Abstract

Four hundred and ninety-five adult patients in 16 medical centres participated in this study aiming to investigate the congruence between the volume CT dose index (CTDIvol) monitor values and measured ones during common CT procedures, performed by 20 systems. Tube output CTDIair measurements were carried out on single and multislice scanners for any kV, slice thickness, mA and FOV combination used. The maximum/minimum ratio of measured CTDIvol values found to be 3.1, 3.5, 7.4, 7.5, 4.2, 11.3 and 5.5 for head base, head cerebrum, thorax, abdomen, pelvis, cervical spine and lumbar spine protocols, respectively. The mean divergence between the measured and displayed CTDIvol values was 4.2, 3.5, −1.0, 2.7, 4.9, −3.9 and −2.8 % for protocols as mentioned above, respectively. From the perspective of the number of detector rows of the scanners, the mean divergence was −6.7, −6.0, 0.8, −1.1, −0.4, −1.9, −5.3, 5.2 and 10.1 % for single, dual, 4, 6, 16, 24, 64, 128 and dual source 256-slice systems, respectively.

You do not currently have access to this article.