Abstract

Single chamber (SC) implantable cardioverter defibrillators (ICDs) have several limitations that might be relevant during follow-up, like atrial pacing requirements, inadequate therapies, sustained atrial tachyarrhythmias and difficulties to achieve an accurate diagnosis of the arrhythmia. Dual chamber (DC) ICDs offer an attractive and rational solution, although controversy remains if the costs and complexity of these devices offer a real clinical advantage. The Dual Chamber & Atrial Tachyarrhythmias Adverse Events Study (DATAS) was designed to analyze the ability of DC ICD, DDED, to reduce clinically significant adverse events compared with SC ICD in a non-selected population with conventional indications for ICD implantation. This is a prospective, multicentre, randomized, open labelled study, with three arms: two of them (simulated SC ICD and true DC ICD) cross-over, and the third (true SC ICD) parallels the other two. The composite primary end point comprises four Clinically Significant Adverse Events (CSAE): (1) all-cause mortality, (2) invasive intervention, hospitalization or prolongation of hospitalization due to cardiovascular cause, (3) inappropriate shocks, and (4) sustained symptomatic atrial tachyarrhythmias that (a) require urgent termination or (b) last more than 48 h leading to therapeutic intervention. Secondary end points constitute each of the individual components of CSAE, cardiovascular status, quality of life and a detailed analysis of atrial and ventricular arrhythmias. To date (June 2003) there have been 343 patients enroled from 947 screened patients. The projected enrolment includes 360 patients and the conclusion of the study is expected at the beginning of 2005.

Introduction

Controlled clinical studies have demonstrated that the implantable cardioverter defibrillator (ICD) is more effective than antiarrhythmic drugs in reducing death (AVID, MADIT, MUSTT) [1–,4], leading to widespread use of these devices. Nevertheless, presently available single chamber (SC) ICDs have at least four main limitations that might be relevant during follow-up: (1) loss of AV synchrony in patients requiring pacing [5,,6]; (2) inappropriate therapies [7,,8] despite the activation of so-called “additional criteria” [9]; (3) sustained atrial tachyarrhythmias for which no therapy is provided [10,,11]; (4) difficulties to achieve an accurate diagnosis of the arrhythmia from the stored electrograms [12,,13].

These problems might be circumvented by using standard antiarrhythmic therapies (drugs, electrical cardioversion, ablation) [14–,16] and/or dual chamber ICDs [17,,18]. Dual chamber (DC) ICDs offer an attractive solution, although there are some concerns about atrial lead related complications, value of atrial therapies and AV discrimination algorithm performance [19,,20]. Recent data from two controlled studies have questioned long-term safety and efficacy of these devices [21,,22]. In fact, to date there is only a consensus on implantation of DC ICD in patients with marked sinus dysfunction or 2nd–3rd degree AV [17,,23], and controversy remains if the costs and complexity of the implants offer a real clinical advantage.

The Dual Chamber & Atrial Tachyarrhythmias Adverse Events Study (DATAS) was designed to analyze the ability of DC ICD, DDED [24], to reduce clinically significant adverse events compared with SC ICD in a non-selected population with conventional indications for ICD implantation.

Justification of the study

Four main situations in which DC ICD is potentially beneficial in comparison with SC ICD can be anticipated to be:

  1. Loss of normal atrial rate and atrioventricular synchrony. Over 15% of ICD patients may need atrial pacing-sensing, half of them undetectable at the time of ICD pre-implant evaluation [25–,29]. Drugs commonly used to treat heart failure or arrhythmias, betablockers, sotalol and amiodarone [30,,31], frequently cause bradycardia at optimal doses for which VVI pacing may offer a suboptimal backup [32,,33].

  2. Inappropriate discrimination of supraventricular and ventricular events. Inappropriate therapies and shocks may hamper quality of life [34], cause proarrhythmia, hospitalization and even death [35,,36]. The percentage of patients with inadequate shocks in single chamber devices may reach 40% [7,,10], the most frequent causes being a fast ventricular rhythm caused by atrial fibrillation (AF) or sinus tachycardia. Nevertheless, although enhanced criteria (i.e. sudden onset, stability and width) can improve the specificity of the diagnostics to more than 90%, the sensitivity can be reduced to 80% or less [37–,39].

  3. Accuracy in the diagnosis of arrhythmias by stored electrograms. Analysis of morphology of stored ventricular electrograms can provide some clues as to the site of origin of arrhythmias [13,,40], however, the ability for such differential diagnosis is limited. Direct recording of atrial rhythm by a dedicated atrial lead should allow more accurate discrimination [41]. Recognition of more than one tachycardia (a condition not rare in ICD recipients) is not possible without atrial electrograms [42].

