Abstract

Aim

Left ventricular (LV) lead placement in the latest activated region is an important determinant of response to cardiac resynchronization therapy (CRT). We investigated the feasibility of coronary venous electroanatomic mapping (EAM) to guide LV lead placement to the latest activated region.

Methods and results

Twenty-five consecutive CRT candidates with left bundle-branch block underwent intra-procedural coronary venous EAM using EnSite NavX. A guidewire was used to map the coronary veins during intrinsic activation, and to test for phrenic nerve stimulation (PNS). The latest activated region, defined as the region with an electrical delay >75% of total QRS duration, was located anterolaterally in 18 (basal, n = 10; mid, n = 8) and inferolaterally in 6 (basal, n = 3; mid, n = 3). In one patient, identification of the latest activated region was impeded by limited coronary venous anatomy. In patients with >1 target vein (n = 12), the anatomically targeted inferolateral vein was rarely the vein with maximal electrical delay (n = 3). A concordant LV lead position was achieved in 18 of 25 patients. In six patients, this was hampered by PNS (n = 4), lead instability (n = 1), and coronary vein stenosis (n = 1).

Conclusion

Coronary venous EAM can be used intraprocedurally to guide LV lead placement to the latest activated region free of PNS. This approach especially contributes to optimization of LV lead electrical delay in patients with multiple target veins. Conventional anatomical LV lead placement strategy does not target the vein with maximal electrical delay in many of these patients.

What's new?

  • We describe a novel method of coronary venous electroanatomic mapping that utilizes the EnSite NavX system in combination with a guidewire that allows mapping and pacing of the coronary veins.

  • This technique can be used at the time of cardiac resynchronization therapy implantation to assess left ventricular (LV) electrical activation pattern and guide LV lead placement to the latest activated region.

  • The technique also enables comprehensive identification of sites that are free of phrenic nerve stimulation.

Introduction

Cardiac resynchronization therapy (CRT) reduces morbidity and mortality and reverses left ventricular (LV) remodelling in heart failure patients with LV systolic impairment and electrical dyssynchrony.1–3 Despite the striking effectiveness of CRT, a substantial proportion of apparently eligible patients fail to benefit.4 Part of this reduced benefit has been attributed to a suboptimal LV lead position.4 There is increasing evidence to suggest that positioning of the LV lead in the region of latest electrical activation provides superior clinical outcome.5,6 The conventional LV lead placement strategy involves an anatomical approach, targeting a coronary venous branch situated on the posterolateral wall.7 This strategy is based on the contention that the posterolateral wall is typically the latest activated site of the ventricle in patients with left bundle-branch block (LBBB). However, studies have shown a considerable variability in the ventricular activation pattern in LBBB, resulting in inter-individual variability in the optimal pacing site.8–10 Therefore, a more patient-specific physiological approach focused on achieving maximal LV lead electrical delay may improve CRT response. Electroanatomic mapping (EAM) is typically used in the electrophysiology lab to guide diagnostic or ablation procedures of cardiac arrhythmias. Recently, the technique has also been applied during CRT implantation to determine the electrical activation pattern of the coronary venous system.11,12 Final LV lead position, however, depends on several factors such as cardiac venous anatomy, performance and stability of the lead, and the absence of phrenic nerve stimulation (PNS). The feasibility of coronary venous EAM to guide LV lead placement to the latest activated region during CRT implantation remains to be assessed. The aim of the present study was to assess the feasibility of positioning the LV lead in the latest activated region guided by coronary venous EAM in patients undergoing CRT implantation.

Methods

Population

From December 2012 to May 2013, consecutive patients referred for CRT device implantation, with LV ejection fraction (LVEF) <35%, New York Heart Association (NYHA) functional class 2, 3, or ambulatory 4, and LBBB according to European Society of Cardiology (ESC) guidelines,13 were prospectively enroled. The study complied with the Declaration of Helsinki. Our locally appointed ethics committee approved the study protocol and waived the need for informed consent.

Electroanatomic mapping and left ventricular lead placement

All patients underwent intra-procedural coronary venous three-dimensional (3D) EAM using EnSite NavX (St Jude Medical). Prior to LV lead placement, an occlusive coronary venogram was recorded under fluoroscopy in right anterior oblique (RAO) and left anterior oblique (LAO) view for optimal visualization of the coronary veins. Subsequently, a 0.014 inch guidewire (Vision Wire, Biotronik SE & Co.KG) was inserted into the coronary sinus and connected to EnSite NavX along with the surface electrocardiogram (ECG). This guidewire has complete polyethylene isolation except for 15 mm at the distal ‘J’-shaped tip and 30 mm at the proximal end permitting unipolar sensing and pacing.14 The guidewire was used to map all target veins located on the anterolateral or inferolateral LV wall as defined by the American Heart Association (AHA) 17-segment heart model.15 A 3D electrical activation map was constructed during intrinsic ventricular activation. Electrical delay was measured in milliseconds from QRS onset on the surface ECG to the peak negative slope on the unipolar intra-cardiac electrogram (EGM) and expressed as a percentage of total QRS duration. All measurements were performed in EnSite NavX at a screen speed of 400 mm/s. The latest activated region of the LV was defined as the region with an electrical delay comprising >75% of total QRS duration. This definition was chosen because epicardial mapping via the coronary veins is limited by coronary venous anatomy, which means that some areas cannot be mapped because they do not contain any veins. Therefore, the latest activated LV region can only be identified using coronary venous mapping by relating the electrical activation time of the anatomical region to its timepoint within the QRS complex. Candidate pacing sites were tested for PNS by 10 V unipolar pacing on the tip of the guidewire and regions with PNS or absence thereof were annotated on the EAM using EnSite NavX. After the mapping procedure, the LV lead was connected to EnSite NavX for real-time visualization and navigation of the lead to the latest activated region free of PNS in the created EAM. Standard bipolar leads of various vendors were used. Figure 1 shows an example of a coronary venous EAM together with the results of PNS testing and the final LV lead position. The latest activated region and final LV lead position were classified according to the AHA 17-segment heart model by detailed evaluation of pre-implantation venograms and post-implantation LV lead fluoroscopy images (Figure 2). The final LV lead position was described as concordant if the lead was located in the same myocardial segment as the latest activated region.

(A) Coronary venous EAM of a patient during intrinsic ventricular activation together with the corresponding unipolar EGMs collected from the regions of earliest and latest activation. The CS and all side branches have been mapped and the anterolateral vein has been tested for PNS. Connecting the LV lead to EnSite NavX allowed real-time visualization and navigation of the lead to the latest activated region in the electroanatomic map. (B) Corresponding coronary venogram. (C) X-ray of final LV lead placement. AIV, anterior inter-ventricular vein; ALV, anterolateral vein; CS, coronary sinus; IIV, inferior inter-ventricular vein; ILV, inferolateral vein; RV, right ventricular; RA, right atrial; Red dot, PNS; Green dot, no PNS.
Figure 1

(A) Coronary venous EAM of a patient during intrinsic ventricular activation together with the corresponding unipolar EGMs collected from the regions of earliest and latest activation. The CS and all side branches have been mapped and the anterolateral vein has been tested for PNS. Connecting the LV lead to EnSite NavX allowed real-time visualization and navigation of the lead to the latest activated region in the electroanatomic map. (B) Corresponding coronary venogram. (C) X-ray of final LV lead placement. AIV, anterior inter-ventricular vein; ALV, anterolateral vein; CS, coronary sinus; IIV, inferior inter-ventricular vein; ILV, inferolateral vein; RV, right ventricular; RA, right atrial; Red dot, PNS; Green dot, no PNS.

