Table 4

Summary of study indication changes in radiation dosages during atrial fibrillation ablations by radiofrequency ablation (A) and single-shot devices (B)

AuthorSystem/techniqueYear of inclusionFluoroscopy time in minKAP changeKAP in cGycm2eED in mSvCXR equivalents
(A) 3D-mapping system (radiofrequency ablation, point-by-point)
 Estner et al.47 (doi:10.1093/europace/eul079)3D-mapping vs. fluoroscopy only200538.9−39%566011.3566
 Lee et al.48 (doi: 10.1093/europace/euv186)Contact force vs. non-contact force2009–20149.5−70%1040.210
 Christoph et al.43 (doi:10.1093/europace/euu334)NFCV vs. conventional 3D-mapping20136.4−49%37267.5373
 Huo et al.49 (doi:10.1016/j.hrthm.2015.05.018)NFCV vs. conventional 3D-mapping20141.8−73%6521.365
 Sommer et al.50 (doi:10.1093/europace/eux378)NFCV: last 250 vs. first 250 patients2012–20170.5−94%1520.315
 Khalaph et al.51 (doi: 10.1111/pace.14555)Visualizable steerable sheath vs. non-visualizable sheath2019–20217.0−35%5071.051
 Knecht et al.52 (doi:10.1093/europace/euv006)Zero-Fluoro after TSP vs. ‘normal-fluoro’ after TSP20144.2−25%13202.6132
 Lehrmann et al.53 (doi:10.1093/europace/euw334)Radiation dose over time200553.046359.3464
20155.0−96%1850.419
 Voskoboinik et al.54 (doi:/10.1016/j.hrthm.2017.02.014)Radiation dose over time201030.69.0450
2015 (CF PVI only)11.4−66%3.1155
 Bourier et al.42 (doi:10.1093/europace/euv364)Optimized X-ray programme vs. previous settings2014–20158.6−77%2000.420
 Attanasio et al.45 (doi:10.1111/pace.14205)Optimized X-ray programme2015–20186.2910.29
 Schreiber et al.55 (doi:/10.1007/s00399-021-00762-7)Optimized X-ray programme20209.41280.313
(B) Single-shot devices for AF ablation
 Hoffmann et al.56 (doi:10.1093/europace/euz155)Cryoballoon vs. RF2011–201623.4+39%24875.0249
 Rubesch-Kütemeyer et al.57 (doi:/10.1007/s10840-019-00564-5)Cryoballoon over time201311.714282.9143
20175.1−57%6171.262
 Reissmann et al.58 (doi:10.1093/europace/eux066)Cryoballoon optimized vs. standard X-ray programme201610.0−82%3890.839
 Kühne et al.59 (doi: 10.3389/fcvm.2021.664538)Cryoballoon without PV occlusion testing vs. standard2017–201911.0−81%3680.737
 Rottner et al.60 (doi:/10.3389/fcvm.2022.967341)Cryoballoon Kodex-EPD version 1.4.6 vs. 1.4.82019–202110.9−58%2940.629
 Huang et al.61 (doi:10.1111/jce.14546)Laserballoon low dose (ICE, 3D-mapping system) vs. standard2018–2019 (standard)16.919804.0198
2018–2019 (low dose)1.7−91%1810.418
 Magni et al.62 (doi:/10.3389/fcvm.2022.959186)Pulsed-field ablation2021–202213.56581.366
 Lemoine et al.63 (doi:/10.1007/s00392-022-02091-2)Pulsed-field ablation2021–202216.05051.051
 Bohnen et al.64 (doi/10.1093/europace/euac111)Pulsed-field ablation202116.01250.313
AuthorSystem/techniqueYear of inclusionFluoroscopy time in minKAP changeKAP in cGycm2eED in mSvCXR equivalents
(A) 3D-mapping system (radiofrequency ablation, point-by-point)
 Estner et al.47 (doi:10.1093/europace/eul079)3D-mapping vs. fluoroscopy only200538.9−39%566011.3566
 Lee et al.48 (doi: 10.1093/europace/euv186)Contact force vs. non-contact force2009–20149.5−70%1040.210
 Christoph et al.43 (doi:10.1093/europace/euu334)NFCV vs. conventional 3D-mapping20136.4−49%37267.5373
 Huo et al.49 (doi:10.1016/j.hrthm.2015.05.018)NFCV vs. conventional 3D-mapping20141.8−73%6521.365
 Sommer et al.50 (doi:10.1093/europace/eux378)NFCV: last 250 vs. first 250 patients2012–20170.5−94%1520.315
 Khalaph et al.51 (doi: 10.1111/pace.14555)Visualizable steerable sheath vs. non-visualizable sheath2019–20217.0−35%5071.051
 Knecht et al.52 (doi:10.1093/europace/euv006)Zero-Fluoro after TSP vs. ‘normal-fluoro’ after TSP20144.2−25%13202.6132
 Lehrmann et al.53 (doi:10.1093/europace/euw334)Radiation dose over time200553.046359.3464
20155.0−96%1850.419
 Voskoboinik et al.54 (doi:/10.1016/j.hrthm.2017.02.014)Radiation dose over time201030.69.0450
2015 (CF PVI only)11.4−66%3.1155
 Bourier et al.42 (doi:10.1093/europace/euv364)Optimized X-ray programme vs. previous settings2014–20158.6−77%2000.420
 Attanasio et al.45 (doi:10.1111/pace.14205)Optimized X-ray programme2015–20186.2910.29
 Schreiber et al.55 (doi:/10.1007/s00399-021-00762-7)Optimized X-ray programme20209.41280.313
(B) Single-shot devices for AF ablation
 Hoffmann et al.56 (doi:10.1093/europace/euz155)Cryoballoon vs. RF2011–201623.4+39%24875.0249
 Rubesch-Kütemeyer et al.57 (doi:/10.1007/s10840-019-00564-5)Cryoballoon over time201311.714282.9143
20175.1−57%6171.262
 Reissmann et al.58 (doi:10.1093/europace/eux066)Cryoballoon optimized vs. standard X-ray programme201610.0−82%3890.839
 Kühne et al.59 (doi: 10.3389/fcvm.2021.664538)Cryoballoon without PV occlusion testing vs. standard2017–201911.0−81%3680.737
 Rottner et al.60 (doi:/10.3389/fcvm.2022.967341)Cryoballoon Kodex-EPD version 1.4.6 vs. 1.4.82019–202110.9−58%2940.629
 Huang et al.61 (doi:10.1111/jce.14546)Laserballoon low dose (ICE, 3D-mapping system) vs. standard2018–2019 (standard)16.919804.0198
2018–2019 (low dose)1.7−91%1810.418
 Magni et al.62 (doi:/10.3389/fcvm.2022.959186)Pulsed-field ablation2021–202213.56581.366
 Lemoine et al.63 (doi:/10.1007/s00392-022-02091-2)Pulsed-field ablation2021–202216.05051.051
 Bohnen et al.64 (doi/10.1093/europace/euac111)Pulsed-field ablation202116.01250.313

