Abstract

In recent years an extension of devices longevity has been obtained for implantable cardioverter-defibrillators (ICDs), including ICDs for cardiac resynchronization therapy (CRT-D) through improved battery chemistry and device technology and this implies important clinical benefits (reduced need for device replacements and associated complications, particularly infections), as well as economic benefits, in line with patient preferences and needs. From a clinical point of view, the availability of this improvement in technology allows to better tune the choice of the device to be implanted, taking into account that the reasons supporting the value of an extended device longevity as a clinical priority may differ according to the clinical setting (purely electrical diseases or left ventricular dysfunction/heart failure, respectively). From an economic point of view, extension of device longevity may have an important impact in reducing long-term costs of device therapy, with substantial daily savings in favour of devices with extended longevity, up to 30%, depending on clinical scenarios. In studies based on projections, an extension of device longevity allowed to calculate that the cost per day of ICDs may be substantially reduced, and this allows to overcome the frequent perception of ICD and CRT-D devices as treatments with unaffordable costs and to overturn the misconception that up-front costs are the only metric with which to value device treatments. In view of its clinical and economic value, device longevity should be a determining factor in device choice by physicians and healthcare commissioners and should be appropriately considered and valued in comparative tenders.

Introduction

Implantable cardioverter-defibrillator (ICD) therapy is effective for both primary and secondary prevention of sudden cardiac death due to ventricular tachycardia or fibrillation while cardiac resynchronization therapy with defibrillation (CRT-D) back-up is effective in improving patient symptoms and outcome in the setting of severe, moderate as well as mild heart failure. The longevity of ICD and CRT-D devices has rarely been discussed as a topic where technology may meet clinical needs in order to improve patient management, reduce device therapy complications, and reduce costs for the healthcare system. The aim of this review is to analyse the clinical and economic value of extending device longevity in the perspective of the patients and the health care system.

The outcome in patients with and without left ventricular dysfunction implanted with an implantable cardioverter-defibrillator and the need for devices with an extended longevity

Since the first relevant cohort studies in ICD patients dating from the mid-eighties, changes in ICD technology and in expanding opportunities for medical treatment, previously considered improbable, have been implemented in daily practice. In parallel to this evolution, indications were expanded from secondary to primary prevention of sudden cardiac death, so that nowadays the majority of ICDs are implanted for primary prevention of sudden cardiac death. The expected survival of patients currently implanted with ICD or CRT-D is of major relevance for assessing the impact of these device therapies at long term.

Whether today’s patients are in a similar overall health status is difficult to determine. In the early phase of ICD therapy, younger patients usually with coronary artery disease (CAD) and more preserved left ventricular ejection fraction (LVEF) were implanted for secondary prevention. The study from Echt et al.,1 for instance, enrolled 77 patients with mean age of 54 ± 13 years, 62% with CAD, and ejection fraction (EF) above 40% in 38%. In this population, 1-year survival was 90%. Similar patient characteristics were reported in a study from Winkle et al.2 published in 1989 (58 ± 12 years, 80% CAD, and EF >40% in 30%). In retrospect, it is noteworthy that the first randomized trials, both on primary prevention with LVEF <35% (MADIT)3 and on secondary prevention independently on LVEF (AVID)4 were published almost 10 years later. When compared with the situation in contemporary patients, heart failure medical treatment was far from being optimal (only 40% of patients were on beta-blocker, and 68% on angiotensin-converting enzyme inhibitors in MADIT). In this early phase comorbidities were not properly addressed. Over time, further randomized trials5–7 and registries8,9 evaluated the benefit of ICDs in primary prevention; however, mortality in the ICD group was never a key issue. Table 1 summarizes survival according to different points in time. As can be appreciated, there have been no major changes over the decades, in spite of all improvements in medical therapy. We can only speculate that this improvement is counterbalanced by ICD implantation in weaker patients with more severe comorbidities. Unfortunately, as will be outlined further below, reported ICD longevity is lower than patient survival at 5 years.9

Table 1

Overall survival rates in ICD and CRT-D patients

StudyYear of publicationPreventionAt 1 year (%)At 2 years (%)At 3 years (%)At 4 years (%)At 5 years (%)
ICD
 Echt et al.11985Sec92NANANANA
 Winkle et al.21989Sec9284828074
 MADIT I31996Inducible VTNANA84NA72
 AVID Investigators41997Sec898276NANA
 MADIT II52002PrimNA8578NANA
 SCD-HeFT62005PrimNA88847772
 Goldenberg et al.72008PrimNA71–93 according to risk scoresNANANA
 van Welsenes et al.82011SecNANA87NA77
PrimNANA82NA75
 Von Gunten et al.92016Prim + SecNANANA8380
CRT-D
 COMPANION10a2004Prim8875NANANA
 Thijssen et al.112012Prim + SecNA88827268
 Perkiomaki et al.122015Prim + SecNA9592NA89
StudyYear of publicationPreventionAt 1 year (%)At 2 years (%)At 3 years (%)At 4 years (%)At 5 years (%)
ICD
 Echt et al.11985Sec92NANANANA
 Winkle et al.21989Sec9284828074
 MADIT I31996Inducible VTNANA84NA72
 AVID Investigators41997Sec898276NANA
 MADIT II52002PrimNA8578NANA
 SCD-HeFT62005PrimNA88847772
 Goldenberg et al.72008PrimNA71–93 according to risk scoresNANANA
 van Welsenes et al.82011SecNANA87NA77
PrimNANA82NA75
 Von Gunten et al.92016Prim + SecNANANA8380
CRT-D
 COMPANION10a2004Prim8875NANANA
 Thijssen et al.112012Prim + SecNA88827268
 Perkiomaki et al.122015Prim + SecNA9592NA89

CRT-D, cardiac resynchronization therapy defibrillator; ICD, implantable cardioverter-defibrillator therapy; NA, not available; Prim, primary; Sec, secondary; VT, ventricular tachycardia.

a

CRT-D group.

Table 1

Overall survival rates in ICD and CRT-D patients

StudyYear of publicationPreventionAt 1 year (%)At 2 years (%)At 3 years (%)At 4 years (%)At 5 years (%)
ICD
 Echt et al.11985Sec92NANANANA
 Winkle et al.21989Sec9284828074
 MADIT I31996Inducible VTNANA84NA72
 AVID Investigators41997Sec898276NANA
 MADIT II52002PrimNA8578NANA
 SCD-HeFT62005PrimNA88847772
 Goldenberg et al.72008PrimNA71–93 according to risk scoresNANANA
 van Welsenes et al.82011SecNANA87NA77
PrimNANA82NA75
 Von Gunten et al.92016Prim + SecNANANA8380
CRT-D
 COMPANION10a2004Prim8875NANANA
 Thijssen et al.112012Prim + SecNA88827268
 Perkiomaki et al.122015Prim + SecNA9592NA89
StudyYear of publicationPreventionAt 1 year (%)At 2 years (%)At 3 years (%)At 4 years (%)At 5 years (%)
ICD
 Echt et al.11985Sec92NANANANA
 Winkle et al.21989Sec9284828074
 MADIT I31996Inducible VTNANA84NA72
 AVID Investigators41997Sec898276NANA
 MADIT II52002PrimNA8578NANA
 SCD-HeFT62005PrimNA88847772
 Goldenberg et al.72008PrimNA71–93 according to risk scoresNANANA
 van Welsenes et al.82011SecNANA87NA77
PrimNANA82NA75
 Von Gunten et al.92016Prim + SecNANANA8380
CRT-D
 COMPANION10a2004Prim8875NANANA
 Thijssen et al.112012Prim + SecNA88827268
 Perkiomaki et al.122015Prim + SecNA9592NA89

CRT-D, cardiac resynchronization therapy defibrillator; ICD, implantable cardioverter-defibrillator therapy; NA, not available; Prim, primary; Sec, secondary; VT, ventricular tachycardia.

a

CRT-D group.

Few data are available on survival in patients with a CRT-D. However, survival is better when compared with the past, in part due to the inclusion of lower risk patients, in part due to the positive effect of cardiac resynchronization therapy (CRT) on the functional status and on LVEF in the majority of patients. There is, however, a discrepancy between survival rates in randomized studies and in registry data (Table 1).

Data on patients with hypertrophic cardiomyopathy (HCM) are scarce and the number of patients studied rather small. In a nationwide Swedish study13 of 342 patients implanted with an ICD, 45 (13.2%) died during a mean follow-up period of 5.4 years. All-cause mortality was higher in HCM patients compared with the age and sex-matched Swedish general population [standardized mortality ratio 3.4, 95% confidence interval (CI) 2.4–4.5; P < 0.001]. Hypertrophic cardiomyopathy was the main cause of death in 76% of the cases, mainly because of progressive heart failure. Similar results were obtained in a study analysing 134 patients with an ICD followed in two tertiary referral clinics.14 Annualized cardiac mortality rate was 3.4% per year and was associated with New York Heart Association (NYHA) Class III or IV [hazard ratio (HR) 5.2, 95% CI 2.0–14; P = 0.002] and cardiac resynchronization therapy (HR 6.3, 95% CI 2.1–20; P = 0.02). Inappropriate ICD intervention occurred in 21 patients (3.7%/year) and in 20 patients device related complications were documented (3.6%/year). Thus, for HCM patients, ICDs almost eliminate premature arrhythmic death and result in a shift to heart failure as the cause of death in the majority of cases. As a consequence, in order to further reduce mortality in HCM patients improvement in prevention and management of heart failure and comorbidities appears as key factors.

Patients implanted with an ICD for inherited channelopathies or cardiomyopathies characterized by an arrhythmic risk, such as arrhythmogenic right ventricular cardiomyopathy (ARVC), had a better survival compared with those with structural heart disease and left ventricular dysfunction, and this finding is associated with a much younger age at implant.15,16 Of 176 Brugada patients followed for mean 84 ± 57 months, only 8 (4.5%) patients died.15 Of 312 patients with ARVC followed for mean 8.8 ± 7.3 years, only 5 (2%) patients died.17 Finally, a meta-analysis published in 2016, which include more than 4400 patients that were followed for mean 51 ± 38 months, reported survival rates of 96.5% for HCM patients, 97.7% for ARVC patients, 99.4% in long QT patients, and 99.9% in Brugada patients.16 It is obvious that especially these young patients with an extremely good survival are the first advocates for maximizing the longevity of ICDs, bearing in mind the risk of infection with every single device replacement (corresponding to 2.6% in 12 800 ICD and CRT-D procedures).18

It is clear from the expected survival of patients implanted with an ICD or a CRT-D device, according to currently accepted indications, that avoidance of any major mismatch between patient survival and device longevity is critical.

Recent evolution of device technology and improvement in projected and actual longevity

Replacement of cardiac implantable devices is associated with a notable risk of complications.19 Moreover, replacements involve incremental costs, and extending longevity has been shown to significantly improve cost-effectiveness of ICD and CRT-D therapy.20

Industry-independent studies on defibrillator longevity revealed large discrepancies in device lifespan depending on device type, generation, and manufacturer. Overall, dual-chamber ICDs exhibited shorter longevity than single-chamber ICDs, but lasted longer than CRT-Ds. Cardiac resynchronization therapy defibrillators are the most demanding antiarrhythmic devices in terms of battery consumption because of the need for continuous biventricular pacing and because pacing thresholds tend to be higher in the left ventricle.21

The results of previous studies comparing longevity of ICDs across manufacturers are frequently contradictory. This can be ascribed to the single-centre nature of the studies, the variety of population, the low number of devices in analysis and mainly to the heterogeneity of device technologies and generations. The results of the studies published before 201222–25 on devices manufactured before 2007 (most of which were no longer available at the time of publication) have been contradicted by more recent observations.9,26–28

Indeed, device longevity is based on battery technology, the efficiency of the electronic circuitry, the availability of specific algorithms for reducing battery consumption, and varies not only among manufacturers but also among generations of devices from the same manufacturer. Figure 1 reports the evolution of battery technology used by manufacturers over time.