Angiographic classification of the latest activated region and final LV lead position. (A) The AHA standardized 17-segment heart model, adapted from Cerqueira et al.15 (B) Coronary venogram in RAO. This view, representative of the long axis of the heart, was used to classify the latest activated region and final LV lead position into basal, mid-ventricular, and apical segments. (C) Coronary venogram in LAO. This view, comparable with the short axis of the heart, was used to divide the LV wall into four equal parts; anterior, anterolateral, inferolateral, and inferior. AIV, anterior inter-ventricular vein; ALV, anterolateral vein; IIV, inferior inter-ventricular vein; ILV, inferolateral vein.
Figure 2

Angiographic classification of the latest activated region and final LV lead position. (A) The AHA standardized 17-segment heart model, adapted from Cerqueira et al.15 (B) Coronary venogram in RAO. This view, representative of the long axis of the heart, was used to classify the latest activated region and final LV lead position into basal, mid-ventricular, and apical segments. (C) Coronary venogram in LAO. This view, comparable with the short axis of the heart, was used to divide the LV wall into four equal parts; anterior, anterolateral, inferolateral, and inferior. AIV, anterior inter-ventricular vein; ALV, anterolateral vein; IIV, inferior inter-ventricular vein; ILV, inferolateral vein.

Statistical analysis

Continuous variables are expressed as mean ± SD. Categorical variables are expressed as observed number and percentage values. Continuous variables were compared using Mann–Whitney U test. Statistical significance was accepted at the 95% confidence interval (P < 0.05). Statistical analysis was performed using SPSS version 20.0 (SPSS Inc.) software.

Results

Patient characteristics

Twenty-five consecutive patients with LBBB referred for CRT device implantation were included in this study. The patient characteristics are described in Table 1.

Table 1

Patient characteristics (n = 25)

Age, years72 ± 6
Male gender (n, %)14 (56)
Ischaemic heart disease (n, %)8 (32)
NYHA functional class (n, %)
 II11 (44)
 III14 (56)
Echocardiography characteristics
 LVEF (%)27 ± 5
 LV end-diastolic diameter (mm)59 ± 4
 LV end-systolic diameter (mm)55 ± 5
 LV end-diastolic volume (mL)146 ± 30
 LV end-systolic volume (mL)107 ± 24
ECG characteristics
 LBBB (n, %)25 (100)
 QRS duration (ms)154 ± 15
Intrinsic rhythm (n, %)
 Sinus rhythm22 (88)
 Atrial fibrillation3 (12)
Treatment (n, %)
 Diuretics17 (68)
 ACE-i/ARB24 (96)
 Beta-blockers22 (88)
 Spironolactone15 (60)
 Digoxin1 (4)
 Amiodarone3 (12)
Age, years72 ± 6
Male gender (n, %)14 (56)
Ischaemic heart disease (n, %)8 (32)
NYHA functional class (n, %)
 II11 (44)
 III14 (56)
Echocardiography characteristics
 LVEF (%)27 ± 5
 LV end-diastolic diameter (mm)59 ± 4
 LV end-systolic diameter (mm)55 ± 5
 LV end-diastolic volume (mL)146 ± 30
 LV end-systolic volume (mL)107 ± 24
ECG characteristics
 LBBB (n, %)25 (100)
 QRS duration (ms)154 ± 15
Intrinsic rhythm (n, %)
 Sinus rhythm22 (88)
 Atrial fibrillation3 (12)
Treatment (n, %)
 Diuretics17 (68)
 ACE-i/ARB24 (96)
 Beta-blockers22 (88)
 Spironolactone15 (60)
 Digoxin1 (4)
 Amiodarone3 (12)

ACE-i, angiotensin-converting enzyme inhibitor; ARB, angiotensin receptor blocker.

Table 1

Patient characteristics (n = 25)

Age, years72 ± 6
Male gender (n, %)14 (56)
Ischaemic heart disease (n, %)8 (32)
NYHA functional class (n, %)
 II11 (44)
 III14 (56)
Echocardiography characteristics
 LVEF (%)27 ± 5
 LV end-diastolic diameter (mm)59 ± 4
 LV end-systolic diameter (mm)55 ± 5
 LV end-diastolic volume (mL)146 ± 30
 LV end-systolic volume (mL)107 ± 24
ECG characteristics
 LBBB (n, %)25 (100)
 QRS duration (ms)154 ± 15
Intrinsic rhythm (n, %)
 Sinus rhythm22 (88)
 Atrial fibrillation3 (12)
Treatment (n, %)
 Diuretics17 (68)
 ACE-i/ARB24 (96)
 Beta-blockers22 (88)
 Spironolactone15 (60)
 Digoxin1 (4)
 Amiodarone3 (12)
Age, years72 ± 6
Male gender (n, %)14 (56)
Ischaemic heart disease (n, %)8 (32)
NYHA functional class (n, %)
 II11 (44)
 III14 (56)
Echocardiography characteristics
 LVEF (%)27 ± 5
 LV end-diastolic diameter (mm)59 ± 4
 LV end-systolic diameter (mm)55 ± 5
 LV end-diastolic volume (mL)146 ± 30
 LV end-systolic volume (mL)107 ± 24
ECG characteristics
 LBBB (n, %)25 (100)
 QRS duration (ms)154 ± 15
Intrinsic rhythm (n, %)
 Sinus rhythm22 (88)
 Atrial fibrillation3 (12)
Treatment (n, %)
 Diuretics17 (68)
 ACE-i/ARB24 (96)
 Beta-blockers22 (88)
 Spironolactone15 (60)
 Digoxin1 (4)
 Amiodarone3 (12)

ACE-i, angiotensin-converting enzyme inhibitor; ARB, angiotensin receptor blocker.

Cardiac resynchronization therapy implantation

In 21 patients, a de novo CRT defibrillator was implanted. Two patients were upgraded from an implantable cardioverter-defibrillator to a CRT defibrillator and two patients received a CRT pacemaker. Left ventricular lead implantation was successful in all patients. There were no procedural complications.

