Attribute . | Medtronic Adapta ADSR01 . | Medtronic Micra . |
---|---|---|
Size | ||
Volume of device (cm3) | 9.7 | 0.8 |
Mass of device (g) | 21.5 | 2.0 |
Total volume of material within body (cm3) | 10.56 (assumes Adapta device + 30 cm, 6 Fr lead) | 0.8 (device with electrodes placed directly on the pacemaker capsule) |
Features | ||
Battery chemistry/capacity | Lithium-iodine, 830 mA h | Lithium silver vanadium oxide/carbon monoflouride, 120 mA h |
Threshold measurement and tracking | Measure 1/day Output = maximum of 2× threshold or 2.0 V | Measure 1/day, verify 1/h Output = threshold + 0.5 V |
Capture management | Evoked response (all leads) Works at 0.4 ms | Evoked response (one pair of low polarization, 18 mm spaced electrodes) Works at 0.24 and 0.4 ms |
Prolonged service period | 3 months | 6 months |
Magnet mode | Yes | No |
Rate response | SubQ accelerometer, (activity) | Cardiac three-axis accelerometer, (activity), individually selectable with one vector used |
Ability to inactivate device at EOS | No | Device can be manually deactivated with programmer and automatically deactives at EOS |
MRI conditional | No | Yes, by design |
Electrodes | Lead dependent | Pacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body |
Steroid elution | Lead dependent | Yes |
Attribute . | Medtronic Adapta ADSR01 . | Medtronic Micra . |
---|---|---|
Size | ||
Volume of device (cm3) | 9.7 | 0.8 |
Mass of device (g) | 21.5 | 2.0 |
Total volume of material within body (cm3) | 10.56 (assumes Adapta device + 30 cm, 6 Fr lead) | 0.8 (device with electrodes placed directly on the pacemaker capsule) |
Features | ||
Battery chemistry/capacity | Lithium-iodine, 830 mA h | Lithium silver vanadium oxide/carbon monoflouride, 120 mA h |
Threshold measurement and tracking | Measure 1/day Output = maximum of 2× threshold or 2.0 V | Measure 1/day, verify 1/h Output = threshold + 0.5 V |
Capture management | Evoked response (all leads) Works at 0.4 ms | Evoked response (one pair of low polarization, 18 mm spaced electrodes) Works at 0.24 and 0.4 ms |
Prolonged service period | 3 months | 6 months |
Magnet mode | Yes | No |
Rate response | SubQ accelerometer, (activity) | Cardiac three-axis accelerometer, (activity), individually selectable with one vector used |
Ability to inactivate device at EOS | No | Device can be manually deactivated with programmer and automatically deactives at EOS |
MRI conditional | No | Yes, by design |
Electrodes | Lead dependent | Pacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body |
Steroid elution | Lead dependent | Yes |
Attribute . | Medtronic Adapta ADSR01 . | Medtronic Micra . |
---|---|---|
Size | ||
Volume of device (cm3) | 9.7 | 0.8 |
Mass of device (g) | 21.5 | 2.0 |
Total volume of material within body (cm3) | 10.56 (assumes Adapta device + 30 cm, 6 Fr lead) | 0.8 (device with electrodes placed directly on the pacemaker capsule) |
Features | ||
Battery chemistry/capacity | Lithium-iodine, 830 mA h | Lithium silver vanadium oxide/carbon monoflouride, 120 mA h |
Threshold measurement and tracking | Measure 1/day Output = maximum of 2× threshold or 2.0 V | Measure 1/day, verify 1/h Output = threshold + 0.5 V |
Capture management | Evoked response (all leads) Works at 0.4 ms | Evoked response (one pair of low polarization, 18 mm spaced electrodes) Works at 0.24 and 0.4 ms |
Prolonged service period | 3 months | 6 months |
Magnet mode | Yes | No |
Rate response | SubQ accelerometer, (activity) | Cardiac three-axis accelerometer, (activity), individually selectable with one vector used |
Ability to inactivate device at EOS | No | Device can be manually deactivated with programmer and automatically deactives at EOS |
MRI conditional | No | Yes, by design |
Electrodes | Lead dependent | Pacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body |
Steroid elution | Lead dependent | Yes |
Attribute . | Medtronic Adapta ADSR01 . | Medtronic Micra . |
---|---|---|
Size | ||
Volume of device (cm3) | 9.7 | 0.8 |
Mass of device (g) | 21.5 | 2.0 |
Total volume of material within body (cm3) | 10.56 (assumes Adapta device + 30 cm, 6 Fr lead) | 0.8 (device with electrodes placed directly on the pacemaker capsule) |
Features | ||
Battery chemistry/capacity | Lithium-iodine, 830 mA h | Lithium silver vanadium oxide/carbon monoflouride, 120 mA h |
Threshold measurement and tracking | Measure 1/day Output = maximum of 2× threshold or 2.0 V | Measure 1/day, verify 1/h Output = threshold + 0.5 V |
Capture management | Evoked response (all leads) Works at 0.4 ms | Evoked response (one pair of low polarization, 18 mm spaced electrodes) Works at 0.24 and 0.4 ms |
Prolonged service period | 3 months | 6 months |
Magnet mode | Yes | No |
Rate response | SubQ accelerometer, (activity) | Cardiac three-axis accelerometer, (activity), individually selectable with one vector used |
Ability to inactivate device at EOS | No | Device can be manually deactivated with programmer and automatically deactives at EOS |
MRI conditional | No | Yes, by design |
Electrodes | Lead dependent | Pacing electrodes placed directly on the pacemaker capsule: Cathode: 2.5 mm2 TiN-coated and sintered, located at tip. Anode: 22 mm2 TiN-coated, located on ring on body |
Steroid elution | Lead dependent | Yes |