Baseline characteristics . | MR-conditional system . | Non-MR conditional system . | P . | ||
---|---|---|---|---|---|
No. of scans | 533 | 615 | |||
No. of patients | 462 | 509 | |||
Patient characteristics | |||||
Age (years), median (IQR) | 65 | (50, 75) | 73 | (65, 79) | <0.001 |
Male sex | 369 | (69%) | 430 | (70%) | 0.6305 |
Hospitalized inpatient | 39 | (7%) | 31 | (5%) | 0.1079 |
Previous MRI with a CIED | 71 | (13%) | 106 | (17%) | 0.0973 |
Pacing-dependent | 70 | (13%) | 98 | (16%) | 0.2092 |
MRI examination | |||||
Cardiac | 321 | (54%) | 185 | (27%) | <0.001 |
Spine | 122 | (20%) | 158 | (23%) | |
Head | 93 | (16%) | 202 | (30%) | |
Abdomen or pelvis | 46 | (8%) | 91 | (14%) | |
Extremity or joint | 9 | (1%) | 36 | (5%) | |
Other | 6 | (1%) | 3 | (1%) | |
Device characteristics | |||||
PPM | 332 | (62%) | 330 | (54%) | 0.0352 |
ICD | 149 | (28%) | 168 | (27%) | |
CRT-P | 15 | (3%) | 26 | (4%) | |
CRT-D | 37 | (7%) | 91 | (15%) | |
Pulse generator characteristics | |||||
MR-conditional | 533 | (100%) | 122 | (20%) | – |
Age (years), median (IQR) | 1 | (1, 2) | 4 | (2, 6) | <0.001 |
Number older than 10 years | 0 | (0%) | 13 | (2%) | |
Pulse generator manufacturer | |||||
Boston Scientific | 139 | (26%) | 183 | (30%) | <0.001 |
Abbott | 44 | (8%) | 151 | (25%) | |
Biotronik | 64 | (12%) | 21 | (3%) | |
Medtronic | 282 | (53%) | 241 | (39%) | |
Sorin | 4 | (1%) | 17 | (3%) | |
Lead characteristics | MR-conditional lead | Non-MR conditional lead | |||
RA lead | 739 | (43%) | 228 | (42%) | 0.0746 |
RV lead | 559 | (32%) | 177 | (32%) | |
HV (ICD) lead | 312 | (18%) | 88 | (16%) | |
LV lead | 112 | (7%) | 53 | (10%) | |
Age (years) | 2 | (1, 4) | 8 | (4, 12) | <0.001 |
No. greater than 10 years old | 7 | 0 (4%) | 168 | (31%) | — |
Abandoned leads | — | 40 | (7%) | — |
Baseline characteristics . | MR-conditional system . | Non-MR conditional system . | P . | ||
---|---|---|---|---|---|
No. of scans | 533 | 615 | |||
No. of patients | 462 | 509 | |||
Patient characteristics | |||||
Age (years), median (IQR) | 65 | (50, 75) | 73 | (65, 79) | <0.001 |
Male sex | 369 | (69%) | 430 | (70%) | 0.6305 |
Hospitalized inpatient | 39 | (7%) | 31 | (5%) | 0.1079 |
Previous MRI with a CIED | 71 | (13%) | 106 | (17%) | 0.0973 |
Pacing-dependent | 70 | (13%) | 98 | (16%) | 0.2092 |
MRI examination | |||||
Cardiac | 321 | (54%) | 185 | (27%) | <0.001 |
Spine | 122 | (20%) | 158 | (23%) | |
Head | 93 | (16%) | 202 | (30%) | |
Abdomen or pelvis | 46 | (8%) | 91 | (14%) | |
Extremity or joint | 9 | (1%) | 36 | (5%) | |
Other | 6 | (1%) | 3 | (1%) | |
Device characteristics | |||||
PPM | 332 | (62%) | 330 | (54%) | 0.0352 |
ICD | 149 | (28%) | 168 | (27%) | |
CRT-P | 15 | (3%) | 26 | (4%) | |
CRT-D | 37 | (7%) | 91 | (15%) | |
Pulse generator characteristics | |||||
MR-conditional | 533 | (100%) | 122 | (20%) | – |
Age (years), median (IQR) | 1 | (1, 2) | 4 | (2, 6) | <0.001 |
Number older than 10 years | 0 | (0%) | 13 | (2%) | |