Evolution along the time of usable battery capacity and battery chemistry used in CRT-D by manufacturers, on the basis of data reported in product manuals. The usable rather than the total battery capacities have been used for consistency in reporting across manufacturers. From Lau.29 Biotronik was the first manufacturer to adopt the Li/SVO-CFx chemistry. Both Boston Scientific and Biotronik adopted in recent years Li/MnO2 chemistry batteries, while St Jude Medical (Abbot), Sorin (LivaNova) and Medtronic. Adopted the Li/SVO-CFx chemistry. Overall, the usable battery capacity between 2006 and 2016 increased with all the manufacturers, although with some differences. CFx, carbon monofluoride; CRT-D, cardiac resynchronization therapy defibrillator; Li, Lithium; MnO2, manganese dioxide; SVO, silver vanadium oxide.
Figure 1

Evolution along the time of usable battery capacity and battery chemistry used in CRT-D by manufacturers, on the basis of data reported in product manuals. The usable rather than the total battery capacities have been used for consistency in reporting across manufacturers. From Lau.29 Biotronik was the first manufacturer to adopt the Li/SVO-CFx chemistry. Both Boston Scientific and Biotronik adopted in recent years Li/MnO2 chemistry batteries, while St Jude Medical (Abbot), Sorin (LivaNova) and Medtronic. Adopted the Li/SVO-CFx chemistry. Overall, the usable battery capacity between 2006 and 2016 increased with all the manufacturers, although with some differences. CFx, carbon monofluoride; CRT-D, cardiac resynchronization therapy defibrillator; Li, Lithium; MnO2, manganese dioxide; SVO, silver vanadium oxide.

Alam et al.26 determined longevity of CRT-Ds of three major manufacturers in 646 patients implanted between 2008 and 2010. They found important differences among manufacturers, with a 4-year survival rate ranging from 67% to 94%. These results have been confirmed by Landolina et al.27 The authors stratified 1726 consecutive CRT-D devices by manufacturers and time of market release (before or since 2007) and demonstrated an overall 5-year longevity of 54%, with significant differences among currently available CRT-D systems from different manufacturers. Moreover, they also demonstrated significant longevity improvements in recent devices from all manufacturers. In their study, battery longevity was independent of the burden of defibrillator therapy, confirming previous reports,22–24,26 but was significantly associated with the left ventricular output, confirming previous findings by Thijssen et al.24 and Alam et al.26 and with the unipolar left ventricular pacing configuration. Among the device characteristics, Lithium Manganese Dioxide cell chemistry and batteries from the manufacturer who adopted a 2.0 Ah cell capacity were protective factors against early device replacement.

More recently, Von Gunten et al.9 performed a similar analysis in patients who received devices at two institutions from 1994 to 2014. They included not only patients who received CRT-D, but also single-chamber and dual-chamber ICDs. They confirmed that recently all manufacturers have improved longevity of their devices, and that large differences in longevity still exist. More importantly, their data also confirmed that in spite of these improvements, survival of patients continued to be better than ICD longevity and not vice versa as should be the case.

In order to overcome recurring problems with transvenous leads, a subcutaneous ICD (S-ICD) has recently been developed. A Class IIa recommendation for S-ICD has been added to the most recent European Society of Cardiology (ESC) Guidelines for patients with ventricular arrhythmias.30 At present, no data are available with regard to the longevity of these new devices, except for the analysis of the patients enrolled in the European Regulatory Trial recently published by Theuns et al.31 They reported that compared with contemporary transvenous ICD systems, longevity of the first generation S-ICD was shorter, with a median longevity of 5.0 years, and not affected by the number of delivered shocks. Nonetheless, also for S-ICD a new generation of devices is currently on the market, designed to provide greater longevity. Moreover, the severity of possible complications associated with replacement of an S-ICD are expected to be lower than with a transvenous ICD.

Further confirmation of the importance of ICD longevity comes from the medical technology guidance published in March 2017 by the National Institute of Health and Care Excellence (NICE) in the framework of the Medical Technologies Evaluation Programme.32 The NICE committee evaluated the literature about currently available ICD battery technology, and concluded that there is good evidence to support the clinical benefit of longer battery life and the associated reduction in device replacements. According to the document, the available evidence suggests that the advantages of longer battery life have not been surpassed by other types of technical advances. The committee therefore concluded that adopting devices powered by batteries with 2 Ah capacity and manganese dioxide chemistry is likely to reduce the number of avoidable replacement procedures a patient may have to undergo, thereby offering improved outcomes for patients and potential savings to the National Health System. Moreover, the committee encouraged further studies that provide data on the battery life of cardiac devices in clinical practice.

Data and projections from the industry on devices longevities

Although real-world industry-independent longevity data are crucial for the assessment of the performance of implantable cardiac devices, the main limitation of all published studies on ICD and CRT-D longevity in clinical practice is the applicability of results to currently available ICD models, due to the high rate of release onto the market of newer device families.

Other information on the performance of devices is provided by the industry and comes from product performance reports. These reports are continuously updated with data from the analysis of returned devices and are periodically published.33–37 These data suffer from the same limitation of industry-independent longevity publications, i.e. they may no longer apply to newer device families. Moreover, cases of premature battery depletion may be underreported, and clinical events such as removal for infection, system upgrade, heart transplantation, or patient death are not considered and then censored and are estimated in analyses only by applying statistical adjustments and correction factors. Indeed, Alam et al.38 recently demonstrated significant discrepancies between actual battery longevity of CRT-Ds and data reported by manufacturers, with product performance reports highly overestimating longevity.

At the time of the release onto the market of a new ICD, the manufacturer provides projections on battery longevity, based on laboratory testing under controlled conditions. Apart from the possible mismatch between real-life conditions and accelerated battery life tests, data presented in the device manuals are difficult to interpret, due to the lack of homogeneity in the assumptions for longevity projections and the paucity of information about the impact of specific features on battery drain. As example, Table 2 summarizes the information reported in the device manual of some currently available devices from different manufacturers.

Table 2

Heterogeneity among the different manufacturers of ICD and CRT-D devices with regard to the assumptions for longevity projections and considerations on the information provided with regard to the impact of specific features on battery drain

Collection of electrogramsShocks and capacitor reformationsPacing optionsRemote monitoring
ABBOTT Quadra Assura MP™Assumption for service life projectionsEGM always ON3/yearDDD mode, 0.4 and 0.5 ms pulse width, RA/RV/LV amplitude at 2.5, 3.5, 5.0 V, Pacing impedances: 300, 500, 900 Ohm, at 60 and 70 min−1 lower rate, with and without atrial pacing
  • No info about remote monitoring assumptions for projection of longevity

  • Two hours of RF communication time at implant

Additional considerationsMultiPoint Pacing activation reduces longevity by 1–1.8 yearsa (four comparable scenarios with and without MultiPoint)
BIOTRONIK Intica 7 HF-T (QP)™Assumption for service life projections:EGM always ON4/year
  • DDD mode, 0.4 ms pulse width, 2.5 V amplitude, pacing impedance 500 Ohm, at 60 min−1 lower rate

  • Table with: pacing percentage from 0% to 100%, shocks/year from 4 to 20

Home Monitoring: ON, 1 device message each day and 24 IEGM online HD transmissions per year
Additional considerationsFour additional shocks per year reduce longevity by 0.9–1.7 years *
  • MultiPole Pacing activation reduces longevity by 0.3–0.7 yearsa (single pacing setting and multiple shock scenarios with and without MultiPole)

  • ‘MultiPole Pacing needs more power, which leads to various lengths of service time’

‘Consumption during implantation corresponds to approximately 7 days of service time and consumption during a 20-min follow-up corresponds to approximately 2 days’
BOSTON SCIENTIFIC RESONATE™ HF CRT-DAssumption for service life projections:EGM always ON2/year
  • DDDR mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 70 min−1 lower rate, sensors ON

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.0 to 3.5 V

  • Pacing impedances: 500, 700 Ohm

  • MultiSite Pacing: Off and On

  • Projections with and without monitoring under following conditions: daily device check on, quarterly scheduled remote follow ups, and other typical interrogations

  • Telemetry use for 2 h at implant and for 40 min annually for in-clinic checks

Additional considerationsAn additional maximum-energy shock reduces longevity by approximately 16 daysMultiSite Pacing activation reduces longevity by 1.1–2.3 yearsa (six comparable scenarios with and without MultiSite)
  • ‘When the latitude communicator is not used for the life of the device, longevity is increased by approximately 4 months’

  • ‘Five patient-initiated latitude communicator interrogations per week for a year reduces longevity by approximately 31 days’

  • ‘One hour of additional telemetry reduces longevity by approximately 7 days’

LIVANOVA PLATINIUM 4LV SonR CRT-D™Assumption for service life projections:EGM always ON2/year
  • DDD mode, 0.35 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate

  • Atrial pacing at 1%, 15%, 30%, and 100%

  • Table with: RA/RV amplitude from 2.5 to 4.5 V, LV amplitude from 2.5 to 4.5 V

  • Pacing impedances: 500, 600, 700 Ohm

  • Remote monitoring: ON, daily check, four follow-ups and five full alert reports per year

  • RF telemetry: ON, 120 min at implantation + 15 min at discharge + 15 min for in-clinic quarterly follow-ups

Additional considerationsTen additional shocks reduce longevity by approximately 0.5 yearsbIf Multipoint LV Pacing (MP) is set to ON, an additional LV pacing vector is activated. Consequently, the device longevity is reduced from 0.8 to 1.6 year
  • ‘An excessive number of remote follow-ups can have a non-negligible impact on device longevity’

  • Graph showing longevity as a function of yearly remote follow-ups (weekly transmissions reduce longevity by approximately 1 year)

  • ‘1 h of additional RF programming session reduces the device longevity from 1 to 2 weeks depending on the device functioning mode (no pacing, 100% pacing)’

MEDTRONIC CLARIA MRI™ QUAD CRT-DAssumption for service life projectionsPre-arrhythmia EGM storage On for a 6-month period over the entire life of the device2/year
  • DDD mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate and 70 min−1 atrial tracking

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.5 to 4.0 V

  • Pacing impedances: 500, 600 Ohm

  • Projections with monitoring under following conditions: a quarterly schedule of remote transmissions

  • Three hours of wireless telemetry during implant, 1 h of in-office wireless telemetry annually

Additional considerationsFull-time use of pre-arrhythmia EGM storage reduces projected service life by approximately two additional months per year or 16%Each additional full-energy charge due to therapy shock or device testing reduces projected service life by approximately 24 days
  • ‘Dual cathode LV pacing—it is not recommended to program a dual site LV pace polarity unless there is evidence that the patient is not responding to single site LV pacing. Long-term programming of a dual LV polarity will affect device longevity’

  • Projected increase in service life with AdaptivCRT: 0.2–0.8 yearsa

  • ‘Monthly transmissions over the life of the device reduce projected service life by 46 days or 2%

  • Weekly transmissions over the life of the device reduce projected service life by 223 days or 8%

  • Daily transmissions over the life of the device reduce projected service life by 992 days or 37%

  • A single additional transmission reduces projected service life by approximately 0.6 days or 0.02%’

‘Each additional hour of wireless telemetry use (in-office or implant) reduces the projected service life by approximately 5.7 days or 0.25%’
Collection of electrogramsShocks and capacitor reformationsPacing optionsRemote monitoring
ABBOTT Quadra Assura MP™Assumption for service life projectionsEGM always ON3/yearDDD mode, 0.4 and 0.5 ms pulse width, RA/RV/LV amplitude at 2.5, 3.5, 5.0 V, Pacing impedances: 300, 500, 900 Ohm, at 60 and 70 min−1 lower rate, with and without atrial pacing
  • No info about remote monitoring assumptions for projection of longevity

  • Two hours of RF communication time at implant

Additional considerationsMultiPoint Pacing activation reduces longevity by 1–1.8 yearsa (four comparable scenarios with and without MultiPoint)
BIOTRONIK Intica 7 HF-T (QP)™Assumption for service life projections:EGM always ON4/year
  • DDD mode, 0.4 ms pulse width, 2.5 V amplitude, pacing impedance 500 Ohm, at 60 min−1 lower rate

  • Table with: pacing percentage from 0% to 100%, shocks/year from 4 to 20

Home Monitoring: ON, 1 device message each day and 24 IEGM online HD transmissions per year
Additional considerationsFour additional shocks per year reduce longevity by 0.9–1.7 years *
  • MultiPole Pacing activation reduces longevity by 0.3–0.7 yearsa (single pacing setting and multiple shock scenarios with and without MultiPole)

  • ‘MultiPole Pacing needs more power, which leads to various lengths of service time’

‘Consumption during implantation corresponds to approximately 7 days of service time and consumption during a 20-min follow-up corresponds to approximately 2 days’
BOSTON SCIENTIFIC RESONATE™ HF CRT-DAssumption for service life projections:EGM always ON2/year
  • DDDR mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 70 min−1 lower rate, sensors ON

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.0 to 3.5 V

  • Pacing impedances: 500, 700 Ohm

  • MultiSite Pacing: Off and On

  • Projections with and without monitoring under following conditions: daily device check on, quarterly scheduled remote follow ups, and other typical interrogations

