Abstract

Background

Silicone breast implant design has evolved over the last 50 years. Regulatory bodies including the FDA require data to support the modifications designed to improve the safety, efficacy, longevity, and biocompatibility of breast implants.

Objectives

The authors reviewed the 3-year data on the safety and effectiveness of Motiva (Establishment Labs Holdings, Inc., Alajuela, Costa Rica) SmoothSilk silicone gel-filled breast implants submitted to the FDA. The current submitted data include the primary breast augmentation and revisional augmentation cohorts.

Methods

The Motiva IDE is a prospective, single-arm, multicenter, 10-year pivotal study in which data are collected on breast augmentation, reconstruction, and revisional surgery. Three-year data were submitted to the FDA on adverse events, reoperations, patient and physician satisfaction, connective tissue diseases, and quality of life validated instruments. A subset of the patients underwent annual magnetic resonance imaging (MRI) at years 1, 2, and 3 to screen for implant rupture.

Results

A total of 451 patients were implanted in the primary augmentation cohort and 109 patients in the revisional augmentation cohort. There were 218 patients enrolled in the MRI cohort. Reported rates for reoperation for any reason were 6.1% in the primary augmentation cohort (92.4% follow-up) and 25.8% in the revisional augmentation cohort (88.7% follow-up).

Discussion

Motiva implants were first introduced in 2010. The 3-year Motiva data suggests that the leading cause of revisional surgery has shifted from capsular contracture and rupture to more subjective indications for reoperation such as malposition and size change.

Conclusions

Three-year data from the primary augmentation and revisional augmentation cohorts submitted to the FDA demonstrate the safety and efficacy of the Motiva implants. There were low complication rates for implant-related complications and high surgeon and patient satisfaction.

Level of Evidence: 2

graphic

Establishment Labs Holdings, Inc. (Alajuela, Costa Rica) is a global medical technology company. The original prototypes for the Motiva current breast implant technology were developed 20 years ago in Costa Rica. Design and development activities were finalized with the launch of the Motiva Round implants in 2010. In 2011 the company obtained a CE mark, and in 2012 the second generation of Motiva Round implants was launched with changes in the gel and the addition of a new barrier technology (BluSeal) to reduce potential silicone leakage. In 2014, the Motiva Ergonomix implant was introduced. From 2019 to 2023, Establishment Labs was approved to market Motiva implants in more than 85 countries.

The US pivotal study included 2 styles of Motiva breast implants, the Motiva SmoothSilk implant with ProgressiveGel Plus (referred to as round) and the Motiva SmoothSilk Ergonomix ProgressiveGel Ultima implants (referred to as Ergonomix). All devices in the study were filled with the same sourced medical-grade long-term implantable silicone manufactured by Avantor-NuSil Silicone Technology (Carpinteria, CA) but mixed in different ratios to alter their rheological properties. All study implants were manufactured with the SmoothSilk surface and patch and contained a pigmented barrier layer (BluSeal). The pigmented barrier layer was purported to facilitate quality control during manufacture and to allow visual inspection of the implants by surgeons before implantation. The pigment represented less than 0.05% of the barrier layer material (NuSil Silicone Technologies). TrueMonobloc was the proprietary mechanism of bonding the shell, patch, and gel. Investigator surgeons could select implants with or without a microtransponder which would respond with a unique identification number when interrogated externally with a handheld device. This number coulc then be cross-referenced with Establishment Labs records to obtain device-related information as to implant type, surface, manufacturing date, serial number, and, if the device was registered, the date of implantation. This Motiva Q Inside Safety Technology (RFID) consisted of a passive microtransponder that was embedded in the silicone near the patch of the implant. It measured approximately 2 mm × 9 mm and was previously cleared by the FDA with the de Novo process (K033440; 21 CRF §880.6300) as an implantable device.1

Establishment Labs has developed a proprietary method of manufacturing the surface of the Motiva SmoothSilk shell. The shell surface is produced with a 3-dimensional (3D) inverted negative imprinting technology directly on the polydimethylsiloxane material.2 This is described as mandrel imprinting, in which the mandrel's surface architecture is transferred to the silicone during curing and the shell is subsequently turned inside out. In the last few years attention has been focused on shell properties that may affect the complex tissue interactions at the surface, including the ingrowth of soft tissues, the inflammatory response, and the ability of bacteria to adhere to the surface of the device.3-5 The SmoothSilk surface of the implant has a uniform surface with an average roughness of 4 microns that can be considered a modification of a smooth surface (average roughness of approximately 1 micron). The International Organization for Standardization (ISO) is currently updating their surface classification to include surface complexity. This classification scheme is based on the texturing technique, surface roughness, surface complexity (measure of added surface area compared to completely smooth surface), and pore size of the breast implant outer surface, as measured by scanning electron microscopy (SEM; Appendix, located online at www.aestheticsurgeryjournal.com).

METHODS

The investigational device exemption (IDE) for the Motiva study on the safety and effectiveness of the Motiva silicone SmoothSilk gel-filled breast implants in patients undergoing primary breast augmentation, primary breast reconstruction, and revisional surgery was submitted to the FDA July 2017 and approved in March 2018. The study design was posted on clinicaltrials.gov (NCT03579901) in July 2018 and last updated on March 27, 2023. All patients signed an informed consent, and the study was designed following the principles of the Declaration of Helsinki. The study was IRB approved in 2018 through the Center for IRB Intelligence (CIRBI), Advarra, Columbia, MD. All approved investigators and their sites were required to notify the IRB of any unanticipated problems involving risks to patients or others, adverse device effects, and protocol deviations that might affect patient rights or the completeness and accuracy of the results. Twenty-five investigator sites and investigators underwent site training before patient enrollment, and the first patient was enrolled in the study on April 17, 2018. The study included patients aged 22 years and over for primary breast augmentation to increase breast size and revisional procedures to correct or improve the results of a previous breast augmentation procedure. Study indications for breast reconstruction included the replacement of breast tissue removed due to cancer, trauma, or severe breast abnormality and revisional surgery to correct or improve the original reconstructive procedure.

Study enrollment was completed in the augmentation cohorts on July 26, 2019, with a total of 451 patients implanted in the primary augmentation cohort and 109 patients in the revisional augmentation cohort. As part of the Motiva trial, a magnetic resonance imaging (MRI) substudy was designed to include up to 250 patients who would undergo a breast MRI scan at 1, 2, 3, 5, 7, and 10 years post surgery. MRI designated sites were selected in close proximity to the investigator's office to further encourage follow-up. Participation in the MRI subgroup was offered to all eligible patients, with compensation for their time at fair market value (FMV). A subgroup of 218 patients from the primary and revisional augmentation cohorts was enrolled. Patients could be discontinued from the MRI cohort due to claustrophobia concerns or a patient's decision to discontinue while remaining enrolled in the larger study.

Clinical data were collected at each investigator site on standardized paper case report forms at baseline, 3 to 6 weeks, and scheduled annual yearly follow-up visits. Data underwent double data entry in validated electronic data capture. Optional virtual patient monitoring became available in April 2020, a necessary modification of the protocol that allowed patients and investigators to complete annual patient follow-up during the peak of the COVID-19 pandemic.6

Patient demographic data were collected, along with previous lactation history, menopausal status, and history of connective tissue diseases (CTD). CTD signs and symptoms were recorded at each scheduled time point to evaluate if prevalence increased with exposure to breast implants. Self-reported measures of health were recorded with the SF-36v2 Health Survey (2000). The BREAST-Q (version 2.0) was added to the protocol in July 2021 and will continue to collect data through 10 years.

A screening baseline mammogram was required for all patients over the age of 35 before study enrollment. Any postoperative mammography results were evaluated, and Kaplan-Meier analysis estimated the cumulative incidence of new postoperative abnormal mammograms (defined as BI-RADS 4 suspicious abnormality). In addition to standardized photography at baseline and all follow-up visits, 3D imaging (Vectra XT 3D Imaging System, Canfield Scientific, Parsippany, NJ; Divina 3-D Imaging System, Establishment Labs, Alajuela, Costa Rica) was an additional study requirement. The Vectra 3D scans continue to be required on all patients in the primary augmentation cohort, while the optional Divina was discontinued in 2020.

The study investigators were able to select from the 2 styles of Motiva implants available, the SmoothSilk Ergonomix and the SmoothSilk Round, in consultation with their patients, and determined the optimal style, size, and fill for each patient (Figure 1). Surgical procedural data collected on the day of surgery included incision approach, pocket location, the use of funnels, placement of drains, the device style and volume, and specifics of pocket irrigation.

(A) Motiva SmoothSilk Round and (B) Motiva Ergonomix implants.
Figure 1.

(A) Motiva SmoothSilk Round and (B) Motiva Ergonomix implants.

