Abstract

Aims

Wolff–Parkinson–White (WPW) syndrome is a cardiovascular disease characterized by abnormal atrioventricular conduction facilitated by accessory pathways (APs). Invasive catheter ablation of the AP represents the primary treatment modality. Accurate localization of APs is crucial for successful ablation outcomes, but current diagnostic algorithms based on the 12-lead electrocardiogram (ECG) often struggle with precise determination of AP locations. In order to gain insight into the mechanisms underlying localization failures observed in current diagnostic algorithms, we employ a virtual cardiac model to elucidate the relationship between AP location and ECG morphology.

Methods and results

We first introduce a cardiac model of electrophysiology that was specifically tailored to represent antegrade APs in the form of a short atrioventricular bypass tract. Locations of antegrade APs were then automatically swept across both ventricles in the virtual model to generate a synthetic ECG database consisting of 9271 signals. Regional grouping of antegrade APs revealed overarching morphological patterns originating from diverse cardiac regions. We then applied variance-based sensitivity analysis relying on polynomial chaos expansion on the ECG database to mathematically quantify how variation in AP location and timing relates to morphological variation in the 12-lead ECG. We utilized our mechanistic virtual model to showcase the limitations of AP localization using standard ECG-based algorithms and provide mechanistic explanations through exemplary simulations.

Conclusion

Our findings highlight the potential of virtual models of cardiac electrophysiology not only to deepen our understanding of the underlying mechanisms of WPW syndrome but also to potentially enhance the diagnostic accuracy of ECG-based algorithms and facilitate personalized treatment planning.

What’s new?
  • We present a physiologically detailed virtual model of cardiac electrophysiology that represents Wolff–Parkinson–White syndrome with an accessory pathway manifesting as an antegrade bypass tract.

  • We quantified how sensitive the morphology of the 12-lead electrocardiogram (ECG) is to the location and timing of pre-ventricular ventricular excitation of an accessory pathway manifesting as an antegrade bypass tract using sensitivity analysis.

  • We demonstrate failure mechanisms of ECG-based localization algorithms to accurately localize accessory pathways inserting on the right and left side of the septum using exemplary simulations.

Introduction

Wolff–Parkinson–White (WPW) syndrome is a heart rhythm disorder that is characterized by the presence of at least one or more accessory pathways (APs) facilitating abnormal atrioventricular conduction. WPW typically manifests in children and affects between 0.03% and 0.5%in the general population.1,2 Without treatment, the condition can lead to paroxysmal palpitations and morbidity through supraventricular tachycardias or sudden cardiac arrest.1,3 Treatment of WPW syndrome typically involves invasive catheter ablation in efforts to restore the normal heart rhythm. Approximately, 6% of ablations are still unsuccessful.4,2,5

From a clinical perspective, accurate prediction of the location of the AP is thus of great value as it relates to procedural planning, risk prediction, and procedural outcome.6 Determining whether an antegrade AP exists in the right or left ventricles (RV or LV) guides the decision to perform trans-septal puncture or retro-aortic approach, respectively. Discrimination of left-sided APs also allows for pre-procedural informed consent emphasizing trans-septal/retro-aortic approach and associated risks (e.g. arterial injury) compared to strictly right-sided catheterization. Right-sided lateral pathways may implicate lower success of ablation due to frequently appearing arborization of APs.7 Right-sided anterolateral or anterior APs may represent as a duplication of the His–Punkinje system necessitating more extensive ablation and different tools (e.g. steerable sheaths) compared to ‘standard’ AP ablation procedures.8 Accurately predicting septal pathways is also of special interest to physicians because of the potential vicinity to the intrinsic conduction system, especially the His-bundle.9 Parahisian AP ablation procedures may require extensive mapping of atrial and ventricular insertion sites aiming to find a target site far enough distant to the His-bundle to safely perform ablation. Procedure time may increase significantly for these specific AP locations. The risk of injuring the intrinsic conduction system and subsequent pacemaker implantation is also higher in e.g. anteroseptal/parahisian APs compared to free wall APs.

Established clinical diagnostic trees for pre-procedural localization of the AP inferred from the 12-lead electrocardiogram (ECG) exist.10–15 These algorithms depend on identifying morphological features in the 12-lead ECG that uniquely indicate pre-excitation within anatomical sites across the atrioventricular junction region and thus maximize differentiation. Current ECG algorithms and diagnostic trees are primarily built upon data acquired through general clinical studies that aim to understand the influence of WPW manifestation on clinically relevant ECG metrics, typically within the delta wave and QRS complex.16,17 For example, both El Hamriti et al.11 and Pambrun et al.12 rely on the polarity of the V1 lead to differentiate between the mitral and tricuspid valves, respectively, relating to the LV and RV. The Easy-WPW algorithm by El Hamriti et al.11 then relies on what leads exhibit the most positive delta waves, and the QRS transition in the precordial leads. Pambrun et al.12 subsequently rely on the number of positive inferior leads (II, aVF, and III). Further differentiation in right-sided APs relies on the polarity of V3, while the V1/I ratio and lead II morphology are utilized within the LV.

Current ECG algorithms for AP localization prediction show acceptable accuracy for the determination of left- vs. right-sided APs, but still suffer difficulties predicting the exact location.18 Furthermore, such diagnostic trees may lack sensitivity and specificity within regions such as the septum, are not as accurate on a paediatric population,11,12 and fail under more complex manifestations of the disease, especially when compounded with multiple pathways or additional cardiac disorders.19 In situations where localization fails with a standard ablation catheter, an electrophysiological study with electro-anatomical mapping is performed to better elucidate AP location. Doing so results in increased procedural time and radiation exposure to both patient and personnel, as well as increased cost due to the use of a mapping catheter.

Virtual models of cardiac electrophysiology that are capable of representing WPW syndrome could be used to overcome such limitations. In general, virtual models of cardiac electrophysiology are becoming increasingly feasible for exploring the mechanisms behind various cardiac diseases, serving as clinical tools, and generating data to supplement clinical data for machine learning approaches.20–22 When the virtual models are personalized in terms of both anatomy and electrophysiology, often referred to as a cardiac digital twin or patient-specific model, they can be used for precision medicine in the form of prognostics, diagnostics, and treatment planning. While computational models have been used to study WPW,23,24 such studies have not been performed in a whole-heart personalized model capable of capturing the patient-specific electrophysiological detail of the disease.

Furthermore, sensitivity analysis (SA) methods on patient-specific models of WPW syndrome have not been yet deployed to understand how the location of the AP influences the morphology of the 12-lead ECG and, therefore, the success of localization algorithms. Performing sensitivity analysis and related uncertainty quantification on the virtual models provides a systematic framework to mathematically quantify how variation in the electrophysiological excitation (or a derived 12-lead ECG signal) is related to underlying model parameters. SA and uncertainty quantification have, thus, been recently deployed to the virtual models in various forms to different cardiac problems at different scales to better understand parameter-output relationships.25–28 Employing the variance-based SA method called polynomial chaos expansion (PCE)29 provides the global sensitivities for a given set of model evaluations. PCE has been recently applied to cardiac modelling,30–34 providing parameter sensitivities in the form of Sobol coefficients over the entire 12-lead ECG, enabling the localization of the signal change with respect to both timing and lead.

We utilized a virtual model of cardiac electrophysiology of WPW to provide insight into the genesis of morphological changes in 12-lead ECG that could influence pre-procedural localization. A whole-heart patient-specific model was constructed using an automated pipeline for a single male patient with normal sinus rhythm and adapted for WPW in previous work.35,36 A short atrioventricular bypass tract allowing antegrade conduction was inserted within the patient-specific model to allow abnormal conduction from the atria to the ventricles to facilitate pre-ventricular excitation. Various locations of antegrade APs were automatically swept using latin-hyper-cube sampling to generate a synthetic 12-lead ECG database. All ECGs within the database were grouped according to regions within the clinically standard bullseye plot to observe regional variation in 12-lead ECG morphology. How strongly AP location influences morphology in the 12-lead ECG was quantified by SA utilizing a PCE framework. Exemplary simulations of APs located on either side of the septum were conducted to showcase possible mechanisms behind localization failure of common ECG algorithms. All results are discussed in the context of ECG criteria as determined by clinically utilized algorithms of Pambrun et al.12 and El Hamriti et al.11

Methods

A model of WPW assuming antegrade AP conduction was first generated within a pre-existing patient-specific model for a single male subject. A synthetic ECG database was then generated by automatically varying the AP location. The database was generated using a modelling pipeline constructed within the openCARP framework.37 To quantify how strongly AP location influenced variation in signal morphology within the 12-lead ECG, the 12-lead ECGs were regionally grouped according to a bullseye plot and UQ with PCE was applied. A graphical abstract provides the methodological overview (Figure 1).

