Abstract

Aims

The Lead EvaluAtion for Defibrillation and Reliability (LEADR) trial evaluated the small-diameter (4.7 Fr), integrated bipolar OmniaSecure defibrillation lead. As previously reported, the trial exceeded primary safety and efficacy objective thresholds, demonstrating favourable performance and zero fractures through ∼12 months follow-up, with patients in ongoing follow-up. Longer-term follow-up of the LEADR trial with emphasis on the sensing and detection capabilities of the OmniaSecure lead is reported here.

Methods and results

Patients with indications for de novo implantable cardioverter-defibrillators/cardiac resynchronisation therapy defibrillators were implanted with the OmniaSecure lead in standard right ventricle (RV) locations and followed at pre-specified intervals along with CareLink™ remote monitoring transmissions, where available. Throughout follow-up, the lead was evaluated for safety, efficacy, and reliability along with sensing and detection performance. There were 643/657 (97.9%) patients successfully implanted with the OmniaSecure lead with mean follow-up of 18.2 ± 5.5 months. There was a 96.9% freedom from major study lead-related complications at 24 months. Inappropriate shock rate was 2.7 and 3.8% at 12 and 24 months, respectively. At 24 months, 17.6% of patients received appropriate therapies (shock and/or ATP) with a 76.5% ATP efficacy. There have been zero fractures during follow-up along with chronically stable pacing capture threshold, pacing impedance, and R-wave amplitudes. There were four patients with an adverse event related to PWOS (0.6%), none of which was associated with inappropriate shock. There were four patients with an adverse event related to TWOS (0.6%), of which three patients were associated with inappropriate shock (0.5%). Oversensing was resolved predominantly by programming the RV sensitivity to less sensitive settings. During VF induction at implant, 97.6% (120/123) of patients showed appropriate VF episode detection at the least sensitive setting of 1.2 mV, with the remaining having detection at more sensitive settings. In follow-up, 670 VT/VF episodes were appropriately detected and treated in 94 patients with a variety of RV sensitivities and no reports of under-detected episodes. Moreover, a virtual sensitivity analysis also showed no under-detection across different RV sensitivity programming.

Conclusion

Chronic sensing performance of the OmniaSecure defibrillation lead demonstrated R-wave stability with a low rate of P-wave and T-wave oversensing, resolved predominantly by adjusting RV sensitivity. Further, VT/VF detection was successful and was not impacted when programmed to less sensitive settings. The OmniaSecure lead shows robust sensing and detection performance and programmability in ongoing follow-up.

What's new?
  • Safety and efficacy of the 4.7 Fr ICD lead—the OmniaSecure defibrillation lead.

  • No fractures through an average of 17.8 ± 6 months of follow-up in the LEADR trial.

  • Stable electrical performance.

  • Low incidence of adverse events related to oversensing.

Introduction

Implantable cardioverter-defibrillators (ICDs) are devices that deliver therapy to convert life-threatening arrhythmias, which are composed of two elements, the pulse generator and the leads. Patients with a high risk of sudden cardiac death may receive an ICD as primary prevention without a history of life-threatening arrhythmias or as secondary prevention for patients with a history of life-threatening arrhythmias.1 Despite years of technological advances, the lead remains the ‘Achilles heel’ of the ICD systems due to the nature of the implant conditions and the potential for lead failure that can result in adverse complications for the patient.2,3

The OmniaSecure lead, designed for reliability,4 is a small-diameter (4.7 Fr), lumenless, integrated bipolar defibrillation lead based on the SelectSecure™ SureScan™ MRI Model 3830 pacing lead that has shown reliable long-term performance.5–7 The Lead EvaluAtion for Defibrillation and Reliability (LEADR) trial evaluated the safety and efficacy of the OmniaSecure defibrillation lead. The trial exceeded primary safety and efficacy objective thresholds, demonstrating favourable performance and zero fractures through ∼12 months.8 The LEADR trial patients remain in ongoing follow-up. While there are known advantages of integrated bipolar leads, these leads can be associated with oversensing. The purpose of this paper is to report additional follow-up of the LEADR trial with an emphasis on the sensing and detection capabilities of the OmniaSecure lead.

Methods

Study design, population, and objectives

The LEADR trial is a prospective, multicenter, single-arm, adaptive, pivotal clinical trial (NCT04863664) with 47 participating study sites worldwide. The trial was approved per local regulatory requirements and ethics committees at each site. After providing written informed consent and in accordance with the Declaration of Helsinki as well as local ethics committees, patients were enrolled if they met ACC/AHA/ESC guideline-directed indications for de novo implantation of a primary or secondary prevention ICD or cardiac resynchronisation therapy defibrillator (CRT-D). The primary efficacy objective evaluated defibrillation efficacy at implantation in a subgroup of at least the first 95 consecutive patients completing the testing protocol. Defibrillation efficacy was tested by VF induction, where sensing was evaluated through the ICD/CRT-D for implant with the RV sense value programmed to 1.2 millivolts (mV) per protocol, with allowance to reduce sensitivity value if under-sensing was observed. The primary safety objective evaluated freedom from major complications related to the study lead at 6 months. During continued follow-up, ongoing safety, ambulatory efficacy, and electrical performance were also evaluated.