Pulse generator manufacturer | |||||
Boston Scientific | 139 | (26%) | 183 | (30%) | <0.001 |
Abbott | 44 | (8%) | 151 | (25%) | |
Biotronik | 64 | (12%) | 21 | (3%) | |
Medtronic | 282 | (53%) | 241 | (39%) | |
Sorin | 4 | (1%) | 17 | (3%) | |
Lead characteristics | MR-conditional lead | Non-MR conditional lead | |||
RA lead | 739 | (43%) | 228 | (42%) | 0.0746 |
RV lead | 559 | (32%) | 177 | (32%) | |
HV (ICD) lead | 312 | (18%) | 88 | (16%) | |
LV lead | 112 | (7%) | 53 | (10%) | |
Age (years) | 2 | (1, 4) | 8 | (4, 12) | <0.001 |
No. greater than 10 years old | 7 | 0 (4%) | 168 | (31%) | — |
Abandoned leads | — | 40 | (7%) | — |
Magnetic resonance conditionality stratification is by complete CIED system classification except for lead numbers, which are stratified by individual lead MR-conditionality. Pacesetter leads were labelled as Abbott (St Jude Medical); Intermedics as Boston Scientific; ELA as Sorin; and Vitatron as Medtronic. One patient underwent lead extraction and replacement between MRI scans changing magnetic resonance conditionality of the device, so total number of unique patients is 970. Some patients had multiple body part examinations per individual scan.
CIED, cardiac implantable electronic device; CRT-P/D, cardiac resynchronization therapy-pacemaker/defibrillator; HV, high voltage; ICD, implantable cardioverter-defibrillator; IQR, interquartile range; MRI, magnetic resonance imaging; PPM, permanent pacemaker.
Baseline characteristics . | MR-conditional system . | Non-MR conditional system . | P . | ||
---|---|---|---|---|---|
No. of scans | 533 | 615 | |||
No. of patients | 462 | 509 | |||
Patient characteristics | |||||
Age (years), median (IQR) | 65 | (50, 75) | 73 | (65, 79) | <0.001 |
Male sex | 369 | (69%) | 430 | (70%) | 0.6305 |
Hospitalized inpatient | 39 | (7%) | 31 | (5%) | 0.1079 |
Previous MRI with a CIED | 71 | (13%) | 106 | (17%) | 0.0973 |
Pacing-dependent | 70 | (13%) | 98 | (16%) | 0.2092 |
MRI examination | |||||
Cardiac | 321 | (54%) | 185 | (27%) | <0.001 |
Spine | 122 | (20%) | 158 | (23%) | |
Head | 93 | (16%) | 202 | (30%) | |
Abdomen or pelvis | 46 | (8%) | 91 | (14%) | |
Extremity or joint | 9 | (1%) | 36 | (5%) | |
Other | 6 | (1%) | 3 | (1%) | |
Device characteristics | |||||
PPM | 332 | (62%) | 330 | (54%) | 0.0352 |
ICD | 149 | (28%) | 168 | (27%) | |
CRT-P | 15 | (3%) | 26 | (4%) | |
CRT-D | 37 | (7%) | 91 | (15%) | |
Pulse generator characteristics | |||||
MR-conditional | 533 | (100%) | 122 | (20%) | – |
Age (years), median (IQR) | 1 | (1, 2) | 4 | (2, 6) | <0.001 |
Number older than 10 years | 0 | (0%) | 13 | (2%) | |
Pulse generator manufacturer | |||||
Boston Scientific | 139 | (26%) | 183 | (30%) | <0.001 |
Abbott | 44 | (8%) | 151 | (25%) | |
Biotronik | 64 | (12%) | 21 | (3%) | |
Medtronic | 282 | (53%) | 241 | (39%) | |
Sorin | 4 | (1%) | 17 | (3%) | |
Lead characteristics | MR-conditional lead | Non-MR conditional lead | |||
