Abstract

OBJECTIVES:

EUROMACS is a registry of the European Association for Cardio-Thoracic Surgery (EACTS) whose purpose is to gather clinical data related to durable mechanical circulatory support for scientific purposes and to publish annual reports. Because the treatment of children with end-stage heart failure has several significantly different characteristics than the treatment of adults, data and outcomes of interventions are analysed in this dedicated paediatric report.

METHODS:

Participating hospitals contributed pre-, peri- and long-term postoperative data on mechanical circulatory support implants to the registry. Data for all implants in paediatric patients (≤19 years of age) performed from 1 January 2000 to 31 December 2017 were analysed. This report includes updates of patient characteristics, implant frequency, outcome (including mortality rates, transplants and recovery rates) as well as adverse events.

RESULTS:

Twenty-five hospitals contributed 237 registered implants in 210 patients (81 ♀, 129 ♂) to the registry. The most frequent diagnosis was any form of cardiomyopathy (71.4%) followed by congenital heart disease (18.6%). Overall mean support time on a device was 11.6 months (±16.5 standard deviation). A total of 173 children (82.4%) survived to transplant, recovery or are ongoing; 37 patients (17.6%) died while on support within the observed follow-up time. At 12 months 38% of patients received transplants, 7% were weaned from their device and 15% died. At 24 months, 51% of patients received transplants, 17% died while on support, 22% were on a device and 9% were explanted due to myocardial recovery. The adverse events rate per 100 patient-months was 11.97 for device malfunction, 2.83 for major bleeding, 2.83 for major infection and 1.52 for neurological events within the first 3 months after implantation.

CONCLUSIONS:

The first paediatric EUROMACS report reveals a low transplant rate in European countries within the first 2 years of implantation compared to US data. The 1-year survival rate seems to be satisfactory. Device malfunction including pump chamber changes due to thrombosis was the most frequent adverse event.

INTRODUCTION

The use of durable mechanical circulatory support (MCS) in children in the form of a ventricular assist device (VAD) has increased dramatically over the years and has improved survival for paediatric patients on the waiting list for a heart transplant [1]. Paediatric patients receiving MCS is a unique area of study due to the physical size of the recipient, which not only requires careful selection of an appropriately sized device but also different management techniques than those used in adults. Children require specially adapted pharmacological treatment and the prevention of adverse events requires a very different clinical management from that in adults.

The EUROMACS Committee of the European Association for Cardio-Thoracic Surgery (EACTS) governs the registry, which was launched in 2009 and became operational in 2012. EUROMACS is the only European-based durable MCS registry for all devices with the CE Marking implanted in children and adults (Table 1). The purpose of the registry is to gather clinical data related to durable MCS for scientific purposes and to publish annual reports. From the outset, all possible options in MCS strategy with respect to devices on the market and to data on patients of every age and geographic area were included [2]. This approach enables the registry not only to select paediatric patients as a distinguished patient cohort for analyses of baseline data but also to follow them up even after they have passed the age of 19 years. EUROMACS collects data continuing through the period of VAD support; there are 3 end points: transplantation, weaning and death. The EUROMACS database has been designed in such a way that the patient and the device outcomes will be comparable with the Pedimacs and Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) databases.

Table 1:

Present CE-marked mechanical circulatory support systems registered in the EUROMACS database

MCS type
Durable devices
 Continuous flowBerlin Heart INCOR
CircuLite SYNERGYa
HeartAssist 5
HeartWare HVAD
Jarvik 2000
MicroMed DeBakey
Thoratec HeartMate II
Thoratec HeartMate 3
 Pulsatile extracorporealBerlin Heart EXCOR
Thoratec PVAD
 Total artificial heartSynCardia Cardiowest
Short-term devicesAbiomed AB5000
Medos DeltaStreamb
Levitronix CentriMagb
Maquet CARDIOHELPb
MCS type
Durable devices
 Continuous flowBerlin Heart INCOR
CircuLite SYNERGYa
HeartAssist 5
HeartWare HVAD
Jarvik 2000
MicroMed DeBakey
Thoratec HeartMate II
Thoratec HeartMate 3
 Pulsatile extracorporealBerlin Heart EXCOR
Thoratec PVAD
 Total artificial heartSynCardia Cardiowest
Short-term devicesAbiomed AB5000
Medos DeltaStreamb
Levitronix CentriMagb
Maquet CARDIOHELPb
a

Withdrawn from the market in 2014.

b

These short-term devices can be used with an oxygenator for extracardiac life support/extracellular membrane oxygenation. A provision has been made for devices that were implanted concomitantly (e.g. a temporary right ventricular assist device) with a durable device.

CE: European conformity; EUROMACS: European Registry for Patients with Mechanical Circulatory Support; MCS: mechanical circulatory support.

Table 1:

Present CE-marked mechanical circulatory support systems registered in the EUROMACS database

MCS type
Durable devices
 Continuous flowBerlin Heart INCOR
CircuLite SYNERGYa
HeartAssist 5
HeartWare HVAD
Jarvik 2000
MicroMed DeBakey
Thoratec HeartMate II
Thoratec HeartMate 3
 Pulsatile extracorporealBerlin Heart EXCOR
Thoratec PVAD
 Total artificial heartSynCardia Cardiowest
Short-term devicesAbiomed AB5000
Medos DeltaStreamb
Levitronix CentriMagb
Maquet CARDIOHELPb
MCS type
Durable devices
 Continuous flowBerlin Heart INCOR
CircuLite SYNERGYa
HeartAssist 5
HeartWare HVAD
Jarvik 2000
MicroMed DeBakey
Thoratec HeartMate II
Thoratec HeartMate 3
 Pulsatile extracorporealBerlin Heart EXCOR
Thoratec PVAD
 Total artificial heartSynCardia Cardiowest
Short-term devicesAbiomed AB5000
Medos DeltaStreamb
Levitronix CentriMagb
Maquet CARDIOHELPb
a

Withdrawn from the market in 2014.

b

These short-term devices can be used with an oxygenator for extracardiac life support/extracellular membrane oxygenation. A provision has been made for devices that were implanted concomitantly (e.g. a temporary right ventricular assist device) with a durable device.

