Abstract

Undesirable sensing of external sources of electromagnetic interference by ICDs is well known. A transcutaneous electrical nerve stimulation (TENS) device has been reported to interfere with an ICD resulting in an inappropriate shock and patients with implanted defibrillators or pacemakers have been cautioned about the use of such units to treat chronic pain. We describe a patient regularly using TENS therapy for pain who subsequently received a biventricular ICD for malignant ventricular arrhythmias and medically refractory cardiac failure. He underwent testing for device interaction immediately post-implant. This did not show inappropriate sensing by either ICD or pacemaker component of his heart failure device. However, six months later, the patient complained of dizziness and bradycardia with application of TENS. Further testing did reveal interference with pacemaker function. Thus, even if initial testing is negative and reassuring, patients with a biventricular ICD still require careful follow-up for potential interaction and should be cautioned against the use of TENS, especially if they are pacemaker-dependent.

References

[1]
Philbin
DM
Marieb
MA
Aithal
KH
Schoenfeld
MH
,
Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit
Pacing Clin Electrophysiol
,
1998
, vol.
21
(pg.
2010
-
2011
)
[2]
Mannheimer
C
Carlsson
CA
Emanuelsson
H
Vedin
A
Waagstein
F
Wilhelmsson
C
,
The effects of transcutaneous electrical nerve stimulation in patients with severe angina pectoris
Circulation
,
1985
, vol.
71
(pg.
308
-
316
)
[3]
Cazeau
S
Ritter
P
Bakdach
S
et al.
,
Four chamber pacing in dilated cardiomyopathy
Pacing Clin Electrophysiol
,
1994
, vol.
17
(pg.
1974
-
1979
)
[4]
Rasmussen
MJ
Hayes
DL
Vlietstra
RE
Thorsteinsson
G
,
Can transcutaneous electrical nerve stimulation be safely used in patients with permanent cardiac pacemakers?
Mayo Clin Proc
,
1988
, vol.
63
(pg.
443
-
445
)
[5]
Chen
D
Philip
M
Philip
P
Monga
TN
,
Cardiac pacemaker inhibition by transcutaneous electrical nerve stimulation
Arch Phys Med Rehabil
,
1990
, vol.
71
(pg.
27
-
30
)