To maintain high patient compliance throughout the 10-year study, patients receive compensation at FMV for their annual scheduled follow-ups. Financial incentives have been shown to create higher patient retention in both device and pharmaceutical clinical trials.7 Patients who were explanted without reimplantation were asked to remain in the study and continue their annual follow-up visits out to 10 years. The MRI subpopulation included patients in both augmentation cohorts, with MRI scheduled at years 1, 2, 3, 5, 7, and 10 years for the analysis of rupture. In addition, a subgroup of patients were further defined by the presence or absence of the microtransponder for safety and effectiveness.

Data Collection and Statistical Analysis

Safety Analysis

The safety population included all patients implanted with a Motiva device. Analysis was conducted by cohort and performed at both the patient and implant levels. Safety assessment included Kaplan-Meier analysis of any patient complications, adverse events, and reoperations with 95% confidence intervals. Additional gathered information included analysis of rupture, CTD signs and symptoms, breast imaging including mammography and ultrasound when indicated, breast cancer, and breast implant–associated anaplastic large cell lymphoma (BIA-ALCL). CTD signs and symptoms were not collected at baseline before implantation, but patients with previously diagnosed CTD were excluded from the study. A Cox regression analysis was performed for complications, reoperations, and adverse events.

Explanatory variables included age, race, smoking status, surgical approach, surgical placement, incision size, irrigation solutions (including antibiotic irrigation and Betadine [Purdue Pharma, Stamford, CT]), pocket location, implant volume, and type of implant selected.

Effectiveness Analyses

Effectiveness was based on a performance goal, with the endpoint measured on a 5-point Likert scale rating the degree of patient satisfaction or dissatisfaction. Quality of life assessments and surgeon satisfaction with outcome were calculated at each time point. The BREAST-Q was added to the effectiveness analysis in 2021 and therefore does not capture all patients at baseline enrolled in the augmentation cohorts.

Monitoring During COVID Pandemic

In 2020, investigational sites were faced with unprecedented challenges due to the COVID pandemic and subsequent restrictions placed upon the healthcare system that varied from state to state. A virtual clinical trial visit plan was submitted to the FDA in March 2020 and promptly accepted; this permitted video conferencing for patients and study sites in lockdown. Sites were instructed to adhere to virtual platforms that were HIPAA compliant. Data points that were unable to be collected during virtual follow-up visits and/or procedures that were not performed due to the virtual nature of the visit (digital photography, 3D digital imaging, MRI, and in some instances measurements) were reported in the source documentation. These deviations were not considered to increase the risks or affect the well-being of patients and data points.

RESULTS

Demographics, Follow-up by Cohort, and Procedural Data

Final enrollment included 451 primary augmentation and 109 revisional augmentation female patients, with a total enrollment of 560 patients. Patient compliance with annual follow-up through 3 years was 92.4% in the primary augmentation cohort and 88.7% in the revisional augmentation cohort (Supplemental Tables 1, 2, located online at www.aestheticsurgeryjournal.com).

The age range for primary augmentation patients was 22 to 69 years (mean, 33.5) and for revisional augmentation 22 to 69 years (mean 44.5). Most patients were White, had a college education or higher degree, and were married. (Supplemental Table 3, located online at www.aestheticsurgeryjournal.com).

Both SmoothSilk Ergonomix and SmoothSilk Round implants were included in the device trial, and investigators selected SmoothSilk Ergonomix implants more often in both the primary augmentation cohort (88.7%) and revisional augmentation cohort (87.2%). Surgeons were also given the choice of whether or not to select implants with a radiofrequency identification device (RFID). In the primary augmentation cohort, 30% of implants included the RFID, and in the revisional augmentation cohort 17.4% of implants included the RFID. Implant volumes ranged from 105 to 1050 cc, with a mean volume of 331 cc in the primary augmentation cohort and 409 cc in the revisional augmentation cohort.

The most common incision site for implant insertion in the primary augmentation cohort was the inframammary fold (IMF) (85.3%), followed by transaxillary (9.5%). In the revisional augmentation cohort, an inframammary approach was performed most often (81.7%), followed by periareolar (10.1%), mastopexy (6.4%), and transaxillary (1.8%). In the augmentation cohort, the most frequent pocket location was partial submuscular (dual-plane) (94.6%), and subglandular placement was rare (2.2%). The revisional augmentation cohort surgeons also favored the partial submuscular (dual-plane) (78.9%), but subglandular placement was less uncommon (17.4%).

Comprehensive data collected on the day of surgery included details of pocket irrigation. Some combination of pocket irrigation was employed by 93.8% of investigator surgeons in the primary augmentation cohort and 100% of the surgeons in the revisional augmentation cohort. The most common irrigation solution was the combination of Betadine and triple antibiotics in both the primary augmentation cohort (55.4%) and the revisional augmentation cohort (63.3%). Betadine mixed with saline was the next most frequent irrigation solution in both the primary augmentation cohort (17.1%) and the revisional augmentation cohort (18.3%). Other combinations of antibiotics and saline were utilized in the primary augmentation cohort (5.5%) and revisional augmentation cohort (4.6%). Normal saline alone was an irrigant in the primary augmentation cohort (0.7%) and in the revisional augmentation cohort (2.8%). Drains were rarely placed in either the primary (2%) or revisional (14%) augmentation cohorts. Funnels for implant insertion were also documented for primary augmentation (72%) and for revisional augmentation (61%) (Tables 1, 2).

Table 1.

Operative Characteristics (Procedural Data) of Primary and Revisional Augmentation Cohorts by Implant

Surgical characteristicsPrimary augmentation (n = 901)Revisional augmentation (n = 218)Overall augmentation (n = 1119)
Device Style
 Round102 (11.3%)28 (12.8%)130 (11.6%)
 Ergonomix799 (88.7%)190 (87.2%)989 (88.4%)
Radiofrequency identification device (RFID)
 With RFID270 (30.0%)38 (17.4%)308 (27%)
 Without RFID631 (70%)180 (82.6%)811 (72.5%)
Implant volume
 105-300 cc378 (42%)58 (26.6%)436 (39.0%)
 305-500 cc502 (55.7%)114 (52.3%)616 (55.0%)
 505-800 cc21 (2.3%)40 (18.3%)61 (5.5%)
 805-1050 cc06 (2.8%)6 (0.5%)
Incision site
 Inframammary769 (85.3%)178 (81.7%)947 (84.6%)
 Mastopexy20 (2.2%)14 (6.4%)34 (3.0%0
 Periareolar26 (2.9%)22 (10.1%)48 (4.3%)
 Transaxillary86 (9.5%)4 (1.8%)90 (8.0%)
Placement
 Complete muscle coverage24 (2.7%)4 (1.8%)28 (2.5%)
 Partial submuscular-dual plane852 (94.6%)172 (78.9%)1024 (91.5%)
 Subfascial5 (0.6%)4.0 (1.8%)9 (0.8%)
 Subglandular20 (2.2%)38 (17.4%)58 (5.2%)
Incision length (cm)
 Mean (SD)4.2 (1.69)5.7 (3.14)4.5 (2.12)
 Median (min-max)4.0 (2-17)5.0 (3-30)4.0 (2-30)
Pocket irrigationa
 No pocket irrigation28 (6.2%)028 (5.0%)
 At least 1 irrigation423 (93.8%)109 (100%)532 (95%)
 Antibiotics only8 (1.8%)08 (1.4%)
 Antibiotics/saline25 (5.5%)5 (4.6%)30 (5.4%)
 Antibiotics/Betadine/saline250 (55.4%)69 (63.3%)319 (57.0%)
 Betadine/saline77 (17.1%)20 (18.3%)97 (17.3%)
 Other antiseptic (not Betadine)/saline44 (9.8%)1 (0.9%)45 (8.0%)
 Other antiseptic only (not Betadine)3 (0.7%)03 (.5%)
 Saline only3 (0.7%)3 (2.8%)6 (1.1%)
Surgical characteristicsPrimary augmentation (n = 901)Revisional augmentation (n = 218)Overall augmentation (n = 1119)
Device Style
 Round102 (11.3%)28 (12.8%)130 (11.6%)
 Ergonomix799 (88.7%)190 (87.2%)989 (88.4%)
Radiofrequency identification device (RFID)
 With RFID270 (30.0%)38 (17.4%)308 (27%)
 Without RFID631 (70%)180 (82.6%)811 (72.5%)
Implant volume
 105-300 cc378 (42%)58 (26.6%)436 (39.0%)
 305-500 cc502 (55.7%)114 (52.3%)616 (55.0%)
 505-800 cc21 (2.3%)40 (18.3%)61 (5.5%)
 805-1050 cc06 (2.8%)6 (0.5%)
Incision site
 Inframammary769 (85.3%)178 (81.7%)947 (84.6%)
 Mastopexy20 (2.2%)14 (6.4%)34 (3.0%0
 Periareolar26 (2.9%)22 (10.1%)48 (4.3%)
 Transaxillary86 (9.5%)4 (1.8%)90 (8.0%)
Placement
 Complete muscle coverage24 (2.7%)4 (1.8%)28 (2.5%)
 Partial submuscular-dual plane852 (94.6%)172 (78.9%)1024 (91.5%)
 Subfascial5 (0.6%)4.0 (1.8%)9 (0.8%)
 Subglandular20 (2.2%)38 (17.4%)58 (5.2%)
Incision length (cm)
 Mean (SD)4.2 (1.69)5.7 (3.14)4.5 (2.12)
 Median (min-max)4.0 (2-17)5.0 (3-30)4.0 (2-30)
Pocket irrigationa
 No pocket irrigation28 (6.2%)028 (5.0%)
 At least 1 irrigation423 (93.8%)109 (100%)532 (95%)
 Antibiotics only8 (1.8%)08 (1.4%)
 Antibiotics/saline25 (5.5%)5 (4.6%)30 (5.4%)
 Antibiotics/Betadine/saline250 (55.4%)69 (63.3%)319 (57.0%)
 Betadine/saline77 (17.1%)20 (18.3%)97 (17.3%)
 Other antiseptic (not Betadine)/saline44 (9.8%)1 (0.9%)45 (8.0%)
 Other antiseptic only (not Betadine)3 (0.7%)03 (.5%)
 Saline only3 (0.7%)3 (2.8%)6 (1.1%)

