Summary of inhibition test results for 10 MV FFF. A dose rate ≥8 Gy/min induced pacing inhibition in all CIEDs. Assurity MRI induced pacing inhibition dependent on the dose rate. I, Pacing inhibition; −, normal operation
Dose rate (Gy/min) . | 4 . | 8 . | 12 . | 16 . | 20 . | 24 . | |
---|---|---|---|---|---|---|---|
Medtronic | Sensia | I | I | I | I | I | I |
Medtronic | INSYNC III | I | I | I | I | I | I |
St. Jude Medical | Assurity MRI | − | I | I | I | I | I |
St. Jude Medical | FRONTIER II | I | I | I | I | I | I |
Dose rate (Gy/min) . | 4 . | 8 . | 12 . | 16 . | 20 . | 24 . | |
---|---|---|---|---|---|---|---|
Medtronic | Sensia | I | I | I | I | I | I |
Medtronic | INSYNC III | I | I | I | I | I | I |
St. Jude Medical | Assurity MRI | − | I | I | I | I | I |
St. Jude Medical | FRONTIER II | I | I | I | I | I | I |
Summary of inhibition test results for 10 MV FFF. A dose rate ≥8 Gy/min induced pacing inhibition in all CIEDs. Assurity MRI induced pacing inhibition dependent on the dose rate. I, Pacing inhibition; −, normal operation
Dose rate (Gy/min) . | 4 . | 8 . | 12 . | 16 . | 20 . | 24 . | |
---|---|---|---|---|---|---|---|
Medtronic | Sensia | I | I | I | I | I | I |
Medtronic | INSYNC III | I | I | I | I | I | I |
St. Jude Medical | Assurity MRI | − | I | I | I | I | I |
St. Jude Medical | FRONTIER II | I | I | I | I | I | I |
Dose rate (Gy/min) . | 4 . | 8 . | 12 . | 16 . | 20 . | 24 . | |
---|---|---|---|---|---|---|---|
Medtronic | Sensia | I | I | I | I | I | I |
Medtronic | INSYNC III | I | I | I | I | I | I |
St. Jude Medical | Assurity MRI | − | I | I | I | I | I |
St. Jude Medical | FRONTIER II | I | I | I | I | I | I |