Lead and procedure description

The OmniaSecure™ SureScan™ MRI defibrillation lead (Figure 1) is a 4.7 Fr, integrated bipolar (tip-to-coil polarity), single-coil, lumenless, catheter-delivered, active-fixation transvenous lead with a right ventricular (RV) defibrillation coil electrode (6.1 cm length, 371 mm2 surface area, 12 mm tip-to-coil spacing).

Cross-sectional view of the OmniaSecure defibrillation lead, adapted from Crossley et al.8
Figure 1

Cross-sectional view of the OmniaSecure defibrillation lead, adapted from Crossley et al.8

Patients were implanted with the lead connected by DF4 connector to a commercially available pulse generator (Medtronic SureScanTM ICD or CRT-D). Lead implantation was allowed only in standard RV positions and not in the left bundle branch locations, per protocol. Care was taken to ensure that the entire defibrillation coil was in the right ventricle after lead tip fixation. Electrical testing included assessment of R-wave sensing, pacing impedance, and pacing capture threshold. Tachyarrhythmia detection and therapies as well as pacing settings were programmed per the physician’s discretion. After implantation, follow-up assessments were conducted at 3 months, 6 months, and will be conducted every 6 months thereafter until study closure. Data were collected at clinical follow-up visits and via CareLink™ remote monitoring transmissions, where available, throughout the study. Throughout follow-up, sensing performance was evaluated by the physician via standard in-office monitoring and via CareLink remote monitoring transmissions (if applicable, i.e. patient on CareLink network), including pacing capture threshold testing and review of real-time EGM, spontaneous episodes, CareAlerts, and other device-recorded data such as sensing integrity counters and pacing percentages. The presence of an integrated bipolar lead continues to underscore the importance of evaluating oversensing in the device function, which includes the auto-adjusting sensitivity feature. Therefore, physicians were trained to carefully monitor and report oversensing, independent of whether it was associated with an adverse event or whether any clinical action was taken.

Virtual sensitivity analysis

A post hoc method was developed to evaluate the safety of the programmable sensitivity settings for detection of VF.9 Virtual sensitivity analysis is a method of evaluating ICD device sensing and detection performance of previously recorded spontaneous ventricular arrhythmias. The virtual analysis utilizes data from recorded patient episodes and assesses them at various sensitivity values in the device to determine the potential for under-detection or delays to detection at non-nominal programmable settings.

EGM data were collected via ICD/CRT-D recordings throughout follow-up of the LEADR trial. These spontaneous episodes were adjudicated by an independent Episode Review Committee consisting of physicians to classify the recorded episodes for the true rhythm of each event. From the adjudicated results, episodes classified as polymorphic VF were selected for further analysis, as these types of episodes require high-voltage shock therapy to defibrillate.

Data from these episodes were recorded to the defibrillator, which captures the unfiltered EGM to be used as input for the virtual sensitivity analysis. Episodes with a minimum duration of 30 sensed R-R intervals, exceeding the minimum rate for VF detection by the ICD, were used for virtual analysis. These episodes were processed and input to the virtual simulation system (see Supplementary material online, Figure S1), where the ICD component was programmed to nominal VF detection parameters [VF detection zone: Number of Intervals for Detection (NID) at 30/40 with R-R interval ≤ 320 ms]. The detected episodes were recorded from each of the input episodes at each of the following RV sensitivity values: 0.3, 0.45, 0.6, 0.9, and 1.2 mV. The time to detect VF was measured for each tested sensitivity setting for all input episodes used in the simulation. The simulated detection time at 0.3 mV was compared with that of the other sensitivity values. From previous publications, a delay in the detection time was considered clinically significant if >2.5 s.9 VF episode under-detection was determined when the device did not reach the 30/40 NID within the input data duration.

Statistical analysis

Descriptive statistics are presented using mean ± standard deviation for continuous variables, while categorical variables are presented as percentages.

Safety was assessed through freedom from RV-lead related major complications via Kaplan-Meier method, efficacy was assessed as ambulatory therapy efficacy (shock and ATP) during all available device follow-up data using simple proportions, and reliability was assessed through lead electrical measurements and fracture-free performance of the OmniaSecure lead. The rates of appropriate and inappropriate therapy were also determined via Kaplan-Meier analysis. In addition, lead electrical performance was characterized using descriptive statistics. Statistical analyses were performed using SAS version 9.4 and R-version R-4.0.3.

Results

Overall clinical trial results

In total, 643/657 (97.9%) patients were successfully implanted with the OmniaSecure defibrillation lead in the standard RV location (26.0% female; 61.9 ± 12.9 years; 35.9% single-chamber ICD, 40.7% dual-chamber ICD, and 23.3% CRT-D), with 99.5% of leads placed in the desired location. The LEADR trial passed the primary efficacy and safety objectives, exceeding the pre-specified thresholds.8 The study reported a 97.5% defibrillation efficacy at implant and a 97.1% freedom from study lead-related major complications at 6 and 12 months [95% Confidence interval (CI): 95.4–98.1%]. There was a 96.9% (95% CI: 95.2–98.0%) freedom from study lead-related major complications at 24 months (average follow-up of 17.8 ± 6.0 months) (Figure 2). There have been 670 ambulatory VT/VF episodes in 94 patients that received appropriate therapy (shock and/or ATP) through device follow-up of 19.3 ± 5.8 months. Termination of ambulatory events with shock was successful in 94% (124/132 episodes), with the remainder ATP terminated (6), self-terminated (1), or associated with other patient factors (1) (see Supplementary material online, Appendix). ATP efficacy was 76.5% (485/634 episodes) in 76 patients. A total of 10.7% (95% CI: 8.5–13.4%) of patients received an appropriate therapy by 12 months and 17.6% (95% CI: 14.2–21.7%) by 24 months (Figure 3). The inappropriate shock rate was 2.7% at 12 months (95% CI: 1.7–4.3%) and 3.8% at 24 months (95% CI: 2.4–5.9%), see Supplementary material online, Appendix for further details. There were zero fractures of the OmniaSecure lead through 17.8 ± 6.0 month follow-up.

