ABSTRACT

Magnetometry plays an important role in exploring the deep sea, which is one of the Earth’s final unknown frontiers. However, the complexity of the marine environment and the limitations of conventional magnetometers restrict its in-depth application. The nitrogen-vacancy (NV) center in diamond offers a potential solution to encompass and transcend conventional ocean magnetometers. Its unique advantages, such as precise vector measurement and tolerance to extreme environments, make it well suited for deep-sea applications like navigation. This work introduces the first deep-sea quantum vector magnetometer based on NV centers. The performance of this magnetometer is effectively validated by a series of field tests on the manned submersible Shenhai Yongshi during a cruise in the South China Sea, including an experimental underwater navigation using the diamond quantum sensor as a magnetic compass. This successful deep-sea application marks a milestone for transforming this promising solid-state spin quantum system into a practical sensor for real-world marine applications.

INTRODUCTION

The ability of magnetometry to utilize Earth’s magnetic field and identify weak magnetic signals is essential in marine exploration. This technology has been widely applied across many scientific and engineering domains, including mineral exploration [1], magnetic mapping [2], underwater navigation [3], geophysical research [4] and volcano observation [5]. The history of ocean magnetometry dates back to the 1950s. Early ocean magnetometers were primarily based on a fluxgate sensor [6], and modern magnetometers tend to employ various sensing technologies for more specific applications, such as the proton precession magnetometer [6], the atomic vapor cell magnetometer [7], the superconducting quantum interference device (SQUID) [8] and the microelectromechanical systems (MEMS) magnetometer [9]. For a typical underwater magnetic survey, the magnetometers need to perform measurement at their own spatial positions, so various types of underwater vehicles are employed as movable ocean sensor platforms, including ship-towed vehicles [10,11], underwater gliding vehicles [7], remotely operated vehicles [12], autonomous underwater vehicles [13] and manned submersibles. The wide-range applications and sensor platforms in marine environments necessitate the selection of different marine magnetometers based on specific requirements.

Marine scalar magnetometers have been well developed and widely used in ocean magnetometry. These magnetometers measure the total magnitude of the magnetic field without directional information. Atomic vapor cell magnetometers are distinguished among these sensors for their high sensitivity, which have been increasingly used in oceanographic and geophysical applications in recent years [15]. Meanwhile, performing full-vector magnetometry on moving platforms still appears to be challenging [16,17]. The resource consumption and system complexity strictly limit the high-sensitivity magnetometers like SQUID [8,18]. The MEMS magnetometers and fluxgate magnetometers have the advantage of low power consumption and miniaturized size, but the relatively low sensitivity constrains their broader adoption [9,19]. Moreover, full-vector measurement based on these magnetometers cannot normally be achieved using a single sensor. Instead, it requires the integration of multiple sensors, each positioned at different locations and orientations, which is sensitive to slight attitude changes and sensor vibrations. This setup introduces systematic errors such as misalignment and inconsistency, which limits the full-vector magnetometry in practical applications [19,20]. Besides, ocean magnetometers also face general challenges such as high hydrostatic pressure, varying attitude in the Earth’s magnetic field and electromagnetic noise from the vehicle itself [10].

Facing these challenges, the nitrogen-vacancy (NV) center in diamond emerges as a competitive candidate of the next-generation general ocean magnetometer. The diamond lattice provides intrinsic directional reference that allows for full-vector measurements with a single sensor, thereby overcoming the common challenge faced by vector magnetometers [21,22]. The diamond NV center also possesses high dynamic range [23], high sensitivity [24,25] and compatibility in extreme environments [26]. These features position it as a well-suited sensor for full-vector marine magnetometry. Compared to atomic vapor cell magnetometers, diamond magnetometers could directly provide directional information of the magnetic field with a single sensor. Besides, the diamond magnetometers do not have the dead zone that disables the output in certain orientations [15], which is a common issue for vapor cell magnetometer devices in marine sensing. Research on atomic vapor cell magnetometers has led to some laboratory setups that achieve similar capabilities with increased system complexity [27,28], but they have not been used in applications like marine sensing. Furthermore, the solid-state feature of NV center sensors allows for a potentially miniaturized design on chip [29], enabling their compact integration into various underwater vehicles without the severe trade-off between performance, power consumption and size in conventional magnetometers.

In this work, we report a quantum sensor using NV centers in diamond for deep-sea applications, as shown in Fig. 1. The diamond quantum magnetometer is installed on a manned deep submersible vehicle named Shenhai Yongshi in the strapdown configuration [30]. During a cruise in the South China Sea, the device’s capability of dynamic vector magnetometry was validated through a series of field tests. In the development of the device, we craft a compact quantum control and readout system for the NV center, and integrate the system into independent functional modules. These components, along with the diamond probe and electronic devices, are encapsulated within a watertight cylinder. This sensor integration is specifically designed for deep-sea full-vector magnetometry applications. An integrated optical system with light-trapping diamond waveguide is designed to excite the NV centers and effectively collect the fluorescence carrying physical quantity information [31]. A multi-channel microwave system is developed to manipulate the NV centers aligning along all four crystallographic axes in the diamond lattice. Quantum states of the NV centers along these directions are readout simultaneously using a frequency-division multiplexing scheme and a proportional–integral–derivative (PID) frequency-locking algorithm. A miniaturized lock-in amplifier device developed in previous work is employed in the system [32]. By further exploiting the potential of this innovative quantum sensor, we could extend our capabilities in underwater exploration and elevate our understanding of the ocean’s depths to a new level in the future.

