Abstract

Aims

Arrhythmic burden after discharge in patients with new-onset persistent left bundle branch block (NOP-LBBB) following transcatheter aortic valve replacement (TAVR) with Evolut devices remains largely unknown. The aim of this study is to assess the incidence and type of arrhythmias at 2-year follow-up in patients with NOP-LBBB post-TAVR.

Methods and results

This is a prospective multicentre study including 88 patients with LBBB persisting for ≥3 days post-implantation. Before discharge, an implantable loop recorder (REVEAL XT/LINQ) was implanted; patients had continuous monitoring for 2 years. Arrhythmic events were adjudicated in a central core lab. Of the arrhythmic events, 411 were detected in 58 patients [65.9%; 2 (1–4) events per patient]. Symptoms were reported in 12/58 (20.7%), and therapy was changed in 25/58 (43.1%). There were 101 bradyarrhythmic events in 33 patients [35 high-grade atrioventricular block (HAVB) and 66 severe bradycardia]. The HAVB incidence was higher in the early (4-week) phase and remained stable over time, whereas severe bradycardia increased after 1 year. Permanent pacemaker was required in 11 (12.5%) patients (6.8% and 5.7% in the first and second year, respectively). There were 310 tachyarrhythmic events in 29 patients (120 AF/AFL, 111 AT, 72 SVT, 6 NSVT, and 1 VT); its incidence decreased throughout the 2 years. New AF/AFL episodes occurred in 20/69 patients [29%; symptomatic in 2/20 (10%)].

Conclusion

Patients with NOP-LBBB post-TAVR with Evolut devices exhibited a high burden of late arrhythmias, with events occurring in two-thirds of patients and leading to treatment changes in about half of them. These data should inform future studies on cardiac monitoring devices for follow-up and treatment optimization in this challenging population.

What’s new?
  • Late-appearing arrhythmic events are particularly common in the long term in patients showing new-onset persistent LBBB after TAVR with the Evolut system.

  • Treatment changes due to arrhythmia are still needed at long-term follow-up in about half of the patients.

  • Late need for PPM occurs frequently at long-term follow-up after TAVR, with sinus node disease becoming a more important cause than typical immediate high-degree atrioventricular block.

  • New AF/AFL episodes are frequent and usually require the onset of an anticoagulation regimen.

  • Ventricular arrhythmias are rare in the long-term post-TAVR.

Introduction

The occurrence of conduction disturbances, particularly left bundle branch block, remains the most common complication of transcatheter aortic valve replacement (TAVR),1 and it has been associated with an increased risk of mortality and heart failure hospitalization at follow-up.2,3 Thus, timely identification and appropriate management are critical in this high-risk group of patients. The use of some self-expanding valves such as the CoreValve system (Medtronic, Minneapolis, MN) has been classically associated with an increased risk of conduction disturbances compared to balloon-expandable valves.4,5 Major changes in the stent frame have been implemented to the newer generation CoreValve Evolut system, and controversial data have been reported regarding the changes in the incidence of conduction disturbances with the use of this new valve platform.6–10 However, in those patients with new-onset persistent left bundle branch block (NOP-LBBB), data on the occurrence of arrhythmias (including life-threatening events) detected by continuous electrocardiogram (ECG) monitoring beyond the in-hospital period have been scarce and limited to a relatively short (1 month) follow-up period.11,12

The use of implantable cardiac monitoring devices has been shown to be useful for detecting cardiac arrhythmias in different clinical scenarios,13 but data in the TAVR field have been restricted to studies including a mix of older and newer valve generations and different valve types.14 Thus, the aim of this study was to determine both the incidence and type of arrhythmias at 2-year follow-up with the use of continuous ECG monitoring in patients with NOP-LBBB following TAVR with the Evolut valve system.

Methods

This was a prospective multicentre study including consecutive patients who underwent TAVR with the Evolut valve system between September 2014 and January 2022 in 11 centres in Canada, the USA, and Europe and had NOP-LBBB that persisted ≥3 days following the procedure. Patients receiving other types of valve prostheses were excluded from the analysis. An implantable loop recorder (ILR), either a Reveal ICM XT or a LINQ (Medtronic) device, was implanted before discharge in all patients. Left bundle branch block (LBBB) was defined according to the American College of Cardiology/American Heart Association/Heart Rhythm Society (ACC/AHA/HRS) recommendations.15 A 2-year follow-up was performed in all patients with outpatient visits and 12-lead ECG being recorded at 1, 12, and 24 months. Data from patients who received the Reveal LINQ device were obtained through automatic wireless transmission, whereas data from patients with a Reveal XT device were generated by device interrogation every 3 months. To address the scheduling constraint of the Reveal XT device, the systematic recording of atrial fibrillation events was deactivated by the centres for patients known to have chronic atrial fibrillation in order to avoid missing the recording of other new events. No adjustment was required for the Reveal LINQ device, since events are transmitted on a daily basis. Valve Academic Research Consortium 2 (VARC-2) criteria were used to define clinical events.16

The primary endpoint of the study was the incidence of arrhythmic events leading to a treatment change at 2-year follow-up. Secondary endpoints included significant arrhythmias defined according to the ACC/AHA/HRS guidelines.17 These were classified as (i) significant bradyarrhythmia [high-degree atrioventricular block (HAVB), severe bradycardia (heart rate < 30 bpm in four consecutive beats or pause > 3 s)]; (ii) atrial fibrillation (AF)/atrial flutter (AFL)/atrial tachycardia (AT)/supraventricular tachycardia (SVT) episodes lasting >30 s; (iii) ventricular tachycardia (VT) (non-sustained: lasting between 6 and 30 s; sustained: lasting >30 s); and (iv) ventricular fibrillation (VF). More specifically, high-degree AVB was defined as the presence of ≥2 consecutive, non-conducted P waves (high-grade second degree) and as unrelated P waves to the QRS complexes (third degree). Arrhythmic burden adjudication was performed in a central core lab. Indications for PPM implantation were also based on ACC/AHA/HRS recommendations. Class I, IIa, and IIb recommendations were followed by managing physicians at each participating centre for sinus node dysfunction, acquired atrioventricular block, and chronic bifascicular block.

The initial diagnosis and management of arrhythmic events were the responsibility of the investigators of each participating centre, based on the mentioned guidelines and on current clinical practice.

The study was approved by the institutional ethics committee of participating centres, and all patients provided signed informed consent. The results including the arrhythmic burden at 1- and 2-year follow-up of 35 patients included in the present study had been previously reported.18

Statistical analysis

Numbers and frequencies were used to report categorical variables. Continuous variables were presented as mean ± standard deviation when normally distributed and as median (interquartile range) if non-normally distributed. χ² tests (or Fisher’s tests when appropriate) were used to compare categorical variables. Numerical variables were compared using the Student’s t-test or Wilcoxon test as appropriate. Kaplan–Meier curves were plotted for event rates over time. Statistical significance was set at P < 0.05. All analyses were performed with the SPSS statistical software package version 29.0.

Results

A total of 88 Evolut recipients (Evolut R: 87, Evolut PRO: 1) with NOP-LBBB after TAVR were included. The main baseline characteristics and procedural data are shown in Table 1. Median (IQR) age of the study population was 83 (80–86) years; 72.7% were women; and the median (IQR) STS-PROM score was 4.6 (3.3–7.0).

Table 1

Baseline and procedural characteristics of the population

 Overall cohort (n = 88)
Age (years)83 (80–86)
Female, n (%)64 (72.7)
Diabetes mellitus30 (34.1)
Atrial fibrillation/flutter, n (%)19 (21.6)
Paroxysmal8 (42.1)
Permanent11 (57.9)
STS-PROM score (%)4.6 (3.3–7.0)
CHADS-VASc2 score (%)4.5 (4.0–5.0)
ECG
 PR interval (ms)167 (153–192)
 QRS duration (ms)94 (80.0–100.0)
Echocardiography
 LVEF (%)60 (55–61.2)
 Mean gradient (mmHg)40.5 (30.9–48)
 Aortic valve area (cm2)0.71 (0.54–0.87)
Approach
 Transfemoral84 (95.5)
 Transapical/transaortic1 (1.1)
 Subclavian/transcarotid3 (3.4)
New-onset persistent LBBB
PR interval (ms)194 ± 46
QRS duration (ms)140 (130–156)
Time to implantable monitor, days post-TAVR4.8 (3–6)
Type of device, LINQ/XT84%/16%
Hospitalization length (days)6 (4–7)
 Overall cohort (n = 88)
Age (years)83 (80–86)
Female, n (%)64 (72.7)
Diabetes mellitus30 (34.1)
Atrial fibrillation/flutter, n (%)19 (21.6)
Paroxysmal8 (42.1)
Permanent11 (57.9)
STS-PROM score (%)4.6 (3.3–7.0)
CHADS-VASc2 score (%)4.5 (4.0–5.0)
ECG
 PR interval (ms)167 (153–192)
 QRS duration (ms)94 (80.0–100.0)
Echocardiography
 LVEF (%)60 (55–61.2)
 Mean gradient (mmHg)40.5 (30.9–48)
 Aortic valve area (cm2)0.71 (0.54–0.87)
Approach
 Transfemoral84 (95.5)
 Transapical/transaortic1 (1.1)
 Subclavian/transcarotid3 (3.4)
New-onset persistent LBBB
PR interval (ms)194 ± 46
QRS duration (ms)140 (130–156)
Time to implantable monitor, days post-TAVR4.8 (3–6)
Type of device, LINQ/XT84%/16%
Hospitalization length (days)6 (4–7)

Values are shown as n (%) for qualitative variables; mean ± SD for normally distributed quantitative variables; or median (interquartile range) for non-normally distributed quantitative variables.