Kaplan-Meier estimated freedom from study lead-related major complications through 2 years post-implant.
Figure 2

Kaplan-Meier estimated freedom from study lead-related major complications through 2 years post-implant.

Kaplan-Meier estimated incidence of first appropriate shock, ATP, or any therapy through 2 years. Values in Figure provide rates at 12 and 24 months.
Figure 3

Kaplan-Meier estimated incidence of first appropriate shock, ATP, or any therapy through 2 years. Values in Figure provide rates at 12 and 24 months.

Electrical performance

During follow-up, the functional electrical performance of the OmniaSecure defibrillation lead was evaluated (Figure 4). The mean pacing impedance was 635.1 ± 175.0 Ohms at implant, 394.4 ± 64.5 Ohms at 3-month follow-up, and remained stable thereafter. The mean pacing capture threshold was 0.60 ± 0.26 V at implant and remained stable throughout follow-up.

Electrical measurements for pacing impedance, pacing capture threshold, and R-wave amplitude through 24 months. Mean ± standard deviation is presented for each visit. The values in the figure provide the mean while the errors bars represent the standard deviation.
Figure 4

Electrical measurements for pacing impedance, pacing capture threshold, and R-wave amplitude through 24 months. Mean ± standard deviation is presented for each visit. The values in the figure provide the mean while the errors bars represent the standard deviation.

The OmniaSecure lead sensing was evaluated by R-wave amplitude monitoring at implant and through follow-up. The recommended implant acceptance criteria for R-wave amplitude was ≥5 mV. The R-wave amplitude at implant was on average 10.2 ± 4.8 mV (25–75th percentile: 6–13 mV) and remained stable throughout follow-up.

Physiologic oversensing

While sensing was adequate and remained stable, oversensing was occasionally observed in some patients. Physicians were trained to carefully monitor and report oversensing, independent of whether it was associated with an adverse event or if any subsequent action was taken. Overall, there were 74 patients with reports of physiologic oversensing, irrespective of time since implant or clinical implications (PWOS N = 19, TWOS N = 38, TWOS and PWOS N = 1, R-wave double counting N = 4, and unspecified N = 14; see Supplementary material online, Appendix). The vast majority of patients, 87.8% (65/74), did not have an adverse event associated with oversensing. The breakdown by device type is shown in Table 1.

Table 1

Breakdown of device type for the overall LEADR trial and for those patients with reports of oversensing

 Single chamber ICDDual chamber ICDCRT-D
All LEADR patients (N = 643)35.9% (231)40.7% (262)23.3% (150)
PWOS (N = 19)a26.3% (5)57.9% (11)15.8% (3)
TWOS (N = 38)a26.3% (10)23.7% (9)50% (19)
PWOS and TWOS (N = 1)N/AN/A100% (1)
R-wave double counting (N = 4)75% (3)N/A25% (1)
Unspecified (N = 14)a50% (7)43% (6)7% (1)
 Single chamber ICDDual chamber ICDCRT-D
All LEADR patients (N = 643)35.9% (231)40.7% (262)23.3% (150)
PWOS (N = 19)a26.3% (5)57.9% (11)15.8% (3)
TWOS (N = 38)a26.3% (10)23.7% (9)50% (19)
PWOS and TWOS (N = 1)N/AN/A100% (1)
R-wave double counting (N = 4)75% (3)N/A25% (1)
Unspecified (N = 14)a50% (7)43% (6)7% (1)

aThere were 74 unique patients with physiologic oversensing, one patient had PWOS in addition to unspecified oversensing and another patient had TWOS in addition to unspecified oversensing.

Table 1

Breakdown of device type for the overall LEADR trial and for those patients with reports of oversensing

 Single chamber ICDDual chamber ICDCRT-D
All LEADR patients (N = 643)35.9% (231)40.7% (262)23.3% (150)
PWOS (N = 19)a26.3% (5)57.9% (11)15.8% (3)
TWOS (N = 38)a26.3% (10)23.7% (9)50% (19)
PWOS and TWOS (N = 1)N/AN/A100% (1)
R-wave double counting (N = 4)75% (3)N/A25% (1)
Unspecified (N = 14)a50% (7)43% (6)7% (1)
 Single chamber ICDDual chamber ICDCRT-D
All LEADR patients (N = 643)35.9% (231)40.7% (262)23.3% (150)
PWOS (N = 19)a26.3% (5)57.9% (11)15.8% (3)
TWOS (N = 38)a26.3% (10)23.7% (9)50% (19)
PWOS and TWOS (N = 1)N/AN/A100% (1)
R-wave double counting (N = 4)75% (3)N/A25% (1)
Unspecified (N = 14)a50% (7)43% (6)7% (1)

aThere were 74 unique patients with physiologic oversensing, one patient had PWOS in addition to unspecified oversensing and another patient had TWOS in addition to unspecified oversensing.