Overview of the diamond ocean magnetometer and deep-sea experiments. The figure showcases a deep-sea exploration in the South China Sea using the full-vector magnetometer based on a diamond NV center. The red icon in the figure denotes the experiment’s diving site, known as the Haima Cold Seep. In this setup, the diamond magnetometer is mounted on a deep-sea submersible and works as a realtime sensor. The device incorporates a suite of integrated subsystems designed for the control and readout of the quantum states of diamond NV centers, all housed within a robust, cylindrical, watertight compartment crafted from titanium alloy. The bathymetric data are obtained from the GEBCO grid [14].
Figure 1.

Overview of the diamond ocean magnetometer and deep-sea experiments. The figure showcases a deep-sea exploration in the South China Sea using the full-vector magnetometer based on a diamond NV center. The red icon in the figure denotes the experiment’s diving site, known as the Haima Cold Seep. In this setup, the diamond magnetometer is mounted on a deep-sea submersible and works as a realtime sensor. The device incorporates a suite of integrated subsystems designed for the control and readout of the quantum states of diamond NV centers, all housed within a robust, cylindrical, watertight compartment crafted from titanium alloy. The bathymetric data are obtained from the GEBCO grid [14].

DIAMOND MAGNETOMETRY

The NV center is a photoluminescent defect in the diamond lattice, composed of a nitrogen atom substituting a carbon site and an adjacent vacancy in the lattice structure [33–35]. The negatively charged NV center is sensitive to multiple physical quantities and can be readout by an optically detected magnetic resonance (ODMR) method. The extraordinarily long spin lifetime of the NV center at room temperature makes it a unique choice for quantum information and quantum sensing [24]. These features of the NV center position it as a promising magnetometer, offering high sensitivity and intrinsic crystallographic axes for full-vector measurement capability [36–38]. The stable property of diamond also allows it to work in extreme environments [39].

As demonstrated in Fig. 2a, the NV center features a ground-state electronic spin in the form of triplet |$^3A$|⁠. The energy levels |$\mathinner {|{m_s=0}\rangle }$| and |$\mathinner {|{m_s=\pm 1}\rangle }$| are split by zero-field splitting, |$D_{\it gs}\approx 2.87\ \text{GHz}$|⁠. This splitting results from spin-spin interactions within the color center and is dependent on temperature. The effective Hamiltonian describing the NV center’s spin system is given by

where |$\vec{S}=(S_x,S_y,S_z)$| is the dimensionless electronic spin-1 operator with |$\hat{z}$| parallel to the NV axis, |$g$| is the electronic |$g$| factor of the NV center and |$\mu _B$| is the Bohr magneton.

Schematic of vector magnetometry performed by the NV centers in diamond. (a) Energy level diagram of the NV center in diamond. The zero-field splitting $D_{\it gs}$ denotes the energy gap between the ground electronic spin levels $\mathinner {|{m_s=0}\rangle }$ and $\mathinner {|{m_s=\pm 1}\rangle }$. When a magnetic field is applied, the degeneracy of the $\mathinner {|{m_s=\pm 1}\rangle }$ states is lifted, causing Zeeman splitting, which enables vector magnetic field sensing. A 520-nm green laser can be used to excite the NV center from the ground state to the excited state, while the resultant red photoluminescence (PL) from spontaneous radiation serves as an optical readout for magnetic resonance. By applying a microwave field that matches the NV center’s transition frequency, its state can be manipulated. (b) The four crystalline structure figures demonstrate the different NV centers along all four crystallographic orientations. (c) The first-order derivative ODMR spectrum of the NV centers obtained by simultaneous microwave (MW) frequency sweeping with different orientations, each denoted by corresponding colors as in (b). The black line represents the fitted spectrum using Lorentzian profile. The hyperfine energy levels are simultaneously manipulated to enhance the signal contrast. The four resonant frequencies $f_{1\!-\!4}$ utilized in full-vector magnetometry are highlighted in the figure with a gray background.
Figure 2.

Schematic of vector magnetometry performed by the NV centers in diamond. (a) Energy level diagram of the NV center in diamond. The zero-field splitting |$D_{\it gs}$| denotes the energy gap between the ground electronic spin levels |$\mathinner {|{m_s=0}\rangle }$| and |$\mathinner {|{m_s=\pm 1}\rangle }$|⁠. When a magnetic field is applied, the degeneracy of the |$\mathinner {|{m_s=\pm 1}\rangle }$| states is lifted, causing Zeeman splitting, which enables vector magnetic field sensing. A 520-nm green laser can be used to excite the NV center from the ground state to the excited state, while the resultant red photoluminescence (PL) from spontaneous radiation serves as an optical readout for magnetic resonance. By applying a microwave field that matches the NV center’s transition frequency, its state can be manipulated. (b) The four crystalline structure figures demonstrate the different NV centers along all four crystallographic orientations. (c) The first-order derivative ODMR spectrum of the NV centers obtained by simultaneous microwave (MW) frequency sweeping with different orientations, each denoted by corresponding colors as in (b). The black line represents the fitted spectrum using Lorentzian profile. The hyperfine energy levels are simultaneously manipulated to enhance the signal contrast. The four resonant frequencies |$f_{1\!-\!4}$| utilized in full-vector magnetometry are highlighted in the figure with a gray background.

