Table 3

Predischarge tests with immediate therapeutic relevant results

PatientR-wave sensing (mV)Pacing threshold ([email protected] ms)Pacing impedance (Ω)HV lead impedance (Ω)Fluoroscopy (lead position)DER (J)Intervention
Impl.PDTImpl.PDTImpl.PDTImpl.PDTPDTImpl.PDT
119.824.10.60.66999794035Unchanged17>31aReprogramming
210.35.30.40.810937894648Unchanged1731bReprogramming
311.92.60.61.094972149n.d.Dislodgement17n.d.Lead revision
416.43.61.05.56505183533Unchanged17n.d.Lead revision
54.28.11.01.411008834037Unchanged2131bLead revision
68.84.61.01.27677834531Unchanged2131bLead revision
PatientR-wave sensing (mV)Pacing threshold ([email protected] ms)Pacing impedance (Ω)HV lead impedance (Ω)Fluoroscopy (lead position)DER (J)Intervention
Impl.PDTImpl.PDTImpl.PDTImpl.PDTPDTImpl.PDT
119.824.10.60.66999794035Unchanged17>31aReprogramming
210.35.30.40.810937894648Unchanged1731bReprogramming
311.92.60.61.094972149n.d.Dislodgement17n.d.Lead revision
416.43.61.05.56505183533Unchanged17n.d.Lead revision
54.28.11.01.411008834037Unchanged2131bLead revision
68.84.61.01.27677834531Unchanged2131bLead revision

Impl. = implantation, PDT = predischarge testing, HV = high voltage, DER = defibrillation energy requirement, n.d. = not done. Maximum shock energy of the implanted devices was 31 J.

aVentricular fibrillation was terminated by an external applied monophasic 360J rescue shock.

bInternal 21J shock failed.

Table 3

Predischarge tests with immediate therapeutic relevant results

PatientR-wave sensing (mV)Pacing threshold ([email protected] ms)Pacing impedance (Ω)HV lead impedance (Ω)Fluoroscopy (lead position)DER (J)Intervention
Impl.PDTImpl.PDTImpl.PDTImpl.PDTPDTImpl.PDT
119.824.10.60.66999794035Unchanged17>31aReprogramming
210.35.30.40.810937894648Unchanged1731bReprogramming
311.92.60.61.094972149n.d.Dislodgement17n.d.Lead revision
416.43.61.05.56505183533Unchanged17n.d.Lead revision
54.28.11.01.411008834037Unchanged2131bLead revision
68.84.61.01.27677834531Unchanged2131bLead revision
PatientR-wave sensing (mV)Pacing threshold ([email protected] ms)Pacing impedance (Ω)HV lead impedance (Ω)Fluoroscopy (lead position)DER (J)Intervention
Impl.PDTImpl.PDTImpl.PDTImpl.PDTPDTImpl.PDT
119.824.10.60.66999794035Unchanged17>31aReprogramming
210.35.30.40.810937894648Unchanged1731bReprogramming
311.92.60.61.094972149n.d.Dislodgement17n.d.Lead revision
416.43.61.05.56505183533Unchanged17n.d.Lead revision
54.28.11.01.411008834037Unchanged2131bLead revision
68.84.61.01.27677834531Unchanged2131bLead revision

Impl. = implantation, PDT = predischarge testing, HV = high voltage, DER = defibrillation energy requirement, n.d. = not done. Maximum shock energy of the implanted devices was 31 J.

aVentricular fibrillation was terminated by an external applied monophasic 360J rescue shock.

bInternal 21J shock failed.