We develop a patient-specific model of Wolff–Parkinson–White (WPW) syndrome from clinical magnetic resonances images that has been personalized with the 12-lead electrocardiogram (ECG) under sinus rhythm. Using efficient cardiac simulation, we perform three computational studies. AP, accessory pathway; MRI, magnetic resonances image.
Figure 1.

We develop a patient-specific model of Wolff–Parkinson–White (WPW) syndrome from clinical magnetic resonances images that has been personalized with the 12-lead electrocardiogram (ECG) under sinus rhythm. Using efficient cardiac simulation, we perform three computational studies. AP, accessory pathway; MRI, magnetic resonances image.

Acquisition of clinical data

Clinical data were acquired for a single male patient of 45 years under normal sinus rhythm. Two magnetic resonance images (MRIs) were acquired sequentially: (i) a 3D whole-heart cardiac MRI with isotropic resolution 0.7×0.7×  0.7mm3 and (ii) a 4-stack torso MRI with resolution 1.3×1.3×  3.0mm3. Both scans were taken at 3T (Magnetom Skyra, Siemens Healthcare, Erlangen, Germany). Before MRI acquisition, the 12-lead ECG was acquired using MRI-compatible electrodes that were left intact during acquisition. The clinical 12-lead ECG had been filtered with a 150 Hz low pass filter, a 50 Hz bandstop filter, and a high pass filter of 0.05 Hz. The MRI study was approved by the ethical review board of the Medical University of Graz (EKNr: 24–126 ex 11/12). The subject gave written informed consent.

Construction of the patient-specific model

To generate a patient-specific anatomical model, automatic segmentation of the cardiac MRI was performed using a convolutional neural network.38 The heart segmentation was corrected manually in open-source Seg3D with the assistance of algorithms predominately based on intensity thresholds.39 This included removing disconnected tissue, closing any holes, and making sure all surfaces intersected properly. The torso and lungs were also segmented from the torso MRI stack using semi-automatic approaches based on thresholds and registered with the heart using a built-in iterative closest point algorithm within Seg3D (see Tools/Point Set Registration).40 From the joint segmentation, a finite-element volumetric mesh with tetrahedral elements of approximately 900μm resolution within the heart and up to 4500μm in the torso and lungs was generated using Tarantula41(CAE Software Solutions, Eggenburg, Austria). The resolution is reported as average edge length across the mesh. The model was subsequently equipped with rule-based myocardial fibres generated in both the ventricles42 and the atria,43 separately. To facilitate automated control and variation of electrophysiological parameters, such as the location of the AP, universal cardiac coordinates were generated for the ventricles44,45 and the atria.46 An electrical isolation layer between the atria and the ventricles was enforced using nodal splitting. The RV outflow tract was assumed to be electrically inert. Electrode locations of the 12-lead ECG were recovered on the torso surface from the location of the MRI-compatible electrodes.

Aspects of the cardiac electrophysiology of the patient-specific model were subsequently personalized to generate a 12-lead ECG replicating the measured beat-averaged 12-lead sinus rhythm ECG.36 Within the atria, the anatomical features of the sino-atrial node, Bachmann’s bundle, and the foramen ovale within the atria were placed to facilitate faster inter-atrial conduction per the measured P-wave duration and morphology.36 Generic parameters were used for atrial electrophysiology, and no calibration was performed. The timings and locations of the primary root sites of the five fascicles of the His–Purkinje system47,48 facilitating ventricular activation were personalized using a refined sampling procedure.45 The optimized sites and timings were then replaced with a physiological His–Purkinje System.47 Repolarization gradients were prescribed using a linear mapping between activation and action potential duration as detailed in36 leading to a realistic T-wave. The personalized signal resulted in an average L2-norm difference of 0.08 under normal sinus rhythm as reported in previous work.36 Full details of the model and other electrophysiological parameter settings are available in.45,47,36

Representation of the antegrade accessory pathway

A short atrioventricular bypass tract allowing antegrade conduction was added into the patient-specific model to replicate ventricular pre-excitation caused by WPW. Corresponding insertion and exit points of the AP were defined using anatomically intuitive universal cardiac coordinates, respectively. Exact localization of the sites into nodes existing on the mesh was facilitated through open-source meshtool that relies on a custom-built K-Dimensional tree algorithm.49 A geodesic path was then found between each corresponding insertion and exit site to determine the length of the pathway d. The cable of the pathway was prescribed a given conduction velocity CVAP. Given this representation, a parameter vector ωAP consisting of a total of seven parameters can be used to define the antegrade AP leading to pre-excitation within the ventricles:

(1)

where xatria and xven are defined with universal cardiac coordinates and CVAP is the conduction velocity of the AP. The sites xven and xatira were localized to the closest node in the mesh using meshtool. The prescribed timing of the pre-excitation in the ventricles (tven) could, therefore, be computed from the activation time at the atrial insertion site plus the time required to traverse the pathway.

where tatria is the time at which the excitation wavefront reaches xatria. For full details on the construction of the pathways, please refer to Supplementary material online, S1.

Sampling schematic

A synthetic ECG database was generated by varying AP locations across both ventricles. All APs were assumed to propagate activation from the atria to the ventricles leading to ventricular pre-excitation. The RV outflow tract was taken to be inert and thus APs were neglected within this area. Latin hypercube sampling was used to generate the sampling schematic for xatria and xven defined in Eq. (1). The maximal allowed distance apart was 6cm to ensure a uniform sampling of xven across the top of the ventricles in the upper two apicobasal regions in the bullseye plot. A conduction velocity in the accessory pathway (CVAP) of 2.0ms1 was set to replicate Purkinje fibres in agreement with electrophysiological studies.50 A final total of 4678 APs in the LV and 4593 in the RV were thus actually sampled across the atrioventricular junction region. The database was split into LV and RV datasets for subsequent processing. For full details on the sampling schematic, please refer to the Supplementary material online, S2.

Cardiac simulation

All electrophysiological simulations during both normal sinus rhythm and in the presence of AP were run with CARPentry51 using 16 cores on a desktop machine. All simulations were run for 700m s. Cardiac sources were represented using the reaction-Eikonal method in monodomain mode without diffusion to recover membrane voltages within the heart.52 Note that use of an Eikonal-based approach allows for simulating pre-ventricular activation from a single node in the mesh, i.e. for xven, without a source-sink mismatch that could occur in traditional reaction-diffusion approaches.

Electrical potentials at electrodes with known positions on the torso surface from the MRI were acquired using lead field projection of the simulated membrane voltages.53 The 12-lead ECG was then constructed using the traditional combination of the electrical potentials.54 The cardiac simulation framework is described in detail in Gillette et al.45 and utilized previously in Gillette et al.35 All simulated 12-lead ECGs were filtered with a 150 Hz low pass filter according to clinical settings used during clinical ECG acquisition. A global scaling factor of 0.32 was previously computed to minimize the L2 norm between the synthetic personalized 12-lead ECG and the measured one during normal sinus rhythm in previous work36 and was applied to all simulated 12-lead ECGs.

Regional separation of the 12-lead ECGs

The sources simulated 12-lead ECGs were regionally grouped according to a standard 17-segment bullseye plot within the LV that was extrapolated to the RV (Figure 2B). For exact details on the construction of the bullseye plot using universal cardiac coordinates, we refer to Supplementary material online, S3. Within each region, the mean signal and the mean signal plus or minus two standard deviations were computed for all 12-lead ECGs within that particular region to demonstrate overall signal variation.

(A) Sampling sites were selected across the upper sections of the ventricles within the timings used to construct the 12-lead ECG Database. (B) Sampled sites are organized into segments on the bullseye plot. (C) The two septal APs corresponding to the exemplary simulations are located on either side of the septum, but in the area of the His–Purkinje System. AP, accessory pathway; ECG, electrocardiogram.
Figure 2.

(A) Sampling sites were selected across the upper sections of the ventricles within the timings used to construct the 12-lead ECG Database. (B) Sampled sites are organized into segments on the bullseye plot. (C) The two septal APs corresponding to the exemplary simulations are located on either side of the septum, but in the area of the His–Purkinje System. AP, accessory pathway; ECG, electrocardiogram.