Coronary venous electroanatomic mapping

In all patients, occlusive coronary venography revealed at least one target vein on the anterolateral or inferolateral LV wall. In 12 of 25 patients, two target veins were available. All target veins were successfully mapped. Three-dimensional electrical activation maps were generated from an average of 46 ± 28 unique anatomical points. Mapping time was 19 ± 6 min and fluoroscopy time during the entire procedure was 20 ± 6 min. Figure 3 shows the distribution of the latest activated regions in all patients. The latest activated region was located anterolaterally in 18 (basal, n = 10; mid, n = 8) and inferolaterally in 6 (basal, n = 3; mid, n = 3). In one patient (No. 7), identification of the latest activated region was impeded by limited coronary venous anatomy.

Distribution of the latest activated region. Each circled number represents a patient. Numbers refer to the order of inclusion. The latest activated region was mainly located on the anterolateral LV wall. In one patient (No. 7), the latest activated region could not be identified by coronary venous EAM due to limited coronary venous anatomy. A, anterior; AL, anterolateral; AS, anteroseptal; I, inferior; IL, inferolateral; IS, inferoseptal; L, lateral; S, septal.
Figure 3

Distribution of the latest activated region. Each circled number represents a patient. Numbers refer to the order of inclusion. The latest activated region was mainly located on the anterolateral LV wall. In one patient (No. 7), the latest activated region could not be identified by coronary venous EAM due to limited coronary venous anatomy. A, anterior; AL, anterolateral; AS, anteroseptal; I, inferior; IL, inferolateral; IS, inferoseptal; L, lateral; S, septal.

Left ventricular lead placement

Figure 4 shows the distribution of the final LV lead position in all patients. The LV lead was positioned anterolaterally in 16 (basal, n = 9; mid, n = 7) and inferolaterally in 9 (basal, n = 4; mid, n = 5). The final LV lead position was concordant with the latest activated region in 18 of 25 patients. In six patients, concordant LV lead placement was hampered by PNS (n = 4, one of them depicted in Figure 5), lead instability (n = 1), and coronary vein stenosis (n = 1). In the patient in whom the latest activated region could not be identified (No. 7), the LV lead was positioned at the site where maximal electrical delay was measured with coronary venous EAM. An overview of the latest activated regions and final LV lead positions of all patients together with the corresponding electrical delays at these sites is provided in Table 2. In patients with a concordant LV lead position, electrical delay was 132 ± 15 ms (85 ± 6% of QRS duration) at the latest activated region and 128 ± 13 ms (83 ± 5% of QRS duration) at the final lead position (P = 0.17). In patients with a discordant LV lead position, electrical delay was 125 ± 10 ms (84 ± 3% of QRS duration) at the latest activated region and 94 ± 15 ms (63 ± 7% of QRS duration) at the final lead position (P = 0.002). Left ventricular lead electrical delay was significantly longer in patients with a concordant compared with patients with a discordant LV lead position (P < 0.001).

Table 2

Overview of the latest activated regions, final LV lead positions, and empirical target veins of all patients together with the corresponding electrical delays

Patient no.No. of target veinsLatest activated regionElectrical delayFinal LV lead positionElectrical delayReason for discordanceEmpirical target vein for LV leadMaximal electrical delay
12B-AL139 (93%)B-AL131 (87%)IL vein90 (60%)
21B-AL136 (86%)B-AL126 (80%)
31M-IL122 (79%)M-IL122 (79%)
41M-AL131 (84%)M-AL128 (82%)
51B-IL131 (85%)B-IL128 (83%)
61M-AL110 (85%)M-IL72 (56%)PNS
71Not identifiedM-IL103 (63%)
82M-AL126 (87%)M-AL126 (87%)AL vein
91B-AL137 (84%)M-AL103 (63%)Lead instability
101B-AL156 (96%)B-AL149 (92%)
112B-AL130 (85%)B-AL128 (83%)IL vein70 (45%)
122B-AL119 (91%)B-AL119 (91%)IL vein72 (55%)
132M-AL124 (86%)B-IL94 (65%)Coronary vein stenosisIL vein94 (65%)
142B-AL149 (93%)B-AL138 (86%)IL vein106 (66%)
152M-AL116 (91%)M-AL111 (87%)IL vein80 (63%)
161M-IL131 (78%)M-IL131 (78%)
171M-AL111 (76%)M-AL111 (78%)
182M-AL118 (79%)M-IL94 (63%)PNSIL vein94 (63%)
192M-IL130 (83%)B-AL116 (74%)PNSIL vein
201B-AL128 (83%)B-AL127 (82%)
211B-IL143 (80%)B-IL141 (79%)
221B-AL128 (85%)M-AL87 (57%)PNS
232M-AL160 (84%)M-AL155 (81%)IL vein113 (59%)
242B-AL140 (83%)B-AL133 (79%)IL vein90 (54%)
252B-IL103 (80%)B-IL101 (78%)IL vein
Patient no.No. of target veinsLatest activated regionElectrical delayFinal LV lead positionElectrical delayReason for discordanceEmpirical target vein for LV leadMaximal electrical delay
12B-AL139 (93%)B-AL131 (87%)IL vein90 (60%)
21B-AL136 (86%)B-AL126 (80%)
31M-IL122 (79%)M-IL122 (79%)
41M-AL131 (84%)M-AL128 (82%)
51B-IL131 (85%)B-IL128 (83%)
61M-AL110 (85%)M-IL72 (56%)PNS
71Not identifiedM-IL103 (63%)
82M-AL126 (87%)M-AL126 (87%)AL vein
91B-AL137 (84%)M-AL103 (63%)Lead instability
101B-AL156 (96%)B-AL149 (92%)
112B-AL130 (85%)B-AL128 (83%)IL vein70 (45%)
122B-AL119 (91%)B-AL119 (91%)IL vein72 (55%)
132M-AL124 (86%)B-IL94 (65%)Coronary vein stenosisIL vein94 (65%)
142B-AL149 (93%)B-AL138 (86%)IL vein106 (66%)
152M-AL116 (91%)M-AL111 (87%)IL vein80 (63%)
161M-IL131 (78%)M-IL131 (78%)
171M-AL111 (76%)M-AL111 (78%)
182M-AL118 (79%)M-IL94 (63%)PNSIL vein94 (63%)
192M-IL130 (83%)B-AL116 (74%)PNSIL vein
201B-AL128 (83%)B-AL127 (82%)
211B-IL143 (80%)B-IL141 (79%)
221B-AL128 (85%)M-AL87 (57%)PNS
232M-AL160 (84%)M-AL155 (81%)IL vein113 (59%)
242B-AL140 (83%)B-AL133 (79%)IL vein90 (54%)
252B-IL103 (80%)B-IL101 (78%)IL vein

Patient numbers correspond with the numbers provided in Figures 3 and 4. Electrical delay is reported in milliseconds (percentage of total QRS duration). In Patient No. 6, the anterolateral vein anastomosed with the inferior inter-ventricular vein, which made it possible to position the LV lead inferolaterally (via the anastomosis) despite the patient having only one target vein on the anterolateral LV wall. In Patient No. 7, identification of the latest activated region was impeded by limited coronary venous anatomy. Patient No. 8 had two target veins on the anterolateral wall. For patients with two target veins, the second column from the right reports the target vein for the LV lead if the conventional anatomical lead placement approach would be used. The right-most column provides the maximal electrical delay measured within that vein. B, basal; M, mid; AL, anterolateral; IL, inferolateral; PNS, phrenic nerve stimulation.

