Abstract

OBJECTIVES

The wearable cardioverter defibrillator (WCD) is an established, safe, effective solution, protecting patients at risk of sudden cardiac death. We specifically investigated WCD use in cardiac surgery patients since data for this patient group are rare.

METHODS

Retrospective data analysis in 10 German cardiac surgery centres was performed. Cardiac surgery patients with left ventricular ejection fraction (LVEF) ≤35% or after implantable cardioverter defibrillator (ICD) explantation who received WCD between 2010 and 2020 were assessed using LifeVest Network data.

RESULTS

A total of 1168 patients with a median age of 66 years [interquartile range (IQR) 57–73] were enrolled; 87% were male. Clinical indications included coronary artery bypass grafting (43%), valve surgery (16%), combined coronary artery bypass graft/valve surgery (15%), ICD explantation (24%) and miscellaneous (2%). The median wear time of WCD was 23.4 h/day (IQR 21.7–23.8). A total of 106 patients (9.1%) exhibited ventricular tachycardia. A total of 93.2% of episodes occurred within the first 3 months. Eighteen patients (1.5%) received 26 adequate shocks. The inadequate shock rate was low (8 patients, 0.7%). LVEF improved from a median of 28% (IQR 22–32%) before WCD prescription to 35% (IQR 28–42%) during follow-up. Excluding ICD explant patients, 37% of patients received an ICD.

CONCLUSIONS

The risk of sudden cardiac death is substantial within the first 3 months after cardiac surgery. Patients were protected effectively by WCD. Due to significant LVEF improvement, the majority did not require ICD implantation after WCD use. Compliance was high despite sternotomy. This multicentre experience confirms existing data regarding effectiveness, safety and compliance. Therefore, WCD should be considered in cardiac surgery patients with severely reduced LVEF.

INTRODUCTION

Sudden cardiac death (SCD) is one of the leading causes of death in Europe, with ∼600 000 deaths annually [1]. Over 50% are due to life-threatening ventricular tachyarrhythmia (VT)/ventricular fibrillation (VF), often with underlying cardiac disease [2]. The vast majority of patients have reduced left ventricular ejection fraction (LVEF) [3, 4]. A correlation between the occurrence of life-threatening VT/VF and reduced LVEF has been demonstrated, making severely reduced LVEF one of the most important predictors for SCD [3, 4].

Cardiac surgery patients often demonstrate severely reduced LVEF even preoperatively. Advances in surgical therapeutic techniques have reduced the operative risk [5]. However, there is still a high peri- and postoperative mortality risk, mostly due to VT/VF [6, 7] This risk is highest in the first 3 months postoperatively and decreases significantly over time [6–9]. Recovery correlates with regeneration of the heart during this time, which leads to a substantial LVEF improvement in a significant proportion of cases. Guideline-directed medical therapy is recommended for at least 3 months for optimal adjustment [10, 11]. In cardiac surgery patients, significant LVEF improvement has been observed up to 6 months postoperatively [12–14].

Implantable cardioverter defibrillators (ICD) can effectively terminate VT/VF. However, current guidelines of the European Society of Cardiology and the American Heart Association recommend a final decision on ICD implantation not earlier than 3 months after the diagnosis of severe heart failure and exclusion of reversible causes [10, 11, 15].

To close the gap between hospital discharge and ICD decision, guidelines recommend the temporary use of a wearable cardioverter defibrillator (WCD) in the vulnerable phase to protect patients at risk from SCD [10, 11, 16, 17]. The WCD is a non-invasive device, which can detect and terminate haemodynamically relevant VT/VFs [18, 19]. The first randomized controlled trial (VEST) showed a significant overall mortality reduction, although a significant arrhythmic mortality reduction was only found in the per-protocol and as-treated analyses [20]. Results underscore the need for clear and direct patient education about the risks of SCD and the benefits of wearing a WCD to ensure good compliance [21]. Several large real-life studies consistently confirmed safety, effectiveness and high WCD compliance [18–27]. Based on these data, the German Society for Thoracic and Cardiovascular Surgery (DGTHG) recently published a position paper recommending WCD use [17]. So far, only a small number of cardiac surgery patients have been included in studies and, thus far, have never been selectively investigated in a larger patient population. Therefore, the aim of this study was to investigate safety, effectiveness and compliance of the WCD, as well as the incidence of VT/VFs, the development of LVEF and potential differences between patients who underwent certain types of cardiac surgery or ICD explantation.

METHODS

Ethics statement

According to the Ethics Committee of the Hannover Medical School (No. 1673-2013), Hannover, Germany, a retrospective analysis like ours handling with anonymized patient data does not require a specific assessment by the ethics committee. Patient data were anonymized.

Study design and patient population

In this retrospective study, 10 German cardiac surgery centres included 1168 patients discharged with a WCD (LifeVest® Wearable Cardioverter Defibrillator; ZOLL, Pittsburgh, PA, USA) between 2010 and 2020. Based on current guidelines, cardiac surgery patients with postoperative LVEF ≤35% or after ICD explantation due to infection were prescribed with a WCD. Diagnosis and therapy strategy was assessed by a heart team of surgeons, clinical cardiologists and electrophysiologists, whenever possible, to obtain optimal decision-making.

Materials (wearable cardioverter defibrillator and LifeVest Network)

The WCD is a non-invasive external defibrillator, first introduced in the USA and Germany as early as 2001. It consists of 4 electrocardiogram electrodes for continuous rhythm analysis and 3 shock electrodes as well as a monitoring system, which reliably detects and treats life-threatening VT/VF without bystander intervention. All detected arrhythmias, as well as several other parameters, such as heart rate, electrocardiogram, arrhythmic events and wear time, are documented and transferred to the LifeVest Network (LVN). Based on this information, the physician can adjust therapy and medication as needed. Patients are trained to press 2 response buttons when conscious during an initiated shock sequence, thereby avoiding inadequate and unnecessary shocks. Components and detailed functionality of the WCD are described elsewhere [28].

Data availability statement

All relevant data are within the manuscript and its supporting information files. Further information can be accessed via the corresponding author.

Data collection and study end points

Clinical data were retrospectively analysed by the study centres. The ethics committee approved the study [decision of the ‘Ethics commitee Medizinische Hochschule Hannover (MHH)’, August 2019].

Data were collected on age, sex, type of cardiac surgery, time of surgery, diagnosis, LVEF at the time of prescription and at follow-up (FU) after 3 months, as well as previous atrial fibrillation (AF) events. In addition, data were collected on WCD wear time, VT events, time of event occurrence and WCD treatments. The LVN was used for assessment and analysis, which automatically processes and stores data.

The main aim of the study was to examine the outcome of cardiac surgery patients wearing a WCD, including follow-up from time of surgery until end of WCD therapy.

We thought to investigate the occurrence of VT/VF, appropriate WCD treatments and their respective temporal relationship to surgery, as well as the LVEF improvement and its influence on a subsequent ICD indication. Furthermore, compliance for the WCD was assessed.

