Abstract

To conduct a systematic review and meta-analysis to compare the effectiveness and safety of cryoballoon ablation of atrial fibrillation (AF) performed using a single freeze strategy in comparison to an empiric double (‘bonus’) freeze strategy. We systematically searched MEDLINE, EMBASE, and CENTRAL databases from inception to 12 July 2020, for prospective and retrospective studies of patients undergoing cryoballoon for paroxysmal or persistent AF comparing a single vs. bonus freeze strategy. The main outcome was atrial arrhythmia-free survival and eligible studies required at least 12 months of follow-up; the primary safety outcome was a composite of all complications. Study quality was assessed using the Cochrane risk of bias tool and the Newcastle–Ottawa Scale. Thirteen studies (3 randomized controlled trials and 10 observational studies) comprising 3163 patients were eligible for inclusion (64% males, 71.5% paroxysmal AF, mean CHA2DS2-VASc score 1.3 ± 0.9). There was no significant difference in pooled effectiveness between single freeze strategy compared to double freeze strategy [relative risk (RR) 1.03; 95% confidence interval (CI): 0.98–1.07; I2 = 0%]. Single freeze procedures were associated with a significantly lower adverse event rate (RR 0.72; 95% CI: 0.53–0.98; I2 = 0%) and shorter average procedure time (90 ± 27 min vs. 121 ± 36 min, P < 0.001). A trend for lower risk of persistent phrenic nerve palsy was observed (RR 0.61; 95% CI: 0.37–1.01; I2 = 0%). The quality of included studies was moderate/good, with no evidence of significant publication bias. Single freeze strategy for cryoballoon of AF is as effective as an empiric double (‘bonus’) freeze strategy while appearing safer and probably quicker (PROSPERO registration number CRD42020158696).

What’s new?
  • This systematic review identified 13 studies comparing single freeze to double (‘bonus’) freeze cryoballoon strategy, including three randomized controlled trials.

  • During a ≥12-month follow-up, the single freeze strategy of cryoballoon ablation of atrial fibrillation (AF) was as effective as the double freeze strategy.

  • The single freeze strategy was associated with a lower risk of any adverse events and a strong trend for lower risk of persistent phrenic nerve palsy.

  • The single freeze strategy probably allows to shorten procedure duration.

  • These findings question the need for routine bonus freeze during cryoballoon ablation of AF.

Introduction

Catheter ablation is the most effective rhythm control strategy for atrial fibrillation (AF) patients, with the potential to improve prognosis in selected populations.1–3 Electrical isolation of the pulmonary veins is considered a crucial endpoint of AF ablation and can be achieved by a variety of ablation techniques, with radiofrequency ablation (RFA) and cryoballoon ablation being the most widely used.1,4 A large body of evidence indicates clinical equivalence of both ablation techniques in terms of arrhythmia recurrence, albeit with significant differences in procedure duration or safety.1,4–6 In two pivotal trials which compared cryoballoon ablation to RFA, all pulmonary veins were subject to at least two cryoballoon applications (double or ‘bonus’ freeze strategy) with a fixed time of freezing.7,8 The cryoballoon ablation technique has evolved with time, leading to a reduction in both the number and duration of applications.9–11 Recent studies report that even a single 180 s application with or without time-to-isolation (TTI) guidance (single freeze strategy) may be sufficient for a favourable long-term effect.10,12 Based on the published data, it is unclear how this evolution may influence the effectiveness and safety of cryoballoon ablation, especially considering that two recent multicentre trials investigating cryoballoon ablation effectiveness as a first-line treatment of AF used a ‘bonus’ freeze strategy.13–15

Therefore, we conducted a systematic review and meta-analysis of randomized and observational studies with a control group to evaluate the effectiveness and safety of a single freeze cryoballoon ablation of AF in comparison to the double (‘bonus’) freeze strategy.

Methods

This systematic review followed the Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines.16 The protocol is registered in the international prospective register of systematic reviews (CRD42020158696).

Search strategy and selection criteria

MEDLINE (Medical Literature Analysis and Retrieval System Online) via PubMED, EMBASE (Excerpta Medical Database), and CENTRAL (Cochrane Central Register of Controlled Trials) databases were searched from inception to the 25 July 2019 using the following search string (cryo* AND ablation AND ‘atrial fibrillation’). Due to delay in completion of the review, the MEDLINE search was updated on 12 July 2020. Reference lists of eligible studies were searched for additional sources of information. Full-text publications of conference abstracts or registered clinical trials were sought. We also contacted key opinion leaders in the field of cryoballoon ablation to guarantee the inclusion of all relevant studies.

We included all prospective and retrospective studies with a control group which assessed the first-time cryoballoon ablation performed in adult patients with either paroxysmal or persistent AF using second or newest generation cryoballoons. Generations 2 to 4 are currently commercially available and do not differ significantly in terms of clinical performance. Only studies published in English were included.

The intervention assessed was cryoballoon ablation of AF. The comparison of interest was intended single freeze cryoballoon procedure (‘single freeze group’) vs. intended two cryoballoon applications or an empirical ‘bonus’ application after demonstrating pulmonary vein isolation (‘double/bonus freeze group’) regardless of the method used for assessment of pulmonary vein isolation (spiral catheter, TTI, etc.) or length of application. An additional freeze was allowed in the ‘single freeze group’ when it was felt that isolation was either not achieved or was achieved very late in the freeze, causing concerns for lesion durability.

Two investigators (M.M.F. and M.K.) independently screened and selected potentially eligible studies based on title and abstract. Final eligibility was decided after evaluation of full-text publications. All disagreements were resolved via discussion or through the involvement of a third referee (D.G.).

Data extraction and quality assessment

Data extraction was done independently by two investigators (M.M.F. and M.K.) and all disagreements were resolved via discussion, or through the involvement of a third referee (D.G.). A standardized form was used to extract the following information from each study: (i) study design and methodology, (ii) details of the ablation procedure (single or double application, TTI), (iii) information on the assessment of the main clinical outcome (surface ECG, Holter monitoring), including length of follow-up, (iv) baseline characteristics of participants (age, CHA2DS2-VASc score, paroxysmal and persistent AF, left atrial diameter), and (v) measures of effect and safety as stated in the protocol of the current meta-analysis. In case of missing data on main outcomes, authors of the original publications were contacted via email.

The risk of bias in randomized controlled trials (RCTs) was assessed using the Cochrane risk of bias tool,17 while the Newcastle–Ottawa Scale (NOS) was used to assess the quality of non-randomized studies. Critical assessments on the risk of bias (high, low, unclear) were done separately for each domain. The risk of bias assessment was done independently by two investigators (M.M.F. and M.K.) and all disagreements were resolved via discussion, or through the involvement of a third referee (D.G.). A trial was considered of high quality if no domains scored as high risk, or low quality if three or more domains scored as high risk. High-quality non-randomized studies were defined as those with a Newcastle–Ottawa score of ≥7.

Outcome measures

The main clinical outcome was atrial arrhythmia-free survival assessed at least 12 months from the date of the ablation, defined as the lack of any atrial arrhythmia (AF or other) lasting more than 30 s as assessed by repeated ECG and/or Holter monitoring. The main safety outcome was any adverse effects of the ablation. Secondary outcome measures included: persistent phrenic nerve palsy (PNP), defined as palsy lasting longer than until the end of the index ablation, transient PNP, defined as palsy which resolved before the end of the index ablation, tamponade, atrio-oesophageal fistula formation, procedure duration, and fluoroscopy time.

Data synthesis and analysis

Data were synthesized if reported in at least two included studies. Continuous variables were presented as mean with standard deviation (SD). If means and SD were not reported, these were estimated from sample size, medians and quartiles.18,19

We used random effect models to calculate pooled: (i) relative risk (RR) for freedom from any atrial arrhythmia in a ≥12-month follow-up, (ii) RR of ablation-associated persistent PNP, (iii) RR of ablation-associated transient PNP, (iv) RR of any ablation-associated adverse event, (v) difference in procedure duration time, and (vi) difference in fluoroscopy time, all with 95% confidence intervals (95% CIs).