  • Telemetry use for 2 h at implant and for 40 min annually for in-clinic checks

Additional considerationsAn additional maximum-energy shock reduces longevity by approximately 16 daysMultiSite Pacing activation reduces longevity by 1.1–2.3 yearsa (six comparable scenarios with and without MultiSite)
  • ‘When the latitude communicator is not used for the life of the device, longevity is increased by approximately 4 months’

  • ‘Five patient-initiated latitude communicator interrogations per week for a year reduces longevity by approximately 31 days’

  • ‘One hour of additional telemetry reduces longevity by approximately 7 days’

LIVANOVA PLATINIUM 4LV SonR CRT-D™Assumption for service life projections:EGM always ON2/year
  • DDD mode, 0.35 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate

  • Atrial pacing at 1%, 15%, 30%, and 100%

  • Table with: RA/RV amplitude from 2.5 to 4.5 V, LV amplitude from 2.5 to 4.5 V

  • Pacing impedances: 500, 600, 700 Ohm

  • Remote monitoring: ON, daily check, four follow-ups and five full alert reports per year

  • RF telemetry: ON, 120 min at implantation + 15 min at discharge + 15 min for in-clinic quarterly follow-ups

Additional considerationsTen additional shocks reduce longevity by approximately 0.5 yearsbIf Multipoint LV Pacing (MP) is set to ON, an additional LV pacing vector is activated. Consequently, the device longevity is reduced from 0.8 to 1.6 year
  • ‘An excessive number of remote follow-ups can have a non-negligible impact on device longevity’

  • Graph showing longevity as a function of yearly remote follow-ups (weekly transmissions reduce longevity by approximately 1 year)

  • ‘1 h of additional RF programming session reduces the device longevity from 1 to 2 weeks depending on the device functioning mode (no pacing, 100% pacing)’

MEDTRONIC CLARIA MRI™ QUAD CRT-DAssumption for service life projectionsPre-arrhythmia EGM storage On for a 6-month period over the entire life of the device2/year
  • DDD mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate and 70 min−1 atrial tracking

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.5 to 4.0 V

  • Pacing impedances: 500, 600 Ohm

  • Projections with monitoring under following conditions: a quarterly schedule of remote transmissions

  • Three hours of wireless telemetry during implant, 1 h of in-office wireless telemetry annually

Additional considerationsFull-time use of pre-arrhythmia EGM storage reduces projected service life by approximately two additional months per year or 16%Each additional full-energy charge due to therapy shock or device testing reduces projected service life by approximately 24 days
  • ‘Dual cathode LV pacing—it is not recommended to program a dual site LV pace polarity unless there is evidence that the patient is not responding to single site LV pacing. Long-term programming of a dual LV polarity will affect device longevity’

  • Projected increase in service life with AdaptivCRT: 0.2–0.8 yearsa

  • ‘Monthly transmissions over the life of the device reduce projected service life by 46 days or 2%

  • Weekly transmissions over the life of the device reduce projected service life by 223 days or 8%

  • Daily transmissions over the life of the device reduce projected service life by 992 days or 37%

  • A single additional transmission reduces projected service life by approximately 0.6 days or 0.02%’

‘Each additional hour of wireless telemetry use (in-office or implant) reduces the projected service life by approximately 5.7 days or 0.25%’

CRT-D, cardiac resynchronization therapy defibrillator; EGM, electrograms; ICD, implantable cardioverter-defibrillator therapy; LV, left ventricle; RV, right ventricle.

a

Data derived from tables.

b

Data derived from graphs.

Table 2

Heterogeneity among the different manufacturers of ICD and CRT-D devices with regard to the assumptions for longevity projections and considerations on the information provided with regard to the impact of specific features on battery drain

Collection of electrogramsShocks and capacitor reformationsPacing optionsRemote monitoring
ABBOTT Quadra Assura MP™Assumption for service life projectionsEGM always ON3/yearDDD mode, 0.4 and 0.5 ms pulse width, RA/RV/LV amplitude at 2.5, 3.5, 5.0 V, Pacing impedances: 300, 500, 900 Ohm, at 60 and 70 min−1 lower rate, with and without atrial pacing
  • No info about remote monitoring assumptions for projection of longevity

  • Two hours of RF communication time at implant

Additional considerationsMultiPoint Pacing activation reduces longevity by 1–1.8 yearsa (four comparable scenarios with and without MultiPoint)
BIOTRONIK Intica 7 HF-T (QP)™Assumption for service life projections:EGM always ON4/year
  • DDD mode, 0.4 ms pulse width, 2.5 V amplitude, pacing impedance 500 Ohm, at 60 min−1 lower rate

  • Table with: pacing percentage from 0% to 100%, shocks/year from 4 to 20

Home Monitoring: ON, 1 device message each day and 24 IEGM online HD transmissions per year
Additional considerationsFour additional shocks per year reduce longevity by 0.9–1.7 years *
  • MultiPole Pacing activation reduces longevity by 0.3–0.7 yearsa (single pacing setting and multiple shock scenarios with and without MultiPole)

  • ‘MultiPole Pacing needs more power, which leads to various lengths of service time’

‘Consumption during implantation corresponds to approximately 7 days of service time and consumption during a 20-min follow-up corresponds to approximately 2 days’
BOSTON SCIENTIFIC RESONATE™ HF CRT-DAssumption for service life projections:EGM always ON2/year
  • DDDR mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 70 min−1 lower rate, sensors ON

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.0 to 3.5 V

  • Pacing impedances: 500, 700 Ohm

  • MultiSite Pacing: Off and On

  • Projections with and without monitoring under following conditions: daily device check on, quarterly scheduled remote follow ups, and other typical interrogations

  • Telemetry use for 2 h at implant and for 40 min annually for in-clinic checks

Additional considerationsAn additional maximum-energy shock reduces longevity by approximately 16 daysMultiSite Pacing activation reduces longevity by 1.1–2.3 yearsa (six comparable scenarios with and without MultiSite)
  • ‘When the latitude communicator is not used for the life of the device, longevity is increased by approximately 4 months’

  • ‘Five patient-initiated latitude communicator interrogations per week for a year reduces longevity by approximately 31 days’

  • ‘One hour of additional telemetry reduces longevity by approximately 7 days’

LIVANOVA PLATINIUM 4LV SonR CRT-D™Assumption for service life projections:EGM always ON2/year
  • DDD mode, 0.35 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate

  • Atrial pacing at 1%, 15%, 30%, and 100%

  • Table with: RA/RV amplitude from 2.5 to 4.5 V, LV amplitude from 2.5 to 4.5 V

  • Pacing impedances: 500, 600, 700 Ohm

  • Remote monitoring: ON, daily check, four follow-ups and five full alert reports per year

  • RF telemetry: ON, 120 min at implantation + 15 min at discharge + 15 min for in-clinic quarterly follow-ups

Additional considerationsTen additional shocks reduce longevity by approximately 0.5 yearsbIf Multipoint LV Pacing (MP) is set to ON, an additional LV pacing vector is activated. Consequently, the device longevity is reduced from 0.8 to 1.6 year
  • ‘An excessive number of remote follow-ups can have a non-negligible impact on device longevity’

  • Graph showing longevity as a function of yearly remote follow-ups (weekly transmissions reduce longevity by approximately 1 year)

  • ‘1 h of additional RF programming session reduces the device longevity from 1 to 2 weeks depending on the device functioning mode (no pacing, 100% pacing)’

MEDTRONIC CLARIA MRI™ QUAD CRT-DAssumption for service life projectionsPre-arrhythmia EGM storage On for a 6-month period over the entire life of the device2/year
  • DDD mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate and 70 min−1 atrial tracking

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.5 to 4.0 V

  • Pacing impedances: 500, 600 Ohm

  • Projections with monitoring under following conditions: a quarterly schedule of remote transmissions

  • Three hours of wireless telemetry during implant, 1 h of in-office wireless telemetry annually

Additional considerationsFull-time use of pre-arrhythmia EGM storage reduces projected service life by approximately two additional months per year or 16%Each additional full-energy charge due to therapy shock or device testing reduces projected service life by approximately 24 days
  • ‘Dual cathode LV pacing—it is not recommended to program a dual site LV pace polarity unless there is evidence that the patient is not responding to single site LV pacing. Long-term programming of a dual LV polarity will affect device longevity’

  • Projected increase in service life with AdaptivCRT: 0.2–0.8 yearsa

  • ‘Monthly transmissions over the life of the device reduce projected service life by 46 days or 2%

  • Weekly transmissions over the life of the device reduce projected service life by 223 days or 8%

  • Daily transmissions over the life of the device reduce projected service life by 992 days or 37%

  • A single additional transmission reduces projected service life by approximately 0.6 days or 0.02%’

‘Each additional hour of wireless telemetry use (in-office or implant) reduces the projected service life by approximately 5.7 days or 0.25%’
Collection of electrogramsShocks and capacitor reformationsPacing optionsRemote monitoring
ABBOTT Quadra Assura MP™Assumption for service life projectionsEGM always ON3/yearDDD mode, 0.4 and 0.5 ms pulse width, RA/RV/LV amplitude at 2.5, 3.5, 5.0 V, Pacing impedances: 300, 500, 900 Ohm, at 60 and 70 min−1 lower rate, with and without atrial pacing
  • No info about remote monitoring assumptions for projection of longevity

  • Two hours of RF communication time at implant

Additional considerationsMultiPoint Pacing activation reduces longevity by 1–1.8 yearsa (four comparable scenarios with and without MultiPoint)
BIOTRONIK Intica 7 HF-T (QP)™Assumption for service life projections:EGM always ON4/year
  • DDD mode, 0.4 ms pulse width, 2.5 V amplitude, pacing impedance 500 Ohm, at 60 min−1 lower rate

  • Table with: pacing percentage from 0% to 100%, shocks/year from 4 to 20

Home Monitoring: ON, 1 device message each day and 24 IEGM online HD transmissions per year
Additional considerationsFour additional shocks per year reduce longevity by 0.9–1.7 years *
  • MultiPole Pacing activation reduces longevity by 0.3–0.7 yearsa (single pacing setting and multiple shock scenarios with and without MultiPole)

  • ‘MultiPole Pacing needs more power, which leads to various lengths of service time’

‘Consumption during implantation corresponds to approximately 7 days of service time and consumption during a 20-min follow-up corresponds to approximately 2 days’
BOSTON SCIENTIFIC RESONATE™ HF CRT-DAssumption for service life projections:EGM always ON2/year
  • DDDR mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 70 min−1 lower rate, sensors ON

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.0 to 3.5 V

  • Pacing impedances: 500, 700 Ohm

  • MultiSite Pacing: Off and On

  • Projections with and without monitoring under following conditions: daily device check on, quarterly scheduled remote follow ups, and other typical interrogations

  • Telemetry use for 2 h at implant and for 40 min annually for in-clinic checks

Additional considerationsAn additional maximum-energy shock reduces longevity by approximately 16 daysMultiSite Pacing activation reduces longevity by 1.1–2.3 yearsa (six comparable scenarios with and without MultiSite)
  • ‘When the latitude communicator is not used for the life of the device, longevity is increased by approximately 4 months’

  • ‘Five patient-initiated latitude communicator interrogations per week for a year reduces longevity by approximately 31 days’

  • ‘One hour of additional telemetry reduces longevity by approximately 7 days’

LIVANOVA PLATINIUM 4LV SonR CRT-D™Assumption for service life projections:EGM always ON2/year
  • DDD mode, 0.35 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate

  • Atrial pacing at 1%, 15%, 30%, and 100%

  • Table with: RA/RV amplitude from 2.5 to 4.5 V, LV amplitude from 2.5 to 4.5 V

  • Pacing impedances: 500, 600, 700 Ohm

  • Remote monitoring: ON, daily check, four follow-ups and five full alert reports per year

  • RF telemetry: ON, 120 min at implantation + 15 min at discharge + 15 min for in-clinic quarterly follow-ups

Additional considerationsTen additional shocks reduce longevity by approximately 0.5 yearsbIf Multipoint LV Pacing (MP) is set to ON, an additional LV pacing vector is activated. Consequently, the device longevity is reduced from 0.8 to 1.6 year
  • ‘An excessive number of remote follow-ups can have a non-negligible impact on device longevity’

  • Graph showing longevity as a function of yearly remote follow-ups (weekly transmissions reduce longevity by approximately 1 year)

  • ‘1 h of additional RF programming session reduces the device longevity from 1 to 2 weeks depending on the device functioning mode (no pacing, 100% pacing)’

MEDTRONIC CLARIA MRI™ QUAD CRT-DAssumption for service life projectionsPre-arrhythmia EGM storage On for a 6-month period over the entire life of the device2/year
  • DDD mode, 0.4 ms pulse width, 15% atrial pacing with 100% biventricular pacing at 60 min−1 lower rate and 70 min−1 atrial tracking

  • Table with: RA/RV amplitude from 2.0 to 3.5 V, LV amplitude from 2.5 to 4.0 V

  • Pacing impedances: 500, 600 Ohm

  • Projections with monitoring under following conditions: a quarterly schedule of remote transmissions

  • Three hours of wireless telemetry during implant, 1 h of in-office wireless telemetry annually

Additional considerationsFull-time use of pre-arrhythmia EGM storage reduces projected service life by approximately two additional months per year or 16%Each additional full-energy charge due to therapy shock or device testing reduces projected service life by approximately 24 days
  • ‘Dual cathode LV pacing—it is not recommended to program a dual site LV pace polarity unless there is evidence that the patient is not responding to single site LV pacing. Long-term programming of a dual LV polarity will affect device longevity’

  • Projected increase in service life with AdaptivCRT: 0.2–0.8 yearsa

  • ‘Monthly transmissions over the life of the device reduce projected service life by 46 days or 2%

  • Weekly transmissions over the life of the device reduce projected service life by 223 days or 8%

  • Daily transmissions over the life of the device reduce projected service life by 992 days or 37%

  • A single additional transmission reduces projected service life by approximately 0.6 days or 0.02%’

‘Each additional hour of wireless telemetry use (in-office or implant) reduces the projected service life by approximately 5.7 days or 0.25%’

CRT-D, cardiac resynchronization therapy defibrillator; EGM, electrograms; ICD, implantable cardioverter-defibrillator therapy; LV, left ventricle; RV, right ventricle.

a

Data derived from tables.

b

Data derived from graphs.