Statistical analysis

Continuous variables were reported as median with interquartile range, and categorical variables were reported as numbers and percentages for each group and combined groups, respectively. Kruskal–Wallis rank-sum test followed by pairwise Wilcoxon rank-sum tests (adjusted by the Bonferroni method) was applied to test group differences. Kaplan–Meier curves and log-rank tests were performed by the R-package ‘Survival’. For all statistical analyses, the statistical software R-4.1.1 (R Core Team, 2021) was used (R Core Team. R: A Language and Environment for Statistical Computing. R Foundation for Statistical Computing, Vienna, Austria, 2021. https://www.R-project.org/).

RESULTS

We identified 1168 patients, discharged with a WCD after cardiac surgery intervention between 2010 and 2020. The median age in this patient cohort was 66 years [interquartile range (IQR) 57–73]. A total of 1016 (87%) were male and 152 (13%) female. A total of 678 patients received coronary artery bypass graft (CABG) surgery (172 in combination with heart valve intervention, and 506 as isolated coronary revascularization). In addition, 186 patients underwent isolated valve surgery. In 280 patients, ICD explantation was performed (see Table 1).

Table 1:

Baseline characteristics depending on the surgical procedure

OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Age (years), median [IQR]66.0 [57.0, 73.0]66.0 [59.0, 73.0]67.0 [59.0, 74.0]64.0 [54.0, 73.0]65.0 [55.0, 74.0]52.5 [35.5, 60.8]
Gender = female, n (%)152 (13.0)51 (10.1)13 (7.6)38 (20.4)45 (16.1)5 (20.8)
Days in hospital, median [IQR]15.0 [11.0, 20.0]14.0 [11.0, 18.0]17.0 [13.0, 21.0]17.0 [13.0, 23.0]14.0 [8.0, 26.0]14.0 [11.0, 25.0]
General ward, median [IQR]9.0 [5.0, 13.0]9.0 [5.0, 12.0]9.0 [5.0, 13.0]12.0 [7.0, 15.0]8.0 [3.0, 15.0]9.0 [5.0, 11.5]
Intermediate care, median [IQR]2.0 [0.0, 6.0]3.0 [0.0, 6.0]4.0 [1.0, 8.0]3.0 [0.0, 7.0]0.0 [0.0, 2.0]1.0 [0.0, 3.5]
Intensive care, median [IQR]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [1.0, 4.0]0.0 [0.0, 2.0]0.5 [0.0, 6.0]
VT_pre-OP, n (%)73 (8.6)25 (6.7)8 (6.5)3 (2.2)33 (15.7)4 (28.6)
VF_pre-OP, n (%)63 (7.4)35 (9.4)5 (4.1)4 (3.0)14 (6.7)5 (33.3)
nsVT_pre-OP, n (%)37 (4.4)17 (4.7)3 (2.5)2 (1.5)14 (6.9)1 (7.7)
VT_post-OP, n (%)81 (8.3)34 (8.1)12 (8.5)18 (11.0)16 (7.0)1 (6.2)
VF_post-OP, n (%)49 (5.0)20 (4.8)12 (8.4)13 (7.9)2 (0.9)2 (11.8)
nsVT_ post-OP, n (%)45 (4.6)14 (3.3)8 (5.6)12 (7.3)11 (4.8)0 (0.0)
AF history, n (%)220 (18.8)73 (14.4)38 (22.1)33 (17.7)71 (25.4)5 (20.8)
WCD usage time per day (h)/ compliance, median [IQR]23.4 [21.7, 23.8]23.4 [21.6, 23.8]23.5 [22.4, 23.9]23.4 [22.5, 23.8]23.4 [20.2, 23.8]22.2 [20.6, 23.7]
Usage period (days), median [IQR]65.0 [34.0, 90.0]74.0 [38.0, 92.0]69.5 [42.5, 93.0]74.0 [40.2, 91.0]48.0 [27.0, 74.2]41.0 [26.8, 64.8]
OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Age (years), median [IQR]66.0 [57.0, 73.0]66.0 [59.0, 73.0]67.0 [59.0, 74.0]64.0 [54.0, 73.0]65.0 [55.0, 74.0]52.5 [35.5, 60.8]
Gender = female, n (%)152 (13.0)51 (10.1)13 (7.6)38 (20.4)45 (16.1)5 (20.8)
Days in hospital, median [IQR]15.0 [11.0, 20.0]14.0 [11.0, 18.0]17.0 [13.0, 21.0]17.0 [13.0, 23.0]14.0 [8.0, 26.0]14.0 [11.0, 25.0]
General ward, median [IQR]9.0 [5.0, 13.0]9.0 [5.0, 12.0]9.0 [5.0, 13.0]12.0 [7.0, 15.0]8.0 [3.0, 15.0]9.0 [5.0, 11.5]
Intermediate care, median [IQR]2.0 [0.0, 6.0]3.0 [0.0, 6.0]4.0 [1.0, 8.0]3.0 [0.0, 7.0]0.0 [0.0, 2.0]1.0 [0.0, 3.5]
Intensive care, median [IQR]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [1.0, 4.0]0.0 [0.0, 2.0]0.5 [0.0, 6.0]
VT_pre-OP, n (%)73 (8.6)25 (6.7)8 (6.5)3 (2.2)33 (15.7)4 (28.6)
VF_pre-OP, n (%)63 (7.4)35 (9.4)5 (4.1)4 (3.0)14 (6.7)5 (33.3)
nsVT_pre-OP, n (%)37 (4.4)17 (4.7)3 (2.5)2 (1.5)14 (6.9)1 (7.7)
VT_post-OP, n (%)81 (8.3)34 (8.1)12 (8.5)18 (11.0)16 (7.0)1 (6.2)
VF_post-OP, n (%)49 (5.0)20 (4.8)12 (8.4)13 (7.9)2 (0.9)2 (11.8)
nsVT_ post-OP, n (%)45 (4.6)14 (3.3)8 (5.6)12 (7.3)11 (4.8)0 (0.0)
AF history, n (%)220 (18.8)73 (14.4)38 (22.1)33 (17.7)71 (25.4)5 (20.8)
WCD usage time per day (h)/ compliance, median [IQR]23.4 [21.7, 23.8]23.4 [21.6, 23.8]23.5 [22.4, 23.9]23.4 [22.5, 23.8]23.4 [20.2, 23.8]22.2 [20.6, 23.7]
Usage period (days), median [IQR]65.0 [34.0, 90.0]74.0 [38.0, 92.0]69.5 [42.5, 93.0]74.0 [40.2, 91.0]48.0 [27.0, 74.2]41.0 [26.8, 64.8]

AF: atrial fibrillation; CABG: coronary artery bypass graft; ICD: implantable cardioverter defibrillator; IQR: interquartile range; OP: surgery; VF: ventricular fibrillation; VT: ventricular tachycardia; WCD: wearable cardioverter defibrillator.