Analyses were conducted separately for randomized and non-randomized studies. Subgroups were compared using the Z statistics. Heterogeneity between study populations was calculated using I2 statistics, where values of less than 25%, 50%, and 75% were regarded as evidence of low, moderate, and high levels of heterogeneity, respectively.20 Funnel plots were used for evaluating the presence of publication bias.

Analyses were performed using Statistica 13.3 (TIBCO Software Inc., CA, USA).

Results

Selection and description of studies

The PRISMA flow diagram in Figure 1 summarizes the study selection process. Primary searches identified 3930 citations and, after removal of duplicates, 2666 abstracts were independently screened by two investigators. Taking also into account additional searches (MEDLINE update, reference lists), a total of 16 potentially relevant studies were selected for full-text examination. Three studies were excluded based on the wrong comparator (no single freeze group in all cases13,29,30) and, finally, 13 studies were included in the data synthesis: 3 RCTs11,12,21 and 10 non-RCTs.9,10,12,22–28

PRISMA flow diagram for study selection process. a Update of the MEDLINE search, manual search of reference lists of included studies, contact with key opinion leaders. RCT, randomized controlled trial.
Figure 1

PRISMA flow diagram for study selection process. a Update of the MEDLINE search, manual search of reference lists of included studies, contact with key opinion leaders. RCT, randomized controlled trial.

Characteristics of included studies

The characteristics of included studies comprising methodology, demographic data, cryoballoon ablation details, and follow-up strategies are provided in Tables 1 and2. Five studies11,21,22,25,31 used a single freeze strategy based on the TTI principle, whereas investigators of two studies10,26 changed the ablation technique from empirical to TTI-guided applications over the course of their studies and six others used empirical 180/240 s applications.9,12,23,24,27,28 In all cases, pulmonary vein isolation was confirmed before the bonus freeze application. The diaphragmatic compound motor action potential assessment, an additional precaution measure for preventing PNP, was used in five studies.12,22,24–26 The definition of primary efficacy and safety endpoints were homogenous among included studies. Apart from two studies,21,31 effectiveness was defined as freedom from any atrial arrhythmia lasting >30 s recorded after a 3-month blanking period. Effectiveness was assessed during clinical visits and through repeated elective electrocardiograms (ECG) or Holter ECG monitoring of varying duration. In all studies, persistent PNP was defined as PNP lasting longer than completion of the index procedure.

Table 1

Summary of methodology and cryoablation details of included studies

First author, year, referenceStudy design (number of centres)Sample size
Description of the procedure
Demographic data
Study groupControl groupStudy groupControl groupTTICMAPPatients’ mean age
Male sex (%)
Paroxysmal AF (%)
Study groupControl groupStudy groupControl groupStudy groupControl group
Heeger 2016Prospective cohort study (1)6060Single 240 s application1 bonus 240 s application after PVI0161 ± 1162 ± 1163608375
Tebbenjohanns 2016Prospective cohort study with a historical control group (2)53139Single 240 s application, adenosine challengeTwo 240 s applications0066 ± 1061 ± 1151547263
Chun 2017Randomized controlled trial (1)5050TTI <75 s or within 25 s after a pull-down, application 240 s; TTI >75 s or not recorded, application 240 s and 240 s bonusTwo 240 s applications1066 ± 1063 ± 126058100100
Ekizler 2017Prospective cohort study (1)5680Single application 240 s1 bonus application after PVI0058 (48–67)62 (49–68)5755100100
Aryana 2017Prospective cohort study (5)355400TTI ≤60 s, one application TTI + 120 s; TTI 60–90 s, one application TTI + 120s and bonus 120 s; TTI >90 s, application aborted; no TT-PVI, 180 s application plus 120 s bonus2–3 applications lasting 2–4 min1164 ± 1163 ± 1169747274
Ströker 2018Case-control study with propensity scoring (2)256256Single 180/240 min application; bonus application if: temperature >−40°C within 1 min, no PVI or early spontaneous PV reconnection1–2 bonus application(s) (240 s/180s) after PVI0059 ± 1260 ± 1162678279
Pott 2018Case-control study (1)100100TTI <30 s, application 120 s; TTI 30–60 s, single application 180 s; TTI >60 s, 180 s application + 180 s bonus; no TTI recording, single 180 s applicationTwo 240 s applications1165 ± 10.965.3 ± 11.356576569
Rottner 2018aProspective cohort study (2)352211Single 240 s application; later single application TTI + 120 s and if no TTI recording, single 180 s applicationTwo 240 s applications1163.3 ± 10.963.3 ± 10.964645858
Mortsell 2018Randomized controlled trial (1)6970Single 240 s application; either TTI or temp. ≤40°C within 120 sTwo 240 s applications1061.9 ± 9.0868.3 ± 10.0707749.340.0
Yoshiga 2019Prospective cohort study (1)6733Single application ≥180 min1 bonus 120 s application after PVI0065.1 ± 10.067.5 ± 8.35870100100
Cordes 2019Prospective cohort study (1)3535TTI + 120 s; if TTI not recorded, single 180 s application with target temp. <40°C; if TTI >90 s, application abortedTwo 180 s applications1058 (IQR 16)60 (IQR 20)71806057
Miyamoto 2019Randomized controlled trial (3)5555Single 180 s application1 bonus 180 s application after PVI0163.1 ± 11.864.0 ± 11.065.563.6100100
Koektuerk 2019Case-control study (1)7792Single 240 s applicationTwo 240 s applications0061 ± 1064 ± 1063.685.9100100
First author, year, referenceStudy design (number of centres)Sample size
Description of the procedure
Demographic data
Study groupControl groupStudy groupControl groupTTICMAPPatients’ mean age
Male sex (%)
Paroxysmal AF (%)
Study groupControl groupStudy groupControl groupStudy groupControl group
Heeger 2016Prospective cohort study (1)6060Single 240 s application1 bonus 240 s application after PVI0161 ± 1162 ± 1163608375
Tebbenjohanns 2016Prospective cohort study with a historical control group (2)53139Single 240 s application, adenosine challengeTwo 240 s applications0066 ± 1061 ± 1151547263
Chun 2017Randomized controlled trial (1)5050TTI <75 s or within 25 s after a pull-down, application 240 s; TTI >75 s or not recorded, application 240 s and 240 s bonusTwo 240 s applications1066 ± 1063 ± 126058100100
Ekizler 2017Prospective cohort study (1)5680Single application 240 s1 bonus application after PVI0058 (48–67)62 (49–68)5755100100
Aryana 2017Prospective cohort study (5)355400TTI ≤60 s, one application TTI + 120 s; TTI 60–90 s, one application TTI + 120s and bonus 120 s; TTI >90 s, application aborted; no TT-PVI, 180 s application plus 120 s bonus2–3 applications lasting 2–4 min1164 ± 1163 ± 1169747274
Ströker 2018Case-control study with propensity scoring (2)256256Single 180/240 min application; bonus application if: temperature >−40°C within 1 min, no PVI or early spontaneous PV reconnection1–2 bonus application(s) (240 s/180s) after PVI0059 ± 1260 ± 1162678279
Pott 2018Case-control study (1)100100TTI <30 s, application 120 s; TTI 30–60 s, single application 180 s; TTI >60 s, 180 s application + 180 s bonus; no TTI recording, single 180 s applicationTwo 240 s applications1165 ± 10.965.3 ± 11.356576569
Rottner 2018aProspective cohort study (2)352211Single 240 s application; later single application TTI + 120 s and if no TTI recording, single 180 s applicationTwo 240 s applications1163.3 ± 10.963.3 ± 10.964645858
Mortsell 2018Randomized controlled trial (1)6970Single 240 s application; either TTI or temp. ≤40°C within 120 sTwo 240 s applications1061.9 ± 9.0868.3 ± 10.0707749.340.0
Yoshiga 2019Prospective cohort study (1)6733Single application ≥180 min1 bonus 120 s application after PVI0065.1 ± 10.067.5 ± 8.35870100100
Cordes 2019Prospective cohort study (1)3535TTI + 120 s; if TTI not recorded, single 180 s application with target temp. <40°C; if TTI >90 s, application abortedTwo 180 s applications1058 (IQR 16)60 (IQR 20)71806057
Miyamoto 2019Randomized controlled trial (3)5555Single 180 s application1 bonus 180 s application after PVI0163.1 ± 11.864.0 ± 11.065.563.6100100
Koektuerk 2019Case-control study (1)7792Single 240 s applicationTwo 240 s applications0061 ± 1064 ± 1063.685.9100100

AF, atrial fibrillation; AT, atrial tachycardia; CMAP, compound motor action potentials; IQR, interquartile range; PVI, pulmonary vein isolation, TTI, time to isolation.

a

Rottner 2018 reported aggregated demographic data for all patient groups.

