Table 2

Baseline patient and cardiac implantable electronic device characteristics

Baseline characteristicsMR-conditional systemNon-MR conditional systemP
No. of scans533615
No. of patients462509
Patient characteristics
 Age (years), median (IQR)65(50, 75)73(65, 79)<0.001
 Male sex369(69%)430(70%)0.6305
 Hospitalized inpatient39(7%)31(5%)0.1079
 Previous MRI with a CIED71(13%)106(17%)0.0973
 Pacing-dependent70(13%)98(16%)0.2092
MRI examination
 Cardiac321(54%)185(27%)<0.001
 Spine122(20%)158(23%)
 Head93(16%)202(30%)
 Abdomen or pelvis46(8%)91(14%)
 Extremity or joint9(1%)36(5%)
 Other6(1%)3(1%)
Device characteristics
 PPM332(62%)330(54%)0.0352
 ICD149(28%)168(27%)
 CRT-P15(3%)26(4%)
 CRT-D37(7%)91(15%)
Pulse generator characteristics
 MR-conditional533(100%)122(20%)
 Age (years), median (IQR)1(1, 2)4(2, 6)<0.001
 Number older than 10 years0(0%)13(2%)
Pulse generator manufacturer
 Boston Scientific139(26%)183(30%)<0.001
 Abbott44(8%)151(25%)
 Biotronik64(12%)21(3%)
 Medtronic282(53%)241(39%)
 Sorin4(1%)17(3%)
Lead characteristicsMR-conditional leadNon-MR conditional lead
RA lead739(43%)228(42%)0.0746
RV lead559(32%)177(32%)
HV (ICD) lead312(18%)88(16%)
LV lead112(7%)53(10%)
Age (years)2(1, 4)8(4, 12)<0.001
No. greater than 10 years old70 (4%)168(31%)
Abandoned leads40(7%)
Baseline characteristicsMR-conditional systemNon-MR conditional systemP
No. of scans533615
No. of patients462509
Patient characteristics
 Age (years), median (IQR)65(50, 75)73(65, 79)<0.001
 Male sex369(69%)430(70%)0.6305
 Hospitalized inpatient39(7%)31(5%)0.1079
 Previous MRI with a CIED71(13%)106(17%)0.0973
 Pacing-dependent70(13%)98(16%)0.2092
MRI examination
 Cardiac321(54%)185(27%)<0.001
 Spine122(20%)158(23%)
 Head93(16%)202(30%)
 Abdomen or pelvis46(8%)91(14%)
 Extremity or joint9(1%)36(5%)
 Other6(1%)3(1%)
Device characteristics
 PPM332(62%)330(54%)0.0352
 ICD149(28%)168(27%)
 CRT-P15(3%)26(4%)
 CRT-D37(7%)91(15%)
Pulse generator characteristics
 MR-conditional533(100%)122(20%)
 Age (years), median (IQR)1(1, 2)4(2, 6)<0.001
 Number older than 10 years0(0%)13(2%)
Pulse generator manufacturer
 Boston Scientific139(26%)183(30%)<0.001
 Abbott44(8%)151(25%)
 Biotronik64(12%)21(3%)
 Medtronic282(53%)241(39%)
 Sorin4(1%)17(3%)
Lead characteristicsMR-conditional leadNon-MR conditional lead
RA lead739(43%)228(42%)0.0746
RV lead559(32%)177(32%)
HV (ICD) lead312(18%)88(16%)
LV lead112(7%)53(10%)
Age (years)2(1, 4)8(4, 12)<0.001
No. greater than 10 years old70 (4%)168(31%)
Abandoned leads40(7%)

Magnetic resonance conditionality stratification is by complete CIED system classification except for lead numbers, which are stratified by individual lead MR-conditionality. Pacesetter leads were labelled as Abbott (St Jude Medical); Intermedics as Boston Scientific; ELA as Sorin; and Vitatron as Medtronic. One patient underwent lead extraction and replacement between MRI scans changing magnetic resonance conditionality of the device, so total number of unique patients is 970. Some patients had multiple body part examinations per individual scan.

CIED, cardiac implantable electronic device; CRT-P/D, cardiac resynchronization therapy-pacemaker/defibrillator; HV, high voltage; ICD, implantable cardioverter-defibrillator; IQR, interquartile range; MRI, magnetic resonance imaging; PPM, permanent pacemaker.