In conclusion, an effort should be made to allow for technology comparison, and public health authorities should promote standardization in the assessment and reporting of device performance. For this purpose, since 2015 the French Health Technology Assessment body (Haute Autorité de Santé—HAS) has been requiring manufacturers to predict longevity with exactly the same settings.39 Moreover, the HAS has defined the minimal longevity requirements based on the specified settings: 7 years for single-chamber ICD, 6.5 years for dual-chamber ICD, 5.5 years for CRT-D.

At the same time, an effort should be made by the manufacturers in providing data as transparent and complete as possible, and the availability of tools for the automatic calculation of longevity projections under specific working conditions may be useful.40

Battery technology

Battery depletion is the dominant cause for ICD replacements.27,41 The high power demands of ICDs necessitate a battery, which is capable of delivering high current pulses of 2–3 A in order to rapidly charge the capacitors of the device.42 Furthermore, the battery must supply a constant low current to power the other functions of the ICD, i.e. data logging, telemetric communication, standard pacing, biventricular pacing for CRT, anti-tachycardia pacing, and the integration of innovative physiologic sensors. The capacity of a battery is measured in ampere-hour (A⋅h) that is the amount of electric charge that will support a current of 1 ampere for 1 h (1 A⋅h = 3, 600 coulombs). In theory, the capacity can be made as high as required but at the expense of a larger can size. In the last decade, ICD development has focused on enlarging the battery capacity, which has almost doubled without major increase in the can size. Currently available ICDs have capacities ranging from 1 to 2.0 A⋅h, and among them a larger battery capacity was shown to be associated with prolonged device longevity in clinical use.27

Electrochemically, a battery consists of an anode and a cathode physically separated. All batteries used today to power ICDs are lithium anode based, while the cathode is made of silver vanadium oxide (SVO), carbon monofluoride (CFx), SVO-CFx hybrid, or manganese dioxide (MnO2), with temporal differences in adoption of these types of batteries by each specific manufacturer, as shown in Figure 1.

Since the late 1980s, the power source chemistry used in ICDs has been the Li/SVO. It provides greater energy density than previous chemistries and over the years has undergone numerous improvements, becoming the most commonly used battery for ICDs. However, the discharge curve of Li/SVO is non-linear and makes it difficult to determine elective replacement. A similar behaviour is shown by the more recent SVO-CFx hybrid cells, whose cathode consists of CFx sandwiched between two layers of SVO material.29 Moreover, the internal cell impedance changes and this leads to variable capacitor charge times over the life of the ICD. By contrast, modern MnO2 batteries are able to maintain a high battery voltage and a relatively stable internal resistance during most of their discharge. This allowed to increase the level of usability from 70% of the total capacity obtained with standard SVO cells to 90% with MnO2 batteries. According to data collected in clinical practice,27 MnO2 batteries were shown to be independent protective factors against early depletion in clinical practice.

Factors affecting device longevity

In addition to the cell capacity, the current drain of the system represents a major factor affecting device longevity. Modern ICD and CRT-D devices are endowed with several diagnostic and therapeutic algorithms and features that may have a relevant impact on the device longevity. Table 2 summarizes the information reported in the device manual of currently available devices from different manufacturers. Table 3 summarizes the variables identified as predictors of early battery depletion in the literature.

Table 3

Variables identified as predictors of early battery depletion in the literature

Published articleDevice typeFactors associated with early battery depletion
Biffi et al.22
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Manufacturer

Schaer et al.23
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • More frequent capacitor charges

  • Manufacturer

Thijssen et al.24
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Percentage pacing

  • Pacing output

  • Manufacturer

Horlbeck et al.25
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

Alam et al.26
  • CRT-D

  • Manufacturer

  • LV pacing output

  • LV pacing impedance

Landolina et al.27
  • CRT-D

  • Manufacturer

  • Earlier device generations

  • Battery chemistry

  • LV pacing output

  • Unipolar LV pacing

von Gunten et al.9
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Manufacturer

Theuns et al.31
  • S-ICD

  • No association with the number of shocks

Alam et al.38
  • CRT-D

  • Manufacturer

Zanon et al.41
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

  • Percentage pacing

Boriani et al.43
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Number of shocks delivered and diverted

  • LV and RV pacing output

  • Percentage pacing

Published articleDevice typeFactors associated with early battery depletion
Biffi et al.22
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Manufacturer

Schaer et al.23
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • More frequent capacitor charges

  • Manufacturer

Thijssen et al.24
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Percentage pacing

  • Pacing output

  • Manufacturer

Horlbeck et al.25
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

Alam et al.26
  • CRT-D

  • Manufacturer

  • LV pacing output

  • LV pacing impedance

Landolina et al.27
  • CRT-D

  • Manufacturer

  • Earlier device generations

  • Battery chemistry

  • LV pacing output

  • Unipolar LV pacing

von Gunten et al.9
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Manufacturer

Theuns et al.31
  • S-ICD

  • No association with the number of shocks

Alam et al.38
  • CRT-D

  • Manufacturer

Zanon et al.41
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

  • Percentage pacing

Boriani et al.43
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Number of shocks delivered and diverted

  • LV and RV pacing output

  • Percentage pacing

CRT-D, cardiac resynchronization therapy defibrillator; ICD, implantable cardioverter-defibrillator therapy; S-ICD, subcutaneous ICD.

Table 3

Variables identified as predictors of early battery depletion in the literature

Published articleDevice typeFactors associated with early battery depletion
Biffi et al.22
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Manufacturer

Schaer et al.23
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • More frequent capacitor charges

  • Manufacturer

Thijssen et al.24
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Percentage pacing

  • Pacing output

  • Manufacturer

Horlbeck et al.25
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

Alam et al.26
  • CRT-D

  • Manufacturer

  • LV pacing output

  • LV pacing impedance

Landolina et al.27
  • CRT-D

  • Manufacturer

  • Earlier device generations

  • Battery chemistry

  • LV pacing output

  • Unipolar LV pacing

von Gunten et al.9
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Manufacturer

Theuns et al.31
  • S-ICD

  • No association with the number of shocks

Alam et al.38
  • CRT-D

  • Manufacturer

Zanon et al.41
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

  • Percentage pacing

Boriani et al.43
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Number of shocks delivered and diverted

  • LV and RV pacing output

  • Percentage pacing

Published articleDevice typeFactors associated with early battery depletion
Biffi et al.22
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Manufacturer

Schaer et al.23
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • More frequent capacitor charges

  • Manufacturer

Thijssen et al.24
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Percentage pacing

  • Pacing output

  • Manufacturer

Horlbeck et al.25
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Percentage pacing

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

Alam et al.26
  • CRT-D

  • Manufacturer

  • LV pacing output

  • LV pacing impedance

Landolina et al.27
  • CRT-D

  • Manufacturer

  • Earlier device generations

  • Battery chemistry

  • LV pacing output

  • Unipolar LV pacing

von Gunten et al.9
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Dual-chamber and CRT-D (vs. single-chamber ICD)

  • Earlier device generations

  • Manufacturer

Theuns et al.31
  • S-ICD

  • No association with the number of shocks

Alam et al.38
  • CRT-D

  • Manufacturer

Zanon et al.41
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • CRT-D (vs. dual-chamber and single-chamber ICD)

  • Manufacturer

  • Percentage pacing

Boriani et al.43
  • Single-chamber ICD

  • Dual-chamber ICD

  • CRT-D

  • Number of shocks delivered and diverted

  • LV and RV pacing output

  • Percentage pacing

CRT-D, cardiac resynchronization therapy defibrillator; ICD, implantable cardioverter-defibrillator therapy; S-ICD, subcutaneous ICD.

Background current

The background current needed to run the device’s electronic circuitry is typically in the μA range.29 Over the last years, more efficient circuitry has allowed to decrease energy consumption regardless of the extra functions incorporated into devices.29

Collection and storage of electrograms

The collection and storage of electrograms before tachycardia detection is intended to facilitate the subsequent clinical diagnosis of the arrhythmia. However, continuous storage of pre-episode electrograms, in addition to the episode summary data and the post-episode electrogram, may significantly reduce device longevity. According to the information reported in the device manual of currently available ICD and CRT-D from different manufacturers, electrograms collection and storage may reduce longevity as much as 16%.44

Capacitor reformation

In the high-voltage capacitors for ICDs, an oxide layer is formed on the anode surface to provide the dielectric for the capacitor. However, this oxide layer can degrade through chemical and physical processes over time, causing possible capacitor deformations, heating and even explosions.

To avoid degradation, an electric current is passed through the capacitor periodically (capacitor reformation process). In modern ICDs two reformations per year are required in absence of any shock therapy. However, this process requires energy and each reformation may consume up to 24 days of the total device longevity.44

Previous studies showed an association between the frequency of capacitor reformations and the time to battery depletion,23 and trials demonstrated that reducing capacitor charges prolonged battery longevity.45 Similarly, a recent analysis showed an association between battery drain and the number of shocks delivered and diverted.43 Therefore, increasing the detection rate or lengthening the monitoring delay before shock delivery could be effective programming strategies not only to reduce inappropriate therapies and improve outcome,46,47 but also to save energy.

Pacing

Among the factors associated with battery drain, pacing parameters play an important role but differences exist among device types. The impact of pacing on battery drain is negligible for single-chamber ICD, while it may be relevant for dual-chamber ICD and in particular for CRT-D, because of the need for continuous biventricular pacing and because pacing thresholds tend to be higher in the left ventricle.27,43 Indeed, when the required output for a pacing channel exceeds 2.5 V, voltage amplification of the battery voltage is needed and this is not always an energy-efficient process.29 Careful programming of pacing output is therefore crucial, and algorithms for automatic verification of myocardial capture and continuous optimization of pacing outputs have been developed. These are primarily intended to protect the patient against loss of pacing capture in case of rise in the chronic threshold, but also to prolong device life.48

For CRT devices, alternative pacing options have also been proposed that may have an impact on device longevity. The Medtronic Adaptive CRT algorithm periodically measures intrinsic conduction and delivers single-chamber left ventricular pacing in case of preserved atrioventricular conduction. According to the estimations of the manufacturer,44 the activation of this feature that essentially reduces the percentage of right ventricular pacing may potentially increase the device longevity by 0.2–0.8 years.

Quadripolar left ventricular leads and multisite pacing

Quadripolar left ventricular leads have become the standard of care in CRT. The availability of multiple pacing options allows the implanter to manage issues represented by phrenic nerve stimulation and high pacing capture thresholds. A quadripolar system allows to choose among up to 17 pacing vectors, increasing the chance (at the time of implantation and during follow-up) of finding a low left ventricular capture threshold, which may ensure effective CRT delivery and possibly prolong device longevity.