Table 1

Summary of methodology and cryoablation details of included studies

First author, year, referenceStudy design (number of centres)Sample size
Description of the procedure
Demographic data
Study groupControl groupStudy groupControl groupTTICMAPPatients’ mean age
Male sex (%)
Paroxysmal AF (%)
Study groupControl groupStudy groupControl groupStudy groupControl group
Heeger 2016Prospective cohort study (1)6060Single 240 s application1 bonus 240 s application after PVI0161 ± 1162 ± 1163608375
Tebbenjohanns 2016Prospective cohort study with a historical control group (2)53139Single 240 s application, adenosine challengeTwo 240 s applications0066 ± 1061 ± 1151547263
Chun 2017Randomized controlled trial (1)5050TTI <75 s or within 25 s after a pull-down, application 240 s; TTI >75 s or not recorded, application 240 s and 240 s bonusTwo 240 s applications1066 ± 1063 ± 126058100100
Ekizler 2017Prospective cohort study (1)5680Single application 240 s1 bonus application after PVI0058 (48–67)62 (49–68)5755100100
Aryana 2017Prospective cohort study (5)355400TTI ≤60 s, one application TTI + 120 s; TTI 60–90 s, one application TTI + 120s and bonus 120 s; TTI >90 s, application aborted; no TT-PVI, 180 s application plus 120 s bonus2–3 applications lasting 2–4 min1164 ± 1163 ± 1169747274
Ströker 2018Case-control study with propensity scoring (2)256256Single 180/240 min application; bonus application if: temperature >−40°C within 1 min, no PVI or early spontaneous PV reconnection1–2 bonus application(s) (240 s/180s) after PVI0059 ± 1260 ± 1162678279
Pott 2018Case-control study (1)100100TTI <30 s, application 120 s; TTI 30–60 s, single application 180 s; TTI >60 s, 180 s application + 180 s bonus; no TTI recording, single 180 s applicationTwo 240 s applications1165 ± 10.965.3 ± 11.356576569
Rottner 2018aProspective cohort study (2)352211Single 240 s application; later single application TTI + 120 s and if no TTI recording, single 180 s applicationTwo 240 s applications1163.3 ± 10.963.3 ± 10.964645858
Mortsell 2018Randomized controlled trial (1)6970Single 240 s application; either TTI or temp. ≤40°C within 120 sTwo 240 s applications1061.9 ± 9.0868.3 ± 10.0707749.340.0
Yoshiga 2019Prospective cohort study (1)6733Single application ≥180 min1 bonus 120 s application after PVI0065.1 ± 10.067.5 ± 8.35870100100
Cordes 2019Prospective cohort study (1)3535TTI + 120 s; if TTI not recorded, single 180 s application with target temp. <40°C; if TTI >90 s, application abortedTwo 180 s applications1058 (IQR 16)60 (IQR 20)71806057
Miyamoto 2019Randomized controlled trial (3)5555Single 180 s application1 bonus 180 s application after PVI0163.1 ± 11.864.0 ± 11.065.563.6100100
Koektuerk 2019Case-control study (1)7792Single 240 s applicationTwo 240 s applications0061 ± 1064 ± 1063.685.9100100
First author, year, referenceStudy design (number of centres)Sample size
Description of the procedure
Demographic data
Study groupControl groupStudy groupControl groupTTICMAPPatients’ mean age
Male sex (%)
Paroxysmal AF (%)
Study groupControl groupStudy groupControl groupStudy groupControl group
Heeger 2016Prospective cohort study (1)6060Single 240 s application1 bonus 240 s application after PVI0161 ± 1162 ± 1163608375
Tebbenjohanns 2016Prospective cohort study with a historical control group (2)53139Single 240 s application, adenosine challengeTwo 240 s applications0066 ± 1061 ± 1151547263
Chun 2017Randomized controlled trial (1)5050TTI <75 s or within 25 s after a pull-down, application 240 s; TTI >75 s or not recorded, application 240 s and 240 s bonusTwo 240 s applications1066 ± 1063 ± 126058100100
Ekizler 2017Prospective cohort study (1)5680Single application 240 s1 bonus application after PVI0058 (48–67)62 (49–68)5755100100
Aryana 2017Prospective cohort study (5)355400TTI ≤60 s, one application TTI + 120 s; TTI 60–90 s, one application TTI + 120s and bonus 120 s; TTI >90 s, application aborted; no TT-PVI, 180 s application plus 120 s bonus2–3 applications lasting 2–4 min1164 ± 1163 ± 1169747274
Ströker 2018Case-control study with propensity scoring (2)256256Single 180/240 min application; bonus application if: temperature >−40°C within 1 min, no PVI or early spontaneous PV reconnection1–2 bonus application(s) (240 s/180s) after PVI0059 ± 1260 ± 1162678279
Pott 2018Case-control study (1)100100TTI <30 s, application 120 s; TTI 30–60 s, single application 180 s; TTI >60 s, 180 s application + 180 s bonus; no TTI recording, single 180 s applicationTwo 240 s applications1165 ± 10.965.3 ± 11.356576569
Rottner 2018aProspective cohort study (2)352211Single 240 s application; later single application TTI + 120 s and if no TTI recording, single 180 s applicationTwo 240 s applications1163.3 ± 10.963.3 ± 10.964645858
Mortsell 2018Randomized controlled trial (1)6970Single 240 s application; either TTI or temp. ≤40°C within 120 sTwo 240 s applications1061.9 ± 9.0868.3 ± 10.0707749.340.0
Yoshiga 2019Prospective cohort study (1)6733Single application ≥180 min1 bonus 120 s application after PVI0065.1 ± 10.067.5 ± 8.35870100100
Cordes 2019Prospective cohort study (1)3535TTI + 120 s; if TTI not recorded, single 180 s application with target temp. <40°C; if TTI >90 s, application abortedTwo 180 s applications1058 (IQR 16)60 (IQR 20)71806057
Miyamoto 2019Randomized controlled trial (3)5555Single 180 s application1 bonus 180 s application after PVI0163.1 ± 11.864.0 ± 11.065.563.6100100
Koektuerk 2019Case-control study (1)7792Single 240 s applicationTwo 240 s applications0061 ± 1064 ± 1063.685.9100100

AF, atrial fibrillation; AT, atrial tachycardia; CMAP, compound motor action potentials; IQR, interquartile range; PVI, pulmonary vein isolation, TTI, time to isolation.

a

Rottner 2018 reported aggregated demographic data for all patient groups.