Table 1:

Baseline characteristics depending on the surgical procedure

OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Age (years), median [IQR]66.0 [57.0, 73.0]66.0 [59.0, 73.0]67.0 [59.0, 74.0]64.0 [54.0, 73.0]65.0 [55.0, 74.0]52.5 [35.5, 60.8]
Gender = female, n (%)152 (13.0)51 (10.1)13 (7.6)38 (20.4)45 (16.1)5 (20.8)
Days in hospital, median [IQR]15.0 [11.0, 20.0]14.0 [11.0, 18.0]17.0 [13.0, 21.0]17.0 [13.0, 23.0]14.0 [8.0, 26.0]14.0 [11.0, 25.0]
General ward, median [IQR]9.0 [5.0, 13.0]9.0 [5.0, 12.0]9.0 [5.0, 13.0]12.0 [7.0, 15.0]8.0 [3.0, 15.0]9.0 [5.0, 11.5]
Intermediate care, median [IQR]2.0 [0.0, 6.0]3.0 [0.0, 6.0]4.0 [1.0, 8.0]3.0 [0.0, 7.0]0.0 [0.0, 2.0]1.0 [0.0, 3.5]
Intensive care, median [IQR]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [1.0, 4.0]0.0 [0.0, 2.0]0.5 [0.0, 6.0]
VT_pre-OP, n (%)73 (8.6)25 (6.7)8 (6.5)3 (2.2)33 (15.7)4 (28.6)
VF_pre-OP, n (%)63 (7.4)35 (9.4)5 (4.1)4 (3.0)14 (6.7)5 (33.3)
nsVT_pre-OP, n (%)37 (4.4)17 (4.7)3 (2.5)2 (1.5)14 (6.9)1 (7.7)
VT_post-OP, n (%)81 (8.3)34 (8.1)12 (8.5)18 (11.0)16 (7.0)1 (6.2)
VF_post-OP, n (%)49 (5.0)20 (4.8)12 (8.4)13 (7.9)2 (0.9)2 (11.8)
nsVT_ post-OP, n (%)45 (4.6)14 (3.3)8 (5.6)12 (7.3)11 (4.8)0 (0.0)
AF history, n (%)220 (18.8)73 (14.4)38 (22.1)33 (17.7)71 (25.4)5 (20.8)
WCD usage time per day (h)/ compliance, median [IQR]23.4 [21.7, 23.8]23.4 [21.6, 23.8]23.5 [22.4, 23.9]23.4 [22.5, 23.8]23.4 [20.2, 23.8]22.2 [20.6, 23.7]
Usage period (days), median [IQR]65.0 [34.0, 90.0]74.0 [38.0, 92.0]69.5 [42.5, 93.0]74.0 [40.2, 91.0]48.0 [27.0, 74.2]41.0 [26.8, 64.8]
OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Age (years), median [IQR]66.0 [57.0, 73.0]66.0 [59.0, 73.0]67.0 [59.0, 74.0]64.0 [54.0, 73.0]65.0 [55.0, 74.0]52.5 [35.5, 60.8]
Gender = female, n (%)152 (13.0)51 (10.1)13 (7.6)38 (20.4)45 (16.1)5 (20.8)
Days in hospital, median [IQR]15.0 [11.0, 20.0]14.0 [11.0, 18.0]17.0 [13.0, 21.0]17.0 [13.0, 23.0]14.0 [8.0, 26.0]14.0 [11.0, 25.0]
General ward, median [IQR]9.0 [5.0, 13.0]9.0 [5.0, 12.0]9.0 [5.0, 13.0]12.0 [7.0, 15.0]8.0 [3.0, 15.0]9.0 [5.0, 11.5]
Intermediate care, median [IQR]2.0 [0.0, 6.0]3.0 [0.0, 6.0]4.0 [1.0, 8.0]3.0 [0.0, 7.0]0.0 [0.0, 2.0]1.0 [0.0, 3.5]
Intensive care, median [IQR]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [0.0, 3.0]1.0 [1.0, 4.0]0.0 [0.0, 2.0]0.5 [0.0, 6.0]
VT_pre-OP, n (%)73 (8.6)25 (6.7)8 (6.5)3 (2.2)33 (15.7)4 (28.6)
VF_pre-OP, n (%)63 (7.4)35 (9.4)5 (4.1)4 (3.0)14 (6.7)5 (33.3)
nsVT_pre-OP, n (%)37 (4.4)17 (4.7)3 (2.5)2 (1.5)14 (6.9)1 (7.7)
VT_post-OP, n (%)81 (8.3)34 (8.1)12 (8.5)18 (11.0)16 (7.0)1 (6.2)
VF_post-OP, n (%)49 (5.0)20 (4.8)12 (8.4)13 (7.9)2 (0.9)2 (11.8)
nsVT_ post-OP, n (%)45 (4.6)14 (3.3)8 (5.6)12 (7.3)11 (4.8)0 (0.0)
AF history, n (%)220 (18.8)73 (14.4)38 (22.1)33 (17.7)71 (25.4)5 (20.8)
WCD usage time per day (h)/ compliance, median [IQR]23.4 [21.7, 23.8]23.4 [21.6, 23.8]23.5 [22.4, 23.9]23.4 [22.5, 23.8]23.4 [20.2, 23.8]22.2 [20.6, 23.7]
Usage period (days), median [IQR]65.0 [34.0, 90.0]74.0 [38.0, 92.0]69.5 [42.5, 93.0]74.0 [40.2, 91.0]48.0 [27.0, 74.2]41.0 [26.8, 64.8]

AF: atrial fibrillation; CABG: coronary artery bypass graft; ICD: implantable cardioverter defibrillator; IQR: interquartile range; OP: surgery; VF: ventricular fibrillation; VT: ventricular tachycardia; WCD: wearable cardioverter defibrillator.