Sensitivity analysis

Quantifying the sensitivity of the 12-lead ECGs from the database with respect to the four different input parameters of ρ, z, ϕ, and tven in ωAP was performed by employing PCE29 in each ventricle separately. We applied the Python toolbox ‘PyThia UQ’ that determines PCE expansion coefficients via a multi-linear least-squares regression.55

To compute the time-dependent sensitivity S^(t) of each parameter in ωAP, we multiply the obtained Sobol indices S(t)i—where i stands for a multi-index numbering all possible parameter combinations—with the standard deviation of the signal σ(t) at each point in time:

(2)

From that, the respective integrated sensitivities, as a measure of overall parameter importance, are obtained by integrating S^(t) over the signal length:

(3)

The integrated sensitivities are given in [mVs], i.e. measurable signal change in mV over the signal length, and best reflect which parameter (combinations) introduce measurable change in the 12-lead ECG signal. By normalizing them, i.e. by dividing through the time-integrated σ=σ(t)dt, one obtains relative proportions of parameter importance Si. Per definition, the sum of all Si is 1, similar to the Sobol coefficients.

(4)

To gauge the surrogate reconstruction error, we reconstructed a random number of ECGs using the PCE with parameters used to generate ECGs in the synthetic ECG database. The L2 norm was then computed between each ECG pair for comparison, and the mean and standard deviation were computed for each ventricle. For full details on the sensitivity analysis, please refer to Supplementary material online, S4.

ECG genesis in septal accessory pathways

Three simulations were selected to exemplify the genesis of the 12-lead ECG in right-sided and left-sided septal accessory pathways. The first simulation was the personalized sinus rhythm generated by the patient-specific model of the single subject. Simulations of two APs in the septum were then selected from the 12-lead ECG database. The two septal APs had similar insertion sites within the atria but exit sites on opposed sides of the septum to modulate left-sided and right-sided pathways. Both exit sites were located in close proximity to the existing His–Purkinje system, but not directly connected. The ventricular activation time in both APs was set to be a value of 125ms from the onset of the sino-atrial node activation facilitating a realistic PR interval. The parameters for the two exit sites thus become:

(5)
(6)

For all three simulations, the membrane voltages, electrical potentials, and 12-lead ECGs were computed using the simulation framework described in Cardiac simulation section run in pseudo-bidomain mode. A body-surface potential map could be constructed from the electrical potentials. The minimum dV/dt was computed in the downstroke of the S wave for lead V2 for both septal APs.

Results

An automatic workflow to generate a whole-heart patient-specific model of electrophysiology and incorporate an AP with limited manual user interaction could be constructed from non-invasive clinical data (Figure 1). Construction and localization of APs required approximately 1 h. Subsequent generation of the synthetic ECG database of 9,271 signals of left-sided and right-sided APs required under 3d using 16 cores on a desktop machine.

Regional variation in the 12-lead ECG database

A database was generated by varying AP locations across both ventricles (see Figure 2). Sampled sites and their corresponding 12-lead ECGs were organized into regional assignments according to the bullseye plot segmentation (Figure 2B). Locations of APs evenly cover the atrioventricular junction region on the ventricular side (Figure 2A) in the upper two apicobasal regions such that the number of regional groupings could be reduced to only 12 within the RV and LV each. The computed tven were within the range of 58 to 195ms.

Expected morphological features are present within the 12-lead ECGs for the single subject for both right-sided and left-sided APs. Namely, within right-sided APs (Figure 3), the polarity of the V1 lead is predominately negative and within the LV, there is less consistency pertaining to the polarity of the V1 precordial lead. Sites within the anterior free wall of the LV exhibit morphological similarities across multiple rotational regions.

Regional variation in 12-lead ECG morphology stemming from right-sides APs within the 12 regions in the RV (columns). Insertion sites xven for each AP are indicated within the corresponding region of the bullseye plot (right). The mean signal is black, and the coloured signals represent the mean ± 2 SD. AP, accessory pathway; ECG, electrocardiogram; RV, right ventricle.
Figure 3.

Regional variation in 12-lead ECG morphology stemming from right-sides APs within the 12 regions in the RV (columns). Insertion sites xven for each AP are indicated within the corresponding region of the bullseye plot (right). The mean signal is black, and the coloured signals represent the mean ± 2 SD. AP, accessory pathway; ECG, electrocardiogram; RV, right ventricle.

Sensitivity analysis

The relative errors for the performed PCEs were 16.8%±2.7% for the LV dataset and 16.3%±3.2% for the RV dataset averaged over all leads of the 12-lead ECG. Following Winkler et al.30, we use the relative error with respect to the signal variance over the whole dataset to quantify the convergence of the PCE surrogate on the chosen parameter interval.30

In terms of measurable changes in signal amplitude (Figure 5B,C), the statistical analysis shows that large measurable changes can be best observed in lead V2 for the LV and the RV datasets due to the high signal variation caused by adding an AP. For the single parameter variations, the ventricular activation time tven and the rotational placement ϕ are important, whereas ρ and z result in no significant signal change. For the second-order variations, the combined contributions of (tven,ϕ) and (tven,ρ) give the biggest contributions. More specifically for LV data, the (tven,ρ) induced variations that are largest for V2–V5. The joint contribution of (tven,ϕ) induced changes the leads aVL and the inferior leads II, aVF, and III. For the RV data, (tven,ρ) gives the strongest sensitivity for lead V2 and smaller values for V3, V4, and V5. The combined variation of (tven,ϕ) causes strong changes in all leads of the RV dataset, except for lead I and -aVR.

Regarding the relative sensitivities Si (Figure 5D), second-order sensitivities dominate. Within left-sided APs, highest second-order sensitives are seen in leads aVL, aVF, and III stemming from the contribution (tven,ϕ). First-order sensitivities are observed in leads -aVR, V2, and V4 to V6 due to variations in excitation time (tven) in the LV. Within right-sided APs, all leads except V2 to V5 are highly sensitive to the joint contribution of (tven,ϕ). Minor second-order sensitivities are observed in leads V4 and II within the LV from the joint-contribution of time and transmural depth (tven,ρ). Some second-order sensitives are also observed in RV for V2 and V3 depending on (tven,ρ). With the exception of -aVR, the rotational placement of the AP influences all limb and Goldberger within the RV.

ECG genesis in septal accessory pathways

The locations of the right-sided and left-sided septal antegrade APs can be visualized in Figure 2C. The downstroke of the S wave in V2 is steeper in the right-sided AP with a minimum dV/dt of 0.37 in comparison to 0.23 in the LV septal AP.

The 12-lead ECGs generated from septal APs have very similar morphological features (Figure 6) that result in both APs being localized to the RV by most existing diagnostic trees. Most importantly, the signals both have a negative (or biphasic) V1 deflection and entirely positive inferior leads (leads II, aVF, and III). Both signals have similar R-wave progression, with the most positive delta wave occurring in lead II. Differences include a longer PR interval in a right-sided AP that can be contributed to the delayed capture of the left bundle of the His–Purkinje System. Noticeable amplitude and morphological differences are observed in leads V1 and V2. The left-sided AP has a weaker S-wave deflection in V1 through V3 and a rSr’ or M shape within V1. The right-sided AP exhibits an rS morphology in V1 and V2, as well as more positive inferior leads. The precordial QRS transition, defined as the occurrence of the R-wave exceeding the S-wave amplitude, occurs in V2 and V3 for the LV and RV sites, respectively. Both sites are located in the fourth regional grouping as displayed in Figure 4.

Regional variation in 12-lead ECG morphology resulting from left-sided APs within the 12 regions in the LV (columns). Insertion sites xven for each AP are indicated within the corresponding region of the bullseye plot (right). The mean signal is black, and the coloured signals represent the mean plus or minus two standard deviations. AP, accessory pathway; ECG, electrocardiogram; LV, left ventricle.
Figure 4.

Regional variation in 12-lead ECG morphology resulting from left-sided APs within the 12 regions in the LV (columns). Insertion sites xven for each AP are indicated within the corresponding region of the bullseye plot (right). The mean signal is black, and the coloured signals represent the mean plus or minus two standard deviations. AP, accessory pathway; ECG, electrocardiogram; LV, left ventricle.