CE: European conformity; EUROMACS: European Registry for Patients with Mechanical Circulatory Support; MCS: mechanical circulatory support.

So far, 2 EUROMACS reports have been published [2, 3] analysing the adult population. This EUROMACS report is the first to focus on patients ≤19 years of age. Its goal is to report outcomes of children supported with MCS from a European perspective.

METHODS

As per 31 December 2017, 25 centres from 14 different countries (Table 2) submitted to EUROMACS data on patients ≤19 years of age. The participating centres are advised to enter data (of the patients or of the parents who have given consent in writing) of the patients who received an MCS device since 1 January 2011. Thus, newly enrolled centres will retrospectively enter data through that date. Some centres have chosen to submit data from an earlier date, and 35 patients were registered before 1 January 2011.

Table 2:

Participating paediatric units providing data for this report

CountryCity, hospital
AustriaInnsbruck, Innsbruck University Clinics
BelarusMinsk, Republican Scientific and Practical Center Cardiology
BelgiumGent, Universitair Ziekenhuis Gent
Leuven, Universitair Ziekenhuis UZ Leuven
Czech RepublicBrno, Center for Cardiovascular and Transplant Surgery
Prague, Institute for Clinical and Experimental Medicine
FranceLe Plessis Robinson, Centre Chirurgical Marie- Lannelongue
GermanyBad Oeynhausen, Herz und Diabeteszentrum Nordrhein-Westfalen
Berlin, Deutsches Herzzentrum Berlin
Freiburg, University Heart Center Freiburg Bad Krozingen
HungaryBudapest, Gottsegen Hungarian Institute of Cardiology
ItalyRome, Ospedale Pediatrico Bambino Gesù
Bergamo, Ospedale Papa Giovanni XIII
Bologna, San Orsola Hospital
Torino, Regina Margherita Children's Hospital
KazakhstanAstana, National Research Cardiac Surgery Center
NetherlandsRotterdam, Erasmus Medisch Center
Utrecht, Universitair Medisch Centrum Utrecht
PolandWarsaw, Childrens Memorial Hospital
SpainMadrid, Hospital La Paz
SwitzerlandZürich, Kinderspital Zürich
Bern, University Hospital Bern (Inselspital)
TurkeyAnkara, Baskent University Hospital
Izmir, Ege University Hospital
Istanbul, Florence Nightingale University Hospital
CountryCity, hospital
AustriaInnsbruck, Innsbruck University Clinics
BelarusMinsk, Republican Scientific and Practical Center Cardiology
BelgiumGent, Universitair Ziekenhuis Gent
Leuven, Universitair Ziekenhuis UZ Leuven
Czech RepublicBrno, Center for Cardiovascular and Transplant Surgery
Prague, Institute for Clinical and Experimental Medicine
FranceLe Plessis Robinson, Centre Chirurgical Marie- Lannelongue
GermanyBad Oeynhausen, Herz und Diabeteszentrum Nordrhein-Westfalen
Berlin, Deutsches Herzzentrum Berlin
Freiburg, University Heart Center Freiburg Bad Krozingen
HungaryBudapest, Gottsegen Hungarian Institute of Cardiology
ItalyRome, Ospedale Pediatrico Bambino Gesù
Bergamo, Ospedale Papa Giovanni XIII
Bologna, San Orsola Hospital
Torino, Regina Margherita Children's Hospital
KazakhstanAstana, National Research Cardiac Surgery Center
NetherlandsRotterdam, Erasmus Medisch Center
Utrecht, Universitair Medisch Centrum Utrecht
PolandWarsaw, Childrens Memorial Hospital
SpainMadrid, Hospital La Paz
SwitzerlandZürich, Kinderspital Zürich
Bern, University Hospital Bern (Inselspital)
TurkeyAnkara, Baskent University Hospital
Izmir, Ege University Hospital
Istanbul, Florence Nightingale University Hospital
Table 2:

Participating paediatric units providing data for this report

CountryCity, hospital
AustriaInnsbruck, Innsbruck University Clinics
BelarusMinsk, Republican Scientific and Practical Center Cardiology
BelgiumGent, Universitair Ziekenhuis Gent
Leuven, Universitair Ziekenhuis UZ Leuven
Czech RepublicBrno, Center for Cardiovascular and Transplant Surgery
Prague, Institute for Clinical and Experimental Medicine
FranceLe Plessis Robinson, Centre Chirurgical Marie- Lannelongue
GermanyBad Oeynhausen, Herz und Diabeteszentrum Nordrhein-Westfalen
Berlin, Deutsches Herzzentrum Berlin
Freiburg, University Heart Center Freiburg Bad Krozingen
HungaryBudapest, Gottsegen Hungarian Institute of Cardiology
ItalyRome, Ospedale Pediatrico Bambino Gesù
Bergamo, Ospedale Papa Giovanni XIII
Bologna, San Orsola Hospital
Torino, Regina Margherita Children's Hospital
KazakhstanAstana, National Research Cardiac Surgery Center
NetherlandsRotterdam, Erasmus Medisch Center
Utrecht, Universitair Medisch Centrum Utrecht
PolandWarsaw, Childrens Memorial Hospital
SpainMadrid, Hospital La Paz
SwitzerlandZürich, Kinderspital Zürich
Bern, University Hospital Bern (Inselspital)
TurkeyAnkara, Baskent University Hospital
Izmir, Ege University Hospital
Istanbul, Florence Nightingale University Hospital
CountryCity, hospital
AustriaInnsbruck, Innsbruck University Clinics
BelarusMinsk, Republican Scientific and Practical Center Cardiology
BelgiumGent, Universitair Ziekenhuis Gent
Leuven, Universitair Ziekenhuis UZ Leuven
Czech RepublicBrno, Center for Cardiovascular and Transplant Surgery
Prague, Institute for Clinical and Experimental Medicine
FranceLe Plessis Robinson, Centre Chirurgical Marie- Lannelongue
GermanyBad Oeynhausen, Herz und Diabeteszentrum Nordrhein-Westfalen
Berlin, Deutsches Herzzentrum Berlin
Freiburg, University Heart Center Freiburg Bad Krozingen
HungaryBudapest, Gottsegen Hungarian Institute of Cardiology
ItalyRome, Ospedale Pediatrico Bambino Gesù
Bergamo, Ospedale Papa Giovanni XIII
Bologna, San Orsola Hospital
Torino, Regina Margherita Children's Hospital
KazakhstanAstana, National Research Cardiac Surgery Center
NetherlandsRotterdam, Erasmus Medisch Center
Utrecht, Universitair Medisch Centrum Utrecht
PolandWarsaw, Childrens Memorial Hospital
SpainMadrid, Hospital La Paz
SwitzerlandZürich, Kinderspital Zürich
Bern, University Hospital Bern (Inselspital)
TurkeyAnkara, Baskent University Hospital
Izmir, Ege University Hospital
Istanbul, Florence Nightingale University Hospital

Data quality checks and audits

To ensure the best quality of data and to exclude the under-reporting of suboptimal outcomes, the EUROMACS Registry applies several methods. Incoming data are analysed on a regular basis. Individual hospitals are approached, and guidance is offered to complete or correct their data. Entries are adapted to adhere to the standard. Twice a year, each centre receives a file in which an overview of patients whose statuses need to be updated and whose changes/answers have to be monitored is presented. Statistical consistency and plausibility checks are performed, and the records containing the inconsistent data of the participating centres are identified. Data that are not plausible require checking and confirmation by the participating centres. The average number of follow-up records per patient is calculated on a per centre basis and serves as an indicator for homogeneity and completeness of recording. In addition, random on-site audits of participating centres are carried out.

Statistical analysis

We checked for the chronological plausibility of the records and eliminated or corrected implausible records by queries to on-site data managers. Data are presented as the mean ± the standard deviation (SD) or frequency with percentage. To examine mortality after implant, Kaplan–Meier estimates of cumulative probabilities were calculated, including 95% confidence intervals as a measure of certainty, because we did not truncate the curves. Kaplan–Meier curves were censored at explantation due to transplant or recovery. A patient is considered at risk until explantation because the patient received a transplant, has been weaned from the device, has died or is alive. To determine these values, cumulative incidences were calculated using competing outcomes methods and are presented for the first 2 years after the device is implanted. To avoid any censored individuals, only patients with a follow-up period of 2 years were considered for the competing outcome analysis. The user-written programme ‘STCOMPET’ in STATA was used to calculate the cumulative incidence [4]. Statistical analyses and figures were constructed using Stata 15.0 (StataCorp, College Station, TX, USA).

RESULTS

Patient population

Between January 2000 and December 2017, 237 implants in 210 patients were registered (Fig. 1), 129 (61.4%) of which were male and 81 (38.6%) of which were female. The mean age was 9.3 years (±7.0 SD), and it ranged from 0 weeks to 19 years. Almost one-fifth of the patients were below 1 year of age, and half of the population was above 10 years of age. Baseline characteristics can be seen in Table 3. Primary diagnoses at admission included cardiomyopathy (including myocarditis) in 150 (71.4%), congenital heart disease in 39 (18.6%) and other in 21 (10%) (Table 4). VAD implantation was performed primarily in patients with INTERMACS levels 1, 2 and 3 with 44 (21.0%) patients at INTERMACS profile 1.

Paediatric patients registered in the EUROMACS Registry. LVAD: left ventricular assist device; RVAD: right ventricular assist device; TAH: total artificial heart.
Figure 1:

Paediatric patients registered in the EUROMACS Registry. LVAD: left ventricular assist device; RVAD: right ventricular assist device; TAH: total artificial heart.

Table 3:

Patient characteristics preimplant

CharacteristicsTotal (n = 210)
Age (years), mean ± SD (median, range)9.3 ± 7.0 (10.5, 0–19)
Preoperative creatinine level (mg/dl), mean ± SD (median, range)0.83 ± 0.51 (0.70, 0.19 – 3.74)
Preoperative total bilirubin level (mg/dl), mean ± SD (median, range)0.1 ± 0.1 (0.06, 0.001–0.9)
Body mass index (kg/m2), mean ± SD (median, range)17.87 ± 5.08 (16.4, 9.78–37.65)
Age categories, n (%)
 <1 year38 (18.1)
 1–5 years45 (21.4)
 6–10 years22 (10.5)
 >10 years105 (50.0)
 Total210
Gender, n (%)
 Male129 (61.4)
 Female81 (38.6)
CharacteristicsTotal (n = 210)
Age (years), mean ± SD (median, range)9.3 ± 7.0 (10.5, 0–19)
Preoperative creatinine level (mg/dl), mean ± SD (median, range)0.83 ± 0.51 (0.70, 0.19 – 3.74)
Preoperative total bilirubin level (mg/dl), mean ± SD (median, range)0.1 ± 0.1 (0.06, 0.001–0.9)
Body mass index (kg/m2), mean ± SD (median, range)17.87 ± 5.08 (16.4, 9.78–37.65)
Age categories, n (%)
 <1 year38 (18.1)
 1–5 years45 (21.4)
 6–10 years22 (10.5)
 >10 years105 (50.0)
 Total210
Gender, n (%)
 Male129 (61.4)
 Female81 (38.6)

SD: standard deviation.