  4. Prevention and/or therapy of significantly prevalent atrial tachyarrhythmias. In the AVID population one fourth of the patients had AF [1], and the same high prevalence has been found in other series of patients after ICD implantation [7,,8]. Data from SOLVD studies suggest that atrial fibrillation is associated with increasing mortality by impairing left ventricular systolic function [43]. Therefore, prevention and management of AF may improve outcome. Atrial therapies, have demonstrated a high success rate in suppressing atrial arrhythmias [44–,48]. Efficacy of atrial antitachycardia pacing ranges from 55% to 66% and the adjusted success rate of atrial shock for atrial fibrillation is more than 60% [46]. Adequate timing of therapies for AF can avoid the dangerous consequences of conversion to sinus rhythm in patients without anticoagulation [49].

Reasons commonly used not to implant DC ICDs include [20,,50,,51]: (1) higher cost of DC devices, (2) higher complication rate, related to the insertion of an atrial lead, (3) prolongation of operation and hospitalization time, (4) higher complexity of follow-up procedures, (5) low compliance of patients with the atrial shocks, and (6) uncertainties about a clinically significant reduction in adverse events during follow-up.

The main issue to be clarified is whether tolerability and reduction of significant clinical problems of DC ICD are offset by the potential benefits in the majority of ICD patients. It should be emphasized that the investigators feel strongly that it is this clinical benefit that needs to be measured as opposed to “pure electrical phenomena” with uncertain clinical implications.

Study objectives

This clinical study is designed to assess the ability of DDED DC ICD to reduce clinically significant adverse events in patients with class I indication for VVEV-VVI conventional ICD therapy. The primary end point is a composite comprising four so-denominated “Clinically Significant Adverse Events” (CSAE): (1) all-cause mortality, (2) invasive intervention, hospitalization (>24 h) or prolongation of hospitalization due to cardiovascular cause, (3) inappropriate shocks (two or more episodes with inappropriate shocks), and (4) sustained symptomatic atrial tachyarrhythmias that (a) require urgent termination or (b) last more than 48 h leading to therapeutic intervention. CSAE will not include cardiac catheterization with/without revascularization scheduled routinely during follow-up. The primary objective of this trial is to determine whether use of DDED ICD results in a significant decrease in the number of primary end points.

Secondary end points constitute each of the individual components of CSAE, cardiovascular status (functional and echo parameters), quality of life and a detailed analysis of atrial arrhythmias (types, burden, time course, consequences, etc). Secondary objectives are:

  1. Number of each of the components of the CSAE.

  2. Arrhythmia related: atrial tachyarrhythmia, frequency and burden, ventricular tachyarrhythmia frequency and burden number of appropriate shocks, number of inappropriate shocks, need for reprogramming, need for medication/RFA for arrhythmia control, pacemaker syndrome and development of dual chamber pacing indication.

  3. Cardiovascular related: NYHA functional class, exercise capacity, left ventricular ejection fraction (LVEF), reduction of medication (diuretics…).

  4. Quality of life: evaluated by the SF-36, Minnesota living test, with heart failure and Symptom Checklist instruments.

Study design, randomization and data collection

The DATAS is a, prospective, multicentre, randomized, open labelled study, with three arms (two of them cross-over and the third parallels the other two) (Fig. 1). About 36 centres in Spain, Germany, Italy, United Kingdom, Portugal and Israel will participate. The projected enrolment includes 360 patients. The first implant took place in November 2000 and the duration of the study (including patient enrolment, follow-up, data analysis, and regulatory approval) was expected to be 48 months. To date (June 2003) there are 343 patients enroled from 947 screened patients. Documented approval of the investigational plan by the Institutional Review Board (IRB) or Ethics Committee (EC) affiliated with the study centre is required for starting the study. Patient written informed consent is also required before enrolment.

Elements of the cross-over design. SC = single chamber; DC = dual chamber; ARS = atrial rate stabilization; atrial ATP = atrial therapies (ATP + shock); stab = stability.
Figure 1

Elements of the cross-over design. SC = single chamber; DC = dual chamber; ARS = atrial rate stabilization; atrial ATP = atrial therapies (ATP + shock); stab = stability.

A web site (http://www.rdes.com/datas) has been specifically designed for this study by an independent company (Remote Data Entry System S.L., Barcelona, Spain). Through this site the randomization and data entry with electronic case report forms are completed and sent to a central database, where the Data Revision Committee can review them.

The study design is depicted in Fig. 1. Patients will be randomized to one of three arms: SC ICD (“SC true arm”), DC ICD initially programmed as SC (SC simulated) ICD (“SC sim arm”) and DC ICD initially programmed as a DDED (“DC true arm”).