Table 2

Overview of the latest activated regions, final LV lead positions, and empirical target veins of all patients together with the corresponding electrical delays

Patient no.No. of target veinsLatest activated regionElectrical delayFinal LV lead positionElectrical delayReason for discordanceEmpirical target vein for LV leadMaximal electrical delay
12B-AL139 (93%)B-AL131 (87%)IL vein90 (60%)
21B-AL136 (86%)B-AL126 (80%)
31M-IL122 (79%)M-IL122 (79%)
41M-AL131 (84%)M-AL128 (82%)
51B-IL131 (85%)B-IL128 (83%)
61M-AL110 (85%)M-IL72 (56%)PNS
71Not identifiedM-IL103 (63%)
82M-AL126 (87%)M-AL126 (87%)AL vein
91B-AL137 (84%)M-AL103 (63%)Lead instability
101B-AL156 (96%)B-AL149 (92%)
112B-AL130 (85%)B-AL128 (83%)IL vein70 (45%)
122B-AL119 (91%)B-AL119 (91%)IL vein72 (55%)
132M-AL124 (86%)B-IL94 (65%)Coronary vein stenosisIL vein94 (65%)
142B-AL149 (93%)B-AL138 (86%)IL vein106 (66%)
152M-AL116 (91%)M-AL111 (87%)IL vein80 (63%)
161M-IL131 (78%)M-IL131 (78%)
171M-AL111 (76%)M-AL111 (78%)
182M-AL118 (79%)M-IL94 (63%)PNSIL vein94 (63%)
192M-IL130 (83%)B-AL116 (74%)PNSIL vein
201B-AL128 (83%)B-AL127 (82%)
211B-IL143 (80%)B-IL141 (79%)
221B-AL128 (85%)M-AL87 (57%)PNS
232M-AL160 (84%)M-AL155 (81%)IL vein113 (59%)
242B-AL140 (83%)B-AL133 (79%)IL vein90 (54%)
252B-IL103 (80%)B-IL101 (78%)IL vein
Patient no.No. of target veinsLatest activated regionElectrical delayFinal LV lead positionElectrical delayReason for discordanceEmpirical target vein for LV leadMaximal electrical delay
12B-AL139 (93%)B-AL131 (87%)IL vein90 (60%)
21B-AL136 (86%)B-AL126 (80%)
31M-IL122 (79%)M-IL122 (79%)
41M-AL131 (84%)M-AL128 (82%)
51B-IL131 (85%)B-IL128 (83%)
61M-AL110 (85%)M-IL72 (56%)PNS
71Not identifiedM-IL103 (63%)
82M-AL126 (87%)M-AL126 (87%)AL vein
91B-AL137 (84%)M-AL103 (63%)Lead instability
101B-AL156 (96%)B-AL149 (92%)
112B-AL130 (85%)B-AL128 (83%)IL vein70 (45%)
122B-AL119 (91%)B-AL119 (91%)IL vein72 (55%)
132M-AL124 (86%)B-IL94 (65%)Coronary vein stenosisIL vein94 (65%)
142B-AL149 (93%)B-AL138 (86%)IL vein106 (66%)
152M-AL116 (91%)M-AL111 (87%)IL vein80 (63%)
161M-IL131 (78%)M-IL131 (78%)
171M-AL111 (76%)M-AL111 (78%)
182M-AL118 (79%)M-IL94 (63%)PNSIL vein94 (63%)
192M-IL130 (83%)B-AL116 (74%)PNSIL vein
201B-AL128 (83%)B-AL127 (82%)
211B-IL143 (80%)B-IL141 (79%)
221B-AL128 (85%)M-AL87 (57%)PNS
232M-AL160 (84%)M-AL155 (81%)IL vein113 (59%)
242B-AL140 (83%)B-AL133 (79%)IL vein90 (54%)
252B-IL103 (80%)B-IL101 (78%)IL vein

Patient numbers correspond with the numbers provided in Figures 3 and 4. Electrical delay is reported in milliseconds (percentage of total QRS duration). In Patient No. 6, the anterolateral vein anastomosed with the inferior inter-ventricular vein, which made it possible to position the LV lead inferolaterally (via the anastomosis) despite the patient having only one target vein on the anterolateral LV wall. In Patient No. 7, identification of the latest activated region was impeded by limited coronary venous anatomy. Patient No. 8 had two target veins on the anterolateral wall. For patients with two target veins, the second column from the right reports the target vein for the LV lead if the conventional anatomical lead placement approach would be used. The right-most column provides the maximal electrical delay measured within that vein. B, basal; M, mid; AL, anterolateral; IL, inferolateral; PNS, phrenic nerve stimulation.

Distribution of the final LV lead position. Each circled number represents a patient. Numbers refer to the order of inclusion and correspond with the numbers in Figure 3. A, anterior; AL, anterolateral; AS, anteroseptal; I, inferior; IL, inferolateral; IS, inferoseptal; L, lateral; S, septal.
Figure 4

Distribution of the final LV lead position. Each circled number represents a patient. Numbers refer to the order of inclusion and correspond with the numbers in Figure 3. A, anterior; AL, anterolateral; AS, anteroseptal; I, inferior; IL, inferolateral; IS, inferoseptal; L, lateral; S, septal.

(A) Coronary venous 3D electrical activation map during intrinsic ventricular activation of a patient in whom PNS hampered placement of the LV lead in the latest activated region. Mapping of the anterolateral and inferolateral vein revealed the latest activated region at the mid-anterolateral LV wall. Unfortunately, this was also a region with PNS. The LV lead was eventually placed as close as possible to the latest activated region and pacing was performed at the ring of the LV lead, which in contrast to the tip still offered some space between ventricular capture and PNS. (B) Corresponding coronary venogram. (C) X-ray of final LV lead placement. Red dot, PNS; Green dot, no PNS; R, LV lead ring; T, LV lead tip; AIV, anterior inter-ventricular vein; ALV, anterolateral vein; IIV, inferior inter-ventricular vein; ILV, inferolateral vein; RV, right ventricular.
Figure 5

(A) Coronary venous 3D electrical activation map during intrinsic ventricular activation of a patient in whom PNS hampered placement of the LV lead in the latest activated region. Mapping of the anterolateral and inferolateral vein revealed the latest activated region at the mid-anterolateral LV wall. Unfortunately, this was also a region with PNS. The LV lead was eventually placed as close as possible to the latest activated region and pacing was performed at the ring of the LV lead, which in contrast to the tip still offered some space between ventricular capture and PNS. (B) Corresponding coronary venogram. (C) X-ray of final LV lead placement. Red dot, PNS; Green dot, no PNS; R, LV lead ring; T, LV lead tip; AIV, anterior inter-ventricular vein; ALV, anterolateral vein; IIV, inferior inter-ventricular vein; ILV, inferolateral vein; RV, right ventricular.