Table 2:

Wearable cardioverter defibrillator recordings

OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Patients with VT/VF (%)106 (9.1)36 (7.1)19 (11.0)17 (9.1)34 (12.1)0 (0.0)
 Appropriate shock, n (%)18 (1.5)7 (1.4)1 (0.6)2 (1.1)8 (2.9)0 (0.0)
 Withholding response, n (%)17 (1.5)3 (0.6)7 (4.1)4 (2.2)3 (1.1)0 (0.0)
 Non-sustained VT, n (%)63 (5.4)20 (3.9)11 (6.4)10 (5.3)22 (7.8)0 (0.0)
 Below detection rate, n (%)7 (0.6)5 (1.0)0 (0.0)1 (0.5)1 (0.4)0 (0.0)
No. VT/VF episodes, n (%)250 (21.4)81 (16.0)75 (43.6)28 (15.1)66 (23.6)0 (0.0)
No. appropriate shocks, n (%)26 (2.2)11 (2.2)3 (1.7)2 (1.1)10 (3.6)0 (0.0)
AF newly detected, n (%)92 (7.9)34 (6.7)21 (12.2)24 (12.9)13 (4.6)0 (0.0)
OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Patients with VT/VF (%)106 (9.1)36 (7.1)19 (11.0)17 (9.1)34 (12.1)0 (0.0)
 Appropriate shock, n (%)18 (1.5)7 (1.4)1 (0.6)2 (1.1)8 (2.9)0 (0.0)
 Withholding response, n (%)17 (1.5)3 (0.6)7 (4.1)4 (2.2)3 (1.1)0 (0.0)
 Non-sustained VT, n (%)63 (5.4)20 (3.9)11 (6.4)10 (5.3)22 (7.8)0 (0.0)
 Below detection rate, n (%)7 (0.6)5 (1.0)0 (0.0)1 (0.5)1 (0.4)0 (0.0)
No. VT/VF episodes, n (%)250 (21.4)81 (16.0)75 (43.6)28 (15.1)66 (23.6)0 (0.0)
No. appropriate shocks, n (%)26 (2.2)11 (2.2)3 (1.7)2 (1.1)10 (3.6)0 (0.0)
AF newly detected, n (%)92 (7.9)34 (6.7)21 (12.2)24 (12.9)13 (4.6)0 (0.0)

AF: atrial fibrillation; CABG: coronary artery bypass graft; ICD: implantable cardioverter defibrillator; VF: ventricular fibrillation; VT: ventricular tachycardia.

Table 2:

Wearable cardioverter defibrillator recordings

OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Patients with VT/VF (%)106 (9.1)36 (7.1)19 (11.0)17 (9.1)34 (12.1)0 (0.0)
 Appropriate shock, n (%)18 (1.5)7 (1.4)1 (0.6)2 (1.1)8 (2.9)0 (0.0)
 Withholding response, n (%)17 (1.5)3 (0.6)7 (4.1)4 (2.2)3 (1.1)0 (0.0)
 Non-sustained VT, n (%)63 (5.4)20 (3.9)11 (6.4)10 (5.3)22 (7.8)0 (0.0)
 Below detection rate, n (%)7 (0.6)5 (1.0)0 (0.0)1 (0.5)1 (0.4)0 (0.0)
No. VT/VF episodes, n (%)250 (21.4)81 (16.0)75 (43.6)28 (15.1)66 (23.6)0 (0.0)
No. appropriate shocks, n (%)26 (2.2)11 (2.2)3 (1.7)2 (1.1)10 (3.6)0 (0.0)
AF newly detected, n (%)92 (7.9)34 (6.7)21 (12.2)24 (12.9)13 (4.6)0 (0.0)
OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
Patients with VT/VF (%)106 (9.1)36 (7.1)19 (11.0)17 (9.1)34 (12.1)0 (0.0)
 Appropriate shock, n (%)18 (1.5)7 (1.4)1 (0.6)2 (1.1)8 (2.9)0 (0.0)
 Withholding response, n (%)17 (1.5)3 (0.6)7 (4.1)4 (2.2)3 (1.1)0 (0.0)
 Non-sustained VT, n (%)63 (5.4)20 (3.9)11 (6.4)10 (5.3)22 (7.8)0 (0.0)
 Below detection rate, n (%)7 (0.6)5 (1.0)0 (0.0)1 (0.5)1 (0.4)0 (0.0)
No. VT/VF episodes, n (%)250 (21.4)81 (16.0)75 (43.6)28 (15.1)66 (23.6)0 (0.0)
No. appropriate shocks, n (%)26 (2.2)11 (2.2)3 (1.7)2 (1.1)10 (3.6)0 (0.0)
AF newly detected, n (%)92 (7.9)34 (6.7)21 (12.2)24 (12.9)13 (4.6)0 (0.0)

AF: atrial fibrillation; CABG: coronary artery bypass graft; ICD: implantable cardioverter defibrillator; VF: ventricular fibrillation; VT: ventricular tachycardia.

Table 3:

Left ventricular ejection fraction progress

OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
LVEF
LVEF_pre-OP, median [IQR]29.0 [20.5, 37.0]29.0 [23.5, 35.0]29.0 [22.0, 40.0]29.5 [20.0, 40.0]27.5 [20.0, 45.0]20.0 [15.0, 21.5]
LVEF_post-OP, median [IQR]28.0 [22.0, 32.0]28.0 [24.0, 30.0]27.0 [20.0, 30.0]26.0 [21.0, 30.0]30.0 [22.0, 40.5]27.0 [20.0, 32.0]
LVEF_end of use, median [IQR]35.0 [28.0, 42.0]35.0 [30.0, 41.0]36.0 [30.0, 40.8]40.0 [30.0, 41.0]31.0 [25.0, 45.0]27.5 [23.8, 41.2]
End of WCD use
Improvement in LVEF, n (%)471 (42.2)265 (54.4)78 (47.9)103 (58.2)22 (8.3)3 (13.0)
ICD (re)implantation, n (%)546 (48.6)184 (37.7)70 (42.9)59 (32.8)217 (80.4)16 (69.6)
Miscellaneous,an (%)124 (10.6)49 (9.7)21 (12.2)22 (11.8)29 (10.4)3 (12.5)
No data27832122
OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
LVEF
LVEF_pre-OP, median [IQR]29.0 [20.5, 37.0]29.0 [23.5, 35.0]29.0 [22.0, 40.0]29.5 [20.0, 40.0]27.5 [20.0, 45.0]20.0 [15.0, 21.5]
LVEF_post-OP, median [IQR]28.0 [22.0, 32.0]28.0 [24.0, 30.0]27.0 [20.0, 30.0]26.0 [21.0, 30.0]30.0 [22.0, 40.5]27.0 [20.0, 32.0]
LVEF_end of use, median [IQR]35.0 [28.0, 42.0]35.0 [30.0, 41.0]36.0 [30.0, 40.8]40.0 [30.0, 41.0]31.0 [25.0, 45.0]27.5 [23.8, 41.2]
End of WCD use
Improvement in LVEF, n (%)471 (42.2)265 (54.4)78 (47.9)103 (58.2)22 (8.3)3 (13.0)
ICD (re)implantation, n (%)546 (48.6)184 (37.7)70 (42.9)59 (32.8)217 (80.4)16 (69.6)
Miscellaneous,an (%)124 (10.6)49 (9.7)21 (12.2)22 (11.8)29 (10.4)3 (12.5)
No data27832122
a

Miscellaneous (multiple answers possible) = died (18, 1.5%), refuses ICD (16, 1.4%), non-compliance (23, 2.0%), discomfort (6, 0.5%), refuses WCD (15, 1.3%), and others (51, 4.4%).

CABG: coronary artery bypass graft; ICD: implantable cardioverter defibrillator; IQR: interquartile range; LVEF: left ventricular ejection fraction; OP: surgery; WCD: wearable cardioverter defibrillator.