Table 3:

Patient characteristics preimplant

CharacteristicsTotal (n = 210)
Age (years), mean ± SD (median, range)9.3 ± 7.0 (10.5, 0–19)
Preoperative creatinine level (mg/dl), mean ± SD (median, range)0.83 ± 0.51 (0.70, 0.19 – 3.74)
Preoperative total bilirubin level (mg/dl), mean ± SD (median, range)0.1 ± 0.1 (0.06, 0.001–0.9)
Body mass index (kg/m2), mean ± SD (median, range)17.87 ± 5.08 (16.4, 9.78–37.65)
Age categories, n (%)
 <1 year38 (18.1)
 1–5 years45 (21.4)
 6–10 years22 (10.5)
 >10 years105 (50.0)
 Total210
Gender, n (%)
 Male129 (61.4)
 Female81 (38.6)
CharacteristicsTotal (n = 210)
Age (years), mean ± SD (median, range)9.3 ± 7.0 (10.5, 0–19)
Preoperative creatinine level (mg/dl), mean ± SD (median, range)0.83 ± 0.51 (0.70, 0.19 – 3.74)
Preoperative total bilirubin level (mg/dl), mean ± SD (median, range)0.1 ± 0.1 (0.06, 0.001–0.9)
Body mass index (kg/m2), mean ± SD (median, range)17.87 ± 5.08 (16.4, 9.78–37.65)
Age categories, n (%)
 <1 year38 (18.1)
 1–5 years45 (21.4)
 6–10 years22 (10.5)
 >10 years105 (50.0)
 Total210
Gender, n (%)
 Male129 (61.4)
 Female81 (38.6)

SD: standard deviation.

Table 4:

Primary diagnosis

Diagnosisn%
Cardiomyopathy11755.7
Myocarditis3315.7
Congenital heart disease3918.6
Coronary artery disease10.5
Valvular heart disease31.4
Cancer10.5
Unknown/missing167.6
210
Diagnosisn%
Cardiomyopathy11755.7
Myocarditis3315.7
Congenital heart disease3918.6
Coronary artery disease10.5
Valvular heart disease31.4
Cancer10.5
Unknown/missing167.6
210
Table 4:

Primary diagnosis

Diagnosisn%
Cardiomyopathy11755.7
Myocarditis3315.7
Congenital heart disease3918.6
Coronary artery disease10.5
Valvular heart disease31.4
Cancer10.5
Unknown/missing167.6
210
Diagnosisn%
Cardiomyopathy11755.7
Myocarditis3315.7
Congenital heart disease3918.6
Coronary artery disease10.5
Valvular heart disease31.4
Cancer10.5
Unknown/missing167.6
210

A total of 70.5% of all children were on inotropic support prior to VAD implantation. Extracardiac life support was used in 17.6% of the patients prior to VAD implantation. Twenty-two patients received a 2nd VAD implant after the 1st one, 3 patients a 3rd and 2 patients a 4th implant (Table 5). The majority of the patients (73.8%) were treated with the intention to transplant (i.e. bridge to transplant or possible bridge to transplant), and this was true for all age groups (Table 6).

Table 5:

Primary and subsequently implanted devices

Devices1st2nd3rd4thTotal
BiVAD36238
LVAD163121176
LVAD and RVAD819
RVAD162211
Total artificial heart112
Unknown11
Total2102232237
Devices1st2nd3rd4thTotal
BiVAD36238
LVAD163121176
LVAD and RVAD819
RVAD162211
Total artificial heart112
Unknown11
Total2102232237

BiVAD: biventricular assist device; LVAD: left ventricular assist device; RVAD: right ventricular assist device.

Table 5:

Primary and subsequently implanted devices

Devices1st2nd3rd4thTotal
BiVAD36238
LVAD163121176
LVAD and RVAD819
RVAD162211
Total artificial heart112
Unknown11
Total2102232237
Devices1st2nd3rd4thTotal
BiVAD36238
LVAD163121176
LVAD and RVAD819
RVAD162211
Total artificial heart112
Unknown11
Total2102232237

BiVAD: biventricular assist device; LVAD: left ventricular assist device; RVAD: right ventricular assist device.

Table 6:

Device strategy at time of implantation, stratified by age categories

<11–56–10>10Total
n (%)n (%)n (%)n (%)n (%)
Bridge to recovery4 (8.9)4 (8.0)2 (7.1)7 (6.1)17 (7.2)
Bridge to transplant18 (40.0)23 (46.0)9 (32.1)51 (44.7)101 (42.6)
Destination therapy0 (0.0)0 (0.0)0 (0.0)1 (0.9)1 (0.4)
Possible bridge to transplant14 (31.1)13 (26.0)11 (39.3)36 (31.6)74 (31.2)
Rescue therapy3 (6.7)7 (14.0)2 (7.1)7 (6.1)19 (8.0)
Unknown6 (13.3)3 (6.0)4 (14.3)12 (10.5)25 (10.5)
Total45 (100)50 (100)28 (100)114 (100.0)237 (100)
<11–56–10>10Total
n (%)n (%)n (%)n (%)n (%)
Bridge to recovery4 (8.9)4 (8.0)2 (7.1)7 (6.1)17 (7.2)
Bridge to transplant18 (40.0)23 (46.0)9 (32.1)51 (44.7)101 (42.6)
Destination therapy0 (0.0)0 (0.0)0 (0.0)1 (0.9)1 (0.4)
Possible bridge to transplant14 (31.1)13 (26.0)11 (39.3)36 (31.6)74 (31.2)
Rescue therapy3 (6.7)7 (14.0)2 (7.1)7 (6.1)19 (8.0)
Unknown6 (13.3)3 (6.0)4 (14.3)12 (10.5)25 (10.5)
Total45 (100)50 (100)28 (100)114 (100.0)237 (100)
Table 6:

Device strategy at time of implantation, stratified by age categories

<11–56–10>10Total
n (%)n (%)n (%)n (%)n (%)
Bridge to recovery4 (8.9)4 (8.0)2 (7.1)7 (6.1)17 (7.2)
Bridge to transplant18 (40.0)23 (46.0)9 (32.1)51 (44.7)101 (42.6)
Destination therapy0 (0.0)0 (0.0)0 (0.0)1 (0.9)1 (0.4)
Possible bridge to transplant14 (31.1)13 (26.0)11 (39.3)36 (31.6)74 (31.2)
Rescue therapy3 (6.7)7 (14.0)2 (7.1)7 (6.1)19 (8.0)
Unknown6 (13.3)3 (6.0)4 (14.3)12 (10.5)25 (10.5)
Total45 (100)50 (100)28 (100)114 (100.0)237 (100)
<11–56–10>10Total
n (%)n (%)n (%)n (%)n (%)
Bridge to recovery4 (8.9)4 (8.0)2 (7.1)7 (6.1)17 (7.2)
Bridge to transplant18 (40.0)23 (46.0)9 (32.1)51 (44.7)101 (42.6)
Destination therapy0 (0.0)0 (0.0)0 (0.0)1 (0.9)1 (0.4)
Possible bridge to transplant14 (31.1)13 (26.0)11 (39.3)36 (31.6)74 (31.2)
Rescue therapy3 (6.7)7 (14.0)2 (7.1)7 (6.1)19 (8.0)
Unknown6 (13.3)3 (6.0)4 (14.3)12 (10.5)25 (10.5)
Total45 (100)50 (100)28 (100)114 (100.0)237 (100)

A total of 46.8% of the patients were supported with the Berlin Heart Excor® (Berlin Heart, Berlin, Germany), 5.9% with the Heart Mate II® (Thoratec Corp., Pleasanton, CA, USA), 0.8% with HeartAssist5® (MicroMed, Houston, TX, USA) and 27.0% with HeartWare HVAD® (HeartWare Ltd., Framingham, MA, USA) (Table 7). In 67 patients, a concomitant cardiac procedure (21 congenital and valve procedures and 46 other procedures) was performed.

Table 7:

Type of ventricular assist devices per age group

<11–56–10>10Total
LVAD alone
 Pulsatile323071483
 Continuous2296881
 Unspecified111012
LVAD, temporary RVAD
 Continuous LVAD, continuous RVAD66
 Pulsatile LVAD, continuous RVAD213
BiVAD
 Pulsatile61151335
 Continuous33
RVAD333211
Total artificial heart
 Pulsatile22
Unknown11
237
<11–56–10>10Total
LVAD alone
 Pulsatile323071483
 Continuous2296881
 Unspecified111012
LVAD, temporary RVAD
 Continuous LVAD, continuous RVAD66
 Pulsatile LVAD, continuous RVAD213
BiVAD
 Pulsatile61151335
 Continuous33
RVAD333211
Total artificial heart
 Pulsatile22
Unknown11
237

BiVAD: biventricular assist device; LVAD: left ventricular assist device; RVAD: right ventricular assist device.

Table 7:

Type of ventricular assist devices per age group

<11–56–10>10Total
LVAD alone
 Pulsatile323071483
 Continuous2296881
 Unspecified111012
LVAD, temporary RVAD
 Continuous LVAD, continuous RVAD66
 Pulsatile LVAD, continuous RVAD213
BiVAD
 Pulsatile61151335
 Continuous33
RVAD333211
Total artificial heart
 Pulsatile22
Unknown11
237
<11–56–10>10Total
LVAD alone
 Pulsatile323071483
 Continuous2296881
 Unspecified111012
LVAD, temporary RVAD
 Continuous LVAD, continuous RVAD66
 Pulsatile LVAD, continuous RVAD213
BiVAD
 Pulsatile61151335
 Continuous33
RVAD333211
Total artificial heart
 Pulsatile22
Unknown11
237

BiVAD: biventricular assist device; LVAD: left ventricular assist device; RVAD: right ventricular assist device.

Outcomes

The mean support time on a device was 11.6 months (±16.5 SD). The mean stay in the intensive care unit was 37.0 days (±54.5 SD). Ninety-three (44.3%) patients were discharged either to their homes or to a rehabilitation facility. A total of 173 children (82.4%) survived to transplant, recovery, or are on ongoing treatment until the last follow-up. At 6 months, 33% of the patients and at the 1st year 38% of the children received a transplant. This percentage climbed to 51% at 2 years post VAD implantation. Thirty-seven patients (17.6%) died while on support within the observed follow-up time (Table 8).

Table 8:

Current device strategy stratified by the end point

End point
On deviceDeadReceived transplantWeanedTotal
Missing13111
Bridge to recovery2211116
Bridge to transplant221260296
Possible bridge to transplant101638670
Rescue therapy265417
Total423710724210
End point
On deviceDeadReceived transplantWeanedTotal
Missing13111
Bridge to recovery2211116
Bridge to transplant221260296
Possible bridge to transplant101638670
Rescue therapy265417
Total423710724210
Table 8:

Current device strategy stratified by the end point

End point
On deviceDeadReceived transplantWeanedTotal
Missing13111
Bridge to recovery2211116
Bridge to transplant221260296
Possible bridge to transplant101638670
Rescue therapy265417
Total423710724210
End point
On deviceDeadReceived transplantWeanedTotal
Missing13111
Bridge to recovery2211116
Bridge to transplant221260296
Possible bridge to transplant101638670
Rescue therapy265417
Total423710724210

A total of 37 patients (17.6%) died, of which 24.3% died of cerebrovascular accidents. Five patients (13.5%) died of multiorgan failure. The primary cause of death was not specified for 14 patients (Table 9).