In 9 of 12 patients with two available target veins (Nos. 1, 11, 12, 13, 14, 15, 18, 23, and 24), the latest activated region was located in a different vein than the anatomically targeted inferolateral vein (Table 2). In these patients, maximal electrical delay was 89 ± 14 ms (58 ± 6% of QRS duration) in the anatomically targeted vein vs. 133 ± 15 ms (87 ± 5% of QRS duration) at the latest activated region (P < 0.001). The LV lead was positioned in the latest activated region in seven of nine patients (Nos. 1, 11, 12, 14, 15, 23, and 24). In these patients, maximal electrical delay was 89 ± 16 ms (57 ± 7% of QRS duration) in the anatomically targeted vein vs. 131 ± 14 ms (85 ± 4% of QRS duration) at the final LV lead position (P = 0.001). In the other two patients (Nos. 13 and 18), coronary vein stenosis and PNS hampered concordant LV lead placement. Figure 6 illustrates the additional value of mapping-guided LV lead placement as compared with the conventional anatomical LV lead placement approach. In this example, the coronary venous EAM of Patient No. 24 is shown where a pure anatomical approach would have resulted in a suboptimal LV lead position in the inferolateral vein with a maximal electrical delay of only 90 ms. However, additional mapping of the small anterolateral vein resulted in a potentially far better LV lead position with an electrical delay close to 140 ms.

Coronary venous EAM of Patient No. 24, illustrating the additional value of LV lead placement guided by coronary venous mapping. Pre-implantation coronary venograms in LAO (A) and RAO (B) view revealed a large inferolateral and a small anterolateral vein. The conventional anatomical LV lead placement approach would target the inferolateral vein. However, coronary venous mapping during intrinsic ventricular activation (C) revealed a maximal electrical delay of only 90 ms in the basal inferolateral vein, which further decreased in the more distal segments, whereas a maximal electrical delay of 140 ms could be measured in the anterolateral vein. Based on the outcome of coronary venous mapping, the LV lead was positioned in the anterolateral vein, thereby achieving a clear benefit in LV lead electrical delay. (D) X-ray of final LV lead position in LAO view. (E) X-ray of final LV lead position in RAO view. AIV, anterior inter-ventricular vein; ALV, anterolateral vein; IIV, inferior inter-ventricular vein; ILV, inferolateral vein.
Figure 6

Coronary venous EAM of Patient No. 24, illustrating the additional value of LV lead placement guided by coronary venous mapping. Pre-implantation coronary venograms in LAO (A) and RAO (B) view revealed a large inferolateral and a small anterolateral vein. The conventional anatomical LV lead placement approach would target the inferolateral vein. However, coronary venous mapping during intrinsic ventricular activation (C) revealed a maximal electrical delay of only 90 ms in the basal inferolateral vein, which further decreased in the more distal segments, whereas a maximal electrical delay of 140 ms could be measured in the anterolateral vein. Based on the outcome of coronary venous mapping, the LV lead was positioned in the anterolateral vein, thereby achieving a clear benefit in LV lead electrical delay. (D) X-ray of final LV lead position in LAO view. (E) X-ray of final LV lead position in RAO view. AIV, anterior inter-ventricular vein; ALV, anterolateral vein; IIV, inferior inter-ventricular vein; ILV, inferolateral vein.

Discussion

Pacing at the latest activated region of the LV appears to improve the outcome of CRT. The present study demonstrates that coronary venous EAM using EnSite NavX in combination with a mapping guidewire can be used intraprocedurally to guide LV lead placement to the region of latest electrical activation. In addition, the potential of this mapping approach to facilitate comprehensive identification of regions free of PNS is demonstrated.

Targeted left ventricular lead placement at the latest activated left ventricular region

Several studies have investigated the relationship between pacing in the region of latest activation and the response to CRT. Studies on mechanical activation suggest superior CRT outcome when the LV lead position coincides with the region of latest mechanical contraction compared with discordant positions.16–22 Two recent prospective single-centre trials randomized a total of 407 patients to speckle-tracking echocardiography-guided LV lead placement targeting the site of latest mechanical activation or to standard unguided LV lead placement. In both trials, patients with a LV lead position concordant with the site of latest activation had a higher echocardiographic response rate, more clinical responders, and a reduced risk of mortality and heart failure hospitalization.23,24 Studies focusing on the electrical activation pattern have demonstrated that a greater delay in time from onset of the QRS complex to the local sensed LV lead EGM (Q-LV) is also associated with a greater likelihood of benefit from CRT.25,26 Recently, Zanon et al. measured the Q-LV interval of various pacing sites within patients, and evaluated the haemodynamic effect of pacing at the different sites by invasive measurement of LVdP/dtmax. Pacing the LV at the latest activated site resulted in greatest haemodynamic improvement.6 The choice between targeting the region of latest electrical or latest mechanical activation is still a matter of debate and definitive randomized trials that support the use of either approach to reliably select an optimal LV pacing site are lacking. Accordingly, current ESC guidelines have assigned a class 2b indication to targeted LV lead placement at the latest activated LV segment.7 A slightly stronger recommendation (class 2a) has been assigned to avoiding an apical LV lead position7 despite conflicting results in the literature.27–31 Therefore, larger prospective studies are required to settle the issue of whether pacing the latest activated LV region provides the best outcome of CRT.

While the position of the LV lead with respect to the site of latest activation is a factor that can influence CRT response, several studies have shown that the proximity of the LV lead to an area with myocardial scar is an important factor as well.32–36 Assessment of scar tissue should therefore be considered before CRT, especially in the population with ischaemic heart disease. Integrating information obtained from scar imaging with measurements of electrical and mechanical delay may further optimize LV lead positioning and the clinical response rate.

Feasibility of coronary venous electroanatomic mapping to guide left ventricular lead placement to the latest activated region

Our observation that the location of the latest activated region varies between patients agrees with previous EAM studies in patients with heart failure and LBBB,8–10 and emphasizes the need for a patient-tailored approach to optimize LV lead positioning. The present study demonstrates that coronary venous EAM using EnSite NavX in combination with a mapping guidewire is a feasible method to tailor LV lead placement to the individual patient at the time of implantation. Using this approach, we were able to position the LV lead in the latest activated region in >70% of our patients. Conventional LV lead placement strategy involves an anatomical approach, targeting a coronary vein situated on the posterolateral wall. Approximately half the patients in our study cohort had more than one target vein available on the lateral LV free wall. In most of these patients, the empirically targeted posterolateral vein was not the vein with maximal electrical delay, which is consistent with the findings of a previous study that also used coronary venous EAM to assess LV epicardial activation.37 In the present study, coronary venous EAM resulted in targeting of an alternative vein in many of these patients, significantly increasing LV lead electrical delay from an average of 57–85% of total QRS duration. In addition, in patients with a single target vein, coronary venous EAM enabled targeting of the vein segment with maximal electrical delay. These results demonstrate the additional value of coronary venous EAM-guided LV lead placement as compared with the conventional anatomical LV lead placement approach for optimization of LV lead electrical delay.