No waiting period after implant has been allowed in order to evaluate the lead fixation issues, and the possible beneficial mechanisms for atrial prevention pacing algorithms. On the other hand a 1-month wash out period after the cross-over is included to avoid influence of the remodeling associated with pacing mode.

The randomization procedure will be centralized through the web site, balancing the sample size of all three arms over the different centres. At the beginning of the study there will be at least six consecutive patients to be enroled by a participating centre. A log of all excluded patients is kept in each centre to avoid any possible selection bias. Later enrolment must also include at least three consecutive patients.

Patients will be followed-up at 1, 4, 8, 9, 13 and 17 months. At enrolment, 8 and 17 months, the quality of life questionnaires, 6 min walk test and echo parameters will be recorded, as well as reevaluation of the cardiovascular history.

An Adverse Events Advisory Committee has been created to analyze blindly: (1) the classification of adverse events and validation of primary end points; (2) the classification and validation of device episodes related to adverse events.

Patient eligibility criteria

All patients' candidates for ICD will be initially screened in each centre (Fig. 1). The study population will include those patients who are appropriate candidates for standard Class I criteria for a single chamber implantable cardioverter defibrillator (ICD).

Exclusion criteria will be the accepted indications (symptomatic sinus node disease, all 2nd AV block, except asymptomatic Mobitz I, and all 3rd degree AV block) and contraindications (permanent atrial tachyarrhythmias) for dual chamber pacing. Patients without structural heart disease, an indication for biventricular pacing or with a previous system implanted (ICD or pacemaker) will also be excluded.

ICD devices and programming

DDED devices are to be used for this investigation, which must have automatic atrial tiered antitachycardia therapies and may be programmed in monitor mode to store data from atrial arrhythmic events without delivering atrial therapy. Devices used in the trial included Jewel AF (model 7250) and GEM III AT (model 7276; Medtronic Inc., Minneapolis, MN, USA) as dual chamber devices and GEM family (Medtronic Inc, Minneapolis, MN, USA) as single chamber devices. The dual chamber algorithm (PR Logic) discriminates between supraventricular and ventricular tachycardias, and detects AT/AF accurately and continuously [45,,52].

The atrial lead could be any commercially available bipolar atrial pace/sense lead with tip-ring spacing ≤ 10 mm, and the ventricular lead could be any commercially available ventricular pace/sense leads with two defibrillation coils.

Intraoperative implantation procedures will be performed according to the standards of each centre. Programming the ICD should also meet clinical standards in every centre, but some general rules were strongly recommended:

  • Supraventricular arrhythmia discrimination in SC by using stability criteria

  • A VT zone with ATP therapies

  • Low pacing rates (50 bpm or less) in SC and, in DC, 70 bpm was encouraged to prevent atrial tachyarrhythmias, but with prolonged AV intervals (recommended values: paced AV 230 ms and sensed AV 200 ms) in an attempt to reduce right ventricle pacing.

  • Therapies for atrial arrhythmias were programmed as follows: nominal pacing therapies with no delay after detection; for atrial arrhythmias as soon as detected, first high energy shock automatically delivered in less than 24 h after episode onset, and programming for subsequent episodes at the investigator's discretion and according to patient tolerance.

Statistical analysis

The main objective is reduction of the number of CSAE. The primary response variable is the sum of the four CSAEs. The main comparison is SC ICD (SC true) versus DDD ICD (DC true). It should be noted that: (a) the DC true group will double the number of patients since it will pool together patients from the two cross-over arms (DC true SC sim and SC sim-DC true) and (b) the SC true group will be followed for a longer period of time (17 months). In order to get a single response measure, the addition of the number of any of these four CSAE was defined as the primary end point. This variable will be sensitive both to reductions in the number of patients with CSAE as well as in the number of CSAEs per patient. SC simulated will be used for a precise evaluation of the impact of dual chamber therapy on atrial tachyarrhythmia burden and SVT/VT discrimination.

The main statistical analysis (p-values and confidence intervals) will be based on the ordinal Mann–Whitney–Wilcoxon statistical method [53]. To take into account the different follow-up for the two groups (twice 8 months versus 17 months), the ranks will be calculated for the rate of CSAE instead of the proportion of CSAE.

Sample size and power for the main analysis

The assumed effect of the DC treatment is a reduction of 15% both in the proportion (Π) of patients with tendency to develop a CSAE (from 30% to 15%), as well as in the mean (M) of CSAE over all patients (from 6 to 5.1). The estimated sample size was 200 (DC true) versus 100 (SC true) patients followed for 8 months, with a two sided α=0.05, and a power of 88.8%. The power of the study has been explored through simulation of 10,000 samples with the SAS system. The number of replications allows for a confidence interval (1−α=0.95) in the estimation of the power with an amplitude equal to ±1%. To protect against dropouts, observational compliance, losses to follow-up, or protocol deviations the sample size was fixed at 360 patients (120 patients per arm).

