours. We have designed a simple automated apparatus that exploits this phenomenon and that permits routine calibration verification of thermistor temperature probes throughout the laboratory day. We describe the physics of the gallium melt, and the design and use of the apparatus.