The use of EnSite NavX to create an EAM of the coronary veins at the time of CRT implantation has previously been described by Ryu et al.11 In addition, Del Greco et al.12 demonstrated that lead placement guided by the 3D navigation system of EnSite NavX is beneficial in terms of reducing X-ray exposure during CRT implantation. Recently, Fujiwara et al.37 compared electrical and mechanical dyssynchrony using coronary venous EAM and speckle-tracking echocardiography. However, sample sizes of these studies were small, LV lead placement was not targeted at the latest activated region, and constraints imposed by coronary venous anatomy for identifying the latest activated LV region were not taken into account. Therefore, the present study is the first to demonstrate, in a larger and representative CRT population, the feasibility of positioning the LV lead in the latest activated region guided by coronary venous EAM.

Unfortunately, concordant LV lead placement was still hampered in one-quarter of patients due to PNS, lead instability, and unfavourable coronary venous anatomy. The use of additional technologies may increase the success rate of targeted LV lead placement on the basis of electrical delay. Venous angioplasty, for example, may have benefited one of our patients in whom coronary venous stenosis hampered LV lead placement in the latest activated region. In addition, the use of new lead technologies, such as active-fixation leads and multipolar leads, could overcome the problem of lead instability and avoid PNS, which hindered concordant LV lead positioning in several other patients in our study. When confronted with PNS, also alternative pacing strategies such as thoracoscopic epicardial or LV endocardial pacing may be considered to stimulate the site of optimal electrical delay.

Our mapping approach was easy to implement intraprocedurally using a commercially available EAM system in combination with a guidewire that allows mapping and pacing of the coronary veins, making the approach applicable in any catheter laboratory that has access to the relevant instruments. In addition, the same wire used to guide the LV lead was also used to map the coronary veins, which obviated the need for inserting additional electrophysiology mapping catheters during implantation. Echocardiography and other imaging modalities have also been proposed as methods to assess the LV activation pattern and guide LV lead placement,16–18,20,21,38 but these techniques are challenging to apply in the catheter laboratory and often require a separate pre-operative assessment.

In current clinical practice, most implanters that tend to target LV lead placement to a region with significant electrical delay do not perform full EAM in favour of a simpler approach where electrical mapping of the anatomically targeted region is performed using the LV lead to test if electrical delay is sufficiently late. However, using this approach the latest activated region may be missed because other potential LV pacing sites are generally not tested when electrical delay at the empirically targeted LV region is considered late enough. Previous reports suggest that a Q-LV >95 ms or >50% of total QRS duration is sufficient to increase the odds of a favourable response.5,39 However, further increase of the Q-LV to >120 ms has been shown to considerably improve these odds,5 suggesting that implanters should aim to optimize LV lead electrical delay to maximize the chance of a positive CRT outcome. Besides, with simple electrical mapping, measurements are generally performed after final LV lead implantation. If the measured electrical delay is judged unsatisfactory, the implanter may have to relocate the LV lead to an alternative target vein, thus extending procedure and fluoroscopy time. Compared with simple electrical mapping using the LV lead, our 3D EAM approach has important advantages that can significantly contribute to LV lead placement optimization. The EnSite NavX system can be used to create an anatomical geometry of the coronary veins while electrical delay and PNS are assessed for any anatomical region accessible with the mapping guidewire. The small and flexible mapping guidewire allows for more detailed data collection by easy and quick mapping of multiple coronary veins and multiple points per vein, which provides information on electrical activation timing and PNS of all potential LV pacing sites. Thus, all information about coronary venous anatomy, LV electrical activation, and PNS can be readily obtained prior to LV lead placement. The LV lead can then be targeted to the latest activated region free of PNS in a single approach, thus reducing placement attempts at unsuitable sites. Targeted LV lead placement is further facilitated by real-time visualization and navigation of the lead in the NavX-created EAM.

Avoiding phrenic nerve stimulation

Phrenic nerve stimulation accounts for failure to deliver proper CRT or urgent need for LV lead relocation in a substantial number of patients. Studies addressing the issue have reported PNS at implantation in 7.6–37% of patients, resulting in failure to pace the target site in 10–14%.40–42 PNS at follow-up has been reported in 11–33% of patients, requiring LV lead replacement in a substantial proportion of patients and eventual turn-off of CRT in some.40–42 In the present study, PNS was assessed by systematic pacing of the coronary veins with the tip of the mapping guidewire. The usefulness of this approach for detecting PNS has previously been demonstrated by de Cock et al.14 In addition, the EAM system used in our study allowed comprehensive annotation of PNS-free areas on the created coronary venous geometry, consequently providing 3D visualization of the anatomical location of these areas with respect to the target LV pacing site and the position of the LV lead tip. Subsequently, real-time navigation of the LV lead in the geometry allowed targeting of a PNS-free zone with optimal electrical delay. This approach proved to be effective to avoid PNS at medium-term follow-up (4 ± 1 months) in all study patients.

Limitations

The sample size of this study was relatively small. However, the study was performed on consecutive patients and the study population resembled the typical population of patients with LBBB and heart failure referred for CRT implantation. Only standard bipolar LV leads were used. The use of new lead technologies, such as multipolar leads and active-fixation leads, may increase the success rate of targeted LV lead placement at the latest activated LV region by improving lead stability and avoiding PNS. Unfortunately, most device vendors did not have multipolar LV leads available at the time the current study was conducted. Cardiac magnetic resonance imaging was not performed in this study. Therefore, we could not assess the distribution and extent of scar and its potential effect on electrical activation. Also, the mapping method used in the current study did not allow for reliable differentiation between scar and healthy myocardium. Thus, the position of the LV lead with respect to regions of myocardial scar was not taken into account during LV lead placement. The current study demonstrates a feasible method to guide LV lead placement to the region of latest electrical activation. However, the clinical value of this lead placement strategy was not proven by acute haemodynamic measurements or long-term echocardiographic or clinical follow-up. Subsequent larger and long-term follow-up studies are therefore required to evaluate the impact of coronary venous mapping-guided LV lead placement on CRT outcome.

Conclusion

Coronary venous EAM using EnSite NavX in combination with a mapping guidewire can be used at the time of CRT implantation to guide LV lead placement to the latest activated region free of PNS. This approach especially contributes to optimization of LV lead electrical delay in patients with multiple target veins. The conventional anatomical LV lead placement strategy does not target the vein with maximal electrical delay in many of these patients. The clinical value of coronary venous mapping-guided LV lead placement needs prospective evaluation.

Acknowledgements

We gratefully thank Peter Bakker and Arjan Bennink from St Jude Medical for their technical support during the mapping procedures.