Table 3:

Left ventricular ejection fraction progress

OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
LVEF
LVEF_pre-OP, median [IQR]29.0 [20.5, 37.0]29.0 [23.5, 35.0]29.0 [22.0, 40.0]29.5 [20.0, 40.0]27.5 [20.0, 45.0]20.0 [15.0, 21.5]
LVEF_post-OP, median [IQR]28.0 [22.0, 32.0]28.0 [24.0, 30.0]27.0 [20.0, 30.0]26.0 [21.0, 30.0]30.0 [22.0, 40.5]27.0 [20.0, 32.0]
LVEF_end of use, median [IQR]35.0 [28.0, 42.0]35.0 [30.0, 41.0]36.0 [30.0, 40.8]40.0 [30.0, 41.0]31.0 [25.0, 45.0]27.5 [23.8, 41.2]
End of WCD use
Improvement in LVEF, n (%)471 (42.2)265 (54.4)78 (47.9)103 (58.2)22 (8.3)3 (13.0)
ICD (re)implantation, n (%)546 (48.6)184 (37.7)70 (42.9)59 (32.8)217 (80.4)16 (69.6)
Miscellaneous,an (%)124 (10.6)49 (9.7)21 (12.2)22 (11.8)29 (10.4)3 (12.5)
No data27832122
OverallCABGCABG with valveValve surgeryICD explantationMisc
n116850617218628024
LVEF
LVEF_pre-OP, median [IQR]29.0 [20.5, 37.0]29.0 [23.5, 35.0]29.0 [22.0, 40.0]29.5 [20.0, 40.0]27.5 [20.0, 45.0]20.0 [15.0, 21.5]
LVEF_post-OP, median [IQR]28.0 [22.0, 32.0]28.0 [24.0, 30.0]27.0 [20.0, 30.0]26.0 [21.0, 30.0]30.0 [22.0, 40.5]27.0 [20.0, 32.0]
LVEF_end of use, median [IQR]35.0 [28.0, 42.0]35.0 [30.0, 41.0]36.0 [30.0, 40.8]40.0 [30.0, 41.0]31.0 [25.0, 45.0]27.5 [23.8, 41.2]
End of WCD use
Improvement in LVEF, n (%)471 (42.2)265 (54.4)78 (47.9)103 (58.2)22 (8.3)3 (13.0)
ICD (re)implantation, n (%)546 (48.6)184 (37.7)70 (42.9)59 (32.8)217 (80.4)16 (69.6)
Miscellaneous,an (%)124 (10.6)49 (9.7)21 (12.2)22 (11.8)29 (10.4)3 (12.5)
No data27832122
a

Miscellaneous (multiple answers possible) = died (18, 1.5%), refuses ICD (16, 1.4%), non-compliance (23, 2.0%), discomfort (6, 0.5%), refuses WCD (15, 1.3%), and others (51, 4.4%).

CABG: coronary artery bypass graft; ICD: implantable cardioverter defibrillator; IQR: interquartile range; LVEF: left ventricular ejection fraction; OP: surgery; WCD: wearable cardioverter defibrillator.

In 106 (9.1%) patients, 250 VT/VF episodes occurred (Table 2). Among these, 26 episodes in 18 patients (1.5%) were appropriately defibrillated by the WCD. All shocks were successful. Eight (0.7%) patients received a total number of 9 inadequate shocks, of whom 4 patients had supraventricular tachycardias with rapid transition. In 1 patient, the supraventricular tachycardia was of haemodynamic relevance. Four inadequate shocks were caused by artefact. A period of asystole was documented in 7 (0.6%) patients.

Of 506 patients with isolated CABG, 36 (7.1%) patients developed 81 VT/VF episodes. Of these, 7 (1.4%) patients received 11 adequate WCD shocks, whereas in 3 patients (0.6%) the response buttons to withhold a shock during consciousness were pressed. Twenty (3.9%) patients had non-sustained ventricular tachycardias (nsVTs). In the combined CABG and valve replacement group (n = 172), 19 (11%) patients developed 75 VT/VF episodes, of whom 1 (0.6%) patient received 3 adequate shocks, 7 (4.1%) patients pressed the response buttons to suppress impending defibrillation and 11 (6.4%) patients had nsVTs. Patients with isolated valve surgery (n = 186) developed 28 VT/VF episodes in 17 (9.1%) cases. Among these, 2 (1.1%) patients received adequate shocks due to VT/VF, whereas 4 (2.2%) patients pressed the response buttons. Ten (5.3%) patients had nsVTs. Of 280 patients undergoing ICD explantation, 34 patients (12.1%) had 66 VT/VF episodes, of whom 8 (2.9%) patients received 10 adequate shocks and 3 (1.1%) used the response buttons to prevent defibrillation. Moreover, 22 (7.8%) patients had nsVTs. No VT/VF episode was detected in the group of miscellaneous surgical procedures. The prevalence of VT/VF episodes in relation to the time of surgical intervention is shown in Fig. 1. The probability of WCD shock delivery depending on the surgical method is displayed in Fig. 2. The estimated survival probability without wearable cardioverter defibrillator is shown in Fig. 3 and Supplementary Material.

The prevalence of ventricular tachycardia/ventricular fibrillation episodes in relation to the time of surgical intervention.
Figure 1:

The prevalence of ventricular tachycardia/ventricular fibrillation episodes in relation to the time of surgical intervention.

Probability of wearable cardioverter defibrillator shock depending on the surgical method.
Figure 2:

Probability of wearable cardioverter defibrillator shock depending on the surgical method.

(A) Estimated survival probability without wearable cardioverter defibrillator per procedure. The vertical axis was cut for better readability of the data curves. (B) Estimated survival probability without wearable cardioverter defibrillator of implantable cardioverter defibrillator explant patients versus all other patients. The vertical axis was cut for better readability of the data curves.
Figure 3:

(A) Estimated survival probability without wearable cardioverter defibrillator per procedure. The vertical axis was cut for better readability of the data curves. (B) Estimated survival probability without wearable cardioverter defibrillator of implantable cardioverter defibrillator explant patients versus all other patients. The vertical axis was cut for better readability of the data curves.

(A) Left ventricular ejection fraction development per surgical procedure. Shown are medians previous to surgery, at discharge, and at end of wearable cardioverter defibrillator use. (B) Box plots. Left ventricular ejection fraction differences from discharge (0) to end of wearable cardioverter defibrillator use per procedure.
Figure 4:

(A) Left ventricular ejection fraction development per surgical procedure. Shown are medians previous to surgery, at discharge, and at end of wearable cardioverter defibrillator use. (B) Box plots. Left ventricular ejection fraction differences from discharge (0) to end of wearable cardioverter defibrillator use per procedure.

The median time to first VT/VF episode with and without shock was 17 days (IQR 6–40) after start of WCD therapy and 29 days (IQR 17–51) after surgery.