In the presence of an external magnetic field |$\vec{B}_0$| aligned with the direction of the NV center, the transition between |$\mathinner {|{m_s=\pm 1}\rangle }$| and |$\mathinner {|{m_s=0}\rangle }$| experiences a Zeeman splitting and the degeneracy is lifted. By measuring both the transition frequencies of |$\mathinner {|{m_s=0}\rangle }\leftrightarrow \mathinner {|{m_s=+1}\rangle }$| and |$\mathinner {|{m_s=0}\rangle }\leftrightarrow \mathinner {|{m_s=-1}\rangle }$|⁠, the magnetic field and zero-field splitting can be simultaneously extracted. The temperature dependence of zero-field splitting |$dD_{\it gs}/dT\approx -74\;\ \text{kHz/K}$| can also be utilized to detect the temperature change from the transition frequencies [40,41]. After the NV center is excited to state |$^3 E$| by a 520-nm green laser as in Fig. 2a, its spin state can be optically readout by the red fluorescence emitted during the NV center’s spontaneous emission to the ground state. The NV center emits red fluorescence at a higher rate in the |$\mathinner {|{m_s=0}\rangle }$| state compared to the |$\mathinner {|{m_s=\pm 1}\rangle }$| states. This is due to the increased probability of non-radiative transitions through the intermediate singlet states |$^1A$| and |$^1E$| at the |$\mathinner {|{m_s=\pm 1}\rangle }$| states. Therefore, the quantum states of this spin system can be identified from the intensity of the red fluorescence.

To implement vector magnetometry with a diamond sensor, it is necessary to separately extract the photoluminescence (PL) signal from the NV centers oriented along different crystallographic axes in the diamond lattice. Here we adopt a frequency-division multiplexing scheme based on microwave modulation technology. [21,42,43] The applied microwave field resonant with different NV centers is frequency modulated by corresponding preset square waves, which presents as a periodical change of the fluorescence signal in measurement. A photodiode is used to collect the fluorescence, and the resulting photocurrent is subsequently demodulated to extract the spin-state information of the four differently oriented NV centers from the overall photocurrent. The crystal structure of the NV centers is depicted in Fig. 2b, showcasing the diamond’s face-centered-cubic crystal symmetry. By applying an external bias magnetic field with appropriate projections along all diamond crystallographic axes, it is possible to separate the resonant frequencies, thereby enabling simultaneous NV center manipulation. As demonstrated in Fig. 2b, the energy diagram represents the NV centers aligning along four orientations in the presence of an external magnetic field of about |$5\ \text{mT}$|⁠. As an example, a continuous-wave optically detected magnetic resonance (CW-ODMR) spectrum of the NV color centers in the diamond ensemble is presented in Fig. 2(c), which includes eight sets of resonance frequencies from NV centers in four orientations. Among these resonant frequencies, four frequencies |$f_1, f_2, f_3, f_4$| along different crystallographic axes are selected to perform vector measurement and suppress common-mode temperature drift.

The frequency-modulation method is employed to modulate the PL signal into higher frequency domains, thereby effectively circumventing the influence of low-frequency electronic noise. When the central frequency of the applied microwave field remains invariant, the dynamic range of the magnetometer is limited by the linewidth of the CW-ODMR spectrum. Consequently, even a slight change in the magnetometer’s orientation within the geomagnetic field could exceed its effective dynamic range. Sweeping the microwave frequency and capturing a complete CW-ODMR spectrum for the center resonant frequency can expand the dynamic range, but the additional time consumption will restrict the sensor’s real-time performance. Here we successfully improved the dynamic range of magnetometry by employing a frequency-locking technique that utilizes a homebuilt field-programmable-gate-array- (FPGA) based PID controller [32,44]. High-speed switching of microwave frequencies and a built-in-hardware feedback algorithm are applied to track the zero-crossing points of the CW-ODMR spectrum, obtaining the resonant frequencies for spin-state transitions. Subsequently, the full-vector magnetic field and temperature information sensed by diamond could be simultaneously obtained from the resonant frequencies. In the functionality test shown in Figs. 3a and 3b, a square-wave magnetic field with an amplitude of |$(4.5\ \mu {\text{T}},7.7\ \mu {\text{T}}, 5.3\ \mu {\text{T}})$| is applied to the magnetometer, showcasing the full-vector measurement capability of the diamond magnetometer. The specific design and operating details of the sensor system are provided in the supplementary materials (SM).