Table 2

Summary of the follow-up strategies among included studies

First author, year, referenceAssessment of the primary end pointStandard AF ablation endpointaMean follow-up time
AADs managementInformation on redo procedures
Study groupControl groupAll patients
Heeger 2016ECG and 24 h Holter at 3, 6, 12 months and in 6-month intervals thereafter; regular telephonic interviews, additional outpatient visits in symptomatic patients1848 ± 101 days849 ± 74 daysAADs continued for 3 monthsA total of 26/34 (76 %) patients suffering from atrial arrhythmia recurrences underwent a second ablation
Tebbenjohanns 2016Clinical visit and 24 h Holter at 3, 6, 12, and 18 months; external event recording for 4 weeks in symptomatic patients1458 ± 107 daysNot reportedNot reported
Chun 2017Visits at 3, 6, 12 months; ECG and 72-h Holter ECG; additional telephone interviews and event recording in symptomatic patients1372 (351–455) days378(361–483) daysAADs discontinued after the procedure. Resumption of AADs in case of AF relapse during blanking periodNot reported
Ekizler 2017ECG and 24 h Holter at 1, 3, 6 and 12 month, biannually thereafter112 ± 3 months after blanking period13 ± 3 months after blanking periodDiscontinued at the end of the blanking periodA redo procedure was performed in five patients in Study group and in six patients in Control group
Aryana 2017ECG during each follow-up visit; 2- to 4-week ambulatory electrocardiographic monitoring at 6 weeks, 3 and 6 months115 ± 2 month16 ± 3 monthAntiarrhythmic therapy was discontinued within 6 weeks of ablation35 patients (9.9%) in Study group and 63 patients (15.7%) in Control group underwent a repeat catheter ablation during the study period
Ströker 2018ECG, Holter at 1, 3, 6 month and every 6 month after or in case of symptoms118 ± 10 monthsAAD’s were discontinued 3 months post-ablation if no recurrenceA redo procedure was performed in 44 (66%) patients in the Control group and 37 (50%) patients in the Study group
Pott 2018Clinical assessment, echocardiography, ECG, and 7-day-Holter-monitoring at 1, 3, and 6 months and thereafter every 6 months1436 ± 184 daysAADs withheld either after the procedure or after blanking periodNot reported
Rottner 2018No information provided; paper focused on procedure safetyNANANANAAADs recommended to be continued for 3 monthsNA
Mortsell 2018Clinical visit at 3, 6, and 12 months; a 7 day Holter at 6 and 12 months1 (1 month blanking period)12 months12 monthsAntiarrhythmic drugs were withdrawn after 3 months in asymptomatic patients free from clinical AF recurrencesRe-ablations were performed in 7 of 70 (10%) patients in the Study group and in 8 of 70 (11.4%) patients in the Control group
Yoshiga 2019ECG and Holter at 1, 3, 6, and 12 months; event monitoring in symptomatic patients112 months12 monthsIn all the patients the anti-arrhythmic drugs were discontinued after the procedureNot reported
Cordes 219Standardized telephone interview at 3 and 6 monthsb6 months6 monthsNot reportedNot reported
Miyamoto 2019ECG at 1 month, and then every 1–3 months; 24 h Holter at 3 and 12 month; event recorder in symptomatic patients112 months12 monthsDiscontinuation of antiarrhythmic drugs was recommended after the ablationNot reported
Koektuerk 2019Outpatient visit at 1, 3, 6, 12 months or earlier if symptomatic; 7-day Holter recording at 3 and 6 months and 24-h Holter later; telephone interview at the end of the follow-up116.4 ± 7.519.0 ± 8.6AADs continued for at least 3 months after ablation7 patients in the Study group and 14 in the Control group underwent repeated ablation
First author, year, referenceAssessment of the primary end pointStandard AF ablation endpointaMean follow-up time
AADs managementInformation on redo procedures
Study groupControl groupAll patients
Heeger 2016ECG and 24 h Holter at 3, 6, 12 months and in 6-month intervals thereafter; regular telephonic interviews, additional outpatient visits in symptomatic patients1848 ± 101 days849 ± 74 daysAADs continued for 3 monthsA total of 26/34 (76 %) patients suffering from atrial arrhythmia recurrences underwent a second ablation
Tebbenjohanns 2016Clinical visit and 24 h Holter at 3, 6, 12, and 18 months; external event recording for 4 weeks in symptomatic patients1458 ± 107 daysNot reportedNot reported
Chun 2017Visits at 3, 6, 12 months; ECG and 72-h Holter ECG; additional telephone interviews and event recording in symptomatic patients1372 (351–455) days378(361–483) daysAADs discontinued after the procedure. Resumption of AADs in case of AF relapse during blanking periodNot reported
Ekizler 2017ECG and 24 h Holter at 1, 3, 6 and 12 month, biannually thereafter112 ± 3 months after blanking period13 ± 3 months after blanking periodDiscontinued at the end of the blanking periodA redo procedure was performed in five patients in Study group and in six patients in Control group
Aryana 2017ECG during each follow-up visit; 2- to 4-week ambulatory electrocardiographic monitoring at 6 weeks, 3 and 6 months115 ± 2 month16 ± 3 monthAntiarrhythmic therapy was discontinued within 6 weeks of ablation35 patients (9.9%) in Study group and 63 patients (15.7%) in Control group underwent a repeat catheter ablation during the study period
Ströker 2018ECG, Holter at 1, 3, 6 month and every 6 month after or in case of symptoms118 ± 10 monthsAAD’s were discontinued 3 months post-ablation if no recurrenceA redo procedure was performed in 44 (66%) patients in the Control group and 37 (50%) patients in the Study group
Pott 2018Clinical assessment, echocardiography, ECG, and 7-day-Holter-monitoring at 1, 3, and 6 months and thereafter every 6 months1436 ± 184 daysAADs withheld either after the procedure or after blanking periodNot reported
Rottner 2018No information provided; paper focused on procedure safetyNANANANAAADs recommended to be continued for 3 monthsNA
Mortsell 2018Clinical visit at 3, 6, and 12 months; a 7 day Holter at 6 and 12 months1 (1 month blanking period)12 months12 monthsAntiarrhythmic drugs were withdrawn after 3 months in asymptomatic patients free from clinical AF recurrencesRe-ablations were performed in 7 of 70 (10%) patients in the Study group and in 8 of 70 (11.4%) patients in the Control group
Yoshiga 2019ECG and Holter at 1, 3, 6, and 12 months; event monitoring in symptomatic patients112 months12 monthsIn all the patients the anti-arrhythmic drugs were discontinued after the procedureNot reported
Cordes 219Standardized telephone interview at 3 and 6 monthsb6 months6 monthsNot reportedNot reported
Miyamoto 2019ECG at 1 month, and then every 1–3 months; 24 h Holter at 3 and 12 month; event recorder in symptomatic patients112 months12 monthsDiscontinuation of antiarrhythmic drugs was recommended after the ablationNot reported
Koektuerk 2019Outpatient visit at 1, 3, 6, 12 months or earlier if symptomatic; 7-day Holter recording at 3 and 6 months and 24-h Holter later; telephone interview at the end of the follow-up116.4 ± 7.519.0 ± 8.6AADs continued for at least 3 months after ablation7 patients in the Study group and 14 in the Control group underwent repeated ablation

AADs, antiarrhythmic drugs; AF, atrial fibrillation; AT, atrial tachycardia.

a

Defined as AF/AT episode lasting more than 30 s recorded after a 3 month blanking period.

b

Only 6 month follow-up results available.