aSum of individual irrigation types may be greater than total because a single patient may have multiple irrigants. Max, maximum; min, minimum; SD, standard deviation.

Table 1.

Operative Characteristics (Procedural Data) of Primary and Revisional Augmentation Cohorts by Implant

Surgical characteristicsPrimary augmentation (n = 901)Revisional augmentation (n = 218)Overall augmentation (n = 1119)
Device Style
 Round102 (11.3%)28 (12.8%)130 (11.6%)
 Ergonomix799 (88.7%)190 (87.2%)989 (88.4%)
Radiofrequency identification device (RFID)
 With RFID270 (30.0%)38 (17.4%)308 (27%)
 Without RFID631 (70%)180 (82.6%)811 (72.5%)
Implant volume
 105-300 cc378 (42%)58 (26.6%)436 (39.0%)
 305-500 cc502 (55.7%)114 (52.3%)616 (55.0%)
 505-800 cc21 (2.3%)40 (18.3%)61 (5.5%)
 805-1050 cc06 (2.8%)6 (0.5%)
Incision site
 Inframammary769 (85.3%)178 (81.7%)947 (84.6%)
 Mastopexy20 (2.2%)14 (6.4%)34 (3.0%0
 Periareolar26 (2.9%)22 (10.1%)48 (4.3%)
 Transaxillary86 (9.5%)4 (1.8%)90 (8.0%)
Placement
 Complete muscle coverage24 (2.7%)4 (1.8%)28 (2.5%)
 Partial submuscular-dual plane852 (94.6%)172 (78.9%)1024 (91.5%)
 Subfascial5 (0.6%)4.0 (1.8%)9 (0.8%)
 Subglandular20 (2.2%)38 (17.4%)58 (5.2%)
Incision length (cm)
 Mean (SD)4.2 (1.69)5.7 (3.14)4.5 (2.12)
 Median (min-max)4.0 (2-17)5.0 (3-30)4.0 (2-30)
Pocket irrigationa
 No pocket irrigation28 (6.2%)028 (5.0%)
 At least 1 irrigation423 (93.8%)109 (100%)532 (95%)
 Antibiotics only8 (1.8%)08 (1.4%)
 Antibiotics/saline25 (5.5%)5 (4.6%)30 (5.4%)
 Antibiotics/Betadine/saline250 (55.4%)69 (63.3%)319 (57.0%)
 Betadine/saline77 (17.1%)20 (18.3%)97 (17.3%)
 Other antiseptic (not Betadine)/saline44 (9.8%)1 (0.9%)45 (8.0%)
 Other antiseptic only (not Betadine)3 (0.7%)03 (.5%)
 Saline only3 (0.7%)3 (2.8%)6 (1.1%)
Surgical characteristicsPrimary augmentation (n = 901)Revisional augmentation (n = 218)Overall augmentation (n = 1119)
Device Style
 Round102 (11.3%)28 (12.8%)130 (11.6%)
 Ergonomix799 (88.7%)190 (87.2%)989 (88.4%)
Radiofrequency identification device (RFID)
 With RFID270 (30.0%)38 (17.4%)308 (27%)
 Without RFID631 (70%)180 (82.6%)811 (72.5%)
Implant volume
 105-300 cc378 (42%)58 (26.6%)436 (39.0%)
 305-500 cc502 (55.7%)114 (52.3%)616 (55.0%)
 505-800 cc21 (2.3%)40 (18.3%)61 (5.5%)
 805-1050 cc06 (2.8%)6 (0.5%)
Incision site
 Inframammary769 (85.3%)178 (81.7%)947 (84.6%)
 Mastopexy20 (2.2%)14 (6.4%)34 (3.0%0
 Periareolar26 (2.9%)22 (10.1%)48 (4.3%)
 Transaxillary86 (9.5%)4 (1.8%)90 (8.0%)
Placement
 Complete muscle coverage24 (2.7%)4 (1.8%)28 (2.5%)
 Partial submuscular-dual plane852 (94.6%)172 (78.9%)1024 (91.5%)
 Subfascial5 (0.6%)4.0 (1.8%)9 (0.8%)
 Subglandular20 (2.2%)38 (17.4%)58 (5.2%)
Incision length (cm)
 Mean (SD)4.2 (1.69)5.7 (3.14)4.5 (2.12)
 Median (min-max)4.0 (2-17)5.0 (3-30)4.0 (2-30)
Pocket irrigationa
 No pocket irrigation28 (6.2%)028 (5.0%)
 At least 1 irrigation423 (93.8%)109 (100%)532 (95%)
 Antibiotics only8 (1.8%)08 (1.4%)
 Antibiotics/saline25 (5.5%)5 (4.6%)30 (5.4%)
 Antibiotics/Betadine/saline250 (55.4%)69 (63.3%)319 (57.0%)
 Betadine/saline77 (17.1%)20 (18.3%)97 (17.3%)
 Other antiseptic (not Betadine)/saline44 (9.8%)1 (0.9%)45 (8.0%)
 Other antiseptic only (not Betadine)3 (0.7%)03 (.5%)
 Saline only3 (0.7%)3 (2.8%)6 (1.1%)

aSum of individual irrigation types may be greater than total because a single patient may have multiple irrigants. Max, maximum; min, minimum; SD, standard deviation.

Table 2.

Operative Characteristics (Procedural Data) of Primary and Revisional Augmentation Cohorts by Patient

Surgical characteristicsPrimary augmentation (n = 451)Revisional augmentation (n = 109)Overall augmentation (n = 560)
Funnel use during surgery
 Funnel325 (72.1%)66 (60.6%)391 (69.8%)
 No funnel126 (27.9%)43 (39.4%)169 (30.2%)
Drain placement during surgery
 Both breasts7 (1.6%)15 (13.8%)22 (3.9%)
 Right breast03 (2.8%)3 (0.5%)
 Left breast01 (0.9%)1 (0.2%)
 No drains444 (98.4%)90 (82.6%)534 (95.4%)
Surgical characteristicsPrimary augmentation (n = 451)Revisional augmentation (n = 109)Overall augmentation (n = 560)
Funnel use during surgery
 Funnel325 (72.1%)66 (60.6%)391 (69.8%)
 No funnel126 (27.9%)43 (39.4%)169 (30.2%)
Drain placement during surgery
 Both breasts7 (1.6%)15 (13.8%)22 (3.9%)
 Right breast03 (2.8%)3 (0.5%)
 Left breast01 (0.9%)1 (0.2%)
 No drains444 (98.4%)90 (82.6%)534 (95.4%)
Table 2.