Nineteen patients had reports of PWOS, of which eight patients had reports at implant and 11 patients at follow-up. Importantly, there were no inappropriate shocks due to PWOS. The average R-wave amplitude for the 19 patients with PWOS was stable through follow-up (Figure 4). PWOS could be mitigated via programming adjustments in 17/19 patients (ICD: N = 14 and CRT-D: N = 3). Of the remaining two, one patient had the lead repositioned at implant, and the other had the lead replaced in chronic follow-up (more information below).

Four (0.6%) of the nineteen patients had a PWOS-associated adverse event. Of the four patients with a PWOS adverse event, one patient had PWOS resolved by programming the RV sensitivity to 1.2 mV with sensing confirmed at this sensitivity value via VF induction in hospital (34 days post-implant), another patient had a system revision to replace the lead with a non-study lead (119 days post-implant), and the remaining two patients showed PWOS resolution by programing adjustments, however underwent an elective replacement of the lead with a non-study lead at physician discretion (at implant and 1 day post-implant). (See Supplementary material online, Appendix for follow-up details pertaining to the remaining patients that did not have an associated adverse event).

Among patients with PWOS that did not undergo a lead replacement, the final RV sensitivity at most recent follow-up was 0.3 (2), 0.45 (5), 0.6 (4), 0.9 (3), or 1.2 mV (2). Of the two patients programmed at 1.2 mV RV sensitivity, induction testing at this setting was completed for one (due to PWOS adverse event).

Thirty-eight patients had reports of TWOS, of which six patients had reports at implant and 32 patients at follow-up. Among these, there were only three patients with inappropriate shock due to TWOS, 0.5%, subsequently described. The average R-wave amplitude for the 38 patients with TWOS was stable through follow-up (Figure 4). There were 21 patients with TWOS characterized as post-pace TWOS, 16 patients with post-sense TWOS, and one patient with both post-pace and post-sense, see Table 2 for breakdown by device type. TWOS was resolved via programming adjustments in 34 of 38 patients (ICD: N = 16 and CRT-D: N = 18). Of the remaining, one resulted in an unsuccessful implant, one was resolved via repositioning at implant, one had no clinical action taken (387 days post-implant), and one patient had the lead replaced (described below).

Table 2

Distribution of post-pace and post-sense TWOS by device type

 Single chamber ICD (N = 10)Dual chamber ICD (N = 9)CRT-D (N = 19)
PP-TWOS (N = 21)9.5% (2)19% (4)71.5% (15)
PS-TWOS (N = 16)50% (8)25% (4)25% (4)
PP and PS-TWOS (N = 1)N/A100% (1)N/A
 Single chamber ICD (N = 10)Dual chamber ICD (N = 9)CRT-D (N = 19)
PP-TWOS (N = 21)9.5% (2)19% (4)71.5% (15)
PS-TWOS (N = 16)50% (8)25% (4)25% (4)
PP and PS-TWOS (N = 1)N/A100% (1)N/A
Table 2

Distribution of post-pace and post-sense TWOS by device type

 Single chamber ICD (N = 10)Dual chamber ICD (N = 9)CRT-D (N = 19)
PP-TWOS (N = 21)9.5% (2)19% (4)71.5% (15)
PS-TWOS (N = 16)50% (8)25% (4)25% (4)
PP and PS-TWOS (N = 1)N/A100% (1)N/A
 Single chamber ICD (N = 10)Dual chamber ICD (N = 9)CRT-D (N = 19)
PP-TWOS (N = 21)9.5% (2)19% (4)71.5% (15)
PS-TWOS (N = 16)50% (8)25% (4)25% (4)
PP and PS-TWOS (N = 1)N/A100% (1)N/A

Four (0.6%) of the 38 patients had a TWOS-associated adverse event. Of these four patients, one patient had a system revision to replace the lead with a non-study lead due to post-pace TWOS that could not be resolved with programing adjustments (359 days); however, TWOS continued after replacement with a non-study lead, requiring further revision of lead placement (see Supplementary material online, Appendix). Three patients had an inappropriate shock due to post-sense TWOS (118, 374, and 630 days) that were mitigated by programming adjustments, after which no further events have been reported. (See Supplementary material online, Appendix for follow-up details pertaining to the remaining patients that did not have an associated adverse event).

Among patients with TWOS that did not undergo a lead replacement, the final RV sensitivity value at most recent follow-up was 0.3 (1), 0.45 (22), 0.6 (12), and 1.2 mV (1). In the patient programmed to 1.2 mV, induction testing was not conducted; however, there was a spontaneous episode successfully detected and treated while programmed at 1.2 mV.

Tachyarrhythmia detection performance

During induced VF as part of defibrillation testing at implant, 97.6% (120/123) of patients showed appropriate detection at the least sensitive setting (1.2 mV) with the remaining three successful at more sensitive settings. Among these three, detection was confirmed at 0.9 mV for one patient while the other two patients were adjusted to 0.3 mV without further testing at other sensitivity values, per physician discretion (see Supplementary material online, Appendix for further information). There were 670 ambulatory VT/VF episodes appropriately treated in 94 patients, all of which were successfully detected across a variety of programmed sensitivities (Table 3) with no reports of under-detection of spontaneous VT/VF episodes.