Other considerations

If the treating physician decides that a patient could benefit from programming to the opposite treatment arm, the premature cross-over should be authorized by the adverse event committee. The intention to treat principle will be assured by counting CSAE in the original arm as it represents a normal medical situation of success and failure of delivering the planned therapy. The main analysis will be fully specified in a blind statistical analysis plan approved by the steering committee before the end of the trial.

Discussion

Key issues in the study design

Since the primary objective is to compare DDED ICD (DC true) to VVEV ICD (SC true) in terms of reducing CSAE, at first glance, a simple two-arm comparison might suffice. However, some of the questions related to atrial arrhythmias and VT/SVT discrimination could only be answered with direct information about atrial activity, i.e. through direct recordings of atrial electrograms. In addition, a cross-over design provides intraindividual comparisons. These considerations would argue in favour of a study design in which all patients get a DC device but half of them are initially programmed as an SC device (as far as therapy is concerned but providing full information about events), crossing over later to DC therapy. Nevertheless, this study design would penalize the “SC programmed” population with all the risks associated with DC devices, mainly a more complex implantation procedure and a potential for a higher surgical and electrode-related complication rate. In addition, it is conceivable that the atrial electrode could transiently “irritate” the atrial chambers.

Therefore a three-arm design was chosen, having a true SC arm and two DC arms. One DC arm will be programmed as an “SC only therapy” (or “simulated SC arm”) and the other as a DC therapy (“true DC arm”). These two DC arms will undergo a cross-over programming after half of the projected follow-up period (8 months). With this design all three groups of therapy may be compared in similar conditions for number and length of follow-up, and the very important aspects included in the secondary objectives will be appropriately analyzed.

Other studies and DATAS

Two somewhat similar studies have recently been published. The DAVID trial [22] assessed the hypothesis that dual chamber pacing was superior to backup ventricular pacing in patients with standard indications for ICD implantation and LVEF of 40% or less without indications for antibradycardia pacing. The study was prematurely stopped because the conditional power for the original alternative (DDDR-70 being better than VVI-40) was less than 10%. In fact, “…a trend (p<0.03) towards a worse outcome with DDDR-70 pacing was established…”. It was also found that “Patients who survived to the 3-month follow-up have worse 12-month event-free rates when the percentage of right ventricular pacing by ICD interrogation was 41–100% (75.9%) than when less than 40% (86.9%) (p=0.09)”. The authors feel that their own along with recently published data [54] suggest that right ventricular stimulation is inappropriate and causes increased heart failure by the mechanism of ventricular desynchronization. The specific programming choices (AV interval was not controlled) could have affected their results, since nearly 60% of all ventricular beats were paced in the DDDR-70 group.

In the study of Deisenhofer et al. [21], 92 patients were randomly assigned to a VVI-ICD or a DDD ICD during a mean follow-up of 7.5 months. Although DDD ICDs allowed a better rhythm classification, the applied detection algorithms offer no advantages in avoiding inappropriate therapies during SVTs. It should be emphasized that in these series, 75% of inappropriate therapies in the DDD ICD group were due to atrial sensing problems, either oversensing or undersensing. This seems to be an unexpectedly high rate. In the DATAS trial we will enrol patients implanted in 36 European centres and we will be able to assess the actual rate of atrial lead related complications in a large number of expert centres.

In contrast, in DATAS: (a) the recommended programming includes an AV interval of 200 ms (instead of 180 ms) or whatever was needed to avoid ventricular pacing; (b) the devices can treat atrial tachyarrhythmias electrically; (c) no cut-off value of LVEF was required; (d) resynchronization indication was an exclusion criterion; and (e) each arm will have a 17 month follow-up.

Final considerations

When the DAVID was published, the Steering Committee of DATAS felt that the potentially deleterious effect of ventricular pacing should be checked. The mean percentage of ventricular paced beats in the 62 patients assigned to the DDD arm, who completed the first 8-month follow-up period, was 36%, far below the 60% found in DAVID, and below the 40% security limit described in MOST and DAVID. Since safety is not of concern in DATAS, and in view of some design differences (i.e., a true VVI arm) there is no reason to stop the study.

The DATAS design (with a VVI true arm and longer follow-up) will provide important information about the effects of AV discrimination algorithms, atrial antibradycardia and antitachycardia therapies (with long AV delays to prevent unnecessary ventricular stimulation) and will help to define the actual indications for dual chamber defibrillators.

This work is supported by a grant from Medtronic Inc (Minneapolis, USA).