Operative Characteristics (Procedural Data) of Primary and Revisional Augmentation Cohorts by Patient

Surgical characteristicsPrimary augmentation (n = 451)Revisional augmentation (n = 109)Overall augmentation (n = 560)
Funnel use during surgery
 Funnel325 (72.1%)66 (60.6%)391 (69.8%)
 No funnel126 (27.9%)43 (39.4%)169 (30.2%)
Drain placement during surgery
 Both breasts7 (1.6%)15 (13.8%)22 (3.9%)
 Right breast03 (2.8%)3 (0.5%)
 Left breast01 (0.9%)1 (0.2%)
 No drains444 (98.4%)90 (82.6%)534 (95.4%)
Surgical characteristicsPrimary augmentation (n = 451)Revisional augmentation (n = 109)Overall augmentation (n = 560)
Funnel use during surgery
 Funnel325 (72.1%)66 (60.6%)391 (69.8%)
 No funnel126 (27.9%)43 (39.4%)169 (30.2%)
Drain placement during surgery
 Both breasts7 (1.6%)15 (13.8%)22 (3.9%)
 Right breast03 (2.8%)3 (0.5%)
 Left breast01 (0.9%)1 (0.2%)
 No drains444 (98.4%)90 (82.6%)534 (95.4%)

There were 218 patients enrolled overall in the MRI substudy, 176 patients (352 implants) in the primary augmentation cohort and 42 patients (84 implants) in the revisional augmentation cohort. At year 3, 25 patients exited the study due to claustrophobia or other patient decisions, with 88.2% follow-up. At year 3 there were 1 confirmed rupture in the primary augmentation cohort and 0 confirmed ruptures in the revisional augmentation cohort.

Complications and Reoperations

Kaplan-Meier risk analysis at 3 years for local breast complications was reported by patient and by implant. In the primary augmentation cohort, reported breast complications at 3 years included size change (6.9%), implant malposition (3.2%), infection (0.9%), hematoma (0.7%), breast pain (0.7%), suspected or confirmed implant rupture (0.6%), capsular contracture grade III or IV (0.5%), glandular ptosis (0.2%), animation deformity (0.2%), and asymmetry (0.2%). In the revisional augmentation cohort, reported complications at 3 years included size change (20.7%), capsular contracture grade III or IV (6.7%), implant malposition (4.9%), glandular ptosis (4.8%), asymmetry (3.9%), hematoma (1.8%), and infection (0.9%). Explantation without replacement occurred in 1 patient (0.2%) in the primary augmentation cohort and 3 patients (2.9%) in the revisional augmentation cohort. There was 1 confirmed rupture in the primary augmentation cohort, and there were no late seromas (Table 3).

Table 3.

Kaplan-Meyer Risk Analysis at Year 3 by Patient

EventPrimary augmentation (n = 451)Revisional augmentation (n = 109)
Reoperation27 (6.1%)27 (25.8%)
Explantation7 (1.6%)17 (16.5%)
Explantation with replacement6 (1.4%)14 (13.8%)
Explantation without replacement1 (0.2%)3 (2.9%)
Number of patients with at least 1 reoperation27 (6.0%)27 (24.8%)
Risk of any complicationa37 (8.4%)30 (28.4%)
Breast pain3 (0.7%)1 (1.0%)
Animation deformity1 (0.2%)0
Asymmetry1 (0.2%)4 (3.9%)
Breast cancer (new)00
Capsular contracture grade II with surgical intervention1 (0.2%)2 (1.9%)
Capsular contracture III/IV2 (0.5%)7 (6.7%)
BIA-ALCL00
Double capsule01 (1.0%)
Infection4 (0.9%)1 (0.9%)
Rupture suspected/confirmed1 (0.6%)0
Size change2 (6.9%)6 (20.7%)
Delayed wound healing1 (0.2%)0
Hematoma3 (0.7%)2 (1.8%)
Iatrogenic injury to implant01 (1.0%)
Implant extrusion01 (1.0%)
Malposition14 (3.2)5 (4.9%)
Palpability/visibility1 (0.2%)0
Breast mass/cyst/lump1 (0.2%)2 (2.4%)
Ptosis1 (0.2%)5 (4.8%)
Nipple complication1 (0.2%)0
Skin rash1 (0.2%)0
Seroma (late >1 year)00
Wrinkling/rippling2 (0.5%)0
EventPrimary augmentation (n = 451)Revisional augmentation (n = 109)
Reoperation27 (6.1%)27 (25.8%)
Explantation7 (1.6%)17 (16.5%)
Explantation with replacement6 (1.4%)14 (13.8%)
Explantation without replacement1 (0.2%)3 (2.9%)
Number of patients with at least 1 reoperation27 (6.0%)27 (24.8%)
Risk of any complicationa37 (8.4%)30 (28.4%)
Breast pain3 (0.7%)1 (1.0%)
Animation deformity1 (0.2%)0
Asymmetry1 (0.2%)4 (3.9%)
Breast cancer (new)00
Capsular contracture grade II with surgical intervention1 (0.2%)2 (1.9%)
Capsular contracture III/IV2 (0.5%)7 (6.7%)
BIA-ALCL00
Double capsule01 (1.0%)
Infection4 (0.9%)1 (0.9%)
Rupture suspected/confirmed1 (0.6%)0
Size change2 (6.9%)6 (20.7%)
Delayed wound healing1 (0.2%)0
Hematoma3 (0.7%)2 (1.8%)
Iatrogenic injury to implant01 (1.0%)
Implant extrusion01 (1.0%)
Malposition14 (3.2)5 (4.9%)
Palpability/visibility1 (0.2%)0
Breast mass/cyst/lump1 (0.2%)2 (2.4%)
Ptosis1 (0.2%)5 (4.8%)
Nipple complication1 (0.2%)0
Skin rash1 (0.2%)0
Seroma (late >1 year)00
Wrinkling/rippling2 (0.5%)0

aDoes not include complications assessed as mild severity, except for capsular contracture III/IV, implant extrusion, and suspected or confirmed rupture. BIA-ALCL, breast implant–associated anaplastic large cell lymphoma.

Table 3.

Kaplan-Meyer Risk Analysis at Year 3 by Patient

EventPrimary augmentation (n = 451)Revisional augmentation (n = 109)
Reoperation27 (6.1%)27 (25.8%)
Explantation7 (1.6%)17 (16.5%)
Explantation with replacement6 (1.4%)14 (13.8%)
Explantation without replacement1 (0.2%)3 (2.9%)
Number of patients with at least 1 reoperation27 (6.0%)27 (24.8%)
Risk of any complicationa37 (8.4%)30 (28.4%)
Breast pain3 (0.7%)1 (1.0%)
Animation deformity1 (0.2%)0
Asymmetry1 (0.2%)4 (3.9%)
Breast cancer (new)00
Capsular contracture grade II with surgical intervention1 (0.2%)2 (1.9%)
Capsular contracture III/IV2 (0.5%)7 (6.7%)
BIA-ALCL00
Double capsule01 (1.0%)
Infection4 (0.9%)1 (0.9%)
Rupture suspected/confirmed1 (0.6%)0
Size change2 (6.9%)6 (20.7%)
Delayed wound healing1 (0.2%)0
Hematoma3 (0.7%)2 (1.8%)
Iatrogenic injury to implant01 (1.0%)
Implant extrusion01 (1.0%)
Malposition14 (3.2)5 (4.9%)
Palpability/visibility1 (0.2%)0
Breast mass/cyst/lump1 (0.2%)2 (2.4%)
Ptosis1 (0.2%)5 (4.8%)
Nipple complication1 (0.2%)0
Skin rash1 (0.2%)0
Seroma (late >1 year)00
Wrinkling/rippling2 (0.5%)0
EventPrimary augmentation (n = 451)Revisional augmentation (n = 109)
Reoperation27 (6.1%)27 (25.8%)
Explantation7 (1.6%)17 (16.5%)
Explantation with replacement6 (1.4%)14 (13.8%)
Explantation without replacement1 (0.2%)3 (2.9%)
Number of patients with at least 1 reoperation27 (6.0%)27 (24.8%)
Risk of any complicationa37 (8.4%)30 (28.4%)
Breast pain3 (0.7%)1 (1.0%)
Animation deformity1 (0.2%)0
Asymmetry1 (0.2%)4 (3.9%)
Breast cancer (new)00
Capsular contracture grade II with surgical intervention1 (0.2%)2 (1.9%)
Capsular contracture III/IV2 (0.5%)7 (6.7%)
BIA-ALCL00
Double capsule01 (1.0%)
Infection4 (0.9%)1 (0.9%)
Rupture suspected/confirmed1 (0.6%)0
Size change2 (6.9%)6 (20.7%)
Delayed wound healing1 (0.2%)0
Hematoma3 (0.7%)2 (1.8%)
Iatrogenic injury to implant01 (1.0%)
Implant extrusion01 (1.0%)
Malposition14 (3.2)5 (4.9%)
Palpability/visibility1 (0.2%)0
Breast mass/cyst/lump1 (0.2%)2 (2.4%)
Ptosis1 (0.2%)5 (4.8%)
Nipple complication1 (0.2%)0
Skin rash1 (0.2%)0
Seroma (late >1 year)00
Wrinkling/rippling2 (0.5%)0

aDoes not include complications assessed as mild severity, except for capsular contracture III/IV, implant extrusion, and suspected or confirmed rupture. BIA-ALCL, breast implant–associated anaplastic large cell lymphoma.