Table 3

Proportion of successfully treated ambulatory events by programmed RV sensitivity

Available Programing (mV)Programming at time of successfully treated ambulatory event (N = 670 episodes in 94 patients)
0.3 [nom]69.3%
0.4521.9%
0.68.7%
0.90.0%
1.20.2%
Available Programing (mV)Programming at time of successfully treated ambulatory event (N = 670 episodes in 94 patients)
0.3 [nom]69.3%
0.4521.9%
0.68.7%
0.90.0%
1.20.2%
Table 3

Proportion of successfully treated ambulatory events by programmed RV sensitivity

Available Programing (mV)Programming at time of successfully treated ambulatory event (N = 670 episodes in 94 patients)
0.3 [nom]69.3%
0.4521.9%
0.68.7%
0.90.0%
1.20.2%
Available Programing (mV)Programming at time of successfully treated ambulatory event (N = 670 episodes in 94 patients)
0.3 [nom]69.3%
0.4521.9%
0.68.7%
0.90.0%
1.20.2%

Virtual sensitivity analysis results

There were 33 episodes from 20 patients adjudicated to have true polymorphic VT/VF that could be processed and input to the simulation test environment for the virtual sensitivity analysis. The simulated time to VF detection increased by ≤0.7 s on average from the nominal (0.3 mV) to the highest sensitivity value (1.2 mV) (Table 4). The difference in simulated time to detection between the nominal and the next sensitivity level (0.45 mV) was on average 0.11 s and at the maximum 0.58 s. There were no cases of complete under-detection at any RV sensitivity level.

Table 4

Virtual sensitivity analysis results

RV sensitivityProportion of clinical programming level (N = 33 episodes)Simulated time to VF detection (s) [mean ± SD]Range of simulated time to VF detection (s) [Min–Max]Difference in simulated detection of clinical VF time from simulated detection of clinical VF time at 0.3 mV (s) [mean ± SD]Proportion of episodes with difference in simulated detection of clinical VF time at 0.3 mV > 2.5s
0.328/337.06 ± 1.664.97–11.60N/AN/A
0.455/337.17 ± 1.664.96–11.590.11 ± 0.190/33
0.60/337.22 ± 1.585.07–10.780.16 ± 0.340/33
0.90/337.44 ± 1.765.07–11.080.38 ± 0.550/33
1.20/337.76 ± 2.275.06–13.250.70 ± 1.702/33
RV sensitivityProportion of clinical programming level (N = 33 episodes)Simulated time to VF detection (s) [mean ± SD]Range of simulated time to VF detection (s) [Min–Max]Difference in simulated detection of clinical VF time from simulated detection of clinical VF time at 0.3 mV (s) [mean ± SD]Proportion of episodes with difference in simulated detection of clinical VF time at 0.3 mV > 2.5s
0.328/337.06 ± 1.664.97–11.60N/AN/A
0.455/337.17 ± 1.664.96–11.590.11 ± 0.190/33
0.60/337.22 ± 1.585.07–10.780.16 ± 0.340/33
0.90/337.44 ± 1.765.07–11.080.38 ± 0.550/33
1.20/337.76 ± 2.275.06–13.250.70 ± 1.702/33
Table 4

Virtual sensitivity analysis results

RV sensitivityProportion of clinical programming level (N = 33 episodes)Simulated time to VF detection (s) [mean ± SD]Range of simulated time to VF detection (s) [Min–Max]Difference in simulated detection of clinical VF time from simulated detection of clinical VF time at 0.3 mV (s) [mean ± SD]Proportion of episodes with difference in simulated detection of clinical VF time at 0.3 mV > 2.5s
0.328/337.06 ± 1.664.97–11.60N/AN/A
0.455/337.17 ± 1.664.96–11.590.11 ± 0.190/33
0.60/337.22 ± 1.585.07–10.780.16 ± 0.340/33
0.90/337.44 ± 1.765.07–11.080.38 ± 0.550/33
1.20/337.76 ± 2.275.06–13.250.70 ± 1.702/33
RV sensitivityProportion of clinical programming level (N = 33 episodes)Simulated time to VF detection (s) [mean ± SD]Range of simulated time to VF detection (s) [Min–Max]Difference in simulated detection of clinical VF time from simulated detection of clinical VF time at 0.3 mV (s) [mean ± SD]Proportion of episodes with difference in simulated detection of clinical VF time at 0.3 mV > 2.5s
0.328/337.06 ± 1.664.97–11.60N/AN/A
0.455/337.17 ± 1.664.96–11.590.11 ± 0.190/33
0.60/337.22 ± 1.585.07–10.780.16 ± 0.340/33
0.90/337.44 ± 1.765.07–11.080.38 ± 0.550/33
1.20/337.76 ± 2.275.06–13.250.70 ± 1.702/33

Discussion

In ICD systems, the transvenous lead has remained the ‘Achilles heel’ of the system due to the frequency of lead-related complications.2,3 The novel, lumenless, integrated bipolar, small-diameter OmniaSecure defibrillation lead may address this need due to the potential of small-diameter leads having reduced lead-related complications.10–12 Moreover, the OmniaSecure lead is based on the SelectSecure 3830 pacing lead that has demonstrated reliable long-term performance.5–7 The LEADR trial was conducted to evaluate the safety, efficacy, and reliability of the OmniaSecure lead implanted in the standard RV location.4 The primary efficacy and safety objectives were met, and results exceeded the pre-specified performance thresholds.8 The LEADR trial continues to demonstrate ongoing safety, efficacy, and reliability of the novel OmniaSecure lead through longer-term follow-up. Chronic sensing performance of the OmniaSecure defibrillation lead demonstrated R-wave stability with a low rate of P-wave and T-wave oversensing, predominantly resolved by reprogramming RV sensitivity. Furthermore, VT/VF detection was successful and was not impacted when reprogrammed to less sensitive settings. The OmniaSecure lead shows robust sensing and detection performance and programmability in ongoing follow-up.