Table 2

Summary of the follow-up strategies among included studies

First author, year, referenceAssessment of the primary end pointStandard AF ablation endpointaMean follow-up time
AADs managementInformation on redo procedures
Study groupControl groupAll patients
Heeger 2016ECG and 24 h Holter at 3, 6, 12 months and in 6-month intervals thereafter; regular telephonic interviews, additional outpatient visits in symptomatic patients1848 ± 101 days849 ± 74 daysAADs continued for 3 monthsA total of 26/34 (76 %) patients suffering from atrial arrhythmia recurrences underwent a second ablation
Tebbenjohanns 2016Clinical visit and 24 h Holter at 3, 6, 12, and 18 months; external event recording for 4 weeks in symptomatic patients1458 ± 107 daysNot reportedNot reported
Chun 2017Visits at 3, 6, 12 months; ECG and 72-h Holter ECG; additional telephone interviews and event recording in symptomatic patients1372 (351–455) days378(361–483) daysAADs discontinued after the procedure. Resumption of AADs in case of AF relapse during blanking periodNot reported
Ekizler 2017ECG and 24 h Holter at 1, 3, 6 and 12 month, biannually thereafter112 ± 3 months after blanking period13 ± 3 months after blanking periodDiscontinued at the end of the blanking periodA redo procedure was performed in five patients in Study group and in six patients in Control group
Aryana 2017ECG during each follow-up visit; 2- to 4-week ambulatory electrocardiographic monitoring at 6 weeks, 3 and 6 months115 ± 2 month16 ± 3 monthAntiarrhythmic therapy was discontinued within 6 weeks of ablation35 patients (9.9%) in Study group and 63 patients (15.7%) in Control group underwent a repeat catheter ablation during the study period
Ströker 2018ECG, Holter at 1, 3, 6 month and every 6 month after or in case of symptoms118 ± 10 monthsAAD’s were discontinued 3 months post-ablation if no recurrenceA redo procedure was performed in 44 (66%) patients in the Control group and 37 (50%) patients in the Study group
Pott 2018Clinical assessment, echocardiography, ECG, and 7-day-Holter-monitoring at 1, 3, and 6 months and thereafter every 6 months1436 ± 184 daysAADs withheld either after the procedure or after blanking periodNot reported
Rottner 2018No information provided; paper focused on procedure safetyNANANANAAADs recommended to be continued for 3 monthsNA
Mortsell 2018Clinical visit at 3, 6, and 12 months; a 7 day Holter at 6 and 12 months1 (1 month blanking period)12 months12 monthsAntiarrhythmic drugs were withdrawn after 3 months in asymptomatic patients free from clinical AF recurrencesRe-ablations were performed in 7 of 70 (10%) patients in the Study group and in 8 of 70 (11.4%) patients in the Control group
Yoshiga 2019ECG and Holter at 1, 3, 6, and 12 months; event monitoring in symptomatic patients112 months12 monthsIn all the patients the anti-arrhythmic drugs were discontinued after the procedureNot reported
Cordes 219Standardized telephone interview at 3 and 6 monthsb6 months6 monthsNot reportedNot reported
Miyamoto 2019ECG at 1 month, and then every 1–3 months; 24 h Holter at 3 and 12 month; event recorder in symptomatic patients112 months12 monthsDiscontinuation of antiarrhythmic drugs was recommended after the ablationNot reported
Koektuerk 2019Outpatient visit at 1, 3, 6, 12 months or earlier if symptomatic; 7-day Holter recording at 3 and 6 months and 24-h Holter later; telephone interview at the end of the follow-up116.4 ± 7.519.0 ± 8.6AADs continued for at least 3 months after ablation7 patients in the Study group and 14 in the Control group underwent repeated ablation
First author, year, referenceAssessment of the primary end pointStandard AF ablation endpointaMean follow-up time
AADs managementInformation on redo procedures
Study groupControl groupAll patients
Heeger 2016ECG and 24 h Holter at 3, 6, 12 months and in 6-month intervals thereafter; regular telephonic interviews, additional outpatient visits in symptomatic patients1848 ± 101 days849 ± 74 daysAADs continued for 3 monthsA total of 26/34 (76 %) patients suffering from atrial arrhythmia recurrences underwent a second ablation
Tebbenjohanns 2016Clinical visit and 24 h Holter at 3, 6, 12, and 18 months; external event recording for 4 weeks in symptomatic patients1458 ± 107 daysNot reportedNot reported
Chun 2017Visits at 3, 6, 12 months; ECG and 72-h Holter ECG; additional telephone interviews and event recording in symptomatic patients1372 (351–455) days378(361–483) daysAADs discontinued after the procedure. Resumption of AADs in case of AF relapse during blanking periodNot reported
Ekizler 2017ECG and 24 h Holter at 1, 3, 6 and 12 month, biannually thereafter112 ± 3 months after blanking period13 ± 3 months after blanking periodDiscontinued at the end of the blanking periodA redo procedure was performed in five patients in Study group and in six patients in Control group
Aryana 2017ECG during each follow-up visit; 2- to 4-week ambulatory electrocardiographic monitoring at 6 weeks, 3 and 6 months115 ± 2 month16 ± 3 monthAntiarrhythmic therapy was discontinued within 6 weeks of ablation35 patients (9.9%) in Study group and 63 patients (15.7%) in Control group underwent a repeat catheter ablation during the study period
Ströker 2018ECG, Holter at 1, 3, 6 month and every 6 month after or in case of symptoms118 ± 10 monthsAAD’s were discontinued 3 months post-ablation if no recurrenceA redo procedure was performed in 44 (66%) patients in the Control group and 37 (50%) patients in the Study group
Pott 2018Clinical assessment, echocardiography, ECG, and 7-day-Holter-monitoring at 1, 3, and 6 months and thereafter every 6 months1436 ± 184 daysAADs withheld either after the procedure or after blanking periodNot reported
Rottner 2018No information provided; paper focused on procedure safetyNANANANAAADs recommended to be continued for 3 monthsNA
Mortsell 2018Clinical visit at 3, 6, and 12 months; a 7 day Holter at 6 and 12 months1 (1 month blanking period)12 months12 monthsAntiarrhythmic drugs were withdrawn after 3 months in asymptomatic patients free from clinical AF recurrencesRe-ablations were performed in 7 of 70 (10%) patients in the Study group and in 8 of 70 (11.4%) patients in the Control group
Yoshiga 2019ECG and Holter at 1, 3, 6, and 12 months; event monitoring in symptomatic patients112 months12 monthsIn all the patients the anti-arrhythmic drugs were discontinued after the procedureNot reported
Cordes 219Standardized telephone interview at 3 and 6 monthsb6 months6 monthsNot reportedNot reported
Miyamoto 2019ECG at 1 month, and then every 1–3 months; 24 h Holter at 3 and 12 month; event recorder in symptomatic patients112 months12 monthsDiscontinuation of antiarrhythmic drugs was recommended after the ablationNot reported
Koektuerk 2019Outpatient visit at 1, 3, 6, 12 months or earlier if symptomatic; 7-day Holter recording at 3 and 6 months and 24-h Holter later; telephone interview at the end of the follow-up116.4 ± 7.519.0 ± 8.6AADs continued for at least 3 months after ablation7 patients in the Study group and 14 in the Control group underwent repeated ablation

AADs, antiarrhythmic drugs; AF, atrial fibrillation; AT, atrial tachycardia.

a

Defined as AF/AT episode lasting more than 30 s recorded after a 3 month blanking period.

b

Only 6 month follow-up results available.

Risk of bias

The overall quality of the included studies was good or moderate (Supplementary material online, Tables S1 and S2). All three RCTs lacked blinding of participants and/or personnel to the intervention, but otherwise had no other major source of bias. All observational studies had a control group and derived data directly from medical records. The NOS score varied between 8 and 9 which was consistently above the established cut-off for high-quality observational studies; in one case26 there was not sufficient data to conduct a proper classification.

Data synthesis

The 13 included studies comprised 3163 patients, predominantly males (64%) with paroxysmal AF (71.5%) and at low risk of stroke (Table 3). The pooled data revealed that the cryoballoon ablation single freeze strategy was as effective as the double freeze strategy in preventing recurrent atrial arrhythmia [OR 1.09; 95% CI (0.90–1.32), P = 0.386; I2 = 0%, Figure 2A] while associating with a significantly lower risk of overall adverse events [RR 0.72; 95% CI (0.53–0.98); P = 0.037; I2 = 0%; Figure 2B] and borderline lower risk of persistent PNP [RR 0.61; 95% CI (0.37–1.01); P = 0.056; I2 = 0%; Figure 3A]. These results were similar when calculated for RCTs and non-RCTs separately (Figure 2A and B). Furthermore, the duration of single freeze procedures appeared shorter with notably high heterogeneity (90 ± 27 min. vs. 121 ± 36 min.; P < 0.001; I2 = 96%; Figure 3B). There was no difference between both strategies in terms of fluoroscopy times or risk of transient PNP (Supplementary material online, Figures S1 and S2). Funnel plots did not reveal any significant publication bias for the main efficacy and safety endpoints (Supplementary material online, Figure S3).

Forest plots of primary efficacy and safety outcomes. (A) Relative risk of freedom from any atrial arrhythmia in the 12-month follow-up. Test for subgroup differences: Z = 0.437 (P = 0.662). Overall heterogeneity: I2 = 0.00%. (B) Relative risk of any adverse event associated in randomized and non-randomized studies. Test for subgroup differences: Z = −1.183 (P = 0.238). Overall heterogeneity: I2 = 0.00%. 95% CI, 95% confidence interval; RCT, randomized clinical trial; RR, relative risk.
Figure 2

Forest plots of primary efficacy and safety outcomes. (A) Relative risk of freedom from any atrial arrhythmia in the 12-month follow-up. Test for subgroup differences: Z = 0.437 (P = 0.662). Overall heterogeneity: I2 = 0.00%. (B) Relative risk of any adverse event associated in randomized and non-randomized studies. Test for subgroup differences: Z = −1.183 (P = 0.238). Overall heterogeneity: I2 = 0.00%. 95% CI, 95% confidence interval; RCT, randomized clinical trial; RR, relative risk.