The performance test of the diamond magnetometer. (a) The time-domain waveform of the resonant frequency response in the presence of a square-wave magnetic field. Each line represents a different orientation of the NV centers in diamond, as indicated in Fig. 2(c). (b) The time-domain waveform of the external magnetic field in (a) calculated from the four NV resonant frequencies. (c) The total magnetic field derived from the vector components measured by the diamond magnetometer during the deep-sea cruise, showcasing the interference from the submersible thruster. The red line indicates periods when the thruster is operating continuously, and the black line represents it operating infrequently. The electromagnetic noise from the submersible is high due to the close strapdown setup of the magnetometer. (d) The noise density spectrum obtained during the deep-sea cruise, derived from the data in (c).
Figure 3.

The performance test of the diamond magnetometer. (a) The time-domain waveform of the resonant frequency response in the presence of a square-wave magnetic field. Each line represents a different orientation of the NV centers in diamond, as indicated in Fig. 2(c). (b) The time-domain waveform of the external magnetic field in (a) calculated from the four NV resonant frequencies. (c) The total magnetic field derived from the vector components measured by the diamond magnetometer during the deep-sea cruise, showcasing the interference from the submersible thruster. The red line indicates periods when the thruster is operating continuously, and the black line represents it operating infrequently. The electromagnetic noise from the submersible is high due to the close strapdown setup of the magnetometer. (d) The noise density spectrum obtained during the deep-sea cruise, derived from the data in (c).

RESULTS

Experiment environment

The deep-sea magnetometry experiments were carried out using a 4500-m-level manned submersible during the TS2-18-7 cruise. The experiments were executed at two sites near the Haima Cold Seep. In each experiment, the diamond quantum magnetometer carried by the submersible descended to a depth of approximately |$1300\ \text{m}$|⁠, and performed magnetometry measurements above the seafloor. The environmental temperature and pressure data of the sensor are listed in SM Fig. S2.

Magnetometer performance

During the experiments, we conducted an in-field compensation experiment (see SM) and assessed the diamond sensor's measurement capabilities both on land and at the sea-bottom. After the magnetometer was deployed on the submersible and descended to the target depth, the initial test conducted was the verification of the magnetic noise measurement, with the results showcased in Figs. 3c and 3d. However, due to the magnetometer’s close strapdown configuration near the submersible’s electrical system, the electromagnetic interference from the submersible itself could not be completely eliminated.

The diamond magnetometer conducted the measurement of the submersible’s background static magnetic noise by deactivating the operational electronic devices and maintained quiescent at the bottom of the seafloor. The comparison between the data obtained during the submersible’s static situation and dynamic situation is demonstrated, from which electromagnetic interference from the submersible’s propulsion system and the magnetometer’s realtime capability in different motion attitudes can be observed. The detected magnetic noise levels were |$11.4\ \text{nT}/\sqrt{\text{Hz}}$| when the submersible’s electric propulsion engine was activated and |$2.0\ \text{nT}/\sqrt{\text{Hz}}$| when it was turned off.

Additionally, the stability and robustness of the diamond magnetometer were qualified during the two 8-h deep-sea diving experiments. The microwave signal could always follow the resonance frequency of the NV centers in continuous attitude change.

Magnetic compass

The capability of the diamond magnetometer in performing stable and continuous full-vector magnetic field measurements was also validated. This was achieved during the rapidly changing conditions of the geomagnetic field by conducting a 1080|$^{\circ }$| turn at the submersible’s minimum turning radius. The successful completion of this maneuver is illustrated in Fig. 4a. The diamond magnetometer accurately recorded the magnetic field during the continuous turning motion of the submersible throughout the cruise, confirming its dynamic range exceeding the geomagnetic field level of |$\pm 40\ \mu {\text{T}}$|⁠.

Demonstration of the diamond NV center magnetometer’s application capability in a deep-sea geomagnetic environment. (a) The vector magnetic field measured by the diamond magnetometer when the submersible continuously rotated for three circles at the minimum turning radius. Each line represents a vector component of the measured magnetic field. (b) A comparison between the results of the NV compass and the submersible’s shipboard compass during this rotation maneuvering. (c) The experimental navigation result given by the inertial measurement unit (IMU) system and the diamond magnetometer. The yellow points represent positioning conducted by the USBL system alone, while the red line represents combined navigation positioning carried out jointly by the USBL system, the IMU system and the diamond magnetometer. (d) The geomagnetic inclination and geomagnetic field intensity’s combined distribution recorded by the diamond magnetometer measured at diving site SY523.
Figure 4.

Demonstration of the diamond NV center magnetometer’s application capability in a deep-sea geomagnetic environment. (a) The vector magnetic field measured by the diamond magnetometer when the submersible continuously rotated for three circles at the minimum turning radius. Each line represents a vector component of the measured magnetic field. (b) A comparison between the results of the NV compass and the submersible’s shipboard compass during this rotation maneuvering. (c) The experimental navigation result given by the inertial measurement unit (IMU) system and the diamond magnetometer. The yellow points represent positioning conducted by the USBL system alone, while the red line represents combined navigation positioning carried out jointly by the USBL system, the IMU system and the diamond magnetometer. (d) The geomagnetic inclination and geomagnetic field intensity’s combined distribution recorded by the diamond magnetometer measured at diving site SY523.