A total of 220 patients (18.8%) had a prior history of AF. In addition, the WCD detected AF for the first time in 92 patients (7.9%). New diagnosis of AF most frequently occurred in the groups of isolated valve surgery (n = 21, 12.2%) and combined CABG and valve surgery (n = 24, 12.9%).

The median wearing period was 65 days (IQR: 34–90). The longest wearing period was after valve surgery with a median of 74 days (IQR: 40–91) and after isolated CABG with a median of 74 days (IQR: 38–92). ICD explant patients showed the shortest wearing periods with 48 days (IQR: 27–74). Daily wear time in the entire patient cohort was a median of 23.4 h per day (IQR: 21.7–23.8). There were no significant differences in daily wear time depending on clinical procedure nor sex.

Postoperatively, the median LVEF of the total population was 28% (IQR: 22–32) and improved to a median of 35% (IQR: 28–42) at FU examination (Table 3). Patients in the CABG, valve and CABG plus valve groups displayed a significant median LVEF difference of between +8% and 10%. The ICD explant group did not show LVEF increase in median. However, LVEF improvement seemed to be quite individual. While patients in the ICD explant and miscellaneous cohort showed an LVEF increase in median from pre-OP to post-OP and end of FU, the LVEF of patients in the CABG, valve and combined CABG with valve groups decreased from pre-OP to post-OP with a pronounced improvement to FU. The left ventricular ejection fraction development per surgical procedure is shown in Fig. 4.

After WCD use, ICD implantation was performed in 546 (48.6%) patients of the entire population. In 471 (42.2%) patients, an ICD implantation could be avoided due to an improved LVEF.

In the ICD explant patient cohort, reimplantation was necessary in 217 (80.4%) patients, while ICD reimplantation could be avoided due to LVEF improvement in 22 (8.3%) individuals.

In the cohort without ICD explantation, 329 (37%) patients finally received an ICD, while further device-based prevention prior to SCD was no longer necessary in 449 (50.6%) patients.

Most ICDs could be avoided after isolated CABG and isolated valve surgery. Here, WCD therapy could be terminated due to an LVEF improvement in 265 (54.4%) and 103 (58.2%) patients. ICD implantation was still performed in 184 (37.7%) and 59 (32.8%) patients.

Most patients could be allocated to the end-of-WCD-use reasons ICD implantation or LVEF improvement. However, 124 (10.6%) patients had different reasons and for 27 patients, no data were available for this parameter. Reasons for "Miscellaneous end of use" included patient refuses WCD, death, non-compliance, discomfort, still active use, end of prescription and others. Furthermore, a reason for a not implanted ICD was refusal of an ICD by the patient.

A total of 18 patients died during the observation period. All patients were in hospital, not wearing the WCD at the time of death. Mortality was significantly higher for explant patients compared to all other patients.

DISCUSSION

This retrospective, multicentre study presents WCD experience in a large, strictly cardiac surgery patient population. Effectiveness, safety and high WCD compliance in nonsurgical patients have been consistently demonstrated in recent years. However, direct transferability of the favourable clinical results in mixed patient populations to a purely cardiac surgery collective does not appear to be imperative because of the specific periprocedural circumstances such as sternotomy or the use of a heart–lung machine and the need for rapid rehabilitation. We therefore investigated the extent to which the available evidence can be transferred to cardiac surgery patients.

Ventricular arrhythmias and shocks

The rate of patients developing VT/VFs is substantial (250 ventricular arrhythmias; 9.1%). The highest proportion of patients with VT/VF occurred in the ICD explant group, followed by CABG with valve, valve surgery alone and CABG alone. The most patients with appropriate shocks were found in the ICD explant group, followed by CABG alone, valve surgery and CABG with valve.

All recorded VT/VFs occurred within the first 180 postoperative days and could be detected from the first day of wearing the WCD. However, the majority of arrhythmic events occurred within the first 90 days.

The rate of 1.5% of adequately treated patients found in our collective aligns with current European studies in various large patient collectives. Wäßnig (n = 6043), Garcia (n = 1157) and Veltmann (n = 781) reported similar adequate shock rates in mixed patient populations (1.6%; 1.6%; 1.3%, respectively). Our study documented an inadequate shock rate of 0.7%, which is also comparable to previous studies (0.4%, 0.7% and 0.3%, respectively) [19, 22, 23].

The cause of inadequate shocks in 4 of 8 patients was supraventricular arrhythmias and artificial rhythm detection in the remaining 4. More recent WCD studies show a trend towards decreasing rates of inadequate shock delivery, which could be attributed to an update of the WCD detection algorithm. According to the manufacturer ZOLL, this was accomplished in 2018 to reduce artefacts that led to misinterpretation of the detected heart rhythm.

Wearing time

The daily wearing time of our collective was overall very good in all patient cohorts, with a median of 23.4 h per day. This is especially remarkable as sternotomy was frequent and wound-healing processes were still ongoing. Similarly, patients participated in rehabilitation activities that did not prevent them from wearing the WCD except when swimming. No difference was observed between sexes. The total median wearing time of 65 days did not vary noticeably between the subgroups. However, there was an exception for patients who underwent ICD explantation. Depending on an already existing preoperatively ICD indication, patients received an ICD much more quickly and therefore showed a shorter WCD usage with a median of 48 days. Overall, recorded total wearing times and in particular daily wearing times were in line with previous study results in non-surgical or mixed patient populations, ranging from a median of 22.6–23.5 h per day [19, 23–25].

Thus, it can be concluded that the WCD shows a good acceptance in cardiac surgery patients, which does not differ from other patient collectives despite specific postoperative challenges. In general, detailed patient education and training is crucial. Online patient management with the LVN can be helpful to ensure compliance.

Development of left ventricular ejection fraction and follow-up care after wearable cardioverter defibrillator use

Postoperatively, many cardiac surgery patients show severely reduced LVEF, which is associated with elevated risk of SCD. In our study, the initial median LVEF of 28% improved during ∼3 months of WCD use to a median of 35% (28–42%). Except for patients with ICD explantation, LVEF subgroup analyses did not show significant differences in the increase of pump function between the various clinical procedures. Box plot analyses showed LVEF differences between discharge and FU of +8–10% for patients with CABG and valve surgery. While patients with ICD explant and miscellaneous interventions increased their LVEF from pre- to post-surgery and FU, patients with CABG and valve surgery experienced an LVEF decrease post-surgery with a steep and prominent increase to FU.

Remarkably, a substantial proportion of all patients recovered above the threshold value of an LVEF >35%, which represents a criterion for permanent ICD indication. Excluding the ICD explantation cohort, ICD implantation was no longer indicated in 50.6% of patients. This is accomplished by drug therapy, which must be slowly adjusted over several months, and by the progressive regenerative processes of the heart itself. The results in terms of LVEF recovery are also consistent with previous studies [22–24].