The OmniaSecure lead has a simplified lumenless, single-coil design that enables a small diameter of 4.7 Fr. This simplified lumenless design replaces the central stylet lumen with a fracture-resistant flexible cable conductor that improves durability.12 In addition, the OmniaSecure lead incorporates a helix design similar to the SelectSecure 3830 pacing lead to preserve capture threshold performance. There have been no fractures of the OmniaSecure lead through available clinical follow-up, and reliability modelling has projected the 10-year fracture-free survival rate of the OmniaSecure lead to be 98.2%.12 The LEADR trial demonstrated average R-wave amplitudes of >10 mV; R-wave amplitude along with pacing capture threshold and pacing impedance remained stable throughout follow-up.

The OmniaSecure lead is an integrated bipolar lead.4,8 Literature notes many advantages of an integrated bipolar lead design, including the ability to downsize the lead and thereby reduce patient complications,4 increased reliability due to a simplified design,13 and increased R-wave amplitudes due to wider electrode spacing.14–16

However, integrated bipolar leads can be prone to oversensing.17–21 Physiologic oversensing, namely PWOS and TWOS, while rare, may result in inappropriate therapy and/or inhibition of pacing. In most cases, oversensing can be mitigated by adjusting the sensitivity of the ventricular lead with considerations for tachyarrhythmia detection.22

In the LEADR trial, PWOS was observed, however, the overall rate of PWOS associated with an adverse event was low at 0.6% (4/657), and there were no inappropriate shocks related to PWOS. TWOS was also reported with a low overall rate of TWOS being associated with an adverse event of 0.6% (4/657), including three patients that had inappropriate shock due to TWOS. This TWOS-related rate of patients with inappropriate shock is in line with the PainFree SST study that included predominantly true bipolar ventricular sensing leads as well as the transvenous arm of the PRAETORIAN trial.23,24

Programming adjustments are an effective tool for mitigating oversensing. In the LEADR trial, oversensing was resolved via programming, mitigating PWOS in 89.5% (17/19) of patients and TWOS in 89.5% (34/38) of patients where reprogramming was attempted. The RV sensitivity value of 0.45 mV was the most common programming adjustment that resolved the oversensing events. When adjusting RV sensitivity, there should also be consideration for tachyarrhythmia detection. Through follow-up, all of the 670 ambulatory VT/VF episodes appropriately treated were successfully detected across a variety of programmed ventricular sensitivities with no reports of spontaneous ventricular arrythmia under-detection.

Virtual sensitivity simulation is an effective tool to evaluate the sensing and detection performance at various RV sensitivity values for a given episode with recorded EGM data from the study. There was no under-detection in the simulations regardless of RV sensitivity value programming. When sensitivity was changed from 0.3 to 0.45 mV, the most common adjustment observed in the trial, the time to detection minimally increased, by an average of 110 ms, which is a clinically insignificant delay.9 Even when the sensitivity was increased to 0.6 mV, these favourable observations were maintained. These data support the option to program to a progressively less sensitive setting to optimize sensing, with minimal impact on VF detection. Therefore, clinicians may consider programming the RV sensitivity to 0.45 mV, which may be an effective means of proactively mitigating potential oversensing. Overall, this detailed analysis of sensing and detection performance demonstrates a low rate of sensing-related adverse events and the flexibility to resolve sensing observations non-invasively through programming.

Limitations

The LEADR trial remains in ongoing follow-up and not all patients have reached the 24-month follow-up. Therefore, 24-month results may vary slightly as follow-up continues. Average follow-up time was 17.8 ± 6.0 months, including 153 patients with 24-month follow-up and 24 patients with 30-month follow-up. The OmniaSecure lead is a DF4 only configuration, and comparisons to leads with other connector types were not evaluated.

Observed oversensing in this study was as reported by the study investigators for whom all study data from follow-up and CareLink transmissions (if applicable) were made available. However, there may be variation among investigators for the criteria of reportable oversensing.

The virtual sensitivity simulations were restricted to recorded episodes adjudicated as polymorphic VT/VF; this limited the sample size to 33 episodes from 20 patients in the trial. Therefore, these results may not be generalizable to all arrhythmias or to all patients in the study.

Sensing was evaluated with the OmniaSecure lead in combination with a Medtronic ICD/CRT-D; therefore, extrapolating results individually to the lead or the device may not be possible since both were studied as part of a whole system.

Conclusion

The LEADR trial demonstrates safety, efficacy, and reliability of the novel, small-diameter OmniaSecure lead, with low complication rates, high ambulatory efficacy, and reliable performance. Chronic sensing performance of the OmniaSecure defibrillation lead demonstrated R-wave stability with a low rate of oversensing, predominantly resolved by adjusting the RV sensitivity programming, with no clinically relevant impact. The OmniaSecure lead shows robust sensing and detection performance and programmability.

Supplementary material

Supplementary material is available at Europace online.