Throughout the entire cruise, the diamond magnetometer functioned as a magnetic compass experimentally, and its effectiveness was validated with the built-in magnetic compass on the submersible. As shown in Fig. 4b, during the turning test, the consistency between the yaw angle calculated by the diamond magnetometer and the yaw angle measured by the submersible’s onboard magnetic compass was maintained within 5|$^{\circ }$|⁠. This test validates the vector measurement capability of the diamond magnetometer in a moving condition. Also, it demonstrates the potential of using the diamond NV center quantum system for navigation tools such as magnetic compasses for the first time.

Combined navigation

Additionally, by utilizing the vector diamond magnetometer in a deep-sea geomagnetic field, we conduct a short-distance experimental combined navigation demonstration using the diamond magnetometer as an attitude sensor. These results are compared with the submersible’s ultrasonic positioning system, validating the system’s capability to transition from absolute navigation to offline navigation for a short time, as shown in Fig. 4c. The validation experiment of combined navigation based on the ultra-short baseline (USBL) underwater positioning system and the diamond magnetometer was conducted. The yellow points represent positioning conducted by the USBL system alone, while the red line represents combined navigation positioning carried out jointly by the USBL system, the inertial measurement unit (IMU) system and the diamond magnetometer. As the variation of the local magnetic field in the test seafloor area was very small compared to the Earth’s magnetic field, this impact of the geological surroundings on the local magnetic field is neglectable during the navigation. The navigation algorithm was based on an extended Kalman filter (EKF) algorithm with the NV centers’ spin-state information as the main parameters, together with an IMU module to measure the necessary motion data. This method integrates USBL data, IMU data and attitude data from the diamond magnetometer with gravity accelerometer results. The USBL data indicate the submersible’s position, the IMU data provide velocity measurements in the submersible’s body coordinate and the diamond magnetometer provides geomagnetic information for attitude measurement. The vehicle’s state is encapsulated by a six-dimensional vector encompassing position and velocity, and a rotation matrix is computed for each time step, converting raw velocity vectors into a world reference frame. This multimodal data integration enhances the precision of state estimation. The EKF algorithm operates in two primary cycles. The predicted cycle projects the current state estimate forward in time, while the update cycle adjusts this prediction based on new measurements from either the IMU or USBL system, employing a specific measurement matrix for different sensor types.

The navigation was achieved by utilizing the diamond magnetometer as an attitude sensor, thereby showcasing its utility from merely magnetic field measurement to multi-sensor fusion area. This comparison underscores the system’s capability to perform offline navigation for short time intervals, which is useful for underwater vehicle navigation. Because of the presence of constraints such as sensor drift in the IMU system, this experimental navigation system that integrates diamond magnetometers with the IMU system is not sufficient to enable underwater offline navigation for now, resulting in real-time positioning inaccuracies of about |$20\ \text{m}$|⁠, as demonstrated in Fig. 4c. Looking forward, we aim to further utilize the full-vector measurement capability of diamond magnetometers and incorporate vector geomagnetic navigation into the framework of underwater offline navigation, offering a viable and efficient solution for this realm.

Geomagnetic measurement

We conducted geomagnetic field measurements at a diving site in the South China Sea with the diamond magnetometer. The geomagnetic inclination angle at the experiment diving site was detected and proven to be consistent with the International Geomagnetic Reference Field (IGRF) model. The observation accords with the IGRF geomagnetic model, and the comparison is shown in Fig. 4d. The geomagnetic information for this area can be calculated using the IGRF model, which predicts a theoretical magnetic inclination of |$22.6^\circ$| and a theoretical geomagnetic field strength of |$43.3\ \mu {\text{T}}$|⁠. Meanwhile, the measured central value of the magnetic inclination |$\theta _I$|’s distribution was approximately |$23.6^\circ$|⁠, and the measured central value of the geomagnetic field strength was |$42.1\ \mu {\text{T}}$|⁠. These results demonstrate the consistency between the geomagnetic data measured by the diamond magnetometer in the deep-sea area and the IGRF model, validating its capability for vector magnetic measurements in geographical environments. The orientation reference for vertical geographic coordinates was provided by a MEMS accelerometer. The geomagnetic data obtained by the diamond magnetometer were processed to obtain the submersible’s heading information.

CONCLUSIONS

In conclusion, we successfully demonstrated a vector diamond quantum sensor as a deep-sea magnetometer. A compact quantum control and readout system based on the NV centers was developed and optimized to achieve the sensor functionality. The functionality validation was carried out through a series of tests at the 1300-m-depth bottom of the South China Sea, effectively proving the capability of the diamond sensor in comparison to previous shipboard efforts [45–47]. The diamond sensor’s primary functionality to perform simultaneous full-vector magnetic field measurements underwater was successfully demonstrated; it was mounted on a manned deep submersible vehicle and managed to perform electromagnetic disturbance monitoring, experimental navigation and geomagnetic measurement.

We plan to further exploit the diamond sensor’s unique features like extreme pressure resistance in future works. Although the diamond probe itself can resist extreme pressure [26], the auxiliary electronics in this setup cannot currently operate when exposed to the deep-sea environment. A device encapsulation method using pressure-resistant soft materials could protect the auxiliary electronics and further exploit the unique extreme environment compatibility of diamond [48]. This method could extend the diamond magnetometer’s applicability in extreme environments, and give rise to a new in situ deep-sea magnetometry.