Sensitivity analysis for APs within the RV and LV. (A) Mean and standard deviation of the 12-lead ECGs stemming from APs within the LV (left) and RV free wall (right). Time for the ECG is not shown, but lasts 700ms. (B) Measure of overall parameter importance in terms of measurable changes in ECG amplitude by integrating S^(t) over the duration of the signal. The normalized total sum across all sensitivities for each lead is provided underneath. Time-dependent sensitivity S^(t) of each parameter is not directly shown. (C) Relative proportions of parameter importance Si. AP, accessory pathway; ECG, electrocardiogram; LV, left ventricle; RV, right ventricle.
Figure 5.

Sensitivity analysis for APs within the RV and LV. (A) Mean and standard deviation of the 12-lead ECGs stemming from APs within the LV (left) and RV free wall (right). Time for the ECG is not shown, but lasts 700ms. (B) Measure of overall parameter importance in terms of measurable changes in ECG amplitude by integrating S^(t) over the duration of the signal. The normalized total sum across all sensitivities for each lead is provided underneath. Time-dependent sensitivity S^(t) of each parameter is not directly shown. (C) Relative proportions of parameter importance Si. AP, accessory pathway; ECG, electrocardiogram; LV, left ventricle; RV, right ventricle.

Cardiac activation during normal sinus rhythm (left, red) compared with right-sided AP (center, green) and left-sided APs (right, blue) that are activated at tven=125.0ms and located on opposites sides of the septum at similar apicobasal depth. For all cases, the membrane voltages are shown within the heart and the electrical potentials are displayed on the torso surface at various time points through the cardiac cycle. Locations of the AP can be seen in Figure 2C. The video representation of this figure is provided in Supplementary material online, S5. AP, accessory pathway.
Figure 6.

Cardiac activation during normal sinus rhythm (left, red) compared with right-sided AP (center, green) and left-sided APs (right, blue) that are activated at tven=125.0ms and located on opposites sides of the septum at similar apicobasal depth. For all cases, the membrane voltages are shown within the heart and the electrical potentials are displayed on the torso surface at various time points through the cardiac cycle. Locations of the AP can be seen in Figure 2C. The video representation of this figure is provided in Supplementary material online, S5. AP, accessory pathway.

Both common algorithms of Pambrun et al.12 and El Hamriti et al.11 were unable to properly locate an AP inserting on the endocardial LV septum within the patient-specific model of WPW as demonstrated through exemplary simulations of two different APs. Within the diagnostic tree from Pambrun et al.,12 the right-sided septal site would be localized to the appropriate anterior RV septum (NH) due to three positive precordial leads, a negative V1, and a positive V3. The left-sided AP would be localized to the RA for similar morphological reasons in combination with a negative V3. According to Easy-WPW by El Hamriti et al.,11 both AP sites would be localized to the anteroseptal within the RV. This stems from the negative V1, a precordial QRS transition observed starting in V3, and the most positive delta wave occurring in lead II.

Discussion

We present a detailed patient-specific model of cardiac electrophysiology that has been adapted to represent an antegrade AP in the form of a short atrioventricular bypass tract in WPW syndrome generated within an efficient workflow. The virtual model was used to provide mechanistic insights into the morphological genesis of the 12-lead ECG for improving both sensitivity and specificity in clinically deployed ECG-based algorithms. Improvements in ECG-based algorithms allow for more accurate non-invasive AP localization to avoid the need for cardiac imaging or electrophysiological studies during the ablation procedure. The efficient workflow for generating a patient-specific model of WPW also has the potential to become a clinically viable approach for personalized care for patients with complex disease representations. Patient-specific models could facilitate tailored ablation strategies, assess the risk of sudden cardiac death, or guide management plans for asymptomatic patients.3,56

Analysis of ECG-based localization algorithms

Performing sensitivity analysis on virtual models provides a novel means to obtain insight into the morphological genesis of the 12-lead ECG within WPW in high fidelity. Signal-derived SA highlights the locations in the 12-lead ECG that depend on parameters relating to the AP placement and timing (Sensitivity analysis section). First, the joint contribution from the rotational placement and activation time of the AP creates the strongest measurable change in the V2 precordial lead for sites located within the RV free wall. Measurable amplitude change in V2 for all APs is also tied to ventricular activation time of the site (tven). When considering the general placement of V2 in proximity to the heart, delayed activation of the LV free wall in comparison to the RV as seen in right-sided APs leads to a generally larger slurred S wave in V2 (Figure 3). This is further exaggerated when the AP is located in the RV free wall. Inversely, a more positive S deflection stems from sites located in the LV free wall (Figure 4). Second, both left- and right-sided APs induce larger morphological changes in the inferior leads (II, aVF, and III) in both left- and right-sided APs (Figure 5). This stems from the AP-triggered ventricular dyssynchrony that causes a slight shift in the axis of the heart leading to a shift in the QRS transition and R-wave progression. Most standard ECG localization algorithms thus expectantly encode the electrical heart-axis observed in these leads in some way; El Hamriti et al.11 use the lead with the most positive delta wave and precordial QRS transition and12 relies on the number of positive leads.

Regarding the resolution of current ECG algorithms, El Hamriti et al.11 states that the use of 10 regions is sufficient to differentiate between characteristics within the 12-lead ECG, and the use of further regions complicates clinical decision-making. Our computational modelling supports this. The majority of signal variation (Figures 3 and 4) can be observed within less than 10 rotational regions within the LV and RV combined. Additionally, as the insertion depth plays little role in signal morphology as indicated in Figure 5, a pure rotational-based regional separation in an ECG algorithm may suffice. Some anatomical regions present in ECG algorithms, such as the deep coronary sinus,12 were not accounted for within the patient-specific model of WPW.

As indicated by the example simulations of APs located on either side of the septum (ECG genesis in septal accessory pathways section), differentiating between a bundle branch block pattern in leads V1 and V2, instead of relying on the polarity of the V1 lead alone, may improve differentiation between left- and right-sided APs. Within the right-sided AP, the RV activates slightly prior to the LV even with an intact His–Purkinje system. The right-sided pathway thus exhibited a stronger S-wave deflection within V1 and V2 (Figure 6). This corresponds with a slight left bundle branch block pattern. A later precordial QRS transition is also observed in the right-sided site, which is indicative of left bundle branch block. Inversely, the left-sided AP experienced a rS shape in both V1 and V2 stemming from later activation of the RV. These morphological traits are more characteristic of right bundle branch block pattern. Algorithms for localizing idiopathic ventricular tachycardia typically rely on similar criteria and principles.57 Further differentiation could be done by looking at the downstroke of the S wave in lead V2, where right-sided APs will have a stronger S wave due to delayed activation of the left ventricle.

Various algorithms based on machine learning also show promise for the purpose of automatic localization,58,59 but suffer from similar limitations and lack of sufficient clinical data for training. Supplementing training data for such approaches with synthetically trained 12-lead ECG data may be useful in training algorithms. Our 12-lead ECG database on WPW is the first publicly available database of its kind that could be utilized towards such an end. The synthetic ECG database generated for this study can be found on Zenodo with DOI 10.5281/zenodo.10949804.

Towards the aim of generating even larger synthetic ECG databases for machine learning, the UQ approach relies on constructing a surrogate model that allows for instantaneous simulation. The underlying PCE-based surrogate model in this study, however, is likely not suitable to replace the forward simulations with sufficient fidelity given the high signal-averaged relative errors of 16.8%±2.7% for the LV and 16.3%±3.2% for the RV dataset. Surrogate models generally converge worse with an increasing number of parameters, larger parameter ranges, and substantial signal variance caused by specific parameter changes. These large errors reflect the complex dependency of the ECG on the parameter variations and the possible non-uniqueness of the problem. Performing SA using predetermined ECG-derived metrics such as V1 polarity may result in better approximations, and reduce errors. While the estimated surrogate errors are still sufficient for a meaningful SA, different PCE terms used could lead to slightly different sensitivities.

