Table 1

Size and features in comparison to conventional pacemaker

AttributeMedtronic Adapta ADSR01Medtronic Micra
Size
 Volume of device (cm3)9.70.8
 Mass of device (g)21.52.0
 Total volume of material within body (cm3)10.56 (assumes Adapta device + 30 cm, 6 Fr lead)0.8 (device with electrodes placed directly on the pacemaker capsule)
Features
 Battery chemistry/capacityLithium-iodine, 830 mA hLithium silver vanadium oxide/carbon monoflouride, 120 mA h
 Threshold measurement and trackingMeasure 1/day
Output = maximum of 2× threshold or 2.0 V
Measure 1/day, verify 1/h
Output = threshold + 0.5 V
 Capture managementEvoked response (all leads)
Works at 0.4 ms
Evoked response (one pair of low polarization, 18 mm spaced electrodes)
Works at 0.24 and 0.4 ms
 Prolonged service period3 months6 months
 Magnet modeYesNo
 Rate responseSubQ accelerometer, (activity)Cardiac three-axis accelerometer, (activity), individually selectable with one vector used
 Ability to inactivate device at EOSNoDevice can be manually deactivated with programmer and automatically deactives at EOS
 MRI conditionalNoYes, by design
 ElectrodesLead dependentPacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body
 Steroid elutionLead dependentYes
AttributeMedtronic Adapta ADSR01Medtronic Micra
Size
 Volume of device (cm3)9.70.8
 Mass of device (g)21.52.0
 Total volume of material within body (cm3)10.56 (assumes Adapta device + 30 cm, 6 Fr lead)0.8 (device with electrodes placed directly on the pacemaker capsule)
Features
 Battery chemistry/capacityLithium-iodine, 830 mA hLithium silver vanadium oxide/carbon monoflouride, 120 mA h
 Threshold measurement and trackingMeasure 1/day
Output = maximum of 2× threshold or 2.0 V
Measure 1/day, verify 1/h
Output = threshold + 0.5 V
 Capture managementEvoked response (all leads)
Works at 0.4 ms
Evoked response (one pair of low polarization, 18 mm spaced electrodes)
Works at 0.24 and 0.4 ms
 Prolonged service period3 months6 months
 Magnet modeYesNo
 Rate responseSubQ accelerometer, (activity)Cardiac three-axis accelerometer, (activity), individually selectable with one vector used
 Ability to inactivate device at EOSNoDevice can be manually deactivated with programmer and automatically deactives at EOS
 MRI conditionalNoYes, by design
 ElectrodesLead dependentPacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body
 Steroid elutionLead dependentYes
Table 1

Size and features in comparison to conventional pacemaker

AttributeMedtronic Adapta ADSR01Medtronic Micra
Size
 Volume of device (cm3)9.70.8
 Mass of device (g)21.52.0
 Total volume of material within body (cm3)10.56 (assumes Adapta device + 30 cm, 6 Fr lead)0.8 (device with electrodes placed directly on the pacemaker capsule)
Features
 Battery chemistry/capacityLithium-iodine, 830 mA hLithium silver vanadium oxide/carbon monoflouride, 120 mA h
 Threshold measurement and trackingMeasure 1/day
Output = maximum of 2× threshold or 2.0 V
Measure 1/day, verify 1/h
Output = threshold + 0.5 V
 Capture managementEvoked response (all leads)
Works at 0.4 ms
Evoked response (one pair of low polarization, 18 mm spaced electrodes)
Works at 0.24 and 0.4 ms
 Prolonged service period3 months6 months
 Magnet modeYesNo
 Rate responseSubQ accelerometer, (activity)Cardiac three-axis accelerometer, (activity), individually selectable with one vector used
 Ability to inactivate device at EOSNoDevice can be manually deactivated with programmer and automatically deactives at EOS
 MRI conditionalNoYes, by design
 ElectrodesLead dependentPacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body
 Steroid elutionLead dependentYes
AttributeMedtronic Adapta ADSR01Medtronic Micra
Size
 Volume of device (cm3)9.70.8
 Mass of device (g)21.52.0
 Total volume of material within body (cm3)10.56 (assumes Adapta device + 30 cm, 6 Fr lead)0.8 (device with electrodes placed directly on the pacemaker capsule)
Features
 Battery chemistry/capacityLithium-iodine, 830 mA hLithium silver vanadium oxide/carbon monoflouride, 120 mA h
 Threshold measurement and trackingMeasure 1/day
Output = maximum of 2× threshold or 2.0 V
Measure 1/day, verify 1/h
Output = threshold + 0.5 V
 Capture managementEvoked response (all leads)
Works at 0.4 ms
Evoked response (one pair of low polarization, 18 mm spaced electrodes)
Works at 0.24 and 0.4 ms
 Prolonged service period3 months6 months
 Magnet modeYesNo
 Rate responseSubQ accelerometer, (activity)Cardiac three-axis accelerometer, (activity), individually selectable with one vector used
 Ability to inactivate device at EOSNoDevice can be manually deactivated with programmer and automatically deactives at EOS
 MRI conditionalNoYes, by design
 ElectrodesLead dependentPacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body
 Steroid elutionLead dependentYes
Close