RA lead | 739 | (43%) | 228 | (42%) | 0.0746 |
RV lead | 559 | (32%) | 177 | (32%) | |
HV (ICD) lead | 312 | (18%) | 88 | (16%) | |
LV lead | 112 | (7%) | 53 | (10%) | |
Age (years) | 2 | (1, 4) | 8 | (4, 12) | <0.001 |
No. greater than 10 years old | 7 | 0 (4%) | 168 | (31%) | — |
Abandoned leads | — | 40 | (7%) | — |
Baseline characteristics . | MR-conditional system . | Non-MR conditional system . | P . | ||
---|---|---|---|---|---|
No. of scans | 533 | 615 | |||
No. of patients | 462 | 509 | |||
Patient characteristics | |||||
Age (years), median (IQR) | 65 | (50, 75) | 73 | (65, 79) | <0.001 |
Male sex | 369 | (69%) | 430 | (70%) | 0.6305 |
Hospitalized inpatient | 39 | (7%) | 31 | (5%) | 0.1079 |
Previous MRI with a CIED | 71 | (13%) | 106 | (17%) | 0.0973 |
Pacing-dependent | 70 | (13%) | 98 | (16%) | 0.2092 |
MRI examination | |||||
Cardiac | 321 | (54%) | 185 | (27%) | <0.001 |
Spine | 122 | (20%) | 158 | (23%) | |
Head | 93 | (16%) | 202 | (30%) | |
Abdomen or pelvis | 46 | (8%) | 91 | (14%) | |
Extremity or joint | 9 | (1%) | 36 | (5%) | |
Other | 6 | (1%) | 3 | (1%) | |
Device characteristics | |||||
PPM | 332 | (62%) | 330 | (54%) | 0.0352 |
ICD | 149 | (28%) | 168 | (27%) | |
CRT-P | 15 | (3%) | 26 | (4%) | |
CRT-D | 37 | (7%) | 91 | (15%) | |
Pulse generator characteristics | |||||
MR-conditional | 533 | (100%) | 122 | (20%) | – |
Age (years), median (IQR) | 1 | (1, 2) | 4 | (2, 6) | <0.001 |
Number older than 10 years | 0 | (0%) | 13 | (2%) | |
Pulse generator manufacturer | |||||
Boston Scientific | 139 | (26%) | 183 | (30%) | <0.001 |
Abbott | 44 | (8%) | 151 | (25%) | |
Biotronik | 64 | (12%) | 21 | (3%) | |
Medtronic | 282 | (53%) | 241 | (39%) | |
Sorin | 4 | (1%) | 17 | (3%) | |
Lead characteristics | MR-conditional lead | Non-MR conditional lead | |||
RA lead | 739 | (43%) | 228 | (42%) | 0.0746 |
RV lead | 559 | (32%) | 177 | (32%) | |
HV (ICD) lead | 312 | (18%) | 88 | (16%) | |
LV lead | 112 | (7%) | 53 | (10%) | |
Age (years) | 2 | (1, 4) | 8 | (4, 12) | <0.001 |
No. greater than 10 years old | 7 | 0 (4%) | 168 | (31%) | — |
Abandoned leads | — | 40 | (7%) | — |
Magnetic resonance conditionality stratification is by complete CIED system classification except for lead numbers, which are stratified by individual lead MR-conditionality. Pacesetter leads were labelled as Abbott (St Jude Medical); Intermedics as Boston Scientific; ELA as Sorin; and Vitatron as Medtronic. One patient underwent lead extraction and replacement between MRI scans changing magnetic resonance conditionality of the device, so total number of unique patients is 970. Some patients had multiple body part examinations per individual scan.
CIED, cardiac implantable electronic device; CRT-P/D, cardiac resynchronization therapy-pacemaker/defibrillator; HV, high voltage; ICD, implantable cardioverter-defibrillator; IQR, interquartile range; MRI, magnetic resonance imaging; PPM, permanent pacemaker.