Recently, CRT devices with quadripolar left ventricular leads have been endowed with the possibility to deliver simultaneous or timed pacing from two electrodes of the left ventricular lead, in order to improve CRT response by simultaneously recruiting a larger volume of myocardium. Preliminary results seem positive49 and additional large studies are currently ongoing. However, the addition of a second left ventricular pacing channel may have a relevant impact on the longevity of the device and the expected clinical benefits from this pacing strategy must justify the premature ICD depletion and the consequent anticipated device replacement. According to the information reported in the device manual of CRT-D systems from different manufacturers, multi-site left ventricular pacing may reduce longevity by more than 1 year.50–52 This has been recently confirmed by Akerstrom et al.53 who estimated the impact of multi-site pacing activation on battery longevity. Based on a simulation using real-life device interrogation lead and programming data, they found that multi-site pacing activation significantly reduces battery longevity. Moreover, maximal interelectrode cathode spacing for the excitation of larger myocardial area could only be obtained by accepting configurations with high pacing thresholds. Therefore, they concluded that balancing the shorter time to generator replacement against the potential benefits of multi-site pacing is a rather difficult task.

Device interrogations and remote transmissions

The ICD exchanges data with a programmer at the time of implantation and in-office visits, and with a base station in the patient’s home if remote monitoring is adopted and enabled. The short-range (around 5 cm) communication performed by applying the programmer wand on the patient chest is based on inductive telemetry. This communication involves not only the exchange of data between ICD and programmer, but also the transmission of energy to the ICD for the communication. Therefore, this type of interrogation does not consume ICD battery energy. By contrast, longer-range communications with the programmer and with the remote monitoring base stations are based on radio frequency telemetry, in accordance with international standards [Industrial, Scientific, and Medical (ISM) or the Medical Implant Communication Service (MICS) standards]. Radio frequency transmission between the external device and the antenna housed in the plastic header of the ICD is extremely energy-consuming.29 Manufacturers adopted different schemes to initiate and complete the communication, in order to save as much as possible battery energy by minimizing the time the ICD must spend in an active state. The remote station can ‘wake up’ the ICD by sending signals, and in this case the ICD will spend energy to periodically detect possible calls. Alternatively, the ICD powers up periodically to transmit a notification and, if the remote station is in proximity and the communication is established, the transmission occurs. This second scheme may have a higher impact on device longevity and this is the reason why ICDs adopting it are typically programmed to attempt transmissions only for few hours per day, when the probability to be near the remote station is higher (e.g. during the night when the patient is in his bed).

In addition to periodical transmissions, the majority of remote monitoring systems allow to perform daily ICD checks to verify possible alert conditions without performing a complete data transmission. In general, this has a minor impact on device longevity.

Estimations of the possible impact of remote data transmissions on device longevity are reported in Table 2.

However, although remote transmission is one of the possible factors associated with battery drain, it has been shown that the appropriate use of remote monitoring may have a favourable impact on device longevity. The Effectiveness and Cost of ICDs Follow-up Schedule with Telecardiology (ECOST) trial showed that remote monitoring was associated with a significant saving in longevity, as it facilitated optimal device programming and consequent reduction in inappropriate shocks.45 More recently, Campana et al.54 showed that remote monitoring facilitated the identification of the proper time of ICD replacement, reducing at the same time the frequency of scheduled visits. Therefore, in clinical practice remote monitoring enables patients to be closely monitored, ensuring that replacement occurs before an unsafe status ensues, but also possibly decreasing the high proportion of devices that are replaced before elective replacement indicator notification.55

Clinical benefit of extending implantable cardioverter-defibrillator/cardiac resynchronization therapy longevity from a patient perspective

Extending device longevity is important for three principle reasons; (i) to reduce complications, (ii) to comply with patient preference, and (iii) to improve cost efficacy. Complications may be prevented by delaying or avoiding altogether a generator replacement. Whilst the surgical procedure itself imparts risk, the patients at the time of replacement are older and have more comorbidities than at initial device implantation. Actually, they have an increased risk of complications and an increased likelihood of dying of non-arrhythmic causes. The risk/benefit ratio of device replacement is unclear which is compounded by the lack of data supporting decision making at time of ICD/CRT-D replacement compared with the evidence for indications at implant. Thus, it is difficult to provide patients with accurate information at the time of consent under which circumstances their preference cannot be ignored and becomes a key issue. Patients must be supported if they decide not to proceed with an ICD/CRT-D replacement but must also understand the potential consequences.

Complications

The overall complications that characterize cardiac electronic implantable device (CIED) generator replacements are different compared with primary implantations because most complications at the time of the implant are related with the leads (dislodgement or perforation) while replacements are associated with an increased risk of infections.

Data from the literature on the occurrence of major complications in case of device replacement are shown in Table 4. The REPLACE registry prospectively studied patients undergoing pacemaker and ICD replacements in 72 US academic and private practice centres.19 Major complications were reported in 4.0% (95% CI 2.9–5.4) of 1031 patients without lead insertion or replacement including a 6 month infection rate of 1.4%. Overall complications were higher with ICD compared with pacemaker generator replacements.

Table 4

Data from the literature on the occurrence of major complications in case of device replacement

Published articleDevice/settingNumber of patientsFollow up (months)Major complications (%)
Poole et al.19PM, ICD, CRT—box change only10316/124.0
Poole et al.19PM, ICD, CRT—box and lead change7136/1215.3
Krahn et al.56ICD, CRT—box change only108145/74.3
Kirkfeldt et al.57PM, ICD, CRT—box change only11366/123.5
Gould et al.59ICD—box change only under advisory45112/129.1
Uslan et al.60ICD—box change only17446/121.3 infections only
Prutkin et al.61ICD, CRT—box change only43 6506/121.9 infections only
Published articleDevice/settingNumber of patientsFollow up (months)Major complications (%)
Poole et al.19PM, ICD, CRT—box change only10316/124.0
Poole et al.19PM, ICD, CRT—box and lead change7136/1215.3
Krahn et al.56ICD, CRT—box change only108145/74.3
Kirkfeldt et al.57PM, ICD, CRT—box change only11366/123.5
Gould et al.59ICD—box change only under advisory45112/129.1
Uslan et al.60ICD—box change only17446/121.3 infections only
Prutkin et al.61ICD, CRT—box change only43 6506/121.9 infections only

CRT, cardiac resynchronization therapy; ICD, implantable cardioverter-defibrillator therapy; PM, pacemaker.

Table 4

Data from the literature on the occurrence of major complications in case of device replacement

Published articleDevice/settingNumber of patientsFollow up (months)Major complications (%)
Poole et al.19PM, ICD, CRT—box change only10316/124.0
Poole et al.19PM, ICD, CRT—box and lead change7136/1215.3
Krahn et al.56ICD, CRT—box change only108145/74.3
Kirkfeldt et al.57PM, ICD, CRT—box change only11366/123.5
Gould et al.59ICD—box change only under advisory45112/129.1
Uslan et al.60ICD—box change only17446/121.3 infections only
Prutkin et al.61ICD, CRT—box change only43 6506/121.9 infections only
Published articleDevice/settingNumber of patientsFollow up (months)Major complications (%)
Poole et al.19PM, ICD, CRT—box change only10316/124.0
Poole et al.19PM, ICD, CRT—box and lead change7136/1215.3
Krahn et al.56ICD, CRT—box change only108145/74.3
Kirkfeldt et al.57PM, ICD, CRT—box change only11366/123.5
Gould et al.59ICD—box change only under advisory45112/129.1
Uslan et al.60ICD—box change only17446/121.3 infections only
Prutkin et al.61ICD, CRT—box change only43 6506/121.9 infections only

CRT, cardiac resynchronization therapy; ICD, implantable cardioverter-defibrillator therapy; PM, pacemaker.

In a further prospective, multicentre, population-based registry of 1081 lone ICD replacement patients at 18 centres in Canada, a 4.3% complication rate was found at 45 days.56 The most common complications were infection (2.1%), lead revision (3.2%), electrical storm (1.3%), and pulmonary oedema (1.2%). Risk factors associated with them were the use of antiarrhythmic therapy (adjusted HR 6.29, 95% CI 2.07–19.09), implanter volume (adjusted HR 10.4, CI 1.32–82.14 for <60/year vs. <120/year; P = 0.026), and Canadian Cardiovascular Society angina class (adjusted HR 3.00, CI 1.11–8.15 for class 2–4 vs. 0–1). A population-based cohort study57 in all Danish patients, who underwent a CIED procedure from May 2010 to April 2011, reported lower risk of any major complication after generator replacement (3.5%) compared with new implants (5.8%). Generator replacement procedures were associated with increased infection risk compared with first implants (1.5%, 95% CI 0.8–2.2; P = 0.001), confirming previous findings. A 4.05% median rate of major complications, with a range between 0.55% and 7.37%, was found in a recent systematic review of all available literature.58 Major complications included infections, haematoma, reoperation for other reasons, and stroke. Infections requiring antibiotic therapy and/or lead extraction had an incidence of 2.2%, whilst those for haematoma requiring evacuation and for reoperation due to any other reason (lead damage, repositioning of the device due to intolerable pain, and pocket erosion) were 0.7% and 2.0% respectively. Mortality ranged between 0% and 0.44%. Minor complications, such as incisional infections, haematoma or discomfort were reported in a further 3.5% of cases. Some of the studies found more complications when the replacement was done by a single operator or by operators with non-electrophysiology training.

Other studies have found that the number of previous procedures at the same pocket site was associated with an increase in complications.59 This emphasises the importance of extending device longevity. The same study also found that combined consultant and fellow operators were associated with a decrease in complications compared with a single operator.

Infections

Infections are the most common complication after ICD and CRT replacement with an incidence of approximately 1.5–2%. The 6-month infection rates found in the REPLACE registry were 1.4% (95% CI 0.7–2.3) and 1.1% (95% CI 0.5–2.2), respectively.60 All patients received preoperative intravenous antibiotics and 68.7% received post-operative systemic antibiotic therapy. Patients with infections were more likely to have had post-operative haematomas (22.7% vs. 0.98%, P = 0.002). Centres with infection rates >5% were more likely to use povidone-iodine for topical antisepsis and had lower implantation volume.

In the population-based cohort study on Danish patients where detailed information on complications after CIED procedures was collected by systematic review of all patient charts and was therefore particularly accurate, the highest infections rates were found in case of upgrade/lead revisions (1.9%, 95% CI 0.6–3.2 when compared with 0.6%, 95% CI 0.3–0.8 for normal implants).57

Evaluation of more than 200 000 ICD procedures in the US National Cardiovascular Data Registry has shown that ICD generator replacement results in a higher rate of infection compared with initial implant (1.9% vs. 1.6%, P < 0.001).61 This has been supported by the work of Borleffs et al.,62 who showed that occurrence of surgical re-intervention due to infections in ICD replacements was overall 2.5 (95% CI 1.6–3.7) times higher compared with first implanted ICDs. This study also demonstrated the increasing risk of sequential replacements, ranging from 1.5% for the first to 8.1% for the fourth implanted ICD. A one-third of ICD patients undergoing generator replacement or lead revision have an asymptomatic bacterial colonization of the generator pocket.63 After the revision, 7.5% of these patients develop a device infection with the same species of microorganism. The most common bacteria isolated was coagulase negative staphylococci (68%). Povidone-iodine irrigation of the subcutaneous pocket did not alter the rates of pocket infection after pacemaker/defibrillator implantation.64

Additional clinical considerations

There are no randomized controlled trials testing the benefits and true risk reduction of ICD/CRT-D replacements. There is however, observational data about all-cause mortality in a large registry of 111 826 patients, which showed 9.8%, 27.0% and 41.2% mortality rate at 1, 3, and 5 year follow-up, respectively.65

Other investigators have studied the incidence of appropriate therapies following ICD/CRT-D replacement. In these reports, patients were divided in two groups: those with an ongoing indication for ICD due to LVEF <35% and/or history or appropriate therapies by the device, and those not fulfilling these criteria. A retrospective review of 231 primary prevention patients found that 26% no longer had indications for an ICD at the time of replacement.66 These patients received appropriate ICD therapies at a significantly lower rate than patients with ongoing ICD indications following generator replacement (2.8% vs. 10.7% annually, P < 0.001). Similar findings were observed in another study of primary prevention ICD patients 3.3 years (median) after generator replacement.67 Patients with EF ≤35% were more likely to experience ICD therapy compared with those with EF >35% (12% vs. 5% per year; HR 3.57; P = 0.001).