Beyond the results we demonstrated in this work, the diamond magnetometer still possesses significant potential for performance improvement in sensitivity, calibration-free vector magnetometry [22,42,49,50], chip-level miniaturization based on a micro-nano machining technique, direct deployment in a high-pressure seawater environment [26,51,52] and multiple physical quantity measurements [53–55]. There are also available opportunities for improvement in the sensor deployment method employed in this ocean experiment, considering the electromagnetic interference observed during the expedition. An external rod or distant cable connection could be applied to the diamond sensor for a measurement with less magnetic disturbance [11]. We could expect continuous improvement in the performance of the diamond quantum magnetometer and its deployment in broader fields, creating practical magnetic sensors capable of covering a wide range of application scenarios than traditional magnetometers.

ACKNOWLEDGEMENTS

The authors thank the crew members of Tansuo-2 and the Shenhai Yongshi pilot crew, as well as the researchers, for their invaluable assistance in the research, development and sea testing of the device. The authors are also grateful to the anonymous reviewers for providing inspiration and valuable comments that improved the manuscript quality.

FUNDING

This work was supported by the National Key R&D Program of China (2021YFB3202800), the Chinese Academy of Sciences (XDC07000000), the Hefei Comprehensive National Science Center, the Zhejiang Provincial Natural Science Foundation of China (LDQ24A040001) and the Fundamental Research Funds for the Central Universities (WK3540000011, 226-2023-00137, 226-2023-00138).

AUTHOR CONTRIBUTION

Z.Y. and Y.Z. contributed equally to this work. Z.Y. contributed to the sensor simulation analysis, ocean experiments, data processing and the program writing. Y.X., Y.Z., Z.Y. and J.K. contributed to the sensor design and engineering integration. W.Z. contributed to the marine experiments with the sensor and provided guidance on the electronics. C.C. and S.W. participated in the ocean experiments and offered guidance in marine science. Y.X., X.R. and J.D. contributed to the sensor design ideas and supervised the work. All authors contributed to the writing of this manuscript.

Conflict of interest statement. None declared.

REFERENCES

1.

Barckhausen U and Roeser
 
HA
.
Seafloor spreading anomalies in the South China Sea revisited
. In:
Clift
 
P
,
Kuhnt
 
W
,
Wang
 
P
,
Hayes
 
D
(eds.).
Continent-Ocean Interactions Within East Asian Marginal Seas
.
Washington, DC
:
American Geophysical Union, Vol. 149
,
2004
,
121
5
.

2.

Hemant
 
K
,
Thébault
 
E
,
Mandea
 
M
 et al.  
Magnetic anomaly map of the world: merging satellite, airborne, marine and ground-based magnetic data sets
.
Earth Planet Sci Lett
 
2007
;
260
:
56
71
.

3.

Goldenberg
 
F
.
Geomagnetic navigation beyond the magnetic compass
. In:
2006 IEEE/ION Position, Location, And Navigation Symposium
.
Piscataway, NJ
:
IEEE Press
,
2006
,
684
94
.

4.

Boyce
 
JI
,
Reinhardt
 
EG
,
Raban
 
A
 et al.  
Marine magnetic survey of a submerged Roman harbour, Caesarea Maritima, Israel
.
Int J Naut Archaeol
 
2004
;
33
:
122
36
.

5.

Caratori Tontini
 
F
,
Bortoluzzi
 
G
,
Carmisciano
 
C
 et al.  
Near-bottom magnetic signatures of submarine hydrothermal systems at Marsili and Palinuro volcanoes, southern Tyrrhenian Sea, Italy
.
Econ Geol
 
2014
;
109
:
2119
28
.

6.

Heirtzler
 
JR
.
Magnetic measurements near the deep ocean floor
.
Deep-Sea Res Oceanogr Abstr
 
1964
;
11
:
891
8
.

7.

Page
 
BR
,
Lambert
 
R
,
Mahmoudian
 
N
 et al.  
Compact quantum magnetometer system on an agile underwater glider
.
Sensors
 
2021
;
21
:
1092
.

8.

Chwala
 
A
,
Schmelz
 
M
,
Zakosarenko
 
V
 et al.  
Underwater operation of a full tensor SQUID gradiometer system
.
Supercond Sci Technol
 
2019
;
32
:
024003
.

9.

Yan
 
R
,
Zhang
 
F
,
Chen
 
H
.
A MEMS-based magnetometer calibration approach in AUV navigation system
. In:
OCEANS 2019 - Marseille
.
Piscataway, NJ
:
IEEE Press
,
2019
,
1
6
.

10.

Gee JS and Cande
 
SC
.
A surface-towed vector magnetometer
.
Geophys Res Lett
 
2002
;
29
:
1670
.

11.

Ge
 
J
,
Luo
 
W
,
Dong
 
H
 et al.  
Towed overhauser marine magnetometer for weak magnetic anomaly detection in severe ocean conditions
.
Rev Sci Instrum
 
2020
;
91
:
035112
.

12.

Amara-Madi
 
S
,
Price
 
CA
,
Bensaoula
 
A
 et al.  
Autonomous sensor system for deep-sea pipeline monitoring
. In:
2013 IEEE 11th International New Circuits and Systems Conference (NEWCAS)
.
Piscataway, NJ
:
IEEE Press
,
2013
,
1
4
.