Appendix

DATAS Participating Investigators, Centers and Enrolments: B. Lüderitz, J. Schwab, T. Lewalter, R. Schimpf, University Hospital, Bonn, Germany (30); M. Santini, R. Ricci, C. Pignalberi, M. Russo, San Filippo Neri, Rome, Italy (27); P. Hanrath, Ch. Stellbrink, K. Mischke, R. Koos, University Hospital RWTH, Aachen, Germany (24); J. Brugada, L. Mont, M. Matas, H. Clinic I Provincial, Barcelona, Spain (18); J. Gill, R. Simon, A. Rinaldi, N. Gall, St Thomas' Hospital, London, UK (18); M. Glikson, Sheba Medical Center, Tel-Hashomer, Israel (15); C. Muto, M. Canciello, G. Carreras, B. Tuccillo, Loreto Mare Hospital, Naples, Italy (15); J. Roda, S. Villalba, V. Palanca, J. Belchi, H. General Universitario, Valencia, Spain (15); M. Borggrefe, S. Spehl, 1st Department of Medicine–Cardiology, University Hospital Mannheim, Mannheim, Germany (12); J.L. Merino, R. Peinado, H. La Paz, Madrid, Spain (12); A. Arenal, E. Gonzalez-Torrecillas, F. Atienza, H. Gregorio Marañon, Madrid, Spain (11); J.C. Rodriguez, O. Medina, J. García, H. Insular de Gran Canaria, Las Palmas, Spain (10); G. Steinbeck, E. Hoffman, S. Janko, U. Dorwarth, Ludwig-Maximilian-University Hospital, München, Germany (9); F. Morgado, Santa Cruz, Lisbon, Portugal (9); I. Lozano, J. Toquero, R. Arroyo, H. Puerta de Hierro, Madrid, Spain (9); J.M. Ormaetxe, M. Arkotxa, H. de Basurto, Bilbao, Spain (9); M. Geist, V. Turkisher, Wolfson Medical Center, Holon, Israel (6); P. Della Bella, G. Fassini, C. Carbucicchio, F. Giraldi, Centro Cardiologico Monzino, Milano, Italy (6); P. Golino, M. Viscusi, F. Mascia, Hospedale Civile, Caserta, Italy (6); L. Tercedor, M. Alvarez, H. Virgen de las Nieves, Granada, Spain (6); J.G. Martinez, A. Ibañez, H. General Universitario, Alicante, Spain (6); A. Moya, E. Rodriguez, C. Alonso, H. Valle Hebron, Barcelona, Spain (6); M. Lopez Gil, J. Sanz, H. 12 Octubre, Madrid, Spain (6); R. Garcia-Civera, R. Ruiz, S. Morell, R. SanJuan, H. Clinico Universitario, Valencia, Spain (6); A. García-Alberola, J. Martinez, J.J. Sanchez, H. Virgen de la Arrixaca, Murcia, Spain (6); S. Hohnloser, G. Grönefeld, Johann Wolfgang Goethe University, Frankfurt, Germany (5); J.J. Manzano, A. Medina, E. Caballero, F. Wangüemert, H. General Dr. Negrín, Las Palmas, Spain (5); B. Lemke, T. Lawo, T. Deneke, S. Holt, BG Kliniken Bergmannsheil, Bochum, Germany (5); M. Manz, D. Burkhardt, A. Markewitz, Krankenhaus Marienhof, Koblenz, Germany (5); E. Castellanos, L. Rodriguez-Padial, H. Virgen de la Salud, Toledo, Spain (5); W. Jung, N. Schwick, B. Roggenbuck-Schwilk, Klinikum Villingen-Schwenningen, Villingen, Germany (4); T Fuchs, Assaf Harofe Medical Center, Tzerifin, Israel (4); M. Sassara, A. Achilli, E. Scabbia, Civile Hospital, Viterbo, Italy (4); J. Olagüe, J.E. Pareja, M.J. Sancho-Tello, H. La Fe, Valencia, Spain (4); A. Maresta, S. Silvani, D. Cornacchia, E. Tampieri, Civile Hospital, Ravenna, Italy (3); G. Baumann, H. Bondke, M. Claus, Campus Charite Mitte, Berlin, Germany (3).

DATAS Data Coordination and Review Committee: M. Ortiz, Madrid, Spain.

DATAS Adverse Event Committee: J.A. Garcia Robles, Madrid, Spain; T. Korte, Hannover, Germany; X. Viñolas, Barcelona, Spain; M. Zoni Berisso, Genova, Italy.

DATAS Study Statistician and Analyst: E. Cobo, Dept. Estadística e Investigación Operativa, Universidad Politécnica de Catalunya, Spain.