Patients with CRT-D often demonstrate an improvement in left ventricular EF by the time of device generator replacement, and this makes unclear whether to replace the device with a CRT-D or a CRT-P. In the Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) trial, compared with pharmacological therapy alone, CRT-P reduced the incidence of the primary endpoint (death or all-cause hospitalization) by 34%, and CRT-D reduced the incidence by 40%, suggesting that CRT provides robust benefit, even in the absence of defibrillator therapy.10 Sebag et al.68 studied 107 patients with an initial primary prevention CRT-D and found that 37% had left ventricular EF ≥40%, and no history of appropriate ICD therapy at the time of device generator replacement. After 26.4 ± 14.4 months follow-up post-device replacement, 95% patients of these latter patients were free of appropriate therapy vs. 49 of 68 (72%) patients with ongoing ICD indication (P = 0.007). The only independent predictor of appropriate therapy after generator replacement was an ongoing ICD indication (HR 6.43; P = 0.01).

Patients preference

The option of no device replacement or replacement by a pacing only device should be considered in patients with significant comorbidity or with a short life expectancy. A recent study of explanted ICDs after death found that 35% of patients received ICD shocks within the last hour of life, but in only 13% of cases was death considered of an arrhythmic cause.69 This suggests that the delivery of shocks in this setting is unlikely to change the ultimate outcome. This has to be considered whenever the end-of-life of the patient period is foreseeable months in advance because of advancing non-cardiac conditions, such as such as dementia, progressive lung disease, cancer, or frailty/debility syndromes.

Device replacement is however, not only associated with complications and cost but with surgical stress and discomfort to the patient. It has thus been proposed that patient preference should be considered and discussed at the time of all ICD/CRT-D replacements.70

A detailed survey found that the vast majority (90.1% vs. 9.9%, P < 0.0001) of patients would prefer a long lasting device with a larger size rather than a smaller device with a shorter battery life.41 This preference was consistent even in thin (79.5%), female (89.2%), and elderly (>72 years) (90.5%) patients.

Pedersen et al.,71 found that most elderly patients were unaware that it was possible to decline device replacement. Furthermore, they had confused ideas regarding the risk/benefit balance with replacement and about the consequences of ending life with an ICD. However, the majority of ICD patients seemed to favour device deactivation at the end of life, primarily due to the wish for a dignified, pain free death. This indicates that patients think about deactivation but not avoiding ICD replacement near the end of life and might welcome discussing it with their physician.41 Healthcare professionals have an ethical and moral responsibility to discuss this with them.

Economic benefit of extending the longevity of implantable cardioverter-defibrillator/cardiac resynchronization therapy defibrillator devices

Healthcare systems need to face an increasing demand for costly medical interventions and this includes also medical devices. Clinicians, public health physicians, regulatory institutions, and policy makers need to find a reasonable balance between what is best for the individual patient and what society can realistically afford. Health technology assessment (HTA) has been proposed as a bridge between evidence and policy-making, with the aim to offer guidance to decision-making in a multidimensional approach, and with a circular pathway as depicted in Figure 2.72–74

The circular process of health technology assessment. From Boriani et al.72
Figure 2

The circular process of health technology assessment. From Boriani et al.72

The approach of HTA appears particularly appropriate in the case of CIEDs, which typically imply an asymmetrical, discontinuous distribution of costs, quite different from the pattern of costs due to drug therapies, usually associated with a progressive cumulative increase of resources consumption.72 In an economic approach, the budget impact of a given therapy must consider the upfront costs as well as the overall costs over the duration of therapy.

In the case of ICD and CRT-D devices the costs for the device and leads, the cost of hospitalization for the implant and potential complications, and the cost for device replacement (device purchase and cost of procedure) are all important determinants of the costs at long term, as shown in Figure 3. In consideration of the characteristics of current technology, of patient age at implant and of expected patient survival, an extension of current device longevity appears an important priority in the case of ICD and CRT-D devices, while a less pronounced impact is expected in the case of pacemakers implanted for bradycardia.75,76

Factors affecting the costs at long term of device therapy, with ICD of CRT-D systems, and possibility to favourably modulate the overall costs over time. CRT-D, cardiac resynchronization therapy defibrillator; ICD, implantable cardioverter-defibrillator therapy.
Figure 3

Factors affecting the costs at long term of device therapy, with ICD of CRT-D systems, and possibility to favourably modulate the overall costs over time. CRT-D, cardiac resynchronization therapy defibrillator; ICD, implantable cardioverter-defibrillator therapy.

The value of extending the longevity of ICDs and CIEDs devices can be appreciated by using economic models. A specific model has been proposed, the Longevity Model77 for calculating the costs associated with ICD and CRT-D device therapy, in a hospital perspective, according to varying device longevities (ranging from 4 to 15 years) in a ‘real-world’ clinical practice setting. The model included inputs corresponding to the costs of complications related to initial implantation and replacement of devices, as well as inputs related to life expectancy of different categories of potential device recipients (patients with or without left ventricular dysfunction and heart failure). The economic model assumed that, other patients’ characteristics being equal, the longer the device longevity, the lower the number of device replacements that patients will face during lifetime. The modelling study showed that, in a 15-year time window, the extension of device longevity (e.g. from 5 to 9 years for single-chamber ICDs) can have an important impact in reducing long-term costs of device therapy, with substantial daily savings in favour of devices with extended longevity, in the range of 29–34%, depending on clinical scenarios. The specific economic advantage of any increase in device longevity for specific patient characteristics, in the setting of single-chamber ICDs and CRT-D, respectively, can be detected in the Supplementary material online, Figures S1 and S2. The model allowed also to calculate that the cost per day of a single-chamber ICD with a longevity of 9 years is in the range of €3.8, while for a CRT-D device with a longevity of 7 years it may correspond to €5.5. These data clearly allow to have a different approach to device treatment and to overcome the frequent perception of ICD and CRT-D devices as treatments with unaffordable costs, higher than costs of other commonly accepted treatments.22,78,79 Conversely, these data could help to overturn the misconception that up-front costs are the only metric with which to value device treatments.

All the projections on longevity of devices and related costs are based on modelling and obviously have some limitations, but data from large datasets appear to confirm the reliability of these estimates based on device battery consumption.43 National registries play a major role in providing appropriate feedback on actual longevity of implanted devices, with appropriate comparisons between predicted longevities and actual performances.21,80,81

In a real-world analysis of around 1800 CRT-D implants published in 201681 the need for device replacements at 6 years was reduced by use of most recent generation devices from 83% to 68%, and this had a major impact on costs since the need for replacement increased total therapy costs by more than 50% over the initial implantation cost for hospitals and by more than 30% for the healthcare system. In consideration of the differences in battery chemistry adopted by different devices, the maximum difference in therapy costs between manufacturers was 40% for hospitals and 19% for the healthcare system.

Furthermore, a national study based on real world data from Sweden and data from the Swedish Pacemaker and ICD registry, based on data collected between 2008 and 2010 showed at base-case analysis that up to 603 replacements and up to SEK 60.4 million cumulative-associated costs (around €6.04 million) could be avoided over 6 years by using devices with extended battery life.82 The assumption at the basis of this analysis and the methods for cost calculation are shown in Supplementary material online, Table S1 and Figure S3. The pattern of savings over time suggests that savings are modest initially but they increase rapidly beginning in the third year of Follow-up with each year’s cumulative savings 2–3 times the previous year (Figure 4 for CRT-D). Evaluating CRT-D, ICD-VR, and ICD-DR devices together over a longer 10-year period, the sensitivity analysis showed 2820 fewer replacement procedures and associated cost savings of SEK 249.3 million (around €24.93 million) for all defibrillators with extended battery life.

Yearly costs of CRT-D device replacements* in the base-case analysis of data from the Swedish registry. From Gadler et al.82 All costs reported in 2015 Swedish krona (SEK) discounted at 3% per annum (1 SEK = around €0.10). *Costs of CRT-D device replacements include costs of device replacement and associated procedures, costs of managing complications related to replacement procedures, and costs of follow-up visits specific to the replacements. BSC, Boston Scientific Corporation; CRT-D, cardiac resynchronization therapy defibrillator; SJM, St. Jude Medical/Abbott; MDT, Medtronic; SEK, Swedish krona.
Figure 4

Yearly costs of CRT-D device replacements* in the base-case analysis of data from the Swedish registry. From Gadler et al.82 All costs reported in 2015 Swedish krona (SEK) discounted at 3% per annum (1 SEK = around €0.10). *Costs of CRT-D device replacements include costs of device replacement and associated procedures, costs of managing complications related to replacement procedures, and costs of follow-up visits specific to the replacements. BSC, Boston Scientific Corporation; CRT-D, cardiac resynchronization therapy defibrillator; SJM, St. Jude Medical/Abbott; MDT, Medtronic; SEK, Swedish krona.

The economic impact of an improved longevity of CIEDs has also been assessed in a macro-economic perspective. NICE considered a cost modelling based on published data of CIEDs, and found that the price and lifespan of the CRT-D have the greatest effect on overall treatment costs.32 According to these projections, adoption of the most advanced technology in batteries of CRT-D devices (batteries with Li/MnO2 chemistry) would result in the perspective of the UK National Health Care system in a saving of around £6 million in the first 5 years.32

The marked reduction in the daily costs of device therapy related to extended device longevity should lead to consider device longevity as a determining factor in device choice by physicians and healthcare commissioners and should be appropriately considered and valued in comparative tenders.

Open issues and areas of research

An important open issue that limits any comparative analysis of data on projected devices longevities provided from the different manufacturers derives by the lack of a standard in the assumptions used for making projections. It is time to establish a standard in order to allow proper comparisons. This is an important limitation for comparative analysis in tenders, where device longevity should be appropriately evaluated and valued.

Moreover, it will be important to collect independent data from ‘real world’ registries in order to validate the data obtained by projections on ICD and CRT-D devices longevities.

By lengthening device longevity, it could be hypothesized that the evolution of ICDs and CRT-Ds in terms of therapeutic algorithms will not become available to patients for an extended time period, until the next device replacement. This is an issue that could be solved, in the future, by the possibility to update devices’ software thus making available the most advanced device technology in terms of software.

The use of devices with extended longevity is currently not valued by reimbursement practices,83 but a revision should be considered in view of available data.

We mainly discussed technical data related to transvenous ICD, with limited focus on S-ICDs, for whom, anyway, an additional improvement in device longevity is desirable.

Finally as an important weak point of ICD/CRT-D systems is represented by the leads, a parallel improvement of the long term performance of ICD and CRT-D leads when compared with devices has to be pursued in order to guarantee an extended reliability of the entire device system.

Conclusions

The importance of extending the longevity of ICDs is quite obvious in the setting of purely electrical diseases (i.e. without left ventricular dysfunction), a setting where patient survival is expected to be quite similar to that of the normal population, provided that sudden arrhythmic death is prevented by the ability of the ICD to terminate life-threatening ventricular tachyarrhythmias. However, the relevance of extending the longevity of ICDs and CRT-D devices is currently actual also in the setting of patients with left ventricular dysfunction and heart failure. Indeed, long-term outcome of heart failure patients has been recently improved when compared with previously established drug treatments84 or will even more improve with new agents under appropriate evaluation,85 even if the contribution of comorbidities has to be considered for all the unselected patients treated in ‘real world’ practice.86–88

Extension of device longevity may reduce the mismatch that existed in the past between patient survival and device service life. This lead to an increased need for device replacements, with increased risk of complications, particularly increased risk of infections. An extension of device longevity has been obtained by improved battery chemistry and device technology and allows to obtain both clinical and economic benefits, in line with patient preferences and needs. From a clinical point of view, the availability of this improvement in technology allows to better tune the choice of the device to be implanted, taking into account that the reasons supporting the value of an extended device longevity as a clinical priority may differ according to the clinical setting, as shown in Figure 5. From an economic point of view, extension of device longevity may have an important impact in reducing long-term costs of device therapy, with substantial daily savings in favour of devices with extended longevity, up to 30%, depending on clinical scenarios. In studies based on projections, an extension of device longevity allowed to calculate that the cost per day of a single-chamber ICD with a longevity of at least 9 years is in the range of €3.8, while for a CRT-D device with a longevity of at least 7 years it may correspond to €5.5. These values imply that we could consider a different approach to device treatment and to overcome the frequent perception of ICD and CRT-D devices as treatments with unaffordable costs, higher than the costs of other commonly accepted treatments and to overturn the misconception that up-front costs are the only metric with which to value device treatments. In view of its clinical and economic value, device longevity should be a determining factor in device choice by physicians and healthcare commissioners and should be appropriately considered and valued in comparative tenders.

Importance of an individualized approach to device choice, taking into account that the reasons supporting the value of an extended device longevity as a clinical priority may differ according to the clinical setting, being maximized in young patients with channelopathies. HF, heart failure; LV, left ventricular.
Figure 5

Importance of an individualized approach to device choice, taking into account that the reasons supporting the value of an extended device longevity as a clinical priority may differ according to the clinical setting, being maximized in young patients with channelopathies. HF, heart failure; LV, left ventricular.