Patients with a history of CTD or rheumatologic illness were excluded from enrollment. Enrolled patients were asked to document any new onset of CTD or rheumatologic illness or symptoms at each study time point. There were no new reported cases of either CTD or rheumatologic illness in either the primary or revisional augmentation cohorts (Table 4).

Table 4.

Risk of Nonlocal Complications: Connective Tissue Diseases and Rheumatologic Illness

Complication/Time pointNumber of events (cumulative)Kaplan-Meier risk95% CI
Primary breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Revisional breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Complication/Time pointNumber of events (cumulative)Kaplan-Meier risk95% CI
Primary breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Revisional breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)

aRheumatologic and connective tissue diseases include rheumatoid arthritis, systemic lupus erythematosus, discoid lupus, scleroderma, vasculitis, dermatomyositis, Raynaud's phenomenon, Sjogren's syndrome, CREST syndrome, carpal tunnel syndrome, multiple sclerosis–like syndrome, multiple myeloma–like syndrome, chronic fatigue syndrome, fibromyalgia. CI, confidence interval.

Table 4.

Risk of Nonlocal Complications: Connective Tissue Diseases and Rheumatologic Illness

Complication/Time pointNumber of events (cumulative)Kaplan-Meier risk95% CI
Primary breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Revisional breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Complication/Time pointNumber of events (cumulative)Kaplan-Meier risk95% CI
Primary breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Revisional breast augmentation
Rheumatic disease, anya
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Rheumatoid arthritis
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Systemic lupus
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)
Other rheumatologic
 Week 60 (0)
 Year 10 (0)
 Year 20 (0)
 Year 30 (0)

aRheumatologic and connective tissue diseases include rheumatoid arthritis, systemic lupus erythematosus, discoid lupus, scleroderma, vasculitis, dermatomyositis, Raynaud's phenomenon, Sjogren's syndrome, CREST syndrome, carpal tunnel syndrome, multiple sclerosis–like syndrome, multiple myeloma–like syndrome, chronic fatigue syndrome, fibromyalgia. CI, confidence interval.

At 3 years there were 6 patients discontinued from the primary augmentation cohort: 1 unrelated death, 1 patient explanted and replaced with a nonstudy device, 2 requested to be discontinued, and 2 were lost to follow-up. In the revisional augmentation cohort, there were 5 patients discontinued: 1 unrelated death, 2 explanted with nonstudy device replacement, 2 requested to be discontinued, and 0 were lost to follow-up.

In the primary augmentation cohort, there were 29 reoperations in 27 patients (6.0%) at 3 years. The indications for explantation with or without replacement for these 29 reoperations included reoperation for malposition in 13 patients (44.8%), grade III/IV capsular contracture in 3 patients (10.3%), size change in 2 patients (6.9%), and infection in 2 patients (6.9%). In the revisional augmentation cohort, there were 29 reoperations in 27 patients (24.8%) over the 3-year period. The indications for these 29 reoperations included revision for size change in 6 patients (20.7%), grade III/IV capsular contracture in 6 patients (20.7%), glandular ptosis in 5 patients (17.2%), and hematoma in 4 patients (13.8%). Animation deformity as a cause for reoperation was rare, occurring in 1 patient in the primary augmentation cohort (Table 5).

Table 5.

Primary Reason for Reoperation

Primary reason for reoperationaPrimary augmentation
(n = 29)
Revisional augmentation
(n = 29)
Overall augmentation
(n = 58)
Infection2 (6.9%)1 (3.4%)3 (5.2%)
Capsular contracture3 (10.3%)6 (20.7%)9 (15.5%)
Implant extrusion01 (3.4%)1 (1.7%)
Hematoma3 (10.3%)4 (13.8%)7 (12.1%)
Breast pain01 (3.4%)1 (1.7%)
Implant malposition13 (44.8%)2 (6.9%)15 (25.9%)
Upper pole fullness01 (3.4%)1 (1.7%)
Animation deformity1 (3.4%)01 (1.7%)
Asymmetry1 (3.4%)1 (3.4%)2 (3.4%)
Ptosis1 (3.4%)5 (17.2%)6 (10.3%)
Hypertrophic/abnormal scarring2 (6.9%)1 (3.4%)3 (5.2%)
Mass/cyst/lump1 (3.4%)01 (1.7%)
Size change/patient choice2 (6.9%)6 (20.7%)8 (13.8%)
Primary reason for reoperationaPrimary augmentation
(n = 29)
Revisional augmentation
(n = 29)
Overall augmentation
(n = 58)
Infection2 (6.9%)1 (3.4%)3 (5.2%)
Capsular contracture3 (10.3%)6 (20.7%)9 (15.5%)
Implant extrusion01 (3.4%)1 (1.7%)
Hematoma3 (10.3%)4 (13.8%)7 (12.1%)
Breast pain01 (3.4%)1 (1.7%)
Implant malposition13 (44.8%)2 (6.9%)15 (25.9%)
Upper pole fullness01 (3.4%)1 (1.7%)
Animation deformity1 (3.4%)01 (1.7%)
Asymmetry1 (3.4%)1 (3.4%)2 (3.4%)
Ptosis1 (3.4%)5 (17.2%)6 (10.3%)
Hypertrophic/abnormal scarring2 (6.9%)1 (3.4%)3 (5.2%)
Mass/cyst/lump1 (3.4%)01 (1.7%)
Size change/patient choice2 (6.9%)6 (20.7%)8 (13.8%)

aIf multiple reasons were provided, then primary was assigned in the order that it appears in the table. Primary reasons that were not reported (eg, rupture) do not appear on this table.

Table 5.

Primary Reason for Reoperation

Primary reason for reoperationaPrimary augmentation
(n = 29)
Revisional augmentation
(n = 29)
Overall augmentation
(n = 58)
Infection2 (6.9%)1 (3.4%)3 (5.2%)
Capsular contracture3 (10.3%)6 (20.7%)9 (15.5%)
Implant extrusion01 (3.4%)1 (1.7%)
Hematoma3 (10.3%)4 (13.8%)7 (12.1%)
Breast pain01 (3.4%)1 (1.7%)
Implant malposition13 (44.8%)2 (6.9%)15 (25.9%)
Upper pole fullness01 (3.4%)1 (1.7%)
Animation deformity1 (3.4%)01 (1.7%)
Asymmetry1 (3.4%)1 (3.4%)2 (3.4%)
Ptosis1 (3.4%)5 (17.2%)6 (10.3%)
Hypertrophic/abnormal scarring2 (6.9%)1 (3.4%)3 (5.2%)
Mass/cyst/lump1 (3.4%)01 (1.7%)
Size change/patient choice2 (6.9%)6 (20.7%)8 (13.8%)
Primary reason for reoperationaPrimary augmentation
(n = 29)
Revisional augmentation
(n = 29)
Overall augmentation
(n = 58)
Infection2 (6.9%)1 (3.4%)3 (5.2%)
Capsular contracture3 (10.3%)6 (20.7%)9 (15.5%)
Implant extrusion01 (3.4%)1 (1.7%)
Hematoma3 (10.3%)4 (13.8%)7 (12.1%)
Breast pain01 (3.4%)1 (1.7%)
Implant malposition13 (44.8%)2 (6.9%)15 (25.9%)
Upper pole fullness01 (3.4%)1 (1.7%)
Animation deformity1 (3.4%)01 (1.7%)
Asymmetry1 (3.4%)1 (3.4%)2 (3.4%)
Ptosis1 (3.4%)5 (17.2%)6 (10.3%)
Hypertrophic/abnormal scarring2 (6.9%)1 (3.4%)3 (5.2%)
Mass/cyst/lump1 (3.4%)01 (1.7%)
Size change/patient choice2 (6.9%)6 (20.7%)8 (13.8%)

aIf multiple reasons were provided, then primary was assigned in the order that it appears in the table. Primary reasons that were not reported (eg, rupture) do not appear on this table.

BIA-ALCL, BIA-SCC, Breast Cancer, and Patient Deaths

There were no new reported breast cancers in either augmentation cohort at 3 years. There were also no reports of either BIA-ALCL or breast implant–associated squamous cell carcinoma (BIA-SCC) in either of the breast augmentation cohorts at year 3. There was 1 reported death in the primary breast augmentation cohort as a complication of rectal cancer at year 2 and 1 death at year 1 in the revisional augmentation cohort due to trauma.

Clinical Effectiveness and Patient Satisfaction

Clinical effectiveness in this study was measured based upon patient and physician satisfaction gathered at each scheduled study visit, based on a 5-point Likert scale (Table 6). Patient satisfaction at 3 years for primary augmentation was (97.1%) and for revisional augmentation was (87.5%), yielding high overall patient satisfaction at year 3 (95.4%). Physician satisfaction was very high at year 3 for both primary augmentation (99.0%) and revisional augmentation (95.5%). Patient-reported satisfaction measures were also documented with the BREAST-Q for both primary and revisional augmentation cohorts (Table 7).