Forest plots of most important secondary outcomes. (A) Relative risk of ablation-associated persistent phrenic nerve palsy in randomized and non-randomized studies. Test for subgroup differences: Z = −0.627 (P = 0.531). Overall heterogeneity: I2 = 0.00%. (B) Difference between double freeze and single freeze procedure duration in randomized and non-randomized studies (presented in minutes). Test for subgroup differences: Z = 1.957 (P = 0.050). Overall heterogeneity: I2 = 95.72%. 95% CI, 95% confidence interval; RCT, randomized clinical trial; RR, relative risk.
Figure 3

Forest plots of most important secondary outcomes. (A) Relative risk of ablation-associated persistent phrenic nerve palsy in randomized and non-randomized studies. Test for subgroup differences: Z = −0.627 (P = 0.531). Overall heterogeneity: I2 = 0.00%. (B) Difference between double freeze and single freeze procedure duration in randomized and non-randomized studies (presented in minutes). Test for subgroup differences: Z = 1.957 (P = 0.050). Overall heterogeneity: I2 = 95.72%. 95% CI, 95% confidence interval; RCT, randomized clinical trial; RR, relative risk.

Table 3

Baseline patient characteristics and procedural details

Overall
NValue
Age (years)316361.3 ± 10.8
Male sex31632025 (64.0)
Paroxysmal AF31632261 (71.5)
CHA2DS2-VASC score20691.3 ± 0.90
BMI (kg/m2)233928.0 ± 5.3
Left atrial diameter (mm)281842.7 ± 6.9
Single freezeRepeated freeze
NvalueNvalue
Fluoroscopy time (min)170016.6 ± 7.5139322.3 ± 9.7
Procedure time (min)170090.5 ± 27.01393122 ± 35.2
Periprocedural death17351 (0.06)14280 (0.00)
Periprocedural stroke or TIA17357 (0.40)14283 (0.21)
Tamponade17352 (0.12)14284 (0.28)
Atrio-oesophageal fistula17350 (0.00)14280 (0.0)
Overall
NValue
Age (years)316361.3 ± 10.8
Male sex31632025 (64.0)
Paroxysmal AF31632261 (71.5)
CHA2DS2-VASC score20691.3 ± 0.90
BMI (kg/m2)233928.0 ± 5.3
Left atrial diameter (mm)281842.7 ± 6.9
Single freezeRepeated freeze
NvalueNvalue
Fluoroscopy time (min)170016.6 ± 7.5139322.3 ± 9.7
Procedure time (min)170090.5 ± 27.01393122 ± 35.2
Periprocedural death17351 (0.06)14280 (0.00)
Periprocedural stroke or TIA17357 (0.40)14283 (0.21)
Tamponade17352 (0.12)14284 (0.28)
Atrio-oesophageal fistula17350 (0.00)14280 (0.0)
Table 3

Baseline patient characteristics and procedural details

Overall
NValue
Age (years)316361.3 ± 10.8
Male sex31632025 (64.0)
Paroxysmal AF31632261 (71.5)
CHA2DS2-VASC score20691.3 ± 0.90
BMI (kg/m2)233928.0 ± 5.3
Left atrial diameter (mm)281842.7 ± 6.9
Single freezeRepeated freeze
NvalueNvalue
Fluoroscopy time (min)170016.6 ± 7.5139322.3 ± 9.7
Procedure time (min)170090.5 ± 27.01393122 ± 35.2
Periprocedural death17351 (0.06)14280 (0.00)
Periprocedural stroke or TIA17357 (0.40)14283 (0.21)
Tamponade17352 (0.12)14284 (0.28)
Atrio-oesophageal fistula17350 (0.00)14280 (0.0)
Overall
NValue
Age (years)316361.3 ± 10.8
Male sex31632025 (64.0)
Paroxysmal AF31632261 (71.5)
CHA2DS2-VASC score20691.3 ± 0.90
BMI (kg/m2)233928.0 ± 5.3
Left atrial diameter (mm)281842.7 ± 6.9
Single freezeRepeated freeze
NvalueNvalue
Fluoroscopy time (min)170016.6 ± 7.5139322.3 ± 9.7
Procedure time (min)170090.5 ± 27.01393122 ± 35.2
Periprocedural death17351 (0.06)14280 (0.00)
Periprocedural stroke or TIA17357 (0.40)14283 (0.21)
Tamponade17352 (0.12)14284 (0.28)
Atrio-oesophageal fistula17350 (0.00)14280 (0.0)

Discussion

The main findings of this systematic review are: (i) a single freeze cryoballoon ablation is as effective as a bonus/double-freeze strategy, while (ii) associating with a lower risk of adverse events (with a strong trend for lower risk of persistent PNP) and (iii) a shorter procedure duration. These findings question the need for routine bonus freeze and suggest that a single freeze strategy should be the preferred technique for cryoballoon ablation of AF.

Cryoballoon ablation is widely considered as equivalent to RFA of AF in terms of efficacy.1,7,8 Recent publications indicate that AF ablation using the cryoballoon might become the first-line treatment of choice for patients with paroxysmal AF as it is more effective than antiarrhythmic drug therapy.14,15 However, in these multicentre RCTs, the cryoballoon ablation was routinely conducted using a double freeze strategy.

A growing body of evidence suggests the possibility of reducing the number of freezes to shorten the procedure and lower the risk of persistent PNP, a relatively common complication of cryoballoon ablation, without compromising the long-term effectiveness.11,12,21 This systematic review and meta-analysis summarizes the current knowledge on the effectiveness and safety of the cryoballoon ablation single freeze strategy. Included studies had good overall quality and described similar cryoballoon ablation techniques, had mandatory confirmation of pulmonary vein isolation, and used relatively homogeneous definitions of effectiveness, safety outcomes, and follow-up methods, which adds credence to the results. We highlight that patients included in these studies were relatively healthy, with a low average CHA2DS2-VASc score and predominantly with paroxysmal AF, and therefore it is unclear whether our results can be extrapolated to cohorts of persistent AF patients.

In terms of effectiveness, none of the individual studies reported a significant difference between strategies, yet the pooled analysis as well as the subgroup analysis based on study design confirmed that a single freeze strategy is as effective as using a routine bonus freeze (Figures 2 and3). Simultaneously, the single-freeze strategy was associated with a lower risk of adverse events, which was seen in both RCTs and non-RCTs and was largely driven by a borderline significant reduction in persistent PNPs (Figures 2B and3A). Given that this lower risk of adverse events was seen consistently across studies, with no significant heterogeneity between studies or evidence of significant publication bias, it reinforces the idea that adding a routine second freeze may be deleterious and should not be generally pursued. The single freeze procedures were also on average 20 min shorter, which may be useful to improve EP lab workflow without compromising patient safety and treatment efficacy.

This systematic review does not, however, provide any indications for the optimal technique for a single cryoballoon application. Included studies differed in terms of utilization of TTI guidance or length of cryoballoon applications—180/240 s in general (Table 1). However, due to occasionally short TTI, the total application times may have been even lower than 180 s in selected cases.22,25,31 It must be emphasized that studies using TTI-guided cryoablation had a formal protocol for quality control of the lesion formation which anticipated prolongation of the application based on prespecified TTI thresholds.10,11,21,22,25,26,31 Still, there was virtually no heterogeneity between studies in terms of effectiveness and safety regardless of the technique used during the single freeze strategy (Figure 2). While the influence of the length of cryoballoon application on the long-term effectiveness of AF ablation was clearly beyond the scope of this review, it should be bore in mind that there is evidence suggesting a significant association between longer freeze times and better rate of durable pulmonary vein isolation.32

Although the results of this systematic review and meta-analysis support the use of a single freeze strategy, our findings must be interpreted in the broader setting and should also take into account the data obtained at the time of re-do ablation in patients with recurrent AF.32–36 Published data suggest that gaps identified during re-do AF procedures after initial cryoballoon ablations tend to accumulate in the vicinity of the left atrial appendage/left superior pulmonary vein ridge, the superior aspect of right superior and inferior aspect of right inferior pulmonary veins. While we could argue that additional ‘bonus’ freezes in those areas may improve durability of cryoballoon-derived pulmonary vein isolation, the present meta-analysis suggests that additional applications in the vicinity of the right pulmonary veins should be discouraged and operators should instead take time to properly position the cryoballoon so it achieves optimal occlusion at the time of the first lesion. As far as the left pulmonary veins are concerned, the potential for fistula formation associated with very low temperatures and repeat freezes, especially in the inferior vein, has been previously reported.37 Our study does not support the need for routine bonus freeze in this area.