The table only includes studies that reported kerma area product (KAP). The bold numbers show the reduction of Kerma area product (KAP) by the use of 3 D mapping systems.

CXR, chest X-ray; eED, estimated effective dose; ICE, intracardiac echocardiography; NFCV, non-fluoroscopic catheter visualization; PV, pulmonary vein; RF, radiofrequency; TSP, transseptal puncture.

Table 4

Summary of study indication changes in radiation dosages during atrial fibrillation ablations by radiofrequency ablation (A) and single-shot devices (B)

AuthorSystem/techniqueYear of inclusionFluoroscopy time in minKAP changeKAP in cGycm2eED in mSvCXR equivalents
(A) 3D-mapping system (radiofrequency ablation, point-by-point)
 Estner et al.47 (doi:10.1093/europace/eul079)3D-mapping vs. fluoroscopy only200538.9−39%566011.3566
 Lee et al.48 (doi: 10.1093/europace/euv186)Contact force vs. non-contact force2009–20149.5−70%1040.210
 Christoph et al.43 (doi:10.1093/europace/euu334)NFCV vs. conventional 3D-mapping20136.4−49%37267.5373
 Huo et al.49 (doi:10.1016/j.hrthm.2015.05.018)NFCV vs. conventional 3D-mapping20141.8−73%6521.365
 Sommer et al.50 (doi:10.1093/europace/eux378)NFCV: last 250 vs. first 250 patients2012–20170.5−94%1520.315
 Khalaph et al.51 (doi: 10.1111/pace.14555)Visualizable steerable sheath vs. non-visualizable sheath2019–20217.0−35%5071.051
 Knecht et al.52 (doi:10.1093/europace/euv006)Zero-Fluoro after TSP vs. ‘normal-fluoro’ after TSP20144.2−25%13202.6132
 Lehrmann et al.53 (doi:10.1093/europace/euw334)Radiation dose over time200553.046359.3464
20155.0−96%1850.419
 Voskoboinik et al.54 (doi:/10.1016/j.hrthm.2017.02.014)Radiation dose over time201030.69.0450
2015 (CF PVI only)11.4−66%3.1155
 Bourier et al.42 (doi:10.1093/europace/euv364)Optimized X-ray programme vs. previous settings2014–20158.6−77%2000.420
 Attanasio et al.45 (doi:10.1111/pace.14205)Optimized X-ray programme2015–20186.2910.29
 Schreiber et al.55 (doi:/10.1007/s00399-021-00762-7)Optimized X-ray programme20209.41280.313
(B) Single-shot devices for AF ablation
 Hoffmann et al.56 (doi:10.1093/europace/euz155)Cryoballoon vs. RF2011–201623.4+39%24875.0249
 Rubesch-Kütemeyer et al.57 (doi:/10.1007/s10840-019-00564-5)Cryoballoon over time201311.714282.9143
20175.1−57%6171.262
 Reissmann et al.58 (doi:10.1093/europace/eux066)Cryoballoon optimized vs. standard X-ray programme201610.0−82%3890.839
 Kühne et al.59 (doi: 10.3389/fcvm.2021.664538)Cryoballoon without PV occlusion testing vs. standard2017–201911.0−81%3680.737
 Rottner et al.60 (doi:/10.3389/fcvm.2022.967341)Cryoballoon Kodex-EPD version 1.4.6 vs. 1.4.82019–202110.9−58%2940.629
 Huang et al.61 (doi:10.1111/jce.14546)Laserballoon low dose (ICE, 3D-mapping system) vs. standard2018–2019 (standard)16.919804.0198
2018–2019 (low dose)1.7−91%1810.418
 Magni et al.62 (doi:/10.3389/fcvm.2022.959186)Pulsed-field ablation2021–202213.56581.366
 Lemoine et al.63 (doi:/10.1007/s00392-022-02091-2)Pulsed-field ablation2021–202216.05051.051
 Bohnen et al.64 (doi/10.1093/europace/euac111)Pulsed-field ablation202116.01250.313
AuthorSystem/techniqueYear of inclusionFluoroscopy time in minKAP changeKAP in cGycm2eED in mSvCXR equivalents