Table 6.

Patient Satisfaction: 5-Point Likert Scale at Year 3

Satisfaction assessmentaPrimary augmentation (n = 410)Revisional augmentation (n = 88)Overall augmentation
Patient satisfaction
 Satisfaction response
  Very satisfied332 (81.0%)64 (72.7%)396 (79.5%)
  Satisfied66 (16.1%)13 (14.8%)79 (15.9%)
  Neither dissatisfied nor satisfied5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied5 (1.2%)4 (4.5%)9 (1.8%)
  Very dissatisfied2 (0.5%)2 (2.3%)4 (0.8%)
 Satisfaction summary
  Satisfied398 (97.1%)77 (87.5%)475 (95.4%)
  Neither5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied7 (1.7%)6 (6.8%)13 (2.6%)
 Satisfaction analysis
  Percentage satisfied97.1%87.5%95.4%
  95% confidence interval(94.9%, 98.5%)(78.7%, 93.6%)(93.2%, 97.1%)
Investigator satisfaction
 Satisfaction response
  Very satisfied364 (88.8%)72 (81.8%)436 (87.6%)
  Satisfied42 (10.2%)12 (13.6%)54 (10.8%)
  Neither dissatisfied nor satisfied3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)1 (1.1%)2 (0.4%)
  Very dissatisfied01 (1.1%)1 (0.2%)
 Satisfaction summary
  Satisfied406 (99.0%)84 (95.5%)490 (98.4%)
  Neither3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)2 (2.3%)3 (0.6%)
 Satisfaction analysis
  Percentage satisfied99.0%95.5%98.4%
  95% confidence interval(97.5%, 99.7%)(88.8%, 98.7%)(96.9%, 99.3%)
Satisfaction assessmentaPrimary augmentation (n = 410)Revisional augmentation (n = 88)Overall augmentation
Patient satisfaction
 Satisfaction response
  Very satisfied332 (81.0%)64 (72.7%)396 (79.5%)
  Satisfied66 (16.1%)13 (14.8%)79 (15.9%)
  Neither dissatisfied nor satisfied5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied5 (1.2%)4 (4.5%)9 (1.8%)
  Very dissatisfied2 (0.5%)2 (2.3%)4 (0.8%)
 Satisfaction summary
  Satisfied398 (97.1%)77 (87.5%)475 (95.4%)
  Neither5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied7 (1.7%)6 (6.8%)13 (2.6%)
 Satisfaction analysis
  Percentage satisfied97.1%87.5%95.4%
  95% confidence interval(94.9%, 98.5%)(78.7%, 93.6%)(93.2%, 97.1%)
Investigator satisfaction
 Satisfaction response
  Very satisfied364 (88.8%)72 (81.8%)436 (87.6%)
  Satisfied42 (10.2%)12 (13.6%)54 (10.8%)
  Neither dissatisfied nor satisfied3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)1 (1.1%)2 (0.4%)
  Very dissatisfied01 (1.1%)1 (0.2%)
 Satisfaction summary
  Satisfied406 (99.0%)84 (95.5%)490 (98.4%)
  Neither3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)2 (2.3%)3 (0.6%)
 Satisfaction analysis
  Percentage satisfied99.0%95.5%98.4%
  95% confidence interval(97.5%, 99.7%)(88.8%, 98.7%)(96.9%, 99.3%)

aData include only patients without missing responses for primary implants at this follow-up year 3.

Table 6.

Patient Satisfaction: 5-Point Likert Scale at Year 3

Satisfaction assessmentaPrimary augmentation (n = 410)Revisional augmentation (n = 88)Overall augmentation
Patient satisfaction
 Satisfaction response
  Very satisfied332 (81.0%)64 (72.7%)396 (79.5%)
  Satisfied66 (16.1%)13 (14.8%)79 (15.9%)
  Neither dissatisfied nor satisfied5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied5 (1.2%)4 (4.5%)9 (1.8%)
  Very dissatisfied2 (0.5%)2 (2.3%)4 (0.8%)
 Satisfaction summary
  Satisfied398 (97.1%)77 (87.5%)475 (95.4%)
  Neither5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied7 (1.7%)6 (6.8%)13 (2.6%)
 Satisfaction analysis
  Percentage satisfied97.1%87.5%95.4%
  95% confidence interval(94.9%, 98.5%)(78.7%, 93.6%)(93.2%, 97.1%)
Investigator satisfaction
 Satisfaction response
  Very satisfied364 (88.8%)72 (81.8%)436 (87.6%)
  Satisfied42 (10.2%)12 (13.6%)54 (10.8%)
  Neither dissatisfied nor satisfied3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)1 (1.1%)2 (0.4%)
  Very dissatisfied01 (1.1%)1 (0.2%)
 Satisfaction summary
  Satisfied406 (99.0%)84 (95.5%)490 (98.4%)
  Neither3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)2 (2.3%)3 (0.6%)
 Satisfaction analysis
  Percentage satisfied99.0%95.5%98.4%
  95% confidence interval(97.5%, 99.7%)(88.8%, 98.7%)(96.9%, 99.3%)
Satisfaction assessmentaPrimary augmentation (n = 410)Revisional augmentation (n = 88)Overall augmentation
Patient satisfaction
 Satisfaction response
  Very satisfied332 (81.0%)64 (72.7%)396 (79.5%)
  Satisfied66 (16.1%)13 (14.8%)79 (15.9%)
  Neither dissatisfied nor satisfied5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied5 (1.2%)4 (4.5%)9 (1.8%)
  Very dissatisfied2 (0.5%)2 (2.3%)4 (0.8%)
 Satisfaction summary
  Satisfied398 (97.1%)77 (87.5%)475 (95.4%)
  Neither5 (1.2%)5 (5.7%)10 (2.0%)
  Dissatisfied7 (1.7%)6 (6.8%)13 (2.6%)
 Satisfaction analysis
  Percentage satisfied97.1%87.5%95.4%
  95% confidence interval(94.9%, 98.5%)(78.7%, 93.6%)(93.2%, 97.1%)
Investigator satisfaction
 Satisfaction response
  Very satisfied364 (88.8%)72 (81.8%)436 (87.6%)
  Satisfied42 (10.2%)12 (13.6%)54 (10.8%)
  Neither dissatisfied nor satisfied3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)1 (1.1%)2 (0.4%)
  Very dissatisfied01 (1.1%)1 (0.2%)
 Satisfaction summary
  Satisfied406 (99.0%)84 (95.5%)490 (98.4%)
  Neither3 (0.7%)2 (2.3%)5 (1.0%)
  Dissatisfied1 (0.2%)2 (2.3%)3 (0.6%)
 Satisfaction analysis
  Percentage satisfied99.0%95.5%98.4%
  95% confidence interval(97.5%, 99.7%)(88.8%, 98.7%)(96.9%, 99.3%)

aData include only patients without missing responses for primary implants at this follow-up year 3.

Table 7.

Satisfaction With Breast Implants Grom BREAST-Q at Year 3: Primary Augmentation and Revision Augmentation Subjects