ECG, electrocardiogram; LBBB, left bundle branch block; LVEF, left ventricular ejection fraction; STS-PROM, Society of Thoracic Surgeons Predicted Risk of Mortality.

Table 1

Baseline and procedural characteristics of the population

 Overall cohort (n = 88)
Age (years)83 (80–86)
Female, n (%)64 (72.7)
Diabetes mellitus30 (34.1)
Atrial fibrillation/flutter, n (%)19 (21.6)
Paroxysmal8 (42.1)
Permanent11 (57.9)
STS-PROM score (%)4.6 (3.3–7.0)
CHADS-VASc2 score (%)4.5 (4.0–5.0)
ECG
 PR interval (ms)167 (153–192)
 QRS duration (ms)94 (80.0–100.0)
Echocardiography
 LVEF (%)60 (55–61.2)
 Mean gradient (mmHg)40.5 (30.9–48)
 Aortic valve area (cm2)0.71 (0.54–0.87)
Approach
 Transfemoral84 (95.5)
 Transapical/transaortic1 (1.1)
 Subclavian/transcarotid3 (3.4)
New-onset persistent LBBB
PR interval (ms)194 ± 46
QRS duration (ms)140 (130–156)
Time to implantable monitor, days post-TAVR4.8 (3–6)
Type of device, LINQ/XT84%/16%
Hospitalization length (days)6 (4–7)
 Overall cohort (n = 88)
Age (years)83 (80–86)
Female, n (%)64 (72.7)
Diabetes mellitus30 (34.1)
Atrial fibrillation/flutter, n (%)19 (21.6)
Paroxysmal8 (42.1)
Permanent11 (57.9)
STS-PROM score (%)4.6 (3.3–7.0)
CHADS-VASc2 score (%)4.5 (4.0–5.0)
ECG
 PR interval (ms)167 (153–192)
 QRS duration (ms)94 (80.0–100.0)
Echocardiography
 LVEF (%)60 (55–61.2)
 Mean gradient (mmHg)40.5 (30.9–48)
 Aortic valve area (cm2)0.71 (0.54–0.87)
Approach
 Transfemoral84 (95.5)
 Transapical/transaortic1 (1.1)
 Subclavian/transcarotid3 (3.4)
New-onset persistent LBBB
PR interval (ms)194 ± 46
QRS duration (ms)140 (130–156)
Time to implantable monitor, days post-TAVR4.8 (3–6)
Type of device, LINQ/XT84%/16%
Hospitalization length (days)6 (4–7)

Values are shown as n (%) for qualitative variables; mean ± SD for normally distributed quantitative variables; or median (interquartile range) for non-normally distributed quantitative variables.

ECG, electrocardiogram; LBBB, left bundle branch block; LVEF, left ventricular ejection fraction; STS-PROM, Society of Thoracic Surgeons Predicted Risk of Mortality.

Global arrhythmic burden

Overall, 411 new arrhythmic events were detected in 58 patients (65.9%) within the total 2-year post-TAVR follow-up, with a median number of 2 (1–4) events per patient. Symptoms were reported in 12 out of 58 patients (20.7%) and a change in treatment was required in 25 out of 58 patients (43.1%).

The following types of arrhythmic events were found: 310 tachyarrhythmic events including 120 AF/AFL, 111 AT, 72 SVT, 6 non-sustained ventricular tachycardia (NSVT), and 1 VT and 101 bradyarrhythmic events in 33 patients, including 35 HAVB events and 66 severe bradycardia events. Complete details of arrhythmic events can be found in Table 2 and Supplementary material online, Table S1. Duration and symptoms of all severe bradycardia, non-sustained ventricular tachycardia, and atrial fibrillation/atrial flutter events are shown in Supplementary material online, Table S2.

Table 2

Number of patients with arrhythmic events in early (≤1 year) and late follow-up (>1 year) (n = 88)

 Overall periodFirst yearSecond year
Global arrhythmic burden
 Patients with arrhythmic events583325
 Arrhythmic events per patient2 (1–4)2 (1–4)3 (1–4)
 Patients with arrhythmic events requiring treatment25169
Bradyarrhythmias
 Patients with bradyarrhythmic events331917
 Patients with high-degree atrioventricular block1486
 Patients with severe bradycardia191111
 Patients with bradyarrhythmias requiring treatment15105
 Pacemaker implantationa1165
 Change in medical treatment440
Tachyarrhythmias
 Atrial fibrillation/atrial flutter
  Patients with new episodes of atrial fibrillation/atrial flutterb20/6913/697/69
  Patients with new episodes of atrial fibrillation/atrial flutter leading to anticoagulation therapy954
 Ventricular tachycardia
  Patients with episodes of ventricular tachycardia642
  Ventricular tachycardia episodes per patient1 (1–1)1 (1–1)1 (1–1)
  Patients with ventricular tachycardia episodes who had a treatment modification220
  Cardiac resynchronization therapy-defibrillator (CRT-D)c202
 Overall periodFirst yearSecond year
Global arrhythmic burden
 Patients with arrhythmic events583325
 Arrhythmic events per patient2 (1–4)2 (1–4)3 (1–4)
 Patients with arrhythmic events requiring treatment25169
Bradyarrhythmias
 Patients with bradyarrhythmic events331917
 Patients with high-degree atrioventricular block1486
 Patients with severe bradycardia191111
 Patients with bradyarrhythmias requiring treatment15105
 Pacemaker implantationa1165
 Change in medical treatment440
Tachyarrhythmias
 Atrial fibrillation/atrial flutter
  Patients with new episodes of atrial fibrillation/atrial flutterb20/6913/697/69
  Patients with new episodes of atrial fibrillation/atrial flutter leading to anticoagulation therapy954
 Ventricular tachycardia
  Patients with episodes of ventricular tachycardia642
  Ventricular tachycardia episodes per patient1 (1–1)1 (1–1)1 (1–1)
  Patients with ventricular tachycardia episodes who had a treatment modification220
  Cardiac resynchronization therapy-defibrillator (CRT-D)c202

Values are expressed as n, n/n, or median (interquartile range).

aAll pacemakers were implanted with an RV apical lead; no leadless or conduction system pacing was used.

bOnly patients without prior atrial fibrillation are considered in the denominator for the percentage.

cCRT-D was implanted based on heart failure, not on ventricular arrhythmias.

Table 2

Number of patients with arrhythmic events in early (≤1 year) and late follow-up (>1 year) (n = 88)

 Overall periodFirst yearSecond year
Global arrhythmic burden
 Patients with arrhythmic events583325
 Arrhythmic events per patient2 (1–4)2 (1–4)3 (1–4)
 Patients with arrhythmic events requiring treatment25169
Bradyarrhythmias
 Patients with bradyarrhythmic events331917
 Patients with high-degree atrioventricular block1486
 Patients with severe bradycardia191111
 Patients with bradyarrhythmias requiring treatment15105
 Pacemaker implantationa1165
 Change in medical treatment440
Tachyarrhythmias
 Atrial fibrillation/atrial flutter
  Patients with new episodes of atrial fibrillation/atrial flutterb20/6913/697/69
  Patients with new episodes of atrial fibrillation/atrial flutter leading to anticoagulation therapy954
 Ventricular tachycardia
  Patients with episodes of ventricular tachycardia642
  Ventricular tachycardia episodes per patient1 (1–1)1 (1–1)1 (1–1)
  Patients with ventricular tachycardia episodes who had a treatment modification220
  Cardiac resynchronization therapy-defibrillator (CRT-D)c202
 Overall periodFirst yearSecond year
Global arrhythmic burden
 Patients with arrhythmic events583325
 Arrhythmic events per patient2 (1–4)2 (1–4)3 (1–4)
 Patients with arrhythmic events requiring treatment25169
Bradyarrhythmias
 Patients with bradyarrhythmic events331917
 Patients with high-degree atrioventricular block1486
 Patients with severe bradycardia191111
 Patients with bradyarrhythmias requiring treatment15105
 Pacemaker implantationa1165
 Change in medical treatment440
Tachyarrhythmias
 Atrial fibrillation/atrial flutter
  Patients with new episodes of atrial fibrillation/atrial flutterb20/6913/697/69
  Patients with new episodes of atrial fibrillation/atrial flutter leading to anticoagulation therapy954
 Ventricular tachycardia
  Patients with episodes of ventricular tachycardia642
  Ventricular tachycardia episodes per patient1 (1–1)1 (1–1)1 (1–1)
  Patients with ventricular tachycardia episodes who had a treatment modification220
  Cardiac resynchronization therapy-defibrillator (CRT-D)c202

Values are expressed as n, n/n, or median (interquartile range).

aAll pacemakers were implanted with an RV apical lead; no leadless or conduction system pacing was used.

bOnly patients without prior atrial fibrillation are considered in the denominator for the percentage.

cCRT-D was implanted based on heart failure, not on ventricular arrhythmias.