13.

Armstrong
 
B
,
Pentzer
 
J
,
Odell
 
D
 et al.  
Field measurement of surface ship magnetic signature using multiple AUVs
. In:
OCEANS 2009
.
Piscataway, NJ
:
IEEE Press
,
2009
,
1
9
.

15.

Bai
 
X
,
Wen
 
K
,
Peng
 
D
 et al.  
Atomic magnetometers and their application in industry
.
Front Phys
 
2023
;
11
:
1212368
.

16.

Humphrey
 
KP
,
Horton
 
TJ
,
Keene
 
MN
.
Detection of mobile targets from a moving platform using an actively shielded, adaptively balanced squid gradiometer
.
IEEE Trans Appl Supercond
 
2005
;
15
:
753
6
.

17.

Wahdan
 
A
,
Georgy
 
J
,
Abdelfatah
 
WF
 et al.  
Magnetometer calibration for portable navigation devices in vehicles using a fast and autonomous technique
.
IEEE Trans Intell Transp Syst
 
2014
;
15
:
2347
52
.

18.

Hirota
 
M
,
Furuse
 
T
,
Ebana
 
K
 et al.  
Magnetic detection of a surface ship by an airborne LTS SQUID MAD
.
IEEE Trans Appl Supercond
 
2001
;
11
:
884
87
.

19.

Ripka
 
P
.
Advances in fluxgate sensors
.
Sens Actuators A: Phys
 
2003
;
106
:
8
14
.

20.

Engels
 
M
,
Barckhausen
 
U
,
Gee
 
JS
.
A new towed marine vector magnetometer: methods and results from a Central Pacific cruise
.
Geophys J Int
 
2008
;
172
:
115
29
.

21.

Schloss
 
JM
,
Barry
 
JF
,
Turner
 
MJ
 et al.  
Simultaneous broadband vector magnetometry using solid-state spins
.
Phys Rev Appl
 
2018
;
10
:
034044
.

22.

Chen
 
B
,
Hou
 
X
,
Ge
 
F
 et al.  
Calibration-free vector magnetometry using nitrogen-vacancy center in diamond integrated with optical vortex beam
.
Nano Lett
 
2020
;
20
:
8267
72
.

23.

Waldherr
 
G
,
Beck
 
J
,
Neumann
 
P
 et al.  
High-dynamic-range magnetometry with a single nuclear spin in diamond
.
Nat Nanotechnol
 
2012
;
7
:
105
8
.

24.

Acosta
 
VM
,
Bauch
 
E
,
Ledbetter
 
MP
 et al.  
Diamonds with a high density of nitrogen-vacancy centers for magnetometry applications
.
Phys Rev B
 
2009
;
80
:
115202
.

25.

Taylor
 
JM
,
Cappellaro
 
P
,
Childress
 
L
 et al.  
High-sensitivity diamond magnetometer with nanoscale resolution
.
Nat Phys
 
2008
;
4
:
810
6
.

26.

Hilberer
 
A
,
Toraille
 
L
,
Dailledouze
 
C
 et al.  
Enabling quantum sensing under extreme pressure: nitrogen-vacancy magnetometry up to 130 GPa
.
Phys Rev B
 
2023
;
107
:
L220102
.

27.

Huang
 
H
,
Dong
 
H
,
Hu
 
X
 et al.  
Three-axis atomic magnetometer based on spin precession modulation
.
Appl Phys Lett
 
2015
;
107
,
182403

28.

Ben-Kish A and Romalis
 
M
.
Dead-zone-free atomic magnetometry with simultaneous excitation of orientation and alignment resonances
.
Phys Rev Lett
 
2010
;
105
:
193601
.

29.

Toyli
 
DM
,
Weis
 
CD
,
Fuchs
 
GD
 et al.  
Chip-scale nanofabrication of single spins and spin arrays in diamond
.
Nano Lett
 
2010
;
10
:
3168
72
.

30.

Cui
 
W
.
An overview of submersible research and development in China
.
J Mar Sci Appl
 
2018
;
17
:
459
70
.

31.

Clevenson
 
H
,
Trusheim
 
ME
,
Teale
 
C
 et al.  
Broadband magnetometry and temperature sensing with a light-trapping diamond waveguide
.
Nat Phys
 
2015
;
11
:
393
7
.

32.

Tong
 
Y
,
Zhang
 
W
,
Qin
 
X
 et al.  
A customized control and readout device for vector magnetometers based on nitrogen-vacancy centers
.
Rev Sci Instrum
 
2023
;
94
:
014709
.

33.

Doherty
 
MW
,
Dolde
 
F
,
Fedder
 
H
 et al.  
Theory of the ground-state spin of the NV center in diamond
.
Phys Rev B
 
2012
;
85
:
205203
.

34.

Doherty
 
MW
,
Manson
 
NB
,
Delaney
 
P
 et al.  
The nitrogen-vacancy colour centre in diamond
.
Phys Repo
 
2013
;
528
:
1
45
.

35.

Degen
 
C
,
Reinhard F and Cappellaro
 
P
.
Quantum sensing
.
Rev Mod Phys
 
2017
;
89
:
035002
.

36.