Limitations

This study has the typical limitations of systematic reviews and meta-analyses. Most of the included studies were observational and, although these were generally of good quality, they cannot replace large scale RCTs. However, our results were consistent across RCTs and non-RCTs, which adds robustness to our findings. In addition, where non-randomized studies are concerned, a learning curve effect cannot entirely be ruled out; some studies noted that a single-freeze strategy was adopted later than a double-freeze strategy. However, this notion is not supported by outcomes of included randomized trials which were all conducted by experienced investigators and generally reported similar effectiveness and safety to observational studies. High heterogeneity in terms of procedure duration (I2 = 96%) does not allow to consider the calculated difference of 20 min fully conclusive. However, all but one RCT and two observational studies reported that the time of single freeze procedure was significantly shorter than the double-freeze.

Conclusion

A single freeze strategy for cryoballoon ablation of AF is as effective as an empiric double (‘bonus’) freeze strategy while appearing safer and probably quicker (PROSPERO registration number CRD42020158696).

Supplementary material

Supplementary material is available at Europace online.

Conflict of interest: M.M.F. and M.P. received speaker/proctoring fees from Abbott Medical Poland, Medtronic Poland. S.B. received training grants from Biosense and Biotronik. D.G. holds shares in the company, which delivers consulting services to Abbott Medical Poland, Medtronic and Zoll. A.A. is a consultant for Farapulse Inc, Galaxy Medical Inc. A.A. received contracted research funds from Boston Scientific Inc, Farapulse Inc, Galaxy Medical Inc and Biosense&Webster. C.d.A. receive compensation for teaching purposes and proctoring from AF solutions, Medtronic, Abbott, Biotronik, Atricure, Boston Scientific, Microport, Acutus, Biosense Webster, Daiichi Sankyo, Schwarzer Cardiotek GmbH, and research grants on behalf of the centre from Biotronik, Medtronic, Abbott, Microport, Boston Scientific, Biosense Webster, Daiichi Sankyo, Atricure, Schwarzer Cardiotek GmbH Boehringer Ingeheim, BMS/Pfizer. D.A.L. received investigator-initiated educational grants from Bristol-Myers Squibb (BMS), speaker fees from Boehringer Ingeheim and BMS/Pfizer and consultancy from BMS/Pfizer, Boehringer Ingelheim, and Daiichi-Sankyo. S.Bo. is consultant for Medtronic, Boston Scientific, Microport and Zoll.

Acknowledgements

The production of this document is under the responsibility of the Scientific Initiatives Committee of the European Heart Rhythm Association: Serge Boveda (Chair), Giulio Conte (Co-Chair), Ante Anic, Sergio Barra, Julian K.R. Chun, Carlo de Asmundis, Nikolaos Dagres, Michal M. Farkowski, Jose Guerra, Konstantinos E. Iliodromitis, Kristine Jubele, Jedrzej Kosiuk, Eloi Marijon, Rui Providencia, Frits Prinzen.

References

1

Hindricks
G
,
Potpara
T
,
Dagres
N
,
Arbelo
E
,
Bax
JJ
,
Blomstrom-Lundqvist
C
 et al. ; ESC Scientific Document Group.
2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association of Cardio-Thoracic Surgery (EACTS)
.
Eur Heart J
 
2021
;
42
:
373
498
.

2

Barra
S
,
Baran
J
,
Narayanan
K
,
Boveda
S
,
Fynn
S
,
Heck
P
 et al.  
Association of catheter ablation for atrial fibrillation with mortality and stroke: a systematic review and meta-analysis
.
Int J Cardiol
 
2018
;
266
:
136
42
.

3

Marrouche
NF
,
Kheirkhahan
M
,
Brachmann
J.
 
Catheter ablation for atrial fibrillation with heart failure
.
N Engl J Med
 
2018
;
379
:
492
.

4

Calkins
H
,
Hindricks
G
,
Cappato
R
,
Kim
YH
,
Saad
EB
,
Aguinaga
L
 et al.  
2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation: executive summary
.
Europace
 
2018
;
20
:
157
208
.

5

Patel
N
,
Patel
K
,
Shenoy
A
,
Baker
WL
,
Makaryus
AN
,
El-Sherif
N.
 
Cryoballoon ablation for the treatment of atrial fibrillation: a meta-analysis
.
Curr Cardiol Rev
 
2019
;
15
:
230
8
.

6

Murray
MI
,
Arnold
A
,
Younis
M
,
Varghese
S
,
Zeiher
AM.
 
Cryoballoon versus radiofrequency ablation for paroxysmal atrial fibrillation: a meta-analysis of randomized controlled trials
.
Clin Res Cardiol
 
2018
;
107
:
658
69
.

7

Kuck
KH
,
Brugada
J
,
Furnkranz
A
,
Metzner
A
,
Ouyang
F
,
Chun
KR
 et al.  
Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation
.
N Engl J Med
 
2016
;
374
:
2235
45
.

8

Luik
A
,
Radzewitz
A
,
Kieser
M
,
Walter
M
,
Bramlage
P
,
Hormann
P
 et al.  
Cryoballoon versus open irrigated radiofrequency ablation in patients with paroxysmal atrial fibrillation: the prospective, randomized, controlled, noninferiority FreezeAF Study
.
Circulation
 
2015
;
132
:
1311
9
.

9

Koektuerk
B
,
Koektuerk
O
,
Yorgun
H
,
Guelker
JE
,
Turan
C
,
Gorr
E
 et al.  
Long-term impact of a bonus freeze on clinical outcome: analysis of effective and non-effective bonus freezes in cryoballoon ablation
.
PLoS One
 
2019
;
14
:
e0214231
.

10

Ströker
E
,
Kupics
K
,
de Asmundis
C
,
Mugnai
G
,
de Regibus
V
,
De Cocker
J
 et al.  
Atrial fibrillation ablation with the second generation cryoballoon: multicenter propensity score matched comparison between freezing strategies
.
Int J Cardiol
 
2018
;
253
:
78
81
.

11

Chun
KRJ
,
Stich
M
,
Fürnkranz
A
,
Bordignon
S
,
Perrotta
L
,
Dugo
D
 et al.  
Individualized cryoballoon energy pulmonary vein isolation guided by real-time pulmonary vein recordings, the randomized ICE-T trial
.
Heart Rhythm
 
2017
;
14
:
495
500
.

12

Miyamoto
K
,
Doi
A
,
Hasegawa
K
,
Morita
Y
,
Mishima
T
,
Suzuki
I
 et al.  
Multicenter study of the validity of additional freeze cycles for cryoballoon ablation in patients with paroxysmal atrial fibrillation
.
Circ Arrhythm Electrophysiol
 
2019
;
12
:
e006989
.

13

Andrade
JG
,
Champagne
J
,
Dubuc
M
,
Deyell
MW
,
Verma
A
,
Macle
L
 et al. ; For the CIRCA-DOSE Study Investigators.
Cryoballoon or radiofrequency ablation for atrial fibrillation assessed by continuous monitoring: a randomized clinical trial
.
Circulation
 
2019
;
140
:
1779
88
.

14

Andrade
JG
,
Wells
GA
,
Deyell
MW
,
Bennett
M
,
Essebag
V
,
Champagne
J
 et al.  
Cryoablation or drug therapy for initial treatment of atrial Ffbrillation
.
N Engl J Med
 
2021
;
384
:
305
15
.