Patient-specific models of WPW syndrome

Accurate prediction of patient-specific AP localization using patient-specific models could aid decision-making regarding initiation of medical treatment,60 especially in currently asymptomatic patients,56 but also regarding indication of an electrophysiological study. Given real patient data, the model could be personalized to the 12-lead ECG of a patient with WPW to quantify the most probable location of an AP to reduce the need for costly electrophysiological studies during an intervention. This would amount to solving a specific ECG inverse problem as seen in electrocardiographic imaging.60 Note that the achieved accuracy is adequate for identifying the major parameter sensitivities, however, it is insufficient for employing the surrogate in solving the inverse problem. Challenges in model personalization must first be overcome as later discussed, but recent efforts in this direction have been made.61,36,62

Uncertainties in patient-specific modelling arising during both anatomical model generation and calibration may influence simulation predictions as well as clinical outcomes.63 In terms of anatomy, the patient-specific model in this work was generated from MRIs that are not standard clinically and changes in imaging modality influence the accuracy of the anatomical model.64 The anatomical model generation pipeline has been extended to work on various other imaging data including standard cardiac MRI and computed tomography.65 The MRIs were registered together resulting in an error on the order of 1 or 2cm stemming from breathing displacement, segmentation errors, and the registration itself. Calibration errors must also be considered when performing patient-specific modelling and may influence outcomes. Sources include generically calibrated parameter values (as seen in the atria in the patient-specific model), assumptions in the biophysical model that cause deviations from reality, and the non-uniqueness of the inverse problem. Further validation is needed to explore the influence of anatomical model generation and calibration.

Testing the viability of using the patient-specific models of WPW to do pre-procedural planning could be carried out within the context of a clinical study. While the 12-lead ECG results support a valid representation of WPW syndrome, model evaluation on real clinical data has not yet been performed. The current model was parameterized under sinus rhythm and then extended for WPW syndrome. Cardiac imaging in these patients of any imaging modality is however not standard and thus it is difficult to build patient-specific models. Furthermore, clinical datasets consisting of electro-anatomical mapping that are needed for validation are not recorded except under the circumstance of a co-existing condition necessitating additional treatment. Even when invasive procedural data is available, it is typically only available as an electro-anatomical map on a limited endocardial manifold on either side of the atrioventricular junction. To assess model credibility for clinical use, SA may also be useful.66

Translational impact

The computational modelling platform we have developed has significant translational impact. In general, personalized models in the form of cardiac digital twins or patient-specific models that accurately reflect individual patients in terms of both cardiac anatomy and electrophysiology enable highly personalized treatment planning, surgical planning, and risk stratification.21 This personalized approach can also be extended to other arrhythmias and structural heart diseases with simple extensions, enhancing the precision and efficacy of interventions. This patient-specific model closely resembles an idiopathic ventricular tachycardia and can be used to help train algorithms for localization of ectopic or focal beats stemming from the ventricles.67,60 Extensions to a cardiac solver allowing re-entry would be required for modelling atrial fibrillation and ventricular tachycardia that coincide with WPW. Many patient-specific models have also been developed for assisting in ablation and treatment protocols68,22 and these conditions often coincide with WPW.5 In patients with heart failure, our patient-specific models could simulate the effects of cardiac re-synchronization therapy and optimize lead placement for improved therapeutic outcomes.69,70 Patient-specific models are also particularly valuable for conditions like hypertrophic cardiomyopathy that can occur with familial WPW, where the risk of sudden cardiac death can be assessed more accurately through personalized simulations. By simulating various clinical scenarios, platforms for generating patient-specific models in a cohort can be used for in silico clinical trials62 and stratifying risk profiles. This can also aid in pre-surgical planning and postoperative management.

Other broader potential benefits span across education and training, research and development, pharmacological studies, and public health and policy. In medical education and training, the platform can serve as an advanced educational tool for medical students, residents, and fellows as it provides a dynamic way to learn about cardiac electrophysiology, arrhythmias, and the impact of various interventions. For electrophysiologists and cardiologists, the platform offers a sophisticated simulation-based training environment. Practitioners can hone their skills in identifying APs, interpreting ECGs, and planning ablation procedures without risk to patients. Towards pharmacological studies, in silico models can accelerate drug development and assess potential anti-arrhythmic therapies. The platform can also be extended to study the effects of various pharmacological agents on cardiac electrophysiology in a controlled in silico environment.71 This would warrant including complex cellular dynamics between drugs and various ion channels.72 By aggregating data from multiple patient-specific simulations, our platform can contribute to large-scale epidemiological studies, providing insights into the prevalence and management of various cardiac conditions. The evidence generated through simulations can also inform health policy decisions, particularly in optimizing resource allocation for cardiac care and improving guidelines for arrhythmia management.

Future work and limitations

All the clinical applications of AP localization prediction become even more important when dealing with paediatric patients. Although most algorithms to predict AP localization were derived from adult patients, the paediatric population carries the highest burden of WPW syndrome with age-related peaks of arrhythmias appearing during infancy and adolescence. Incorporation of paediatric ECG data on a whole-heart model of a paediatric patient suffering from WPW syndrome may allow for more accurate algorithms in the future. This study, however, was performed within a single patient-specific model of a healthy adult male who was equipped with an antegrade AP to represent WPW. The methodology should be extended to a cohort of children and adolescents, the majority of the clinical population, as changes in the heart axis due to age may influence both results and the efficacy of such algorithms in children.11 As acquisition of cardiac images in patients with WPW is not clinically standard thus limiting cohort generation, constructing shape-models or atlases of paediatric73 or adult74 hearts from a limited number of scans may be a viable option for larger population studies on WPW and ECG localization algorithms.These models have been also paired together with SA.28

Assessment of how variations in AP conduction and placement impact a clinical ECG diagnostic method is needed. We assumed a maximal AP length of d=6cm of the sites in order to investigate the role of AP placement and excitation across the top portion of the ventricles with no restrictions on atrial placement. In future work, further physiological restrictions in universal cardiac coordinates on both xven and xatria defining the APs can be done to explore specific types of pathways of varying lengths and types.75,50 Furthermore, the conduction velocity in the AP (CVAP) was prescribed to be fairly fast at 2.0ms1 based on reported electrophysiological properties and behaviour of atriofascicular pathways in patients with pre-excitation syndromes and Mahaim fibres.50 Conduction may be slower in APs that are located in different areas of the heart.76 Variations in conduction velocity, and thus tven, should therefore be explored. Multiple APs with different conduction velocities could also be considered.75,19

While not available at the time of the study, re-entrant simulations would facilitate modelling APs allowing typical atrioventricular conduction, but retrograde propagation through the atrioventricular junction region, as well as simulations of multiple heartbeats. Re-entrant circuits would necessitate accounting for restitution in conduction velocity and action potential duration stemming from rapid cycle lengths. We reported a technique for tuning these properties in previous work.77 The patient-specific model could then be used to explore disease mechanisms and morphological changes in the 12-lead ECG when WPW is in conjunction with other cardiac diseases and manifestations such as atrial fibrillation, bundle branch blocks, ventricular tachycardia.1,5

The ECG database could be generated in under 3 days, but using simpler models or alternative sampling schematics could decrease these computational costs. Simpler and lighter patient-specific models without full representations of the HPS may be capable of capturing similar morphological elements on the ECG as indicated in previous work47 and can thus be useful in understanding basic mechanisms of the disease or even directly for pre-excitation localization as shown in Berger et al.60 using electrocardiographic imaging.60 They may thus be more efficient for population studies without a loss of fidelity. Further investigation into model differences on ECG morphology and underlying mechanisms would be needed. A more informed sampling of APs could also decrease the ECG database generation, as it is known that APs are not distributed uniformly across the atrioventricular plane.76 Alternative sampling approaches may be more efficient than latin-hyper-cub sampling, but a speed-up was not observed in similar work using Sobol sampling.30 For procedural planning in individual patients, a highly detailed patient-specific model with a full representation of the HPS is warranted especially in the instance of compounding conditions or retrograde APs.

Supplementary material

Supplementary material is available at Europace online.

Authors' contributions

K.G. both constructed and parameterized the virtual cardiac model of electrophysiology. She then performed all cardiac simulations including both validation and synthetic ECG database generation. She was also involved in all aspects relating to manuscript drafting and revision. B.W. conducted SA on the synthetic 12-lead ECG database via PCE and performed analysis on the results. He was involved in manuscript drafting and revision. S.K. provided clinical guidance and direction throughout the project. He also facilitated clinical data acquisition and was involved in manuscript drafting and revision. D.S. was involved in manuscript drafting and revision. E.V. helped in construction of the computational modelling framework. He was involved in manuscript drafting and revision. M.B. was involved in manuscript drafting and revision. G.P. was involved in manuscript drafting and revision.

Funding

This research was funded in whole or in part by the Austrian Science Fund (FWF) [grant 10.55776/ESP592] awarded to K.G. This research was also supported by funds received from BioTechMed-Graz under the ILearnHeart Project and under grant I2760-B30 from the Austrian Science Fund awarded to G.P. This project conducted under MedalCare 18HLT07 has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme. For the purpose of open access, the author has applied a CC BY public copyright licence to any Author Accepted Manuscript version arising from this submission.