Table 2

Baseline patient and cardiac implantable electronic device characteristics

Baseline characteristicsMR-conditional systemNon-MR conditional systemP
No. of scans533615
No. of patients462509
Patient characteristics
 Age (years), median (IQR)65(50, 75)73(65, 79)<0.001
 Male sex369(69%)430(70%)0.6305
 Hospitalized inpatient39(7%)31(5%)0.1079
 Previous MRI with a CIED71(13%)106(17%)0.0973
 Pacing-dependent70(13%)98(16%)0.2092
MRI examination
 Cardiac321(54%)185(27%)<0.001
 Spine122(20%)158(23%)
 Head93(16%)202(30%)
 Abdomen or pelvis46(8%)91(14%)
 Extremity or joint9(1%)36(5%)
 Other6(1%)3(1%)
Device characteristics
 PPM332(62%)330(54%)0.0352
 ICD149(28%)168(27%)
 CRT-P15(3%)26(4%)
 CRT-D37(7%)91(15%)
Pulse generator characteristics
 MR-conditional533(100%)122(20%)
 Age (years), median (IQR)1(1, 2)4(2, 6)<0.001
 Number older than 10 years0(0%)13(2%)
Pulse generator manufacturer
 Boston Scientific139(26%)183(30%)<0.001
 Abbott44(8%)151(25%)
 Biotronik64(12%)21(3%)
 Medtronic282(53%)241(39%)
 Sorin4(1%)17(3%)
Lead characteristicsMR-conditional leadNon-MR conditional lead
RA lead739(43%)228(42%)0.0746
RV lead559(32%)177(32%)
HV (ICD) lead312(18%)88(16%)
LV lead112(7%)53(10%)
Age (years)2(1, 4)8(4, 12)<0.001
No. greater than 10 years old70 (4%)168(31%)
Abandoned leads40(7%)
Baseline characteristicsMR-conditional systemNon-MR conditional systemP
No. of scans533615
No. of patients462509
Patient characteristics
 Age (years), median (IQR)65(50, 75)73(65, 79)<0.001
 Male sex369(69%)430(70%)0.6305
 Hospitalized inpatient39(7%)31(5%)0.1079
 Previous MRI with a CIED71(13%)106(17%)0.0973
 Pacing-dependent70(13%)98(16%)0.2092
MRI examination
 Cardiac321(54%)185(27%)<0.001
 Spine122(20%)158(23%)
 Head93(16%)202(30%)
 Abdomen or pelvis46(8%)91(14%)
 Extremity or joint9(1%)36(5%)
 Other6(1%)3(1%)
Device characteristics
 PPM332(62%)330(54%)0.0352
 ICD149(28%)168(27%)
 CRT-P15(3%)26(4%)
 CRT-D37(7%)91(15%)
Pulse generator characteristics
 MR-conditional533(100%)122(20%)
 Age (years), median (IQR)1(1, 2)4(2, 6)<0.001
 Number older than 10 years0(0%)13(2%)
Pulse generator manufacturer
 Boston Scientific139(26%)183(30%)<0.001
 Abbott44(8%)151(25%)
 Biotronik64(12%)21(3%)
 Medtronic282(53%)241(39%)
 Sorin4(1%)17(3%)
Lead characteristicsMR-conditional leadNon-MR conditional lead
RA lead739(43%)228(42%)0.0746
RV lead559(32%)177(32%)
HV (ICD) lead312(18%)88(16%)
LV lead112(7%)53(10%)
Age (years)2(1, 4)8(4, 12)<0.001
No. greater than 10 years old70 (4%)168(31%)
Abandoned leads40(7%)

Magnetic resonance conditionality stratification is by complete CIED system classification except for lead numbers, which are stratified by individual lead MR-conditionality. Pacesetter leads were labelled as Abbott (St Jude Medical); Intermedics as Boston Scientific; ELA as Sorin; and Vitatron as Medtronic. One patient underwent lead extraction and replacement between MRI scans changing magnetic resonance conditionality of the device, so total number of unique patients is 970. Some patients had multiple body part examinations per individual scan.

CIED, cardiac implantable electronic device; CRT-P/D, cardiac resynchronization therapy-pacemaker/defibrillator; HV, high voltage; ICD, implantable cardioverter-defibrillator; IQR, interquartile range; MRI, magnetic resonance imaging; PPM, permanent pacemaker.

Close
This Feature Is Available To Subscribers Only

Sign In or Create an Account

Close

This PDF is available to Subscribers Only

View Article Abstract & Purchase Options

For full access to this pdf, sign in to an existing account, or purchase an annual subscription.

Close