nDissatisfiedSatisfied
Primary Augmentation
 Amount of cleavage when wearing a bra3835.5%94.5%
 Bra fit3834.2%95.8%
 Breast firmness3834.7%95.3%
 Breast shape without a bra3836.8%93.2%
 Breast size3826.5%93.5%
 Breast size matches the rest of the body3823.4%96.6%
 Breast feels to touch3822.4%97.6%
 How close breasts are when not wearing a bra38311.7%88.3%
 How close breasts match with each other3833.9%96.1%`
 Implant evenness/position relative to each other3815.0%95.0%
 Implant position on the chest3837.8%92.2%
 Look in the mirror clothed3833.7%96.3%
 Look in the mirror unclothed38110.5%89.5%
 Natural look of breasts3782.9%97.1%
 Naturally breasts sit/hang3806.8%93.2%
Revision Augmentation
 Amount of cleavage when wearing a bra767.9%92.1%
 Bra fit767.9%92.1%
 Breast firmness761.3%98.7%
 Breast shape without a bra7610.5%89.5%
 Breast size767.9%92.1%
 Breast size matches the rest of the body758.0%92.0%
 Breast feels to touch761.3%98.7%
 How close breasts are when not wearing a bra767.9%92.1%
 How close breasts match with each other7615.8%84.2%
 Implant evenness/position relative to each other7614.5%85.5%
 Implant position on the chest766.6%93.4%
 Look in the mirror clothed763.9%96.1%
 Look in the mirror unclothed7611.8%88.2%
 Natural look of breasts7610.5%89.5%
 Naturally breasts sit/hang766.6%93.4%
nDissatisfiedSatisfied
Primary Augmentation
 Amount of cleavage when wearing a bra3835.5%94.5%
 Bra fit3834.2%95.8%
 Breast firmness3834.7%95.3%
 Breast shape without a bra3836.8%93.2%
 Breast size3826.5%93.5%
 Breast size matches the rest of the body3823.4%96.6%
 Breast feels to touch3822.4%97.6%
 How close breasts are when not wearing a bra38311.7%88.3%
 How close breasts match with each other3833.9%96.1%`
 Implant evenness/position relative to each other3815.0%95.0%
 Implant position on the chest3837.8%92.2%
 Look in the mirror clothed3833.7%96.3%
 Look in the mirror unclothed38110.5%89.5%
 Natural look of breasts3782.9%97.1%
 Naturally breasts sit/hang3806.8%93.2%
Revision Augmentation
 Amount of cleavage when wearing a bra767.9%92.1%
 Bra fit767.9%92.1%
 Breast firmness761.3%98.7%
 Breast shape without a bra7610.5%89.5%
 Breast size767.9%92.1%
 Breast size matches the rest of the body758.0%92.0%
 Breast feels to touch761.3%98.7%
 How close breasts are when not wearing a bra767.9%92.1%
 How close breasts match with each other7615.8%84.2%
 Implant evenness/position relative to each other7614.5%85.5%
 Implant position on the chest766.6%93.4%
 Look in the mirror clothed763.9%96.1%
 Look in the mirror unclothed7611.8%88.2%
 Natural look of breasts7610.5%89.5%
 Naturally breasts sit/hang766.6%93.4%
Table 7.

Satisfaction With Breast Implants Grom BREAST-Q at Year 3: Primary Augmentation and Revision Augmentation Subjects

nDissatisfiedSatisfied
Primary Augmentation
 Amount of cleavage when wearing a bra3835.5%94.5%
 Bra fit3834.2%95.8%
 Breast firmness3834.7%95.3%
 Breast shape without a bra3836.8%93.2%
 Breast size3826.5%93.5%
 Breast size matches the rest of the body3823.4%96.6%
 Breast feels to touch3822.4%97.6%
 How close breasts are when not wearing a bra38311.7%88.3%
 How close breasts match with each other3833.9%96.1%`
 Implant evenness/position relative to each other3815.0%95.0%
 Implant position on the chest3837.8%92.2%
 Look in the mirror clothed3833.7%96.3%
 Look in the mirror unclothed38110.5%89.5%
 Natural look of breasts3782.9%97.1%
 Naturally breasts sit/hang3806.8%93.2%
Revision Augmentation
 Amount of cleavage when wearing a bra767.9%92.1%
 Bra fit767.9%92.1%
 Breast firmness761.3%98.7%
 Breast shape without a bra7610.5%89.5%
 Breast size767.9%92.1%
 Breast size matches the rest of the body758.0%92.0%
 Breast feels to touch761.3%98.7%
 How close breasts are when not wearing a bra767.9%92.1%
 How close breasts match with each other7615.8%84.2%
 Implant evenness/position relative to each other7614.5%85.5%
 Implant position on the chest766.6%93.4%
 Look in the mirror clothed763.9%96.1%
 Look in the mirror unclothed7611.8%88.2%
 Natural look of breasts7610.5%89.5%
 Naturally breasts sit/hang766.6%93.4%
nDissatisfiedSatisfied
Primary Augmentation
 Amount of cleavage when wearing a bra3835.5%94.5%
 Bra fit3834.2%95.8%
 Breast firmness3834.7%95.3%
 Breast shape without a bra3836.8%93.2%
 Breast size3826.5%93.5%
 Breast size matches the rest of the body3823.4%96.6%
 Breast feels to touch3822.4%97.6%
 How close breasts are when not wearing a bra38311.7%88.3%
 How close breasts match with each other3833.9%96.1%`
 Implant evenness/position relative to each other3815.0%95.0%
 Implant position on the chest3837.8%92.2%
 Look in the mirror clothed3833.7%96.3%
 Look in the mirror unclothed38110.5%89.5%
 Natural look of breasts3782.9%97.1%
 Naturally breasts sit/hang3806.8%93.2%
Revision Augmentation
 Amount of cleavage when wearing a bra767.9%92.1%
 Bra fit767.9%92.1%
 Breast firmness761.3%98.7%
 Breast shape without a bra7610.5%89.5%
 Breast size767.9%92.1%
 Breast size matches the rest of the body758.0%92.0%
 Breast feels to touch761.3%98.7%
 How close breasts are when not wearing a bra767.9%92.1%
 How close breasts match with each other7615.8%84.2%
 Implant evenness/position relative to each other7614.5%85.5%
 Implant position on the chest766.6%93.4%
 Look in the mirror clothed763.9%96.1%
 Look in the mirror unclothed7611.8%88.2%
 Natural look of breasts7610.5%89.5%
 Naturally breasts sit/hang766.6%93.4%

DISCUSSION

Motiva breast implants were first introduced to the market in 2011 and have undergone several modifications since then. They are currently approved in over 85 countries. This report contains the 3-year data from the first prospective, multicenter clinical trial to study the safety and efficacy of the Motiva SmoothSilk Ergonomix silicone gel-filled breast implants and the Motiva SmoothSilk Round gel implants. The implants included in this core study are designed with several new brand-named technologies, including ProgressiveGel Ultima, TrueMonobloc, and BluSeal.

Investigator surgeons received both in-person and online training before patient enrollment. Investigator surgeons were selected based on their years in practice, support staff required to follow the study protocol, and their previous performance as clinical investigators in premarket approval (PMA) studies. Training included educational materials on the novel implant matrix and a series of videos and in-person meetings with European and Costa Rican surgeons with more than 5 years of clinical experience with Motiva implants.2,8 Investigator training was designed to reduce the potential learning curve associated with these investigational implants.9

At 3 years, the reoperation rate for any reason was 6.0% in the primary augmentation cohort and 24.8% in the revisional augmentation cohort. The reoperation rate was higher in the revisional augmentation cohort, consistent with previous core studies showing that revisional patients are at higher risk of additional surgical procedures.10,11 Six patients (6.7%) in the revisional augmentation cohort underwent reoperation for capsular contracture grade III/IV by year 3. Five of these 6 patients had either previous grade III/IV capsular contracture or a ruptured device. It should also be noted that the study protocol prohibited either acellular dermal matrix or Scaffold in any revisional augmentation procedure.

Surgeon investigators were required to document the irrigation agents utilized during surgery and specifically provide intraoperative details regarding antibiotic solutions, Betadine, and funnels at the time of implantation. Most investigators reported some contamination mitigation techniques in both primary and revisional augmentation. These mitigation strategies included an IMF approach in the primary augmentation cohort (85.3%) and revisional augmentation cohort (81.7%), and pocket irrigation by the surgeon reported as combinations of saline, Betadine, and antibiotics, overall 95.0%. Additionally, the investigator surgeons reported the use of funnels in the primary augmentation cohort (72.1%) and revisional augmentation cohort (60.6%), an absent or uncommon practice when previous manufacturer IDE studies were undertaken. These techniques have been shown to reduce pocket contamination and potentially reduce capsular contracture and have become increasingly incorporated into breast implant procedures.12

Previously published core studies on the safety and efficacy of silicone breast implants initiated 10 to 20 years ago reported high revision rates for the leading drivers of reoperation: capsular contracture, size change, and malposition.13,14 It is evident that changes in surgeon practices over the last 2 decades, including better preoperative patient education, more precise surgical planning, and modifications in surgical techniques that include steps to reduce pocket contamination, have led to improvements in patient outcomes, as demonstrated in more modern clinical trials.15-17 The 3-year Motiva data demonstrate that the leading reasons for revisional surgery in both primary and revisional breast augmentation surgery have shifted from an objective classification of capsular contracture (grade III or IV) to more subjective indications such as malposition and size change, for which both the physician and the patient may determine the necessity of a revisional procedure.