Table 9:

Primary cause of death

Primary cause of deathn%
Bleeding25.4
Cardiopulmonary failure25.4
Cerebrovascular accident924.3
Device failure12.7
Multiorgan failure513.5
Other cause of death12.7
Right heart failure12.7
Sepsis25.4
Unknown/missing1437.8
37
Primary cause of deathn%
Bleeding25.4
Cardiopulmonary failure25.4
Cerebrovascular accident924.3
Device failure12.7
Multiorgan failure513.5
Other cause of death12.7
Right heart failure12.7
Sepsis25.4
Unknown/missing1437.8
37
Table 9:

Primary cause of death

Primary cause of deathn%
Bleeding25.4
Cardiopulmonary failure25.4
Cerebrovascular accident924.3
Device failure12.7
Multiorgan failure513.5
Other cause of death12.7
Right heart failure12.7
Sepsis25.4
Unknown/missing1437.8
37
Primary cause of deathn%
Bleeding25.4
Cardiopulmonary failure25.4
Cerebrovascular accident924.3
Device failure12.7
Multiorgan failure513.5
Other cause of death12.7
Right heart failure12.7
Sepsis25.4
Unknown/missing1437.8
37

Survival

Event-free survival of all paediatric patients on MCS was 81% at 6 months, 78% at 12 months and 66% at 2 years with censoring at time of explantation for transplant or recovery (Fig. 2). When stratified by device type, i.e. left ventricular assist device (LVAD) or a biventricular assist device, 81% survival was observed in the 1st year for LVADs and 63% for biventricular assist devices (P = 0.06) (Fig. 3).

Survival of paediatric patients after primary left or biventricular assist device implantation.
Figure 2:

Survival of paediatric patients after primary left or biventricular assist device implantation.

Survival of paediatric patients after primary implantation of a left ventricular assist device or a biventricular assist device, stratified by age.
Figure 3:

Survival of paediatric patients after primary implantation of a left ventricular assist device or a biventricular assist device, stratified by age.

When stratified by age, the oldest age group (11–19 years) had an 86% survival rate at the end of the 1st year and 76% at the end of the 2nd year; the age group 6–10 years had an 86% 1-year and 72% 2-year survival rate and the age group 1–5 years had a 69% survival rate at the end of the 1st year and 55% at the end of the 2nd year. Patients ≤1 year old showed the poorest outcome: 54% had a 1-year and 43%, a 2-year survival rate (Fig. 4). However, the latter survival rates showed poor statistical significance (P = 0.22). Figure 5 shows the survival rate stratified by device strategy.

Survival of paediatric patients after primary implantation of a left ventricular assist device or a biventricular assist device, stratified by the device implant strategy used. HTX: heart transplant.
Figure 4:

Survival of paediatric patients after primary implantation of a left ventricular assist device or a biventricular assist device, stratified by the device implant strategy used. HTX: heart transplant.

Survival of paediatric patients after primary implantation of a left ventricular assist device or a biventricular assist device, stratified by Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) level.
Figure 5:

Survival of paediatric patients after primary implantation of a left ventricular assist device or a biventricular assist device, stratified by Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) level.

Competing outcomes

Within 2 years after an implant, 51% of the patients received a heart transplant and 17% died. Only 9% could be weaned from the device and 22% had ongoing device support (Fig. 6).

Competing outcomes.
Figure 6:

Competing outcomes.

Adverse events

Overall, 151 major adverse events were reported during VAD support. Major adverse events are defined using the INTERMACS definitions [5]. These included infection, device malfunction, bleeding and neurological events (Table 10). Within the first 3 months after VAD implantation, 88 events occurred whereas 63 occurred after 3 months.

Table 10:

Major adverse event rates

Within 3 months after implant
More than 3 months after implant
Event CountsEvents per 100 patient-monthsEvents per patient-yearEvent CountsEvents per 100 patient-monthsEvents per patient-yearTotal
Device malfunction5511.971.44392.270.2794
Major bleeding132.830.3420.120.0115
Major infection132.830.34191.110.1332
Neurological event71.520.1830.170.0210
Total8863151
Within 3 months after implant
More than 3 months after implant
Event CountsEvents per 100 patient-monthsEvents per patient-yearEvent CountsEvents per 100 patient-monthsEvents per patient-yearTotal
Device malfunction5511.971.44392.270.2794
Major bleeding132.830.3420.120.0115
Major infection132.830.34191.110.1332
Neurological event71.520.1830.170.0210
Total8863151
Table 10:

Major adverse event rates

Within 3 months after implant
More than 3 months after implant
Event CountsEvents per 100 patient-monthsEvents per patient-yearEvent CountsEvents per 100 patient-monthsEvents per patient-yearTotal
Device malfunction5511.971.44392.270.2794
Major bleeding132.830.3420.120.0115
Major infection132.830.34191.110.1332
Neurological event71.520.1830.170.0210
Total8863151
Within 3 months after implant
More than 3 months after implant
Event CountsEvents per 100 patient-monthsEvents per patient-yearEvent CountsEvents per 100 patient-monthsEvents per patient-yearTotal
Device malfunction5511.971.44392.270.2794
Major bleeding132.830.3420.120.0115
Major infection132.830.34191.110.1332
Neurological event71.520.1830.170.0210
Total8863151

The most frequent major adverse event was device malfunction, which included as per definition pump exchanges from extracorporeal devices due to pump thrombosis. Device malfunction occurred 55 times in the first 3 months. In the same period, the device malfunction rate was 11.97 per 100 patient-months and 2.27 per 100 patient-months after 3 months.

Infections were the 2nd most frequent adverse event (n = 32; event rate 21.2%). Infections were divided into VAD-related and non-VAD related.