The time to first arrhythmic event is shown in Kaplan–Meier curves in Figure 1. Of the 411 arrhythmic events, 104 (25.3%) were detected in 25 out of 58 patients (43.1%) after the first year. Further details regarding distribution of events within the first or second year of follow-up are shown in Supplementary material online, Table S3.

Time to first arrhythmic episode in patients receiving Evolut system devices. (A) Time to the first arrhythmic (brady- or tachyarrhythmia) episode. (B) Time to the first episode of bradyarrhythmia. (C) Time to the first episode of high-degree atrioventricular block. (D) Time to the first episode of severe bradycardia. (E) Time to the first episode of tachyarrhythmia. (F) Time to the first episode of atrial fibrillation/atrial flutter.
Figure 1

Time to first arrhythmic episode in patients receiving Evolut system devices. (A) Time to the first arrhythmic (brady- or tachyarrhythmia) episode. (B) Time to the first episode of bradyarrhythmia. (C) Time to the first episode of high-degree atrioventricular block. (D) Time to the first episode of severe bradycardia. (E) Time to the first episode of tachyarrhythmia. (F) Time to the first episode of atrial fibrillation/atrial flutter.

Bradyarrhythmic events and need for pacemaker implantation

All bradyarrhythmic events occurring within the 2-year period are reported in Supplementary material online, Table S1. A bradyarrhythmic event was seen in 33 patients, but only 9 were symptomatic. The proportion of patients with severe bradycardia was higher than the proportion of patients with HAVB both in first and second years. Time to first bradyarrhythmic episode is shown in Figure 1B. Details on patients developing complete AVB can be found in Supplementary material online, Table S4. Bradyarrhythmic events showed different occurrence patterns in the first and second year, with a stable trend in HAVB events and an increasing trend for severe bradycardia in the second year of follow-up.

During the first year, bradyarrhythmic events occurred in 19 patients, with 10 patients requiring a change in treatment and only 3 being symptomatic. After the first year, bradyarrhythmic events occurred in 17 patients, with 5 patients requiring treatment. The PPM implantation in order to treat bradyarrhythmia was needed in six patients within the first year and in five patients during the second year. Only four of them showed symptoms. All patients received a standard right ventricular (RV) apical pacing. There were no cases of leadless or conduction system pacing. It is worth mentioning that not all patients with severe bradycardia underwent pacemaker implantation.

A cardiac resynchronization therapy defibrillator (CRT-D) was needed in two patients. One of them had a baseline ventricular dysfunction due to known ischaemic cardiomyopathy; the other one showed progressive ventricular dysfunction during follow-up after TAVR. While PPM implantation was distributed similarly between Year 1 and Year 2, all CRT-D implantations occurred within the second year of follow-up, due to a systolic dysfunction caused by ventricular dyssynchrony secondary to LBBB.

The reasons for PPM/CRT-D implantation were mainly HAVB or sinus node disease (SND). The PPM implantations within the first month were all due to HAVB, whereas indications in patients who required a PPM at a later stage were mainly due to either HAVB or SND in similar proportions. Of note, LBBB resolved only temporary in one patient and permanently in two patients receiving a PPM. Further details on post-procedural and follow-up ECGs in patients requiring a PPM or a CRT-D are shown in Table 3.

Table 3

Electrocardiograms post-procedure, at 30 days, 12 months, and 24 months for all patients requiring a permanent pacemaker or a cardiac resynchronization therapy defibrillator

Pat.TimingDays after TAVRPacemaker indicationSymptomsHeart rateHeart rhythmPR durationAVBQRS durationQRS morphologyLBBB resolutionQRS axis
1Post-procedure220Prolonged ventricular pauses
Discharge85AFNone132LBBBNoNormal
30 days81AF127LBBBNoNormal
12 months94AF138LBBBNoNormal
24 months66AF136LBBBNoNormal
2aPost-procedure393HAVB and left ventricular dysfunction
Discharge100NoneLBBBNoLeft
30 days
12 monthsSinusalLBBB
24 months
3Post-procedure48Slow atrial fibrillation
DischargeAFNone134LBBBNoLeft
30 days93AFNone120ILBBBNoLeft
12 months96AFNone86Left
24 months80AF
4Post-procedure7CHBSyncope
Discharge73Sinusal2021st degree146LBBBNoLeft
30 days66Sinusal188None146LBBBNoLeft
12 months64Sinusal2141st degree142LBBBNoNormal
24 months
5aPost-procedure596LBBB and left ventricular dysfunction
Discharge78Sinusal2241st degree178LBBBNoLeft
30 days91Sinusal2041st degree164LBBBNoLeft
12 months75Sinusal2201st degree174LBBBNoLeft
24 months84Paced164148
6Post-procedure41CHBLipothymia/syncope
Discharge75Sinusal2401st degree124LBBBNoLeft
30 days65Sinusal232116
12 months
24 months87Paced192120
7Post-procedure9AVBSyncope
Discharge101Sinusal2021st degree130LBBBNo
30 days
12 months
24 months
8Post-procedure571Sinus dysfunction, significant pauses
Discharge57Sinusal
30 days75Sinusal173None154LBBBNoUndetermined
12 months60Sinusal174None157LBBBNoLeft
24 months55Paced2021st degree161PacedNoNormal
9Post-procedure483Pause > 6 s
Discharge55Sinusal
30 days53Sinusal183158LBBBNo
12 months75SinusalNoneLBBBNoNormal
24 months81AFNone144LBBBYesNormal
10Post-procedure633Pause > 20 sPre-syncope
Discharge55AFNone180LBBBNoNormal
30 days60AFNone100LBBBYesNormal
12 months60AFNone100ILBBBYesNormal
24 months70AF110ILBBBYesNormal
11Post-procedure195HAVB
Discharge51AF2501st degree152LBBBNoNormal
30 days
12 months
24 months72PacedPacedLeft
12Post-procedure461CHB
Discharge75Sinusal2501st degreeLBBBNoUndetermined
30 days72Sinusal2101st degree110IVCDYesNormal
12 months67Sinusal2281st degree166LBBBNoLeft
24 months60Sinusal2001st degree162LBBBNoNormal
13Post-procedure399HAVB
Discharge81Sinusal188140LBBBNoLeft
30 days83Sinusal158144LBBBNoLeft
12 months73Sinusal162146LBBBNoLeft
24 months72Sinusal188132LBBBNoLeft
Pat.TimingDays after TAVRPacemaker indicationSymptomsHeart rateHeart rhythmPR durationAVBQRS durationQRS morphologyLBBB resolutionQRS axis
1Post-procedure220Prolonged ventricular pauses
Discharge85AFNone132LBBBNoNormal
30 days81AF127LBBBNoNormal
12 months94AF138LBBBNoNormal
24 months66AF136LBBBNoNormal
2aPost-procedure393HAVB and left ventricular dysfunction
Discharge100NoneLBBBNoLeft
30 days
12 monthsSinusalLBBB
24 months
3Post-procedure48Slow atrial fibrillation
DischargeAFNone134LBBBNoLeft
30 days93AFNone120ILBBBNoLeft
12 months96AFNone86Left
24 months80AF
4Post-procedure7CHBSyncope
Discharge73Sinusal2021st degree146LBBBNoLeft
30 days66Sinusal188None146LBBBNoLeft
12 months64Sinusal2141st degree142LBBBNoNormal
24 months
5aPost-procedure596LBBB and left ventricular dysfunction
Discharge78Sinusal2241st degree178LBBBNoLeft
30 days91Sinusal2041st degree164LBBBNoLeft
12 months75Sinusal2201st degree174LBBBNoLeft
24 months84Paced164148
6Post-procedure41CHBLipothymia/syncope
Discharge75Sinusal2401st degree124LBBBNoLeft
30 days65Sinusal232116
12 months
24 months87Paced192120
7Post-procedure9AVBSyncope
Discharge101Sinusal2021st degree130LBBBNo
30 days
12 months
24 months
8Post-procedure571Sinus dysfunction, significant pauses
Discharge57Sinusal
30 days75Sinusal173None154LBBBNoUndetermined
12 months60Sinusal174None157LBBBNoLeft
24 months55Paced2021st degree161PacedNoNormal
9Post-procedure483Pause > 6 s
Discharge55Sinusal
30 days53Sinusal183158LBBBNo
12 months75SinusalNoneLBBBNoNormal
24 months81AFNone144LBBBYesNormal
10Post-procedure633Pause > 20 sPre-syncope
Discharge55AFNone180LBBBNoNormal
30 days60AFNone100LBBBYesNormal
12 months60AFNone100ILBBBYesNormal
24 months70AF110ILBBBYesNormal
11Post-procedure195HAVB
Discharge51AF2501st degree152LBBBNoNormal
30 days
12 months
24 months72PacedPacedLeft
12Post-procedure461CHB
Discharge75Sinusal2501st degreeLBBBNoUndetermined
30 days72Sinusal2101st degree110IVCDYesNormal
12 months67Sinusal2281st degree166LBBBNoLeft
24 months60Sinusal2001st degree162LBBBNoNormal
13Post-procedure399HAVB
Discharge81Sinusal188140LBBBNoLeft
30 days83Sinusal158144LBBBNoLeft
12 months73Sinusal162146LBBBNoLeft
24 months72Sinusal188132LBBBNoLeft

Some ECGs were not available due to the COVID pandemic.