Barry
 
JF
,
Schloss
 
JM
,
Bauch
 
E
 et al.  
Sensitivity optimization for NV-diamond magnetometry
.
Rev Mod Phys
 
2020
;
92
:
015004
.

37.

Zhuang
 
M
,
Chen
 
S
,
Huang
 
J
 et al.  
Quantum vector DC magnetometry via selective phase accumulation
.
Sci China Phys Mech Astron
 
2024
;
67
:
100312
.

38.

Xie
 
Y
,
Yu
 
H
,
Zhu
 
Y
 et al.  
A hybrid magnetometer towards femtotesla sensitivity under ambient conditions
.
Sci Bull
 
2021
;
66
:
127
32
.

39.

Scheidegger
 
PJ
,
Diesch
 
S
,
Palm
 
ML
 et al.  
Scanning nitrogen-vacancy magnetometry down to 350 mk
.
Appl Phys Lett
 
2022
;
120
:
224001
.

40.

Acosta
 
VM
,
Bauch
 
E
,
Ledbetter
 
MP
 et al.  
Temperature dependence of the nitrogen-vacancy magnetic resonance in diamond
.
Phys Rev Lett
 
2010
;
104
:
070801
.

41.

Doherty
 
MW
,
Acosta
 
VM
,
Jarmola
 
A
 et al.  
Temperature shifts of the resonances of the NV center in diamond
.
Phys Rev B
 
2014
;
90
:
041201
.

42.

Dolan
 
PR
,
Li
 
X
,
Storteboom
 
J
 et al.  
Complete determination of the orientation of NV centers with radially polarized beams
.
Opt Express
 
2014
;
22
:
4379
87
.

43.

Zhang
 
C
,
Yuan
 
H
,
Zhang
 
N
 et al.  
Vector magnetometer based on synchronous manipulation of nitrogen-vacancy centers in all crystal directions
.
J Phys D: Appl Phys
 
2018
;
51
:
155102
.

44.

Clevenson
 
H
,
Pham
 
LM
,
Teale
 
C
 et al.  
Robust high-dynamic-range vector magnetometry with nitrogen-vacancy centers in diamond
.
Appl Phys Lett
 
2018
;
112
:
252406
.

45.

Frontera
 
P
,
Alessandrini
 
S
,
Stetson
 
J
.
Shipboard calibration of a diamond nitrogen vacancy magnetic field sensor
. In:
2018 IEEE/ION Position, Location and Navigation Symposium (PLANS)
.
Piscataway, NJ
:
IEEE Press
,
2018
,
497
504
.

46.

Fleig T and Frontera
 
P
.
Maritime magnetic anomaly mapping with a diamond nitrogen vacancy sensor
. In:
2018 IEEE/ION Position, Location and Navigation Symposium (PLANS)
.
Piscataway, NJ
:
IEEE Press
,
2018
,
1107
12
.

47.

Wang
 
X
,
Li
 
W
,
Moran
 
B
 et al.  
Quantum diamond magnetometry for navigation in GNSS denied environments
. In:
Gravity, Positioning and Reference Frames (REFAG 2022)
.
Cham
:
Springer
,
2023
,
87
92
.

48.

Li
 
G
,
Chen
 
X
,
Zhou
 
F
 et al.  
Self-powered soft robot in the Mariana Trench
.
Nature
 
2021
;
591
:
66
71
.

49.

Weggler
 
T
,
Ganslmayer
 
C
,
Frank
 
F
 et al.  
Determination of the three-dimensional magnetic field vector orientation with nitrogen vacancy centers in diamond
.
Nano Lett
 
2020
;
20
:
2980
5
.

50.

Vetter
 
PJ
,
Marshall
 
A
,
Genov
 
GT
 et al.  
Zero- and low-field sensing with nitrogen-vacancy centers
.
Phys Rev Appl
 
2022
;
17
:
044028
.

51.

Fu
 
KMC
,
Iwata
 
GZ
,
Wickenbrock
 
A
 et al.  
Sensitive magnetometry in challenging environments
.
AVS Quantum Sci
 
2020
;
2
:
044702
.

52.

Pillai
 
A
,
Elanchezhian
 
M
,
Virtanen
 
T
 et al.  
Electron-to-nuclear spectral mapping via dynamic nuclear polarization
.
J Chem Phy
 
2023
;
159
:
154201
.

53.

Soshenko
 
VV
,
Bolshedvorskii
 
SV
,
Rubinas
 
O
 et al.  
Nuclear spin gyroscope based on the nitrogen vacancy center in diamond
.
Phys Rev Lett
 
2021
;
126
:
197702
.

54.

Ajoy A and Cappellaro
 
P
.
Stable three-axis nuclear-spin gyroscope in diamond
.
Phys Rev A
 
2012
;
86
:
062104
.

55.

Xu
 
M
,
Zhang
 
X
,
Chai
 
X
 et al.  
A marine salinity sensor based on boron-doped diamond film electrode
. In:
International Conference on Optoelectronic Materials and Devices (ICOMD 2021)
.
Bellingham, WA
:
Society of Photo-Optical Instrumentation Engineers
,
2022
,
30
7
.

Author notes

Equally contributed to this work.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.

Supplementary data