Supplementary material

Supplementary material is available at Europace online.

Acknowledgements

The authors thank EHRA Scientific Documents Committee: Gregory Y.H. Lip, Laurent Fauchier, David Arnar, Carina Blomstrom-Lundqvist, Zbigniew Kalarus, Gulmira Kudaiberdieva, Georges H. Mairesse, Tatjana Potpara, Irina Savelieva, Jesper Hastrup Svendsen, Vassil B. Traykov.

Funding

This review was prepared by a writing group that was invited on behalf of the European Heart Rhythm Association (EHRA) Education Committee. The document was supported with an educational grant from Boston Scientific. However, the content has not been influenced in any way by its sponsor.

Conflict of interest: G.B. reported speaker's fees from Biotronik, Boston and Medtronic; J.L.M. reported speaker's fees and honoraria from Abbott, Bayer, Biotronik, Biosense Webster, Boston, Cardiome, Daiichi Sankyo, Medtronic, Sorin and research grants from Boehringer-Ingelheim, Boston, Daiichi Sankyo, Medtronic, Pfizer; D.J.W. has received research grants and consultancy fees from Medtronic and Boston Scientific and educational fees from St Jude Medical; F.G. reported honoraria from Abbott/St.Jude Medical and Medtronic; B.S. reported to be part of Boston and Medtronic speaker’s bureau; M.L. reported to be part of speakers’ bureau appointment with Medtronic, LivaNova and Boston Scientific and an advisory board relationship with Medtronic.

References

1

Echt
DS
,
Armstrong
K
,
Schmidt
P
,
Oyer
PE
,
Stinson
EB
,
Winkle
RA.
Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator
.
Circulation
1985
;
71
:
289
96
.

2

Winkle
RA
,
Mead
RH
,
Ruder
MA
,
Gaudiani
VA
,
Smith
NA
,
Buch
WS
et al.
Long-term outcome with the automatic implantable cardioverter-defibrillator
.
J Am Coll Cardiol
1989
;
13
:
1353
61
.

3

Moss
AJ
,
Hall
WJ
,
Cannom
DS
,
Daubert
JP
,
Higgins
SL
,
Klein
H
et al.
Multicenter Automatic Defibrillator Implantation Trial Investigators
.
Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia
.
N Engl J Med
1996
;
335
:
1933
40
.

4

The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators
.
A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias
.
N Engl J Med
1997
;
337
:
1576
83
.

5

Moss
AJ
,
Zareba
W
,
Hall
WJ
,
Klein
H
,
Wilber
DJ
,
Cannom
DS
et al.
Multicenter Automatic Defibrillator Implantation Trial II Investigators
. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction
.
N Engl J Med
2002
;
346
:
877
83
.

6

Bardy
GH
,
Lee
KL
,
Mark
DB
,
Poole
JE
,
Packer
DL
,
Boineau
R
et al.
Sudden Cardiac Death in Heart Failure (SCD-HeFT) Investigators
. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure
.
N Engl J Med
2005
;
352
:
225
37
.

7

Goldenberg
I
,
Vyas
AK
,
Hall
WJ
,
Moss
AJ
,
Wang
H
,
He
H
et al.
MADIT-II Investigators
.
Risk stratification for primary implantation of a cardioverter-defibrillator in patients with ischemic left ventricular dysfunction
.
J Am Coll Cardiol
2008
;
51
:
288
96
.

8

van Welsenes
GH
,
van Rees
JB
,
Borleffs
JW
,
Cannegieter
SC
,
Bax
JJ
,
van Erven
L
et al.
Long-term follow-up of primary and secondary prevention implantable defibrillator patients
.
Europace
2011
;
13
:
389
94
.

9

Von Gunten
S
,
Schaer
BA
,
Yap
SC
,
Szili-Torok
T
,
Kühne
M
,
Sticherling
C
et al.
Longevity of implantable cardioverter defibrillators: a comparison among manufacturers and over time
.
Europace
2016
;
18
:
710
7
.

10

Bristow
MR
,
Saxon
LA
,
Boehmer
J
,
Krueger
S
,
Kass
DA
,
De Marco
T
et al.
Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Investigators
.
Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure
.
N Engl J Med
2004
;
350
:
2140
50
.

11

Thijssen
J
,
van Rees
JB
,
Venlet
J
,
Borleffs
CJ
,
Höke
U
,
Putter
H
et al.
The mode of death in implantable cardioverter-defibrillator and cardiac resynchronization therapy with defibrillator patients: results from routine clinical practice
.
Heart Rhythm
2012
;
9
:
1605
12
.

12

Perkiomaki
JS
,
Ruwald
AC
,
Kutyifa
V
,
Ruwald
MH
,
Mcnitt
S
,
Polonsky
B
et al.
MADIT-CRT Investigators
. Risk factors and the effect of cardiac resynchronization therapy on cardiac and non-cardiac mortality in MADIT-CRT
.
Europace
2015
;
17
:
1816
22
.

13

Magnusson
P
,
Gadler
F
,
Liv
P
,
Mörner
S.
Causes of death and mortality in hypertrophic cardiomyopathy patients with implantable cardioverter defibrillators
.
J Cardiovasc Med
2016
;
17
:
478
84
.

14

Vriesendorp
PA
,
Schinkel
AF
,
Van Cleemput
J
,
Willems
R
,
Jordaens
LJ
,
Theuns
DA
et al.
Implantable cardioverter-defibrillators in hypertrophic cardiomyopathy: patient outcomes, rate of appropriate and inappropriate interventions, and complications
.
Am Heart J
2013
;
166
:
496
502
.

15

Conte
G
,
Sieira
J
,
Ciconte
G
,
de Asmundis
C
,
Chierchia
GB
,
Baltogiannis
G
et al.
Implantable cardioverter-defibrillator therapy in Brugada syndrome
.
J Am Coll Cardiol
2015
;
65
:
879
88
.

16

Olde Nordkamp
LR
,
Postema
PG
,
Knops
RE
,
van Dijk
N
,
Limpens
J
,
Wilde
AA
et al.
Implantable cardioverter-defibrillator harm in young patients with inherited arrhythmia syndromes
.
Heart Rhythm
2016
;
13
:
443
54
.

17

Orgeron
GM
,
James
CA
,
Te Riele
A
,
Tichnell
C
,
Murray
B
,
Bhonsale
A
et al.
Implantable cardioverter-defibrillator therapy in arrhythmogenic right ventricular dysplasia/cardiomyopathy: predictors of appropriate therapy, outcomes, and complications
.
J Am Heart Assoc
2017
;
6
:
e006242.

18

Sohail
MR
,
Eby
EL
,
Ryan
MP
,
Gunnarsson
C
,
Wright
LA
,
Greenspon
AJ.
Incidence, treatment intensity, and incremental annual expenditures for patients experiencing a cardiac implantable electronic device infection
.
Circ Arrhythm Electrophysiol
2016
;
9
:
e003929.

19

Poole
JE
,
Gleva
MJ
,
Mela
T
,
Chung
MK
,
Uslan
DZ
,
Borge
R
et al.
Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry
.
Circulation
2010
;
122
:
1553
61
.

20

Sanders
GD
,
Hlatky
MA
,
Owens
DK.
Cost-effectiveness of implantable cardioverter-defibrillators
.
N Engl J Med
2005
;
353
:
1471
80
.

21

Landolina
M
,
Gasparini
M
,
Lunati
M
,
Iacopino
S
,
Boriani
G
,
Bonanno
C
et al.
Long-term complications related to biventricular defibrillator implantation: rate of surgical revisions and impact on survival: insights from the Italian Clinical Service Database
.
Circulation
2011
;
123
:
2526
35
.

22

Biffi
M
,
Ziacchi
M
,
Bertini
M
,
Sangiorgi
D
,
Corsini
D
,
Martignani
C
et al.
Longevity of implantable cardioverter-defibrillators: implications for clinical practice and health care systems
.
Europace
2008
;
10
:
1288
95
.

23

Schaer
BA
,
Koller
MT
,
Sticherling
C
,
Altmann
D
,
Joerg
L
,
Osswald
S.
Longevity of implantable cardioverter-defibrillators, influencing factors, and comparison to industry-projected longevity
.
Heart Rhythm
2009
;
6
:
1737
43
.

24

Thijssen
J
,
Borleffs
CJ
,
van Rees
JB
,
Man
S
,
de Bie
MK
,
Venlet
J
et al.
Implantable cardioverter-defibrillator longevity under clinical circumstances: an analysis according to device type, generation, and manufacturer
.
Heart Rhythm
2012
;
9
:
513
9
.

25

Horlbeck
FW
,
Mellert
F
,
Kreuz
J
,
Nickenig
G
,
Schwab
JO.
Real-world data on the lifespan of implantable cardioverter-defibrillators depending on manufacturers and the amount of ventricular pacing
.
J Cardiovasc Electrophysiol
2012
;
23
:
1336
42
.

26

Alam
MD
,
Munir
MB
,
Rattan
R
,
Flanigan
S
,
Adelstein
E
,
Jain
SK
et al.
Battery longevity in cardiac resynchronization therapy implantable cardioverter defibrillators
.
Europace
2014
;
16
:
246
51
.

27

Landolina
M
,
Curnis
A
,
Morani
G
,
Vado
A
,
Ammendola
E
,
D’Onofrio
A
et al.
Longevity of implantable cardioverter-defibrillators for cardiac resynchronization therapy in current clinical practice: an analysis according to influencing factors, device generation, and manufacturer
.
Europace
2015
;
17
:
1251
8
.

28

Evans
JM
,
Cleves
A
,
Morgan
H
,
Millar
L
,
Carolan-Rees
G.
ENDURALIFE-powered cardiac resynchronisation therapy defibrillator devices for treating heart failure: a NICE Medical Technology Guidance
.
Appl Health Econ Health Policy
2017
; doi:10.1007/s40258-017-0354-6.

29

Lau
EW.
Technologies for prolonging cardiac implantable electronic device longevity
.
Pacing Clin Electrophysiol
2017
;
40
:
75
96
.

30

Priori
SG
,
Blomström-Lundqvist
C
,
Mazzanti
A
,
Blom
N
,
Borggrefe
M
,
Camm
J
et al.
2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: the Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology (ESC)Endorsed by: Association for European Paediatric and Congenital Cardiology (AEPC)
.
Europace
2015
;
17
:
1601
87
.

31

Theuns
DA
,
Crozier
IG
,
Barr
CS
,
Hood
MA
,
Cappato
R
,
Knops
RE
et al.
Longevity of the subcutaneous implantable defibrillator: long-term follow-up of the European Regulatory Trial Cohort
.
Circ Arrhythm Electrophysiol
2015
;
8
:
1159
63
.

32

NICE Guidance. ENDURALIFE Powered CRT-D Devices for Treating Heart Failure Medical Technologies Guidance Published. nice.org.uk/guidance/mtg33 (17 March 2017, date last accessed).

38

Alam
MB
,
Munir
MB
,
Rattan
R
,
Adelstein
E
,
Jain
S
,
Saba
S.
Battery longevity from cardiac resynchronization therapy defibrillators: differences between manufacturers and discrepancies with published product performance reports
.
Europace
2017
;
19
:
421
4
.

41

Zanon
F
,
Martignani
C
,
Ammendola
E
,
Menardi
E
,
Narducci
ML
,
De Filippo
P
et al.
Device longevity in a contemporary cohort of ICD/CRT-D patients undergoing device replacement
.
J Cardiovasc Electrophysiol
2016
;
27
:
840
5
.

42

Bock
DC
,
Marschilok
AC
,
Takeuchi
KJ
,
Takeuchi
ES.
Batteries used to power implantable biomedical devices
.
Electrochim Acta
2012
;
84
:
155
64
.

43

Boriani
G
,
Ritter
P
,
Biffi
M
,
Ziacchi
M
,
Diemberger
I
,
Martignani
C
et al.
Battery drain in daily practice and medium-term projections on longevity of cardioverter defibrillators: an analysis from a remote monitoring database
.
Europace
2016
;
18
:
1366
73
.

44

Medtronic CLARIA MRI™ QUAD CRT-D SURESCAN™ DTMA2QQ Device Manual.

45

Guédon-Moreau
L
,
Lacroix
D
,
Sadoul
N
,
Clémenty
J
,
Kouakam
C
,
Hermida
JS
et al.
A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trial
.
Eur Heart J
2013
;
34
:
605
14
.