The high burden of VT/VF within the first months might raise the question of early ICD implantation. However, our study results clearly underline the recommended waiting times required by current guidelines as well as the reliable exclusion of reversible causes prior to permanent ICD implantation. In particular, patients who recover after CABG or valve surgery show a significant improvement in LVEF within 3–6 months and at the same time confirm their willingness to wear a WCD with a high therapy compliance. Therefore, wearing the device during this temporary high-risk phase can provide safe and effective protection from SCD and simultaneously avoid permanent ICD implantation in a relevant proportion of patients. Prematurely implanted ICDs may lose their justification after the end of the recovery period and, if there is no indication, only result in complications such as device infections or lead failure. Thus, the use of a WCD can overall improve the patient's quality of life and reduce the financial burden on health care systems. A recent health technology assessment on WCD demonstrated cost-effectiveness in patients after myocardial infarction and cost-savings after ICD explantation [29].

CONCLUSION

In conclusion, this study demonstrates that the risk of VT and life-threatening tachyarrhythmias is highest within the first 3 months in cardiac surgery patients with LVEF ≤35%, and patients with ICD explant, regardless of the procedure. These patients can be protected with a WCD. During the observation period, LVEF improved significantly in non-ICD explant patients, avoiding permanent ICD implantation in a substantial part of patients. Furthermore, compliance was very good despite sternotomy.

Available clinical evidence consistently demonstrates high efficacy, safety and compliance with a WCD in patients at increased transient risk of SCD due to severely impaired LVEF. This particularly applies to the postoperative care of cardiac surgery patients for whom, depending on the ventricular function and the individual risk of SCD, temporary WCD protection should be recommended.

SUPPLEMENTARY MATERIAL

Supplementary material is available at EJCTS online.

Conflict of interest: Christian Kuehn reports honorarium fees from ZOLL. The other authors report no conflict of interest.

Author contributions

Christian Kuehn: Conceptualization; Formal analysis; Writing—original draft. Stefan Ruemke: Data curation; Investigation; Writing—review & editing. Philipp Rellecke: Data curation; Formal analysis; Investigation; Writing—review & editing. Artur Lichtenberg: Formal analysis; Investigation; Writing—review & editing. Dominik Joskowiak: Data curation; Formal analysis; Investigation; Writing—review & editing. Christian Hagl: Formal analysis; Investigation; Writing—review & editing. Mohamed Hassan: Data curation; Writing—review & editing. Rainer G. Leyh: Formal analysis; Investigation; Writing—review & editing. Stefan Erler: Data curation; Formal analysis; Writing—review & editing. Jens Garbade: Conceptualization; Investigation; Validation; Writing—review & editing. Sandra Eifert: Data curation; Investigation; Writing—review & editing. Philippe Grieshaber: Data curation; Investigation; Writing—review & editing. Andreas Boening: Conceptualization; Investigation; Supervision; Writing—review & editing. Torsten Doenst: Conceptualization; Investigation; Supervision; Writing—review & editing. Ilia Velichkov: Data curation; Writing—review & editing. Tomas Madej: Data curation; Writing—review & editing. Michael Knaut: Conceptualization; Data curation; Supervision; Writing—review & editing. Andreas Hain: Data curation; Investigation; Validation; Writing—review & editing. Heiko Burger: Conceptualization; Investigation; Validation; Writing—original draft.

Reviewer information

European Journal of Cardio-Thoracic Surgery thanks Pradeep Narayan, Ulrich Otto von Oppell and the other, anonymous reviewer(s) for their contribution to the peer review process of this article.

Presented at the 49th Annual Meeting of the German Society for Thoracic and Cardiovascular Surgery, Wiesbaden, Germany, 29 February–3 March 2020.

REFERENCES

1

Gräsner
J-T
,
Lefering
R
,
Koster
RW
,
Masterson
S
,
Böttiger
BW
,
Herlitz
J
et al.
EuReCa ONE-27 Nations, ONE Europe, ONE Registry: a prospective one month analysis of out-of-hospital cardiac arrest outcomes in 27 countries in Europe
.
Resuscitation
2016
;
105
:
188
95
.

2

Srinivasan
NT
,
Schilling
RJ.
Sudden cardiac death and arrhythmias
.
Arrhythm Electrophysiol Rev
2018
;
7
:
111
7
.

3

Moss
AJ
,
Zareba
W
,
Hall
WJ
,
Klein
H
,
Wilber
DJ
,
Cannom
DS
et al. ; Multicenter Automatic Defibrillator Implantation Trial II Investigators.
Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction
.
N Engl J Med
2002
;
346
:
877
83
.

4

Moss
AJ
,
Hall
WJ
,
Cannom
DS
,
Daubert
JP
,
Higgins
SL
,
Klein
H
et al.
Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators
.
N Engl J Med
1996
;
335
:
1933
40
.

5

Appoo
J
,
Norris
C
,
Merali
S
,
Graham
MM
,
Koshal
A
,
Knudtson
ML
et al.
Long-term outcome of isolated coronary artery bypass surgery in patients with severe left ventricular dysfunction
.
Circulation
2004
;
110
:
II-13
II-17
.

6

Rao
MP
,
Al-Khatib
SM
,
Pokorney
SD
,
She
L
,
Romanov
A
,
Nicolau
JC
et al. ; STICH Trial Investigators.
Sudden cardiac death in patients with ischemic heart failure undergoing coronary artery bypass grafting: results from the STICH randomized clinical trial (Surgical Treatment for Ischemic Heart Failure)
.
Circulation
2017
;
135
:
1136
44
.

7

Heimeshoff
J
,
Merz
C
,
Ricklefs
M
,
Kirchhoff
F
,
Haverich
A
,
Bara
C
et al.
Wearable cardioverter–defibrillators following cardiac surgery—a single-center experience
.
Thorac Cardiovasc Surg
2019
;
67
:
92
7
.

8

Kaul
TK
,
Fields
BL
,
Riggins
LS
,
Wyatt
DA
,
Jones
CR.
Ventricular arrhythmia following successful myocardial revascularization: incidence, predictors and prevention
.
Eur J Cardiothorac Surg
1998
;
13
:
629
36
.

9

Weintraub
WS
,
Grau-Sepulveda
MV
,
Weiss
JM
,
Delong
ER
,
Peterson
ED
,
O'Brien
SM
et al.
Prediction of long-term mortality after percutaneous coronary intervention in older adults: results from the National Cardiovascular Data Registry
.
Circulation
2012
;
125
:
1501
10
.

10

Priori
SG
,
Blomström-Lundqvist
C
,
Mazzanti
A
,
Blom
N
,
Borggrefe
M
,
Camm
J
et al. ; ESC Scientific Document Group.
2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death
.
Eur Heart J
2015
;
36
:
2793
867
.

11

Al-Khatib
SM
,
Stevenson
WG
,
Ackerman
MJ
,
Bryant
WJ
,
Callans
DJ
,
Curtis
AB
et al.
2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death
.
Circulation
2018
;
138
:
e272
e391
.