DATAS Steering Committee: A. Quesada, Valencia, Spain; J. Almendral, F. Arribas, Madrid, Spain; M. Santini, Rome, Italy; C. Wolpert, Mannheim, Germany; P. Adragao, Carnaxide, Portugal; X. Navarro, Barcelona, Spain.

References

[1]

The AVID investigators
A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias
N Engl J Med
1997
337
1576
1583

[2]

Moss
A.J.
 
Hall
J.
 
Cannom
D.S.
 et al.
Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmias
N Engl J Med
1996
335
1933
1940

[3]

Buxton
A.E.
 
Lee
K.L.
 
Fisher
J.D.
 et al.
A randomized study of the prevention of sudden death in patients with coronary artery disease
N Engl J Med
1999
341
1882
1890

[4]

Connolly
S.J.
 
Hallstrom
A.
 
Cappato
R.
 et al. on behalf of
the investigators or the AVID, CASH and CIDS studies
Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials
Eur Heart J
2000
21
2071
2078

[5]

Santini
M.
 
Ansalone
G.
 
Auriti
A.
 
Magris
B.
 
Pandozi
C.
 
Altamura
G.
Indications for dual chamber (DDD) pacing in ICD patients
Am J Cardiol
1996
78
116
118

[6]

Best
P.J.
 
Hayes
D.L.
 
Stanton
M.S.
The potential usage of dual chamber pacing in patients with implantable cardioverter defibrillators
Pacing Clin Electrophysiol
1999
22
79
85

[7]

Grimm
W.
 
Flores
B.
 
Marchlinski
F.E.
Electrocardiographically documented unnecessary, spontaneous shocks in 241 patients with implantable cardioverter defibrillators
Pacing Clin Electrophysiol
1992
15
1667
1673

[8]

Ryan
S.
 
Siemon
G.
 
Drogemuller
A.
 et al.
2 year follow-up of 321 patients with an implantable cardioverter-defibrillator: comparison of patients with and without atrial fibrillation
Z Kardiol
2001
90
906
915

[9]

Nunain
S.O.
 
Roelke
M.D.
 
Trouton
T.
 et al.
Limitations and late complications of third generation automatic cardioverter defibrillators
Circulation
1995
91
2204
2213

[10]

Schmitt
C.
 
Montero
M.
 
Melichercik
J.
Significance of supraventricular tachyarrhythmias in patients with implanted pacing cardioverter defibrillators
Pacing Clin Electrophysiol
1994
17
295
302

[11]

Neuzner
J.
 
Petschner
H.F.
Managing the problem of atrial tachyarrhythmias in patients with ICDs. Practice and progress in cardiac pacing and electrophysiology
1996
Kluwer Academic Publishers
353
360

[12]

Neuzner
J.
 
Pitschner
H.F.
 
Konig
S.
 
Stohring
R.
 
Schlepper
M.
Stored electrogram in cardioverter-defibrillator therapy: accuracy of rhythm diagnosis in 335 spontaneous arrhythmia episodes
J Ambulatory Monit
1995
8
1
9

[13]

Marchlinski
F.E.
 
Callans
D.J.
 
Gottlieb
C.D.
 
Schwartzman
D.
 
Preminger
M.
Benefits and lessons learned from stored electrogram information in implantable defibrillators
J Cardiovasc Electrophysiol
1995
832
851

[14]

Epstein
A.E.
 
Kay
G.N.
 
Plumb
V.J.
 
Shepard
R.B.
 
Kirklin
J.K.
Combined automatic implantable cardioverter defibrillator and pacemaker systems: implantation techniques and follow-up
J Am Coll Cardiol
1989
13
121
131

[15]

Calkins
H.
 
Brinker
J.
 
Veltri
E.
 
Levine
J.H.
Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillators
J Am Coll Cardiol
1990
16
666
673

[16]

Hook
B.G.
 
Callans
D.J.
 
Kleiman
R.B.
 
Flores
B.
 
Marchlinski
F.E.
Implantable cardioverter defibrillators therapy in the absence of significant symptoms. Rhythm diagnosis and management aided by stored electrogram analysis
Circulation
1993
87
1897
1906

[17]

Trappe
H.
 
Achtelik
M.
 
Pfitzner
P.
 
Voigt
B.
 
Weismuller
P.
Single-chamber versus dual-chamber implantable cardioverter defibrillator: indications and clinical results
Am J Cardiol
1999
83
8D
16D

[18]

Osswald
S.
 
Cron
T.A.
 
Sterns
L.
 
Alt
E.
 
Stotts
L.
 
Coenen
M.
New ICD-technologies: first clinical experience with dual-chamber sensing for differentiation of supraventricular tachyarrhythmias
Pacing Clin Electrophysiol
1998
21
292
295

[19]

Olson
W.
Singer
I.
 
Barold
S.S.
 