Data availability

The synthetic ECG database of a specific type of WPW syndrome generated within this study can be found on Zenodo with DOI 10.5281/zenodo.10949804. The model and clinical data used within this study are not publicly available. Many tools used to generate the pipeline are available in the openCARP framework37,49. All other tools require specific collaboration agreements.

References

1

Lu
 
C-W
,
Wu
 
M-H
,
Chen
 
H-C
,
Kao
 
F-Y
,
Huang
 
S-K
.
Epidemiological profile of Wolff–Parkinson–White syndrome in a general population younger than 50 years of age in an era of radiofrequency catheter ablation
.
Int J Cardiol
 
2014
;
174
:
530
534
.

2

Pærregaard
 
MM
,
Hartmann
 
J
,
Sillesen
 
A-S
,
Pihl
 
C
,
Dannesbo
 
S
,
Kock
 
TO
, et al.
The Wolff–Parkinson–White pattern in neonates: results from a large population-based cohort study
.
Europace
 
2023
;
25
:
euad165
.

3

Paul
 
T
,
Krause
 
U
,
Sanatani
 
S
,
Etheridge
 
SP
.
Advancing the science of management of arrhythmic disease in children and adult congenital heart disease patients within the last 25 years
.
Europace
 
2023
;
25
:
euad155
.

4

Sacher
 
F
,
Wright
 
M
,
Tedrow
 
UB
,
O’Neill
 
MD
,
Jais
 
P
,
Hocini
 
M
, et al.
Wolff–Parkinson–White ablation after a prior failure: a 7-year multicentre experience
.
Europace
 
2010
;
12
:
835
841
.

5

Borregaard
 
R
,
Lukac
 
P
,
Gerdes
 
C
,
Møller
 
D
,
Mortensen
 
PT
,
Pedersen
 
L
, et al.
Radiofrequency ablation of accessory pathways in patients with the Wolff–Parkinson–White syndrome: the long-term mortality and risk of atrial fibrillation
.
EP Europace
 
2015
;
17
:
117
122
.

6

Pappone
 
C
,
Vicedomini
 
G
,
Manguso
 
F
,
Saviano
 
M
,
Baldi
 
M
,
Pappone
 
A
, et al.
Wolff-Parkinson-White syndrome in the era of catheter ablation: insights from a registry study of 2169 patients
.
Circulation
 
2014
;
130
:
811
819
.

7

Sherdia
 
AFIA
,
Abdelaal
 
SA
,
Hasan
 
MT
,
Elsayed
 
E
,
Mare’y
 
M
,
Nawar
 
AA
.
The success rate of radiofrequency catheter ablation in Wolff-Parkinson- syndrome patients: a systematic review and meta-analysis
.
Indian Heart Journal
 
2023
;
75
:
98
107
.

8

Sternick
 
EB
,
Sanchez-Quintana
 
D
,
Wellens
 
HJ
,
Anderson
 
RH
.
Mahaim revisited
.
Arrhythm Electrophysiol Rev
 
2022
;
11
:
e14
.

9

Macedo
 
PG
,
Patel
 
SM
,
Bisco
 
SE
,
Asirvatham
 
SJ
.
Septal accessory pathway: anatomy, causes for difficulty, and an approach to ablation
.
Indian Pacing Electrophysiol J
 
2010
;
10
:
292
.

10

Arruda
 
MS
,
McClelland
 
JH
,
Wang
 
X
,
Beckman
 
KJ
,
Widman
 
LE
,
Gonzalez
 
MD
, et al.
Development and validation of an ECG algorithm for identifying accessory pathway ablation site in Wolff-Parkinson-White syndrome
.
J Cardiovasc Electrophysiol
 
1998
;
9
:
2
12
.

11

El Hamriti
 
M
,
Braun
 
M
,
Molatta
 
S
,
Imnadze
 
G
,
Khalaph
 
M
,
Lucas
 
P
, et al.
EASY-WPW: a novel ECG-algorithm for easy and reliable localization of manifest accessory pathways in children and adults
.
Europace
 
2023
;
25
:
600
609
.

12

Pambrun
 
T
,
El Bouazzaoui
 
R
,
Combes
 
N
,
Combes
 
S
,
Sousa
 
P
,
Le Bloa
 
M
, et al.
Maximal pre-excitation based algorithm for localization of manifest accessory pathways in adults
.
JACC Clin Electrophysiol
 
2018
;
4
:
1052
1061
.

13

Milstein
 
S
,
Sharma
 
AD
,
Guiraudon
 
GM
,
Klein
 
GJ
.
An algorithm for the electrocardiographic localization of accessory pathways in the Wolff-Parkinson-White syndrome
.
Pacing Clin Electrophysiol
 
1987
;
10
:
555
563
.

14

d’Avila
 
A
,
Brugada
 
J
,
Skeberis
 
V
,
Andries
 
E
,
Sosa
 
E
,
Brugada
 
P
.
A fast and reliable algorithm to localize accessory pathways based on the polarity of the QRS complex on the surface ECG during sinus rhythm
.
Pacing Clin Electrophysiol
 
1995
;
18
:
1615
1627
.

15

Fitzpatrick
 
AP
,
Gonzales
 
RP
,
Lesh
 
MD
,
Lee
 
RJ
,
Scheinman
 
MM
.
New algorithm for the localization of accessory atrioventricular connections using a baseline electrocardiogram
.
J Am Coll Cardiol
 
1994
;
23
:
107
116
.

16

Haissaguerre
 
M
,
Marcus
 
F
,
Poquet
 
F
,
Gencel
 
L
,
Le Metayer
 
P
,
Clementy
 
J
.
Electrocardiographic characteristics and catheter ablation of parahissian accessory pathways
.
Circulation
 
1994
;
90
:
1124
1128
.

17

Fananapazir
 
L
,
German
 
L
,
Gallagher
 
J
,
Lowe
 
J
,
Prystowsky
 
E
.
Importance of preexcited QRS morphology during induced atrial fibrillation to the diagnosis and localization of multiple accessory pathways
.
Circulation
 
1990
;
81
:
578
585
.

18

Kurath-Koller
 
S
,
Manninger
 
M
,
Öffl
 
N
,
Köstenberger
 
M
,
Sallmon
 
H
,
Will
 
J
, et al.
Accuracy of algorithms predicting accessory pathway localization in pediatric patients with Wolff-Parkinson-White syndrome
.
Children
 
2022
;
9
:
1962
.

19

Fujino
 
T
,
De Ruvo
 
E
,
Grieco
 
D
,
Scará
 
A
,
Borrelli
 
A
,
De Luca
 
L
, et al.
Clinical characteristics of challenging catheter ablation procedures in patients with WPW syndrome: a 10 year single-center experience
.
J Cardiol
 
2020
;
76
:
420
426
.

20

Niederer
 
SA
,
Lumens
 
J
,
Trayanova
 
NA
.
Computational models in cardiology
.
Nat Rev Cardiol
 
2019
;
16
:
100
111
.

21

Corral-Acero
 
J
,
Margara
 
F
,
Marciniak
 
M
,
Rodero
 
C
,
Loncaric
 
F
,
Feng
 
Y
, et al.
The ‘digital twin’ to enable the vision of precision cardiology
.
Eur Heart J
 
2020
;
41
:
4556
4564
.

22

Dasí
 
A
,
Nagel
 
C
,
Pope
 
MT
,
Wijesurendra
 
RS
,
Betts
 
TR
,
Sachetto
 
R
.
In silico trials guide optimal stratification of atrial fibrillation patients to catheter ablation and pharmacological medication: the i-stratification study
.
EP Europace
 
2024
;
26
:
euae150
.

23

Zeile
 
C
,
Scholz
 
E
,
Sager
 
S
.
A simplified 2D heart model of the Wolff-Parkinson-White syndrome
.
IFAC-PapersOnLine
 
2016
;
49
:
26
31
.

24

Wei
 
D
,
Yamada
 
G
,
Musha
 
T
,
Tsunakawa
 
H
,
Tsutsumi
 
T
,
Harumi
 
K-i
.
Computer simulation of supraventricular tachycardia with the Wolff-Parkinson-White syndrome using three-dimensional heart models
.
J Electrocardiol
 
1990
;
23
:
261
273
.