Acknowledgements

We would like to thank the patients for their participation in this trial as well as the participating sites. We would also like to thank Cathy Condie, Mackenzie Hall, Robert Musto (employees of Medtronic, Inc.) for their contributions to the virtual sensitivity analysis, and Roberto Carta (employees of Medtronic, Inc.) for statistical analyses.

Funding

This work was supported by Medtronic, Inc.

Data availability

All data requests will be made through the corresponding author. Data requests will require details of the analysis and usage to be performed and will need to be approved by the study steering committee. If approved and within the confines of the Human Research Ethics approval, data will be made available.

References

1

Ursaru
 
AM
,
Petris
 
AO
,
Costache
 
II
,
Nicolae
 
A
,
Crisan
 
A
,
Tesloianu
 
ND
.
Implantable cardioverter defibrillator in primary and secondary prevention of SCD-what we still don't know
.
J Cardiovasc Dev Dis
 
2022
;
9
:
120
.

2

Swerdlow
 
CD
,
Kalahasty
 
G
,
Ellenbogen
 
KA
.
Implantable cardiac defibrillator lead failure and management
.
J Am Coll Cardiol
 
2016
;
67
:
1358
68
.

3

Defaye
 
P
,
Biffi
 
M
,
El-Chami
 
M
,
Boveda
 
S
,
Glikson
 
M
,
Piccini
 
J
 et al.  
Cardiac pacing and lead devices management: 25 years of research at EP Europace journal
.
Europace
 
2023
;
25
:
euad202
.

4

Crossley
 
GH
,
Sanders
 
P
,
De Filippo
 
P
,
Tarakji
 
KG
,
Hansky
 
B
,
Shah
 
M
 et al.  
Rationale and design of the lead evaluation for defibrillation and reliability study: safety and efficacy of a novel ICD lead design
.
J Cardiovasc Electrophysiol
 
2023
;
34
:
257
67
.

5

De Filippo
 
P
,
Giofre
 
F
,
Leidi
 
C
,
Senni
 
M
,
Ferrari
 
P
.
Transvenous pacing in pediatric patients with bipolar lumenless lead: ten-year clinical experience
.
Int J Cardiol
 
2018
;
255
:
45
9
.

6

Bansal
 
N
,
Samuel
 
S
,
Zelin
 
K
,
Karpawich
 
PP
.
Ten-year clinical experience with the lumenless, catheter-delivered, 4.1-Fr diameter pacing lead in patients with and without congenital heart
.
Pacing Clin Electrophysiol
 
2017
;
40
:
17
25
.

7

Zou
 
J
,
Chen
 
K
,
Liu
 
X
,
Xu
 
Y
,
Jiang
 
L
,
Dai
 
Y
 et al.  
Clinical use conditions of lead deployment and simulated lead fracture rate in left bundle branch area pacing
.
J Cardiovasc Electrophysiol
 
2023
;
34
:
718
25
.

8

Crossley
 
GH
,
Sanders
 
P
,
Hansky
 
B
,
De Filippo
 
P
,
Shah
 
MJ
,
Morio
 
S
 et al.  
Safety, efficacy, and reliability evaluation of a novel small-diameter defibrillation lead: global LEADR pivotal trial results
.
Heart Rhythm
 
2024
;
21
:
1914
22
.

9

Swerdlow
 
CD
,
Zhang
 
X
,
Sawchuk
 
R
,
Greenhut
 
S
,
Grinberg
 
Y
,
Liu
 
Y
 et al.  
Design and preliminary results of sensing and detection for an extravascular implantable cardioverter-defibrillator
.
JACC Clin Electrophysiol
 
2021
;
7
:
1387
99
.

10

Bharmanee
 
A
,
Zelin
 
K
,
Sanil
 
Y
,
Gupta
 
P
,
Karpawich
 
PP
.
Comparative chronic valve and venous effects of lumenless versus stylet-delivered pacing leads in patients with and without congenital heart
.
Pacing Clin Electrophysiol
 
2015
;
38
:
1343
50
.

11

Tatum
 
R
,
Maynes
 
EJ
,
Wood
 
CT
,
Deb
 
AK
,
Austin
 
MA
,
O'Malley
 
TJ
 et al.  
Tricuspid regurgitation associated with implantable electrical device insertion: a systematic review and meta-analysis
.
Pacing Clin Electrophysiol
 
2021
;
44
:
1297
302
.

12

Crossley
 
GH
,
Mason
 
PK
,
Hansky
 
B
,
De Filippo
 
P
,
Shah
 
MJ
,
Philippon
 
F
 et al.  
High predicted durability for the novel small-diameter OmniaSecure defibrillation lead
.
Heart Rhythm
 
2025
;
22
:
302
10
.

13

Richardson
 
TD
,
Himes
 
A
,
Marshall
 
M
,
Crossley
 
GH
.
Rationale for and use of the lumenless 3830 pacing lead
.
J Cardiovasc Electrophysiol
 
2023
;
34
:
769
74
.

14

Krahn
 
AD
,
Pickett
 
RA
,
Sakaguchi
 
S
,
Shaik
 
N
,
Cao
 
J
,
Norman
 
HS
 et al.  
R-wave sensing in an implantable cardiac monitor without ECG-based preimplant mapping: results from a multicenter clinical trial
.
Pacing Clin Electrophysiol
 
2014
;
37
:
505
11
.