AF, atrial fibrillation; CHB, complete heart block; CRT-D, cardiac resynchronization therapy-defibrillator; ECG, electrocardiogram; HAVB, high-degree atrioventricular bloc; ILBBB, incomplete left bundle branch block; IVCD, intraventricular conduction delay; LBBB, left bundle branch block; PPM, pacemaker implantation; TAVR, transcatheter aortic valve replacement.

aPatients with CRT-D.

Table 3

Electrocardiograms post-procedure, at 30 days, 12 months, and 24 months for all patients requiring a permanent pacemaker or a cardiac resynchronization therapy defibrillator

Pat.TimingDays after TAVRPacemaker indicationSymptomsHeart rateHeart rhythmPR durationAVBQRS durationQRS morphologyLBBB resolutionQRS axis
1Post-procedure220Prolonged ventricular pauses
Discharge85AFNone132LBBBNoNormal
30 days81AF127LBBBNoNormal
12 months94AF138LBBBNoNormal
24 months66AF136LBBBNoNormal
2aPost-procedure393HAVB and left ventricular dysfunction
Discharge100NoneLBBBNoLeft
30 days
12 monthsSinusalLBBB
24 months
3Post-procedure48Slow atrial fibrillation
DischargeAFNone134LBBBNoLeft
30 days93AFNone120ILBBBNoLeft
12 months96AFNone86Left
24 months80AF
4Post-procedure7CHBSyncope
Discharge73Sinusal2021st degree146LBBBNoLeft
30 days66Sinusal188None146LBBBNoLeft
12 months64Sinusal2141st degree142LBBBNoNormal
24 months
5aPost-procedure596LBBB and left ventricular dysfunction
Discharge78Sinusal2241st degree178LBBBNoLeft
30 days91Sinusal2041st degree164LBBBNoLeft
12 months75Sinusal2201st degree174LBBBNoLeft
24 months84Paced164148
6Post-procedure41CHBLipothymia/syncope
Discharge75Sinusal2401st degree124LBBBNoLeft
30 days65Sinusal232116
12 months
24 months87Paced192120
7Post-procedure9AVBSyncope
Discharge101Sinusal2021st degree130LBBBNo
30 days
12 months
24 months
8Post-procedure571Sinus dysfunction, significant pauses
Discharge57Sinusal
30 days75Sinusal173None154LBBBNoUndetermined
12 months60Sinusal174None157LBBBNoLeft
24 months55Paced2021st degree161PacedNoNormal
9Post-procedure483Pause > 6 s
Discharge55Sinusal
30 days53Sinusal183158LBBBNo
12 months75SinusalNoneLBBBNoNormal
24 months81AFNone144LBBBYesNormal
10Post-procedure633Pause > 20 sPre-syncope
Discharge55AFNone180LBBBNoNormal
30 days60AFNone100LBBBYesNormal
12 months60AFNone100ILBBBYesNormal
24 months70AF110ILBBBYesNormal
11Post-procedure195HAVB
Discharge51AF2501st degree152LBBBNoNormal
30 days
12 months
24 months72PacedPacedLeft
12Post-procedure461CHB
Discharge75Sinusal2501st degreeLBBBNoUndetermined
30 days72Sinusal2101st degree110IVCDYesNormal
12 months67Sinusal2281st degree166LBBBNoLeft
24 months60Sinusal2001st degree162LBBBNoNormal
13Post-procedure399HAVB
Discharge81Sinusal188140LBBBNoLeft
30 days83Sinusal158144LBBBNoLeft
12 months73Sinusal162146LBBBNoLeft
24 months72Sinusal188132LBBBNoLeft
Pat.TimingDays after TAVRPacemaker indicationSymptomsHeart rateHeart rhythmPR durationAVBQRS durationQRS morphologyLBBB resolutionQRS axis
1Post-procedure220Prolonged ventricular pauses
Discharge85AFNone132LBBBNoNormal
30 days81AF127LBBBNoNormal
12 months94AF138LBBBNoNormal
24 months66AF136LBBBNoNormal
2aPost-procedure393HAVB and left ventricular dysfunction
Discharge100NoneLBBBNoLeft
30 days
12 monthsSinusalLBBB
24 months
3Post-procedure48Slow atrial fibrillation
DischargeAFNone134LBBBNoLeft
30 days93AFNone120ILBBBNoLeft
12 months96AFNone86Left
24 months80AF
4Post-procedure7CHBSyncope
Discharge73Sinusal2021st degree146LBBBNoLeft
30 days66Sinusal188None146LBBBNoLeft
12 months64Sinusal2141st degree142LBBBNoNormal
24 months
5aPost-procedure596LBBB and left ventricular dysfunction
Discharge78Sinusal2241st degree178LBBBNoLeft
30 days91Sinusal2041st degree164LBBBNoLeft
12 months75Sinusal2201st degree174LBBBNoLeft
24 months84Paced164148
6Post-procedure41CHBLipothymia/syncope
Discharge75Sinusal2401st degree124LBBBNoLeft
30 days65Sinusal232116
12 months
24 months87Paced192120
7Post-procedure9AVBSyncope
Discharge101Sinusal2021st degree130LBBBNo
30 days
12 months
24 months
8Post-procedure571Sinus dysfunction, significant pauses
Discharge57Sinusal
30 days75Sinusal173None154LBBBNoUndetermined
12 months60Sinusal174None157LBBBNoLeft
24 months55Paced2021st degree161PacedNoNormal
9Post-procedure483Pause > 6 s
Discharge55Sinusal
30 days53Sinusal183158LBBBNo
12 months75SinusalNoneLBBBNoNormal
24 months81AFNone144LBBBYesNormal
10Post-procedure633Pause > 20 sPre-syncope
Discharge55AFNone180LBBBNoNormal
30 days60AFNone100LBBBYesNormal
12 months60AFNone100ILBBBYesNormal
24 months70AF110ILBBBYesNormal
11Post-procedure195HAVB
Discharge51AF2501st degree152LBBBNoNormal
30 days
12 months
24 months72PacedPacedLeft
12Post-procedure461CHB
Discharge75Sinusal2501st degreeLBBBNoUndetermined
30 days72Sinusal2101st degree110IVCDYesNormal
12 months67Sinusal2281st degree166LBBBNoLeft
24 months60Sinusal2001st degree162LBBBNoNormal
13Post-procedure399HAVB
Discharge81Sinusal188140LBBBNoLeft
30 days83Sinusal158144LBBBNoLeft
12 months73Sinusal162146LBBBNoLeft
24 months72Sinusal188132LBBBNoLeft

Some ECGs were not available due to the COVID pandemic.

AF, atrial fibrillation; CHB, complete heart block; CRT-D, cardiac resynchronization therapy-defibrillator; ECG, electrocardiogram; HAVB, high-degree atrioventricular bloc; ILBBB, incomplete left bundle branch block; IVCD, intraventricular conduction delay; LBBB, left bundle branch block; PPM, pacemaker implantation; TAVR, transcatheter aortic valve replacement.

aPatients with CRT-D.

Tachyarrhythmic events

A total of 29 patients showed a tachyarrhythmic event. All tachyarrhythmic events occurring within the 2-year period are reported in Supplementary material online, Table S1. Time to first tachyarrhythmic episode is shown in Figure 1E. New episodes of AF/AFL were observed in 20 out of 69 patients (29%) and only 2 of them showed symptoms. Initiation of an anticoagulation therapy was required in nine of them.

In late follow-up (after the first year) analysis, recurrent AF/AFL events in patients with a previous AF/AFL event during the early follow-up (in the first year) were excluded. Length of AF was >1 h in six patients and >24 h in two patients, but <1 h in all other patients. Episodes of ventricular tachycardia were found in 7.9% of patients within 2 years, with only one patient being symptomatic. Most of these were non-sustained VTs. Treatment changes (initiation of beta-blockers) were required in two of them, but no implantation of ICD devices was needed.

Further details on change in medical treatment are shown in Supplementary material online, Table S5.

