Abstract

This study aims at testing the INTE ring dosemeter based on MCP-Ns and TLD-100 detectors on users from the field of medical applications, namely radiopharmacists, personnel at a cyclotron facility with corresponding FDG synthesis cells, interventional radiology technologists and radiologists. These users were chosen due to the fact that they have a significantly high risk of exposure to their hands. Following previous results, MCP-Ns TL thin material was used for radiology measurements, whereas TLD-100 was preferred for other applications. The dosemeters were tested to make sure that they were waterproof and that they could be sterilised properly prior to use. Results confirm the need to implement finger dosimetry, mainly for interventional radiologists as finger dose can be >50 times higher than whole-body dose and 3 times higher than wrist dose.