(A) 3D-mapping system (radiofrequency ablation, point-by-point)
 Estner et al.47 (doi:10.1093/europace/eul079)3D-mapping vs. fluoroscopy only200538.9−39%566011.3566
 Lee et al.48 (doi: 10.1093/europace/euv186)Contact force vs. non-contact force2009–20149.5−70%1040.210
 Christoph et al.43 (doi:10.1093/europace/euu334)NFCV vs. conventional 3D-mapping20136.4−49%37267.5373
 Huo et al.49 (doi:10.1016/j.hrthm.2015.05.018)NFCV vs. conventional 3D-mapping20141.8−73%6521.365
 Sommer et al.50 (doi:10.1093/europace/eux378)NFCV: last 250 vs. first 250 patients2012–20170.5−94%1520.315
 Khalaph et al.51 (doi: 10.1111/pace.14555)Visualizable steerable sheath vs. non-visualizable sheath2019–20217.0−35%5071.051
 Knecht et al.52 (doi:10.1093/europace/euv006)Zero-Fluoro after TSP vs. ‘normal-fluoro’ after TSP20144.2−25%13202.6132
 Lehrmann et al.53 (doi:10.1093/europace/euw334)Radiation dose over time200553.046359.3464
20155.0−96%1850.419
 Voskoboinik et al.54 (doi:/10.1016/j.hrthm.2017.02.014)Radiation dose over time201030.69.0450
2015 (CF PVI only)11.4−66%3.1155
 Bourier et al.42 (doi:10.1093/europace/euv364)Optimized X-ray programme vs. previous settings2014–20158.6−77%2000.420
 Attanasio et al.45 (doi:10.1111/pace.14205)Optimized X-ray programme2015–20186.2910.29
 Schreiber et al.55 (doi:/10.1007/s00399-021-00762-7)Optimized X-ray programme20209.41280.313
(B) Single-shot devices for AF ablation
 Hoffmann et al.56 (doi:10.1093/europace/euz155)Cryoballoon vs. RF2011–201623.4+39%24875.0249
 Rubesch-Kütemeyer et al.57 (doi:/10.1007/s10840-019-00564-5)Cryoballoon over time201311.714282.9143
20175.1−57%6171.262
 Reissmann et al.58 (doi:10.1093/europace/eux066)Cryoballoon optimized vs. standard X-ray programme201610.0−82%3890.839
 Kühne et al.59 (doi: 10.3389/fcvm.2021.664538)Cryoballoon without PV occlusion testing vs. standard2017–201911.0−81%3680.737
 Rottner et al.60 (doi:/10.3389/fcvm.2022.967341)Cryoballoon Kodex-EPD version 1.4.6 vs. 1.4.82019–202110.9−58%2940.629
 Huang et al.61 (doi:10.1111/jce.14546)Laserballoon low dose (ICE, 3D-mapping system) vs. standard2018–2019 (standard)16.919804.0198
2018–2019 (low dose)1.7−91%1810.418
 Magni et al.62 (doi:/10.3389/fcvm.2022.959186)Pulsed-field ablation2021–202213.56581.366
 Lemoine et al.63 (doi:/10.1007/s00392-022-02091-2)Pulsed-field ablation2021–202216.05051.051
 Bohnen et al.64 (doi/10.1093/europace/euac111)Pulsed-field ablation202116.01250.313

The table only includes studies that reported kerma area product (KAP). The bold numbers show the reduction of Kerma area product (KAP) by the use of 3 D mapping systems.

CXR, chest X-ray; eED, estimated effective dose; ICE, intracardiac echocardiography; NFCV, non-fluoroscopic catheter visualization; PV, pulmonary vein; RF, radiofrequency; TSP, transseptal puncture.

Close
This Feature Is Available To Subscribers Only

Sign In or Create an Account

Close

This PDF is available to Subscribers Only

View Article Abstract & Purchase Options

For full access to this pdf, sign in to an existing account, or purchase an annual subscription.

Close