Clinical outcomes

The 2-year all-cause mortality was 19.3% (n = 17), with cardiovascular mortality accounting for less than half of fatal outcomes (n = 8). Number and type of arrhythmic events of patients who died are shown in Supplementary material online, Table S6. Two patients died from sudden death within the 2-year follow-up time after TAVR. Specific causes could not be established because their ILR could not be retrieved after death. Both cases occurred late during the follow-up period, one after 20 months and the other at 2 years, and both patients were known to have coronary artery disease. One of them had previously shown an episode of HAVB whereas the other had previously shown 12 episodes of HAVB, 17 severe bradycardia, and 1 NSVT. Treating physicians presumed such episodes were vagal. However, none of these conduction disturbances occurred within a few weeks before the sudden death episode. Unfortunately, no information on potential arrhythmic events was available for the last two days before their sudden death.

No ischaemic stroke events were reported. Two patients (2.3%) had major haemorrhagic strokes. One of them had AF at baseline and was on oral anticoagulation, whereas the other one did not show any AF episode at baseline and during the follow-up period.

Discussion

In patients with NOP-LBBB after TAVR with an Evolut valve, late arrhythmic events occurred in about two-thirds of the patients and translated into a treatment change in about half of them. The PPM implantation was required in 12.5% of patients at 2 years. While tachyarrhythmic events, including AF/AFL, NSVT, AT, and VT, showed a gradual decrease in frequency throughout the 2-year period, bradyarrhythmic events showed a different temporal pattern. The rate of HAVB episodes was mostly early after the TAVR procedure and then remained stable over the 2-year follow-up, whereas episodes of severe bradycardia kept increasing after the first year.

Most studies assessing conduction disturbances after TAVR have used a 2-week continuous ambulatory ECG monitoring,19 except for the recent study by Massouillé et al. which assessed AV conduction disturbances by means of continuous ambulatory ECG monitoring in patients with NOP-LBBB up to 12 months after TAVR, and the study by Nozica et al. assessing new-onset arrhythmias within 12 months using an implantable cardiac monitoring,11 which provided a valuable overall insight on new-onset arrhythmias after TAVR. The current study reports specific data for the new-generation Evolut valves, a main device type with specific characteristics and design (difference in frame, radial force, position of the valve within the annulus, valve positioning manoeuvres, etc.) having a specific impact on the conduction system after TAVR implantation. Another recent study using novel digital technology showed that a wearable smartwatch was able to detect a large proportion of clinical events due to conduction abnormalities. However, the follow-up was still short (30 days).12 To the best of our knowledge, apart from the MARE study,18 only the Reveal study has assessed the incidence of arrhythmic events within 1 year, with AF being the most common arrhythmia in that cohort.20 Our study not only focuses specifically on patients with NOP-LBBB after TAVR, which are known to have a higher risk of conduction disturbances, but reported events at a follow-up of 2 years, showing that new-onset arrhythmic events still occur in a substantial number of patients beyond the first year after TAVR. Thus, selected patients might benefit from a long-term monitoring. Furthermore, our results provide now a better support to the generalizability of the findings, since our study included a number of centres in Canada, the USA, and Europe. This reflects current practice and management of these patients in diverse scenarios.

Nowadays, the need for a PPM due to procedure-induced conduction disturbances remains a major issue of TAVR, especially when expanding the indication to a lower risk population, since this has been shown to increase both mortality and morbidity/heart failure at midterm follow-up, including heart failure hospitalization.21,22 In our study, following an initial 4-week higher risk period, PPM implantations were homogenously distributed throughout the 2-year follow-up period. However, while the majority of patients receiving a PPM during the first year had HAVB, those receiving a PPM during the second year had other indications for implantation such as SND. These results are consistent with previous studies showing the most common cause of immediate (<1 month) post-procedural PPM implantation to be HAVB.23 However, in the long-term, other reasons such as SND become more important; therefore, electrocardiographic monitoring in the post-TAVR years may be necessary beyond the strict monitoring that is warranted immediately after the procedure.4

In the seminal clinical trial with the initial CoreValve device, the need for PPM at 1 year reached 22.3%.24 In subsequent clinical trials, use of CoreValve, Evolut R, or Evolut PRO in low-risk patients resulted in PPM implantation in 19.4% of patients at 1 year,25 whereas use of CoreValve or Evolut R in intermediate-risk patients in SURTAVI trial was associated to a 25.9% rate of PPM need at 30 days.26 Importantly, cusp-overlapping, a new promising valve implantation technique, has recently shown the lowest PPM rate to date (9.8%) in patients receiving the Evolut PRO/PRO+, in the Optimize PRO Study.27

Real-world and registry data on PPM rates after TAVR with successive generations of CoreValve/Evolut prostheses have been controversial. As a consequence of improvements in the stent frame of the valve, several comparative analyses have shown a substantially lower PPM rate with both Evolut R and Evolut PRO as compared to previous generation CoreValve devices.6–8 A systematic review suggested a 16.3–27.7% rate with CoreValve device vs. 14.7–26.7% rate with Evolut R.28 However, in spite of some observational comparative studies showing a significant reduction of PPM need with Evolut PRO as compared to Evolut R,6,29 other comparisons have only shown a non-significant trend to lower PPM rates30,31 or even a numerically higher PPM rate7,9,10 with Evolut PRO as compared to Evolut R.

In the general population, the reported annual rate of PPM implantation in patients with LBBB is 1–2%.32 In our study, the annual rate of PPM implantation was notably higher (6.2%). This difference suggests that TAVR patients with NOP-LBBB are at a higher risk of developing significant bradyarrhythmias and that the interaction of the prosthesis with the conduction system might play a role beyond the immediate effects of the implantation. For conduction disturbances appearing at longer term after TAVR, the pro-arrhythmic substrate that is inherent to the remodelled LV in elderly patients with AVS could play an additional role, besides the procedure itself. Current expansion of TAVR implantation to a younger population could help to elucidate the most probable aetiology of long-term conduction disturbances after TAVR.

Only 4 out of 11 patients receiving a PPM showed symptoms. This could have impacted on the percentage of implanted PPM in our study population, since indications were mainly based on monitoring findings. In line with these findings, the LOOP study showed that ILR screening led to a six-fold increase in bradyarrhythmia diagnoses and to a significant increase in pacemaker implantations compared to usual care, but no changes in the risk of syncope or sudden death were observed.33 These results show that arrhythmic events found in monitoring recordings do not always translate into clinical consequences. Future studies will need to focus on the time point at which action should be taken from a therapeutic point of view for these patients. Recently, measurement of HV intervals is gaining importance as a tool for risk stratification of post-TAVR conduction disorders in patients with persistent LBBB, as shown in a recent study by Massouillé et al.34 Although several studies have suggested different cut-off values (ranging from HV > 55 to 75) that could be useful to stratify this subgroup of patients,34–36 a definite cut-off value is yet to be established. In fact, some heterogeneity still exists in the management of conduction disorders at discharge after TAVR, but both ILR and HV interval measurement will need to be considered as stratification tools in this high-risk subgroup of patients.

The Evolut R valve consists of a self-expanding nitinol frame that conforms to the anatomy of the native annulus.37 Recently, Spaccarotella et al. documented a significant degree of acute expansion of nitinol prostheses immediately after deployment of the valve in patients treated with the Evolut R system.38 The impact of this further expansion in the long-term is still poorly understood with this new-generation device. However, previous studies with the old-generation CoreValve prosthesis suggest that the decrease in paravalvular leak (PVL) over time may be due to a continued expansion of the frame,39 which could translate into a potential increase in PPM rates over time (PVL/PPM trade-off). However, our study suggests that although some differences could appear with the new Evolut PRO, the low number of patients with this specific valve model included in our study prevents us from drawing any solid conclusions.

Some procedural characteristics such as depth of valve implantation are known to have an impact on conduction disturbances after TAVR. Jilaihawi et al. reported that both PPM and LBBB rates showed a substantial decline (from 9.7 to 3% for PPM and from 25.8 to 9% for LBBB) when implanting a self-expanding device (CoreValve/Evolut) above the membranous septum, as measured by multislice computed tomography.40 Thus, implantation depth appears to be a critical consideration in patients who are prone to develop conduction disturbances. Also, the use of a cusp-overlapping projection technique for implantation of the Evolut system has translated into a significant reduction in PPM likely due to the higher (more aortic) implantation of the transcatheter valve.41 Furthermore, some studies have shown septum length and thickness to be predictors for PPM.42

The two cases of CRT-D implantation in our study occurred within the second year of follow-up, due to a systolic dysfunction caused by ventricular dyssynchrony secondary to LBBB. This suggests that patients with a NOP-LBBB after TAVR could also benefit from continuous echocardiographic monitoring in order to identify those patients who could profit from a resynchronization device therapy or an implantable defibrillator (ICD) in the long term. The implantation of a conduction system pacing device (His or left bundle area pacing) could also improve heart failure symptoms. In addition, all patients in this trial received a standard RV apical pacing device which can contribute to left ventricular dysfunction or increase the risk of atrial fibrillation. Recent trials43,44 have shown the feasibility of conduction system pacing in the TAVR population which can prevent left ventricular dysfunction and also decrease the risk of atrial fibrillation and even ventricular arrhythmias.45 This new pacing approach may change the outcome in patients after TAVR who require a PPM.