25

Lazarus
 
A
,
Dalton
 
D
,
Husmeier
 
D
,
Gao
 
H
.
Sensitivity analysis and inverse uncertainty quantification for the left ventricular passive mechanics
.
Biomech Model Mechanobiol
 
2022
;
21
:
953
982
.

26

Karabelas
 
E
,
Longobardi
 
S
,
Fuchsberger
 
J
,
Razeghi
 
O
,
Rodero
 
C
,
Strocchi
 
M
, et al.
Global sensitivity analysis of four chamber heart hemodynamics using surrogate models
.
IEEE Trans Biomed Eng
 
2022
;
69
:
3216
3223
.

27

Strocchi
 
M
,
Longobardi
 
S
,
Augustin
 
CM
,
Gsell
 
MA
,
Petras
 
A
,
Rinaldi
 
CA
, et al.
Cell to whole organ global sensitivity analysis on a four-chamber heart electromechanics model using Gaussian processes emulators
.
PLoS Comput Biol
 
2023
;
19
:
e1011257
.

28

Rodero
 
C
,
Strocchi
 
M
,
Marciniak
 
M
,
Longobardi
 
S
,
Whitaker
 
J
,
O’Neill
 
MD
, et al.
Linking statistical shape models and simulated function in the healthy adult human heart
.
PLoS Comput Biol
 
2021
;
17
:
e1008851
.

29

Sudret
 
B
.
Global sensitivity analysis using polynomial chaos expansions
.
Reliab Eng Syst Saf
 
2008
;
93
:
964
979
.

30

Winkler
 
B
,
Nagel
 
C
,
Farchmin
 
N
,
Heidenreich
 
S
,
Loewe
 
A
,
Dössel
 
O
, et al.
Global sensitivity analysis and uncertainty quantification for simulated atrial electrocardiograms
.
Metrology
 
2022
;
3
:
1
28
.

31

Narayan
 
A
,
Liu
 
Z
,
Bergquist
 
JA
,
Charlebois
 
C
,
Rampersad
 
S
,
Rupp
 
L
, et al.
Uncertainsci: uncertainty quantification for computational models in biomedicine and bioengineering
.
Comput Biol Med
 
2023
;
152
:
106407
.

32

Rupp
 
LC
,
Busatto
 
A
,
Bergquist
 
JA
,
Gillette
 
K
,
Narayan
 
A
,
Plank
 
G
, et al.  
Uncertainty quantification of fiber orientation and epicardial activation. In 2023 Computing in Cardiology (CinC). Vol. 50. IEEE; 2023. p1–4
.

33

Bergquist
 
JA
,
Lange
 
M
,
Zenger
 
B
,
Orkild
 
B
,
Paccione
 
E
,
Kwan
 
E
, et al.  
Uncertainty quantification of the effect of variable conductivity in ventricular fibrotic regions on ventricular tachycardia. In 2023 Computing in Cardiology (CinC). Vol. 50. IEEE; 2023. p1–4
.

34

Busatto
 
A
,
Rupp
 
LC
,
Gillette
 
K
,
Narayan
 
A
,
Plank
 
G
,
MacLeod
 
RS
.
Capturing the influence of conduction velocity on epicardial activation patterns using uncertainty quantification. In 2023 Computing in Cardiology (CinC). Vol. 50. IEEE; 2023. p1–4
.

35

Gillette
 
K
,
Gsell
 
MA
,
Kurath-Koller
 
S
,
Manninger
 
M
,
Prassl
 
AJ
,
Scherr
 
D
, et al.  
Exploring role of accessory pathway location in Wolff-Parkinson-White syndrome in a model of whole heart electrophysiology. In 2022 Computing in Cardiology (CinC). Vol. 498. IEEE; 2022. p1–4
.

36

Gillette
 
K
,
Gsell
 
MA
,
Strocchi
 
M
,
Grandits
 
T
,
Neic
 
A
,
Manninger
 
M
.
A personalized real-time virtual model of whole heart electrophysiology
.
Front Physiol
 
2022
;
13
:
907190
.

37

Plank
 
G
,
Loewe
 
A
,
Neic
 
A
,
Augustin
 
C
,
Huang
 
Y-L
,
Gsell
 
MA
, et al.
The opencarp simulation environment for cardiac electrophysiology
.
Comput Methods Programs Biomed
 
2021
;
208
:
106223
.

38

Payer
 
C
,
Štern
 
D
,
Bischof
 
H
,
Urschler
 
M
.
Multi-label whole heart segmentation using CNNS and anatomical label configurations. In Statistical Atlases and Computational Models of the Heart. ACDC and MMWHS Challenges: 8th International Workshop, STACOM 2017, Held in Conjunction with MICCAI 2017, Quebec City, Canada, September 10–14, 2017, Revised Selected Papers. Springer International Publishing; 2018. p190–198
.

39

CIBC
. Seg3D: Volumetric image segmentation and visualization. Scientific Computing and Imaging Institute (SCI), Download from: http://www.seg3d.org, 2016.

40

Besl
 
PJ
,
McKay
 
ND
.
Method for registration of 3-d shapes. In Sensor Fusion IV: Control Paradigms and Data Structures Vol. 1611. Spie; 1992. p586–606
.

41

Prassl
 
AJ
,
Kickinger
 
F
,
Ahammer
 
H
,
Grau
 
V
,
Schneider
 
JE
,
Hofer
 
E
, et al.
Automatically generated, anatomically accurate meshes for cardiac electrophysiology problems
.
IEEE Trans Biomed Eng
 
2009
;
56
:
1318
30
.

42

Bayer
 
JD
,
Blake
 
RC
,
Plank
 
G
,
Trayanova
 
NA
.
A novel rule-based algorithm for assigning myocardial fiber orientation to computational heart models
.
Ann Biomed Eng
 
2012
;
40
:
2243
2254
.

43

Roney
 
CH
,
Pashaei
 
A
,
Meo
 
M
,
Dubois
 
R
,
Boyle
 
PM
,
Trayanova
 
NA
, et al.
Universal atrial coordinates applied to visualisation, registration and construction of patient specific meshes
.
Med Image Anal
 
2019
;
55
:
65
75
.

44

Bayer
 
J
,
Prassl
 
AJ
,
Pashaei
 
A
,
Gomez
 
JF
,
Frontera
 
A
,
Neic
 
A
, et al.
Universal ventricular coordinates: a generic framework for describing position within the heart and transferring data
.
Med Image Anal
 
2018
;
45
:
83
93
.

45

Gillette
 
K
,
Gsell
 
MA
,
Prassl
 
AJ
,
Karabelas
 
E
,
Reiter
 
U
,
Reiter
 
G
, et al.
A framework for the generation of digital twins of cardiac electrophysiology from clinical 12-leads ECGs
.
Med Image Anal
 
2021
;
71
:
102080
.

46

Roney
 
CH
,
Pashaei
 
A
,
Meo
 
M
,
Dubois
 
R
,
Boyle
 
PM
,
Trayanova
 
NA
, et al.
Universal atrial coordinates applied to visualisation, registration and construction of patient specific meshes
.
Med Image Anal
 
2019
;
55
:
65
75
.

47

Gillette
 
K
,
Gsell
 
MA
,
Bouyssier
 
J
,
Prassl
 
AJ
,
Neic
 
A
,
Vigmond
 
EJ
, et al.
Automated framework for the inclusion of a His–Purkinje system in cardiac digital twins of ventricular electrophysiology
.
Ann Biomed Eng
 
2021
;
49
:
3143
3153
.

48

Costabal
 
FS
,
Hurtado
 
DE
,
Kuhl
 
E
.
Generating purkinje networks in the human heart
.
J Biomech
 
2016
;
49
:
2455
2465
.

49

Neic
 
A
,
Gsell
 
MA
,
Karabelas
 
E
,
Prassl
 
AJ
,
Plank
 
G
.
Automating image-based mesh generation and manipulation tasks in cardiac modeling workflows using meshtool
.
SoftwareX
 
2020
;
11
:
100454
.

50

Sternick
 
EB
,
Sosa
 
EA
,
Timmermans
 
C
,
Filho
 
FEC
,
Rodriguez
 
L-M
,
Gerken
 
LM
, et al.
Automaticity in Mahaim fibers
.
J Cardiovasc Electrophysiol
 
2004
;
15
:
738
744
.

51

Vigmond
 
E
,
Dos Santos
 
RW
,
Prassl
 
A
,
Deo
 
M
,
Plank
 
G
.
Solvers for the cardiac bidomain equations
.
Prog Biophys Mol Biol
 
2008
;
96
:
3
18
.