15

Wazni
OM
,
Dandamudi
G
,
Sood
N
,
Hoyt
R
,
Tyler
J
,
Durrani
S
 et al.  
Cryoballoon ablation as initial therapy for atrial fibrillation
.
N Engl J Med
 
2021
;
384
:
316
24
.

16

Moher
D
,
Liberati
A
,
Tetzlaff
J
,
Altman
DG
; for the PRISMA Group.
Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement
.
BMJ
 
2009
;
339
:
b2535
.

17

Higgins
JP
,
Altman
DG
,
Gotzsche
PC
,
Juni
P
,
Moher
D
,
Oxman
AD
 et al. ; Cochrane Bias Methods Group.
The Cochrane Collaboration's tool for assessing risk of bias in randomised trials
.
BMJ
 
2011
;
343
:
d5928
.

18

Wan
X
,
Wang
W
,
Liu
J
,
Tong
T.
 
Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range
.
BMC Med Res Methodol
 
2014
;
14
:
135
.

19

Luo
D
,
Wan
X
,
Liu
J
,
Tong
T.
 
Optimally estimating the sample mean from the sample size, median, mid-range, and/or mid-quartile range
.
Stat Methods Med Res
 
2018
;
27
:
1785
805
.

20

Higgins
JP
,
Thompson
SG
,
Deeks
JJ
,
Altman
DG.
 
Measuring inconsistency in meta-analyses
.
BMJ
 
2003
;
327
:
557
60
.

21

Mörtsell
D
,
Malmborg
H
,
Lönnerholm
S
,
Jansson
V
,
Blomström-Lundqvist
C.
 
Acute and long-term efficacy and safety with a single cryoballoon application as compared with the standard dual application strategy: a prospective randomized study using the second-generation cryoballoon for pulmonary vein isolation in patients with symptomatic atrial fibrillation
.
Europace
 
2018
;
20
:
1598
605
.

22

Aryana
A
,
Kenigsberg
DN
,
Kowalski
M
,
Koo
CH
,
Lim
HW
,
O'Neill
PG
 et al.  
Verification of a novel atrial fibrillation cryoablation dosing algorithm guided by time-to-pulmonary vein isolation: results from the Cryo-DOSING Study (Cryoballoon-ablation DOSING Based on the Assessment of Time-to-Effect and Pulmonary Vein Isolation Guidance
).
Heart Rhythm
 
2017
;
14
:
1319
25
.

23

Ekizler
A
,
Aras
D
,
Cay
S
,
Ozeke
O
,
Ozcan
F
,
Topaloglu
S.
 
Bonus vs no bonus cryoballoon isolation for paroxysmal atrial fibrillation ablation
.
J Atr Fibrillation
 
2017
;
9
:
1513
.

24

Heeger
CH
,
Wissner
E
,
Wohlmuth
P
,
Mathew
S
,
Hayashi
K
,
Sohns
C
 et al.  
Bonus-freeze: benefit or risk? Two-year outcome and procedural comparison of a "bonus-freeze" and "no bonus-freeze" protocol using the second-generation cryoballoon for pulmonary vein isolation
.
Clin Res Cardiol
 
2016
;
105
:
774
82
.

25

Pott
A
,
Kraft
C
,
Stephan
T
,
Petscher
K
,
Rottbauer
W
,
Dahme
T.
 
Time-to-isolation guided titration of freeze duration in 3rd generation short-tip cryoballoon pulmonary vein isolation—comparable clinical outcome and shorter procedure duration
.
Int J Cardiol
 
2018
;
255
:
80
4
.

26

Rottner
L
,
Fink
T
,
Heeger
CH
,
Schlüter
M
,
Goldmann
B
,
Lemes
C
 et al.  
Is less more? Impact of different ablation protocols on periprocedural complications in second-generation cryoballoon based pulmonary vein isolation
.
Europace
 
2018
;
20
:
1459
67
.

27

Tebbenjohanns
J
,
Höfer
C
,
Bergmann
L
,
Dedroogh
M
,
Gaudin
D
,
Von Werder
A
 et al.  
Shortening of freezing cycles provides equal outcome to standard ablation procedure using second-generation 28 mm cryoballoon after 15-month follow-up
.
Europace
 
2016
;
18
:
206
10
.

28

Yoshiga
Y
,
Okamoto
T
,
Shimizu
A
,
Ueyama
T
,
Ono
M
,
Mito
T
 et al.  
Correlation between asymptomatic gastroesophageal excessive transmural injury after pulmonary vein isolation and a bonus freeze protocol using the second-generation 28-mm cryoballoon for paroxysmal atrial fibrillation
.
J Cardiol
 
2019
;
74
:
494
500
.

29

Ding
J
,
Xu
J
,
Ma
W
,
Chen
B
,
Yang
P
,
Qi
Y
 et al.  
A novel individualized strategy for cryoballoon catheter ablation in patients with paroxysmal atrial fibrillation
.
BMC Cardiovasc Disord
 
2019
;
19
:
299
.

30

Ferrero-De-Loma-Osorio
Á
,
Garciá-Fernández
A
,
Castillo-Castillo
J
,
Izquierdo-De-Francisco
M
,
Ibáñez-Criádo
A
,
Moreno-Arribas
J
 et al.  
Time-to-effect-based dosing strategy for cryoballoon ablation in patients with paroxysmal atrial fibrillation: results of the plusONE Multicenter Randomized Controlled Noninferiority Trial
.
Circ Arrhythm Electrophysiol
 
2017
;
10
:
e005318
.

31

Cordes
F
,
Ellermann
C
,
Dechering
DG
,
Frommeyer
G
,
Kochhäuser
S
,
Lange
PS
 et al.  
Time-to-isolation-guided cryoballoon ablation reduces oesophageal and mediastinal alterations detected by endoscopic ultrasound: results of the MADE-PVI trial
.
Europace
 
2019
;
21
:
1325
33
.

32

Chen
S
,
Schmidt
B
,
Bordignon
S
,
Perrotta
L
,
Bologna
F
,
Chun
KRJ.
 
Impact of cryoballoon freeze duration on long-term durability of pulmonary vein isolation: ICE Re-Map Study
.
JACC Clin Electrophysiol
 
2019
;
5
:
551
9
.

33

Buist
TJ
,
Adiyaman
A
,
Smit
JJJ
,
Ramdat Misier
AR
,
Elvan
A.
 
Arrhythmia-free survival and pulmonary vein reconnection patterns after second-generation cryoballoon and contact-force radiofrequency pulmonary vein isolation
.
Clin Res Cardiol
 
2018
;
107
:
498
506
.

34

Ciconte
G
,
Velagic
V
,
Mugnai
G
,
Saitoh
Y
,
Irfan
G
,
Hunuk
B
 et al.  
Electrophysiological findings following pulmonary vein isolation using radiofrequency catheter guided by contact-force and second-generation cryoballoon: lessons from repeat ablation procedures
.
Europace
 
2016
;
18
:
71
7
.

35

Koektuerk
B
,
Yorgun
H
,
Koektuerk
O
,
Turan
CH
,
Keskin
K
,
Schoett
M
 et al.  
Characterization of electrical reconnection following pulmonary vein isolation using first- and second-generation cryoballoon
.
Pacing Clin Electrophysiol
 
2016
;
39
:
434
42
.

36

Martins
RP
,
Galand
V
,
Cesari
O
,
Milhem
A
,
Behar
N
,
Auffret
V
 et al.  
The second generation cryoballoon has improved durable isolation of left but not right pulmonary veins: new insights from a multicentre study
.
Europace
 
2018
;
20
:
1115
21
.

37

John
RM
,
Kapur
S
,
Ellenbogen
KA
,
Koneru
JN.
 
Atrioesophageal fistula formation with cryoballoon ablation is most commonly related to the left inferior pulmonary vein
.
Heart Rhythm
 
2017
;
14
:
184
9
.

This article is published and distributed under the terms of the Oxford University Press, Standard Journals Publication Model (https://dbpia.nl.go.kr/journals/pages/open_access/funder_policies/chorus/standard_publication_model)

Supplementary data