Nearly one-third of the study population showed a new-onset AF/AFL event within the 2 years of follow-up; AF/AFL episodes occurred in one-fifth of the cohort in the first year and in one-tenth of the cohort in the second year after TAVR. Of note, most of these episodes were clinically silent, probably due to the short duration of the episodes. The 1-year incidence was higher than in previous trials like the US CoreValve High Risk trial (15.9%)24 or the Evolut Low-Risk trial (9.8%).25 On the other hand, a similar incidence was found in the NOTION (Nordic Aortic Valve Intervention) trial, probably due to the use of ILR to detect new-onset AF/AFL events.46 Thus, some events may go unnoticed if patients are asymptomatic or if only standard ECGs are recorded. Data on 2-year AF/AFL incidence after TAVR are scarce. Findings from the REVEAL AF study were consistent with our current results, with 29.3% of patients at high risk for AF showing an episode of AF/AFL within 18 months.20 Even though our data showed a reduction in frequency, a substantial rate of new-onset AF/AFL remains over time. The importance of detecting such events cannot be overemphasized, not only because of the need to prevent a potential risk of stroke but also because most of these patients will require an anticoagulation therapy that can also be associated with adverse effects. Notably, oral anticoagulation was only initiated in about half of patients with new-onset AF/AFL in our study. Many of the newly appeared AF episodes in our study had the characteristics of subclinical AF. The risk of stroke in these patients is slightly higher than in patients without subclinical AF but lower than in patients with clinically overt AF. Initiation of oral anticoagulation in such patients is still a matter of debate, but it can be considered in patients with subclinical AF > 24 h after careful clinical risk evaluation. In any case, the benefit of oral anticoagulation should outweigh the risk of bleeding.47

The GALILEO trial showed a higher risk for both thromboembolic and bleeding events in patients on oral anticoagulation (compared to those receiving an antiplatelet therapy) in patients without an indication for oral anticoagulation48; the selection of those patients with new episodes of AF/AFL requiring anticoagulants should therefore be made very carefully. Two very recent studies would further support this view. A pooled meta-analysis including the ARTESiA49 and NOAH-AFNET 650 randomized trials in patients with short-duration atrial fibrillation (>6 min, <24 h) showed a reduced risk of stroke, but also an increased risk in major bleeding in patients with device-detected or subclinical AF.51 Based on these trials, shared decision-making with patients, discussing the risk of stroke or bleeding, and tailoring anticoagulation prescription to the high risk for stroke are favoured.52,53

Evidence on ventricular arrhythmias after TAVR is scarce, and its real impact on clinical outcomes after TAVR is still unknown. Studies on sudden cardiac death in patients with PPM after TAVR have shown ventricular tachycardia or ventricular fibrillation to be the cause of death in some cases.54,55 A previous study by Tempio et al. using continuous monitoring in patients who received either a CoreValve or a SapienXT prosthesis showed a very low rate of VT episodes (2%) and a significant reduction in ventricular arrhythmias at 1 month after TAVR.56 The cumulative events in our study were slightly higher (5.7% within the first year and 2.3% within the second year), although still decreasing over time. A potential explanation may be the improvement in systolic function resulting from the replacement of the stenotic valve itself, which could possibly reduce ventricular arrhythmic events. Despite the identification of non-sustained ventricular events in our study cohort, medical management was preferred (since patients had a left ventricular ejection fraction > 40%), and no ICD implantation was recommended.

Study limitations

This study was based on a multicentre and international cohort of patients from both North America and Europe, and minor differences in the interpretation and management of arrhythmic events cannot be excluded. However, even though the initial diagnosis and therapy were performed in each individual site, all arrhythmic events were independently adjudicated in a single electrocardiography core lab. Other predictive factors for arrhythmic events could not be adequately assessed due to the non-controlled single-arm design and the limited sample size of the study. Although unlikely, similar arrhythmic findings cannot be excluded in patients without NOP-LBBB due to a missing control group. Even though all contributing centres participated in the early part of the study (up to 2016) with the expected number of patients, most centres stopped their participation in the later part of it due to budget constraints along with clinical restrictions during the COVID-19 pandemic. The potential impact of some anatomical factors such as implantation depth on conduction disturbances after TAVR could not be assessed. Since new pacing modalities such as conduction system pacing or leadless pacing were not used in this study, the impact of these technologies on arrhythmic burden will require further studies. Our results were obtained with the previous Evolut R valve and may not be generalized to the latest generation of the Evolut System (PRO/PRO+ and FX), potentially showing a different HAVB prevalence. However, the stent frame of the Evolut system is similar in all valve types, and the possibility of major differences in late arrhythmic events remains unlikely. Also, the number and the current follow-up length of patients having received these most recent devices are still limited, and it will take some time before a 2-year follow-up study to detect the late arrhythmic burden by means of an implantable recorder can be completed. In the meantime, our results are useful for management decision-taking in the large number of current patients having received the Evolut R device, and could be used as a reference to confirm or assess a potential change of such arrhythmic events after the implantation of newer devices. Finally, it is worth noting that our results are not generalizable to other types of valves such as balloon-expandable prostheses, due to their differences in design and implantation technique, which may imply a different frequency of conduction disturbances.

Conclusions

In patients with NOP-LBBB after TAVR using a new-generation Evolut device, arrhythmic events were found in more than half of the study cohort within a 2-year follow-up period and translated into treatment changes in a significant proportion of cases. Although both tachyarrhythmic and bradyarrhythmic events seemed to decrease over time, PPM/ICD implantation still occurred during the second year of follow-up, which suggests a need for long-term monitoring in this subgroup of high-risk patients. As for events requiring PPM implantation, while the number of HAVB episodes remained relatively stable, SND became more important over time. The AF/AFL episodes represented a substantial proportion of all events, and this has important implications for the shared decision-making to start an anticoagulation regimen. The impact of new pacing modalities (leadless pacing or conduction system pacing) on arrhythmic burden in this population will require further studies. Finally, data on ventricular arrhythmias are still scarce in post-TAVR populations and should be investigated in future studies.

Supplementary material

Supplementary material is available at Europace online.

Funding

S.M.-P. has received a research grant from the ‘Fundación Alfonso Martín Escudero’ (Madrid, Spain). J.R.-C. holds the Research Chair in ‘Fondation Famille Jaques Larivière’ for the Development of Structural Heart Interventions.

Data availability

The data underlying this article will be shared on reasonable request to the corresponding author.

References

1

Auffret
 
V
,
Puri
 
R
,
Urena
 
M
,
Chamandi
 
C
,
Rodriguez-Gabella
 
T
,
Philippon
 
F
 et al.  
Conduction disturbances after transcatheter aortic valve replacement: current status and future perspectives
.
Circulation
 
2017
;
136
:
1049
69
.

2

Faroux
 
L
,
Chen
 
S
,
Muntané-Carol
 
G
,
Regueiro
 
A
,
Philippon
 
F
,
Sondergaard
 
L
 et al.  
Clinical impact of conduction disturbances in transcatheter aortic valve replacement recipients: a systematic review and meta-analysis
.
Eur Heart J
 
2020
;
41
:
2771
81
.

3

Wang
 
J
,
Liu
 
S
,
Han
 
X
,
Chen
 
Y
,
Chen
 
H
,
Wan
 
Z
 et al.  
Prognostic outcome of new-onset left bundle branch block after transcatheter aortic valve replacement in patients with aortic stenosis: a systematic review and meta-analysis
.
Front Cardiovasc Med
 
2022
;
9
:
842929
.

4

Sammour
 
Y
,
Krishnaswamy
 
A
,
Kumar
 
A
,
Puri
 
R
,
Tarakji
 
KG
,
Bazarbashi
 
N
 et al.  
Incidence, predictors, and implications of permanent pacemaker requirement after transcatheter aortic valve replacement
.
JACC Cardiovasc Interv
 
2021
;
14
:
115
34
.

5

Chamandi
 
C
,
Barbanti
 
M
,
Munoz-Garcia
 
A
,
Latib
 
A
,
Nombela-Franco
 
L
,
Gutiérrez-Ibanez
 
E
 et al.  
Long-term outcomes in patients with new-onset persistent left bundle branch block following TAVR
.
JACC Cardiovasc Interv
 
2019
;
12
:
1175
84
.

6

Loewenstein
 
I
,
Merdler
 
I
,
Hochstadt
 
A
,
Zahler
 
D
,
Finkelstein
 
A
,
Banai
 
S
 et al.  
Generational differences in outcomes of self-expanding valves for transcatheter aortic valve replacement
.
J Invasive Cardiol
 
2022
;
34
:
E326
33
.

7

Kroon
 
HG
,
van Gils
 
L
,
Ziviello
 
F
,
van Wiechen
 
M
,
Ooms
 
J
,
Rahhab
 
Z
 et al.  
Impact of baseline and newly acquired conduction disorders on need for permanent pacemakers with 3 consecutive generations of self-expanding transcatheter aortic heart valves
.
Cardiovasc Revasc Med
 
2022
;
34
:
40
5
.

