Table 2

Events trigger setting in the HM web site

PacemakerDual-chamber ICDICD–CRT
Technical dataTechnical dataTechnical data
 Elective replacement indicator Elective replacement indicator Elective replacement indicator
 A and/or V lead impedance <200 Ω and >3000 Ω A and/or V lead impedance <250 Ω and >1500 Ω A, right and left V lead impedance <250 Ω and >1500 Ω
 Increased ventricular threshold ≥4.8 V Shock impedance <25 Ω and >110 Ω Biventricular pacing lead impedance <200 Ω and >750 Ω
 Disabled ventricular capture control Abnormal system status Shock impedance <25 Ω and >110 Ω
 Reduced P or R wave sensing by 50% of safety marginArrhythmias episodes dignostic Abnormal system status
Heart Rhythm VT1, VF, and SVT detectionArrhythmias episodes diagnostic
 MS episode duration ≥10% (2, 5 h) 30 J ineffective shock VT1, VF, and SVT detection
 V run (4–8 consecutive VES)Heart rhythm 30 J ineffective shock
 Unsustained VT (>8 consecutive VES) MS episode duration ≥10% (2, 5 h)CHF diagnostics
 V intrinsic rhythm <90% % CRT pacing <90%
 Mean VES/h >100 MS episode duration ≥10% (2, 5 h)
 Mean V rate at rest >80 bpm
 Mean VES/h >100
PacemakerDual-chamber ICDICD–CRT
Technical dataTechnical dataTechnical data
 Elective replacement indicator Elective replacement indicator Elective replacement indicator
 A and/or V lead impedance <200 Ω and >3000 Ω A and/or V lead impedance <250 Ω and >1500 Ω A, right and left V lead impedance <250 Ω and >1500 Ω
 Increased ventricular threshold ≥4.8 V Shock impedance <25 Ω and >110 Ω Biventricular pacing lead impedance <200 Ω and >750 Ω
 Disabled ventricular capture control Abnormal system status Shock impedance <25 Ω and >110 Ω
 Reduced P or R wave sensing by 50% of safety marginArrhythmias episodes dignostic Abnormal system status
Heart Rhythm VT1, VF, and SVT detectionArrhythmias episodes diagnostic
 MS episode duration ≥10% (2, 5 h) 30 J ineffective shock VT1, VF, and SVT detection
 V run (4–8 consecutive VES)Heart rhythm 30 J ineffective shock
 Unsustained VT (>8 consecutive VES) MS episode duration ≥10% (2, 5 h)CHF diagnostics
 V intrinsic rhythm <90% % CRT pacing <90%
 Mean VES/h >100 MS episode duration ≥10% (2, 5 h)
 Mean V rate at rest >80 bpm
 Mean VES/h >100

A, atrial; CRT, cardiac resynchronization therapy; MS, mode switch; SVT, supraventricular tachycardia; V, ventricular; VES, ventricular extrasistole; VF, ventricular fibrillation; VT, ventricular tachycardia.

Table 2

Events trigger setting in the HM web site

PacemakerDual-chamber ICDICD–CRT
Technical dataTechnical dataTechnical data
 Elective replacement indicator Elective replacement indicator Elective replacement indicator
 A and/or V lead impedance <200 Ω and >3000 Ω A and/or V lead impedance <250 Ω and >1500 Ω A, right and left V lead impedance <250 Ω and >1500 Ω
 Increased ventricular threshold ≥4.8 V Shock impedance <25 Ω and >110 Ω Biventricular pacing lead impedance <200 Ω and >750 Ω
 Disabled ventricular capture control Abnormal system status Shock impedance <25 Ω and >110 Ω
 Reduced P or R wave sensing by 50% of safety marginArrhythmias episodes dignostic Abnormal system status
Heart Rhythm VT1, VF, and SVT detectionArrhythmias episodes diagnostic
 MS episode duration ≥10% (2, 5 h) 30 J ineffective shock VT1, VF, and SVT detection
 V run (4–8 consecutive VES)Heart rhythm 30 J ineffective shock
 Unsustained VT (>8 consecutive VES) MS episode duration ≥10% (2, 5 h)CHF diagnostics
 V intrinsic rhythm <90% % CRT pacing <90%
 Mean VES/h >100 MS episode duration ≥10% (2, 5 h)
 Mean V rate at rest >80 bpm
 Mean VES/h >100
PacemakerDual-chamber ICDICD–CRT
Technical dataTechnical dataTechnical data
 Elective replacement indicator Elective replacement indicator Elective replacement indicator
 A and/or V lead impedance <200 Ω and >3000 Ω A and/or V lead impedance <250 Ω and >1500 Ω A, right and left V lead impedance <250 Ω and >1500 Ω
 Increased ventricular threshold ≥4.8 V Shock impedance <25 Ω and >110 Ω Biventricular pacing lead impedance <200 Ω and >750 Ω
 Disabled ventricular capture control Abnormal system status Shock impedance <25 Ω and >110 Ω
 Reduced P or R wave sensing by 50% of safety marginArrhythmias episodes dignostic Abnormal system status
Heart Rhythm VT1, VF, and SVT detectionArrhythmias episodes diagnostic
 MS episode duration ≥10% (2, 5 h) 30 J ineffective shock VT1, VF, and SVT detection
 V run (4–8 consecutive VES)Heart rhythm 30 J ineffective shock
 Unsustained VT (>8 consecutive VES) MS episode duration ≥10% (2, 5 h)CHF diagnostics
 V intrinsic rhythm <90% % CRT pacing <90%
 Mean VES/h >100 MS episode duration ≥10% (2, 5 h)
 Mean V rate at rest >80 bpm
 Mean VES/h >100

A, atrial; CRT, cardiac resynchronization therapy; MS, mode switch; SVT, supraventricular tachycardia; V, ventricular; VES, ventricular extrasistole; VF, ventricular fibrillation; VT, ventricular tachycardia.