Conflict of interest: Y.B. is a consultant for Medtronic. L.P. is a consultant for Atricure. F.W.P. received research grants from Medtronic, Boston Scientific, EBR Systems, Biological Delivery System Cordis, MSD, and Proteus Medical. H.J.C. received consulting fees from Boehringer Ingelheim, Sanofi-Aventis, and AstraZeneca, grant support from St Jude Medical, Boston Scientific, Boehringer Ingelheim, Sanofi-Aventis, Medapharma, and Merck, and honoraria from Medtronic, Sanofi-Aventis, Medapharma, Merck, Boehringer Ingelheim, and Biosense Webster. K.V. received research grants from Medtronic and is a consultant for Medtronic. All other authors declare no conflict of interest.

References

1
Cleland
JG
Daubert
JC
Erdmann
E
Freemantle
N
Gras
D
Kappenberger
L
et al.
,
The effect of cardiac resynchronization on morbidity and mortality in heart failure
N Engl J Med
,
2005
, vol.
352
(pg.
1539
-
49
)
2
Linde
C
Abraham
WT
Gold
MR
St John Sutton
M
Ghio
S
Daubert
C
,
Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms
J Am Coll Cardiol
,
2008
, vol.
52
(pg.
1834
-
43
)
3
Moss
AJ
Hall
WJ
Cannom
DS
Klein
H
Brown
MW
Daubert
JP
et al.
,
Cardiac-resynchronization therapy for the prevention of heart-failure events
N Engl J Med
,
2009
, vol.
361
(pg.
1329
-
38
)
4
Daubert
JC
Saxon
L
Adamson
PB
Auricchio
A
Berger
RD
Beshai
JF
et al.
,
2012 EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up recommendations and management
Europace
,
2012
, vol.
14
(pg.
1236
-
86
)
5
Gold
MR
Birgersdotter-Green
U
Singh
JP
Ellenbogen
KA
Yu
Y
Meyer
TE
et al.
,
The relationship between ventricular electrical delay and left ventricular remodelling with cardiac resynchronization therapy
Eur Heart J
,
2011
, vol.
32
(pg.
2516
-
24
)
6
Zanon
F
Baracca
E
Pastore
G
Fraccaro
C
Roncon
L
Aggio
S
et al.
,
Determination of the longest intra-patient left ventricular electrical delay may predict acute hemodynamic improvement in cardiac resynchronization therapy patients
Circ Arrhythm Electrophysiol
,
2014
, vol.
7
(pg.
377
-
83
)
7
Brignole
M
Auricchio
A
Baron-Esquivias
G
Bordachar
P
Boriani
G
Breithardt
OA
et al.
,
2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA)
Europace
,
2013
, vol.
15
(pg.
1070
-
118
)
8
Fung
JW
Yu
CM
Yip
G
Zhang
Y
Chan
H
Kum
CC
et al.
,
Variable left ventricular activation pattern in patients with heart failure and left bundle branch block
Heart
,
2004
, vol.
90
(pg.
17
-
9
)
9
Auricchio
A
Fantoni
C
Regoli
F
Carbucicchio
C
Goette
A
Geller
C
et al.
,
Characterization of left ventricular activation in patients with heart failure and left bundle-branch block
Circulation
,
2004
, vol.
109
(pg.
1133
-
9
)
10
Rodriguez
LM
Timmermans
C
Nabar
A
Beatty
G
Wellens
HJ
,
Variable patterns of septal activation in patients with left bundle branch block and heart failure
J Cardiovasc Electrophysiol
,
2003
, vol.
14
(pg.
135
-
41
)
11
Ryu
K
D'Avila
A
Heist
EK
Rosenberg
SP
Chou
J
Yang
M
et al.
,
Simultaneous electrical and mechanical mapping using 3D cardiac mapping system: novel approach for optimal cardiac resynchronization therapy
J Cardiovasc Electrophysiol
,
2010
, vol.
21
(pg.
219
-
22
)
12
Del Greco
M
Marini
M
Bonmassari
R
,
Implantation of a biventricular implantable cardioverter-defibrillator guided by an electroanatomic mapping system
Europace
,
2012
, vol.
14
(pg.
107
-
11
)
13
Cosio
FG
Camm
AJ
,
The electrocardiogram
The ESC Textbook of Cardiovascular Medicine 2009
,
2009
2nd edn
UK
Oxford University Press
14
de Cock
CC
Res
JC
Hendriks
ML
Allaart
CP
,
Usefulness of a pacing guidewire to facilitate left ventricular lead implantation in cardiac resynchronization therapy
Pacing Clin Electrophysiol
,
2009
, vol.
32
(pg.
446
-
9
)
15
Cerqueira
MD
Weissman
NJ
Dilsizian
V
Jacobs
AK
Kaul
S
Laskey
WK
et al.
,
Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association
Circulation
,
2002
, vol.
105
(pg.
539
-
42
)
16
Ypenburg
C
van Bommel
RJ
Delgado
V
Mollema
SA
Bleeker
GB
Boersma
E
et al.
,
Optimal left ventricular lead position predicts reverse remodeling and survival after cardiac resynchronization therapy
J Am Coll Cardiol
,
2008
, vol.
52
(pg.
1402
-
9
)
17
Ansalone
G
Giannantoni
P
Ricci
R
Trambaiolo
P
Fedele
F
Santini
M
,
Doppler myocardial imaging to evaluate the effectiveness of pacing sites in patients receiving biventricular pacing
J Am Coll Cardiol
,
2002
, vol.
39
(pg.
489
-
99
)
18
Becker
M
Altiok
E
Ocklenburg
C
Krings
R
Adams
D
Lysansky
M
et al.
,
Analysis of LV lead position in cardiac resynchronization therapy using different imaging modalities
JACC Cardiovasc Imaging
,
2010
, vol.
3
(pg.
472
-
81
)
19
Becker
M
Franke
A
Breithardt
OA
Ocklenburg
C
Kaminski
T
Kramann
R
et al.
,
Impact of left ventricular lead position on the efficacy of cardiac resynchronisation therapy: a two-dimensional strain echocardiography study
Heart
,
2007
, vol.
93
(pg.
1197
-
203
)
20
Becker
M
Hoffmann
R
Schmitz
F
Hundemer
A
Kuhl
H
Schauerte
P
et al.
,
Relation of optimal lead positioning as defined by three-dimensional echocardiography to long-term benefit of cardiac resynchronization
Am J Cardiol
,
2007
, vol.
100
(pg.
1671
-
6
)
21
Becker
M
Kramann
R
Franke
A
Breithardt
OA
Heussen
N
Knackstedt
C
et al.
,
Impact of left ventricular lead position in cardiac resynchronization therapy on left ventricular remodelling. A circumferential strain analysis based on 2D echocardiography