8

Forrest
 
JK
,
Kaple
 
RK
,
Tang
 
GHL
,
Yakubov
 
SJ
,
Nazif
 
TM
,
Williams
 
MR
 et al.  
Three generations of self-expanding transcatheter aortic valves A report from the STS/ACC TVT registry
.
JACC Cardiovasc Interv
 
2020
;
13
:
170
9
.

9

Hellhammer
 
K
,
Piayda
 
K
,
Afzal
 
S
,
Kleinebrecht
 
L
,
Makosch
 
M
,
Hennig
 
I
 et al.  
The latest evolution of the Medtronic CoreValve system in the era of transcatheter aortic valve replacement: matched comparison of the Evolut PRO and Evolut R
.
JACC Cardiovasc Interv
 
2018
;
11
:
2314
22
.

10

Alvarado
 
T
,
Rivero
 
F
,
Diego
 
G
,
García-Guimaraes
 
M
,
Salamanca
 
J
,
Díez-Villanueva
 
P
 et al.  
Transcatheter aortic valve replacement using the new Evolut-Pro system: a prospective comparison with the Evolut-R device
.
J Thorac Dis
 
2021
;
13
:
4023
32
.

11

Nozica
 
N
,
Siontis
 
GCM
,
Elchinova
 
EG
,
Goulouti
 
E
,
Asami
 
M
,
Bartkowiak
 
J
 et al.  
Assessment of new onset arrhythmias after transcatheter aortic valve implantation using an implantable cardiac monitor
.
Front Cardiovasc Med
 
2022
;
9
:
876546
.

12

Liu
 
X
,
Fan
 
J
,
Guo
 
Y
,
Dai
 
H
,
Xu
 
J
,
Wang
 
L
 et al.  
Wearable smartwatch facilitated remote health management for patients undergoing transcatheter aortic valve replacement
.
JAHA
 
2022
;
11
:
e023219
.

13

Sana
 
F
,
Isselbacher
 
EM
,
Singh
 
JP
,
Heist
 
EK
,
Pathik
 
B
,
Armoundas
 
AA
.
Wearable devices for ambulatory cardiac monitoring: JACC state-of-the-art review
.
J Am Coll Cardiol
 
2020
;
75
:
1582
92
.

14

Rodés-Cabau
 
J
,
Urena
 
M
,
Nombela-Franco
 
L
,
Amat-Santos
 
I
,
Kleiman
 
N
,
Munoz-Garcia
 
A
 et al.  
Arrhythmic burden as determined by ambulatory continuous cardiac monitoring in patients with new-onset persistent left bundle branch block following transcatheter aortic valve replacement: the MARE study
.
JACC Cardiovasc Interv
 
2018
;
11
:
1495
505
.

15

Surawicz
 
B
,
Childers
 
R
,
Deal
 
BJ
,
Gettes
 
LS
.
AHA/ACCF/HRS recommendations for the standardization and interpretation of the electrocardiogram: part III: intraventricular conduction disturbances a scientific statement from the American Heart Association electrocardiography and arrhythmias committee
.
J Am Coll Cardiol
 
2009
;
53
:
976
81
.

16

Kappetein
 
AP
,
Head
 
SJ
,
Généreux
 
P
,
Piazza
 
N
,
Van Mieghem
 
NM
,
Blackstone
 
EH
 et al.  
Updated standardized endpoint definitions for transcatheter aortic valve implantation: the valve academic research consortium-2 consensus document
.
J Am Coll Cardiol
 
2012
;
60
:
1438
54
.

17

Tracy
 
CM
,
Epstein
 
AE
,
Darbar
 
D
,
Dimarco
 
JP
,
Dunbar
 
SB
,
Estes
 
NAM
 et al.  
2012 ACCF/AHA/HRS focused update incorporated into the ACCF/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association task force on practice guide
.
J Am Coll Cardiol
 
2013
;
61
:
e6
75
.

18

Muntané-Carol
 
G
,
Urena
 
M
,
Nombela-Franco
 
L
,
Amat-Santos
 
I
,
Kleiman
 
N
,
Munoz-Garcia
 
A
 et al.  
Arrhythmic burden in patients with new-onset persistent left bundle branch block after transcatheter aortic valve replacement: 2-year results of the MARE study
.
Europace
 
2021
;
23
:
254
63
.

19

Natarajan
 
MK
,
Sheth
 
TN
,
Wijeysundera
 
HC
,
Chavarria
 
J
,
Rodes-Cabau
 
J
,
Velianou
 
JL
 et al.  
Remote ECG monitoring to reduce complications following transcatheter aortic valve implantations: the redirect TAVI study
.
Europace
 
2022
;
24
:
1475
83
.

20

Reiffel
 
JA
,
Verma
 
A
,
Kowey
 
PR
,
Halperin
 
JL
,
Gersh
 
BJ
,
Wachter
 
R
 et al.  
Incidence of previously undiagnosed atrial fibrillation using insertable cardiac monitors in a high-risk population the REVEAL AF study supplemental content
.
JAMA Cardiol
 
2017
;
2
:
1120
7
.

21

Clementy
 
N
,
Bisson
 
A
,
Bodin
 
A
,
Herbert
 
J
,
Lacour
 
T
,
Saint Etienne
 
C
 et al.  
Outcomes associated with pacemaker implantation following transcatheter aortic valve replacement: a nationwide cohort study
.
Hear Rhythm
 
2021
;
18
:
2027
32
.

22

Natanzon
 
SS
,
Fardman
 
A
,
Koren-Morag
 
N
,
Fefer
 
P
,
Maor
 
E
,
Guetta
 
V
 et al.  
Pacing burden and clinical outcomes after transcatheter aortic valve replacement—a real-world registry report
.
Hear Rhythm
 
2022
;
19
:
1508
15
.

23

Malebranche
 
D
,
Bartkowiak
 
J
,
Ryffel
 
C
,
Bernhard
 
B
,
Elsmaan
 
M
,
Nozica
 
N
 et al.  
Validation of the 2019 expert consensus algorithm for the management of conduction disturbances after TAVR
.
JACC Cardiovasc Interv
 
2021
;
14
:
981
91
.

24

Adams
 
DH
,
Popma
 
JJ
,
Reardon
 
MJ
,
Yakubov
 
SJ
,
Coselli
 
JS
,
Deeb
 
GM
 et al.  
Transcatheter aortic-valve replacement with a self-expanding prosthesis
.
N Engl J Med
 
2014
;
370
:
1790
8
.

25

Popma
 
JJ
,
Deeb
 
GM
,
Yakubov
 
SJ
,
Mumtaz
 
M
,
Gada
 
H
,
O’Hair
 
D
 et al.  
Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients
.
N Engl J Med
 
2019
;
380
:
1706
15
.

26

Reardon
 
MJ
,
Van Mieghem
 
NM
,
Popma
 
JJ
,
Kleiman
 
NS
,
Søndergaard
 
L
,
Mumtaz
 
M
 et al.  
Surgical or transcatheter aortic-valve replacement in intermediate-risk patients
.
N Engl J Med
 
2017
;
376
:
1321
31
.

27

Grubb
 
KJ
,
Gada
 
H
,
Mittal
 
S
,
Nazif
 
T
,
Rodés-Cabau
 
J
,
Fraser
 
DGW
 et al.  
Clinical impact of standardized TAVR technique and care pathway insights from the optimize PRO study
.
JACC Cardiovasc Interv
 
2023
;
16
:
558
70
.

28

Van Rosendael
 
PJ
,
Delgado
 
V
,
Bax
 
JJ
.
Pacemaker implantation rate after transcatheter aortic valve implantation with early and new-generation devices: a systematic review
.
Eur Heart J
 
2018
;
39
:
2003
13
.

29

Kalogeras
 
K
,
Ruparelia
 
N
,
Kabir
 
T
,
Jabbour
 
R
,
Naganuma
 
T
,
Vavuranakis
 
M
 et al.  
Comparison of the self-expanding Evolut-PRO transcatheter aortic valve to its predecessor Evolut-R in the real world multicenter ATLAS registry
.
Int J Cardiol
 
2020
;
310
:
120
5
.

30

Rao
 
G
,
Sheth
 
S
,
Donnelly
 
J
,
Scatola
 
A
,
Tariq
 
U
,
Laighold
 
S
 et al.  
Clinical study early real-world experience with CoreValve Evolut PRO and R systems for transcatheter aortic valve replacement
.
J Interv Cardiol
 
2019
;
2019
:
1906814
.

31

Regazzoli
 
D
,
Chiarito
 
M
,
Cannata
 
F
,
Pagnesi
 
M
,
Miura
 
M
,
Ziviello
 
F
 et al.  
Transcatheter self-expandable valve implantation for aortic stenosis in small aortic annuli: the TAVI-SMALL registry
.
JACC Cardiovasc Interv
 
2020
;
13
:
196
206
.

32

Glikson
 
M
,
Nielsen
 
JC
,
Kronborg
 
MB
,
Michowitz
 
Y
,
Auricchio
 
A
,
Barbash
 
IM
 et al.  
2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy
.
Eur Heart J
 
2021
;
42
:
3427
520
.