46

Moss
AJ
,
Schuger
C
,
Beck
CA
,
Brown
MW
,
Cannom
DS
,
Daubert
JP
et al.
Reduction in inappropriate therapy and mortality through ICD programming
.
N Engl J Med
2012
;
367
:
2275
83
.

47

Gasparini
M
,
Lunati
MG
,
Proclemer
A
,
Arenal
A
,
Kloppe
A
,
Martínez Ferrer
JB
et al.
Long detection programming in single chamber defibrillators reduces unnecessary therapies and mortality: the ADVANCEIII trial
.
JACC Clin Electrophysiol
2017
; doi:10.1016/j.jacep.2017.05.001.

48

Pierantozzi
A
,
Landolina
M
,
Agricola
T
,
Lunati
M
,
Pisanò
E
,
Lonardi
G
et al.
Automatic adjustment of stimulation output in resynchronization therapy: impact and effectiveness in clinical practice
.
Europace
2011
;
13
:
1311
8
.

49

Forleo
GB
,
Santini
L
,
Giammaria
M
,
Potenza
D
,
Curnis
A
,
Calabrese
V
et al.
Multipoint pacing via a quadripolar left-ventricular lead: preliminary results from the Italian registry on multipoint left-ventricular pacing in cardiac resynchronization therapy (IRON-MPP)
.
Europace
2017
;
19
:
1170
7
.

50

Boston Scientific RESONATE™ HF CRT-D Physician’s Technical Manual.

51

ABBOTT Quadra Assura MP™ Help Manual.

52

LivaNova PLATINIUM 4LV SonR CRT-D 1844 Implant Manual.

53

Akerström
F
,
Narváez
I
,
Puchol
A
,
Pachón
M
,
Martín-Sierra
C
,
Rodríguez-Mañero
M
et al.
Estimation of the effects of multipoint pacing on battery longevity in routine clinical practice
.
Europace
2017
; doi:10.1093/europace/eux209.

54

Campana
A
,
Giofrè
F
,
Stabile
G
,
Iori
M
,
La Rosa
C
,
Tomasi
C
et al.
Use of remote monitoring in the management of ICD end-of-life: data from the DECODE registry
.
Int J Cardiol
2016
;
221
:
430
2
.

55

Biffi
M
,
Ammendola
E
,
Santamaria
M
,
Campana
A
,
Iori
M
,
Baiocchi
C
et al.
Timing of ICD/CRT-D replacement in real-world practice: need for a behavioral change?
Eur Heart J
2015
;
36
(
Suppl 1
):
54
.

56

Krahn
AD
,
Lee
DS
,
Birnie
D
,
Healey
JS
,
Crystal
E
,
Dorian
P
et al.
Ontario ICDDI
.
Predictors of short-term complications after implantable cardioverter-defibrillator replacement: results from the Ontario ICD Database
.
Circ Arrhythm Electrophysiol
2011
;
4
:
136
42
.

57

Kirkfeldt
RE
,
Johansen
JB
,
Nohr
EA
,
Jørgensen
OD
,
Nielsen
JC.
Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark
.
Eur Heart J
2014
;
35
:
1186
94
.

58

Lewis
KB
,
Stacey
D
,
Carroll
SL
,
Boland
L
,
Sikora
L
,
Birnie
D.
Estimating the Risks and Benefits of implantable cardioverter defibrillator generator replacement: a systematic review
.
Pacing Clin Electrophysiol
2016
;
39
:
709
22
.

59

Gould
PA
,
Gula
LJ
,
Champagne
J
,
Healey
JS
,
Cameron
D
,
Simpson
C
et al.
Outcome of advisory implantable cardioverter-defibrillator replacement: one-year follow-up
.
Heart Rhythm
2008
;
5
:
1675
81
.

60

Uslan
DZ
,
Gleva
MJ
,
Warren
DK
,
Mela
T
,
Chung
MK
,
Gottipaty
V
et al.
Cardiovascular implantable electronic device replacement infections and prevention: results from the REPLACE Registry
.
Pacing Clin Electrophysiol
2012
;
35
:
81
7
.

61

Prutkin
JM
,
Reynolds
MR
,
Bao
H
,
Curtis
JP
,
Al-Khatib
SM
,
Aggarwal
S
et al.
Rates of and factors associated with infection in 200 909 Medicare implantable cardioverter-defibrillator implants: results from the National Cardiovascular Data Registry
.
Circulation
2014
;
130
:
1037
43
.

62

Borleffs
CJ
,
Thijssen
J
,
de Bie
MK
,
van Rees
JB
,
van Welsenes
GH
,
van Erven
L
et al.
Recurrent implantable cardioverter-defibrillator replacement is associated with an increasing risk of pocket-related complications
.
Pacing Clin Electrophysiol
2010
;
33
:
1013
9
.

63

Kleemann
T
,
Becker
T
,
Strauss
M
,
Dyck
N
,
Weisse
U
,
Saggau
W
et al.
Prevalence of bacterial colonization of generator pockets in implantable cardioverter defibrillator patients without signs of infection undergoing generator replacement or lead revision
.
Europace
2010
;
12
:
58
63
.

64

Lakkireddy
D
,
Valasareddi
S
,
Ryschon
K
,
Basarkodu
K
,
Rovang
K
,
Mohiuddin
SM
et al.
The impact of povidone-iodine pocket irrigation use on pacemaker and defibrillator infections
.
Pacing Clin Electrophysiol
2005
;
28
:
789
94
.

65

Kramer
DB
,
Kennedy
KF
,
Spertus
JA
,
Normand
SL
,
Noseworthy
PA
,
Buxton
AE
et al.
Mortality risk following replacement implantable cardioverter-defibrillator implantation at end of battery life: results from the NCDR
.
Heart Rhythm
2014
;
11
:
216
21
.

66

Kini
V
,
Soufi
MK
,
Deo
R
,
Epstein
AE
,
Bala
R
,
Riley
M
et al.
Appropriateness of primary prevention implantable cardioverter-defibrillators at the time of generator replacement: are indications still met?
J Am Coll Cardiol
2014
;
63
:
2388
94
.

67

Madhavan
M
,
Waks
JW
,
Friedman
PA
,
Kramer
DB
,
Buxton
AE
,
Noseworthy
PA
et al.
Outcomes after implantable cardioverter-defibrillator generator replacement for primary prevention of sudden cardiac death
.
Circ Arrhythm Electrophysiol
2016
;
9
:
e003283
.

68

Sebag
FA
,
Lellouche
N
,
Chen
Z
,
Tritar
A
,
O’neill
MD
,
Gill
J
et al.
Positive response to cardiac resynchronization therapy reduces arrhythmic events after elective generator change in patients with primary prevention CRT-D
.
J Cardiovasc Electrophysiol
2014
;
25
:
1368
75
.

69

Kinch Westerdahl
A
,
Sjoblom
J
,
Mattiasson
AC
,
Rosenqvist
M
,
Frykman
V.
Implantable cardioverter-defibrillator therapy before death: high risk for painful shocks at end of life
.
Circulation
2014
;
129
:
422
9
.

70

Kramer
DB
,
Buxton
AE
,
Zimetbaum
PJ.
Time for a change—a new approach to ICD replacement
.
N Engl J Med
2012
;
366
:
291
3
.

71

Pedersen
SS
,
Chaitsing
R
,
Szili-Torok
T
,
Jordaens
L
,
Theuns
DA.
Patients’ perspective on deactivation of the implantable cardioverter-defibrillator near the end of life
.
Am J Cardiol
2013
;
111
:
1443
7
.

72

Boriani
G
,
Maniadakis
N
,
Auricchio
A
,
Müller-Riemenschneider
F
,
Fattore
G
et al.
Health technology assessment in interventional electrophysiology and device therapy: a position paper of the European Heart Rhythm Association
.
Eur Heart J
2013
;
34
:
1869
74
.

73

Fattore
G
,
Maniadakis
N
,
Mantovani
LG
,
Boriani
G.
Health technology assessment: what is it? Current status and perspectives in the field of electrophysiology
.
Europace
2011
;
13
(
Suppl 2
):
ii49
53
.

74

Maniadakis
N
,
Vardas
P
,
Mantovani
LG
,
Fattore
G
,
Boriani
G.
Economic evaluation in cardiology
.
Europace
2011
;
13
(
Suppl 2
):
ii3
8
.

75

Diemberger
I
,
Biffi
M
,
Martignani
C
,
Boriani
G.
From lead management to implanted patient management: indications to lead extraction in pacemaker and cardioverter-defibrillator systems
.
Expert Rev Med Devices
2011
;
8
:
235
55
.

76

Diemberger
I
,
Mazzotti
A
,
Giulia
MB
,
Cristian
M
,
Matteo
M
,
Letizia
ZM
et al.
From lead management to implanted patient management: systematic review and meta-analysis of the last 15 years of experience in lead extraction
.
Expert Rev Med Devices
2013
;
10
:
551
73
.

77

Boriani
G
,
Braunschweig
F
,
Deharo
JC
,
Leyva
F
,
Lubinski
A
,
Lazzaro
C.
Impact of extending device longevity on the long-term costs of implantable cardioverter-defibrillator therapy: a modelling study with a 15-year time horizon
.
Europace
2013
;
15
:
1453
62
.

78

Boriani
G
,
Biffi
M
,
Martignani
C
,
Diemberger
I
,
Valzania
C
,
Bertini
M
et al.
Expenditure and value for money: the challenge of implantable cardioverter defibrillators
.
QJM
2009
;
102
:
349
56
.

79

Biffi
M
,
Ziacchi
M
,
Bertini
M
,
Gardini
B
,
Mazzotti
A
,
Massaro
G
et al.
How to truly value implantable cardioverter-defibrillators technology: up-front cost or daily cost?
Int J Technol Assess Health Care
2011
;
27
:
201
6
.

80

Gadler
F
,
Valzania
C
,
Linde
C.
Current use of implantable electrical devices in Sweden: data from the Swedish pacemaker and implantable cardioverter-defibrillator registry
.
Europace
2015
;
17
:
69
77
.

81

Landolina
M
,
Morani
G
,
Curnis
A
,
Vado
A
,
D’Onofrio
A
,
Bianchi
V
et al.
The economic impact of battery longevity in implantable cardioverter-defibrillators for cardiac resynchronization therapy: the hospital and healthcare system perspectives
.
Europace
2017
;
19
:
1349
56
.

82

Gadler
F
,
Ding
Y
,
Verin
N
,
Bergius
M
,
Miller
JD
,
Lenhart
GM
et al.
Economic impact of longer battery life of cardiac resynchronization therapy defibrillators in Sweden
.
Clinicoecon Outcomes Res
2016
;
8
:
657
66
.

83

Boriani
G
,
Burri
H
,
Mantovani
LG
,
Maniadakis
N
,
Leyva
F
,
Kautzner
J
et al.
Device therapy and hospital reimbursement practices across European countries: a heterogeneous scenario
.
Europace
2011
;
13
(
Suppl 2
):
ii59
65
.

84

McMurray
JJ
,
Packer
M
,
Desai
AS
,
Gong
J
,
Lefkowitz
MP
,
Rizkala
AR
et al.
PARADIGM-HF Investigators and Committees
.
Angiotensin-neprilysin inhibition versus enalapril in heart failure
.
N Engl J Med
2014
;
371
:
993
1004
.

85

Cowie
MR
,
Filippatos
GS
,
Alonso Garcia
MLA
,
Anker
SD
,
Baczynska
A
,
Bloomfield
DM
et al.
New medicinal products for chronic heart failure: advances in clinical trial design and efficacy assessment
.
Eur J Heart Fail
2017
;
19
:
718
27
.

86

Boriani
G
,
Malavasi
VL.
Patient outcome after implant of a cardioverter defibrillator in the ‘real world’: the key role of co-morbidities
.
Eur J Heart Fail
2017
;
19
:
387
90
.

87

Triposkiadis
F
,
Giamouzis
G
,
Parissis
J
,
Starling
RC
,
Boudoulas
H
,
Skoularigis
J
et al.
Reframing the association and significance of co-morbidities in heart failure
.
Eur J Heart Fail
2016
;
18
:
744
58
.

88

Boriani
G
,
Malavasi
VL.
Extending survival by reducing sudden death with implantable cardioverter-defibrillators: a challenging clinical issue in non-ischaemic and ischaemic cardiomyopathies
.
Eur J Heart Fail
2017
; doi:10.1002/ejhf.1080.

This article is published and distributed under the terms of the Oxford University Press, Standard Journals Publication Model (https://dbpia.nl.go.kr/journals/pages/open_access/funder_policies/chorus/standard_publication_model)

Supplementary data