Positive microbiological swap . | Negative microbiological swap/Physiological flora . | ||
---|---|---|---|
Positive result of the Moleculight | 10 (True positive) | 9 (False positive) | PPV 52.63% |
Negative result of the Moleculight | 64 (False negative) | 24 (True negative) | NPV 27.27% |
Sensitivity 13.51% | Specificity 72.73% |
Positive microbiological swap . | Negative microbiological swap/Physiological flora . | ||
---|---|---|---|
Positive result of the Moleculight | 10 (True positive) | 9 (False positive) | PPV 52.63% |
Negative result of the Moleculight | 64 (False negative) | 24 (True negative) | NPV 27.27% |
Sensitivity 13.51% | Specificity 72.73% |
Bold values indicate sensitivity and specificity of the moleculight i:X.