Major infection in paediatric patients occurred less frequently after the first 3 months post implantation (n = 19), i.e. 1.11 per 100 patient-months. During the first 3 months, 13 cases, or 2.83 per patient-months were reported.

Major bleeding, defined as an episode of suspected internal or external bleeding that resulted in death, reoperation, hospitalization or major transfusion, but not including cerebral haemorrhage, occurred in 15 patients (event rate, 6.9%) with 2.83 events per 100 patient-months in the first 3 months and 0.12 events per 100 patient-months after 3 months. Two patients died (0.95%) of a bleeding event.

Neurological events were defined as the occurrence of an ischaemic or a haemorrhagic stroke. Ten patients had a neurological event (event rate: 6.6%). Whereas 1.52 events per 100 patient-months occurred within the first 3 months after implantation, more than 3 months after implantation this was 0.17 events per 100 patients-months. Nine patients (24.3%) died of neurological events, making this the primary cause of death within the whole cohort (see also Table 9). Six patients had heart transplants or were successfully weaned from the device after a neurological event.

DISCUSSION

This report is the first of children supported with durable MCS that has emerged from the EUROMACS database. The EUROMACS registry is the largest database monitoring children supported with VADs in Europe, and enrolment of centres and patients continues. The authors believe it is crucial to add information about the European cohort to the other paediatric MCS database of similar size, Pedimacs, which is restricted to North American data.

One of the most striking differences between the EUROMACS and the Pedimacs cohorts is the waiting time for a heart transplant. Whereas permanent support has long become a reality for adults, bridge to transplantation or transplantability still remains the highest percentage in intention to treat within the paediatric population. Whereas almost 50% [6] of the paediatric patients in North America had a transplant within the first 6 months after a VAD implant, in Europe, only 33% at 6 months had a transplant and 38% patients at 12 months. These numbers reflect the lack of suitable donor organs in Europe, which leads to significantly longer support times. Especially in small countries or in patients under 5 years of age [7, 8], times on the heart transplant waiting list have increased. In the registry of the Eurotransplant International Foundation, the percentage of paediatric patients who receive transplants is 48% at 6 months and 57% at 12 months (personal communication, J. Smits, Eurotransplant). In Switzerland, the number of paediatric heart transplant candidates between 2009 and 2013 increased by a factor of 4 compared to the previous period [9]. In Italy, the mean time on the waiting list is more than 11 months, and in Poland (all patients), the mean waiting time is 12 months. Especially for small countries, international organ exchange among organ procurement organizations is essential. It has a direct positive impact on the possibility of patients receiving a timely, often life-saving, transplant [8]. The longer support times in Europe enable us to provide outcome data beyond 12 months of support.

One important finding of this report is that the cumulative competing incidence of death is 15% by the end of year 1 and 17% by the end of year 2. This result indicates a low mortality rate in the 2nd year of support and makes permanent support in children more feasible.

A total of 44.3% of the patients were discharged on the device. The methods used for quality checks do not indicate that serious infections are under-reported; in fact, the opposite is true. The percentage is high (20.5%), though the specificity with respect to the severity and location is low, which leads to the suspicion that different definitions may have been used.

The implantation strategy of bridge to recovery is low at 7.3%, which is relatively similar to the percentage published in the Pedimacs report (6.3%). In the group categorized as bridge to recovery (n = 16), 11 patients underwent successful explantation (69%). The others are either still on support (n = 2), have died (n = 2) or received a transplant (n = 1). For the whole cohort, 24 patients out of 210 had the device explanted due to weaning (see Table 8). The percentage of devices implanted with the intention to treat for bridge to recovery almost equals the number of devices explanted due to recovery. One reason why this number is so low might be the current lack of standardized guidelines for echocardiographic and haemodynamic criteria for LVAD removal in children [10], although children may have a greater potential for recovery [11] compared to adults.

Adverse events

Neurological events were the leading cause of death in our cohort as well in the North American cohort (24% vs 30%). Blume et al. [6] reported a higher stroke rate of 13 early events per 100 patient months and 2 late events per 100 patient months with paracorporeal devices compared to continuous flow devices (3 and 1 events per 100 patient months, respectively). Almond et al. [12] showed comparably high stroke rates for children on EXCOR VADs during the investigational device exemption trial of 15 events per 100 patient years with 29% of children affected. Although the stroke rate was not investigated within this 1st EUROMACS Paediatric Report, a recent study from the paediatric EUROMACS cohort reported low early and late stroke rates with intracorporeal continuous flow devices (0.03 and 0.4 events per 100 patient months, respectively), independent of body surface area [13]. The stroke rates in children on EXCOR and on continuous flow VADs reported in the EUROMACS registry are remarkably low. However, a recent survey addressing the antithrombotic protocols for children on EXCOR VADs in European centres revealed many modifications of the recommended Edmonton protocol with a trend towards more aggressive antithrombotic therapy [14]. Whether these modifications have contributed to lower stroke rates compared to the investigational device exemption trial is under investigation.

Another frequent adverse event was infection in 20.5%, which is clearly high (Fig. 7). One explanation could be that the definition of infection in the EUROMACS registry includes VAD-specific, VAD-related and non-VAD-related infections. The authors found that the major infection rate 3 months post implant was 1.11 per 100 patient-months compared to 2.83 per 100 patient-months within the first 3 months of implant. The higher early infection rate might be attributed to multiple factors like pre-implant condition of the patient, implant surgery and intensive care treatment. This result could be another effect of the lower transplantation rate and the longer support times in Europe.

Freedom from infection.
Figure 7:

Freedom from infection.