52

Neic
 
A
,
Campos
 
FO
,
Prassl
 
AJ
,
Niederer
 
SA
,
Bishop
 
MJ
,
Vigmond
 
EJ
, et al.
Efficient computation of electrograms and ECGs in human whole heart simulations using a reaction-eikonal model
.
J Comput Phys
 
2017
;
346
:
191
211
.

53

Potse
 
M
.
Scalable and accurate ECG simulation for reaction-diffusion models of the human heart
.
Front Physiol
 
2018
;
9
:
370
.

54

Malmivuo
 
J
,
Plonsey
 
R
.
Bioelectromagnetism: principles and applications of bioelectric and biomagnetic fields
.
USA
:
Oxford University Press
;
1995
.

55

Hegemann
 
N
,
Heidenreich
 
S
.
Pythia: a python package for uncertainty quantification based on non-intrusive polynomial chaos expansions
.
J Open Source Softw
 
2023
;
8
:
5489
.

56

Jemtrén
 
A
,
Saygi
 
S
,
Åkerström
 
F
,
Asaad
 
F
,
Bourke
 
T
,
Braunschweig
 
F
, et al.
Risk assessment in patients with symptomatic and asymptomatic pre-excitation
.
Europace
 
2024
;
26
:
euae036
.

57

Park
 
K-m
,
Kim
 
Y-h
,
Marchlinski
 
FE
.
Using the surface electrocardiogram to localize the origin of idiopathic ventricular tachycardia
.
Pacing Clin Electrophysiol
 
2012
;
35
:
1516
1527
.

58

Senoner
 
T
,
Pfeifer
 
B
,
Barbieri
 
F
,
Adukauskaite
 
A
,
Dichtl
 
W
,
Bauer
 
A
, et al.
Identifying the location of an accessory pathway in pre-excitation syndromes using an artificial intelligence-based algorithm
.
J Clin Med
 
2021
;
10
:
4394
.

59

Nishimori
 
M
,
Kiuchi
 
K
,
Nishimura
 
K
,
Kusano
 
K
,
Yoshida
 
A
,
Adachi
 
K
, et al.
Accessory pathway analysis using a multimodal deep learning model
.
Sci Rep
 
2021
;
11
:
8045
.

60

Berger
 
T
,
Fischer
 
G
,
Pfeifer
 
B
,
Modre
 
R
,
Hanser
 
F
,
Trieb
 
T
, et al.
Single-beat noninvasive imaging of cardiac electrophysiology of ventricular pre-excitation
.
J Am Coll Cardiol
 
2006
;
48
:
2045
2052
.

61

Grandits
 
T
,
Gillette
 
K
,
Neic
 
A
,
Bayer
 
J
,
Vigmond
 
E
,
Pock
 
T
, et al.
An inverse Eikonal method for identifying ventricular activation sequences from epicardial activation maps
.
J Comput Phys
 
2020
;
419
:
109700
.

62

Camps
 
J
,
Berg
 
LA
,
Wang
 
ZJ
,
Sebastian
 
R
,
Riebel
 
LL
,
Doste
 
R
, et al.
Digital twinning of the human ventricular activation sequence to clinical 12-lead ECGs and magnetic resonance imaging using realistic purkinje networks for in silico clinical trials
.
Med Image Anal
 
2024
;
94
:
103108
.

63

Corrado
 
C
,
Razeghi
 
O
,
Roney
 
C
,
Coveney
 
S
,
Williams
 
S
,
Sim
 
I
, et al.
Quantifying atrial anatomy uncertainty from clinical data and its impact on electro-physiology simulation predictions
.
Med Image Anal
 
2020
;
61
:
101626
.

64

Lamata
 
P
,
Casero
 
R
,
Carapella
 
V
,
Niederer
 
SA
,
Bishop
 
MJ
,
Schneider
 
JE
, et al.
Images as drivers of progress in cardiac computational modelling
.
Prog Biophys Mol Biol
 
2014
;
115
:
198
212
.

65

Strocchi
 
M
,
Augustin
 
CM
,
Gsell
 
MA
,
Karabelas
 
E
,
Neic
 
A
,
Gillette
 
K
, et al.
A publicly available virtual cohort of four-chamber heart meshes for cardiac electro-mechanics simulations
.
PLoS One
 
2020
;
15
:
e0235145
.

66

Galappaththige
 
S
,
Gray
 
RA
,
Costa
 
CM
,
Niederer
 
S
,
Pathmanathan
 
P
.
Credibility assessment of patient-specific computational modeling using patient-specific cardiac modeling as an exemplar
.
PLoS Comput Biol
 
2022
;
18
:
e1010541
.

67

Jiang
 
X
,
Missel
 
R
,
Toloubidokhti
 
M
,
Gillette
 
K
,
Prassl
 
AJ
,
Plank
 
G
.
Hybrid neural state-space modeling for supervised and unsupervised electrocardiographic imaging
.
IEEE Trans Med Imaging
 
2024
;
43
:
2733
2744
.

68

Boyle
 
PM
,
Zghaib
 
T
,
Zahid
 
S
,
Ali
 
RL
,
Deng
 
D
,
Franceschi
 
WH
, et al.
Computationally guided personalized targeted ablation of persistent atrial fibrillation
.
Nat Biomed Eng
 
2019
;
3
:
870
879
.

69

Strocchi
 
M
,
Gillette
 
K
,
Neic
 
A
,
Elliott
 
MK
,
Wijesuriya
 
N
,
Mehta
 
V
, et al.
Effect of scar and his–purkinje and myocardium conduction on response to conduction system pacing
.
J Cardiovasc Electrophysiol
 
2023
;
34
:
984
993
.

70

Carpio
 
EF
,
Gomez
 
JF
,
Sebastian
 
R
,
Lopez-Perez
 
A
,
Castellanos
 
E
,
Almendral
 
J
, et al.
Optimization of lead placement in the right ventricle during cardiac resynchronization therapy. a simulation study
.
Front Physiol
 
2019
;
10
:
74
.

71

Gintant
 
G
,
Sager
 
PT
,
Stockbridge
 
N
.
Evolution of strategies to improve preclinical cardiac safety testing
.
Nat Rev Drug Discov
 
2016
;
15
:
457
471
.

72

Grandits
 
T
,
Augustin
 
CM
,
Haase
 
G
,
Jost
 
N
,
Mirams
 
GR
,
Niederer
 
SA
, et al.
Neural network emulation of the human ventricular cardiomyocyte action potential for more efficient computations in pharmacological studies
.
Elife
 
2024
;
12
:
RP91911
.

73

Marciniak
 
M
,
van Deutekom
 
AW
,
Toemen
 
L
,
Lewandowski
 
AJ
,
Gaillard
 
R
,
Young
 
AA
, et al.
A three-dimensional atlas of child’s cardiac anatomy and the unique morphological alterations associated with obesity
.
Eur Heart J Cardiovasc Imaging
 
2022
;
23
:
1645
1653
.

74

Young
 
AA
,
Frangi
 
AF
.
Computational cardiac atlases: from patient to population and back
.
Exp Physiol
 
2009
;
94
:
578
596
.

75

Correa
 
FS
,
Lokhandwala
 
Y
,
Cruz Filho
 
F
,
Sánchez-Quintana
 
D
,
Mori
 
S
,
Anderson
 
RH
, et al.
Part II–Clinical presentation, electrophysiologic characteristics, and when and how to ablate atriofascicular pathways and long and short decrementally conducting accessory pathways
.
J Cardiovasc Electrophysiol
 
2019
;
30
:
3079
3096
.

76

Liu
 
S
,
Yuan
 
S
,
Olsson
 
B
.
Conduction properties of accessory atrioventricular pathways: importance of the accessory pathway location and normal atrioventricular conduction
.
Scand Cardiovasc J
 
2003
;
37
:
43
48
.

77

Gsell
 
MA
,
Neic
 
A
,
Bishop
 
MJ
,
Gillette
 
K
,
Prassl
 
AJ
,
Augustin
 
CM
, et al.
Forcepss—a framework for cardiac electrophysiology simulations standardization
.
Comput Methods Programs Biomed
 
2024
;
251
:
108189
.

Author notes

Conflict of interest: G.P. and E.V. are associated with the company NumeriCor GmbH. All other authors declare no potential conflicts of interest.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.

Supplementary data