15

Swerdlow
 
CD
,
Brown
 
ML
,
Bordachar
 
P.
 4—sensing And detection with cardiac implantable electronic devices. In:
Ellenbogen
 
KA
,
Wilkoff
 
BL
,
Kay
 
GN
,
Lau
 
C-P
,
Auricchio
 
A
, eds.
Clinical Cardiac Pacing, Defibrillation and Resynchronization Therapy
(Fifth Edition). Amsterdam, Netherlands:
Elsevier
;
2017
. p
114
67
.

16

Feldman
 
T
,
Borow
 
KM
,
Neumann
 
A
,
Lang
 
RM
,
Childers
 
RW
.
Relation of electrocardiographic R-wave amplitude to changes in left ventricular chamber size and position in normal subjects
.
Am J Cardiol
 
1985
;
55
:
1168
74
.

17

Borleffs
 
CJ
,
van Erven
 
L
,
van Bommel
 
RJ
,
van der Velde
 
ET
,
van der Wall
 
EE
,
Bax
 
JJ
 et al.  
Risk of failure of transvenous implantable cardioverter-defibrillator leads
.
Circ Arrhythm Electrophysiol
 
2009
;
2
:
411
6
.

18

Dorwarth
 
U
,
Frey
 
B
,
Dugas
 
M
,
Matis
 
T
,
Fiek
 
M
,
Schmoeckel
 
M
 et al.  
Transvenous defibrillation leads: high incidence of failure during long-term follow-up
.
J Cardiovasc Electrophysiol
 
2003
;
14
:
38
43
.

19

Eckstein
 
J
,
Koller
 
MT
,
Zabel
 
M
,
Kalusche
 
D
,
Schaer
 
BA
,
Osswald
 
S
 et al.  
Necessity for surgical revision of defibrillator leads implanted long-term: causes and management
.
Circulation
 
2008
;
117
:
2727
33
.

20

Duray
 
GZ
,
Schmitt
 
J
,
Cicek-Hartvig
 
S
,
Hohnloser
 
SH
,
Israel
 
CW
.
Complications leading to surgical revision in implantable cardioverter defibrillator patients: comparison of patients with single-chamber, dual-chamber, and biventricular devices
.
Europace
 
2009
;
11
:
297
302
.

21

Ellenbogen
 
KA
,
Wood
 
MA
,
Shepard
 
RK
,
Clemo
 
HF
,
Vaughn
 
T
,
Holloman
 
K
 et al.  
Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications
.
J Am Coll Cardiol
 
2003
;
41
:
73
80
.

22

Swerdlow
 
CD
,
Asirvatham
 
SJ
,
Ellenbogen
 
KA
,
Friedman
 
PA
.
Troubleshooting implanted cardioverter defibrillator sensing problems I
.
Circ Arrhythm Electrophysiol
 
2014
;
7
:
1237
61
.

23

Auricchio
 
A
,
Schloss
 
EJ
,
Kurita
 
T
,
Meijer
 
A
,
Gerritse
 
B
,
Zweibel
 
S
 et al.  
Low inappropriate shock rates in patients with single- and dual/triple-chamber implantable cardioverter-defibrillators using a novel suite of detection algorithms: painFree SST trial primary results
.
Heart Rhythm
 
2015
;
12
:
926
36
.

24

Knops
 
RE
,
Nordkamp
 
LRAO
,
Delnoy
 
PHM
,
Boersma
 
LVA
,
Kuschyk
 
J
,
El-Chami
 
MF
 et al.  
Subcutaneous or transvenous defibrillator therapy
.
N Engl J Med
 
2020
;
383
:
526
36
.

Author notes

Conflict of interest: P.S.: reports having served on the advisory boards of Medtronic, Abbott, Boston Scientific, PaceMate, and CathRx; the University of Adelaide has received on his behalf lecture and/or consulting fees from Medtronic, Abbott, and Boston Scientific; and the University of Adelaide has received on his behalf research funding from Medtronic, Abbott, Boston Scientific, and Becton-Dickson. P.K.M.: consultant for Medtronic, Boston Scientific, and Cook as well as honoraria from Medtronic and Cook. B.H.: reports funding from Medtronic and CVRx. P.D.F.: consultant fees and travel support from Medtronic, Boston Scientific, Abbott, and Biotronik as well as honoraria and advisory board participation for Medtronic and Abbott. M.J.S.: consultant for Medtronic and Tenaya Therapeutics. D.P.S.: institutional funding from Medtronic. J.S.Z.: consultant for Medtronic. F.P.: speaker at Medtronic and Boston Scientific; consultant for Medtronic and Boston Scientific; research funding from Medtronic and Boston Scientific. B.T.: honorarium from Boston scientific. R.K.P.: reports that Canberra Heart Rhythm Foundation has received on his behalf lecture and/or consulting fees from Medtronic, Abbott Medical, Boston Scientific, and Biotronik. T.D.R.: consultant for Medtronic, Johnson and Johnson, and Philips; research funding from Medtronic and Abbott. M.F.: consultant for Medtronic. R.D.S: consultant for Medtronic as well as speaking honoraria. I.A.: consultant for Medtronic. A.M.: consultant fees and travel support from Medtronic, Boston Scientific, Abbott, Biotronik and Microport, advisory board participation for Medtronic and Boston. B.B.: consultant for Medtronic. A.E.T., K.A., B.M., and C.B.: employees of Medtronic, Inc. G.H.C.: speaker at Medtronic and Philips; consultant for Medtronic and Boston Scientific.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.

Supplementary data