12

Röger
S
,
Rosenkaimer
SL
,
Hohneck
A
,
Lang
S
,
El-Battrawy
I
,
Rudic
B
et al.
Therapy optimization in patients with heart failure: the role of the wearable cardioverter defibrillator in a real-world setting
.
BMC Cardiovasc Disord
2018
;
18
:
13
.

13

Bax
JJ
,
Visser
FC
,
Poldermans
D
,
Elhendy
A
,
Cornel
JH
,
Boersma
E
et al.
Relationship between preoperative viability and postoperative improvement in LVEF and heart failure symptoms
.
J Nucl Med
2001
;
42
:
79
86
.

14

Søraas
CL
,
Larstorp
AC
,
Mangschau
A
,
Tønnessen
T
,
Kjeldsen
SE
,
Bjørnerheim
R.
Echocardiographic demonstration of improved myocardial function early after coronary artery bypass graft surgery
.
Interact CardioVasc Thorac Surg
2011
;
12
:
946
51
.

15

Hohnloser
SH
,
Kuck
KH
,
Dorian
P
,
Roberts
R
,
Hampton
JR
,
Hatala
R
et al. ; DINAMIT Investigators.
Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction
.
N Engl J Med
2004
;
351
:
2481
8
.

16

Reek
S
,
Burri
H
,
Roberts
PR
,
Perings
C
,
Epstein
AE
,
Klein
HU
et al. ; EHRA Scientific Documents Committee (as external reviewers).
The wearable cardioverter-defibrillator: current technology and evolving indications. EHRA Scientific Documents Committee
.
Europace
2017
;
19
:
335
45
.

17

Burger
H
,
Schmitt
J
,
Knaut
M
,
Eitz
T
,
Starck
CT
,
Hakmi
S
et al.
Einsatz des tragbaren Kardioverter Defibrillators nach kardiochirurgischen Eingriffen [Use of wearable cardioverter defibrillators after cardiac surgery: position paper of the working group on cardiac arrhythmias of German Society of Thoracic and Cardiovascular Surgery]
.
Z Herz Thorax Gefäßchir
2018
;
32
:
286
99
.

18

Kutyifa
V
,
Moss
AJ
,
Klein
H
,
Biton
Y
,
McNitt
S
,
MacKecknie
B
et al.
Use of the wearable cardioverter defibrillator in high-risk cardiac patients: data from the Prospective Registry of Patients Using the Wearable Cardioverter Defibrillator (WEARIT-II Registry)
.
Circulation
2015
;
132
:
1613
9
.

19

Wäßnig
NK
,
Günther
M
,
Quick
S
,
Pfluecke
C
,
Rottstädt
F
,
Szymkiewicz
SJ
et al.
Experience with the wearable cardioverter-defibrillator in patients at high risk for sudden cardiac death
.
Circulation
2016
;
134
:
635
43
.

20

Olgin
JE
,
Pletcher
MJ
,
Vittinghoff
E
,
Wranicz
J
,
Malik
R
,
Morin
DP
et al. ; VEST Investigators.
Wearable cardioverter-defibrillator after myocardial infarction
.
N Engl J Med
2018
;
379
:
1205
15
.

21

Olgin
JE
,
Lee
BK
,
Vittinghoff
E
,
Morin
DP
,
Zweibel
S
,
Rashba
E
et al.
Impact of wearable cardioverter-defibrillator compliance on outcomes in the VEST trial: as-treated and per-protocol analyses
.
J Cardiovasc Electrophysiol
.
2020
;
31
:
1009
18
.

22

Veltmann
C
,
Winter
S
,
Duncker
D
,
Wäßnig
NK
,
Geller
JC
,
Erath
JW
et al.
Protected risk stratification with the wearable cardioverter-defibrillator: results from the WEARIT-II-EUROPE registry
.
Clin Res Cardiol
2021
;
110
:
151
.

23

Garcia
R
,
Combes
N
,
Defaye
P
,
Narayanan
K
,
Guedon-Moreau
L
,
Boveda
S
et al.
Wearable cardioverter-debrillator in patients with a transient risk of sudden cardiac death: the WEARIT-France cohort study
.
Europace
2021
;
23
:
73
81
.

24

Kovacs
B
,
Reek
S
,
Sticherling
C
,
Schaer
B
,
Linka
A
,
Ammann
P
et al.
Use of the wearable cardioverter-defibrillator—the Swiss experience
.
Swiss Med Wkly
2020
;
150
:
w20343
.

25

Nguyen
E
,
Weeda
ER
,
Kohn
CG
,
D'Souza
BA
,
Russo
AM
,
Noreika
S
et al.
Wearable cardioverter-defibrillators for the prevention of sudden cardiac death: a meta-analysis
.
J Innov Card Rhythm Manag
2018
;
9
:
3151
62
.

26

Odeneg
T
,
Ebner
C
,
Mortl
D
,
Keller
H
,
Dirninger
A
,
Stix
G
et al.
Indications for and outcome in patients with the wearable cardioverter-defibrillator in a nurse-based training programme: results of the Austrian WCD Registry
.
Eur J Cardiovasc Nurs
2019
;
18
:
75
83
.

27

Masri
A
,
Altibi
AM
,
Erqou
S
,
Zmaili
MA
,
Saleh
A
,
Al-Adham
R
et al.
Wearable cardioverter-defibrillator therapy for the prevention of sudden cardiac death: a systematic review and meta-analysis
.
JACC Clin Electrophysiol
2019
;
5
:
152
61
.

28

Klein
HU
,
Goldenberg
I
,
Moss
AJ.
Risk stratification for implantable cardioverter defibrillator therapy: the role of the wearable cardioverter-defibrillator
.
Eur Heart J
2013
;
34
:
2230
42
.

29

Cortesi
PA
,
Botto
GL
,
D’Angiolella
LS
,
Scalone
L
,
De Ponti
R
,
Boriani
G
et al.
Health technology assessment on the use of the wearable cardioverter defibrillator in patients with myocardial infarction and with ICD explant
.
Farmeconomia
2021
;
22
:
3
20
.

ABBREVIATIONS

     
  • AF

    Atrial fibrillation

  •  
  • CABG

    Coronary artery bypass graft

  •  
  • FU

    Follow-up

  •  
  • ICD

    Implantable cardioverter defibrillator

  •  
  • IQR

    Interquartile range

  •  
  • LVEF

    Left ventricular ejection fraction

  •  
  • LVN

    LifeVest Network

  •  
  • SCD

    Sudden cardiac death

  •  
  • nsVTs

    Non-sustained ventricular tachycardias

  •  
  • VF

    Ventricular fibrillation

  •  
  • VT

    Ventricular tachycardia

  •  
  • WCD

    Wearable cardioverter defibrillator

This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact [email protected]

Supplementary data