Limitations

The present study does not include all European centres that are implanting MCS devices. Besides the contributing centres, 14 additional hospitals were invited to join EUROMACS and submit data. Considering their positive feedback, it is expected that, in a 2nd EUROMACS Paediatric report, the data from most of these ‘additional’ hospitals will be in the registry. Data collection by means of a registry has per se an important limitation: as in every database, despite all efforts to guarantee data quality and the implementation of audits, under-reporting of adverse events cannot be ruled out.

CONCLUSION

Because EUROMACS is supported by the EACTS, the registry can reach out to an increasing number of participating hospitals to collect baseline and follow-up data on MCS from both adults and children, thus representing European data at the best achievable level. The ability to specify the different factors contributing to the outcomes of MCS in patients enables paediatric medical professionals to benchmark their data against the results of this study. Many questions remain to be addressed, i.e. discharge, additional specifics in anticoagulation management, focus on congenital heart disease and much more, which were beyond the scope of this 1st paediatric EUROMACS report. Further, a comparison with the 2nd Pedimacs report shows that outcome data differ between the registries. Investigating the reasons for these differences may contribute to insights with respect to treatment modalities and thus provide leads to possible improvements both in Europe and elsewhere.

ACKNOWLEDGEMENTS

The authors would like to thank Jacqueline Smits of the Eurotranplant International Foundation for her advice.

Funding

The European Association for Cardio-Thoracic Surgery, EACTS, supported this work. It was also supported in part by the Stiftung Charité [BIH_PRO_430 to O.M.].

Conflictof interest: none declared.

REFERENCES

1

Fraser
CD
,
Jaquiss
RDB
,
Rosenthal
DN
,
Humpl
T
,
Canter
CE
,
Blackstone
EH
et al.
Prospective trial of a pediatric ventricular assist device
.
N Engl J Med
2012
;
367
:
532
41
.

2

de By
TMMH
,
Mohacsi
P
,
Gummert
J
,
Bushnaq
H
,
Krabatsch
T
,
Gustafsson
F
et al.
The European Registry for Patients with Mechanical Circulatory Support (EUROMACS): first annual report
.
Eur J Cardiothorac Surg
2015
;
47
:
770
6;
discussion 776–7.

3

de By
TMMH
,
Mohacsi
P
,
Gahl
B
,
Zittermann
A
,
Krabatsch
T
,
Gustafsson
F
et al.
The European Registry for Patients with Mechanical Circulatory Support (EUROMACS) of the European Association for Cardio-Thoracic Surgery (EACTS): second report
.
Eur J Cardiothorac Surg
2018
;
53
:309--16.

4

Coviello
E.
STCOMPET: Stata Module to Generate Cumulative Incidence in Presence of Competing Events. Statistical Software Components S431301
.
College Station, TX, USA: StataCorp
.
Boston College Department of Economics
,
2012
.

5

Kirklin
JK
,
Holman
WL.
Mechanical circulatory support therapy as a bridge to transplant or recovery (new advances)
.
Curr Opin Cardiol
2006
;
21
:
120
6
.

6

Blume
ED
,
VanderPluym
C
,
Lorts
A
,
Baldwin
JT
,
Morales
DLS
,
Cantor
RS
et al.
Second annual Pediatric Interagency Registry for Mechanical Circulatory Support (Pedimacs) report: pre-implant characteristics and outcomes
.
J Heart Lung Transplant
2018
;
37
:
38
45
.

7

Schweiger
M
,
Stiasny
B
,
Immer
F
,
Bürki
C
,
Schmiady
M
,
Hübler
M
et al.
Cardiac transplantation in a neonate-First case in Switzerland and European overview
.
Clin Transplant
2017
;
31
; doi: 10.1111/ctr.12935.

8

Weiss
J
,
Kocher
M
,
Immer
FF.
International collaboration and organ exchange in Switzerland
.
J Thorac Dis
2015
;
7
:
543
8
.

9

Schweiger
M
,
Stiasny
B
,
Dave
H
,
Cavigelli-Brunner
A
,
Balmer
C
,
Kretschmar
O.
Pediatric heart transplantation
.
J Thorac Dis
2015
;
7
:
552
9
.

10

Cavigelli-Brunner
A
,
Schweiger
M
,
Knirsch
W
,
Stiasny
B
,
Kretschmar
O
,
Hübler
M
et al.
VAD as bridge to recovery in anthracycline-induced cardiomyopathy and HHV6 myocarditis
.
Pediatrics
2014
;
134
:
e894
9
.

11

Jefferies
JL
,
Morales
DL.
Mechanical circulatory support in children: bridge to transplant versus recovery
.
Curr Heart Fail Rep
2012
;
9
:
236
43
.

12

Almond
CS
,
Morales
DL
,
Blackstone
EH
,
Turrentine
MW
,
Imamura
M
,
Massicotte
MP
et al.
Berlin Heart EXCOR pediatric ventricular assist device for bridge to heart transplantation in US children
.
Circulation
2013
;
127
:
1702
11
.

13

Schweiger
M
,
Miera
O
,
de By
TMMH
,
Hübler
M
,
Berger
F
,
Özbaran
M
et al.
Cerebral strokes in children on intracorporeal ventricular assist devices: analysis of the EUROMACS Registry
.
Eur J Cardiothorac Surg
2018
;
53
:
416
21
.

14

Miera
O
,
Schmitt
KL
,
Akintuerk
H
,
Boet
A
,
Cesnjevar
R
,
Chila
T
et al.
Antithrombotic therapy in pediatric ventricular assist devices: multicenter survey of the European EXCOR Pediatric Investigator Group
.
Int J Artif Organs
2018
;
41
:
385
92
.

Author notes

Theo M.M.H. de By and Martin Schweiger contributed equally to this study.

This article is published and distributed under the terms of the Oxford University Press, Standard Journals Publication Model (https://dbpia.nl.go.kr/journals/pages/open_access/funder_policies/chorus/standard_publication_model)