Implant malposition was an initial concern of the study investigators due to published descriptions of explanted Motiva capsules being thin and translucent.4 By year 3, the reported incidence of malposition was 3.2% in the primary augmentation cohort and 4.9% in the revisional augmentation cohort. However, malposition did account for 44.8% of the revisional surgeries in the primary augmentation cohort (n = 13). There were other parameters that required a short learning curve for the investigator surgeons, such as interpreting a new implant sizing matrix and understanding the performance of the 2 gel fills. The fundamentals such as patient selection, implant selection, pocket location, precise surgical dissection, and meticulous closure remained the essential steps in minimizing complications. Malposition was not associated with a learning curve based on a review of the timing of patient enrollment and the experience of the investigators.2,18

Implant malposition results from a combination of factors, including soft tissue quality, implant volume, pocket dissection, adjustments to the IMF, and the closure of the IMF. Additionally, supportive postoperative bras and limitations in postoperative exercise have been associated with a reduction in implant malposition.19 Revision for malposition is based in part on the management of patient expectations, because the range of malposition can vary from a very mild unilateral malposition to significant bilateral malposition. A potential limitation of the study protocol was the lack of defined quantitative guidelines that classified the degree of malposition or the type of malposition. Investigators were not required to report the gradation of malposition in centimeters, nor did they define whether the malposition was inferior, lateral, medial, or superior. Revision for malposition in primary augmentation occurred at only 6 of 25 investigator sites, and this included revision for several patients with superior malposition. Previous published studies from Europe, Asia, and South America have reported very thin capsule formation surrounding these implants.20,21 In a retrospective study by Randquist, inferior and lateral displacement occurred most often in the first 3 years after the transition to Motiva implants. It is important to note that the transition in their study was from macrotextured shaped implants to smooth round devices and required additional refinements in implant selection, pocket dissection, management of the IMF incision, and postoperative care.2,7 For most surgeons in the United States, the transition will be from smooth devices to Motiva.

CONCLUSIONS

The 3-year data on Motiva implants demonstrate very low device-related complications. With high patient follow-up at 3 years (91.7%), the overall findings indicate that Motiva implants are safe and effective. In addition, patient and physician satisfaction was high and the risk of adverse events was low.

Supplemental Material

This article contains supplemental material located online at www.aestheticsurgeryjournal.com.

Acknowledgments

The authors wish to acknowledge the considerable contributions of Dr Scott Spear (1948-2017) in the design of the study protocol and the selection of the clinical investigators. The authors would also like to thank Dr Michael Schwartz for his contribution to this paper.

Disclosures

Dr Glicksman is a clinical investigator and the medical director for the Motiva US breast implant clinical trial (Establishment Labs, Alejuela, Costa Rica) and is the medical director of the BD GalaFlex clinical trial (BD, Franklin Lakes, NJ). Drs Wolfe and McGuire are clinical investigators for Establishment Labs. The authors have no financial interest to declare in Establishment Labs and received no financial support for the preparation of this manuscript.

Funding

The authors received no financial support for the research, authorship, and publication of this article.

REFERENCES

1

US Foot and Drug Administration
. Implantable Radiofrequency Transponder System for Patient Identification and Health Information - Class II Special Controls Guidance Document for Industry and FDA Staff. Published December 10, 2004. Accessed April 13, 2024. https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation-emitting-products/implantable-radiofrequency-transponder-system-patient-identification-and-health-information-class-ii

2

Chacón Quirós
 
M
,
Chacón Bolaños
 
M
,
Fassero
 
J
.
Six-year prospective outcomes of primary breast augmentation with nano surface implants
.
Aesthet Surg J
.
2019
;
39
(
5
):
495
508
. doi:

3

Hu
 
H
,
Jacombs
 
A
,
Vickery
 
K
,
Merten
 
SL
,
Pennington
 
DG
,
Deva
 
AK
.
Chronic biofilm infection in breast implants is associated with an increased T-cell lymphocytic infiltrate: implications for breast implant- associated lymphoma
.
Plast Reconstr Surg
.
2015
;
135
(
2
):
319
329
. doi:

4

Doloff
 
JC
,
Veiseh
 
O
,
de Mezerville
 
R
, et al.  
The surface topography of silicone breast implants mediates the foreign body response in mice, rabbits and humans
.
Nat Biomed Eng
.
2021
;
5
(
10
):
1115
1130
. doi:

5

Atlan
 
M
,
Nuti
 
G
,
Wang
 
H
,
Decker
 
S
,
Perry
 
TA
.
Breast implant surface texture impacts host tissue response
.
J Mech Behav Biomed Mater
.
2018
;
88
:
377
385
. doi :

6

Glicksman
 
CA
.
Managing a device clinical trial or clinical research during a global pandemic
.
Aesthet Surg J
.
2020
;
40
(
7
):
814
817
. doi:

7

Nichter
 
LS
,
Hardesty
 
RA
,
Anigian
 
GM
.
Ideal implant structured breast implants: core study results at 6 years
.
Plast Reconstr Surg
.
2018
;
142
(
1
):
66
75
. doi:

8

Randquist
 
C
,
Jaeger
 
M
,
Stavrou
 
D
.
Six-year evaluation of Motiva round and Ergonomix SmoothSilk surface silicone breast implants: a two-center, two-surgeon outcome analysis of 1053 primary and secondary breast augmentations and augmentation mastopexy
.
Aesthet Surg J
.
2023
;
43
(
3
):
295
307
. doi:

9

Montemurro
 
P
,
Tay
 
VKS
.
Transitioning from conventional textured to nanotextured breast implants: our early experience and modifications for optimal breast augmentation outcomes
.
Aesthet Surg J
.
2021
;
41
(
2
):
189
195
. doi:

10

Sientra FDA
. Summary of safety and effectiveness data (SSED) report. Accessed December 31, 2023. www.accessdata.fda.gov/cdrh_docs/pdf7/p070004b.pdf

11

Cunningham
 
B
.
The mentor core study on silicone MemoryGel breast implants
.
Plast Reconstr Surg
.
2007
;
120
(Supplement 1):
19S
29S
. doi:

12

Adams
 
WP
 Jr,
Conner
 
WC
,
Barton
 
FE
 Jr
,
Rohrich
 
RJ
.
Optimizing breast-pocket irrigation: the post-betadine era
.
Plast Reconstr Surg
.
2001
;
107
(
6
):
1596
1601
. doi:

13

Maxwell
 
GP
,
Van Natta
 
B
,
Bengtson
 
BP
,
Murphy
 
DK
.
Ten-year results the Natrelle 410 anatomical form-stable silicone breast implant core study
.
Aesthet Surg J
.
2015
;
35
(
2
):
145
155
. doi:

14

Spear
 
SL
,
Murphy
 
DK
.
Allergan silicone breast implant U.S. core clinical study group. Natrelle round silicone breast implants: core study results at 10 years
.
Plast Reconstr Surg
.
2014
;
133
(
6
):
1354
1361
. doi:

15

Deva
 
AK
,
Adams
 
WP
 Jr
,
Vickery
 
K
.
The role of bacterial biofilms in device-associated infection
.
Plast Reconstr Surg
.
2013
;
132
(
5
):
1319
1328
. doi:

16

Alderman
 
A
,
Caplin
 
D
,
Hammond
 
DC
,
Keane
 
A
,
Turetzky
 
J
,
Kane
 
WJ
.
Clinical results of mentor MemoryGel Xtra breast implants from the GLOW clinical trial
.
Aesthet Surg J
.
2023
;
43
(
12
):
NP1021
NP1032
. doi:

17

Adams
 
WP
 Jr,
Rios
 
JL
,
Smith
 
SJ
.
Enhancing patient outcomes in aesthetic and reconstructive breast surgery using triple antibiotic breast irrigation: six-year prospective clinical study
.
Plast Reconstr Surg
.
2006
;
118
(
7 Suppl
):
46S
52S
. doi:

18

Adams
 
WP
 Jr,
Small
 
KH
.
The process of breast augmentation with special focus on patient education, patient selection and implant selection
.
Clin Plast Surg
.
2015
;
42
(
4
):
413
426
. doi:

19

Gabriel
 
A
,
Maxwell
 
GP
. Treatment of implant malposition. In:
Kim
 
JYS
, ed.
Managing Common and Uncommon Complications of Aesthetic Breast Surgery
.
Springer Nature
;
2021
:
35
44
.

20

Sforza
 
M
,
Hammond
 
DC
,
Botti
 
G
, et al.  
Expert consensus on the use of a new bioengineered, cell-friendly, smooth surface breast implant
.
Aesthet Surg J
.
2019
;
39
(
Suppl_3
):
S95
S102
. doi:

21

Aitzetmüller-Klietz
 
ML
,
Yang
 
S
,
Wiebringhaus
 
P
, et al.  
Complication rates after breast surgery with the Motiva smooth silk surface silicone gel implants—a systematic review and meta-analysis
.
J Clin Med
.
2023
;
12
(
5
):
1881
. doi:

Author notes

Dr Glicksman is an assistant clinical professor, Hackensack Meridian School of Medicine, Nutley, NJ, USA and is an Aesthetic Breast Surgery section editor for Aesthetic Surgery Journal.

Dr Wolfe is a plastic surgeon in private practice, Golden, CO, USA.

Dr McGuire is a clinical instructor of surgery, Department of Plastic and Reconstructive Surgery, Washington University in St. Louis, St. Louis, MO, USA and is an Aesthetic Breast Surgery section editor for Aesthetic Surgery Journal.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.