33

Svendsen
 
JH
,
Diederichsen
 
SZ
,
Højberg
 
S
,
Krieger
 
DW
,
Graff
 
C
,
Kronborg
 
C
 et al.  
Implantable loop recorder detection of atrial fibrillation to prevent stroke (the LOOP study): a randomised controlled trial
.
Lancet
 
2021
;
398
:
1507
16
.

34

Massoullié
 
G
,
Ploux
 
S
,
Souteyrand
 
G
,
Mondoly
 
P
,
Pereira
 
B
,
Amabile
 
N
 et al.  
Incidence and management of atrioventricular conduction disorders in new-onset left bundle branch block after TAVI: a prospective multicenter study
.
Hear Rhythm
 
2023
;
20
:
699
706
.

35

Badertscher
 
P
,
Knecht
 
S
,
Spies
 
F
,
Auberson
 
C
,
Salis
 
M
,
Jeger
 
RV
 et al.  
Value of periprocedural electrophysiology testing during transcatheter aortic valve replacement for risk stratification of patients with new-onset left bundle-branch block
.
J Am Hear Assoc Cardiovasc Cerebrovasc Dis
 
2022
;
11
:
e026239
.

36

Raad
 
M
,
Greenberg
 
J
,
Altawil
 
M
,
Lee
 
J
,
Wang
 
DD
,
Oudeif
 
A
 et al.  
The transcatheter aortic valve replacement-conduction study: the value of the his-ventricular interval in risk stratification for post-TAVR atrioventricular-block
.
Struct Hear
 
2024
;
8
:
100296
.

37

Bhogal
 
S
,
Rogers
 
T
,
Merdler
 
I
,
Reddy
 
P
,
Ali
 
S
,
Shea
 
C
 et al.  
Evolut PRO/PRO+ versus Evolut R system for transcatheter aortic valve replacement
.
Int J Cardiol
 
2023
;
389
:
131196
.

38

Spaccarotella
 
CAM
,
Sorrentino
 
S
,
Mongiardo
 
A
,
Riverso
 
V
,
Mancuso
 
C
,
Giordano
 
S
 et al.  
Acute expansion of nitinol-based self-expandable device in patients undergoing transfemoral aortic valve implantation
.
J Cardiovasc Med
 
2023
;
24
:
308
12
.

39

Oh
 
JK
,
Little
 
SH
,
Abdelmoneim
 
SS
,
Reardon
 
MJ
,
Kleiman
 
NS
,
Lin
 
G
 et al.  
Regression of paravalvular aortic regurgitation and remodeling of self-expanding transcatheter aortic valve: an observation from the CoreValve U.S. Pivotal trial
.
JACC Cardiovasc Imaging
 
2015
;
8
:
1364
75
.

40

Jilaihawi
 
H
,
Zhao
 
Z
,
Du
 
R
,
Staniloae
 
C
,
Saric
 
M
,
Neuburger
 
PJ
 et al.  
Minimizing permanent pacemaker following repositionable self-expanding transcatheter aortic valve replacement
.
JACC Cardiovasc Interv
 
2019
;
12
:
1796
807
.

41

Pascual
 
I
,
Hernández-Vaquero
 
D
,
Alperi
 
A
,
Almendarez
 
M
,
Avanzas
 
P
,
Kalavrouziotis
 
D
 et al.  
Permanent pacemaker reduction using cusp-overlapping projection in TAVR: a propensity score analysis
.
JACC Cardiovasc Interv
 
2022
;
15
:
150
61
.

42

Na
 
J
,
Nixon
 
B
,
Childress
 
J
,
Han
 
P
,
Norcross
 
G
,
Liskov
 
S
 et al.  
Anatomical characteristics of the membranous septum are predictive of pacemaker requirement in patients undergoing transcatheter aortic valve replacement
.
J Interv Card Electrophysiol
 
2022
;
63
:
449
59
.

43

Shah
 
K
,
Williamson
 
BD
,
Kutinsky
 
I
,
Bhardwaj
 
R
,
Contractor
 
T
,
Turagam
 
MK
 et al.  
Conduction system pacing in prosthetic heart valves
.
J Interv Card Electrophysiol
 
2023
;
66
:
561
6
.

44

Vijayaraman
 
P
,
Cano
 
Ó
,
Koruth
 
JS
,
Subzposh
 
FA
,
Nanda
 
S
,
Pugliese
 
J
 et al.  
His-Purkinje conduction system pacing following transcatheter aortic valve replacement: feasibility and safety
.
JACC Clin Electrophysiol
 
2020
;
6
:
649
57
.

45

Herweg
 
B
,
Sharma
 
PS
,
Cano
 
Ó
,
Ponnusamy
 
SS
,
Zanon
 
F
,
Jastrzebski
 
M
 et al.  
Arrhythmic risk in biventricular pacing compared with left bundle branch area pacing: results from the I-CLAS study
.
Circulation
 
2024
;
149
:
379
90
.

46

Thyregod
 
HGH
,
Steinbrüchel
 
DA
,
Ihlemann
 
N
,
Nissen
 
H
,
Kjeldsen
 
BJ
,
Petursson
 
P
 et al.  
Transcatheter versus surgical aortic valve replacement in patients with severe aortic valve stenosis: 1-year results from the all-comers NOTION randomized clinical trial
.
J Am Coll Cardiol
 
2015
;
65
:
2184
94
.

47

Toennis
 
T
,
Bertaglia
 
E
,
Brandes
 
A
,
Dichtl
 
W
,
Fluschnik
 
N
,
de Groot
 
JR
 et al.  
The influence of atrial high-rate episodes on stroke and cardiovascular death: an update
.
Europace
 
2023
;
25
:
1
10
.

48

Dangas
 
GD
,
Tijssen
 
JGP
,
Wöhrle
 
J
,
Søndergaard
 
L
,
Gilard
 
M
,
Möllmann
 
H
 et al.  
A controlled trial of rivaroxaban after transcatheter aortic-valve replacement
.
N Engl J Med
 
2020
;
382
:
120
9
.

49

Healey
 
JS
,
Lopes
 
RD
,
Granger
 
CB
,
Alings
 
M
,
Rivard
 
L
,
McIntyre
 
WF
 et al.  
Apixaban for stroke prevention in subclinical atrial fibrillation
.
N Engl J Med
 
2024
;
390
:
107
17
.

50

Kirchhof
 
P
,
Toennis
 
T
,
Goette
 
A
,
Camm
 
AJ
,
Diener
 
HC
,
Becher
 
N
 et al.  
Anticoagulation with edoxaban in patients with atrial high-rate episodes
.
N Engl J Med
 
2023
;
389
:
1167
79
.

51

McIntyre
 
WF
,
Benz
 
AP
,
Becher
 
N
,
Healey
 
JS
,
Granger
 
CB
,
Rivard
 
L
 et al.  
Direct oral anticoagulants for stroke prevention in patients with device-detected atrial fibrillation: a study-level meta-analysis of the NOAH-AFNET 6 and ARTESiA trials
.
Circulation
 
2024
;
149
:
981
8
.

52

Mcintyre
 
WF
,
Healey
 
JS
,
Lopes
 
RD
.
When to prescribe oral anticoagulation for patients with device-detected subclinical atrial fibrillation? Insights from the NOAH-AFNET 6 and ARTESiA trials
.
Eur J Intern Med
 
2024
;
123
:
52
4
.

53

Boriani
 
G
,
Gerra
 
L
,
Mei
 
DA
,
Bonini
 
N
,
Vitolo
 
M
,
Proietti
 
M
 et al.  
Detection of subclinical atrial fibrillation with cardiac implanted electronic devices: what decision making on anticoagulation after the NOAH and ARTESiA trials?
 
Eur J Intern Med
 
2024
;
123
:
37
41
.

54

Beinart
 
R
,
Danik
 
S
,
Palacios
 
I
,
Barrett
 
C
,
Inglessis
 
I
,
Agnihotri
 
A
 et al.  
Ventricular tachycardia following trans-apical aortic valve replacement
.
Europace
 
2012
;
14
:
450
2
.

55

Dvir
 
D
,
Haim
 
M
,
Kornowski
 
R
.
Left-ventricular outflow tract ventricular-tachycardia event following corevalve transcatheter aortic-valve implantation
.
Catheter Cardiovasc Interv
 
2012
;
79
:
331
3
.

56

Tempio
 
D
,
Pruiti
 
GP
,
Conti
 
S
,
Romano
 
SA
,
Tavano
 
E
,
Capodanno
 
D
 et al.  
Ventricular arrhythmias in aortic valve stenosis before and after transcatheter aortic valve implantation
.
Europace
 
2015
;
17
:
1136
40
.

Author notes

Conflict of interest: F.A. has equity of Corify Care (Madrid, Spain) and served on the advisory board of Medtronic. M.D. has honoraria from Abbott, Boston Scientific, and Medtronic. J.G.W. is a consultant for Edwards Lifesciences and has received research funding from Medtronic and Edwards Lifesciences. J.R.-C has received institutional research grants and consultant/speaker fees from Medtronic. All other authors have no conflicts of interest related to the subject of the article.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.

Supplementary data