Eur Heart J
,
2007
, vol.
28
(pg.
1211
-
20
)
22
Bakos
Z
Ostenfeld
E
Markstad
H
Werther-Evaldsson
A
Roijer
A
Arheden
H
et al.
,
A comparison between radial strain evaluation by speckle-tracking echocardiography and cardiac magnetic resonance imaging, for assessment of suitable segments for left ventricular lead placement in cardiac resynchronization therapy
Europace
,
2014
, vol.
16
(pg.
1779
-
86
)
23
Khan
FZ
Virdee
MS
Palmer
CR
Pugh
PJ
O'Halloran
D
Elsik
M
et al.
,
Targeted left ventricular lead placement to guide cardiac resynchronization therapy: the TARGET study: a randomized, controlled trial
J Am Coll Cardiol
,
2012
, vol.
59
(pg.
1509
-
18
)
24
Saba
S
Marek
J
Schwartzman
D
Jain
S
Adelstein
E
White
P
et al.
,
Echocardiography-guided left ventricular lead placement for cardiac resynchronization therapy: results of the Speckle Tracking Assisted Resynchronization Therapy for Electrode Region trial
Circ Heart Fail
,
2013
, vol.
6
(pg.
427
-
34
)
25
Singh
JP
Fan
D
Heist
EK
Alabiad
CR
Taub
C
Reddy
V
et al.
,
Left ventricular lead electrical delay predicts response to cardiac resynchronization therapy
Heart Rhythm
,
2006
, vol.
3
(pg.
1285
-
92
)
26
van Gelder
BM
Meijer
A
Bracke
FA
,
Timing of the left ventricular electrogram and acute hemodynamic changes during implant of cardiac resynchronization therapy devices
Pacing Clin Electrophysiol
,
2009
, vol.
32
Suppl 1
(pg.
S94
-
7
)
27
Singh
JP
Klein
HU
Huang
DT
Reek
S
Kuniss
M
Quesada
A
et al.
,
Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial-cardiac resynchronization therapy (MADIT-CRT) trial
Circulation
,
2011
, vol.
123
(pg.
1159
-
66
)
28
Gold
MR
Auricchio
A
Hummel
JD
Giudici
MC
Ding
J
Tockman
B
et al.
,
Comparison of stimulation sites within left ventricular veins on the acute hemodynamic effects of cardiac resynchronization therapy
Heart Rhythm
,
2005
, vol.
2
(pg.
376
-
81
)
29
Foley
PW
Chalil
S
Ratib
K
Smith
R
Prinzen
F
Auricchio
A
et al.
,
Fluoroscopic left ventricular lead position and the long-term clinical outcome of cardiac resynchronization therapy
Pacing Clin Electrophysiol
,
2011
, vol.
34
(pg.
785
-
97
)
30
Jastrzebski
M
Wilinski
J
Fijorek
K
Sondej
T
Czarnecka
D.
,
Mortality and morbidity in cardiac resynchronization patients: impact of lead position, paced left ventricular QRS morphology and other characteristics on long-term outcome
Europace
,
2013
, vol.
15
(pg.
258
-
65
)
31
Saxon
LA
Olshansky
B
Volosin
K
Steinberg
JS
Lee
BK
Tomassoni
G
et al.
,
Influence of left ventricular lead location on outcomes in the COMPANION study
J Cardiovasc Electrophysiol
,
2009
, vol.
20
(pg.
764
-
8
)
32
Bleeker
GB
Schalij
MJ
Van Der Wall
EE
Bax
JJ
,
Postero-lateral scar tissue resulting in non-response to cardiac resynchronization therapy
J Cardiovasc Electrophysiol
,
2006
, vol.
17
(pg.
899
-
901
)
33
Hummel
JP
Lindner
JR
Belcik
JT
Ferguson
JD
Mangrum
JM
Bergin
JD
et al.
,
Extent of myocardial viability predicts response to biventricular pacing in ischemic cardiomyopathy
Heart Rhythm
,
2005
, vol.
2
(pg.
1211
-
7
)
34
Ypenburg
C
Roes
SD
Bleeker
GB
Kaandorp
TA
de Roos
A
Schalij
MJ
et al.
,
Effect of total scar burden on contrast-enhanced magnetic resonance imaging on response to cardiac resynchronization therapy
Am J Cardiol
,
2007
, vol.
99
(pg.
657
-
60
)
35
Ypenburg
C
Schalij
MJ
Bleeker
GB
Steendijk
P
Boersma
E
Dibbets-Schneider
P
et al.
,
Impact of viability and scar tissue on response to cardiac resynchronization therapy in ischaemic heart failure patients
Eur Heart J
,
2007
, vol.
28
(pg.
33
-
41
)
36
Chalil
S
Stegemann
B
Muhyaldeen
SA
Khadjooi
K
Foley
PW
Smith
RE
et al.
,
Effect of posterolateral left ventricular scar on mortality and morbidity following cardiac resynchronization therapy
Pacing Clin Electrophysiol
,
2007
, vol.
30
(pg.
1201
-
9
)
37
Fujiwara
R
Yoshida
A
Fukuzawa
K
Takei
A
Kiuchi
K
Itoh
M
et al.
,
Discrepancy between electrical and mechanical dyssynchrony in patients with heart failure and an electrical disturbance
Pacing Clin Electrophysiol
,
2013
, vol.
37
(pg.
576
-
84
)
38
Boogers
MJ
Chen
J
van Bommel
RJ
Borleffs
CJ
Dibbets-Schneider
P
van der Hiel
B
et al.
,
Optimal left ventricular lead position assessed with phase analysis on gated myocardial perfusion SPECT
Eur J Nucl Med Mol Imaging
,
2011
, vol.
38
(pg.
230
-
8
)
39
Kandala
J
Upadhyay
GA
Altman
RK
Parks
KA
Orencole
M
Mela
T
et al.
,
QRS morphology, left ventricular lead location, and clinical outcome in patients receiving cardiac resynchronization therapy
Eur Heart J
,
2013
, vol.
34
(pg.
2252
-
62
)
40
Gurevitz
O
Nof
E
Carasso
S
Luria
D
Bar-Lev
D
Tanami
N
et al.
,
Programmable multiple pacing configurations help to overcome high left ventricular pacing thresholds and avoid phrenic nerve stimulation
Pacing Clin Electrophysiol
,
2005
, vol.
28
(pg.
1255
-
9
)
41
Biffi
M
Moschini
C
Bertini
M
Saporito
D
Ziacchi
M
Diemberger
I
et al.
,
Phrenic stimulation: a challenge for cardiac resynchronization therapy
Circ Arrhythm Electrophysiol
,
2009
, vol.
2
(pg.
402
-
10
)
42
Seifert
M
Schau
T
Moeller
V
Neuss
M
Meyhoefer
J
Butter
C
,
Influence of pacing configurations, body mass index, and position of coronary sinus lead on frequency of phrenic nerve stimulation and pacing thresholds under cardiac resynchronization therapy
Europace
,
2010
, vol.
12
(pg.
961
-
7
)

Supplementary data