Camm
A.J.
Dual chamber sensing and detection for implantable cardioverter defibrillators
Nonpharmacological therapy of arrhythmias for the 21st century: the state of the art
1998
Armonk, NY
Futura Publishing Co, Inc
385
395

[20]

Sticherling
C.
 
Schaumann
A.
 
Klingenheben
T.
 
Hohnloser
S.
 
for the Ventak AV II DR investigators
First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator
Europace
1999
1
96
99

[21]

Deisenhofer
I.
 
Kolb
Ch.
 
Ndrepepa
G.
 et al.
Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study
J Cardiovasc Electrophysiol
2001
12
134
142

[22]

The dual chamber and VVI implantable defibrillator (DAVID) trial. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator
J Am Med Assoc
2002
288
3115
3123

[23]

Gregoratos
G.
 
Abrams
J.
 
Epstein
A.E.
 et al.
ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines)
Circulation
2002
106
2145
2161

[24]

Bernstein
A.D.
 
Camm
A.J.
 
Fisher
J.D.
 et al.
The North American Society of Pacing and Electrophysiology policy statement. The NASPE/BPEG Defibrillation Code
Pacing Clin Electrophysiol
1993
16
1776
1780

[25]

Geelen
P.
 
Lorga
A.
 
Chauvin
M.
 
Wellens
F.
 
Brugada
P.
The value of DDD pacing in patients with an implantable cardioverter defibrillator
Pacing Clin Electrophysiol
1997
20
177
181

[26]

Mattke
S.
 
Markewitz
A.
 
Muller
D.
 et al.
The combined transvenous implantation of cardioverter defibrillators and permanent pacemakers
Pacing Clin Electrophysiol
1997
20
2775
2782

[27]

Higgins
S.L.
 
Williams
S.K.
 
Pak
J.P.
 
Meyer
D.B.
Indications for implantation of a dual chamber pacemaker combined with an implantable cardioverter-defibrillator
Am J Cardiol
1998
81
1360
1362

[28]

Sutandar
A.
 
Solimon
S.
 
Kaddaha
R.
 
Vloka
M.
 
Steinberg
J.
Should all patients receive a dual chamber device? Unanticipated use of the implantable defibrillator as a pacemaker
J Am Coll Cardiol
2000
35
Suppl A
127

[29]

Proclemer
A.
 
Della Bella
P.
 
Facchin
D.
 et al.
Indications for dual-chamber cardioverter defibrillators at implant and at 1 year follow-up: a retrospective analysis in the single-chamber defibrillator era
Europace
2001
3
132
135

[30]

Pacífico
A.
 
Hohnloser
S.
 
Williams
J.
 
Tao
B.
 
Saksena
S.
 
Henry
P.
 et al.
Prevention of implantable-defibrillator shocks by treatment with sotalol
N Engl J Med
1999
340
1855
1862

[31]

Packer
M.
 
Coats
A.
 
Fowler
M.
 et al.
Effect of carvedilol on survival in patients with severe chronic heart failure
N Engl J Med
2001
344
1651
1658

[32]

Reiter
M.J.
 
Hindman
M.C.
Hemodynamic effects of acute atrioventricular sequential pacing in patients with sick sinus syndrome: the hidden benefits of dual-chamber pacing
J Am Coll Cardiol
1992
19
1542
1549

[33]

Andersen
H.R.
 
Nielsen
J.C.
 
Thomsen
P.E.B.
 
Thuesen
L.
 
Mortensen
P.T.
 
Vesterlund
T.
 et al.
Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome
Lancet
1997
350
1210
1216

[34]

Lüderitz
B.
 
Jung
W.
 
Deister
A.
 
Marneros
A.
 
Manz
A.
Patient acceptance of the implantable cardioverter defibrillator in ventricular tachyarrhythmias
Pacing Clin Electrophysiol
1993
16
1815
1821

[35]

Jones
G.
 
Johnson
G.
 
Troutman
C.
 et al.
Incidence of atrial fibrillation following ventricular defibrillation with transvenous lead systems in man
J Cardiovasc Electrophysiol
1992
411
417

[36]

Birgersdotter-Green
U.
 
Rosenqvist
M.
 
Lindemans
F.
 
Rydén
L.
 
Rådegran
K.
Holter documented sudden death in a patient with an implanted defibrillator
Pacing Clin Electrophysiol
1992
15
1008
1014

[37]

Neuzner
J.
 
Pitschner
H.F.
 
Schlepper
M.
Programmable detection enhancements in implantable cardioverter defibrillator therapy
Pacing Clin Electrophysiol
1995
18
539
547

[38]

Unterberg
C.
 
Stevens
J.
 
Vollmann
D.
 
Hasenfuss
G.
 
Buchwald
A.B.
Long term clinical experience with the EGM width detection criterion for differentiation of supraventricular and ventricular tachycardia in patients with implantable cardioverter defibrillators
Pacing Clin Electrophysiol
2000
23
1611
1617

[39]

Brugada
J.
Is inappropriate therapy a resolved issue with current implantable cardioverter defibrillators?
Am J Cardiol
1999
83
40D
44D

[40]

Roelke
M.
 
Garan
H.
 
McGovern
B.A.
 
Ruskin
J.
Analysis of the initiation of spontaneous ventricular tachycardia by stored electrograms
J Am Coll Cardiol
1994
23
117
122

[41]

Dijkman
B.
 
Wellens
H.J.
Importance of the atrial channel for ventricular arrhythmia therapy in the dual chamber implantable cardioverter defibrillator
J Cardiovasc Electrophysiol
2000
11
1309
1319

[42]

Stein
K.M.
 
Euler
D.E.
 
Mehra
R.
 et al.  
Jewel AF Worldwide Investigators
Do atrial tachyarrhythmias beget ventricular tachyarrhythmias in defibrillator recipients?
J Am Coll Cardiol
2002
40
335
340

[43]

Dries
D.L.
 
Exner
D.V.
 
Gersh
B.J.
 
Domanski
M.J.
 
Waclawiw
M.A.
 
Stevenson
L.W.
Atrial fibrillation is associated with an increased risk for mortality and heart failure progression in patients with asymptomatic and symptomatic left ventricular systolic dysfunction: a retrospective analysis of the SOLVD trials. Studies of left ventricular dysfunction
J Am Coll Cardiol
1998
32
695
703

[44]

Wellens
H.J.J.
 
Lau
C.P.
 
Lüderitz
B.
 
Akhart
M.
 
Waldo
A.L.
 
Camm
A.J.
 et al.
Atrioverter: an implantable device for treatment of atrial fibrillation
Circulation
1998
98
1651
1656

[45]

Schoels
W.
 
Swerdlow
C.D.
 
Jung
W.
 
Stein
K.M.
 
Seidl
K.
 
Haffajee
C.J.
 
The Worldwide Jewel AF Investigators
Worldwide clinical experience with a new dual-chamber implantable cardioverter defibrillator system
J Cardiovasc Electrophysiol
2001
12
521
528

[46]

Ricci
R.
 
Pignalberi
C.
 
Disertori
M.
 et al.
Efficacy of a dual chamber defibrillator with atrial antitachycardia functions in treating spontaneous atrial tachyarrhythmias in patients with life-threatening ventricular tachyarrhythmias
Eur Heart J
2002
23
1471

[47]

Josephson
M.E.
New approaches to the management of atrial fibrillation: the role of the atrial defibrillator
Circulation
1998
98
1594
1596

[48]

Adler
S.W.
 
Wolpert
C.
 
Warman
E.N.
 
Musley
S.K.
 
Koehler
J.L.
 
Euler
D.E.
 
for the Worldwide Jewel AF Investigators
Efficacy of pacing therapies for treating atrial tachyarrhythmias in patients with ventricular arrhythmias receiving a dual-chamber implantable cardioverter defibrillator
Circulation
2001
104
887
892

[49]

Shenasa
H.
 
Shenasa
M.
Conversion of atrial fibrillation to sinus rhythm during implantable cardioverter-defibrillator discharges: incidence and implications
Eur Heart J
1996
17
Abstr Suppl
479

[50]

Kosar
E.
 
Nademance
K.
Complications in the implantation of dual chamber ICD
Cardiac Electrophysiol Rev
2001
5
108
109

[51]

Kühlkamp
V.
 
Dörnberger
V.
 
Mewis
C.
 
Suchalla
R.
 
Bosch
R.
 
Seipel
L.
Clinical experience with the new detection algorithms for atrial fibrillation of a defibrillator with dual chamber sensing and pacing
J Cardiovasc Electrophysiol
1999
10
905
915

[52]

Swerdlow
C.D.
 
Schsls
W.
 
Dijkman
B.
 
Jung
W.
 
Sheth
N.V.
 
Olson
W.H.
 et al.  
for the Worldwide Jewel AF Investigators
Detection of atrial fibrillation and flutter by a dual-chamber implantable cardioverter-defibrillator
Circulation
2000
101
878
885

[53]

Rahlfs
V.
 
Zimmermann
H.
H scores: ordinal data with few categories—how should they be analyzed?
Drug Information J
1993
27
1227
1240

[54]

Moss
A.J.
 
Zareba
W.
 
Hall
W.J.
 et al.
Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction
N Engl J Med
2002
346
877
883

Author notes

*The International Standard Randomised Controlled Trial Number is ISRCTN82728969.